Percutaneous spinal instrumentation was first introduced by Magerl1 in 1977 for the treatment of thoracic and lumbar spinal fractures. Since then, numerous clinical studies have compared this technique to the conventional open approach for various lumbar pathologies and demonstrate numerous benefits including decreased postoperative pain, lower intraoperative blood loss, shorter operative time, and reduced length of hospital stay.2–5 With the emergence of computed tomography (CT)-based navigated systems, minimally invasive surgery demonstrated even greater precision, safety, and accuracy for pedicle screw instrumentation.6,7 Over the last two decades, robot-assisted spine surgery has taken center-stage because of the potential for greater reliability and safety in pedicle instrumentation, especially for those with complex and variable anatomy, as well as decreased radiation exposure for the patient and operative team.8,9
A plethora of literature exists comparing the outcomes between robot-assisted minimally invasive techniques and conventional open approaches; however, the comparison between robot-assisted percutaneous and robot-assisted open surgeries remains largely unknown.10 Determining the clinical differences between these cohorts can better inform surgeons and patients during their preoperative planning phase of care. Current literature on this topic only comments on this comparison in their subanalysis, and are limited by single-surgeon or single-center series, small sample sizes, inpatient complications, and early generation robot systems.11–13
This is the first multicenter study of four geographically diverse institutions on robot-assisted spine surgery to directly compare the outcomes and complications between percutaneously placed pedicle screws and open approach surgeries. With 310 patients and over 1000 robotically placed screws, we compare the robot time per screw, radiation exposure, robot abandonment rate, screw accuracy, and 90-day postoperative complications between these two cohorts for short lumbar fusion (1- and 2-levels). We hypothesize that there will be statistically significant advantages to minimally invasive robot-assisted spine surgery.
MATERIALS AND METHODS
Patient Selection
We included adult patients (≥18 yrs old) who underwent robot-assisted short lumbar fusion (1- and 2-level) surgery from 2015 to 2019 at four independent institutions and among seven spine attending surgeons. The minimum follow-up was 90 days after the index admission. Three robot systems were used during this study's timeframe: Mazor Renaissance, Mazor X, and Mazor Stealth Edition. Cases with missing data were excluded. This study was approved by the institutional review board.
Data Collection
Patient factors included sex, body mass index, Charlson comorbidity index, and indication for surgery. Operative factors included prior spine surgery, number of instrumented levels per patient, pelvic fixation, interbody fusion, robot system, and number of planned robot screws per patient.
The primary outcomes of interest included robot time/ screw (minutes/screw), radiation exposure, robot abandonment, screw exchange for breach, and 90-day postdischarge outcomes. Total operative time, robot time, and fluoroscopy time were recorded by dedicated research coordinators at each hospital site. Robot time was measured as the time from when the robot was brought into the operating field to when the robot was removed. Robot time per screw was calculated by dividing the robot time by the number of screws placed by the robot. Radiation exposure was measured as the fluoroscopy time spent per screw (seconds/ screw). Robot abandonment was defined as discontinuing robot use for a preoperatively planned robot screw, and instead using freehand technique and/or fluoroscopy. The 90-day postdischarge outcomes included any readmission and reoperations.
Other non-robot-related outcomes and complications were recorded as well. These included length of hospital stay, dural tear, loss of intraoperative motor/sensory signals, perioperative blood transfusion, estimated blood loss (mL), and return to the operating room during the same inpatient stay.
Propensity Score Matching
A propensity score matching algorithm was employed to control for the potential selection bias between percutaneous and open surgery. The designation of open versus percutaneous was made based on whether screws (e.g., pedicle or iliac/S2AI screws) were placed through percutaneous incisions or conventional open, larger incisions. Both patient and operative factors were included in this algorithm (Table 1). Bivariate analyses were performed for perioperative factors and outcomes/complications between percutaneous and open surgery. The Chi-square or Fisher exact test (where appropriate) and t tests/ANOVA were used for categorical and continuous variables, respectively. Statistical significance was defined as a P value < 0.05. Propensity score matching was performed in R version 3.0.2 (Vienna, Austria, http://www.R-project.org/) using the Matchlt (Ho, Imai, King, and Stuart, 2011) package. SAS Studio Version 3.4 (SAS Institute Inc, Cary, NC).
