An Augmented-Reality Edge Enhancement Application for Google Glass

Hwang, Alex D.*; Peli, Eli

doi: 10.1097/OPX.0000000000000326
Low Vision and Vision Function: Original Articles

Purpose Google Glass provides a platform that can be easily extended to include a vision enhancement tool. We have implemented an augmented vision system on Glass, which overlays enhanced edge information over the wearer’s real-world view, to provide contrast-improved central vision to the Glass wearers. The enhanced central vision can be naturally integrated with scanning.

Methods Google Glass’ camera lens distortions were corrected by using an image warping. Because the camera and virtual display are horizontally separated by 16 mm, and the camera aiming and virtual display projection angle are off by 10°, the warped camera image had to go through a series of three-dimensional transformations to minimize parallax errors before the final projection to the Glass’ see-through virtual display. All image processes were implemented to achieve near real-time performance. The impacts of the contrast enhancements were measured for three normal-vision subjects, with and without a diffuser film to simulate vision loss.

Results For all three subjects, significantly improved contrast sensitivity was achieved when the subjects used the edge enhancements with a diffuser film. The performance boost is limited by the Glass camera’s performance. The authors assume that this accounts for why performance improvements were observed only with the diffuser filter condition (simulating low vision).

Conclusions Improvements were measured with simulated visual impairments. With the benefit of see-through augmented reality edge enhancement, natural visual scanning process is possible and suggests that the device may provide better visual function in a cosmetically and ergonomically attractive format for patients with macular degeneration.

*PhD

OD, MSc, FAAO

Schepens Eye Research Institute, Massachusetts Eye and Ear, Harvard Medical School, Boston, Massachusetts (both authors).

Alex D. Hwang, Schepens Eye Research Institute 20 Staniford Street Boston, MA 02114 e-mail: alex_hwang@meei.harvard.edu

Most patients with advanced age-related macular degeneration (AMD) experience reduced visual acuity and contrast sensitivity because they have to rely on the residual nonfoveal retina to inspect targets of interest. Although patients’ peri-peripheral vision is sufficient to recognize the gist of scenes,1 their quality of life is significantly affected by the impairment.2–4 The reduced visual function has a large impact on emotional well-being5 and social engagement,6 especially because of its negative impact on tasks such as face recognition,7,8 which require the ability to discriminate fine details or small contrast differences.

The low-vision enhancement system was the first such commercial system using an opaque head-mounted display (HMD). It improved visual acuity and contrast sensitivity by converting the camera image into high contrast-magnified video.9 Later, video-based contrast enhancement and zoom-controlling HMD devices were commercialized, as the Jordy (Enhanced Vision, Huntington Beach, CA) and the SightMate (Vuzix, Rochester, NY). All these devices use full virtual vision HMDs that block the wearer’s natural field of vision.

An alternative approach, augmented reality (AR) using optical see-through HMDs, was proposed by Peli et al.10 for various vision conditions, for example, using wideband image enhancement to enhance the visibility of edges of the scene, as an alternative to magnification.11,12 Luo and Peli13,14 developed a hybrid see-through/opaque HMD device that superimposed a scene edge-view over the wearer’s natural view in see-through mode, but the display could be made opaque to display only a magnified scene edge-view. Using desktop displays, it has been shown that adding high-contrast edges to static images15 or videos16 is preferred by patients with AMD,15,16 and improves visual search performance of older people with simulated central visual loss.15,17

Precise alignment of the augmented edges with the see-through scene is necessary but difficult to achieve in the HMD application, because the camera and the display are usually shifted relative to each other. The parallax produced by that shift varies with distance to the objects of regard. An on-axis HMD-camera configuration was attempted and achieved excellent edge alignment, but the optical system aligning the camera and display reduced display brightness to such an extent that the augmented edges were too dim to provide a significant boost in visibility.18

Google Inc (Mountain View, CA) has recently introduced a novel HMD system, Google Glass, which is intended for an interactive personal communication device. Glass provides a hardware and software development platform that can be applied to vision enhancement for patients with impaired vision. It features a wide-field high-resolution camera, a small optical see-through display positioned above the line of sight, a rechargeable battery, and enough computing power for image processing. The Android-based operating system, which is open for custom app development, supports OpenGL GPU computations and camera access.

