Case Report

Case report: unique failure of a Synthes TFNA fenestrated lag screw in a peritrochanteric nonunion

Smidt, Kevin P. MD; Perry, Kevin J. MD DPT; Andre, Lincoln K. MD; Chauvin, Brad J. MD; Kautz, Steve MD; Massey, Patrick A. MD; Barton, R. Shane MD MPH

Author Information
doi: 10.1097/OI9.0000000000000203
  • Open

Abstract

1 Introduction

Intertrochanteric femur fractures are a common injury seen in orthopaedic trauma. By 2050, the annual incidence of hip fractures is estimated to surpass 6.3 million worldwide.[1] These injuries are usually treated surgically, with both intramedullary or extramedullary devices. Cephalomedullary nails (CMN) have gained popularity due to their biomechanical superiority and minimally invasive implantation.[2–7] In fact, CMNs have now become the dominant mode of fixation for intertrochanteric hip fractures across the United States.[5,6,8,9]

Intramedullary implants used for treatment of intertrochanteric fractures have demonstrated a failure rate between 0% and 22%.[10–15] Implant-related complications include intraoperative peri-implant fracture, postoperative peri-implant fracture, implant cut-out, loss of fixation, anterior perforation of the distal femur, and rarely implant failure.[16–18] The reported incidence of CMN implant breakage is 0.2% to 5.7%.[19–22] The majority of these occur at the proximal nail aperture. There have been reports of rare implant failure at the distal screw aperture, the distal locking screw itself, and the shaft of the CMN. Cephaolmedullary lag screw fracture is a rare mode of failure and can be difficult to manage. We report a case of a fenestrated cephalic lag screw implant failure in a peritrochanteric nonunion using the Trochanteric Fixation Nail—Advanced (TFNA) Proximal Femoral Nailing System (DePuy Synthes, Paoli, Pennsylvania).

2 Case report

Submission to our Institutional Review Board (IRB) determined that this case report is exempt from Institutional Review Board approval. Informed consent was obtained from the patient for reproduction. This report has been approved for publication by our institutions committee on research ethics in accordance with the Declaration of the World Medical Association.

A 70-year-old female sustained a right closed displaced peritrochanteric proximal femur fracture from a high-speed motor vehicle collision (Fig. 1). The patient was previously an independent ambulator and had a past medical history of hypertension and lumbar stenosis status post decompression and fusion. She was overweight with a body mass index of 26.52 kg/ m2 and her American Society of Anesthesiologists score was III. The fracture was classified as OTA/AO 31-A2.2, Evans type III fracture. The lateral wall measured 17.6 mm according to the technique described by Hsu et al.[23]

F1
Figure 1:
Anteroposterior radiograph of the right peritrochanteric proximal femur fracture described in this case report.

The patient was treated operatively on the day following her injury with closed reduction on a fracture table. It was noted intraoperatively that the fracture extended into the region of the superior femoral neck. Her closed reduction was deemed adequate and was held temporarily with 2-mm Kirschner wires during nail insertion to prevent displacement. She underwent uneventful internal fixation using a 130° Synthes TFNA CMN which measured 340 mm in length and 10 mm in diameter with a 90 mm fenestrated cephalic lag screw. The cephalic screw was placed and interfragmentary compression was achieved across the fracture site using the compression nut as described in the manufacturer's technique guide.[24] There were no observed complications during drilling or insertion of the cephalic lag screw. The cephalic screw was set in the static position. Two distal static interlocking screws were placed. The remainder of the procedure was uneventful. The patient also presented with multiple ipsilateral foot fractures that were treated during the index procedure and she was made non-weight bearing postoperatively for her ipsilateral foot injury.

The patient was seen regularly for postoperative follow-up visits. At her 6-week follow-up, adequate radiographic healing of her ipsilateral foot fractures was visualized and her weight bearing precautions were lifted (Fig. 2). She was ambulating without pain by her 3-month follow-up visit (Fig. 3). She was asymptomatic and ambulating without assistive devices at her 6-month follow-up visit and had no complaints but her radiographic images demonstrated a delayed union or nonunion of her peritrochanteric fracture (Fig. 4). Given the patients lack of symptoms, a delayed union/nonunion workup was not initiated at that time. She was scheduled to return for a 1-year postoperative follow-up visit.

F2
Figure 2:
Six-week routine follow-up anteroposterior and lateral radiographs demonstrating right peritrochanteric proximal femur fracture with TFNA fixation.
F3
Figure 3:
Three-month routine anteroposterior and lateral radiographs of right peritrochanteric femur fracture with TFNA fixation.
F4
Figure 4:
Six-month routine anteroposterior and lateral radiographs of right peritrochanteric femur fracture with TFNA fixation.

