Review Article

Diagnostic performance of PET/computed tomography versus PET/MRI and diffusion-weighted imaging in the N- and M-staging of breast cancer patients

de Mooij, Cornelis Maartena,,b,,c; Sunen, Inésa,,d; Mitea, Cristinaa,,c; Lalji, Ulrich C.a; Vanwetswinkel, Sigrida; Smidt, Marjolein L.b,,c; van Nijnatten, Thiemo J.A.a,,c

Author Information
Nuclear Medicine Communications: October 2020 - Volume 41 - Issue 10 - p 995-1004
doi: 10.1097/MNM.0000000000001254

Abstract

Introduction

Breast cancer is the most commonly diagnosed cancer in women worldwide with over two million new cases annually [1]. Despite decreasing mortality rates over the past few decades due to earlier diagnosis and improved treatment modalities, an estimated 600 000 women succumbed to breast cancer in 2018 [1]. Prognosis is mainly determined by the individual tumor burden reflected by the decrease of the 5-year survival rate from 99% in localized, 85% in regional, to 27% in metastasized breast cancer [2]. The median overall survival in patients with distant metastases is only 26 months and while the breast cancer subtype predisposes the site of distant metastases, bone structures are generally the most frequent location of metastatic dissemination followed by the lung and liver [3–5].

Accurate staging is of paramount importance because it determines the ideal treatment for each patient regarding surgery, (neo)adjuvant systemic therapy (NST), and radiotherapy. Breast cancer imaging for disease staging encompasses a wide range of modalities and includes, among others, MRI and PET with computed tomography (PET/CT) using 18F-fluorodeoxyglucose (18F-FDG) as a radioactive tracer [6]. Breast MRI is widely used to assess the locoregional extent of disease in patients as well as monitoring response to NST [7,8]. Complementary, whole-body 18F-FDG PET/CT has become an established imaging technique for the assessment of node positive, distant metastatic, and recurrent breast cancer [9].

Shortcomings of 18F-FDG PET/CT, like its inability of simultaneous acquisition and reduced soft tissue contrast, have recently led to the development of hybrid positron emission tomography/magnetic resonance imaging (PET/MRI) [10]. Theoretically, 18F-FDG PET/MRI could offer an attractive one-stop-shop solution for patients with locally advanced breast cancer with affected axillary lymph nodes who should undergo both breast MRI for locoregional staging and whole-body 18F-FDG PET/CT for nodal and distant staging. 18F-FDG PET/MRI could then subsequently be used to assess the response to NST [11,12]. Moreover, using 18F-FDG PET/MRI for breast cancer staging allows the possibility to perform diffusion-weighted imaging (DWI) as part of the MRI protocol.

DWI, an MRI sequence showing the restricted movement of water molecules in tumoral tissues, was recently introduced as a functional modality to evaluate microstructural characteristics in tumors [13]. DWI can visualize changes occurring at a cellular spatial scale making it an advantageous tool for evaluating changes in the tumor microenvironment, both before and after NST [14]. Since its recent introduction promising results have been reported on the value of DWI in the diagnostic work-up of breast cancer [15,16].

Several publications in recent literature have reported on the diagnostic performance of whole-body 18F-FDG PET/CT, 18F-FDG PET/MRI, and DWI in breast cancer patients. To investigate if 18F-FDG PET/MRI could safely replace 18F-FDG PET/CT, possibly with the addition of a DWI sequence to the MRI protocol, studies comparing these imaging modalities in the same cohort have to be evaluated. Therefore, the aim of the current study is to provide a systematic review summarizing the diagnostic performance of 18F-FDG PET/MRI and DWI compared to 18F-FDG PET/CT with a specific focus on nodal and distant staging in breast cancer patients. 

Methods

Search strategy

For this systematic review, the guidelines of Preferred Reporting Items for Systematic Reviews and Meta-analysis for Diagnostic Test Accuracy Studies (PRISMA-DTA) were followed [17]. The Embase and PubMed databases were searched independently by two authors (C.M.d.M. and I.S.). The last search was run on 28 April 2020. Search terms used for the condition were breast or mamma combined with tumor, cancer, malignancy, and carcinoma. Terms for the imaging modalities were positron emission tomography computed tomography, positron emission tomography magnetic resonance imaging, diffusion weighted imaging, and their corresponding abbreviations (PET/CT, PET/MRI, and DWI). No specific terms were used to search for outcome. Detailed descriptions of the literature search strategies are provided in the Supplementary material, Supplemental digital content 1, http://links.lww.com/NMC/A174. A manual search of the reference lists of the retrieved articles was performed to identify any additional publications.

Eligibility criteria

All primary diagnostic test accuracy studies that evaluated the diagnostic performance regarding nodal and distant staging of breast cancer patients with 18F-FDG PET/CT versus 18F-FDG PET/MRI or 18F-FDG PET/CT versus DWI were considered eligible for inclusion. To avoid selection bias, in- and exclusion criteria were established prior to the literature search. Inclusion criteria were as follows: (1) staging of patients with primary or recurrent breast cancer; (2) reported accuracy of imaging modalities regarding nodal and distant staging; and (3) a comparison of the diagnostic performance of 18F-FDG PET/CT versus 18F-FDG PET/MRI or 18F-FDG PET/CT versus DWI. Exclusion criteria were as follows: (1) response evaluation to NST; (2) publication not available in English; and (3) editorials, conference publications, surveys, case reports, reviews, ex-vivo studies, and animal studies.

Study selection

Two independent reviewers (C.M.d.M. and I.S.) searched for eligible articles and excluded duplicates. After excluding irrelevant articles based on title and abstract, the full-text of the remaining articles were obtained and read thoroughly to check for eligibility. Any disagreements regarding eligibility were resolved by the intervention of a third reviewer (T.J.A.N.).

Data extraction and quality assessment

Data extraction was performed independently by two reviewers (C.M.d.M. and I.S.) using a standardized extraction form, and the final extraction was completed in a consensus meeting. The following data were extracted: first author, year of publication, country, study design (retrospective or prospective), index tests, reference standard, follow-up time, type of cancer (primary or recurrent; local, regional, or metastasized), pathology, magnetic field strength, acquired imaging sequences, contrast agent used, number of reviewers, blinding, and parameters of diagnostic performance such as sensitivity, specificity, and accuracy. The quality of the included studies was independently assessed by two reviewers (C.M.d.M. and I.S.) using the Quality Assessment of Diagnostic Accuracy Studies 2 (QUADAS-2) [18].

Statistical analysis

Diagnostic performance as stated in the primary article was extracted using a standardized extraction form. Due to the high heterogeneity between and within studies and the inability to compute 2 × 2 contingency tables of at least three studies for two or more imaging modalities regarding any of the anatomic locations, no meta-analysis was performed. Instead, descriptive tables were used to provide a clear overview of the diagnostic performance of the primary studies.

