Clinical Outcome of a Wide-diameter Bone-anchored Hearing Implant and a Surgical Technique With Tissue Preservation

Mowinckel, Marius S.; Møller, Martin N.; Wielandt, Kirsten N.; Foghsgaard, Søren

Author Information
doi: 10.1097/MAO.0000000000000990
  • Open

Abstract

Bone-anchored hearing implant surgery was introduced in 1977 by Tjellström et al. (1). Over 30 years later more than 150,000 implantations have been made. The principle of bone-anchored hearing system (BAHS) is that sound vibrations are led directly to the inner ear via the mastoid bone, hereby bypassing the middle ear. This is achieved by an osseointegrated implant and a skin-penetrating abutment, which is coupled with a sound processor.

Traditionally, the surgical technique for installing BAHS includes thinning of the skin, or more precisely removal of dermal and subcutaneous tissue, hair follicles and musculature, surrounding the abutment. The aim of this technique is to achieve a thin hairless skin that is well fixated to underlying periosteum to hinder skin movement around the abutment and to facilitate cleaning. The most common forms of soft tissue reduction techniques in use are the pedicle skin flap, Dermatome and linear techniques. In recent years, promising results with modified surgical techniques with tissue preservation have been presented (2–8). This minimally invasive technique has several benefits compared with more traditional surgical techniques. It is fast, can lead to less numbness and postoperative pain, and no increase in soft tissue reactions compared with techniques with skin thinning (2,9).

Recently, implants with wider diameter have been introduced. The diameter has increased from 3.75 to 4.5 mm (Fig. 1). This provides larger initial surface-to-bone contact, leading to higher stability, and thereby the possibility of loading the implant with the hearing aid at an early time point after surgery. Successful, early loading has been shown in several studies (8,10–13). The broader implant also allows for longer abutment (12 mm), which is important when new surgical techniques without subcutaneous thinning are performed (2). The longer abutments can increase the force transferred to the fixture with higher demand on osseointegration and stability.

FIG. 1
FIG. 1:
The Wide Ponto Implant (∅ 4.5 mm) has 72% more initial contact surface than the previous model (∅ 3.75 mm).

The objective of this study was to assess skin reactions and implant stability with the use of the 4.5-mm Wide Ponto Implant from Oticon Medical and a surgical technique with tissue preservation.

METHODS

The study is a consecutive, prospective case series of patients operated at our tertiary referral center (Copenhagen University Hospital, Rigshospitalet).

The patients were operated from January 18, 2013 to March 19, 2014. To be included in the study, the patients had to be at least 18 years and eligible for bone conduction implant surgery. Prespecified exclusion criteria were skin disease in the operation area and inability to participate in follow-up.

Twenty-four patients were enrolled. The study was reviewed by the regional ethical committee (H-4-2013-FSP).

Intervention

All patients were operated by either of the two senior authors on an outpatient basis using local anesthesia. A thin needle and a ruler were used before local anesthesia to measure skin thickness.

Surgical technique used was minimally invasive using a one-stage linear-incision technique without removal of subcutaneous tissue (2,14). Skin incision was approximately 4-cm long and to the periosteum, which was removed using a raspatory. A hole was drilled and widened with a 3.8-mm-diameter countersink. The implant was placed in a single stage with application of 50-Ncm torque. The incision was sutured in the epidermal layer only and a separate hole was made with a biopsy punch 0.5 to 1 cm from the incision line, allowing the abutment to penetrate the skin. Finally, the implant stability quotient (ISQ) was measured.

Implant

The implant used was a Wide Ponto Implant (diameter, 4.5 mm; length, 4 mm). The implant has a greater diameter (4.5 mm versus 3.75 mm) and a different cutting geometry and threading when compared with the previous generation model. Depending on skin thickness, abutments of lengths 6, 9, and 12 mm were used. The implant is equipped with a traditional Brånemark-type machined titanium implant surface, and the abutments are titanium abutments without coating.

Postoperative Care and Fitting

A healing cap was fixed to the abutment and gauze imbibed with antibiotic-steroid ointment (hydrocortisone 10 mg/g with oxytetracycline 30 mg/g) circulated around the abutment, under the cap. A pressure head dressing was applied overnight to prevent hematoma. The healing cap and gauze were removed after approximately 10 days. The sound processor was fitted from approximately 6 weeks after surgery.

