Institutional members access full text with Ovid®

Share this article on:

A Prospective Clinical Trial Comparing Visible Light Spectroscopy to Handheld Doppler for Postoperative Free Tissue Transfer Monitoring

Mericli, Alexander F. M.D.; Wren, James M.P.H.; Garvey, Patrick B. M.D.; Liu, Jun Ph.D.; Butler, Charles E. M.D.; Selber, Jesse C. M.D., M.P.H.

Plastic and Reconstructive Surgery: September 2017 - Volume 140 - Issue 3 - p 604–613
doi: 10.1097/PRS.0000000000003600
Reconstructive: Trunk: Original Article

Background: Early detection of compromised free flap perfusion is critical. A common modality of thrombosis detection, physical examination augmented with hand-held Doppler, provides only intermittent data and is insensitive to venous compromise. Visible light spectroscopy provides continuous, noninvasive evaluation of tissue perfusion. The authors hypothesized that visible light spectroscopy is a more sensitive and specific monitoring method for early detection of postoperative flap compromise than intermittent hand-held Doppler and clinical examination.

Methods: The authors prospectively conducted a controlled study evaluating the sensitivity, specificity, and accuracy of the T-Stat Model 303 VLS oximeter versus that of intermittent Doppler and clinical examination. The authors prospectively collected and analyzed patient data, complications, reoperations, flap failures, and tissue oxygen saturation.

Results: Sixty-eight patients with 81 flaps completed the study. The majority of flaps (86.4 percent) were either transverse rectus abdominis musculocutaneous or deep inferior epigastric artery perforator flaps. The mean tissue oxygen saturation for all flaps in the study was 56.7 ± 7.1 percent (range, 39.4 to 72.1 percent) and did not differ significantly with patient comorbidity or flap type. During the course of the study, three flaps were returned to the operating room for exploration because of a perfusion abnormality, and the resulting salvage rate was 100 percent. The sensitivity, specificity, and accuracy of visible light spectroscopy were found to be greater than both intermittent Doppler and clinical examination.

Conclusion: Visible light spectroscopy is a reliable, continuous adjunct to free tissue transfer monitoring with advantages over intermittent hand-held Doppler and clinical examination.

CLINICAL QUESTION/LEVEL OF EVIDENCE: Diagnostic, II.

Houston, Texas

From the Department of Plastic Surgery, The University of Texas M. D. Anderson Cancer Center.

Received for publication December 6, 2016; accepted March 2, 2017.

Presented at the Texas Society of Plastic Surgeons Annual Meeting, in San Antonio, Texas, September 9 through 11, 2016.

Disclosure: The authors have no financial interest to declare in relation to the content of this article. No external funding was received.

Jesse C. Selber, M.D., M.P.H., Department of Plastic Surgery, University of Texas M. D. Anderson Cancer Center, 1400 Pressler Street, Unit 1488, Houston, Texas 77030, jcselber@mdanderson.org

©2017American Society of Plastic Surgeons