Here, we report a preliminary exploration of the utility of Google Glass as a visual aid for patients with AMD by developing an app that provides edge enhancement as an augmented view. The real view of the environment through the Glass display is multiplexed with a scaled cartoonized outline of the view, captured by the camera. The user sees enhanced contrast at the location of edges in the real world. We show how the spatial alignment problem between the augmented edges and the real view can be addressed to provide such an augmented view. We also describe how we generate and display augmented edge information, based on user selectable edge enhancement method and range of enhancement, because the optimal settings of these factors depend on the individual user’s impairment and the immediate task that the user anticipates, as well as the visual environment and lighting conditions. The user interface for controlling system parameters must be simple, intuitive, and quick.

Back to Top | Article Outline

METHODS

Google Glass Hardware

We have measured various system parameters that are not yet provided by Google. The image and video resolution details were derived from the properties of the files recorded by Glass, and the field of view dimensions were measured by identifying the horizontal and vertical endpoints of the captured image, video frame, and see-through display, when the camera was aimed perpendicularly at a marked wall. The Google Glass’ wide field of view camera captures 75.7- by 58.3-degree static images at a resolution of 2528 by 1856 pixels. During video capturing, the upper and lower portions of the captured image are cropped to fit the high-definition screen ratio of 16:9 and then encoded to a resolution of 1280 by 720 pixels (720p) at 30 frames per second. The virtual display is only available for the right eye, covering 13° by 7.3° of the visual field, and has a resolution of 640 by 360 pixels (Fig. 1A). The camera and display are horizontally separated by 16 mm but encased in a single rigid compartment. Therefore, adjustment of the position or angle of the display also changes camera position and aiming angle. The optics for the display provides a small eye box angled 7° downward; thus, the virtual display is only visible if the display is positioned at the designated location (7° above the wearer’s line of sight). The camera is aimed 10° downward relative to the virtual display (Fig. 1B). This setting allows more natural camera aiming direction (3° downward) when the user is standing upright and looking straight into the display by tilting the head down 7°. The outer surface of the see-through display is coated with photochromic material, so that the see-through view is dimmer outdoors (when exposed to UV radiation) to improve the contrast/visibility of the displayed image, just like the transition spectacle lenses from Transitions Optical, Inc (Pinellas Park, FL), but the coating is applied to the display’s external surface for better exposure to the UV, and therefore, they darken faster than the spectacle lenses.

FIGURE 1

FIGURE 1

Back to Top | Article Outline

Design Assumptions for Augmented Vision Alignment

Assume the following conditions make AR alignment simpler: (1) the virtual display plane is perpendicular to the wearer’s line of sight, and (2) the line of sight passes through the center of the virtual display. In the case of Google Glass, when aiming straight ahead, the wearer must tuck his or her chin down 7° to position the virtual display perpendicular to and centered on the line of sight. This is similar to the usage of low-vision bioptic telescopes, where the user views the scene through the carrier lens most of the time but glimpses through the telescope by tilting the head downward.19 Bioptic telescopes are typically set at 10° upward.

Back to Top | Article Outline

Matching the Size of the Video Image to the See-Through View

The field of view of the camera for video preview and the virtual display span 75.7° by 42.6° and 13° by 7.3° of visual angle and have 1920 by 1080 and 640 by 360 pixels, respectively. Note that 1080p is the maximum resolution available directly from the camera for video preview, before being encoded to the default video resolution (720p). If the full captured image is drawn to fill the virtual display, as in normal video preview during the recording, the image displayed will be minified by a factor of ×5.8, compared with the see-through view. To match the size of the AR image to the real world, only a small portion of the captured image (330 by 186) needs to be displayed on the virtual display (see the cropped-out area marked in Fig. 1C), and the image should be magnified by a factor of ×1.94 to fill the virtual display.

