Share this article on:

Novel Pressure-Sensing Smart Insole System Used for the Prevention of Pressure Ulceration in the Insensate Foot

Alfonso, Allyson, R., BS, BA*; Rao, Smita, PT, PhD; Everett, Breanne, MD, MBA, BSc; Chiu, Ernest, S., MD, FACS*

Plastic and Reconstructive Surgery – Global Open: December 2017 - Volume 5 - Issue 12 - p e1568
doi: 10.1097/GOX.0000000000001568
Case Report
United States

Summary: Wounds of the foot challenge reconstructive surgeons to manage multiple factors: sensibility, stability, and durability. In this article, we focus on the insensate foot, which poses challenges to wound prevention with its propensity to develop pressure ulceration. The authors present the innovative use of a pressure-sensing smart insole system (SurroSense Rx, Orpyx Medical Technologies Inc., Calgary, Canada) in the management of the insensate foot in a patient following foot reconstruction. The pressure-sensing smart insole system provided unique feedback to both patient and provider in ways that contributed to the prevention of pressure ulcer recurrence, as well as highlight the importance of prescribed footwear in both the affected and unaffected foot. Wearable real-time monitoring and feedback faces the challenge of patient adherence. Future studies are indicated to examine the specific behaviors that are associated with favorable outcomes and long-term behavior changes.

From the *Hansjörg Wyss Department of Plastic Surgery, NYU Langone Health, New York, N.Y.; †Department of Physical Therapy, Steinhardt School of Culture, Education and Human Development, New York University, New York, N.Y.; and ‡University of Calgary, Section of Plastic Surgery and Orpyx Medical Technologies.

Received for publication April 26, 2017; accepted September 22, 2017.

Presented at the NYU HealthTech 2016 Apps, Gadgets, & Gizmos, May 7, 2016 New York, N.Y.

Disclosure: Breanne Everett is the CEO and a shareholder in Orpyx Medical Technologies. The other authors have no financial interest to declare in relation to the content of this article. The Article Processing Charge was paid for by Orpyx Medical Technologies, Inc.

Ernest S. Chiu, MD, FACS, Hansjörg Wyss Department of Plastic Surgery, NYU Langone Health, 240 East 38th Street, 13th Floor, New York, NY 10016, E-mail: ernest.chiu@nyumc.org

This is an open-access article distributed under the terms of the Creative Commons Attribution-Non Commercial-No Derivatives License 4.0 (CCBY-NC-ND), where it is permissible to download and share the work provided it is properly cited. The work cannot be changed in any way or used commercially without permission from the journal.

Wounds involving the plantar foot challenge reconstructive surgeons to find functional solutions that can withstand the repetitive trauma associated with ambulation.1 Optimal plantar reconstruction provides durable coverage with a favorable aesthetic and protective sensation to prevent pressure sores.2 Coverage of lower extremity defects is often achieved with skin-grafted muscle or fasciocutaneous flaps.1 Free flaps can be innervated to provide sensory feedback, though they have not been shown to be superior in durability and gait. Regardless, recurrence of full-thickness ulcers was observed in patients with underlying neuropathy.3 , 4

Prevention of ulcer recurrence is a significant challenge. Specialized orthotics that reduce plantar pressures can help prevent recurrent breakdown5 and contribute to prevention of ulceration in neuropathic patients.6 There is still concern, however, for wound development in the insensate foot with its inability to provide dynamic feedback normally provided by sensate plantar skin.

The authors present the innovative use of a pressure-sensing smart insole system in the management of a patient with an insensate right foot after reconstruction of a chronic plantar foot ulcer. The purpose of this study was to evaluate the insole system’s ability to provide useful feedback to patients and providers to prevent wound development.

Back to Top | Article Outline

CASE REPORT

The patient is a 22-year-old female with spina bifida, history of clubfoot, and 20 operations to her right leg who presented to the wound center for a chronic neuropathic lateral right plantar foot ulcer measuring 3 cm in diameter with surrounding callus. After 3 years of unsuccessful wound management, including offloading and local wound care, the patient underwent surgery for definitive closure using a vertical profunda artery perforator flap. The postoperative course was uneventful. After 6 weeks, the patient returned for flap debulking and revision without further complications. The patient was then fitted with the pressure-sensing smart insole system to enable ongoing plantar pressure feedback.

Back to Top | Article Outline

MATERIALS AND METHODS

The patient was fitted with the pressure-sensing smart insole system (SurroSense Rx, Orpyx Medical Technologies Inc., Calgary, Canada). The device compensates for loss of plantar protective sensation by providing on-demand smartwatch-based cues for offloading. This feedback is based on the identification of critical time/pressure signatures in insole sensor data from 8 pressure sensors distributed across the heel, lateral plantar surface, and medial and lateral metatarsal heads and toes. Device alerts are generated using algorithms that integrate pressure data over time; core algorithms are built on the current understanding of thresholds for pressure ulcer formation (conservatively, over > 35 mm Hg for a minimum of 15 minutes).7 When safe pressure and time thresholds are exceeded, the patient is instructed to go through offloading steps.

For comparison, the patient was evaluated with a current in-office pressure-sensing technique containing 99 sensors with spatial resolution of 1 sensor per square centimeter and temporal resolution of 90-Hz (Pedar-X, Novel Inc, St Paul, Minn.). The TruBlu calibration device set reading pressures between 20 and 600 kPa. Data from the heel, forefoot, and medial and lateral midfoot were collected as the patient walked at a self-selected speed (0.98–1 m/s), which was censored by an infrared timing system (Brower Timing Systems, Draper, Utah).

