Performance and Hemocompatibility of a Novel Polysulfone Dialyzer: A Randomized Controlled Trial : Kidney360

Journal Logo

Original Investigation: Dialysis

Performance and Hemocompatibility of a Novel Polysulfone Dialyzer: A Randomized Controlled Trial

Ehlerding, Götz1; Erlenkötter, Ansgar2; Gauly, Adelheid3; Griesshaber, Bettina3; Kennedy, James2; Rauber, Lena2; Ries, Wolfgang4; Schmidt-Gürtler, Hans1; Stauss-Grabo, Manuela3; Wagner, Stephan5; Zawada, Adam M.2; Zschätzsch, Sebastian5; Kempkes-Koch, Manuela6

Author Information
Kidney360 2(6):p 937-947, June 2021. | DOI: 10.34067/KID.0000302021
  • Open
  • Infographic
  • SDC

Introduction

The majority of patients who receive RRT rely on extracorporeal treatments (1). Here, the dialyzer is the cornerstone for adequate dialysis performance and a central element on which efforts are focused to provide the most hemocompatible extracorporeal circuit.

High-flux dialyzers, applied either in hemodialysis (HD) or in online hemodiafiltration (HDF), deliver the most efficient elimination of uremic retention solutes over the relevant mol wt range. High-flux dialyzers aim at approaching the elimination characteristics of the kidney, and a sharp mol wt cutoff prevents excessive albumin loss. The importance of removing both small- and medium-sized mol wt uremic solutes (2) has been underscored by studies identifying β2-microglobulin as an independent risk factor for mortality among patients on dialysis (3,4). Moreover, the hydraulic characteristics of high-flux dialyzers allow for online HDF, which has gained wide clinical acceptance on the basis of evidence of further improved elimination of uremic solutes (5) and improved survival (6), particularly when treated with high convection volumes (78–9).

The role of dialyzer hemocompatibility for different patient outcomes has been widely investigated. Upon contact of blood with foreign surfaces, cell reactions and release of humoral factors may be triggered. The latter includes the activation of the complement system, with a release of complement factors and the induction of cytokines and coagulation factors (10). Release of humoral factors may contribute to various pathophysiologic processes in patients on HD (11). This hypothesis is supported by recent findings from a population-based cohort of people, without CKD, where the complement factors C5a and sC5b-9 were associated with markers of acute and chronic inflammation processes and of endothelial dysfunction (12). Complement activation during dialysis treatments is most prominent with cellulose-based membranes; however, synthetic membranes also induce transient complement activation (13). Besides this transient reaction, complement activation has been reported to have a long-term effect on the outcomes of patients on dialysis, such as increased infection risk, fibrosis, and cardiovascular events (1415–16). Complement activation has also been discussed as a factor contributing to the higher mortality risk that has been associated with cellulose-based membranes, as compared with synthetic membranes (17).

Therefore, with potential acute and long-term implications through cellular and humoral activation by artificial surfaces, a good hemocompatibility profile is central to the design of new dialysis membranes (10,18). The FX CorAL 600 dialyzer was recently developed as part of the ongoing effort to improve dialyzer hemocompatibility. This dialyzer contains a membrane comprising a blend of polysulfone and polyvinylpyrrolidone, with small amounts of α-tocopherol added to stabilize the blood-side surface of the membrane (19). The objective of this study was to determine the effect of this modification on performance and hemocompatibility of the new dialyzer, in comparison with two commercially available dialyzers.

Materials and Methods

Study Design and Patients

The objective of the study was to investigate the removal rate and clearance of uremic retention solutes, and determine the hemocompatibility and safety of the new dialyzer, in comparison with two commercially available dialyzers.

This was a prospective, randomized, crossover, and multicenter study. It was performed between August and December, 2018, in six study centers in Germany. We enrolled adult patients with chronic kidney failure, a minimum of 3 months on online HDF, fistula or graft as vascular access, and without major illnesses and known allergies to trial products or related products. Patients were studied during three online HDF sessions in the course of a week, with study dialyzers in a randomized order. Patients were allocated to one of the six possible treatment sequences through central randomization, stratified by study center. The new dialyzer FX CorAL 600 (Fresenius Medical Care, Bad Homburg, Germany) was compared with two commercially available dialyzers with good performance in small- and medium-sized molecule removal: one with a synthetic membrane material, the polyarylethersulfone-based Polyflux 170H dialyzer (Baxter Healthcare Corporation, Deerfield, IL) (20); and one cellulose triacetate–based dialyzer, the SureFlux 17UX (Nipro Medical Europe, Mechelen, Belgium) (21) (Table 1). A 1 week follow-up for safety surveillance was included at the end of the study, yielding a total study duration of 4 weeks per patient.

Table 1. - Characteristics of study dialyzers
Characteristic FX CorAL 600 SureFlux 17UX Polyflux 170H
Membrane material Polysulfone-polyvinylpyrrolidone blend Cellulose triacetate Polyamix (polyarylethersulfone, polyvinylpyrrolidone, and polyamide blend)
Membrane surface area (m2) 1.6 1.7 1.7
Inner fiber diameter (µm) 210 200 215
Sterilization method INLINE steam sterilization Dry γ sterilization Steam sterilization (inside out)

Treatments were performed with the 5008 or the 6008 HD system (Fresenius Medical Care) as HDF in online postdilution mode. Treatment parameters (including a blood flow rate of at least 300 ml/min, a dialysate flow rate of ≥500 ml/min, a minimum treatment time of 240 minutes, and a patient-adjusted ultrafiltration rate) remained unchanged between study phases. Substitution flow rate was manually set, and anticoagulation was provided according to center practice.

The study was approved by the ethics committees of the participating centers and the national competent authority. Patients were informed orally and in writing on the purpose, meaning, and risk of the study, and were included in the study after signing and dating the informed consent form. The study was executed in accordance with the Declaration of Helsinki in its current version.