TABLE 1 -
Patient Demographics, Comorbidities, and Perioperative Factors After Propensity Score Matching
|
All |
Percutaneous |
Open |
P Value |
| N |
% |
N |
% |
N |
% |
| Total # of patients |
310 |
100% |
155 |
48.4% |
155 |
50% |
|
| Female |
163 |
53% |
82 |
52.9% |
81 |
52.3% |
0.909 |
| Obese (BMI > 30kg/m2) |
122 |
39% |
63 |
40.6% |
59 |
38.1% |
0.642 |
| CCI, mean (standard deviation, SD) |
1.6 (1.5) |
1.5 (1.4) |
1.7 (1.7) |
0.241 |
| Preoperative diagnosis |
| High grade spondylolisthesis |
150 |
48% |
83 |
53.5% |
67 |
43.2% |
0.398 |
| Degenerative disc disease |
68 |
22% |
33 |
21.3% |
35 |
22.6% |
|
| Spinal stenosis |
76 |
25% |
32 |
20.6% |
44 |
28.4% |
|
| Pseudarthrosis, implant failure |
10 |
3% |
4 |
2.6% |
6 |
3.9% |
|
| Other |
6 |
2% |
3 |
1.9% |
3 |
1.9% |
|
| Operative |
| Prior spine surgery |
9 |
3% |
4 |
2.6% |
5 |
3.2% |
0.735 |
| Instrumented levels per patient, mean (SD) |
1.5 (0.5) |
1.5 (0.5) |
1.4 (0.5) |
0.141 |
| Pelvic fixation |
5 |
2% |
2 |
1.3% |
3 |
1.9% |
0.652 |
| Interbody fusion |
124 |
40% |
54 |
34.8% |
70 |
45.2% |
0.064 |
| Planned robot screws per patient, mean (SD) |
3.6 (1.2) |
3.7 (1.3) |
3.4 (1.2) |
0.155 |
| Robot system |
| Renaissance |
41 |
13% |
22 |
14.2% |
19 |
12.3% |
0.297 |
| X |
231 |
75% |
110 |
71.0% |
121 |
78.1% |
|
| Stealth |
38 |
12% |
23 |
14.8% |
15 |
9.7% |
|
Robotic Systems
Three Mazor robot systems were used and controlled for in the propensity score matching algorithm of this study. First, the Renaissance was used which replaced Mazor's 1st generation robot system (SpineAssist). The Renaissance relies on several mounting platforms and a portable computer-controlled, hexapod device to spatially position and orient surgical tools according to the surgeon's preoperative plan. Second, the Mazor X system was used which had a number of considerable advantages in comparison with prior platforms. These included a more sophisticated three-dimensional software to improve preoperative planning. The Align application combined standing spine x-rays with CT-scan to model the impact of corrective changes, such as osteotomies, on a patient's global alignment (e.g., sagittal and coronal planes). The X robot introduced a robotic arm, which was designedtobeserial, rather than parallel, toincrease the range of motion of the robotic device and an integrated three-dimensional camera with spatial tracking to better self-detect its location and reduce collision error with the operative landscape. Finally, Mazor X Stealth Edition was used which is Mazor's most recent iteration of robot systems. It combines Medtronic's Stealth navigation technology with the Mazor X platform to allow surgeons to have real-time three-dimensional feedback as they drill and place their implants down preoperatively planned trajectories.
RESULTS
After propensity score matching, a total of 310 patients were included in this study. The mean (standard deviation) Charl-son comorbidity index was 1.6 (1.5), 53% (N = 163) of patients were female, and 39% (N = 122) of patients were obese. (Table 1). The most common diagnoses included high-grade spondylolisthesis (grade >2) (48%, N = 150), spinal stenosis (25%, N = 76), and degenerative disc disease (22%, N = 68). Prior spine surgery was observed in 3% (N = 9) of patients and the mean number of instrumented levels was 1.5 (0.5). Pelvic fixation rate was 2% (N = 5), interbody fusion rate was (40%, N = 124), and the mean number of planned robot screws per patient was 3.6 (1.2). Most cases were performed by the Mazor X (Renaissance: 13%, N = 41; X: 75%, N = 231, Stealth: 12%, N = 38). No statistically significant differences were observed among these patient and operative factors between percutaneous and open surgery after propensity score matching (P value > 0.05) (Table 1).