Results

Study selection

After removing 38 duplicates, a total of 561 potentially eligible studies were identified in the primary search. Titles and abstracts of these 561 studies were read and in- and exclusion criteria were applied. Subsequently, 36 studies were reviewed based on full text. Finally, a total of 11 articles published between 2010 and 2020 were selected for this systematic review [19–29]. The search and selection processes are summarized in Fig. 1.

Fig. 1
Fig. 1:
Flowchart of study selection process according to the Preferred Reporting Items for Systematic Reviews and Meta-Analyses (PRISMA).

Characteristics of included studies

Eight of the 11 included studies (73%) had a prospective study design [19,22–26,28,29]. The included studies contained a total of 476 patients (mean 43 per study, range 13–109). Six articles reported results comparing 18F-FDG PET/CT with 18F-FDG PET/MRI [19,21,24,26,27,29] four articles compared 18F-FDG PET/CT with DWI [22,23,25,28] and one article compared 18F-FDG PET/CT with both 18F-FDG PET/MRI and DWI [20]. Detailed information of the characteristics of the included studies is presented in Table 1.

Table 1 - Study characteristics
Author Year Country Study design n Index tests Reference standard Follow-up time Patients Pathology
Grankvist et al. [23] 2010 Denmark P 13 PET/CT, DWI PET/CT nr Suspected or known metastasis, treated and untreated nr
Heusner et al. [25] 2010 Germany P 20 PET/CT, DWI Pathology, concordant imaging findings, follow-up nr Initial and recurrent, untreated IDC (12), ILC (4), others (4)
Ergul et al. [22] 2014 Turkey P 24 PET/CT, DWI Pathology nr Initial (stage I–II), untreated IDC (19), ILC (1), others (4)
Catalano et al. [21] 2015 Italy R 109 PET/CT, PET/MRI Follow-up (109) >11 months Initial and recurrent, treated and untreated IDC (109)
Grueneisen et al. [24] 2015 Germany P 49 PET/CT, PET/MRI Pathology (48), follow-up (1) 629 days Initial, untreated IDC (39), ILC (7), others (4)
Sawicki et al. [29] 2015 Germany P 21 PET/CT, PET/MRI Pathology (10), follow-up, prior imaging 16 months Recurrent, untreated IDC (12), ILC (4), unknown (5)
Melsaether et al. [26] 2016 USA P 51 PET/CT, PET/MRI Pathology (12), clinical follow-up (9), imaging follow-up (40), consensus (2) 19 months Initial and recurrent, treated and untreated nr
Pujara et al. [27] 2016 USA R 35 PET/CT, PET/MRI Follow-up (28), prior imaging (33), consensus 12 months Suspected or known metastasis, untreated IDC (15), ILC (2), unknown (4)
Catalano et al. [20] 2017 USA R 51 PET/CT, PET/MRI, DWI Pathology (42), follow-up (9) >24 months Initial, untreated IDC (51)
Botsikas et al. [19] 2018 Switzerland P 80 PET/CT, PET/MRI Pathology (64), follow-up (16) >12 months Initial and recurrent, untreated IDC (69), ILC (5), others (6)
Rezk et al. [28] 2019 Egypt P 23 PET/CT, DWI Pathology, follow-up 6–12 months Recurrent nr
CT, computed tomography; DWI, diffusion-weighted imaging; IDC, invasive ductal carcinoma; ILC, invasive lobular carcinoma; nr, not reported; P, prospective; R, retrospective.

Methodological quality assessment

Quality assessment of the included studies is summarized in Table 2. Significant risk of bias as well as significant applicability concerns were observed in all of the included studies except one [21]. Overall, the reference standard contained the highest risk of bias. This was mainly caused by no use of pathology, no use of an acceptable duration of follow-up, or the use of consensus between the imaging modalities to serve as a reference standard. Risk of bias regarding patient selection was mainly due the lack of blinding and the risk of recall bias. Concerns regarding applicability for the patient selection were due to the heterogeneous patient populations and the use of early stage breast cancer patients in whom the chances of having regional or distant metastases are generally lower.

Table 2 - Quality assessment of the included articles based on Quality Assessment of Diagnostic Accuracy Studies 2
Risk of bias Applicability concerns
Study Patient selection Index test Reference standard Flow and timing Patient selection Index test Reference standard
Grankvist et al. [23] Low Low High Low High High High
Heusner et al. [25] Low Low High Low Low Low Low
Ergul et al. [22] Low High Low Low High Low Low
Catalano et al. [21] Low Low Low Low Low Low Low
Grueneisen et al. [24] Low High Low Low Low Low Low
Sawicki et al. [29] Low Low High High Low Low Low
Melsaether et al. [26] Low Low High Low High Unclear Low
Pujara et al. [27] Low Low High Low Low Unclear Low
Catalano et al. [20] High Low Low Low Low Low Low
Botsikas et al. [19] Low Low Low Low Low High Low
Rezk et al. [28] High High Unclear Low Low Low High

Technical details

All included studies used 18F-FDG as a radiotracer. In six studies, the field strength was 3.0 T [19,23,24,26,27,29] in three 1.5 T [22,25,28] and in two it was not reported [20,21]. PET and MRI images were simultaneously acquired in all but one study evaluating PET/MRI, the exception being a sequential acquisition [19]. In five studies, reviewers assessed the images of the imaging modalities separately [22,23,25–27] while this was performed by the same reviewers in another five articles [19–21,24,29]. The remaining publication is unclear about image interpretation [28]. Intravenous contrast was administrated in all studies; however, only five studies used contrast in both techniques [20,21,24,25,29]. Detailed descriptions of the imaging sequences used for PET/MRI and DWI are depicted in Table 3.