Outcome Variables and Laboratory Methods

Dialyzer performance was assessed using removal rate and clearance of β2-microglobulin, urea, creatinine, phosphate, myoglobin, and α1-microglobulin. To calculate removal rate, plasma samples were taken at the midweek session from the arterial line pre- and postdialysis. To calculate clearance, samples were taken from both the arterial and venous line at 60 minutes after treatment start.

Complement factors, blood cell counts, markers of cell and immune activation, coagulation factors, and contact activation factors were analyzed in samples taken predialysis, 15 and 60 minutes after treatment start, and postdialysis (at 240 minutes). Thrombin-antithrombin (TAT) was measured predialysis, at 120 minutes, and at 240 minutes after treatment start.

All laboratory analyses were performed following the methods given in Supplemental Table 1.

Statistical Analyses

The primary outcome of this study was noninferiority of the FX CorAL 600 versus the SureFlux 17UX and the Polyflux 170H with respect to the removal rates of β2-microglobulin. The noninferiority margin was set to −2% for the difference between the FX CorAL 600 and either comparator dialyzer.

The sample size estimation was determined on the basis of the hypothesis of noninferiority of the FX CorAL 600 dialyzer compared with the SureFlux 17UX and Polyflux 170H for the outcome measure of β2-microglobulin removal rate. Assuming a lower limit of noninferiority of −2%; an intraclass correlation of 0.4, due to clustering of patients in centers; a one-sided significance level of 1.25%; a drop-out rate of 25%; and a group size divisible by six (the number of study centers), a minimum sample size of 48 patients was estimated.

Removal rates (%) from time=0 minutes (predialysis) to time=240 minutes after start of HDF (postdialysis), and plasma clearances (kb) at 60 minutes after start of HDF, were calculated for β2-microglobulin, myoglobulin, α1-microglobulin, phosphate, creatinine, and urea. Laboratory parameters were corrected for hematocrit, as appropriate. All formulas are given in Supplemental Table 2.

Assuming no carry-over effect, a linear mixed model was used for statistical analysis. This model included the fixed effects “period” and “dialyzer,” and the random effects “center” and “patient.” The comparison of the FX CorAL 600 with each of the other two comparator dialyzers was performed by defining contrasts or estimators of the fixed effect “dialyzer.” Noninferiority comparisons were made on the basis of the two contrasts that allow for the estimation of the mean difference in the primary variable removal rate of β2-microglobulin between the FX CorAL 600 and the SureFlux 17UX and between the FX CorAL 600 and the Polyflux 170H, respectively. Further details on the two-step test and adjustments according to Bonferroni and Holm (22) are given in Supplemental Appendix 1.

Because the intention-to-treat (ITT) and per-protocol (PP) populations were identical, no sensitivity analyses were carried out. Secondary efficacy analyses were conducted on the ITT/PP population. Safety analysis was based on the safety population. Here, all data available from each randomized patient were taken into account.

Sample size estimation and statistical analysis was performed using SAS version 9.4 (SAS Institute Inc., Cary, NC).

Clinical Trial registry name and registration number

Comparative Safety and Clinical Performance of Dialyzers Applied During Post-dilution Online Hemodiafiltration (eMPORA II), NCT03611218

Results

Patients and Treatments

A total of 70 patients from six study centers were recruited. These patients composed the safety population, of which 12 patients were excluded due to dropout before study start (one patient), premature study termination (one patient), and missing or invalid primary efficacy parameter (ten patients). Thus, 58 patients represented the ITT population, which was identical to the PP population because no further criteria to exclude patients from the PP population were applicable.

Patients enrolled in the study were, on average, 68 years old and 74% were male (Table 2). Following the study protocol, treatment parameters during the three study phases were nearly unchanged with blood flow rate of approximately 300 ml/min, dialysate flow rate of 500 ml/min, and treatment time of 4.5 hours; anticoagulation was predominantly performed with unfractionated heparin (Table 3).

Table 2. - Patient characteristics at baseline (ITT/PP)
Parameter Value (N=58)
Age (yr), mean±SD 67.8±13.4
Sex, % male 74
Body weight (kg), mean±SD
 Female 80.5±26.2
 Male 90.1±19.7
Body mass index (kg/m2), mean±SD 28.9±6.6
Primary renal disease, N (%) a
 Diabetes 21 (36)
 Hypertension/large vessel disease 17 (29)
 Glomerulonephritis 7 (12)
 Cystic/hereditary/congenital disease 6 (10)
 Others/unknown 21 (36)
Time on RRT (mo), median (range) 50 (3–212)
ITT, intention to treat; PP, per protocol.
aMore than one disease could be documented.

Table 3. - Treatment parameters applied in study phases using indicated dialyzer
Treatment parameter FX CorAL 600 (N=58) SureFlux 17UX (N=58) Polyflux 170H (N=58)
Parameter, mean±SD
 Effective blood flow rate (ml/min) 306±29 303±25 305±26
 Dialysate flow rate (ml/min) 500±0.1 500±0.3 500±0.0
 Substitution volume (L) 17.6±4.5 15.3±2.8 17.4±4.8
 Ultrafiltration volume (L) 2.2±1.2 2.2±1.3 2.2±1.3
 Effective treatment time (min) 272±29 270±29 270±29
Anticoagulation (% of patients)
 Unfractionated heparin (bolus and infusion) 79 81 79
 Low mol wt heparin (bolus and infusion) 9 7 9
 Low mol wt heparin (bolus) 12 12 12

Dialyzer Performance Measures

The least squares means of the removal rate of β2-microglobulin were 70%, 51%, and 68% for the FX CorAL 600, the SureFlux 17UX, and the Polyflux 170H, respectively (Figure 1). Noninferiority of FX CorAL 600 versus both comparators was statistically significant (P<0.001 versus SureFlux 17UX; P=0.0006 versus Polyflux 170H; Table 4). Further, as derived from the confidence intervals of the difference between the dialyzers, the removal rate of β2-microglobulin was significantly higher with the FX CorAL 600 than with the SureFlux 17UX, and was comparable with the Polyflux 170H (Table 4). The descriptive analysis of further performance parameters is presented in Supplemental Table 3. Clearance of β2-microglobulin with the FX CorAL 600 was significantly higher than with SureFlux 17UX (P<0.001). The clearance and removal rate of myoglobin was lower than with both comparators. The performance of the FX CorAL 600 was not different for α1-microglobulin and phosphate, and was slightly, but not clinically relevantly, lower for urea and creatinine (Supplemental Table 3).

