Abstract

BACKGROUND: 

Managing patients with hydrocephalus and cerebrospinal fluid (CSF) disorders requires repeated head imaging. In adults, it is typically computed tomography (CT) or less commonly magnetic resonance imaging (MRI). However, CT poses cumulative radiation risks and MRI is costly. Ultrasound is a radiation-free, relatively inexpensive, and optionally point-of-care alternative, but is prohibited by very limited windows through an intact skull.

OBJECTIVE: 

To describe our initial experience with transcutaneous transcranial ultrasound through sonolucent burr hole covers in postoperative hydrocephalus and CSF disorder patients.

METHODS: 

Using cohort study design, infection and revision rates were compared between patients who underwent sonolucent burr hole cover placement during new ventriculoperitoneal shunt placement and endoscopic third ventriculostomy over the 1-year study time period and controls from the period 1 year before. Postoperatively, trans-burr hole ultrasound was performed in the clinic, at bedside inpatient, and in the radiology suite to assess ventricular anatomy.

RESULTS: 

Thirty-seven patients with sonolucent burr hole cover were compared with 57 historical control patients. There was no statistically significant difference in infection rates between the sonolucent burr hole cover group (1/37, 2.7%) and the control group (0/57, P = .394). Revision rates were 13.5% vs 15.8% (P = 1.000), but no revisions were related to the burr hole or cranial hardware.

CONCLUSION: 

Trans-burr hole ultrasound is feasible for gross evaluation of ventricular caliber postoperatively in patients with sonolucent burr hole covers. There was no increase in infection rate or revision rate. This imaging technique may serve as an alternative to CT and MRI in the management of select patients with hydrocephalus and CSF disorders.