Table 3 - Technical details of imaging
First author Index tests Field strength PET/MRI acquisition Intravenous contrast Imaging sequences [(PET)-MRI] Reviewers Blinding Delay a
Grankvist et al. [23] PET/CT, DWI 3.0T na Yes, no Neck-thorax, lumbar and pelvic: T1, STIR, T2 fat suppressed. Whole-body: DWI (b-value nr) 2 + 2 Yes na
Heusner et al. [25] PET/CT, DWI 1.5T na Yes, yes Whole-body: DWI (b-value 50/600/800), T2 HASTE, DWI (spine), T2 SPAIR, T1w FLASH, T2 HASTE, contrast-enhanced T1 VIBE. 2 + 2 nr na
Ergul et al. [22] PET/CT, DWI 1.5T na No, yes Breast: T1, T2, TIRM, DWI (b-value 50/400/ 800), contrast-enhanced 3D T1 fat suppressed. 2 + 1 No na
Catalano et al. [21] PET/CT, DWI nr Simultaneous Yes, yes Whole-body: DWI (b-value 50/400/800), STIR, T1 Dixon, T2 HASTE, contrast-enhanced T1 fat suppressed VIBE. 2 nr 6 weeks
Grueneisen et al. [24] PET/CT, PET/MRI 3.0T Simultaneous Yes, yes Whole-body (prone): T2 TIRM, T2 FSE, DWI (b-value 0/500/1000), contrast-enhanced T1 FLASH. 2 Yes 2 weeks
Sawicki et al. [29] PET/CT, PET/MRI 3.0T Simultaneous Yes, yes Whole-body: 3D Dixon VIBE, T2 HASTE, DWI (b-value 0/500/1000), T2 TIRM, contrast-enhanced T1 fat suppressed VIBE. 2 Yes 4 weeks
Melsaether et al. [26] PET/CT, PET/MRI 3.0T Simultaneous No, yes Whole-body (supine): Dixon, contrast-enhanced 3D T1, DWI (b-value 0/350/700). 2 + 2 Yes na
Pujara et al. [27] PET/CT, PET/MRI 3.0T Simultaneous No, yes Whole-body (prone): 3D VIBE for MRAC, 3D T1 VIBE, DWI (b-value 0/350/700). 1 + 1 nr na
Catalano et al. [20] PET/CT, PET/MRI, DWI nr Simultaneous Yes, yes Whole-body: DWI (b-value 50/400/800), STIR, T1 Dixon, T2 HASTE, contrast-enhanced T1 fat suppressed VIBE. 2 Yes 6 weeks
Botsikas et al. [19] PET/CT, PET/MRI 3.0T Sequential No, yes Whole-body (supine): T2 FSE, DWI (b-value 0/1000), 3D T1 Dixon, 3D T1 FFE. Breast (prone): T2 FSE, DWI (b-value 0/1000), contrast-enhanced 3D T1 Dixon, 3D T1 FFE. 2 Yes 3 months
Rezk et al. [28] PET/CT, DWI 1.5T na Yes, no Whole-body: T1 TSE, STIR, DWI (b-value 0/1000). 1 nr na
a, delay in the case of same reviewer of both imaging modalities; b-value (s/mm2); DWI, diffusion-weighted imaging; FFE, fast-field echo; FLASH, fast low-angle shot; FSE, fast spin echo; HASTE, half-Fourier acquisition single-shot turbo-spin echo; MRAC, MRI-based attenuation correction; na, not applicable, nr, not reported; SPAIR, spectral selection attenuated inversion recovery; STIR, short tau inversion recovery; TIRM, turbo inversion recovery magnitude; VIBE, volumetric interpolated breath-hold examination.

PET/computed tomography versus PET/MRI

18F-FDG PET/CT was compared with 18F-FDG PET/MRI in seven studies comprising a total of 396 patients. Evaluated anatomic sites were all lesions, contralateral breast cancer, axillary, internal mammary, and mediastinal lymph nodes, bone, liver, lung, pleura, and brain. A tabular overview of the results is depicted in Table 4.

Table 4 - PET/computed tomography versus PET/MRI
PET/CT PET/MRI
First author Site Analysis Sensitivity (%; 95 CIs) Specificity (%; 95 CIs) Accuracy (%; 95 CIs) Sensitivity (%; 95 CIs) Specificity (%; 95 CIs) Accuracy (%; 95 CIs)
Sawicki All L 96 89 95 100 89 99
Sawicki All P 100 100
Melsaether All P 97A,B (88–99) 77A–82B (64–91) 100A,B (92–100) 86C–90D (76–95)
Catalano Alla P 75 98b
Botsikas All L 77 (67–85) 98 (97–99)c 89 (81–94)d 96 (94–98)
Botsikas Alle P 69 (39–90) 100 (93–100) 85 (54–97) 97 (89–99)
Grueneisen Axillary LNs P 78 (52–94) 94 (79–99) 88 78 (52–94) 90 (74–98) 86
Melsaether Axillary LNs P 88A,B (64–99) 95A,B (88–98) 88D–100C (69–97) 95C,D (88–98)
Botsikas Axillary LNs L 81 (67–90) 92 (85–96) 85 (72–93) 89 (82–94)
Botsikas Axillary LNs P 83 (68–91) 76 (58–89) 87 (73–95) 68 (49–82)
Botsikas Internal mammary LNs L 90 (54–99) 100 (97–100) 90 (54–99) 100 (97–100)
Botsikas Internal mammary LNs P 89 (51–99) 100 (94–100) 89 (51–99) 100 (94–100)
Botsikas Mediastinal LNs L 100 (52–100) 100 (94–100) 100 (52–100) 100 (94–100)
Botsikas Mediastinal LNs P 100 (52–100) 100 (94–100) 100 (52–100) 100 (94–100)
Catalano Bone P 85 (70–96) 96 (87–100) 99 (96–100)
Melsaether Bone L 87B–99A (79–100) 95D–98C (90–100) †
Melsaether Bone P 96A,B (84–100) 97A–100B (90–100) 100C,D (91–100) 100C,D (93–100)
Pujara Bone P 94 100
Botsikas Bone L 69 (48–85) 100 (97–100) ‡ 92 (73–99) † 95 (90–98)
Botsikas Bone P 67 (31–91) 100 (94–100) 89 (51–99) 97 (89–99)
Melsaether Liver L 70B–75A (54–87) 80D–100C (66–100) †
Melsaether Liver P 73A,B (50–88) 100A,B (95–100) 91D–100C (78–100) 98C–100D (92–100)
Pujara Liver P 86 100
Melsaether Distant LNs L 85B–95A (72–99) 92D–95C (80–99)
Melsaether Distant LNs P 82A–91B (66–95) 95B–98A (88–100) 91D–100C (78–100) 98C,D (90–100)
Pujara Distant LNs P 100 100
Botsikas Contralateral BC L 25 (1–78) 99 (92–100) 100 (40–100) 99 (92–100)
Botsikas Contralateral BC P 33 (2–87) 99 (92–100) 100 (31–100) 99 (92–100)
Melsaether Lung L 96B–100A (79–100) 74D–87C (53–96)
Melsaether Lung P 100A,B (74–100) 80A–82B (71–90) 83D–100C (62–100) 89C–91D (81–96) ‡
Melsaether Pleura L 100A,B (80–100) 100A,B (80–100)
Melsaether Pleura P 100A,B (74–100) 100A,B (95–100) 100C,D (74–100) 100C,D (95–100)
Melsaether Brain L 93D–100C (70–100)
A, reader 3 (PET/CT); B, reader 4 (PET/CT); C, reader 1 (PET/MRI); CI, confidence interval; CT, computed tomography; D, reader 2 (PET/MRI); L, lesion level; LNs, lymph nodes; P, patient level.
aStaging performance.
bSignificantly higher accuracy.
cSignificantly higher specificity.
dSignificantly higher sensitivity.
ePatients with bone, liver, pulmonary, mediastinal, pleural and thoracic wall metastases.