F1
Figure 1.:
Removal rate (RR) of β 2-microglobulin with the three tested dialyzers. The interquartile range (IQR; box), mean (square), median (line), minimum and maximum value within IQR±1.5 IQR (whiskers), and outliers (circles) are displayed.
Table 4. - Removal rate of β2-microglobulin
Dialyzer N LSMean β2-Microglobulin Removal Rate (%) SEM 97.5% CI P Value a 95% CI P Value a
FX CorAL 600 58 70.33 1.35

15.82 to 22.17

SureFlux 17UX 58 51.34 1.35
Polyflux 170H 58 67.68 1.35
Difference FX CorAL 600−SureFlux 17UX 18.99 1.40 <0.001 16.22 to 21.76 <0.001
Difference FX CorAL 600−Polyflux 170H 2.66 1.40 −0.52 to 5.83 0.0006 −0.11 to 5.42 0.0006
Least squares mean and test for noninferiority of FX CorAL 600 versus SureFlux 17UX and versus Polyflux 170H. LSMean, least squares mean; SEM, standard error of the mean; 95/97.5% CI, 95/97.5% confidence interval of the differences between investigated dialyzers.
aP value to confirm noninferiority.

Complement Activation

With all dialyzers, a peak of C3a and C5a concentrations at 15 minutes after treatment start was observed, which then declined until the end of the treatment (Figure 2, Supplemental Table 4). Moreover, with all dialyzers, sC5b-9 increased within the first 60 minutes and then also declined until the end of the treatment (Figure 2). The increase from predialysis to peak level of C3a and C5a at 15 minutes was significantly less with the FX CorAL 600 than with the SureFlux 17UX, and was comparable with the Polyflux 170H. The increase of sC5b-9 from predialysis to all time points during treatment was significantly less with the FX CorAL 600 than with the SureFlux 17UX, and was also less, but not significantly so, compared with the Polyflux 170H (Supplemental Table 4). When comparing the three dialyzers at the peak concentrations of C3a and C5a (15 minutes), and of sC5b-9 (60 minutes), the FX CorAL 600 showed significantly lower levels of all complement factors than the SureFlux 17UX (P<0.001 for C3a; P=0.02 for C5a; P<0.001 for sC5b-9; Tables 5 and 6). The difference between the FX CorAL 600 and SureFlux 17UX for sC5b-9 persisted until treatment end. The FX CorAL 600 and Polyflux 170H showed comparable peak levels of C3a and C5a (measured at 15 minutes, P=0.68 and P=0.82, respectively). Peak levels of sC5b-9 (measured at 60 minutes) were higher for the Polyflux 170H than for the FX CorAL 600 (P=0.02).

F2
Figure 2.:
Course of plasma levels of complement factors (A) C3a, (B) C5a, and (C) sC5b-9 over the treatment by investigated dialyzer. Means±SD are displayed. Pre, predialysis.
Table 5. - Least squares mean of differences between dialyzers over time of treatment, comparing the FX CorAL 600 versus SureFlux 17UX
LSMean, µg/L (95% CI) P Value
Parameter Pre 15 Min 60 Min 240 Min Pre 15 Min 60 Min 240 Min
C3a 6.66 (−21.07 to −34.40) 61.96 (34.23 to 89.70) 27.45 (−0.28 to 55.19) 21.42 (−6.31 to 49.16) 0.64 <0.001 0.05 0.13
C5a −0.011 (−0.039 to 0.017) 0.033 (0.005 to 0.061) −0.042 (−0.070 to −0.014) −0.005 (−0.033 to 0.023) 0.45 0.02 0.004 0.72
sC5b-9 1.07 (−18.09 to 20.23) 66.27 (47.12 to 85.43) 58.01 (38.86 to 77.17) 30.71 (11.56 to 49.87) 0.91 <0.001 <0.001 0.002
LSMean, least squares mean; 95% CI, 95% confidence interval of LSMean; pre, predialysis; 15 min, 15 minutes after start of HDF; 60 min, 60 minutes after start of HDF; 240 min, at the end of HDF; HDF, hemodiafiltration.

Table 6. - Least squares mean of differences between dialyzers over time of treatment, comparing the FX CorAL 600 versus Polyflux 170H
LSMean, µg/L (95% CI) P Value
Parameter Pre 15 Min 60 Min 240 Min Pre 15 Min 60 Min 240 Min
C3a 6.36 (−14.39 to 27.10) 4.33 (−16.58 to 25.24) −12.53 (−33.27 to 8.21) 2.72 (−18.02 to 23.47) 0.55 0.68 0.24 0.80
C5a −0.029 (−0.115 to 0.057) −0.010 (−0.097 to 0.076) 0.068 (−0.017 to 0.154) −0.008 (−0.094 to 0.077) 0.51 0.82 0.12 0.85
sC5b-9 0.54 (−14.97 to 16.04) 11.78 (−3.83 to 27.39) 18.25 (2.75 to 33.75) 13.33 (−2.17 to 28.83) 0.95 0.14 0.02 0.09
LSMean, least squares mean; 95% CI, 95% confidence interval of LSMean; pre, predialysis; 15 min, 15 minutes after start of HDF; 60 min, 60 minutes after start of HDF; 240 min, at the end of HDF; HDF, hemodiafiltration.

