Research Article: Clinical Case Report

Use of over-the-scope clip for ulcer hemostasis after failure of both surgery and interventional therapy

A case report

Lai, Yong-Kang MD; Peng, Chun-Yan MD; Guo, Gui-Hai MD; Shu, Xu MD

Editor(s): Saranathan., Maya

Author Information
doi: 10.1097/MD.0000000000022646
  • Open

Abstract

1 Introduction

Upper gastrointestinal bleeding is a significant cause of hospitalization and mortality worldwide,[1] and despite advances in medications and therapeutic techniques, the rates of rebleeding and mortality remain high.[2] Endoscopic hemostasis is the first choice of treatment for patients with upper gastrointestinal bleeding,[3] and surgical and radiological interventions are secondary measures.[4,5] However, surgical and radiological interventions sometimes fail to stop the bleeding. In such cases, it is unclear what the next step should be, such as a second endoscopic hemostasis procedure, additional interventional therapies, or an additional surgery. Research on this topic has been scarce.

The over-the-scope clip (OTSC) system (Ovesco Endoscopy AG, Tübingen, Germany) is a novel endoscopic device that can be used to manage leaks and fistulas,[6,7] remove submucosal tumors,[8] and close perforations.[9] In addition, the efficacy of this system in treating high-risk gastrointestinal bleeding has been widely reported.[10,11] We used the OTSC system to successfully achieve hemostasis in a patient who was still bleeding from a lesion after surgical and interventional treatments. We then performed long-term follow-up observations.

2 Case report

A 47-year-old man had intermittent dull pain in the upper abdomen after eating greasy food for 2 days associated with melena twice a day before he was admitted to the local hospital. During hospitalization, the patient experienced hemorrhagic shock and underwent an initial emergency gastroscopy. An ulcer (0.8 cm × 0.8 cm) was identified on the anterior wall of the duodenal bulb (Fig. 1). The ulcer had a large area, and its anatomical location was difficult to approach. Moreover, the ulcer caused arterial bleeding, thus traditional endoscopic hemostasis was difficult to achieve; thus, effective hemostasis was not attained. Due to the large amount of bleeding, poor endoscopic field of vision, and small operating space, the ulcer could not be treated. Thus, the patient was immediately transferred to surgery, and a gastroenterostomy (Billroth II) + Bancroft operation was performed. During the operation, ulcers with scabs were observed near the descending part of the duodenal bulb and the ulcerated area was deep. After determining the area for surgical excision, it was still difficult to fully expose the intestinal tubes of the duodenal ulcer from the outside; therefore, the surgeon at the local hospital selected the Billroth II + Bancroft operation. The bleeding stopped after the operation.

Figure 1
Figure 1:
Endoscopy performed at a local hospital. Endoscopic images suggested that an ulcer (0.8 cm × 0.8 cm) was on the anterior wall of the duodenal bulb.

Seven days postoperatively, the patient again experienced melena, which suggested that the ulcer had begun bleeding again. He was immediately transferred to the Department of Gastroenterology at the First Affiliated Hospital of Nanchang University, where he underwent endoscopic hemostasis for the second time. However, no lesion was observed. Digital subtraction angiography (DSA) was also performed, and it revealed no signs of hemorrhage (Fig. 2). The patient's condition was stable after blood transfusion and acid inhibition therapy. Five days after he was admitted to our hospital, the patient suddenly vomited a large quantity of blood (approximately 800 mL), and his hemoglobin level dropped to 57 g/L. An emergency endoscopy was performed, and a large amount of fresh blood was observed in the esophagus and residual stomach. Repeated washing of the anastomotic site showed that there were no bleeding points. Then, we used a JI enteroscope (PCF-Q260JI, Olympus, Tokyo, Japan), which can extend deeper than a normal gastroscope. Fortunately, blood clots were observed in the input loop, and residual blood vessels were observed near the blind end (Fig. 3). Next, the positioning for the OTSC system was confirmed above the lesion. By rotating the wheel, which was attached to the endoscope, the OTSC was quickly deployed to the target site and the hemostasis process was completed. Repeated flushing revealed no additional blood (Fig. 4), indicating that the bleeding had stopped. We performed follow-up examinations at 5 months, 10 months, and 1 year after endoscopy (Fig. 5). Rebleeding was not observed in any of these examinations.