Back to Top | Article Outline

Correcting Camera Lens Distortion

The wide field of view camera on Glass introduces a nonuniform optical distortion (a barrel distortion at the image center, gradually turning into a pincushion distortion toward the periphery) with less distortion in the central field of view (Fig 1C, top). The distortion becomes more apparent as the eccentricity from the center of the camera increases. In on-axis camera-display HMD,18 this type of distortion causes less problems because the higher distortion is further from the center and the image portion to be used (with large magnification) is mostly away from the distorted field. However, for an off-axis camera-display AR system like Google Glass, controlling distortions at the image periphery is necessary because the image portion to be displayed on the virtual display (marked with a dashed line in Fig. 1C) is taken from the more distorted upper-left portion of the captured image (attributed to the 10° vertical rotational difference and 16 mm horizontal disparity between camera and the virtual display).

We measured the camera distortion by photographing a square grid perpendicularly, from 8 inches away. All intersections captured were marked, and distortions for each intersection were measured as the horizontal and vertical offsets from the corresponding intersections in a uniform grid. Our custom calibration program produces a black/white grid (with intersections being marked as red) on a desktop monitor and analyzes the captured grid image by checking the position of those red dots. A corrective mesh was created by subtracting the offsets from the corresponding intersection coordinates from the uniform grid. This corrective mesh was used as a vertex for the texture fragment shader process20 and was applied to warp the captured image and compensate for the lens distortion (Fig. 1C, bottom). Unlike the camera, the optical display does not show any apparent spatial distortions; thus, no additional correction was needed.

Back to Top | Article Outline

Selection of Displayed Image Portion

Once the captured image is corrected for camera distortion, the image is projected onto the display plane. Because the virtual display plane is tilted upward 10° relative to the camera aiming direction (Fig. 1B), the image is projected to a rotated (10° downward) display plane to compensate for the relative angular difference between the camera aiming direction and the virtual display orientation. An additional two-dimensional translation must be applied to align the magnified image to the see-through view (Fig. 1A). The displacement between the camera and the virtual display axes produces a parallax, where the magnitude of image misalignment between captured images and the view through the virtual display varies with the distance to the aimed object. Because near objects (less than 10 ft away) are usually easier for AMD patients to recognize, the default operational correction for the parallax effect we applied is for 10 ft. The user may change the parallax correction for closer, default, or farther distances by swiping the Glass touchpad forward or backward, where forward swiping intuitively corrects for farther parallax. The disparity caused by parallax decreases as distance to the object increases, for example, for objects at 1, 3, 10, and 30 ft away, the amount of disparity correction that needs to be applied is 3.0, 1.0, 0.3, and 0.1°, respectively. Fig. 2 shows the overall image processing flow implemented with OpenGL and Shader,20 where the displayed portion of image is defined following the lens distortion correction, and edge extraction is only applied to the smaller displayed portion of the image.

FIGURE 2

FIGURE 2

Back to Top | Article Outline

Edge Enhancement Method

Two types of edge enhancement methods have been implemented so far: the positive and negative Laplacian filters.21 The positive Laplacian filter computes the second order gradient of pixel brightness and suppresses below-threshold pixel gradient values. The results on a see-through display will be enhanced bright edges with clear surroundings (Fig. 3A). The negative Laplacian filter gives the opposite effect; edges become transparent but the surroundings of the edges are highlighted (Fig. 3B).