The mean operative time was significantly longer in the open group (198 min) versus the percutaneous group (167 min, P value = 0.007) (Table 2). However, the overall robot time and mean robot time per screw was similar between cohorts (P value > 0.05). The mean fluoroscopy time per screw for percutaneous surgery (14.4 s) was significantly longer than the open group (10.1 s, P value = 0.021).
TABLE 2 -
Operative Efficiency and Radiation for Open Versus Percutaneous Surgery
| Mean (SD) |
All |
Percutaneous |
Open |
P Value |
| Operative time (minutes) |
182 (92) |
167 (106) |
198 (73) |
0.007 |
| Robot time (minutes) |
24.5 (12.0) |
24.0 (11.9) |
25.6 (12.5) |
0.562 |
| Robot time per screw (minutes/screw) |
7.8 (6.3) |
7.5 (4.2) |
8.5 (9.3) |
0.553 |
| Total fluoroscopy time (seconds) |
39.1 (33.0) |
41.7 (30.8) |
33.6 (37.1) |
0.095 |
| Fluoroscopy time per screw (seconds/screw) |
13.0 (13.6) |
14.4 (14.4) |
10.1 (11.4) |
0.021 |
The exchange rate of robot screws was low and similar between cohorts (Percutaneous: 0.8%, N = 4; Open: 1.3%, N = 7, P value = 0.348) (Table 3). Similarly, the robot abandonment rate was low and similar between groups (Percutaneous: 0.6%, N = 1; Open: 2.6%, N = 4, P value = 0.176). In regards to non-robot related complications, the incidence of dural tear was similar between cohorts (Percutaneous: 1.3%, N = 2; Open: 3.2%, N = 5, P value = 0.251). No cases were recorded to have an intraoperative motor/sensory loss and no cases required a return to the operating room during the same inpatient stay. The estimated blood loss (Percutaneous: 61.3 mL; Open: 146 mL, P value < 0.001) and perioperative blood transfusion rate (Percutaneous: 0%, N = 0; Open: 3.9%, N = 6, P value = 0.013) were significantly greater for the Open group versus the Percutaneous group (Table 3). The 90-day readmission (Percutaneous: 5.8%, N = 9; Open: 5.8%, N = 9, P value = 1) and reoperation rates (Percutaneous: 2.6%, N = 4; Open: 4.5%, N = 7, P value = 0.357) were similar between groups. Similarly, the reasons for reoperation (e.g., wound complication, implant failure) were not different between cohorts (Table 4). The mean length of stay (Percutaneous: 3.0 days; Open: 3.1days, P value = 0.579) were not significantly different.
TABLE 3 -
Robot Technical Errors and Other Surgical Complications for Open Versus Percutaneous Surgery
|
All |
Percutaneous |
Open |
P Value |
| N |
% |
N |
% |
N |
% |
| Total # of executed robot screws |
1051 |
100% |
530 |
50.4% |
521 |
49.6% |
|
| Exchange of malpositioned robot screw (per executed screws) |
11 |
1.0% |
4 |
0.8% |
7 |
1.3% |
0.348 |
| Robot abandonment |
5 |
1.6% |
1 |
0.6% |
4 |
2.6% |
0.176 |
| Registration issue |
3 |
1.0% |
0 |
0.0% |
3 |
1.9% |
0.082 |
| Unreachable anatomy |
2 |
0.6% |
1 |
0.6% |
1 |
0.6% |
1 |
| Other surgical complications |
| Dural tear |
7 |
2.3% |
2 |
1.3% |
5 |
3.2% |
0.251 |
| Loss of motor/sensory function |
0 |
0.0% |
0 |
0.0% |
0 |
0.0% |
|
| Perioperative blood transfusion |
6 |
1.9% |
0 |
0.0% |
6 |
3.9% |
0.013 |
| Estimated blood loss (mL), mean (SD) |
104 (155) |
61.3 (98.4) |
146 (187) |
<0.001 |
| Return to operating room during same inpatient stay |
0 |
0.0% |
0 |
0.0% |
0 |
0.0% |
|
TABLE 4 -
Postdischarge Outcomes and Complications Requiring Reoperation
|
All |
Percutaneous |
Open |
P Value |
| N |
% |
N |
% |
N |
% |
| Any readmission within 90 days after surgery |
18 |
5.8% |
9 |
5.8% |
9 |
5.8% |
1 |
| Any reoperation within 90 days after surgery |
11 |
3.5% |
4 |
2.6% |
7 |
4.5% |
0.357 |
| Wound complication |
9 |
2.9% |
4 |
2.6% |
5 |
3.2% |
0.735 |
| Implant failure |
1 |
0.3% |
0 |
0.0% |
1 |
0.6% |
0.317 |
| New neurologic deficit |
0 |
0.0% |
0 |
0.0% |
0 |
0.0% |
|
| Dura fistula |
0 |
0.0% |
0 |
0.0% |
0 |
0.0% |
|
| Screw malposition |
0 |
0.0% |
0 |
0.0% |
0 |
0.0% |
|
| Other |
1 |
0.3% |
0 |
0.0% |
1 |
0.6% |
0.317 |
| Length of hospital stay, mean (SD) |