Prospective Follow-up Examinations

Follow-up was performed after 10 days, 6 weeks, 6 months, and 12 months. Implant stability and soft tissue reactions, numbness, and pain around the abutment were registered at all the above scheduled visits. Skin reactions were classified according to Holgers grading system (15).

Implant stability was measured using resonance frequency analysis (RFA) obtained by Osstell ISQ (Osstell, Göteborg, Sweden). This is a portable, handheld instrument that includes the use of a magnetic SmartPeg (Osstell) attached to the abutment. The resonance frequency is calculated from the response signal and the result is displayed as ISQ, which is scaled from 1 to 100. The greater the number, the greater the stability.

The measurements were performed manually in two perpendicular directions, which result in two different ISQ values: ISQ high and ISQ low. The same SmartPeg was used on the individual patient throughout the study to minimize the measurement variability. The Ostell ISQ instrument has an accuracy of ±2 ISQ points.

Statistics

For comparisons over time, the Wilcoxon signed rank test was used for continuous variables. A significance level of 0.05 was adopted.

RESULTS

The baseline patient characteristics and implants used are represented in Table 1. The most common reason for implantation was conductive or mixed hearing loss and single-sided deafness. Two patients were previously irradiated, one patient had diabetes and two patients received per oral steroid treatment, because of Mb, Addison, and nasal polyposis, respectively. The most common used abutment length was 9 mm (19 patients). All of these patients had skin thickness between 5 and 6 mm, except one (skin thickness 7 mm). The abutment length was changed from 9 to 12 mm in this patient because of contact between the processor and the skin. Three patients received 12-mm abutments because of thick subcutaneous tissue, and these had skin thickness of 8, 8, and 9 mm respectively.

TABLE 1
TABLE 1:
Baseline characteristics of the study population (n = 24)

There was no major surgical complication, but in one patient a minor bleeding occurred because of drilling into a vein. Surgery time was on average 22.5 (range, 13–30) minutes.

Loading of the implant with the sound processor was done on average after 8.3 weeks (range, 4.0–14.6 wk) after operation (n = 24).

Distributions of skin reactions according to Holgers grading system are summarized in Table 2. The total number of follow-up visits was 90, hereby 88 planned visits and 2 unplanned because of infection. Eight patients missed one follow-up visit and 91.7% of the study group attended the rest of their scheduled follow-up visits. In 87.8% of the visits, there was no skin irritation (Holgers grade 0). Four patients had adverse skin reactions (Holgers ≥2) corresponding to 14% of the patients. In general, these complications were treated successfully with local antibiotics, and granulation tissue was removed with lapis 50%. One patient had three adverse skin reaction (Holgers ≥2) at three out of six follow-ups. This was treated as mentioned above, though supplemented with systemic antibiotics as well. This patient had comorbidity in the form of Waldenströms macroglobulinemia, which may affect the blood circulation because of hyperviscosity of the blood. No infections that required removal of implant or abutment were observed (Holgers grade 4).

TABLE 2
TABLE 2:
Skin reactions (Holgers grading system) at follow-up

Results for numbness and pain at follow-up are presented in Table 3. Grading of pain and numbness around the implant is achieved by visual analogue scale (VAS). The scale is from 0 to 10. Zero is no pain and no numbness respectively, and 10 is worst possible, unbearable pain and numbness, respectively. This 10-point scale was converted to a 4-point scale for more convenient and reader-friendly publishing; none (VAS 0) limited (VAS 1–3), moderate (VAS 4–7), and extensive (VAS 8–10). As illustrated in the table, there were few complaints of pain and numbness at follow-up.

TABLE 3
TABLE 3:
Distribution of pain and numbness at follow-up

The total number analyzed differs in Tables 2 and 3. Unfortunately, there were a few patients who were not registered according to pain and numbness at follow-up.

ISQ high and ISQ low for the different abutment lengths are presented in Figure 2A and B, respectively. The patient who changed abutment was excluded from both figures. There is a significant increase in mean ISQ high and mean ISQ low the first 10 days after surgery, and this continues after loading of the implants and at 6 months and 1 year after surgery (p = <0.0001). No implants were lost. ISQ was registered in total 111 of possible 120 registrations. Eight of these missed registrations were because the patient did not show up, and one was because of oversight.

