Comparison of adenoma detection rate between high-definition colonoscopes with different fields of view: 170 degrees versus 140 degrees : Medicine

Journal Logo

Research Article: Observational Study

Comparison of adenoma detection rate between high-definition colonoscopes with different fields of view: 170 degrees versus 140 degrees

Moon, Sang Yi MDa; Lee, Jong Yoon MDa,*; Lee, Jong Hoon MD, PhDa

Author Information
Medicine 102(2):p e32675, January 13, 2023. | DOI: 10.1097/MD.0000000000032675
  • Open

Abstract

1. Introduction

Colonoscopy is recommended for colorectal cancer (CRC) screening and surveillance.[1] Removing precancerous lesions by endoscopic resection is associated with lower incidence and mortality rates of CRC.[2,3] Despite its widespread use in CRC screening and surveillance, the technique remains suboptimal. Of the diagnosed CRCs, 1.8 to 9.0% are interval type of post-colonoscopy CRCs or interval CRC, which develops in the interval between a first screening colonoscopy and the subsequent surveillance colonoscopy.[4,5] Approximately half of interval CRCs are a result of missed lesions due to poor-quality colonoscopy.[6,7] The adenoma detection rate (ADR) is an important indicator for determining the quality of colonoscopy. It has been reported that the incidence of interval CRC after colonoscopy performed by endoscopists with low ADR is significantly higher than that after colonoscopy performed by endoscopists with high ADR.[8] Corley et al reported that every 1% increase in ADR lowers the rates of interval CRC by 3% and fatal interval CRC by 5%.[9] Therefore, medical societies recommend an average ADR of ≥25% in screening for high-quality colonoscopies.[10,11]

Recent advances in colonoscopy techniques have been aimed at increasing ADR. A withdrawal time of >6 minutes, adequate bowel preparation, and a high cecal insertion rate contributed to an increased ADR.[12–14] Advanced imaging techniques include high-resolution image quality, virtual chromoendoscopy, and wide fields of view (FOV).[15,16] Furthermore, colonoscopes with special equipment, such as Endocuff, EndoRings, and the third eye, are available to expose the additional colonic mucosa.[17–19] Full-spectrum endoscopy (FUSE) is a new endoscopy technique that allows a panoramic 330° view of the colon and improves colonoscopy quality indicators.[20] However, because these studies use expensive equipment that are not commonly used, the FOV of commonly used colonoscopes may not reveal the factors that influence the ADR. To the best of our knowledge, no study has investigated the difference in quality indicators, such as ADR, between colonoscopes with different FOVs with the same resolution, without using additional devices to expand the FOV. In commonly used newer-generation colonoscopes, the FOV varies slightly, between 170° and 140°, depending on the type of colonoscopy. The CV-290 EVIS LUCERA ELITE (CV-290) (Olympus Medical, Tokyo, Japan) and EPK-i7010 (Pentax, Hoya Corporation, Tokyo, Japan) have the same high-definition (HD) resolution with 1080 intercalating vertical pixels; however, CV-290 has a 170° FOV and EPK-i7010 has a 140° FOV. This study aimed to investigate whether quality indicators such as ADR differed between colonoscopes with a 170° and a 140° FOV with same HD resolution.

2. Methods

2.1. Study design and patients

We retrospectively analyzed the medical records of patients who underwent screening or surveillance colonoscopy at the Dong-A University Hospital in Busan, South Korea, between March 2021 and February 2022. All patients who underwent colonoscopy were aged >30 years and had undergone either the first screening colonoscopy or surveillance colonoscopy 3 to 5 years after the initial screening. Colonoscopies were performed by 4 experienced endoscopists with 3 either second- or third-year gastroenterology fellows who could independently perform colonoscopies. Only proper bowel preparations (Boston Bowel Preparation Scale score ≥2 in all segments) were included in this study. Patients with symptoms of active bleeding or a history of inflammatory bowel disease, CRC, polyposis syndrome, or colon surgery were excluded.

2.2. Endoscopic procedures

Dong-A University Hospital is a tertiary hospital that performs colonoscopy using 2 types of HD video processor systems: CV-290 and EPK-i7010. CV-290 and EPK-i7010 have the same HD resolution (1080 intercalating vertical pixels); however, CV-290 has a 170° FOV and EPK-i7010 has a 140° FOV. Patients could schedule a time and date for colonoscopy but could not select an endoscopy system. The patients underwent bowel cleansing with 1 to 2 L of a polyethylene glycol solution containing ascorbic acid and an additional 1 to 2 L of water. Sedation colonoscopy was performed only if the patient desired it, and midazolam (2–5 mg) and/or propofol (10–60 mg) were used as sedatives.