3.0 (2.3) |
3.0 (1.9) |
3.1 (2.6) |
0.579 |
DISCUSSION
The advent of new imaging and surgical technologies continues to evolve the landscape of spine surgery. In recent decades, the concept of minimally invasive spine surgery (MIS) has gained increasing popularity.10,14 Furthermore, the introduction of robot-assisted platforms nearly 20 years ago has brought another paradigm shift within spine surgery. A plethora of literature exists on the comparison between MIS and conventional open pedicle screw insertion. There is ample evidence to suggest that percutaneous pedicle screws can achieve less blood loss, decreased iatro-genic disruption to muscle and soft tissues, less postoperative pain, and shorter length of stay.15–23 However, nonrobot MIS techniques have been associated with increased radiation exposure to patients and operating staff.24–27 The use of robot-assisted techniques is thought to potentially overcome this shortcoming of fluoroscopy-guided MIS by improving the mechanical control of instruments and implant insertion in a precise, preplanned manner without the need for excessive fluoroscopy.28 Although extensive research on the utility of robotic technology in spine surgery is underway,29–31 the comparison between robot-assisted percutaneous instrumentation and robot-assisted open techniques remains limited in current literature.
A few studies have commented on the use of percutaneous pedicle screw insertion using robot platforms. In 2009, Pechlivanis et al32 described the first clinical assessment of percutaneous posterior lumbar interbody fusion with the SpineAssist. In 31 consecutive patients and 133 pedicle screws, a deviation of <2 mm from the preoperative plan was observed in more than 91% of screws. Hyun et al28 performed a prospective randomized clinical trial of 60 patients undergoing 1- or 2-level spinal fusions between robot-guided percutaneous instrumentation (Mazor Renaissance) and fluoroscopic-guided open surgery. They demonstrated a mean reduction of 62.5% in intraoperative radiation exposure for the robot-assisted percutaneous cohort as well as a significant reduction in length of stay (Percutaneous: 6.8 days; Open: 9.4 days, P value = 0.020). It is important to note that the robot's dependence on preoperative CT scan for planning purposes is a potential source for increased radiation exposure to the patient. However, it is not uncommon to routinely preform preoperative CT scans for both robot and non-robot cases. No differences in patient-reported outcomes (i.e., Visual Analog Scale and Oswestry Disability Index), operative time, and screw accuracy were observed; however, the mean facet to screw distance was significantly smaller in the open, freehand cohort, and no proximal facet violations occurred in the robot group.
Similarly, in a prospective cohort study, Zhang et al33 compared the screw accuracy between robot-assisted percutaneous pedicle screw placement (TiRobot, TINAVI Medical Technologies) and open fluoroscopic-guided surgery for transforaminal lumbar interbody fusion in 100 patients. They found that the robot-assisted MIS cohort was associated with a lower intraoperative blood loss, fewer proximal facet joint violations, larger facet to screw distance, and improved pedicle screw accuracy. Iatrogenic injury to the facet could potentially accelerate adjacent level facet degeneration and thus adjacent segment disease. Furthermore, the MIS cohort experienced a greater total radiation exposure (Percutaneous: 65.3 μSv; Open: 30.3 μSv,P value < 0.001) and longer operative time (Percutaneous: 184.7 min; Open: 117.8, P value < 0.001). These authors attributed the increased radiation exposure to the need for additional intraoperative fluoroscopy for robot registration and intraoperative planning. The extended operative time was noted to be largely related to the additional time needed for robot setup.