All lesions were evaluated in four primary studies [19,20,26,29]. In all four studies, the sensitivity of 18F-FDG PET/MRI was at least equal to that of 18F-FDG PET/CT, with only Botsikas et al. [19] reporting a statistically significant difference in favor of 18F-FDG PET/MRI (77% versus 89%, P = 0.0013). Catalano et al. found a significantly higher accuracy for staging of breast cancer patients in favor of 18F-FDG PET/MRI (75% versus 98%, P = 0.005). Contrasting results were found regarding specificity, for which only Botsikas et al. reported a statistically significant difference in favor of 18F-FDG PET/CT (98% versus 96%, P = 0.0075). Three studies reported on the diagnostic performance regarding axillary lymph node staging and none of these studies reported a statistical difference between 18F-FDG PET/CT and 18F-FDG PET/MRI [19,24,26]. In general, there is a trend toward a higher sensitivity for 18F-FDG PET/MRI and a higher specificity for 18F-FDG PET/CT. Four studies compared the diagnostic performance for the detection of bone metastases between 18F-FDG PET/CT and 18F-FDG PET/MRI [19,21,26,27]. Botsikas et al. found a higher sensitivity for 18F-FDG PET/MRI (69% versus 92%, P = 0.0034) and a higher specificity for 18F-FDG PET/CT (100% versus 95%, P = 0.0081) in the lesion detection of bone metastases. Melsaether et al. [26] demonstrated higher sensitivity in a lesion-based analysis in favor of 18F-FDG PET/MRI (99% for reader 3 and 87% for reader 4 versus 98% for reader 1 and 95% for reader 2, P = 0.012). However, the patient-based analysis in this same publication did not reveal a statistically significant difference. Moreover, Catalano et al. demonstrated that 18F-FDG PET/CT revealed a statistically lower amount of osseous lesions compared to 18F-FDG PET/MRI (90 versus 141, P < 0.001). Regarding the detection of metastases in distant lymph nodes, two studies reported similar diagnostic performance between 18F-FDG PET/CT and 18F-FDG PET/MRI [26,27]. Regarding hepatic metastases, Melsaether et al. reported higher sensitivity in favor of 18F-FDG PET/MRI on a lesion-based analysis (75% for reader 3 and 70% for reader 4 versus 100% for reader 1 and 80% for reader 2, P < 0.001), but similar performance in a patient-based analysis (73% for readers 3 and 4 versus 100% for reader 1 and 91% for reader 2, P = 0.095). Pujara et al. [27] found lower sensitivity for PET/CT in a patient-based analysis (86% versus 100%) but did not review the data statistically. Only Melsaether et al. compared the diagnostic accuracy of 18F-FDG PET/CT and 18F-FDG PET/MRI for the detection of lung metastases and did so both lesion- and patient-based. While there was a trend toward superiority for 18F-FDG PET/CT, no statistically significant difference was found regarding sensitivity both in a lesion-based (100% for reader 3 and 96% for reader 4 versus 87% for reader 1 and 74% for reader 2, P = 0.065) and in a patient-based analysis (100% for readers 3 and 4 versus 100% for reader 1 and 83% for reader 2, P > 0.99). Contrarily, 18F-FDG PET/MRI was significantly superior regarding patient-based specificity (80% for reader 3 and 82% for reader 4 versus 89% for reader 1 and 99% for reader 2, P = 0.008).

Contralateral breast cancer, internal mammary lymph nodes, mediastinal lymph nodes, pleural metastases, and brain metastases were compared in one study each [19,26].

PET/computed tomography versus diffusion-weighted imaging

18F-FDG PET/CT was compared with DWI in four studies comprising a total of 131 patients [20,22,25,28]. DWI was also compared to 18F-FDG PET/CT as part of the MRI protocol in all seven studies evaluating the diagnostic performance of 18F-FDG PET/MRI [19–21,24,26,27,29]. Evaluated sites were all lesions, locoregional and distant lymph nodes, bone, lung, liver, and other sites not specified. None of the reviewed studies reported a significant difference between 18F-FDG PET/CT and DWI for any of the evaluated sites. A tabular overview of the results is depicted in Table 5.

Table 5 - PET/computed tomography versus diffusion-weighted imaging
PET/CT DWI
First author Site Analysis Sensitivity (%; 95 CIs) Specificity (%; 95 CIs) Accuracy (%; 95 CIs) Sensitivity (%; 95 CIs) Specificity (%; 95 CIs) Accuracy (%; 95 CIs)
Heusner All L 95 99 98 86 67 71
Heusner All C 94 99 98 91 72 76
Catalano All P 75 84
Rezk All L 85 86 85 82 78 81
Ergul Axillary LNs P 67 89 75 40 100 63
Heusner Regional LNs P 75 100 93 75 64 67
Rezk Regional LNs L 90 82 84 73
Grankvist Bone L 100 100 67A–70B 40B–95A
Heusner Bone P 100 100 100 86 8 35
Heusner Distant LNs P 100 100 100 100 0 30
Rezk Distant LNs L 86 92 83 80
Heusner Lung P 100 100 100 100 100 100
Heusner Liver P 100 100 100 100 82 85
Rezk Distant lesions L 84 84 80 81
Heusner Other organs P 100 94 95 66 94 89
Heusner Organs C 100 98 99 87 75 77
A, STIR and DWI with and without T1; B, DWI only; C, compartment level; CI, confidence interval; CT, computed tomography; DWI, diffusion-weighted imaging; L, lesion level; LNs, lymph nodes; P, patient level; STIR, short tau inversion recovery.

Three studies evaluated the diagnostic performance regarding all metastases [20,25,28]. In the study by Heusner et al., higher accuracy for PET/CT was found both in a lesion-based analysis (98% versus 71%) as well as in a compartment-based analysis (98% versus 76%), and Rezk et al. found similar results in a lesion-based analysis (85% versus 81%, P = 0.66) [25,28]. On the contrary, in the study by Catalano et al., DWI achieved higher accuracy in a patient-based analysis (75% versus 84%, P = 0.27) [20]. Three studies investigated (loco)regional lymph nodes and all studies showed that better diagnostic performance was achieved with 18F-FDG PET/CT [22,25,28]. In a study by Heusner et al., similar sensitivity for the detection of regional lymph node metastases was found (75% versus 75%), while a higher specificity was achieved by 18F-FDG PET/CT (100% versus 64%) [25]. Rezk et al. found better diagnostic performance for 18F-FDG PET/CT regarding sensitivity (90% versus 84%, P = 0.59) and specificity (82% versus 73%, P = 0.47) in a similar analysis [28]. Regarding axillary lymph node metastases, higher accuracy was achieved by 18F-FDG PET/CT (75% versus 63%) in one study [22]. Two studies evaluated metastases in the bone and two in distant lymph nodes; for all analyses, 18F-FDG PET/CT achieved better diagnostic performance than DWI [25,28]. The lung, liver, distant lesions, other organs not specified, and lymphatic system were evaluated by one study each [25,28].