2.3. Endpoints

The primary endpoint was the comparison of ADR between the 2 groups with different FOVs. We defined ADR as the proportion of patients with at least 1 adenoma detected among all examined patients.

The secondary endpoints were polyp detection rate (PDR), sessile serrated PDR (SSPDR), advanced neoplasia detection rate (ANDR), and number of adenomas per colonoscopy (APC). We defined PDR as the proportion of patients with at least 1 polyp, including adenoma and hyperplastic polyp, detected in all patients examined. Advanced neoplasia was defined as any adenoma ≥10 mm with villous histology or high-grade dysplasia and any SSP ≥10 mm with dysplasia. The calculations for SSPDR and ANDR were similar to those for ADR and PDR. APC was defined as the mean number of adenomas detected per colonoscope.

2.4. Statistics and data analysis

Statistical analysis was performed using SPSS software (version 26.0; IBM, Armonk, NY). Statistical significance was set at P < .05. The data were divided by group: the 170° FOV group and the 140° FOV group. Continuous data are presented as mean ± standard deviation and were analyzed using Student t test. Categorical data were analyzed using Pearson chi-squared or Fisher exact tests. A logistical regression model was used for the multivariate analysis to assess the association between clinical factors and adenoma detection. Variables found to be significantly associated in the univariate analysis were used as covariates in the multivariate analysis.

2.5. Ethics statement

The Ethics Committee approved our research protocols through international agreements (World Medical Association Declaration of Helsinki: Ethical principles for medical research involving human subjects). Formal consent was not required for this study. The study protocol was reviewed and approved by the Institutional Review Board of the Dong-A University College of Medicine (DAUHIRB-22-096).

3. Results

3.1. Demographic and clinical characteristics

A total of 1711 patients were included in this study. Approximately 48% (n = 838) of the procedures were performed using colonoscopes with a 170° FOV, while 52% (n = 873) were performed using colonoscopes with a 140° FOV. The demographic and clinical characteristics of the patients are presented in Table 1. There were no significant differences between the 2 groups (170° vs 140° FOV) in terms of age, sex, bowel preparation, sedation, examination day, reason for examination, and endoscopist experience.

Table 1 - Demographic and clinical characteristics in patients who underwent colonoscopy.
Colonoscopes with fields of view
170° (n = 838) 140° (n = 873) P value
Age 62.20 ± 11.07 61.81 ± 12.26 .495
Gender .986
 Male 464 (55.4%) 483 (55.3%)
 Female 374 (44.6%) 390 (44.7%)
Bowel preparation .889
 BBPS 8,9 635 (75.8%) 659 (75.5%)
 BBPS 6,7 203 (24.2%) 214 (24.5%)
Sedation .947
 Yes 800 (95.5%) 834 (95.5%)
 No 38 (4.5%) 39 (4.5%)
Examination of d .803
 Morning 142 (16.9%) 144 (16.5%)
 Afternoon 696 (83.1%) 729 (83.5%)
Reason of examination .415
 Screening 629 (75.1%) 670 (76.7%)
 Surveillance 209 (24.9%) 203 (23.3%)
Endoscopist’s experience .806
 Experienced 527 (62.9%) 554 (63.5%)
 Trainee 311 (37.1%) 319 (36.5%)
BPPS = Boston bowel preparation score.

3.2. Colonoscopy quality indicators between the groups

There were no significant differences in cecal intubation time (421.23 ± 216.96 vs 416.09 ± 227.09; P = .630) or withdrawal time (741.25 ± 420.21 vs 743.37 ± 402.05; P = .915) between colonoscopes with a 170° FOV and a 140° FOV. There were also no significant differences in PDR (56.44 vs 53.49%; P = .220), ADR (43.79 vs 41.92%; P = .434), SSPDR (1.19 vs 0.92%; P = .304), ANDR (5.00 vs 4.58%; P = .735), and number of APC (0.88 ± 1.52 vs 0.89 ± 1.59; P = .987) between the 2 groups (Table 2).