Figure 2
Figure 2:
Digital subtraction angiography (DSA) performed at the First Affiliated Hospital of Nanchang University for treatment. Bleeding was not found in the gastroduodenal artery or superior mesenteric artery.
Figure 3
Figure 3:
JI enteroscope (PCF-Q260JI, Olympus, Tokyo, Japan) was used innovatively. Fortunately, blood clots were observed in the input loop and residual blood vessels were observed near the blind end. Then, the OTSC system was used to treat the lesion. Repeated flushing showed no bleeding. OTSC = over-the-scope-clip.
Figure 4
Figure 4:
OTSC system used to treat the lesion; repeated flushing showed no bleeding. OTSC = over-the-scope-clip.
Figure 5
Figure 5:
One-year follow-up endoscopy showed that the patient did not bleed again.

3 Discussion

We report on a patient treated with the OTSC system and find that the system successfully stopped rebleeding after the failure of both surgical and interventional treatments. In this patient, the ulcer's anatomical location was difficult to approach, and the large amount of bleeding, poor endoscopic field of vision, and small operating space made traditional endoscopic hemostasis difficult to fully achieve. In addition, the patient was in worse condition because of the ulcer's bleeding. The first emergency endoscopy failed to stop the bleeding; consequently, surgery was immediately performed to achieve hemostasis. To avoid the complications of a Billroth II-type gastric resection,[12] the local hospital chose a Bancroft operation to address the duodenal ulcer. Although the Billroth II operation is the cornerstone for the management of patients with uncontrolled bleeding,[13] the ulcer was not completely resected. When the patient underwent an emergency endoscopy for the second time at our hospital, the ulcer was located so deeply that the exact location could not be identified using the normal endoscopy procedure. Therefore, DSA was used to look for the source of the gastrointestinal bleeding; however, we were still unable to identify the ulcer. The reasons for this failure might be as follows: the bleeding arteries were too small to be observed; the lesion bled intermittently and had temporarily stopped bleeding when we performed the DSA procedure; and DSA might not be an effective technique for identifying bleeding areas. Therefore, we made the innovative choice to use the JI enteroscope and finally found the lesion near the blind side when the patient again experienced hematemesis. Then, the OTSC system was used to successfully stop the bleeding.

The OTSC system is a novel and full-thickness suturing device that can close lesions, such as perforations,[14] fistulas,[6] and sites of gastrointestinal bleeding.[15] Its characteristics, including a high-elasticity nickel alloy, shape-memory function, and strong and persistent organizational grip, render the OTSC system effective for the continuous closure of wounds, which is why the OTSC system was able to stop the patient's ulcer from bleeding after the failure of both surgery and DSA.

To date, several studies have reported on the use of the OTSC system as a rescue treatment for refractory bleeding after endoscopic hemostasis.[10,16–18] However, in those studies, surgery was considered the final choice, and whether endoscopic hemostasis could be attempted again after the failure of surgery and interventional therapy was not indicated. Reports of rescue therapy after the failure of surgical and interventional treatments have been rare. This study is innovative in the use of J-type enteroscopy to detect a deep ulcer and the OTSC system to close the wound after the failure of surgical treatment, which could thus serve as an alternative treatment for hemostasis after the failure of surgery and interventional therapy.

In conclusion, the application of the OTSC system is a reasonable option for controlling ulcer rebleeding after the failure of both surgery and interventional therapy, especially for ulcers with deep anatomical locations, large amounts of bleeding, and small bleeding arteries. This is the first report suggesting that surgical and interventional therapies are not the last line of defense against ulcer bleeding. Thus, for cases of ulcers in which both surgical and interventional therapies fail, the OTSC system may represent a treatment option.

Author contributions

Xu Shu designed the report, approved the final submission, and clinically managed the patient; Yong-Kang Lai collected the data, analyzed the relevant information, and wrote the manuscript; and Chun-Yan Peng and Gui-Hai Guo clinically managed the patient.