FIGURE 3

FIGURE 3

Users can choose to enhance one of three contrast-dependent edge levels. Fig. 3C to E illustrates the effects of selecting edge contrast level, when enhancing the view of a Pelli-Robson contrast sensitivity chart22 (Fig. 3F). At the high-threshold setting, only edges with high contrast are enhanced (Fig. 3C); at the medium setting, only the medium-contrast edges will be enhanced whereas high-contrast edges will not be enhanced (Fig. 3D); at the low setting, only weak edges are enhanced (Fig. 3E). This adjustable enhancement (bilevel thresholding) is desired because individual AMD patients have varying levels of contrast sensitivity, and most of them may be able to see strong edges even without enhancement. Enhancing already visible edges may not be helpful or desired. It has also been demonstrated that for some object recognition tasks, like face and character recognition, selectively enhancing the contrast of the critical frequencies band for a specific contrast range results in better preference or performance.12,15,23 Therefore, the user should be able to choose the range of edge contrast for enhancement that fits his or her needs. This approach of selective contrast-level enhancement was described and proposed previously,10,13,14,24,25 but, to our knowledge, this is the first such implementation on a see-through HMD device.

If the system enhances edges with contrast lower than a single threshold, it typically also increases noise in the image. Therefore, the enhancement level should be defined by two contrast thresholds, upper and lower, where the upper threshold is matched with the user’s contrast sensitivity to increase the contrast of just invisible details, and the lower threshold limits enhancement of image noise (Fig. 3C to E).

Back to Top | Article Outline

Contrast Sensitivity Measurement

We conducted a preliminary study to examine the effectiveness of edge enhancement. Three normal-vision subjects (aged 25 to 45 years; visual acuity of 20/30 or better) participated and had their contrast sensitivity measured using a Pelli-Robson contrast sensitivity chart,22 at 1 m away, under recommended lighting conditions (the chart was illuminated as uniformly as possible without any glare, and white areas were ∼85 cd/m2). Normal-vision subjects were used because the effect of edge enhancement depends more on the Glass’ hardware performance (camera and display) than the individual’s contrast sensitivity threshold. With the maximum enhancement setting, which enhances the full range of edge contrasts, the edges drawn on the virtual display produced higher contrast than the highest contrast character on the contrast sensitivity chart (by more than a factor of 2, in terms of pixel values of white background of the chart and white AR edges in Fig. 4).

FIGURE 4

FIGURE 4

For each subject, contrast threshold (the lowest visible contrast) was measured under four conditions: with and without a heavy diffuser filter (which effectively raises the contrast detection threshold of subjects by a factor of 4), and with and without edge enhancement. The diffuser filter was inserted between the Glass’ display compartment and subject’s eye.

Back to Top | Article Outline

RESULTS

Alignment of AR Edges with See-Through View

Some spatial misalignment of the off-axis camera-display system is unavoidable. In our implementation, the misalignment is smaller than 0.3° for edges farther than 10 ft away (less than 2.5 pixels offset in the original image). This misalignment may go unnoticed because the amount is comparable to the thickness of the enhanced edges. Fig. 4 demonstrates indoor and outdoor AR alignment with the maximum enhancement setting. Some misalignment is visible in Fig. 4B because the misalignment for edges located farther than 10 ft is magnified on the virtual display by a factor of 1.94. Note that the misalignment is less visible in Fig. 4C or D, as the offset at farther distances is too small to be visible even with the magnification.

Back to Top | Article Outline

Effects of Edge Enhancement

Subjects’ contrast sensitivity improved substantially with AR edge enhancement, compared to their contrast sensitivity without the Glass. The contrast sensitivity benefits become larger with insertion of a diffuser filter in front of the subjects’ eye. However, this improvement is bound by the sensitivity of the camera and could not improve beyond the log contrast sensitivity of 1.50 (Table 1). Thus, the current system may provide letter contrast sensitivity improvement only for patients with letter contrast sensitivity worse than 1.50.

TABLE 1

TABLE 1

With the maximum edge enhancement setting, the effect of the noise becomes stronger when the scene has low contrast, because the signal-to-noise ratio is decreased. An example of the AR edge enhancement for the contrast sensitivity chart is shown in Fig. 5.