Cell Activation

Leukocytes showed a typical transient decrease at 15 minutes after treatment start, which recovered at 60 minutes, and decreased again toward the end of the treatment (Figure 3). The mean decrease at 15 minutes was significantly less with the FX CorAL 600 than with both other dialyzers (P<0.001 versus both the SureFlux 17UX and Polyflux 170H; Table 7).

F3
Figure 3.:
Course of leukocyte count over the treatment by investigated dialyzer. Means±SD are displayed. Pre, predialysis.
Table 7. - Changes from baseline to 15 minutes (leukocytes) and 240 minutes for other markers of cell and coagulation activation, and contact activation
LSMean P Value
Parameter FX CorAL 600 SureFlux 17UX Polyflux 170H Overall A versus B A versus C B versus C
Change after 15 min
 Leukocytes (109/L) −0.38 −0.87 −0.78 <0.001 <0.001 <0.001 0.40
Change after 240 min
 PMN elastase (µg/L) 16.33 19.34 25.83 0.002 0.26 <0.001 0.02
 Platelets (109/L) −15.3 −10.5 −23.0 <0.001 0.05 0.002 <0.001
 TAT (µg/L) 6.90 5.06 4.02 0.81 0.68 0.52 0.82
 Kallikrein (U/L) 5.13 11.09 10.85 0.16 0.09 0.11 0.95
LSMean, least squares mean; A, FX CorAL 600; B, SureFlux 17UX; C, Polyflux 170H; PMN, polymorphonuclear; TAT, thrombin-antithrombin III.

Polymorphonuclear elastase increased in the first 60 minutes with all three dialyzers, and then remained stable for the rest of the treatment (Supplemental Figure 1). The increase was highest with the Polyflux 170H and, at treatment end, it proved to be significantly higher than with both of the other dialyzers (P<0.001 for comparison with FX CorAL 600; P=0.02 for comparison with the SureFlux 17UX; Table 7).

Platelets decreased slightly in the first 15 minutes and then remained nearly unchanged over the remainder of the treatment. With the Polyflux 170H, the strongest decrease was observed from baseline to 240 minutes (P=0.002 for comparison with the FX CorAL 600; P<0.001 for comparison with the SureFlux 17UX; Table 7).

TAT concentrations showed a slight increase with all three dialyzers throughout the treatment, with no statistically significant difference at 240 minutes (Table 7).

Kallikrein increased slightly throughout the treatment for all dialyzers. After 240 minutes, the FX CorAL 600 had a lower, but statistically nonsignificant, increase (5.13 U/L) than both the SureFlux 17UX (11.09 U/L, P=0.09) and the Polyflux 170H (10.85 U/L, P=0.11).

Adverse Events

For each dialyzer, a comparable number of patients experienced at least one adverse event (AE) (Table 8). In total, 22, 20, and 18 AEs occurred with the FX CorAL 600, the SureFlux 17UX, and the Polyflux 170H, respectively. Of these, one, 12, and eight AEs, respectively, have been classified by the investigator as possibly being related to the dialyzer applied in the respective study phase. One serious AE was reported during the phase with the FX CorAL 600 and SureFlux 17UX, and one further serious AE (shunt thrombosis) occurred before any study-related intervention; neither event was related to the medical procedure or the dialyzer. Most AEs resolved without sequelae; four were ongoing at study termination.

Table 8. - Adverse events
Characteristics of Adverse Events FX CorAL 600 (N=69) SureFlux 17UX (N=69) Polyflux 170H (N=68)
Number of patients with AE in the respective dialyzer phase 14 13 11
Number of AEs in the respective dialyzer phase 22 20 18
Relation of AE to medical procedure, definitively/possibly/not related 0/4/18 1/13/6 0/13/5
Relation of AE to medical device, definitively/possibly/not related 0/1/21 0/12/8 0/8/10
Serious adverse events 1
(road traffic accident)
1
(cystitis)
0
Symptoms of AEs definitively or possibly related to medical device
 Anemia 1
 Palpitations 1
 Flatulence 1
 Asthenia 1
 Feeling hot 1
 Muscle spasms (cramps) 1 1
 Myalgia 2
 Dizziness 2
 Headache 1 1 1
  Involuntary muscle contractions 1
 Thrombosis in device (filter clotting) 1
 Pruritus 1 4
Analysis performed in the safety population. One further serious AE (shunt thrombosis) occurred in week −1, before starting treatment with any study dialyzer; patient was withdrawn from study. One patient prematurely terminated the study before being treated with the Polyflux 170H. AE, adverse event.

Discussion

This randomized, controlled, multicenter trial investigated performance and hemocompatibility of the novel polysulfone-based, high-flux dialyzer, FX CorAL 600. Noninferiority of the removal rate of β2-microglobulin, the primary outcome parameter of the study, was confirmed for FX CorAL 600 versus both comparator dialyzers. Moreover, the removal rate of β2-microglobulin was significantly higher with the FX CorAL 600 than with the cellulose triacetate–based dialyzer, and was comparable with the Polyflux 170H dialyzer. This removal rate has been achieved with moderate blood flow rates of approximately 300 ml/min, and substitution flow rates of around 70 ml/min, and can certainly be further enhanced by applying higher blood flow rates and substitution flow rates, as observed in other studies investigating dialyzer performance (2324–25).

We compared the FX CorAL 600 with the commercially available, polyarylethersulfone-based Polyflux 170H dialyzer and the cellulose triacetate–based SureFlux 17UX, which is often used as an alternative in case of intolerance to synthetic membranes (26,27). All three dialyzers showed the typical pattern of a fast increase of the complement factors C3a and C5a, as measured 15 minutes after treatment start, which was significantly lower with both dialyzers with synthetic membranes compared with the dialyzer containing cellulose triacetate. These differences were more pronounced than previously published comparisons between polysulfone and cellulose triacetate dialyzers; however, the investigated membrane in the FX CorAL 600, and possibly also in the comparator membranes, differ from those studied earlier (2829–30). Our in vivo data are further supported by in vitro investigations confirming less activation of complement factors C3a, C5a, and sC5b-9 with the FX CorAL 600 than with the SureFlux 17UX (19). After the initial increase, C3a and C5a then declined throughout the remainder of the treatment, possibly due to convective elimination, because the molecular mass of C3a and C5a (9–11 kDa) (31,32) is similar to that of β2-microglobulin (11.2 kDa), which is effectively removed by high-flux dialyzers. On the other hand, downregulation within the patient (33) and metabolic processes may also contribute to the decrease over the course of the treatment.