Discussion

Based on the results summarized in this systematic review, 18F-FDG PET/MRI has demonstrated to achieve similar diagnostic performance to 18F-FDG PET/CT with no statistically significant differences in nodal staging and an even higher accuracy in overall distant staging of breast cancer patients. In general, there is a trend toward higher sensitivity and lower specificity for 18F-FDG PET/MRI when compared to 18F-FDG PET/CT. Results on the diagnostic performance of DWI only are conflicting if used separately. It tends to have a complementary value when combined with other MR sequences.

Combining the evidence from the primary studies, it can be concluded that 18F-FDG PET/MRI and 18F-FDG PET/CT have similar diagnostic performance regarding lesion detection with a slight tendency to improved sensitivity for 18F-FDG PET/MRI and higher specificity for 18F-FDG PET/CT [19,20,26,29]. Pace et al. [30] previously concluded that 18F-FDG PET/MR performs equally to 18F-FDG PET/CT in the anatomic allocation and multiple studies show a strong correlation between the standardized uptake values (SUVs) derived from these two hybrid techniques [27]. In line with these findings, the main reason for the higher sensitivity of 18F-FDG PET/MR comes from non-FDG-avid lesions such as permeative osseous metastases and subcentimeter hepatic metastases that are visible on MRI but not on CT [20,29]. Based on the primary articles included in this review, the diagnostic performance of DWI for lesion detection and staging is inconclusive. Of 283 lesions for which a standard of reference was available and that were only seen on DWI and not on 18F-FDG PET/CT in a study by Heusner et al., 231 (82%) were false-positives mainly in the bone and distant lymph nodes [25]. They agree with previous publications stating that DWI only causes false alarm [31].

The determination of the correct clinical nodal status is essential to provide the patient with the appropriate oncological treatment and strongly influences patients’ prognosis [2]. Compared to a sensitivity of 60% and specificity of 97% of 18F-FDG PET/CT reported in a systematic review by Robertson et al. [32], the primary studies in this review achieve on average a higher sensitivity and lower specificity. From another systematic review by Liang et al. [33], it can be concluded that MRI shows superiority over 18F-FDG PET/CT in assessing nodal status with fewer false positive and false negative results. In addition, a recent prospective study by Bruckmann et al. [34] showed that the diagnostic accuracy is further improved by combining 18F-FDG PET with MRI, with a significantly higher diagnostic confidence in lesion characterization. The findings summarized in the current systematic review show that 18F-FDG PET/MRI tends to be noninferior to 18F-FDG PET/CT. Moreover, the addition of an axillary dedicated hybrid protocol as investigated by Van Nijnatten et al. [35] could even further improve the diagnostic performance of 18F-FDG PET/MRI in determining the clinical nodal status. Opposite to the results found by Chung et al. [15], Ergul et al. [22] found a strikingly low sensitivity and high specificity for DWI compared to 18F-FDG PET/CT in early stage breast cancer patients. They state that the low tumor cell burden in their cohort likely rendered DWI-mediated detection difficult [22]. In summary, regarding nodal staging, 18F-FDG PET/CT could safely be replaced by 18F-FDG PET/MRI, while DWI alone shows conflicting results.

The bone is the predominant site of metastatic dissemination in breast cancer patients and occurs in 69% of patients with advanced disease [36–38]. Remarkably more osseous lesions are detected by 18F-FDG PET/MRI compared to 18F-FDG PET/CT and it seems that the majority of the missed lesions on 18F-FDG PET/CT are permeative in nature and lack FDG avidity [19,21,26,39]. Bone marrow replacement in non-FDG-avid lesions revealed by T1-weighted MRI sequences can lead to a positive lesion on PET/MRI, while the low conspicuity due to low-intrinsic tissue contrast on CT would render these lesions negative [19,21]. Moreover, enhancement after gadolinium administration could further increase the diagnostic confidence [21]. The results summarized in this review show a lesion-level preference for 18F-FDG PET/MRI with at least noninferiority on a patient-level [19,21,26,27]. While Grankvist et al. [23] demonstrated that DWI combined with other sequences yields a high specificity and improves diagnostic efficacy, many of the lesions suspicious on DWI were false-positive and Heusner et al. [25] clearly stated that DWI alone generates a lot of discordant and false-positive findings and in its current form it is unsuitable for the detection of osseous metastases [23].

A widespread concern that has kept clinicians hesitant to the use of 18F-FDG PET/MRI as an alternative for 18F-FDG PET/CT for whole-body staging in breast cancer is the diagnostic performance of MRI in the detection of small non-FDG-avid pulmonary lesions [40–43]. Similar to another publication, Melsaether et al. found a higher sensitivity for 18F-FDG PET/CT in pulmonary lesion detection, while this difference could not be established on a patient-level [26,44]. However, a recent study by Martin et al. in a cohort of 1003 cancer patients found that the amount of missed malignant lung lesions was negligibly low (0.8%), and other publications state that the vast majority of small non-FDG-avid pulmonary lesions missed on 18F-FDG PET/MRI remain stable on follow-up and are likely to be benign [41,45,46]. Moreover, 18F-FDG PET/MRI is associated with higher diagnostic confidence likely due to the high soft-tissue contrast of MR compared with CT and especially in large lung lesions the higher intrinsic contrast of MR imaging may provide better conspicuity [40]. These recent findings combined with promising new MRI sequences, such as the ultrashort echo time sequence that is expected to improve pulmonary lesion detection, suggest that 18F-FDG PET/MRI can safely replace 18F-FDG PET/CT regarding the detection of pulmonary lesions [47].

Together with bone and lung, the liver is the most common site of metastatic disease in breast cancer patients. A recent systematic review and meta-analysis by Hong et al. investigated the diagnostic performance of 18F-FDG PET/MRI for hepatic metastases in a diverse group of patients with primary malignancy. Similar to the results in this systematic review, they found very good diagnostic accuracy for the detection of hepatic metastases with a patient-level meta-analytic summery sensitivity of 99.2% and specificity of 98.6% [11,26,48]. Especially in subcentimeter hepatic metastases, the reported sensitivity of 18F-FDG PET/MRI outperforms that achieved by 18F-FDG PET/CT [48]. Because two early publications reported on the usefulness of DWI for the detection of small (≤10 mm) hepatic lesions, the presence of a DWI sequence in MRI protocols could have contributed to the high sensitivity achieved by 18F-FDG PET/MRI [49,50].