FIG. 2
FIG. 2:
A, ISQ high for individual patients and mean interindividual ISQ high for patients with 6-mm abutment (bold, dotted line; n = 2), 9-mm-long abutment (bold, dashed line; n = 19) and 12-mm-long abutment (bold, large dashed line; n = 3). Bold solid line represents mean ISQ high across all abutment lengths. B, ISQ low for individual patients and mean interindividual ISQ low for patients with 6-mm abutment (bold, dotted line; n = 2), 9-mm-long abutment (bold, dashed line; n = 19), and 12-mm-long abutment (bold, large dashed line; n = 3). Bold solid line represents mean ISQ low across all abutment lengths. ISQ indicates implant stability quotient.

DISCUSSION

The main objective of this study was to monitor the clinical outcome of a wide-diameter bone-anchored hearing implant (BAHI) and tissue preservation surgery. This surgical technique was first discussed in 2009 (16,17) and several studies are showing promising results (2–8). This surgical procedure is quick, can be performed under local anesthesia, and gives good cosmetic results with no bald spot around the abutment and less scar tissue (18) (Fig. 3). In addition, several studies have shown less postoperative pain, shorter healing time, and less numbness around the abutment compared with more traditional surgical techniques with soft tissue reduction (2,7,9).

FIG. 3
FIG. 3:
Implant/abutment 6 months after surgery showing good cosmetic result.

Over the past few years several new transcutaneous BAHI have been introduced to the market. These are either passive or active components, which are completely covered by skin. Hence there is no skin-penetrating coupling. Instead of a skin-penetrating coupling, these transcutaneous devices use magnetic coupling between the implant and the sound processor. Advantages of this system are that it may reduce the risk of infection, that it has good postoperative cosmetic results, and that it reduces the necessity of cleaning around the coupling. The latter is especially an interesting alternative with patients who for medical or other reasons have difficulties with cleaning routines (19). Disadvantages with this system compared with percutaneous BAHI can be temporary pain around the implant because of a powerful magnet. In several studies however, the pain disappeared when a magnet with less strength was chosen (19,20). Furthermore, the risk of accidently loosing the external device is higher with magnetic coupling compared with a skin-penetrating coupling. In addition, some studies have shown better hearing thresholds for the percutaneous solution with direct coupling compared with the transcutaneous devices, especially in the high frequencies (20). This is because of soft tissue attenuation and energy loss in the intervening skin layer (19).

The ISQ value represents the resonance frequency of the mechanical arrangement fixated in the bone. Despite its use in research for over a decade, RFA is more a tool for clinical studies and implant research, rather than of clinical and therapeutic importance. There is, for example, no specific ISQ value that has been documented as an acceptance value for loading the implant (21). Loading of the implant is not only depending of ISQ alone, but also bone quality and postoperative wound healing. Nevertheless, ISQ value is very useful following the stability of one specific implant over time (22). On the other hand, it is difficult to compare absolute ISQ measurements between studies, because of different reporting standards. Recently, practical recommendations regarding RFA in scientific clinical studies were published (22).

Several factors influence ISQ, for example, bone quality, design of transducer (i.e., SmartPeg), and the length of implants and abutments. A well-integrated implant in a soft bone can give lower ISQ than an insufficiently integrated implant in a hard bone (23). Comparing the present wide implant stability directly with older more narrow ones is difficult since the SmartPeg for testing ISQ does not fit the older abutments (18). Nevertheless, ISQ in the earlier narrow version (3.75 mm) has shown statistically significant lower values than the present Wide Ponto Implant (13). Furthermore, the use of a wider implant has several advantages. The larger diameter of this wide implant increases initial bone-to-implant contact surface. This is increased by 72% compared with previous narrow versions (3.75 mm) and by 10% compared with other wide implants on the market (also 4.5 mm) (18,24). The larger radius of the implant and increased surface area improve stability (10,25). Second, the Wide Ponto Implant allows smaller drilling hole (3.8 mm versus 4.0 mm for the other available implants), thus reducing the surgical trauma. This is because of a new cutting geometry of the threaded area.

Basic studies in the laboratory with the same implant but different abutment lengths give an estimation of stability loss of 3 to 4 ISQ for each millimeter of longer abutment (23). This was also the case in the present study. Both mean high and low ISQ values are clearly lower in the longer abutments. The average data for the 6- and 12-mm abutment length are only based on two and three patients respectively, and should be interpreted with caution.