FIGURE 5

FIGURE 5

Back to Top | Article Outline

DISCUSSION

Utility of Enhancement over a Small Section of the Visual Field

The 13- by 7.3-degree span of the Glass virtual display may seem to be too small to provide practical benefits to AMD patients. However, considering that the typical distal field of view through a ×3.0 bioptic telescope spans between 6 and 12° in diameter,26 yet it delivers functional utility, we can expect at least a similar level of functionality in terms of visual field size. The foveal and parafoveal visual field (<5° eccentricity) provides the most detailed visual information and is critical for object recognition.27,28 The field within that or just outside of that range provides the same functionality to patients with central scotomas.

The edge enhancement AR approach has some advantages over bioptic telescope magnification. Bioptic magnification creates a ring scotoma that obstructs instant integration of the central view from the wide peripheral view,29,30 making image navigation difficult. For example, when a ×3.0 bioptic telescope with a 12-degree field-of-view is used, 36° of circular retinal surface (shown as a dotted red circle in Fig. 6A) will be filled with the ×3.0 magnified view of the 12-degree field of view in the real world. This “ring scotoma” makes it difficult for the user to see the surrounding environment located between 6 and 18° eccentricity, when looking through the telescope. This is true even if the telescope is used monocularly. With the Glass AR implementation of contrast enhancement, simultaneous integration of both central and peri-peripheral scene perception is possible with no ring scotoma. This implementation affords the patient the benefit of natural head scanning processes, for example, planning of scanning and continuous readjustment of instantaneous aiming direction.

FIGURE 6

FIGURE 6

Fig. 6 illustrates a hypothetical head scanning sequence of an AMD patient using a Google Glass edge enhancement application. With a central scotoma of about 7° in diameter, the patient fixates with preferred retinal locus (PRL) under the scotoma (Fig. 6A), as measured by the Nidek MPI (Padova, Italy). Although high spatial frequency-low contrast features are not recognizable (as illustrated in the blurred images of Fig. 6C to F), the patient can still see the overall layout of the scene using his residual peri-peripheral vision. Therefore, the patient is able to identify the series of possible objects of interest and plan to examine those objects sequentially with the enhanced view using head scanning, as diagrammed in Fig. 6B.

A space variant blurring31 was applied to the scene image (representing the 75.7-degree horizontal camera visual field) to illustrate the reduction in sensitivity with increasing retinal eccentricity. In addition, to illustrate the high brightness of edges on the virtual display, the brightness of the see-through image has been scaled down to fit into the bottom half of the pixel brightness range, whereas the onscreen edge brightness is fit into the upper half of the brightness range. This produces an illusion of a high dynamic range32 AR view, as is experienced by the user.

As scene scanning is executed, detailed observations, aided by the enhanced edge information, can be made at each gaze position (Fig. 6C to F), whereas the location of the next scanning target can be perceived with peripheral vision, and the gist of the scene is undisturbed. This is a clear benefit of the see-through HMD design that provides enhanced vision near the patient’s PRL while preserving peripheral vision.33

In the transition from Fig. 6D to Fig. 6E, the next scanning target may appear to be blocked in part by the physiological scotoma, but this location can be seen usually by the left eye, illustrating another advantage of this method that benefits from the binocular reduction of scotomas. Note that although we made an effort to generate perceptual views of a patient, Fig. 6 is still a conceptual illustration of the functionality and does not represent a veridical simulation. Verification of such a simulation is complex and laborious33–36 and is beyond the scope of this article.

The sequence of head scanning enables the patient to acquire “central” visual information, while using near-peripheral vision, in a process very similar to natural foveation during eye scanning. The natural aspect of this function makes it intuitive and easy to adopt. The same scanning can be applied with smaller head movements to examine details of an object that is slightly larger than the field of view of the display. The ability to fuse the two eyes’ peripheral vision37–39 contributes to the comfortable use of the display and may enable more extended use than the short glimpses, which are typical of bioptic use.