For sC5b-9, the lowest increase, particularly at 60 minutes, was observed with the FX CorAL 600 compared with both the SureFlux 17UX and the Polyflux 170H dialyzer. Because intradialytic elimination likely plays no role, due to sC5b-9’s molecular mass of approximately 1000 kDa, this marker may better reflect the complement activation associated with high-flux dialyzers, such as those used in this study, than small molecules, such as C3a and C5a.

All investigated complement factors, and their intradialytic behavior, underscore better hemocompatibility of the FX CorAL 600 dialyzer with a polysulfone-based membrane when compared with the cellulose triacetate–based dialyzer and, in regards to sC5b-9, to the polyarylethersulfone-based dialyzer. The modification of the polysulfone membrane in the FX CorAL 600 dialyzer, by stabilization of the polyvinylpyrrolidone on the blood-side surface of the membrane, supports the formation of a more robust hydrophilic layer. The hydrophilic layer on the membrane surface enhances hemocompatibility of polysulfone membranes through improved antifouling ability, which yields less protein adsorption and lower coagulation activation (3435–36).

It has been observed that the transiently elevated plasma C3a and C5a levels shortly after treatment start are accompanied by a complementary reduction of peripheral blood leukocyte counts. The three dialyzers showed a similar overall pattern, but there was a significantly lower decrease with the FX CorAL 600 as compared with both other dialyzers, in line with previously published evidence (28). This observation supports an association of complement activation with pulmonary vascular leukocyte sequestration, and also suggests that leukocytes begin to reappear in the peripheral blood only when the complement activation diminishes (37).

The transient increase of complement factors and occurrence of leukopenia has not been reflected by acute symptoms during treatment in our study. However, previous studies showed that complement activation can be associated with various pathophysiologic processes, which may clinically manifest over a longer period of time, and with the development of cardiovascular diseases (38). In a nondialysis population, C5a and sC5b-9 were suggested to be involved in acute and chronic inflammation and in endothelial dysfunction (12). Similarly, complement C3 and its activation product, the anaphylatoxin C3a, was found to be associated with carotid intima–media thickness, ankle-arm BP index, and may, therefore, have a possibly pathophysiologic role in the development of cardiovascular diseases (39). The role that the dialysis membrane can play in inflammatory processes and possible long-term effects for patients on dialysis could not be assessed in this study with an exposure of only 1 week; however, this potential association warrants further study.

Polymorphonuclear elastase is a marker of cell activation and is released upon degranulation of leukocytes (40). With all investigated dialyzers, it increased in the first hour of the treatment—to the lowest extent with the FX CorAL 600, in line with the lowest leukocyte drop—and remained at the same level until the end of the treatment. The observation of an initial increase when complement factors also transiently rise, and of no further stimulation of leukocyte degranulation when complement levels decrease, would support the hypothesis of an activation by complement, although this is still a matter of debate (41).

The activation of coagulation was assessed by platelet count and TAT concentrations. In all cases, the decrease in platelet count was moderate, and less than that found in a study on dialyzers based on similar membrane materials (42). Similar platelet counts before dialysis in all phases hint at both postdialytic recovery of platelet counts and no long-term effect of the intradialytic decrease in platelet counts. TAT increased slightly and steadily during the treatment, as similarly observed in other studies (43), with no significant difference between dialyzers. Thus, some thrombin formation seems to be ongoing, despite systemic anticoagulation, which was applied equally in all three study phases. The absence of macroscopic clotting in the dialyzers (except one case, see Table 8) nevertheless confirms that the applied systemic anticoagulation has been sufficient, despite small increases of TAT, and that the coagulation activation potential of the dialyzers is low.

Kallikrein, a precursor of bradykinin, is generated upon contact activation, i.e., blood contact with negative charges on surfaces (44). There was no evidence of clinically relevant acute contact activation for any of the dialyzers. The FX CorAL 600 had the lowest (but not statistically significant) measured levels of contact activation versus both the SureFlux 17UX and Polyflux 170H. This tendency toward lower contact activation may be related to the lower surface charge observed with the FX CorAL 600 membrane compared with that in the other two dialyzers (19).

AEs were recorded irrespective of being related to the study medical devices or the medical procedure. Given the approximately 200 treatments that were performed with each dialyzer, <10% were affected by an AE. Thus, overall, few AEs occurred during the study; this must also be considered against the background that, in the HD population, intradialytic events that are related to the dialysis procedure, but also to the patients’ pathologic state, are well known and are reported in a wide range of frequencies (45,46). With the new FX CorAL 600, only one AE occurred (headache), which was rated by the investigator to be possibly related to the study dialyzer. No event of hypersensitivity to the new dialyzer was reported; therefore, on the basis of this study, the novel dialyzer can be rated as well tolerated.

The study has been designed to analyze possible acute effects of the membrane material during the dialysis treatment. Therefore, we cannot conclude on long-term effects of performance and hemocompatibility of the new dialyzer on patient outcomes, which requires further studies with longer-term application of this dialyzer.

The FX CorAL 600 dialyzer demonstrated noninferiority in terms of β-microglobulin removal rates compared with two commercially available dialyzers. It efficiently removes small- and medium-molecules during online HDF, shows a favorable hemocompatibility profile, and is associated with a low frequency of AEs in this study of a limited patient number and follow-up time. Clinical studies with longer observation times are warranted to address the effect on hemocompatibility markers associated with inflammatory processes and cardiovascular outcomes. These studies may provide further evidence supporting the use of the new dialyzer in a wide range of therapeutic options and long-term treatments.