The results summarized in this publication should be taken with caution and the large degree of heterogeneity between and within studies causes several limitations. First, a heterogeneous cohort of patients was used varying from primary staging to restaging of patients suspicious for recurrence. Second, different imaging sequences and radiofrequency coils were used. Third, in the majority of studies, 18F-FDG PET/MRI and 18F-FDG PET/CT were performed on the same day on the same dose of 18F-FDG. Differences in acquisition time influence the tumor-to-background ratio and tumor-to-nontumor ratios which have an effect on tumor visibility and possibly on diagnostic performance [51,52]. Multiple studies suggest that delayed acquisition could have a beneficial effect on the sensitivity; therefore, the order of imaging should be taken into account when comparing the diagnostic performance of 18F-FDG PET/CT with 18F-FDG PET/MRI [51–54]. Fourth, in some studies, intravenous contrast was not used in both techniques. For example, in three of the included studies, patients underwent unenhanced whole-body 18F-FDG PET/CT and they were compared with enhanced 18F-FDG PET/MRI, so the results should be interpreted with this important limitation in mind [19,26,27]. Last, due to the high heterogeneity between and within studies and the inability to compute 2 × 2 contingency tables of at least three studies for two or more imaging modalities, no meta-analysis was performed.

18F-FDG PET/MRI is a promising imaging tool for performing accurate nodal and distant staging of breast cancer patients with lower radiation exposure compared to 18F-FDG PET/CT. Melsaether et al. reported an average dose reduction of 50% when using 18F-FDG PET/MRI instead of 18F-FDG PET/CT [26]. Other advantages of 18F-FDG PET/MRI compared to 18F-FDG PET/CT are the soft tissue contrast and motion correction possibilities [55]. SUVs could be valuable in determining prognosis of breast cancer patients and it has been evaluated in both imaging modalities by Pujara et al. who found a strong correlation of the maximum SUV acquired by 18F-FDG PET/MRI compared to 18F-FDG PET/CT [27,56]. Pace et al. compared SUVs for nodal and distant metastases and found that both the maximum SUV and mean SUV were significantly higher on 18F-FDG PET/MRI than on 18F-FDG PET/CT [57]. Van Nijnatten et al. [35] reported comparable values achieved on hybrid PET/MRI for the primary tumor and the most FDG-avid lymph node. PET/MRI, a multimodal and multiparametric technique, may provide more quantitative information than PET/CT [58]. Even though the per-examination cost of 18F-FDG PET/MRI is approximately 50% higher, the possibility of a one-stop-shop solution for, for example, node-positive breast cancer patients could not only reduce costs in initial staging but could also reduce the burden on the patient [59].

To conclude, the relatively new imaging modality 18F-FDG PET/MRI has demonstrated to be noninferior to 18F-FDG PET/CT regarding nodal staging. For distant staging, a trend toward a higher sensitivity in lesion detection for 18F-FDG PET/MRI and a higher specificity for 18F-FDG PET/CT can be observed with the former detecting a significantly higher amount of osseous metastases, which are the most common type of distant metastases in breast cancer. Results on the diagnostic performance of DWI are conflicting, and it tends to have a complementary value when combined with other MR sequences rather than evaluating it separately. We deem 18F-FDG PET/MRI to be accurate enough in nodal staging and distant staging to be able to replace 18F-FDG PET/CT in breast cancer staging, hereby forming an attractive one-stop-shop solution for breast cancer patients for whom both an MRI as well as a PET-CT is indicated.

Acknowledgements

Conflicts of interest

C.M.d.M. received a salary from Health Foundation Limburg. M.L.S. received research funding from Servier for the microbiota study. There are no conflicts of interest for the remaining authors.