After loading of the implant on average 8.3 weeks after surgery we did not observe a fall in ISQ. With the initially high ISQ, and quick wound healing, earlier loading seems feasible. Successful, early loading of wide implants has been documented in several studies where loading was performed after 2, 3, and 4 weeks (8,11,12), as well as 3 weeks with the previous 3.75-mm implant (13). We registered a significant increase in ISQ throughout the 1-year follow-up period and the values remained above baseline values. This is also shown in other studies of wide implants (10,11,13,18,26).

In this study four patients (14%) had Holgers ≥2. This is consistent with previously presented material with the same surgical technique without soft tissue reduction (2,3,20). On the other hand, there are several studies showing no difference regarding frequency of infections when comparing surgical techniques with or without soft tissue reduction, respectively (2,9).

The most common complication in BAHS surgery is soft tissue reactions around the skin-penetrating coupling. The majority of skin inflammation and infection are grade 1 reactions, and can be treated locally with topical treatment. No revision surgery was performed within the 1-year follow-up, and a single abutment change was made to a longer abutment. No other soft-tissue-related complications that have been reported for tissue reduction techniques, such as skin overgrowth or skin necrosis, were encountered. In our group, the only soft-tissue-related complications were Holgers scores, which were all successfully treated. At the 12-month visit, two patients experienced a skin reaction corresponding to Holgers grades 1 and 2 respectively.

Pain and numbness were a subjective evaluation made by the patient and hence subject to a substantial degree of variability. Nevertheless, it is worth taking notice that there were only very few reports of numbness and pain related to the surgical procedure. At the end of the study (12-month follow-up), no patient reported either numbness or pain around the implant. In similar, several studies have shown less pain and numbness around the implant after tissue preservation surgery compared with other surgical techniques with soft tissue reduction (2,4,9).

Limitations of this study were the small number of patients and no control group for comparison.

CONCLUSION

The present prospective, 1-year follow-up study, concerning a wide BAHI in combination with a surgical technique with tissue preservation, reveals ISQ values with good stability as a sign for good osseointegration and no decrease in ISQ after implant loading. In addition, we observed few skin reactions, minimal pain, and numbness. We believe that linear incision with tissue preservation technique is a simple, safe, and reliable method and should be taken into consideration when assessing the patient in the need for bone conductive hearing aid.

Acknowledgments

The authors thank the nurses in the BAHS-outpatient-clinic at the Department of Otorhinolaryngology, Head and Neck Surgery, Copenhagen University Hospital Rigshospitalet, for helping in all phases of the study.