Disclosures

A. Erlenkötter, A. Gauly, B. Griesshaber, J. Kennedy, L. Rauber, M. Stauss-Grabo, and A.M. Zawada report being full-time employees of Fresenius Medical Care, and report personal fees from Fresenius Medical Care, outside the submitted work. M. Kempkes-Koch reports receiving grants, personal fees, and nonfinancial support from Fresenius Medical Care, during the conduct of the study. All remaining authors have nothing to disclose.

Funding

None.

Acknowledgments

The authors gratefully acknowledge the support from the clinical staff in the participating centers, Anke Torp, Wolfgang Legl, Markus Görig, Inga Petersen, Christiane Rode, Sonja Fuchs, Anja Olfe, Ulrike Frenzel, Dr. Ann-Christin Pfannerstill, and Dr. Burkhard Floehr; and the technical assistance in the laboratory analyses from Isolde Werth, Heike Hoffmann, and Birgit Leusch.

The interpretation of study results and revision of this manuscript was conducted by all authors. The decision to submit this manuscript for publication was jointly made by all authors and the manuscript was confirmed to be accurate and approved by all authors.

Author Contributions

G. Ehlerding, M. Kempkes-Koch, J. Kennedy, L. Rauber, W. Ries, H. Schmidt-Gürtler, S. Wagner, and S. Zschätzsch were responsible for investigation; A. Erlenkötter and J. Kennedy were responsible for methodology; A. Erlenkötter and M. Stauss-Grabo conceptualized the study; A. Gauly wrote the original draft; A. Gauly and A. M. Zawada reviewed and edited the manuscript; and B. Griesshaber and M. Stauss-Grabo were responsible for project administration.

Supplemental Material

This article contains the following supplemental material online at http://kidney360.asnjournals.org/lookup/suppl/doi:10.34067/KID.0000302021/-/DCSupplemental.

Supplemental Appendix 1: Statistical analysis—Bonferroni Holm adjustment of linear mixed model.

Supplemental Table 1. Analytical methods of investigated parameters.

Supplemental Table 2. Formula for calculation of removal rates, plasma clearance and correction of laboratory parameters for hematocrit.

Supplemental Table 3. Least squares mean (LSMean) of removal rates and clearance of performance markers.

Supplemental Table 4. Least squares mean (LSMean) of absolute changes from treatment start to 15 min, 60 min, 240 min for C3a (A), C5a (B), sC5b-9 (C).

Supplemental Figure 1. Course of PMN elastase over the treatment by dialyzer.