The lack of magnification also eliminates the vestibular-visual motion signal conflicts, which may cause difficulties in adaptation to the bioptic telescope.40–42 The bioptic telescope depends on magnification, where high spatial frequency image contents are converted to low spatial frequencies, whereas the edge enhancement we implemented for Glass enhances the contrast of edges over a wide range of spatial frequencies (wide-band enhancement).11,12

Back to Top | Article Outline

Limitations of Current Implementation

The illustrations in Fig. 6 also point to a known limitation of the current implementation of the AR edge enhancement. With the optical see-through display, enhanced edges must be drawn as white/bright pixels because dark/black edges in the see-through display become transparent. As a result, the current implementation of edge enhancement increases contrast well in dark areas of the scene (as shown in Fig. 6D, F) but does not provide the same level of contrast enhancement over bright areas (as shown in Fig. 6C, E). A possible solution would be covering the front (and possibly the top and bottom as well) of the Glass display compartment with black masking tape and converting it into an opaque display. With this video see-through configuration, contrast enhancement can be achieved easily by displaying not only the edges but also the full magnification-matched video image enhanced by bipolar (black and white) edges43,44 and possibly additional contrast enhancement approaches.11,45 Such functionality is possible thanks to the small frame-free format of the display and the ease with which it can be taken out of the line of sight when it is not needed. The use of video see-through in the implementation will also assure perfect alignment of the edges to the natural (photographed) view. The parallax will cause a shift of the whole display screen image relative to its surrounding, but that will be small, and importantly, it will not degrade image quality.

It also can be observed that strong edge enhancement may mask weak-contrast, high-frequency visual features if they are close to the strong edges. A more sophisticated algorithm that considers the proximity of local edges should be explored. The enhancement of only moderate-contrast edges discussed above is also a possible solution to this limitation.

Back to Top | Article Outline

CONCLUSIONS

Increased edge contrast has been demonstrated to be preferred by patients with AMD and other low-vision conditions when watching TV12,16,45 or viewing images.15,17 Edge enhancement has also been demonstrated to improve performance in a search task performed on a computer screen.15–17,45 The utility of this approach for enhancing the natural environment has been suggested, but so far has not been implemented because of technological limitations. The Google Glass appears to overcome many of these limitations and thus may provide a valuable platform for implementing a variety of applications that can aid patients with various low-vision conditions (1) at a very reasonable cost, (2) with a cosmetically acceptable and socially desirable format, and (3) with flexibility that supports further innovation. At this stage, we think that a limitation posed to Google Glass in using it as a visual aid is its relatively short battery runtime (∼40 minutes per full charge with our AR edge application). However, the short runtime problem can be easily overcome by adding an external power pack. Many applications may be developed on this platform to meet the various needs of low-vision patients. We are looking forward to evaluating the utility of such applications with patients and improving their function.

Alex D. Hwang

Schepens Eye Research Institute

20 Staniford Street

Boston, MA 02114

e-mail: alex_hwang@meei.harvard.edu

Back to Top | Article Outline

ACKNOWLEDGMENTS

Supported in part by National Institutes of Health Grants R01EY012890 and R01EY005957 and by a gift from Google Inc. EP has patent rights, through a patent assigned to the Schepens Eye Research Institute, for the use of wide-band enhancement for impaired vision in HMD. EP is also a consultant to Google Inc. on the Glass project.

Received February 28, 2014; accepted May 9, 2014.