Eleven months after her index surgery, she suffered a ground level fall in her home. She complained of pain and inability to bear weight and was taken to a nearby hospital by local Emergency Medical Services. She was then transferred to our facility for higher level of care where she was found to have sustained implant failure of the cephalic lag screw portion of the TFNA in the setting of nonunion and varus collapse (Fig. 5). Computed Tomography scan revealed nonunion of the peritrochanteric fracture with varus collapse and breakage of the cephalic lag screw through the femoral neck. The cephalic lag screw appeared to have failed through the fenestrations (Fig. 6). She underwent revision surgery with removal of the broken nail and total hip arthroplasty was performed (Figs. 7 and 8). Explanted hardware was inspected and confirmed failure of the cephalic lag screw through the fenestrations.

F5
Figure 5:
Anteroposterior and lateral radiographs demonstrating implant failure of TFNA through the cephalic lag screw at 11 months postoperatively.
F6
Figure 6:
Coronal reconstructions of sequential CT images demonstrating peritrochanteric fracture nonunion with varus collapse and failure of the cephalic lag screw of the TFNA through the fenestrations. CT, computed tomography.
F7
Figure 7:
Intraoperative and postoperative clinical images showing explanted failed implant with breakage through the cephalic lag screw fenestrations.
F8
Figure 8:
Anteroposterior pelvis and lateral hip radiographs following revision surgery and total hip arthroplasty.

3 Discussion

Cephalomedullary nails are the treatment of choice for unstable peritrochanteric proximal femur fractures and have even become the most utilized mode of internal fixation for stable peritrochanteric fractures across the United States.[9,25] There are several reports in the literature describing cephalomedullary nail breakage. The most common site of nail breakage is at the aperture for the cephalic lag screw or blade. At this location the cross-sectional area of the intramedullary nail is narrowed by 73%.[26–28] Prior reports also show evidence of eccentric drilling at the proximal nail aperture causing implant notching and decreased fatigue strength.[28–31] Other previously reported sites of implant breakage include the distal screw aperture, the distal locking screw itself, and the shaft of the cephalomedullary implant.[19,20,22,28,32] We present a unique failure of the TFNA cephalomedullary nail in a peritrochanteric nonunion.

There are several risk factors that have been associated with nail breakage including a low American Society of Anesthesiologists score (I or II), young age, reverse obliquity intertrochanteric fractures, subtrochanteric fractures, and pathological fractures.[13,16,33] Early implant failures have been attributed to poor insertion technique with notching of the proximal nail aperture with the Gamma nail (Stryker, Kalamazoo, Michigan).[34] Late implant catastrophic failures have been attributed to inadequate reduction, delayed union, or nonunion. As in our patient, any implant will eventually fail in the setting of nonunion based on its fatigue life of the implant and the stresses placed upon it. Being critical of our own work, the closed reduction was imperfect but acceptable. The decision to place a static cephalic screw was made based on the lack of medial cortical contact and concerns of overcompression and loss of femoral neck length. In retrospect, a short femoral neck that collapsed and compressed may have prevented a return to the operating room, but may have also resulted in relative abductor weakness.

The Synthes TFNA system offers both solid and fenestrated helical blade and lag screw options. The fenestrated head elements allow for cement augmentation. There is some evidence that augmented fixation may have a lower rate of failure in osteoporotic individuals, however evidence demonstrating superiority of cement augmentation in individuals with normal bone density is lacking.[35–37] Despite the rare usage of cement augmentation for peritrochanteric fracture fixation, most hospitals, including our institution, only stock the fenestrated implants.[24] There is no difference in cost between the fenestrated and nonfenestrated implants. In this case, a fenestrated lag screw was utilized without cement augmentation. The lag screw appears to have failed through the fenestrated holes. Accepting that the nail would have failed in the setting of nonunion, we question if the nail would have failed at the same location if a solid lag screw were utilized. In response to this case, our institution now stocks nonfenestrated cephalic lag screws and helical blades which are used if cement augmentation is not utilized. Perhaps fenestrated head elements should only be used if cement augmentation is planned.

The cephalic lag screw failure was reported to DePuy Synthes. Engineers at Depuy Synthes were available to discuss the case, the universal adoption of fenestrated head elements, and biomechanical data and testing which supported this decision. DePuy Synthes was willing to share their biomechanical testing methods and the results of early market surveillance after signing a nondisclosure agreement. In response to early surveillance data and finite element analysis, the fenestrated lag screw has recently been redesigned and optimized to improve fatigue strength (Fig. 9).

F9
Figure 9:
Figure provided by Johnson & Johnson DePuy Synthes (Paoli, Pennsylvania) and approved for reprinting, describing new design changes to the fenestrated screw head element in response to biomechanical testing and device surveillance.