References

1. United States Renal Data System. 2015 USRDS Annual Data Report. Epidemiology of kidney disease in the United States. National Institute of Diabetes and Digestive and Kidney Diseases. Chapter 13: International comparisons, Bethesda, MD, 2015. https://usrds.org/annual-data-report/previous-adrs/.Accessed October 22, 2020
2. Palmer SC, Rabindranath K, Craig JC, Roderick PJ, Locatelli F, Strippoli GF: High-flux versus low-flux membranes for end-stage kidney disease. Cochrane Database Syst Rev 2012: CD005016, 2012 https://doi.org/10.1002/14651858.CD005016.pub2
3. Cheung AK, Rocco MV, Yan G, Leypoldt JK, Levin NW, Greene T, Agodoa L, Bailey J, Beck GJ, Clark W, Levey AS, Ornt DB, Schulman G, Schwab S, Teehan B, Eknoyan G: Serum beta-2 microglobulin levels predict mortality in dialysis patients: Results of the HEMO study. J Am Soc Nephrol 17: 546–555, 2006 https://doi.org/10.1681/ASN.2005020132
4. Okuno S, Ishimura E, Kohno K, Fujino-Katoh Y, Maeno Y, Yamakawa T, Inaba M, Nishizawa Y: Serum beta2-microglobulin level is a significant predictor of mortality in maintenance haemodialysis patients. Nephrol Dial Transplant 24: 571–577, 2009 https://doi.org/10.1093/ndt/gfn521
5. Maduell F, Navarro V, Cruz MC, Torregrosa E, Garcia D, Simon V, Ferrero JA: Osteocalcin and myoglobin removal in on-line hemodiafiltration versus low- and high-flux hemodialysis. Am J Kidney Dis 40: 582–589, 2002 https://doi.org/10.1053/ajkd.2002.34918
6. Peters SA, Bots ML, Canaud B, Davenport A, Grooteman MP, Kircelli F, Locatelli F, Maduell F, Morena M, Nubé MJ, Ok E, Torres F, Woodward M, Blankestijn PJ; HDF Pooling Project Investigators: Haemodiafiltration and mortality in end-stage kidney disease patients: A pooled individual participant data analysis from four randomized controlled trials. Nephrol Dial Transplant 31: 978–984, 2016 https://doi.org/10.1093/ndt/gfv349
7. Maduell F, Moreso F, Pons M, Ramos R, Mora-Macià J, Carreras J, Soler J, Torres F, Campistol JM, Martinez-Castelao A; ESHOL Study Group: High-efficiency postdilution online hemodiafiltration reduces all-cause mortality in hemodialysis patients. J Am Soc Nephrol 24: 487–497, 2013 https://doi.org/10.1681/ASN.2012080875
8. Grooteman MP, van den Dorpel MA, Bots ML, Penne EL, van der Weerd NC, Mazairac AH, den Hoedt CH, van der Tweel I, Lévesque R, Nubé MJ, ter Wee PM, Blankestijn PJ; CONTRAST Investigators: Effect of online hemodiafiltration on all-cause mortality and cardiovascular outcomes. J Am Soc Nephrol 23: 1087–1096, 2012 https://doi.org/10.1681/ASN.2011121140
9. Ok E, Asci G, Toz H, Ok ES, Kircelli F, Yilmaz M, Hur E, Demirci MS, Demirci C, Duman S, Basci A, Adam SM, Isik IO, Zengin M, Suleymanlar G, Yilmaz ME, Ozkahya M; Turkish Online Haemodiafiltration Study: Mortality and cardiovascular events in online haemodiafiltration (OL-HDF) compared with high-flux dialysis: Results from the Turkish OL-HDF Study. Nephrol Dial Transplant 28: 192–202, 2013 https://doi.org/10.1093/ndt/gfs407
10. Hakim RM: Clinical implications of biocompatibility in blood purification membranes. Nephrol Dial Transplant 15[Suppl 2]: 16–20, 2000 https://doi.org/10.1093/ndt/15.suppl_1.16
11. Hakim RM: Clinical sequelae of complement activation in hemodialysis. Clin Nephrol 26[Suppl 1]: S9–S12, 1986
12. Hertle E, van Greevenbroek MM, Arts IC, van der Kallen CJ, Feskens EJ, Schalkwijk CG, Stehouwer CD: Complement activation products C5a and sC5b-9 are associated with low-grade inflammation and endothelial dysfunction, but not with atherosclerosis in a cross-sectional analysis: the CODAM study. Int J Cardiol 174: 400–403, 2014 https://doi.org/10.1016/j.ijcard.2014.04.057
13. Martin-Malo A, Castillo D, Castro M, Pérez R, Rios A, Jaraba M, Aljama P: Biocompatibility of dialysis membranes: A comparative study. Nephrol Dial Transplant 6[Suppl 2]: 55–58, 1991
14. Poppelaars F, Faria B, Gaya da Costa M, Franssen CFM, van Son WJ, Berger SP, Daha MR, Seelen MA: The complement system in dialysis: A forgotten story? Front Immunol 9: 71, 2018 https://doi.org/10.3389/fimmu.2018.00071
15. Poppelaars F, Gaya da Costa M, Faria B, Berger SP, Assa S, Daha MR, Medina Pestana JO, van Son WJ, Franssen CFM, Seelen MA: Intradialytic complement activation precedes the development of cardiovascular events in hemodialysis patients. Front Immunol 9: 2070, 2018 https://doi.org/10.3389/fimmu.2018.02070
16. Lines SW, Richardson VR, Thomas B, Dunn EJ, Wright MJ, Carter AM: Complement and cardiovascular disease--The missing link in haemodialysis patients. Nephron 132: 5–14, 2016 https://doi.org/10.1159/000442426
17. Hakim RM, Held PJ, Stannard DC, Wolfe RA, Port FK, Daugirdas JT, Agodoa L: Effect of the dialysis membrane on mortality of chronic hemodialysis patients. Kidney Int 50: 566–570, 1996 https://doi.org/10.1038/ki.1996.350
18. Chenoweth DE: Complement activation produced by biomaterials. Artif Organs 12: 508–510, 1988 https://doi.org/10.1111/j.1525-1594.1988.tb02815.x
19. Melchior P, Erlenkötter A, Zawada AM, Delinski D, Schall C, Stauss-Grabo M, Kennedy J: Complement activation by dialysis membranes and its association with secondary membrane formation and surface charge [published online ahead of print December 16, 2020]. Artif Organs10.1111/aor.13887
20. Santos García A, Macías Carmona N, Vega Martínez A, Abad Estébanez S, Linares Grávalos T, Aragoncillo Sauco I, Verdalles Guzmán U, Panizo González N, Cruzado Vega L, López Gómez JM: Removal capacity of different high-flux dialyzers during postdilution online hemodiafiltration. Hemodial Int 23: 50–57, 2019 https://doi.org/10.1111/hdi.12685
21. Maduell F, Ojeda R, Arias-Guillén M, Fontseré N, Vera M, Rodas L, Gómez M, Huablocho KP, Esquivel F, Mori PD, Hoffmann V, Ugalde J, Rico N: A new generation of cellulose triacetate suitable for online haemodiafiltration. Nefrologia 38: 161–168, 2018 https://doi.org/10.1016/j.nefro.2017.03.011
22. Holm S: A simple sequentially rejective multiple test procedure. Scand Stat Theory Appl 6: 65–70, 1979