A dearth of literature exists comparing robot-assisted MIS and robot-assisted open approach spine surgery. In a single-center, retrospective study, Kantelhardt et al11 compared the perioperative outcomes and screw accuracy in robot-assisted open (SpineAssist), robot-assisted percutaneous, and conventional open approach spine surgery in 112 patients with a mean follow-up of 90 days. In their series, pre- and post-CT scans were routinely preformed for all cases. These authors observed a decrease in intraoperative fluoroscopic radiation for robot-assisted percutaneous cases compared with robot-assisted open procedures, but this was not statistically significant. Furthermore, Kantelhardt et al observed lower opioid analgesic requirements (P < 0.05) and shorter length of stay (–1.5 d, P < 0.05). No differences between robot cohorts were observed for operative time per screw, screw accuracy, short-term adverse events, and revision surgery.
In 2014, Schatlo et al12 performed a single surgeon study on 95 patients to examine the screw accuracy and inpatient complications between robot-assisted percutaneous/open and conventional open lumbar fusion. No significant differences were observed for screw accuracy or inpatient complications, except for a significantly lower blood loss for robot cases. These authors attribute this to the fact that the majority of robot cases were performed percutaneously (69%, N = 38). Unfortunately, no direct comparative analysis was performed between robot-assisted percutaneous and robot-assisted open surgeries.
Most recently, Vaccaro et al13 performed a cadaveric investigation (N = 10) on screw accuracy and fluoroscopy radiation for MIS (fluoroscopic-guided and robot-assisted [ExcelsiusGPS, Globus Medical]) and open approach (fluoroscopic-guided and robot-assisted). Of note, the operative time per screw was 3.1 minutes for conventional open, 3.3 minutes for robot-assisted open, 7.6 minutes for fluoro-scopic MIS, and 3.6 minutes for robot-assisted percutaneous groups. Furthermore, robot assisted groups, regardless of technique, reduced the number of pedicle breaches, and the fluoroscopic radiation compared with conventional MIS and conventional open techniques. Although this study illustrated the advantages of robot-assisted procedures, no direct comparison was made between robot-assisted percutaneous and robot-assisted open groups.
In contrast to prior literature, our study used a multicenter database of seven independent spine surgeons to directly compare percutaneous and open approaches with robotassisted platforms. This study is the first and largest multicenter comparison between these cohorts. Total operative time was significantly longer for the robot-assisted open group (+30 min, P = 0.007). This is likely due to the increased time needed for exposure and closure of the spinal wound for open approaches. However, the mean robot time per screw was not significantly different between cohorts. Of note, prior literature measured operative time divided by number of screws, as opposed to the robot time divided by number of screws. The latter is a more accurate representation of the time needed for robot screw placement. The mean fluoroscopy time per screw was longer for the percutaneous cohort. This is in contrast to the Kantelhardt et al study, which found no difference between groups.11 This discrepancy may be explained by the learning curve and the increasing confidence in robot-assisted technology of a single platform. In comparison, our study involved multiple centers and utilized newer robotic technology over the study's timeframe. Furthermore, the longer radiation time for percutaneous approach may be due to surgeons repeatedly checking the entry point and final screw placements. Interestingly, the screw accuracies for both cohorts were excellent and similar (Percutaneous: 99.2%, N = 526, Open: 98.7%, N = 514, P = 0.348). In addition, the robot abandonment rate for either registration issues or unreachable anatomy was not statistically different between groups. Due to the high accuracy of robot-assisted platforms, it may not be necessary to take additional fluoroscopy shots and the intraoperative radiation exposure may decrease over time as surgeons become more confident in their use of the robot. Of note, the perioperative blood transfusion rates were significantly higher for the open approach cohort, which is consistent with prior literature.12,33 Although prior literature suggests that MIS techniques are associated with decreased hospital stays, this was not observed in our study. The mean length of stay for both cohorts was about 3 days, which is substantially lower than what has been cited in literature for robot-assisted MIS (Hyun et al: 6.8 d; Kantelhardt et al: 10.1 d).11,28 Finally, the 90 day postoperative outcomes were low and similar between groups.
There are a number of limitations to this study. First, the minimum follow-up was 90 days. Longer term follow-up is needed to determine the differences in other complications, such as pseudarthrosis, between open and percutaneous approaches. Second, there was one surgeon who performed only open procedures in our database. The use of a large multicenter database with propensity score matching may diminish potential surgeon-specific biases. Third, patient-reported outcomes have previously been cited as a benefit of MIS techniques. Unfortunately, this data was not included in our database at the time of our study. Fourth, another variable that was not included was using either preoperative CT scan to intraoperative fluoroscopy versus intraoperative CT scan and plan method. It is possible that differences in radiation exposure may be found with these two robot approaches. However, intraoperative CT scan and plan method was not an option for the Renaissance robot platform.