Table 2 - Colonoscopic quality indicators between the groups.
Colonoscopes with fields of view
170° (n = 838) 140° (n = 873) P value
Intubation time (s) 421.23 ± 216.96 416.06 ± 27.06 .630
Withdrawal time (s) 741.25 ± 420.21 743.37 ± 402.05 .915
PDR (%) 56.44 (473/838) 53.49 (467/873) .220
ADR (%) 43.79 (367/838) 41.92 (366/873) .434
SSPDR (%) 1.19 (10/838) 0.92 (8/873) .575
ANDR (%) 5.00 (42/838) 4.58 (40/873) .735
APC 0.88 ± 1.52 0.89 ± 1.59 .987
ADR = adenoma detection rate, ANDR = advanced neoplasm detection rate, APC = adenoma per colonoscopy, PDR = polyp detection rate, SSPDR = sessile serrated polyp detection rate.

3.3. APC based on endoscopic feature, location, and size

Adenomas were classified based on their endoscopic features, their location, and their size. Based on their endoscopic features, adenomas were classified into 2 groups, polypoid and flat. The number of APC, according to their endoscopic features, was not significantly different between colonoscopes with 170° FOV to those with 140° FOV (polypoid: 0.05 ± 0.22 vs 0.05 ± 0.23, P = .822; flat: 0.83 ± 1.44 vs 0.83 ± 1.50, P = .986). Similarly, adenomas based on their location, were classified into 2 groups: proximal and distal. The number of APC, according to their location also did not differ significantly between colonoscopes with 170° FOV and those with 140° FOV (proximal: 0.62 ± 1.22 vs 0.62 ± 1.27, P = .928; distal: 0.26 ± 0.62 vs 0.27 ± 0.68, P = .833). Lastly, adenomas, were classified into 3 groups, based on their size: <5, 5 to 10, and >10 mm. The number of APC according to their size also did not differ significantly between colonoscopes with 170° FOV and those with 140° FOV (<5 mm: 0.75 ± 1.35 vs 0.75 ± 1.23, P = .915; 5–10 mm: 0.10 ± 0.54 vs 0.09 ± 0.60, P = .697; >10 mm: 0.04 ± 0.23 vs 0.04 ± 0.37, P = .725) (Table 3).

Table 3 - Adenomas per colonoscopy by endoscopic feature, location, and size.
Colonoscopes with fields of view
170° (n = 838) 140°s (n = 873) P value
Endoscopic feature
 Polypoid 0.05 ± 0.22 (44/838) 0.05 ± 0.23 (48/873) .822
 Flat 0.83 ± 1.44 (697/838) 0.83 ± 1.50 (725/873) .986
Location
 Proximal 0.62 ± 1.22 (521/838) 0.62 ± 1.27 (538/873) .928
 Distal 0.26 ± 0.62 (220/838) 0.27 ± 0.68 (235/873) .833
Size
 <5 mm 0.75 ± 1.35 (627/838) 0.75 ± 1.23 (659/873) .915
 5–10 mm 0.10 ± 0.54 (81/838) 0.09 ± 0.60 (75/873) .697
 >10 mm 0.04 ± 0.23 (33/838) 0.04 ± 0.37 (39/873) .725
Overall 0.88 ± 1.52 (741/838) 0.89 ± 1.59 (773/873) .987
APC = adenoma per colonoscopy.

3.4. Colonoscopy quality indicators according to endoscopists’ experience

Quality indicators of the colonoscopes with 170° FOV and those with 140° FOV were further compared based on the endoscopists’ experience. For trainee endoscopists, there were no statistically significant differences between colonoscopes with 170° FOV and those with 140° FOV in cecal intubation time (535.56 ± 252.21 vs 529.78 ± 314.68, P = .799), withdrawal time (793.38 ± 462.50 vs 756.99 ± 399.78, P = .291), PDR (52.73 vs 50.78%, P = .624), ADR (41.16 vs 40.13%, P = .792), or APC (0.80 ± 1.38 vs 0.85 ± 1.45, P = .664). Similarly, for experienced endoscopists as well, there were no statistically significant differences between colonoscopes in cecal intubation time (353.76 ± 158.50 vs 350.58 ± 112.25, P = .705), withdrawal time (710.49 ± 390.34 vs 735.52 ± 403.51, P = .301), PDR (58.63 vs 55.05%, P = .235), ADR (45.35 vs 42.96%, P = .429), and APC (0.94 ± 1.60 vs 0.91 ± 1.67, P = .782) (Table 4).