References

[1]. Lau JY, Barkun A, Fan DM, et al. Challenges in the management of acute peptic ulcer bleeding. Lancet 2013;381:2033–43.
[2]. Kim JS, Kim BW, Park SM, et al. Factors associated with rebleeding in patients with peptic ulcer bleeding: analysis of the Korean Peptic Ulcer Bleeding (K-PUB) Study. Gut Liver 2018;12:271–7.
[3]. Chandrasekar VT, Desai M, Aziz M, et al. Efficacy and safety of over-the-scope clips for gastrointestinal bleeding: a systematic review and meta-analysis. Endoscopy 2019;51:941–9.
[4]. Laine L, Jensen DM. Management of patients with ulcer bleeding. Am J Gastroenterol 2012;107:345–60. quiz 361.
[5]. Gralnek IM, Dumonceau JM, Kuipers EJ, et al. Diagnosis and management of nonvariceal upper gastrointestinal hemorrhage: European Society of Gastrointestinal Endoscopy (ESGE) Guideline. Endoscopy 2015;47:a1–46.
[6]. Shoar S, Poliakin L, Khorgami Z, et al. Efficacy and safety of the Over-the-Scope Clip (OTSC) system in the management of leak and fistula after laparoscopic sleeve gastrectomy: a systematic review. Obes Surg 2017;27:2410–8.
[7]. Keren D, Eyal O, Sroka G, et al. Over-the-Scope Clip (OTSC) system for sleeve gastrectomy leaks. Obes Surg 2015;25:1358–63.
[8]. Sarker S, Gutierrez JP, Council L, et al. Over-the-scope clip-assisted method for resection of full-thickness submucosal lesions of the gastrointestinal tract. Endoscopy 2014;46:758–61.
[9]. Daher S, Khoury T, Benson AA, et al. Hospital management of colonic perforations complicating ambulatory outpatient colonoscopy via over-the-scope clips or surgery: a case series. Tech Coloproctol 2019;23:681–5.
[10]. Brandler J, Baruah A, Zeb M, et al. Efficacy of Over-the-Scope Clips in management of high-risk gastrointestinal bleeding. Clin Gastroenterol Hepatol 2018;16:690.e1–6.e1.
[11]. Manno M, Mangiafico S, Caruso A, et al. First-line endoscopic treatment with OTSC in patients with high-risk non-variceal upper gastrointestinal bleeding: preliminary experience in 40 cases. Surg Endosc 2016;30:2026–9.
[12]. Malheiros CA, Moreno CH, Rodrigues FC, et al. Finsterer-Bancroft operation: an option for the treatment of difficult duodenal ulcers. Int Surg 1998;83:111–4.
[13]. Lanas A, Chan FKL. Peptic ulcer disease. Lancet 2017;390:613–24.
[14]. Ngo E, Born P, Strebel H. Endoscopic treatment of an oesophageal rupture using an over-the-scope clip (OTSC): a case report. Arab J Gastroenterol 2015;16:139–41.
[15]. Wedi E, Gonzalez S, Menke D, et al. One hundred and one over-the-scope-clip applications for severe gastrointestinal bleeding, leaks and fistulas. World J Gastroenterol 2016;22:1844–53.
[16]. Kobara H, Mori H, Nishiyama N, et al. Over-the-scope clip system: a review of 1517 cases over 9 years. J Gastroenterol Hepatol 2019;34:22–30.
[17]. Richter-Schrag HJ, Glatz T, Walker C, et al. First-line endoscopic treatment with over-the-scope clips significantly improves the primary failure and rebleeding rates in high-risk gastrointestinal bleeding: a single-center experience with 100 cases. World J Gastroenterol 2016;22:9162–71.
[18]. Schmidt A, Gölder S, Goetz M, et al. Over-the-Scope Clips are more effective than standard endoscopic therapy for patients with recurrent bleeding of peptic ulcers. Gastroenterology 2018;155:674.e6–86.e6.
Keywords:

over-the-scope clip; radiological interventions; surgery; ulcer bleeding; upper gastrointestinal bleeding

Copyright © 2020 the Author(s). Published by Wolters Kluwer Health, Inc.