References

1. Bray F, Ferlay J, Soerjomataram I, Siegel RL, Torre LA, Jemal A. Global cancer statistics 2018: GLOBOCAN estimates of incidence and mortality worldwide for 36 cancers in 185 countries. CA Cancer J Clin. 2018; 68:394–424
2. Noone AM HN KM, Miller D, Brest A, Yu M, Ruhl J, Tatalovich Z, Mariotto A, et al. SEER Cancer Statistics Review, 1975-2015. 2018, Bethesda, MD: National Cancer Institute. https://seer.cancer.gov/csr/1975_2015/. [Accessed 1 April 2020].
3. Soni A, Ren Z, Hameed O, Chanda D, Morgan CJ, Siegal GP, Wei S. Breast cancer subtypes predispose the site of distant metastases. Am J Clin Pathol. 2015; 143:471–478
4. Thomas A, Khan SA, Chrischilles EA, Schroeder MC. Initial surgery and survival in stage IV breast cancer in the United States, 1988-2011. JAMA Surg. 2016; 151:424–431
5. Weigelt B, Peterse JL, van ‘t Veer LJ. Breast cancer metastasis: markers and models. Nat Rev Cancer. 2005; 5:591–602
6. Jafari SH, Saadatpour Z, Salmaninejad A, Momeni F, Mokhtari M, Nahand JS, et al. Breast cancer diagnosis: imaging techniques and biochemical markers. J Cell Physiol. 2018; 233:5200–5213
7. Fowler AM, Mankoff DA, Joe BN. Imaging neoadjuvant therapy response in breast cancer. Radiology. 2017; 285:358–375
8. Kuijs VJ, Moossdorff M, Schipper RJ, Beets-Tan RG, Heuts EM, Keymeulen KB, et al. The role of MRI in axillary lymph node imaging in breast cancer patients: a systematic review. Insights Imaging. 2015; 6:203–215
9. Caresia Aroztegui AP, García Vicente AM, Alvarez Ruiz S, Delgado Bolton RC, Orcajo Rincon J, Garcia Garzon JR, et al.; Oncology Task Force of the Spanish Society of Nuclear Medicine and Molecular Imaging. 18F-FDG PET/CT in breast cancer: evidence-based recommendations in initial staging. Tumour Biol. 2017; 39:1010428317728285
10. Musafargani S, Ghosh KK, Mishra S, Mahalakshmi P, Padmanabhan P, Gulyás B. PET/MRI: a frontier in era of complementary hybrid imaging. Eur J Hybrid Imaging. 2018; 2:12
11. Pujara AC, Kim E, Axelrod D, Melsaether AN. PET/MRI in breast cancer. J Magn Reson Imaging. 2019; 49:328–342
12. Rosenkrantz AB, Friedman K, Chandarana H, Melsaether A, Moy L, Ding YS, et al. Current status of hybrid PET/MRI in oncologic imaging. AJR Am J Roentgenol. 2016; 206:162–172
13. Kitajima K, Miyoshi Y, Yamano T, Odawara S, Higuchi T, Yamakado K. Prognostic value of FDG-PET and DWI in breast cancer. Ann Nucl Med. 2018; 32:44–53
14. Koh DM, Collins DJ. Diffusion-weighted MRI in the body: applications and challenges in oncology. AJR Am J Roentgenol. 2007; 188:1622–1635
15. Chung J, Youk JH, Kim JA, Gweon HM, Kim EK, Ryu YH, Son EJ. Role of diffusion-weighted MRI: predicting axillary lymph node metastases in breast cancer. Acta Radiol. 2014; 55:909–916
16. Komori T, Narabayashi I, Matsumura K, Matsuki M, Akagi H, Ogura Y, et al. 2-[Fluorine-18]-fluoro-2-deoxy-D-glucose positron emission tomography/computed tomography versus whole-body diffusion-weighted MRI for detection of malignant lesions: initial experience. Ann Nucl Med. 2007; 21:209–215
17. McInnes MDF, Moher D, Thombs BD, McGrath TA, Bossuyt PM, Clifford T, et al.; The PRISMA-DTA Group. Preferred reporting items for a systematic review and meta-analysis of diagnostic test accuracy studies: the PRISMA-DTA Statement. JAMA. 2018; 319:388–396
18. Whiting PF, Rutjes AW, Westwood ME, Mallett S, Deeks JJ, Reitsma JB, et al.; QUADAS-2 Group. QUADAS-2: a revised tool for the quality assessment of diagnostic accuracy studies. Ann Intern Med. 2011; 155:529–536
19. Botsikas D, Bagetakos I, Picarra M, Da Cunha Afonso Barisits AC, Boudabbous S, Montet X, et al. What is the diagnostic performance of 18-FDG-PET/MR compared to PET/CT for the N- and M- staging of breast cancer? Eur Radiol. 2019; 29:1787–1798
20. Catalano OA, Daye D, Signore A, Iannace C, Vangel M, Luongo A, et al. Staging performance of whole-body DWI, PET/CT and PET/MRI in invasive ductal carcinoma of the breast. Int J Oncol. 2017; 51:281–288
21. Catalano OA, Nicolai E, Rosen BR, Luongo A, Catalano M, Iannace C, et al. Comparison of CE-FDG-PET/CT with CE-FDG-PET/MR in the evaluation of osseous metastases in breast cancer patients. Br J Cancer. 2015; 112:1452–1460
22. Ergul N, Kadioglu H, Yildiz S, Yucel SB, Gucin Z, Erdogan EB, et al. Assessment of multifocality and axillary nodal involvement in early-stage breast cancer patients using 18F-FDG PET/CT compared to contrast-enhanced and diffusion-weighted magnetic resonance imaging and sentinel node biopsy. Acta Radiol. 2015; 56:917–923
23. Grankvist J, Fisker R, Iyer V, Fründ ET, Simonsen C, Christensen T, et al. MRI and PET/CT of patients with bone metastases from breast carcinoma. Eur J Radiol. 2012; 81:e13–e18
24. Grueneisen J, Nagarajah J, Buchbender C, Hoffmann O, Schaarschmidt BM, Poeppel T, et al. Positron emission tomography/magnetic resonance imaging for local tumor staging in patients with primary breast cancer: a comparison with positron emission tomography/computed tomography and magnetic resonance imaging. Invest Radiol. 2015; 50:505–513
25. Heusner TA, Kuemmel S, Koeninger A, Hamami ME, Hahn S, Quinsten A, et al. Diagnostic value of diffusion-weighted magnetic resonance imaging (DWI) compared to FDG PET/CT for whole-body breast cancer staging. Eur J Nucl Med Mol Imaging. 2010; 37:1077–1086
26. Melsaether AN, Raad RA, Pujara AC, Ponzo FD, Pysarenko KM, Jhaveri K, et al. Comparison of whole-body (18)F FDG PET/MR imaging and whole-body (18)F FDG PET/CT in terms of lesion detection and radiation dose in patients with breast cancer. Radiology. 2016; 281:193–202
27. Pujara AC, Raad RA, Ponzo F, Wassong C, Babb JS, Moy L, Melsaether AN. Standardized uptake values from PET/MRI in metastatic breast cancer: an organ-based comparison with PET/CT. Breast J. 2016; 22:264–273
28. Rezk M, Nasr I, Ali I, Abdelhamed H. Comparative study between (18)F FDG-PET/CT and whole body MRI DWIBS in assessment of recurrent breast cancer (prospective, comparative, cross-sectional study design). Indian J Nucl Med. 2019; 34:1–9
29. Sawicki LM, Grueneisen J, Schaarschmidt BM, Buchbender C, Nagarajah J, Umutlu L, et al. Evaluation of ¹8F-FDG PET/MRI, ¹8F-FDG PET/CT, MRI, and CT in whole-body staging of recurrent breast cancer. Eur J Radiol. 2016; 85:459–465
30. Pace L, Nicolai E, Luongo A, Aiello M, Catalano OA, Soricelli A, Salvatore M. Comparison of whole-body PET/CT and PET/MRI in breast cancer patients: lesion detection and quantitation of 18F-deoxyglucose uptake in lesions and in normal organ tissues. Eur J Radiol. 2014; 83:289–296