Table 4 - Colonoscopic quality indicators according to endoscopists’ experience.
Colonoscopes with fields of view
170° (n = 838) 140° (n = 873) P value
Trainee endoscopists
 Intubation time (s) 535.56 ± 252.21 529.78 ± 314.68 .799
 Withdrawal time, (s) 793.38 ± 462.50 756.99 ± 399.78 .291
 PDR, % 52.73 (164/311) 50.78 (162/319) .624
 ADR, % 41.16 (128/311) 40.13 (128/319) .792
 APC 0.80 ± 1.38 0.85 ± 1.45 .664
Experienced endoscopists
 Intubation time (s) 353.76 ± 158.50 350.58 ± 112.25 .705
 Withdrawal time (s) 710.49 ± 390.34 735.52 ± 403.51 .301
 PDR, % 58.63 (309/527) 55.05 (305/554) .235
 ADR, % 45.35 (239/527) 42.96 (238/554)) .429
 APC 0.94 ± 1.60 0.91 ± 1.67 .782
ADR = adenoma detection rate, APC = adenoma per colonoscopy, PDR = polyp detection rate.

3.5. Univariate and multivariate analysis of factors associated with adenoma detection

Age, male sex, surveillance colonoscopy, and long withdrawal time were significantly associated with adenoma detection in the univariate logistic regression analysis. These factors were then analyzed as covariates in multivariate logistic regression analysis, out of which age, male sex, and long withdrawal time remained statistically significant as opposed to surveillance colonoscopy (Table 5).

Table 5 - Univariate and multivariate analysis for factors related to adenoma detection.
Univariate analysis Multivariate analysis
OR (95% CI) P value OR (95% CI) P value
Age (yr) 1.045 (1.035–1.055) .000 1.037 (1.027–1.048) .000
Gender .000 .000
 Male 1 1
 Female .544 (.447–.661) . .626 (.502–.780)
Indication of exam .003 .581
 Screening 1 1
 Surveillance 1.409 (1.128–1.761) 1.074 (.834–1.383)
Withdrawal time (s) 1.003 (1.003–1.003) .000 1.003 (1.002–1.003) .000
Type of endoscopy .434
 170° 1
 140° 0.926 (0.765–1.122)

4. Discussion

This single-center retrospective study aimed to evaluate the difference in quality indicators such as ADR between 170° and 140° FOV colonoscopes with the same HD resolution. We found that ADR was not affected by variations in the FOV of colonoscopes with the same high-resolution image quality. We also found that other colonoscopy quality indicators, such as PDR, SSPDR, ANDR, and APC, also did not differ based on the FOV.

Several innovations in endoscopic technology have been aimed at improving adenoma detection. Newer colonoscopes (such as Olympus CV-290 or CV-190) have a wider FOV of 170° than the older colonoscopes having a 140° FOV, and also possess features such as higher resolution and near focus.[21] Previous studies comparing older and newer colonoscopes from the same manufacturer have shown that newer colonoscopes have higher ADRs.[22–25] Studies that compared the newer CV-290 or CV-190 (170° FOV) with the older CV-260 or CV-165 (140° FOV) (Olympus Medical, Tokyo, Japan) have also reported differences in ADR, but the differences cannot be solely attributed to the FOV as there were differences also in the resolution.[22,24] Among comparative studies on colonoscopes of different generations, EPK-i7010 and EPK-i (Pentax, Hoya Corporation, Tokyo, Japan) showed higher ADRs in the latest generation. A study in which both generations of colonoscopes had an FOV of 140°, it was found that the difference between them was due to differences in the image quality.[25] In another study that was conducted to confirm the difference in the ADR depending on the manufacturer of the colonoscope, there was no difference in results with FOV 140° for colonoscopes from different manufacturers.[26] It is therefore necessary to determine whether the FOV affects the ADR because there are slight differences in the FOV of the latest generation of colonoscopes for each manufacturer.

To the best of our knowledge, no study has investigated whether the visual field difference of the colonoscope affects quality indicators, such as ADR, without using additional devices to expand the FOV of colonoscopes with the same resolution. The 2 types of colonoscopes used in this study were newer-generation colonoscopes with similar launch times. Our study confirmed that if the image quality was the same, a difference in FOV of 30° did not affect the quality indicators, including the ADR.