This particular patient was salvaged with total hip arthroplasty; however, this failure mode would be difficult to treat if revision open reduction internal fixation were attempted. The broken head element can be removed with a reverse threaded extraction cone from the broken screw removal set. Careful attention would be required not to spin the femoral head during this extraction preventing damage to its blood supply. In a series of patients treated for TFNA failures at the proximal nail aperture, 38% revised to a second TFNA experienced a second implant failure.[16] It is unclear if this trend would be similar for nail failure at the cephalic lag screw.

Close follow-up is recommended for unstable peritrochanteric proximal femur fractures as displacement of a delayed union or nonunion can have catastrophic consequences. CMN case, closer monitoring after the 6 month visit and more timely workup of nonunion may have prevented catastrophic implant failure.

4 Conclusion

Cephalomedullary nails have become the most common treatment for peritrochanteric fracture fixation in the United States. We report a unique failure of the TFNA implant at the fenestrated cephalic lag screw. It is important to recognize and monitor new implants for unique failure modes.

References

1. Cooper C, Campion G, Melton LJ. Hip fractures in the elderly: a worldwide projection. Osteoporos Int 1992; 2:285–289.
2. Schipper IB, Marti RK, Van Der Werken C. Unstable trochanteric femoral fractures: extramedullary or intramedullary fixation. Review of literature. Injury 2004; 35:142–151.
3. Curtis MJ, Jinnah RH, Wilson V, Cunningham BW. Proximal femoral fractures: a biomechanical study to compare intramedullary and extramedullary fixation. Injury 1994; 25:99–104.
4. Aros B, Tosteson ANA, Gottlieb DJ, Koval KJ. Is a sliding hip screw or im nail the preferred implant for intertrochanteric fracture fixation? Clin Orthop Relat Res 2008; 466:2827–2832.
5. Forte ML, Virnig BA, Eberly LE, et al. Provider factors associated with intramedullary nail use for intertrochanteric hip fractures. J Bone Joint Surg Am 2010; 95:1105–1114.
6. Anglen JO, Weinstein JN. Nail or plate fixation of intertrochanteric hip fractures: changing pattern of practice. A review of the American Board of Orthopaedic Surgery database. J Bone Jt Surg Am 2008; 90:700–777.
7. Haynes RC, Poll RG, Miles AW, Weston RB. Failure of femoral head fixation: a cadaveric analysis of lag screw cut-out with the gamma locking nail and AO dynamic hip screw. Injury 1997; 28:337–341.
8. Stern R. Are there advances in the treatment of extracapsular hip fractures in the elderly? Injury 2007; 38 (suppl 3):S77–S87.
9. Reindl R, Harvey EJ, Berry GK, Rahme E. Intramedullary versus extramedullary fixation for unstable intertrochanteric fractures: a prospective randomized controlled trial. J Bone Jt Surg Am 2014; 97:1905–1912.
10. Park SY, Yang KH, Yoo JH, Yoon HK, Park HW. The treatment of reverse obliquity intertrochanteric fractures with the intramedullary hip nail. J Trauma 2008; 65:852–857.
11. Makki D, Matar HE, Jacob N, Lipscombe S, Gudena R. Comparison of the reconstruction trochanteric antigrade nail (TAN) with the proximal femoral nail antirotation (PFNA) in the management of reverse oblique intertrochanteric hip fractures. Injury 2015; 46:2389–2393.
12. Irgit K, Richard RD, Beebe MJ, Bowen TR, Kubiak E, Horwitz DS. Reverse oblique and transverse intertrochanteric femoral fractures treated with the long cephalomedullary nail. J Orthop Trauma 2015; 29:e299–e304.
13. Hao Y, Zhang Z, Zhou F, et al. Risk factors for implant failure in reverse oblique and transverse intertrochanteric fractures treated with proximal femoral nail antirotation (PFNA). J Orthop Surg Res 2019; 14:10–17.
14. Min WK, Kim SY, Kim TK, et al. Proximal femoral nail for the treatment of reverse obliquity intertrochanteric fractures compared with gamma nail. J Trauma 2007; 63:1054–1060.
15. Palm H, Jacobsen S, Sonne-Holm S, Gebuhr P. Integrity of the lateral femoral wall in intertrochanteric hip fractures: an important predictor of a reoperation. J Bone Jt Surg Am 2007; 89:470–475.
16. Lambers A, Rieger B, Kop A, D’Alessandro P, Yates P. Implant fracture analysis of the TFNA proximal femoral nail. J Bone Jt Surg Am 2019; 101:804–811.