23. Maduell F, Arias-Guillen M, Fontseré N, Ojeda R, Rico N, Vera M, Elena M, Bedini JL, Wieneke P, Campistol JM: Elimination of large uremic toxins by a dialyzer specifically designed for high-volume convective therapies. Blood Purif 37: 125–130, 2014 https://doi.org/10.1159/000358214
24. Maduell F, Ojeda R, Belmar L, Munguía P, Sango C, Martinez-Díaz AI, Arias-Guillén M, Vera M, Fontseré N, Gómez M, Rodas L, Rico N: Evaluation of the dialyser inner diameter in online haemodiafiltration. Nefrologia 38: 34–40, 2018 https://doi.org/10.1016/j.nefro.2017.03.013
25. Potier J, Bowry S, Canaud B: Clinical performance assessment of CorDiax filters in hemodialysis and hemodiafiltration. Contrib Nephrol 189: 237–245, 2017 https://doi.org/10.1159/000450810
26. Martin-Navarro J, Esteras R, Castillo E, Carriazo S, Fernández-Prado R, Gracia-Iguacel C, Mas Fontao S, Ortíz A, González-Parra E: Reactions to synthetic membranes dialyzers: Is there an increase in incidence? Kidney Blood Press Res 44: 907–914, 2019 https://doi.org/10.1159/000501035
27. Esteras R, Martín-Navarro J, Ledesma G, Fernández-Prado R, Carreño G, Cintra M, Cidraque I, Sanz I, Tarragón B, Alexandru S, Milla M, Astudillo E, Sánchez E, Mas S, Tejeiro RD, Ortiz A, Sánchez R, González-Parra E: Incidence of hypersensitivity reactions during hemodialysis. Kidney Blood Press Res 43: 1472–1478, 2018 https://doi.org/10.1159/000493662
28. Grooteman MP, Nubé MJ, van Limbeek J, van Houte AJ, Daha MR, van Geelen JA: Biocompatibility and performance of a modified cellulosic and a synthetic high flux dialyzer. A randomized crossover comparison between cellulose triacetate and polysulphon. ASAIO J 41: 215–220, 1995 https://doi.org/10.1097/00002480-199541020-00018
29. Grooteman MPC, Nubé MJ, Daha MR, Van Limbeek J, van Deuren M, Schoorl M, Bet PM, Van Houte AJ: Cytokine profiles during clinical high-flux dialysis: No evidence for cytokine generation by circulating monocytes. J Am Soc Nephrol 8: 1745–1754, 1997
30. Zhang DL, Liu J, Cui WY, Ji DY, Zhang Y, Liu WH: Differences in bio-incompatibility among four biocompatible dialyzer membranes using in maintenance hemodialysis patients. Ren Fail 33: 682–691, 2011 https://doi.org/10.3109/0886022X.2011.589943
31. Sahu A, Lambris JD: Structure and biology of complement protein C3, a connecting link between innate and acquired immunity. Immunol Rev 180: 35–48, 2001 https://doi.org/10.1034/j.1600-065X.2001.1800103.x
32. Fernandez HN, Hugli TE: Primary structural analysis of the polypeptide portion of human C5a anaphylatoxin. Polypeptide sequence determination and assignment of the oligosaccharide attachment site in C5a. J Biol Chem 253: 6955–6964, 1978 https://doi.org/10.1016/S0021-9258(17)38013-4
33. Chenoweth DE, Cheung AK, Henderson LW: Anaphylatoxin formation during hemodialysis: effects of different dialyzer membranes. Kidney Int 24: 764–769, 1983 https://doi.org/10.1038/ki.1983.225
34. Hayama M, Yamamoto K-i, Kohori F, Sakai K: How polysulfone dialysis membranes containing polyvinylpyrrolidone achieve excellent biocompatibility? J Memb Sci 234: 41–49, 2004 https://doi.org/10.1016/j.memsci.2004.01.020
35. Wang H, Yu T, Zhao C, Du Q: Improvement of hydrophilicity and blood compatibility on polyethersulfone membrane by adding polyvinylpyrrolidone. Fibers Polym 10: 1–5, 2009 https://doi.org/10.1007/s12221-009-0001-4
36. Zhu L, Song H, Zhang D, Wang G, Zeng Z, Xue Q: Negatively charged polysulfone membranes with hydrophilicity and antifouling properties based on in situ cross-linked polymerization. J Colloid Interface Sci 498: 136–143, 2017 https://doi.org/10.1016/j.jcis.2017.03.055
37. Chenoweth DE: Complement activation during hemodialysis: Clinical observations, proposed mechanisms, and theoretical implications. Artif Organs 8: 281–290, 1984 https://doi.org/10.1111/j.1525-1594.1984.tb04291.x
38. Hertle E, Stehouwer CD, van Greevenbroek MM: The complement system in human cardiometabolic disease. Mol Immunol 61: 135–148, 2014 https://doi.org/10.1016/j.molimm.2014.06.031
39. Hertle E, van Greevenbroek MM, Arts IC, van der Kallen CJ, Geijselaers SL, Feskens EJ, Jansen EH, Schalkwijk CG, Stehouwer CD: Distinct associations of complement C3a and its precursor C3 with atherosclerosis and cardiovascular disease. The CODAM study. Thromb Haemost 111: 1102–1111, 2014 https://doi.org/10.1160/TH13-10-0831
40. Krieter DH, Morgenroth A, Barasinski A, Lemke HD, Schuster O, von Harten B, Wanner C: Effects of a polyelectrolyte additive on the selective dialysis membrane permeability for low-molecular-weight proteins. Nephrol Dial Transplant 22: 491–499, 2007 https://doi.org/10.1093/ndt/gfl610
41. Stefoni S, Colì L, Cianciolo G, Donati G, Ruggeri G, Ramazzotti E, Pohlmeier R, Lang D: Inflammatory response of a new synthetic dialyzer membrane. A randomised cross-over comparison between polysulfone and helixone. Int J Artif Organs 26: 26–32, 2003 https://doi.org/10.1177/039139880302600105
42. Ojeda R, Arias-Guillén M, Gómez M, Vera M, Fontseré N, Rodas L, Filella X, Reverter JC, Lozano F, Villamor N, Maduell F: Study of biocompatibility of membranes in online hemodiafiltration. Blood Purif 49: 400–408, 2020 https://doi.org/10.1159/000504954
43. François K, Orlando C, Jochmans K, Cools W, De Meyer V, Tielemans C, Wissing KM: Hemodialysis does not induce detectable activation of the contact system of coagulation. Kidney Int Rep 5: 831–838, 2020 https://doi.org/10.1016/j.ekir.2020.03.010
44. Renaux JL, Thomas M, Crost T, Loughraieb N, Vantard G: Activation of the kallikrein-kinin system in hemodialysis: Role of membrane electronegativity, blood dilution, and pH. Kidney Int 55: 1097–1103, 1999 https://doi.org/10.1046/j.1523-1755.1999.0550031097.x
45. Holley JL: A descriptive report of errors and adverse events in chronic hemodialysis units. Nephrol News Issues 20: 57–58, 60–51, 63 passim, 2006
46. Arenas Jiménez MD, Ferre G, Álvarez-Ude F: Strategies to increase patient safety in Hemodialysis: Application of the modal analysis system of errors and effects (FEMA system). Nefrologia 37: 608–621, 2017
Keywords:

dialysis; β2-microglobulin removal rate; complement activation; dialyzer; hemocompatibility; leukocytes; polymers

Copyright © 2021 by the American Society of Nephrology