CONCLUSION
For short lumbar fusion, percutaneous robot-assisted spine surgery can achieve a shorter operative time and lower risk for intraoperative blood loss requiring a blood transfusion. This must be weighed against the significantly increased radiation exposure observed for the percutaneous cohort. However, as surgeons become more confident with robotassisted platforms, it is possible that excessive intraoperative radiation exposure may decrease over time. Furthermore, percutaneous screw placement has comparable outcomes to robot-assisted open approach, including screw accuracy, robot reliability, 90-day reoperations, and length of hospital stay. Future multicenter studies with longer follow-up, greater fusion levels, and patient-reported outcomes are needed.Key Points
- This is the first and largest multicenter study to compare percutaneous and open approaches with robot-assisted platforms.
- For short lumbar fusion (1-, 2-levels), percutaneous robot-assisted spine surgery can achieve a shorter operative time and lower risk for intraoperative blood loss requiring blood transfusion.
- The percutaneous robot-assisted cohort experienced greater radiation exposure; however, as surgeons become more confident with robot-assisted technology, it is possible that excessive intraoperative radiation exposure may decrease overtime.
References
1. Magerl FP. Stabilization ofthe lower thoracic and lumbar spine with external skeletal fixation.
Clin Orthop Relat Res 1984; 125–141.
2. Anderson DG, Samartzis D, Shen FH, et al. Percutaneous instrumentation of the thoracic and lumbar spine.
Orthop Clin North Am 2007; 38:401–408.
3. Isaacs RE, Podichetty VK, Santiago P, et al. Minimally invasive microendoscopy-assisted transforaminal lumbar interbody fusion with instrumentation.
J Neurosurg Spine 2005; 3:98–105.
4. Choi WW, Green BA, Levi AD. Computer-assisted fluoroscopic targeting system for pedicle screw insertion.
Neurosurgery 2000; 47:872–878.
5. Foley KT, Gupta SK. Percutaneous pedicle screw fixation of the lumbar spine: preliminary clinical results.
J Neurosurg 2002; 97: (1 suppl): 7–12.
6. Kosmopoulos V, Schizas C. Pedicle screw placement accuracy: a meta-analysis.
Spine (Phila Pa 1976) 2007; 32:E111–120.
7. Rampersaud YR, Foley KT, Shen AC, et al. Radiation exposure to the spine surgeon during fluoroscopically assisted pedicle screw insertion.
Spine (Phila Pa 1976) 2000; 25:2637–2645.
8. D'Souza M, Gendreau J, Feng A, et al. Robotic-assisted spine surgery: history, efficacy, cost, and future trends.
Robot Surg 2019; 6:9–23.
9. Kochanski RB, Lombardi JM, Laratta JL, et al. Image-guided navigation and robotics in spine surgery.
Neurosurgery 2019; 84:1179–1189.
10. Staub BN, Sadrameli SS. The use of robotics in minimally invasive spine surgery.
J Spine Surg 2019; 5: (suppl 1): S31–40.
11. Kantelhardt SR, Martinez R, Baerwinkel S, et al. Perioperative course and accuracy of screw positioning in conventional, open robotic-guided and percutaneous robotic-guided, pedicle screw placement.
Eur Spine J 2011; 20:860–868.
12. Schatlo B, Molliqaj G, Cuvinciuc V, et al. Safety and accuracy of robot-assisted versus fluoroscopy-guided pedicle screw insertion for degenerative diseases of the lumbar spine: a matched cohort comparison.
J Neurosurg Spine 2014; 20:636–643.
13. Vaccaro AR, Harris JA, Hussain MM, et al. Assessment of surgical procedural time, pedicle screw accuracy, and clinician radiation exposure of a novel robotic navigation system compared with conventional open and percutaneous freehand techniques: a cadaveric investigation.
Global Spine J 2020; 10:814–825.
14. Momin AA, Steinmetz MP. Evolution of minimally invasive lumbar spine surgery.
World Neurosurg 2020; 140:622–626.
15. Park Y, Ha JW. Comparison of one-level posterior lumbar interbody fusion performed with a minimally invasive approach or a traditional open approach.
Spine (Phila Pa 1976) 2007; 32:537–543.