Back to Top | Article Outline

REFERENCES

1. Tran TH, Rambaud C, Despretz P, Boucart M. Scene perception in age-related macular degeneration. Invest Ophthalmol Vis Sci 2010; 51: 6868–74.
2. Hassell JB, Lamoureux EL, Keeffe JE. Impact of age related macular degeneration on quality of life. Br J Ophthalmol 2006; 90: 593–6.
3. Cahill MT, Banks AD, Stinnett SS, Toth CA. Vision-related quality of life in patients with bilateral severe age-related macular degeneration. Ophthalmology 2005; 112: 152–8.
4. Stevenson MR, Hart PM, Montgomery AM, McCulloch DW, Chakravarthy U. Reduced vision in older adults with age related macular degeneration interferes with ability to care for self and impairs role as carer. Br J Ophthalmol 2004; 88: 1125–30.
5. Lamoureux EL, Pallant JF, Pesudovs K, Tennant A, Rees G, O’Connor PM, Keeffe JE. Assessing participation in daily living and the effectiveness of rehabilitation in age related macular degeneration patients using the impact of vision impairment scale. Ophthalmic Epidemiol 2008; 15: 105–13.
6. Bennion AE, Shaw RL, Gibson JM. What do we know about the experience of age related macular degeneration? A systematic review and meta-synthesis of qualitative research. Soc Sci Med 2012; 75: 976–85.
7. Peli E, Goldstein RB, Young GM, Trempe CL, Buzney SM. Image enhancement for the visually impaired. Simulations and experimental results. Invest Ophthalmol Vis Sci 1991; 32: 2337–50.
8. Peli E, Lee E, Trempe CL, Buzney S. Image enhancement for the visually impaired: the effects of enhancement on face recognition. J Opt Soc Am (A) 1994; 11: 1929–39.
9. Massof RW, Baker FH, Dagnelie G, DeRose JL, Alibhai S, Deremeik JT, Ewart C. Low Vision Enhancement System: improvements in acuity and contrast sensitivity. Optom Vis Sci 1995; 72 (Suppl.): 20.
10. Peli E, Luo G, Bowers A, Rensing N. Applications of augmented vision head-mounted systems in vision rehabilitation. J Soc Inf Disp 2007; 15: 1037–45.
11. Peli E. Wide-band image enhancement. US Patent No 6,611,618 B1. Washington, DC: U.S. Patent and Trademark Office; 2003.
12. Peli E, Kim J, Yitzhaky Y, Goldstein RB, Woods RL. Wideband enhancement of television images for people with visual impairments. J Opt Soc Am (A) 2004; 21: 937–50.
13. Luo G, Peli E. Use of an augmented-vision device for visual search by patients with tunnel vision. Invest Ophthalmol Vis Sci 2006; 47: 4152–9.
14. Luo G, Peli E. Development and evaluation of vision rehabilitation devices. Conf Proc IEEE Eng Med Biol Soc 2011; 2011: 5228–31.
15. Satgunam P, Woods RL, Luo G, Bronstad PM, Reynolds Z, Ramachandra C, Mel BW, Peli E. Effects of contour enhancement on low-vision preference and visual search. Optom Vis Sci 2012; 89: E1364–73.
16. Wolffsohn JS, Mukhopadhyay D, Rubinstein M. Image enhancement of real-time television to benefit the visually impaired. Am J Ophthalmol 2007; 144: 436–40.
17. Kwon M, Ramachandra C, Satgunam P, Mel BW, Peli E, Tjan BS. Contour enhancement benefits older adults with simulated central field loss. Optom Vis Sci 2012; 89: 1374–84.
18. Luo G, Rensing N, Weststrate E, Peli E. Registration of an on-axis see-through head-mounted display and camera system. Opt Eng 2005; 44: 024002.
19. Doherty AL, Bowers AR, Luo G, Peli E. Object detection in the ring scotoma of a monocular bioptic telescope. Arch Ophthalmol 2011; 129: 611–7.
20. Bax MR. Real-time lens distortion correction: 3D video graphics cards are good for more than games. Stanford ECJ 2002; 1: 9–13.
21. Watt A, Policarpo F. The Computer Image. Essex, England: Addison Wesley Longman Ltd.; 1998: 248–9.