Back to Top | Article Outline

RESULTS

The patient was monitored with the device for 89 days, throughout which no recurrent ulcerative events were seen (Fig. 1). Both the patient and practitioner had access to the data collected by the device (Fig. 2), which provided information on average daily use, alert offloading effectiveness, and percentage of alerts by foot region.

Fig. 1

Fig. 1

Fig. 2

Fig. 2

Patient adherence was an average of 4.02 h/d (SD = 6.63) over the monitoring period. Daily device use was an average of 8.52 h/d (SD = 7.42). Over periods of device use, the patient triggered more alerts on the contralateral (noninjured) foot than the surgical site (right midfoot; Fig. 3).

Fig. 3

Fig. 3

Current in-office pressure-sensing techniques showed that there was variability of peak pressure as the patient walked (Fig. 4). When looking at the distribution of pressure across the foot, the site of surgery unloaded and exhibited no evidence of focal loading, whereas the contralateral side showed focal loads under the first metatarsal head. The mean pressure was low (179 kPa).

Fig. 4

Fig. 4

Back to Top | Article Outline

DISCUSSION

Recent technological advances have made wearable pressure-sensing solutions available.8 In preliminary studies, these devices show promise in preventing foot ulcers and improving ulcer healing. This case report supports prevention of wound recurrence with the use of a pressure-sensing smart insole system in a patient with an insensate foot, though patient adherence to device use is an obstacle to care. A study of this device showed that at least 1 alert every 2 hours could optimize response to offloading, suggesting there may be a protocol to improve adherence that should be further studied.8

The pressure-sensing smart insole system provided provider- and patient-level utility with respect to the management of loss of plantar protective sensation. From a provider’s perspective, the device was able to monitor and quantify device compliance. Data patterns emerged enabling one to deduce behavior with respect to plantar pressure distribution over time and call attention to the importance of proper footwear in offloading pressures. Initially, the data indicated that the patient applied more pressure to the contralateral foot, with reference to the site of surgery. The in-office pressure-sensing system supported these findings. This overloading suggests the importance of prescribed footwear for both feet.

For the patient, the device provided feedback to guide offloading. The limitation is patient adherence in wearing the device and responding to feedback. In this study, adherence was a clear obstacle, yet on days the device was worn the patient showed fairly high average daily use (8.52 h/d) without ulcer recurrence. Considering long-term effects, real-time biofeedback has been shown to be effective in correcting trunk sway deviations during stance and gait, including carry-over effects extending beyond device use.9 Further evaluation of technology-enhanced interventions is necessary because behavior changes like weight loss have had equivocal, if not diminished, success compared with standard interventions.10 Larger and long-term studies are necessary to evaluate the reproducibility and sustainability of the results observed in this case report.

The absence of ulcerative events over the monitoring period and agreement with in-office techniques was highly encouraging. Patient adherence is a key factor to improve moving forward, but positive findings suggest that further study is merited.

Back to Top | Article Outline

Conclusion

The insensate foot poses challenges to wound prevention with its propensity to develop pressure ulceration. The pressure-sensing smart insole system provided unique feedback to both patient and provider in ways that contributed to the prevention of pressure ulcer recurrence, as well as highlight the importance of prescribed footwear in both the affected and unaffected foot. Wearable real-time monitoring and feedback faces the challenge of patient adherence. Future studies are indicated to examine the specific behaviors that are associated with favorable outcomes and long-term behavior changes.

Back to Top | Article Outline

REFERENCES

1. Soltanian H, Garcia RM, Hollenbeck ST. Current concepts in lower extremity reconstruction. Plast Reconstr Surg. 2015;136:815e–829e.
2. Liebau J, Berger A, Pallua N, et al. Reconstruction of plantar defects. Eur J Plast Surg. 1997;20:300–305.
3. Potparić Z, Rajacić N. Long-term results of weight-bearing foot reconstruction with non-innervated and reinnervated free flaps. Br J Plast Surg. 1997;50:176–181.
4. Ducic I, Hung V, Dellon AL. Innervated free flaps for foot reconstruction: a review. J Reconstr Microsurg. 2006;22:433–442.
5. Rautio J. Resurfacing and sensory recovery of the sole. Clin Plast Surg. 1991;18:615–626.
6. Owings TM, Apelqvist J, Stenström A, et al. Plantar pressures in diabetic patients with foot ulcers which have remained healed. Diabet Med. 2009;26:1141–1146.
7. Bhattacharya S, Mishra RK. Pressure ulcers: current understanding and newer modalities of treatment. Indian J Plast Surg. 2015;48:4–16.
8. Najafi B, Ron E, Enriquez A, et al. Smarter sole survival: will neuropathic patients at high risk for ulceration use a smart insole-based foot protection system? J Diabetes Sci Technol. 2017;11:702–713.
9. Allum JH, Carpenter MG, Horslen BC, et al. Improving impaired balance function: real-time versus carry-over effects of prosthetic feedback. Conf Proc IEEE Eng Med Biol Soc. 2011;1314–8.
10. Jakicic JM, Davis KK, Rogers RJ, et al. Effect of wearable technology combined with a lifestyle intervention on long-term weight loss: The IDEA Randomized Clinical Trial. JAMA. 2016;316:1161–1171.
Copyright © 2017 The Authors. Published by Wolters Kluwer Health, Inc. on behalf of the American Society of Plastic Surgeons. All rights reserved.