17. Kashigar A, Vincent A, Gunton MJ, Backstein D, Safir O, Kuzyk PRT. Predictors of failure for cephalomedullary nailing of proximal femoral fractures. Bone Jt J 2014; 96-B:1029–1034.
18. Whale CS, Hulet DA, Beebe MJ, et al. Cephalomedullary nail versus sliding hip screw for fixation of AO 31 A1/2 intertrochanteric femoral fracture: a 12-year comparison of failure, complications, and mortality. Curr Orthop Pract 2016; 27:604–613.
19. Gaebler C, Stanzl-Tschegg S, Tschegg EK, et al. Implant failure of the gamma nail. Injury 1999; 30:91–99.
20. Tantigate D, Riansuwan K, Mahaisavariya B, Sukjaitham K. Breakage of a lag screw of cephalomedullary nail: a technique of removal. Clin Orthop Surg 2015; 7:261–263.
21. Iwakura T, Niikura T, Lee SY, et al. Breakage of a third generation gamma nail: a case report and review of the literature. Case Rep Orthop 2013; 2013:1–5.
22. Bertrand Alvarez D, Paz Aparicio J, Lopez-Anglada Fernandez E, Gonzalez-Busto Mugica I, Núñez Batalla D, Paz Jiménez J. Implant breakage, a rare complication with the Gamma nail: a review of 843 fractures of the proximal femur treated with a Gamma nail. Acta Orthop Belg 2004; 70:435–443.
23. Hsu CE, Shih CM, Wang CC, Huang KC. Lateral femoral wall thickness: a reliable predictor of post-operative lateral wall fracture in intertrochanteric fractures. Bone Jt J 2013; 95 B:1134–1138.
24. DePuy Synthes. TFN-Advanced Proximal Femoral Nailing System: Surgical Technique.
25. Nherera L, Trueman P, Horner A, Watson T, Johnstone AJ. Comparison of a twin interlocking derotation and compression screw cephalomedullary nail (InterTAN) with a single screw derotation cephalomedullary nail (proximal femoral nail antirotation): a systematic review and meta-analysis for intertrochanteric frac. J Orthop Surg Res 2018; 13:1–10.
26. Giannoudis PV, Ahmad MA, Mineo GV, Tosounidis TI, Calori GM, Kanakaris NK. Subtrochanteric fracture non-unions with implant failure managed with the “diamond” concept. Injury 2013; 44 (Suppl 1):S76–S81.
27. Sbiyaa M, El Alaoui A, Admi M, Lahrach K, Marzouki A, Boutayeb F. Intertrochanteric fracture non-unions with implant failure of the gamma nail. Pan Afr Med J 2016; 23:57.
28. von Rüden C, Hungerer S, Augat P, Trapp O, Bühren V, Hierholzer C. Breakage of cephalomedullary nailing in operative treatment of trochanteric and subtrochanteric femoral fractures. Arch Orthop Trauma Surg 2015; 135:179–185.
29. Klima M. Comparison of early fatigue failure of the TFNa and Gamma 3 cephalomedullary nails in the United States from 2015–2019. J Orthop Trauma 2021; 35:e39–e44.
30. Valverde JA, Alonso MG, Porro JG, Rueda D, Larrauri PM, Soler JJ. Use of the gamma nail in the treatment of fractures of the proximal femur. Clin Orthop Relat Res 1998; 56–61.
31. Fine NF, Sexton SA, Williams DH. The learning curve with a new cephalomedullary femoral nail. Injury 2017; 48:1575–1578.
32. Liu W, Zhou D, Liu F, Weaver MJ, Vrahas MS. Mechanical complications of intertrochanteric hip fractures treated with trochanteric femoral nails. J Trauma Acute Care Surg 2013; 75:304–310.
33. Johnson NA, Uzoigwe C, Venkatesan M, et al. Risk factors for intramedullary nail breakage in proximal femoral fractures: a 10-year retrospective review. Ann R Coll Surg Engl 2017; 99:145–150.
34. Klima ML. Comparison of early fatigue failure of the TFNa and Gamma 3 cephalomedullary nails in the United States from 2015 to 2019. J Orthop Trauma 2021; 35:e39–e44.
35. Goodnough LH, Wadhwa H, Tigchelaar SS, et al. Trochanteric fixation nail advanced with helical blade and cement augmentation: early experience with a retrospective cohort. Eur J Orthop Surg Traumatol 2021; 31:259–264.
36. Kammerlander C, Hem ES, Klopfer T, et al. Cement augmentation of the Proximal Femoral Nail Antirotation (PFNA)—a multicentre randomized controlled trial. Injury 2018; 49:1436–1444.
37. Yee DKH, Lau W, Tiu KL, et al. Cementation: for better or worse? Interim results of a multi-centre cohort study using a fenestrated spiral blade cephalomedullary device for pertrochanteric fractures in the elderly. Arch Orthop Trauma Surg 2020; 140:1957–1964.
Keywords:

case report; implant failure; intertrochanteric fracture; non-union; Synthes TFNA

Copyright © 2022 The Authors. Published by Wolters Kluwer Health, Inc. on behalf of the Orthopaedic Trauma Association.