22. Solomon JA, Pelli DG. The visual filter mediating letter identification. Nature 1994; 369: 395–7.
23. Peli E. Limitations of image enhancement for the visually impaired. Optom Vis Sci 1992; 69: 15–24.
24. Peli E, Luo G, Bowers A, Rensing N. Development and evaluation of vision multiplexing devices for vision impairments. Int J Artif Intell Tools 2009; 18: 365–78.
25. Pelli D, Robson J, Wilkins A. The design of a new letter chart for measuring contrast sensitivity. Clin Vis Sci 1988; 2: 187–99.
26. Nguyen A, Nguyen AT, Hemenger RP, Williams DR. Resolution, field of view, and retinal illuminance of miniaturized bioptic telescopes and their clinical significance. J Vis Rehab 1993; 7: 5–9.
27. Strasburger H, Rentschler I, Juttner M. Peripheral vision and pattern recognition: a review. J Vis 2011; 11: 13.
28. Pomplun M, Reingold EM, Shen J. Investigating the visual span in comparative search: the effects of task difficulty and divided attention. Cognition 2001; 81: B57–67.
29. Doherty AL, Bowers AR, Luo G, Peli E. The effect of strabismus on object detection in the ring scotoma of a monocular bioptic telescope. Ophthalmic Physiol Opt 2013; 33: 550–60.
30. Peli E, Vargas-Martin F. In-the-spectacle-lens telescopic device. J Biomed Opt 2008; 13: 034027.
31. Geisler WS, Perry JS. Real-time simulation of arbitrary visual fields. ETRA’02 Proceedings of the 2002 Symposium on Eye Tracking Research and Applications, New York, NY: ACM; 2002: 83–7.
32. Reinhard E, Heidrich W, Debevec P, Pattanaik S, Ward G, Myszkowski K. High Dynamic Range Imaging: Acquisition, Display, and Image-Based Lighting, 2nd ed. Burlington, MA: Morgan Kaufmann; 2010.
33. Peli E. Vision multiplexing: an engineering approach to vision rehabilitation device development. Optom Vis Sci 2001; 78: 304–15.
34. Peli E, Arend L, Labianca AT. Contrast perception across changes in luminance and spatial frequency. J Opt Soc Am (A) 1996; 13: 1953–9.
35. Peli E. Contrast sensitivity function and image discrimination. J Opt Soc Am (A) 2001; 18: 283–93.
36. Peli E, Geri GA. Discrimination of wide-field images as a test of a peripheral-vision model. J Opt Soc Am (A) 2001; 18: 294–301.
37. Kertesz AE. The effectiveness of wide-angle fusional stimulation in the treatment of convergence insufficiency. Invest Ophthalmol Vis Sci 1982; 22: 690–3.
38. Kertesz AE, Hampton DR. Fusional response to extrafoveal stimulation. Invest Ophthalmol Vis Sci 1981; 21: 600–5.
39. Kertesz AE, Lee HJ. The nature of sensory compensation during fusional response. Vision Res 1988; 28: 313–22.
40. Demer JL, Porter FI, Goldberg J, Jenkins HA, Schmidt K. Adaptation to telescopic spectacles: vestibulo-ocular reflex plasticity. Invest Ophthalmol Vis Sci 1989; 30: 159–70.
41. Chmielowski RJM, Chmielowski PFM. A decentric system for fitting telescopes in home-trial frames. In: Stuen C, Arditi A, Horowitz A, Lang MA, Rosenthal B, Seidman KR, eds. Vision Rehabilitation: Assessment, Intervention, and Outcomes. Lisse, Netherlands: Swets & Zeitlinger; 2000: 191–4.
42. Peli E. The optical functional advantages of an intraocular low-vision telescope. Optom Vis Sci 2002; 79: 225–33.
43. Peli E. Feature detection algorithm based on a visual system model. Proc IEEE 2002; 90: 78–93.
44. Peli E. Recognition performance and perceived quality of video enhanced for the visually impaired. Ophthalmic Physiol Opt 2005; 25: 543–55.
45. Fullerton M, Peli E. Post transmission digital video enhancement for people with visual impairments. J Soc Inf Disp 2006; 14: 15–24.
Keywords:

Google Glass; HMD; vision enhancement; vision rehabilitation; contrast enhancement; low vision

© 2014 American Academy of Optometry