16. Ringel F, Stoffel M, Stüer C, et al. Minimally invasive transmuscular pedicle screw fixation of the thoracic and lumbar spine.
Neurosurgery 2006; 59: (4 suppl 2): ONS361–366.
17. Datta G, Gnanalingham KK, Peterson D, et al. Back pain and disability after lumbar laminectomy: is there a relationship to muscle retraction?
Neurosurgery 2004; 54:1413–1420.
18. Wang B, Fan Y, Dong J, et al. A retrospective study comparing percutaneous and open pedicle screw fixation for thoracolumbar fractures with spinal injuries.
Medicine (Baltimore) 2017; 96:e8104.
19. Tian F, Tu LY, Gu WF, et al. Percutaneous versus open pedicle screw instrumentation in treatment of thoracic and lumbar spine fractures: A systematic review and meta-analysis.
Medicine (Madr) 2018; 97:e12535.
20. Bronsard N, Boli T, Challali M, et al. Comparison between percutaneous and traditional fixation of lumbar spine fracture: intraoperative radiation exposure levels and outcomes.
Orthop Traumatol Surg Res 2013; 99:162–168.
21. Fan SW, Hu ZJ, Fang XQ, et al. Comparison of paraspinal muscle injury in one-level lumbar posterior inter-body fusion: modified minimally invasive and traditional open approaches.
Orthop Surg 2010; 2:194–200.
22. Kasis AG, Marshman LA, Krishna M, et al. Significantly improved outcomes with a less invasive posterior lumbar interbody fusion incorporating total facetectomy.
Spine (Phila Pa 1976) 2009; 34:572–577.
23. Ntilikina Y, Bahlau D, Garnon J, et al. Open versus percutaneous instrumentation in thoracolumbar fractures: magnetic resonance imaging comparison of paravertebral muscles after implant removal.
J Neurosurg Spine 2017; 27:235–241.
24. Ntoukas V, Müller A. Minimally invasive approach versus traditional open approach for one level posterior lumbar interbody fusion.
Minim Invasive Neurosurg 2010; 53:21–24.
25. Peng CW, Yue WM, Poh SY, et al. Clinical and radiological outcomes of minimally invasive versus open transforaminal lumbar interbody fusion.
Spine (Phila Pa 1976) 2009; 34:1385–1389.
26. Wang J, Zhou Y, Zhang ZF, et al. Minimally invasive or open transforaminal lumbar interbody fusion as revision surgery for patients previously treated by open discectomy and decompression of the lumbar spine.
Eur Spine J 2011; 20:623–628.
27. Wang MY, Cummock MD, Yu Y, et al. An analysis of the differences in the acute hospitalization charges following minimally invasive versus open posterior lumbar interbody fusion.
J Neurosurg Spine 2010; 12:694–699.
28. Hyun SJ, Kim KJ, Jahng TA, et al. Minimally invasive robotic versus open fluoroscopic-guided spinal instrumented fusions: a randomized controlled trial.
Spine (Phila Pa 1976) 2017; 42:353–358.
29. Peng YN, Tsai LC, Hsu HC, et al. Accuracy of robot-assisted versus conventional freehand pedicle screw placement in spine surgery: a systematic review and meta-analysis of randomized controlled trials.
Ann Transl Med 2020; 8:824.
30. Panchmatia JR, Vaccaro AR, Wang W, et al. Lumbar percutaneous pedicle screw breach rates: a comparison of robotic navigation platform versus conventional techniques.
Clin Spine Surg 2020; 33:E162–E167.
31. Li HM, Zhang RJ, Shen CL. Accuracy of pedicle screw placement and clinical outcomes of robot-assisted technique versus conventional freehand technique in spine surgery from nine randomized controlled trials: a meta-analysis.
Spine (Phila Pa 1976) 2020; 45:E111–E119.
32. Pechlivanis I, Kiriyanthan G, Engelhardt M, et al. Percutaneous placement of pedicle screws in the lumbar spine using a bone mounted miniature robotic system: first experiences and accuracy of screw placement.
Spine (Phila Pa 1976) 2009; 34:392–398.
33. Zhang Q, Xu YF, Tian W, et al. Comparison of superior-level facet joint violations between robot-assisted percutaneous pedicle screw placement and conventional open fluoroscopic-guided pedicle screw placement.
Orthop Surg 2019; 11:850–856.