REFERENCES

1. Tjellström A, Lindström J, Hallén O, Albrektsson T, Brånemark PI. Osseointegrated titanium implants in the temporal bone. A clinical study on bone-anchored hearing aids. Am J Otol 1981; 2:304–310.
2. Hultcrantz M. Outcome of the bone-anchored hearing aid procedure without skin thinning: A prospective clinical trial. Otol Neurotol 2011; 32:1134–1139.
3. Husseman J, Szudek J, Monksfield P, Power D, O’Leary S, Briggs R. Simplified bone-anchored hearing aid insertion using a linear incision without soft tissue reduction. J Laryngol Otol 2013; 127 (Suppl):S33–S38.
4. Hultcrantz M, Lanis A. A five-year follow-up on the osseointegration of bone-anchored hearing device implantation without tissue reduction. Otol Neurotol 2014; 35:1480–1485.
5. Lanis A, Hultcrantz M. Percutaneous osseointegrated implant surgery without skin thinning in children: A retrospective case review. Otol Neurotol 2013; 34:715–722.
6. Hawley K, Haberkamp TJ. Osseointegrated hearing implant surgery: Outcomes using a minimal soft tissue removal technique. Otolaryngol Head Neck Surg 2013; 148:653–657.
7. Singam S, Williams R, Saxby C, Houlihan FP. Percutaneous bone-anchored hearing implant surgery without soft-tissue reduction: Up to 42 months of follow-up. Otol Neurotol 2014; 35:1596–1600.
8. Høgsbro M, Agger A, Johansen LV. Successful loading of a bone-anchored hearing implant at two weeks after surgery: Randomized trial of two surgical methods and detailed stability measurements. Otol Neurotol 2015; 36:e51–e57.
9. Høgsbro M, Agger A, Johansen LV. Bone-anchored hearing implant surgery: Randomized trial of dermatome versus linear incision without soft tissue reduction–clinical measures. Otol Neurotol 2015; 36:805–811.
10. Dun CAJ, de Wolf MJF, Hol MKS, et al. Stability, survival, and tolerability of a novel baha implant system: Six-month data from a multicenter clinical investigation. Otol Neurotol 2011; 32:1001–1007.
11. McLarnon CM, Johnson I, Davison T, et al. Evidence for early loading of osseointegrated implants for bone conduction at 4 weeks. Otol Neurotol 2012; 33:1578–1582.
12. Nelissen RC, den Besten CA, Faber HT, Dun CAJ, Mylanus EAM, Hol MKS. Loading of osseointegrated implants for bone conduction hearing at 3 weeks: 3-year stability, survival, and tolerability. Eur Arch Otorhinolaryngol 2015; [Epub ahead of print].
13. Nelissen RC, den Besten CA, Mylanus EAM, Hol MKS. Stability, survival, and tolerability of a 4.5-mm-wide bone-anchored hearing implant: 6-month data from a randomized controlled clinical trial. Eur Arch Otorhinolaryngol 2016; 273:105–111.
14. Wide Ponto Implant—Surgical Quick Guide. Available at: http://www.oticonmedical.com/~asset/cache.ashx?id=33037&type=14&format=web. Accessed November 2, 2015.
15. Holgers KM, Tjellström A, Bjursten LM, Erlandsson BE. Soft tissue reactions around percutaneous implants: a clinical study of soft tissue conditions around skin-penetrating titanium implants for bone-anchored hearing aids. Am J Otol 1988; 9:56–59.
16. Hultcrantz M. A new (old) technique to avoid periimplant problems? Abstract presented at the Second International Symposium on Bone Conduction, Hearing-Craniofacial Osseointegration. Gothenburg, Sweden; 2009.
17. Soo G. The Hong Kong incision (direct percutaneous BAHA surgery without soft tissue reduction or skin grafting. Early results. Abstract presented at the Second International Symposium on Bone Conduction, Hearing-Craniaofacial Osseointegration. Gothenburg, Sweden; 2009.
18. Hultcrantz M. Stability testing of a wide bone-anchored device after surgery without skin thinning. Biomed Res Int 2015; 2015:1–6.
19. Briggs R, Van Hasselt A, Luntz M, et al. Clinical performance of a new magnetic bone conduction hearing implant system: Results from a prospective, multicenter, clinical investigation. Otol Neurotol 2015; 36:834–841.
20. Iseri M, Orhan KS, Tuncer U, et al. Transcutaneous bone-anchored hearing aids versus percutaneous ones: Multicenter comparative clinical study. Otol Neurotol 2015; 36:849–853.
21. Sennerby L, Meredith N. Implant stability measurements using resonance frequency analysis: Biological and biomechanical aspects and clinical implications. Periodontol 2000 2008; 47:51–66.
22. Nelissen RC, Wigren S, Flynn MC, Meijer GJ, Mylanus EAM, Hol MKS. Application and interpretation of resonance frequency analysis in auditory osseointegrated implants. Otol Neurotol 2015; 36:1518–1524.
23. Sim CPC, Lang NP. Factors influencing resonance frequency analysis assessed by Osstell mentor during implant tissue integration: I. Instrument positioning, bone structure, implant length. Clin Oral Implants Res 2010; 21:598–604.
24. Westerkull P, Jinton L. The new Wide Ponto Implant design: Clinical and surgical aspects. Available at: http://www.oticonmedical.com/~asset/cache.ashx?id=12639&type=14&format=web. Accessed November 2, 2015.
25. Ivanoff CJ, Sennerby L, Johansson C, Rangert B, Lekholm U. Influence of implant diameters on the integration of screw implants. An experimental study in rabbits. Int J Oral Maxillofac Surg 1997; 26:141–148.
26. Foghsgaard S, Caye-Thomasen P. A new wide-diameter bone-anchored hearing implant-prospective 1-year data on complications, implant stability, and survival. Otol Neurotol 2014; 35:1238–1241.
Keywords:

Bone-anchored hearing systems; Holgers grading system; Implant stability quotient; Osseointegration; Resonance frequency analysis; Skin reactions; Tissue preservation surgery; Wide-implant