Special equipment, such as Endocuff, EndoRings, and FUSE, improve colonoscopic qualities; however, a study has shown that HD colonoscopy with Endocuff and EndoRings had a higher APC than that with FUSE,[27] due to the better HD image quality. The results showed that the image quality could affect adenoma detection more than the FOV. These results are consistent with those of our study, which showed that there was no difference in the ADR with different FOV at the same resolution. The human FOV is approximately 210° horizontally, but with binocular vision only 114° is required for depth perception.[28] Therefore, it can be seen that the expansion of the FOV contributes to improving colonoscopic quality. However, if the FOV is above a certain level, further expansion of the FOV does not proportionally improve the quality of the endoscope.

This study has some limitations. First, there were no significant differences in the variables between the 2 groups for a certain period. Therefore, it was impossible to identify all factors affecting ADR, such as family history, smoking history, and other comorbidities, which are difficult to confirm using chart review. However, this study retrospectively included a large number of patients who underwent screening or surveillance colonoscopy, and most patients were Korean. Therefore, it is estimated that there was no significant difference between the 2 groups for the factors that we have not reviewed. Second, endoscope manufacturers do not disclose the specifications of the video processors. The visual quality of the video processor can be quantified; however, the numerical specifications for color and brightness have not been disclosed. Therefore, only the image quality and FOV were considered, not other subtle differences. However, the image quality specifications of the 2 manufacturers were the same at 1080 intercalating vertical pixels, and the image quality subjectively experienced by the endoscopists in this study was also similar. Among the factors affecting ADR in several studies, it has been proven that image quality is the most important.[22,24,27] This is the first study to confirm whether the difference in the FOV of the colonoscope affects the ADR without the use of special equipment to expand the FOV.

5. Conclusion

This study revealed no significant differences in ADR between HD colonoscopes with a 170° FOV and a 140° FOV. This shows that a slight difference in the FOV at the same resolution does not significantly impact colonoscopic quality.

Author contributions

Conceptualization: Jong Yoon Lee.

Data curation: Sang Yi Moon, Jong Yoon Lee.

Formal analysis: Sang Yi Moon, Jong Yoon Lee.

Funding acquisition: Jong Yoon Lee.

Investigation: Sang Yi Moon, Jong Yoon Lee.

Methodology: Jong Yoon Lee.

Supervision: Jong Yoon Lee, Jong Hoon Lee.

Validation: Jong Yoon Lee.

Visualization: Jong Yoon Lee.

Writing – original draft: Sang Yi Moon, Jong Yoon Lee.

Writing – review & editing: Jong Yoon Lee, Jong Hoon Lee.

Abbreviations:

ADR =
adenoma detection rate
ANDR =
advanced neoplasia detection rate
APC =
adenomas per colonoscopy
CRC =
colorectal cancer
FOV =
field of view
FUSE =
full-spectrum endoscopy
HD =
high-definition
PDR =
polyp detection rate
SSPDR =
sessile serrated PDR