31. Ohno Y, Koyama H, Onishi Y, Takenaka D, Nogami M, Yoshikawa T, et al. Non-small cell lung cancer: whole-body MR examination for M-stage assessment–utility for whole-body diffusion-weighted imaging compared with integrated FDG PET/CT. Radiology. 2008; 248:643–654
32. Robertson IJ, Hand F, Kell MR. FDG-PET/CT in the staging of local/regional metastases in breast cancer. Breast. 2011; 20:491–494
33. Liang X, Yu J, Wen B, Xie J, Cai Q, Yang Q. MRI and FDG-PET/CT based assessment of axillary lymph node metastasis in early breast cancer: a meta-analysis. Clin Radiol. 2017; 72:295–301
34. Bruckmann NM, Sawicki LM, Kirchner J, Martin O, Umutlu L, Herrmann K, et al. Prospective evaluation of whole-body MRI and (18)F-FDG PET/MRI in N and M staging of primary breast cancer patients. Eur J Nucl Med Mol Imaging. 2020
35. van Nijnatten TJA, Goorts B, Vöö S, de Boer M, Kooreman LFS, Heuts EM, et al. Added value of dedicated axillary hybrid 18F-FDG PET/MRI for improved axillary nodal staging in clinically node-positive breast cancer patients: a feasibility study. Eur J Nucl Med Mol Imaging. 2018; 45:179–186
36. Costelloe CM, Rohren EM, Madewell JE, Hamaoka T, Theriault RL, Yu TK, et al. Imaging bone metastases in breast cancer: techniques and recommendations for diagnosis. Lancet Oncol. 2009; 10:606–614
37. Coleman RE, Rubens RD. The clinical course of bone metastases from breast cancer. Br J Cancer. 1987; 55:61–66
38. Brook N, Brook E, Dharmarajan A, Dass CR, Chan A. Breast cancer bone metastases: pathogenesis and therapeutic targets. Int J Biochem Cell Biol. 2018; 96:63–78
39. Çelebi F. What is the diagnostic performance of 18F-FDG-PET/MRI in the detection of bone metastasis in patients with breast cancer? Eur J Breast Health. 2019; 15:213–216
40. Huellner MW, Appenzeller P, Kuhn FP, Husmann L, Pietsch CM, Burger IA, et al. Whole-body nonenhanced PET/MR versus PET/CT in the staging and restaging of cancers: preliminary observations. Radiology. 2014; 273:859–869
41. Raad RA, Friedman KP, Heacock L, Ponzo F, Melsaether A, Chandarana H. Outcome of small lung nodules missed on hybrid PET/MRI in patients with primary malignancy. J Magn Reson Imaging. 2016; 43:504–511
42. Sawicki LM, Grueneisen J, Buchbender C, Schaarschmidt BM, Gomez B, Ruhlmann V, et al. Comparative performance of ¹8F-FDG PET/MRI and ¹8F-FDG PET/CT in detection and characterization of pulmonary lesions in 121 oncologic patients. J Nucl Med. 2016; 57:582–586
43. Chandarana H, Heacock L, Rakheja R, DeMello LR, Bonavita J, Block TK, et al. Pulmonary nodules in patients with primary malignancy: comparison of hybrid PET/MR and PET/CT imaging. Radiology. 2013; 268:874–881
44. Stolzmann P, Veit-Haibach P, Chuck N, Rossi C, Frauenfelder T, Alkadhi H, et al. Detection rate, location, and size of pulmonary nodules in trimodality PET/CT-MR: comparison of low-dose CT and Dixon-based MR imaging. Invest Radiol. 2013; 48:241–246
45. Sawicki LM, Grueneisen J, Buchbender C, Schaarschmidt BM, Gomez B, Ruhlmann V, et al. Evaluation of the outcome of lung nodules missed on 18F-FDG PET/MRI compared with 18F-FDG PET/CT in patients with known malignancies. J Nucl Med. 2016; 57:15–20
46. Martin O, Schaarschmidt BM, Kirchner J, Suntharalingam S, Grueneisen J, Demircioglu A, et al. PET/MRI versus PET/CT in whole-body staging: results from a unicenter observational study in 1003 subsequent examinations. J Nucl Med. 2019
47. Burris NS, Johnson KM, Larson PE, Hope MD, Nagle SK, Behr SC, Hope TA. Detection of small pulmonary nodules with ultrashort echo time sequences in oncology patients by using a PET/MR system. Radiology. 2016; 278:239–246
48. Hong SB, Choi SH, Kim KW, Park SH, Kim SY, Lee SJ, et al. Diagnostic performance of [18F]FDG-PET/MRI for liver metastasis in patients with primary malignancy: a systematic review and meta-analysis. Eur Radiol. 2019; 29:3553–3563
49. Bruegel M, Gaa J, Waldt S, Woertler K, Holzapfel K, Kiefer B, Rummeny EJ. Diagnosis of hepatic metastasis: comparison of respiration-triggered diffusion-weighted echo-planar MRI and five t2-weighted turbo spin-echo sequences. AJR Am J Roentgenol. 2008; 191:1421–1429
50. Coenegrachts K, Delanote J, Ter Beek L, Haspeslagh M, Bipat S, Stoker J, et al. Improved focal liver lesion detection: comparison of single-shot diffusion-weighted echoplanar and single-shot T2 weighted turbo spin echo techniques. Br J Radiol. 2007; 80:524–531
51. Boerner AR, Weckesser M, Herzog H, Schmitz T, Audretsch W, Nitz U, et al. Optimal scan time for fluorine-18 fluorodeoxyglucose positron emission tomography in breast cancer. Eur J Nucl Med. 1999; 26:226–230
52. Wangerin KA, Muzi M, Peterson LM, Linden HM, Novakova A, O’Sullivan F, et al. Effect of 18F-FDG uptake time on lesion detectability in PET imaging of early stage breast cancer. Tomography. 2015; 1:53–60
53. Chen YM, Huang G, Sun XG, Liu JJ, Chen T, Shi YP, Wan LR. Optimizing delayed scan time for FDG PET: comparison of the early and late delayed scan. Nucl Med Commun. 2008; 29:425–430
54. Cheng G, Torigian DA, Zhuang H, Alavi A. When should we recommend use of dual time-point and delayed time-point imaging techniques in FDG PET? Eur J Nucl Med Mol Imaging. 2013; 40:779–787
55. Ehman EC, Johnson GB, Villanueva-Meyer JE, Cha S, Leynes AP, Larson PEZ, Hope TA. PET/MRI: where might it replace PET/CT? J Magn Reson Imaging. 2017; 46:1247–1262
56. Diao W, Tian F, Jia Z. The prognostic value of SUVmax measuring on primary lesion and ALN by 18F-FDG PET or PET/CT in patients with breast cancer. Eur J Radiol. 2018; 105:1–7
57. Pace L, Nicolai E, Luongo A, Aiello M, Catalano OA, Soricelli A, Salvatore M. Comparison of whole-body PET/CT and PET/MRI in breast cancer patients: lesion detection and quantitation of 18F-deoxyglucose uptake in lesions and in normal organ tissues. Eur J Radiol. 2014; 83:289–296
58. Wei L, Osman S, Hatt M, El Naqa I. Machine learning for radiomics-based multimodality and multiparametric modeling. Q J Nucl Med Mol Imaging. 2019; 63:323–338
59. Mayerhoefer ME, Prosch H, Beer L, Tamandl D, Beyer T, Hoeller C, et al. PET/MRI versus PET/CT in oncology: a prospective single-center study of 330 examinations focusing on implications for patient management and cost considerations. Eur J Nucl Med Mol Imaging. 2020; 47:51–60
Keywords:

breast cancer; diffusion-weighted imaging; distant staging; nodal staging; PET/computed tomography; positron emission tomography/magnetic resonance imaging

Supplemental Digital Content

Copyright © 2020 The Author(s). Published by Wolters Kluwer Health, Inc.