References

[1]. Gupta S, Lieberman D, Anderson JC, et al. Recommendations for follow-up after colonoscopy and polypectomy: a consensus update by the US multi-society task force on colorectal cancer. Am J Gastroenterol. 2020;115:415–34.
[2]. Citarda F, Tomaselli G, Capocaccia R, et al. Efficacy in standard clinical practice of colonoscopic polypectomy in reducing colorectal cancer incidence. Gut. 2001;48:812–5.
[3]. Zauber AG, Winawer SJ, O’Brien MJ, et al. Colonoscopic polypectomy and long-term prevention of colorectal-cancer deaths. N Engl J Med. 2012;366:687–96.
[4]. Singh S, Singh PP, Murad MH, et al. Prevalence, risk factors, and outcomes of interval colorectal cancers: a systematic review and meta-analysis. Am J Gastroenterol. 2014;109:1375–89.
[5]. Rutter MD, Beintaris I, Valori R, et al. World Endoscopy Organization consensus statements on post-colonoscopy and post-imaging colorectal cancer. Gastroenterology. 2018;155:909–925.e3.
[6]. le Clercq CM, Bouwens MW, Rondagh EJ, et al. Postcolonoscopy colorectal cancers are preventable: a population-based study. Gut. 2014;63:957–63.
[7]. Robertson DJ, Lieberman DA, Winawer SJ, et al. Colorectal cancers soon after colonoscopy: a pooled multicohort analysis. Gut. 2014;63:949–56.
[8]. Kaminski MF, Regula J, Kraszewska E, et al. Quality indicators for colonoscopy and the risk of interval cancer. N Engl J Med. 2010;362:1795–803.
[9]. Corley DA, Jensen CD, Marks AR, et al. Adenoma detection rate and risk of colorectal cancer and death. N Engl J Med. 2014;370:1298–306.
[10]. Rex DK, Boland CR, Dominitz JA, et al. Colorectal cancer screening: recommendations for physicians and patients from the U.S. multi-society task force on colorectal cancer. Gastrointest Endosc. 2017;86:18–33.
[11]. Kaminski MF, Thomas-Gibson S, Bugajski M, et al. Performance measures for lower gastrointestinal endoscopy: a European society of gastrointestinal endoscopy (ESGE) quality improvement initiative. Endoscopy. 2017;49:378–97.
[12]. Barclay RL, Vicari JJ, Doughty AS, et al. Colonoscopic withdrawal times and adenoma detection during screening colonoscopy. N Engl J Med. 2006;355:2533–41.
[13]. Kluge MA, Williams JL, Wu CK, et al. Inadequate Boston Bowel Preparation Scale scores predict the risk of missed neoplasia on the next colonoscopy. Gastrointest Endosc. 2018;87:744–51.
[14]. Baxter NN, Sutradhar R, Forbes SS, et al. Analysis of administrative data finds endoscopist quality measures associated with postcolonoscopy colorectal cancer. Gastroenterology. 2011;140:65–72.
[15]. Adler A, Aminalai A, Aschenbeck J, et al. Latest generation, wide-angle, high-definition colonoscopes increase adenoma detection rate. Clin Gastroenterol Hepatol. 2012;10:155–9.
[16]. Atkinson NSS, Ket S, Bassett P, et al. Narrow-band imaging for detection of neoplasia at colonoscopy: a meta-analysis of data from individual patients in randomized controlled trials. Gastroenterology. 2019;157:462–71.
[17]. Facciorusso A, Triantafyllou K, Murad MH, et al. Compared abilities of endoscopic techniques to increase colon adenoma detection rates: a network meta-analysis. Clin Gastroenterol Hepatol. 2019;17:2439–2454.e25.
[18]. Leufkens AM, DeMarco DC, Rastogi A, et al. Effect of a retrograde-viewing device on adenoma detection rate during colonoscopy: the TERRACE study. Gastrointest Endosc. 2011;73:480–9.
[19]. Dik VK, Gralnek IM, Segol O, et al. Multicenter, randomized, tandem evaluation of EndoRings colonoscopy--results of the CLEVER study. Endoscopy. 2015;47:1151–8.
[20]. Gralnek IM, Siersema PD, Halpern Z, et al. Standard forward-viewing colonoscopy versus full-spectrum endoscopy: an international, multicentre, randomised, tandem colonoscopy trial. Lancet Oncol. 2014;15:353–60.
[21]. Bhat YM, Abu Dayyeh BK, Chauhan SS, Gottlieb KT, et al. High-definition and high-magnification endoscopes. Gastrointest Endosc. 2014;80:919–27.
[22]. Bond A, O’Toole P, Fisher G, et al. New-generation high-definition colonoscopes increase adenoma detection when screening a moderate-risk population for colorectal cancer. Clin Colorectal Cancer. 2017;16:44–50.
[23]. Jrebi NY, Hefty M, Jalouta T, et al. High-definition colonoscopy increases adenoma detection rate. Surg Endosc. 2017;31:78–84.
[24]. Zimmermann-Fraedrich K, Groth S, Sehner S, et al. Effects of two instrument-generation changes on adenoma detection rate during screening colonoscopy: results from a prospective randomized comparative study. Endoscopy. 2018;50:878–85.
[25]. Lee JY, Koh M, Lee JH. Latest generation high-definition colonoscopy increases adenoma detection rate by trainee endoscopists. Dig Dis Sci. 2021;66:2756–62.
[26]. Chernolesskiy A, Swain D, Lee JC, et al. Comparison of Pentax HiLine and Olympus Lucera systems at screening colonoscopy. World J Gastrointest Endosc. 2013;5:62–6.
[27]. Rex DK, Repici A, Gross SA, et al. High-definition colonoscopy versus Endocuff versus EndoRings versus full-spectrum endoscopy for adenoma detection at colonoscopy: a multicenter randomized trial. Gastrointest Endosc. 2018;88:335–344.e2.
[28]. Howard IP, Rogers BJ. Binocular vision and stereopsis. USA: Oxford University Press. 1995.
Keywords:

adenomatous polyps; colonic polyps; colonoscopy; endoscopy