INTRODUCTION
Total body irradiation (TBI) is a specialized form of magna field radiation therapy where a relatively homogeneous dose of radiation is delivered to the entire body as part of a conditioning regimen before hematopoietic stem cell transplant. TBI serves a dual purpose in a conditioning regimen; immunosuppression and cytotoxicity.[ 1–3 ] It aims to destroy all malignant cells in the host body, prepare niches in the bone marrow to receive the recipient stem cells and finally suppress the host immunity to prevent the immune system from rejecting the donor stem cells that are being transplanted.[ 4 ] The unique advantage of TBI over conditioning regimens based on cytotoxic chemotherapy is that it has better tissue penetration since it is not dependent on either blood supply or hepatic/renal parameters; hence, even sanctuary sites such as central nervous system and testes are not spared.[ 5–8 ] The disadvantages associated with TBI are risks associated with acute and late toxicities including second malignancy which contributes to increased morbidity and mortality, need for specialized equipment, and challenging logistics of administration.[ 9 , 10 ] Recent developments have been focusing on simplifying treatment delivery, decreasing the risk of toxicity while achieving better efficacy.
The target volume for the TBI treatment is the whole body including the skin, the skin can get an adequate dose using the beam spoiler kept at the distance from the 10–20 cm distance from the patient surface.[ 11 ] Many studies were done to achieve homogeneous dose distribution to the whole body.[ 12–14 ] TBI treatment can be delivered conventionally with different techniques such as extended source to surface distance (SSD) with anterior to posterior/posterior to anterior (AP/PA) technique and extended SSD with bilateral treatment technique, which can be either compensator-based (CB) TBI or field-in-field (FIF) TBI. The centers which do not have the facility to perform CB TBI or FIF TBI, deliver TBI through an open field. Problem with bilateral TBI technique is inhomogeneous dose distribution due to the varying thickness of the patient from the lateral view. To overcome this inhomogeneous dose distribution, CB bilateral TBI technique and FIF TBI technique was used. The disadvantage of CB TBI is that the treatment time will be more compared to FIF TBI technique due to placing the compensator in the appropriate place, calculating compensator thickness, and binding the compensators. In vivo dosimetry is the recommended procedure to check the dose at different regions of the patient during TBI treatment technique. Various in vivo dosimeters can be used to measure the dose received at different places in the patient; such as film, TLD chips, Metal–oxide–semiconductor-field-effect-transmitter (MOSFET), and silicon diodes.[ 15–17 ]
FIF-TBI treatment planning is done with the help of treatment planning system (TPS). Using the computed tomography (CT) images and the TPS treatment planning will be done with FIF technique. Using computerized FIF technique, homogeneous dose distribution can be achieved within the tolerance limit Different studies have showed that the FIF technique can be successfully employed to achieve homogeneous distribution.[ 18–20 ] Many centers do not have high-end treatment machine to plan TBI with Tomotherapy or volumetric-modulated arc therapy (VMAT) treatment technique. The limitation of dose calculation at extended SSD using TPS is also another issue. At our center, we faced these issues. This novel technique of MFIF-TBI will be useful for centers with limited resources. The manual field-in-field TBI (MFIF-TBI) can be implemented using multi-leaf collimator (MLC) and jaws to overcome the TPS dose calculation limitation and to avoid time spent on the calculation of compensator thickness and for binding compensators of different regions for each patient. This study aims to implement a novel treatment approach for TBI using MFIF-TBI technique and its dosimetric results were compared with CB-TBI and open field TBI technique.
MATERIALS AND METHODS
Measurement setup
The Elekta™ Synergy Platform treatment machine was used for the analysis of midplane dose for different methods used in the bilateral TBI techniques. The rice flour phantom (RFP) was prepared with the rice flour in the form of the human structure. The shape of the different human anatomy was created separately like skull was made using a single bag; the neck was made using another bag; and so on. All parts were then kept together to form the shape of a human body. Figure 1 depicts the photograph of RFP. The phantom was placed in the supine position with bent knees on the motorized treatment couch at the extended SSD of 385 cm from the source. The percentage depth dose (PDD), half value layer, and beam flatness were measured at 385 cm SSD using gantry angle 270°, field size 40 cm × 40 cm, collimator angle 45°, and 6 MV high energy X-ray.[ 11 ] While the MFIF TBI measurements were done with gantry angle 270°, field size 40 cm × 40 cm, collimator angle 0° and 6 MV high energy X-ray, the CB-TBI measurements were done with all parameters remaining the same except for the collimator, with a rotation of 45°to achieve a larger field size. However, using collimator angle 45° to create FIF will be very difficult and it cannot be manageable. To avoid this issue, the measurements were taken at 0° collimator angle for MFIF TBI. An acrylic sheet beam spoiler with 1 cm thickness was used between phantom surface and source.
Figure 1: The photograph of RFP. RFP: Rice flour phantom
In vivo dosimetry
In vivo dosimetry is necessary to measure the delivered dose to TBI treatment patients.[ 17 , 21–23 ] In this study, we used in vivo silicon diodes (ISORAD™ from Sun Nuclear Corporation) to measure the dose. The silicon diode detectors have a diameter of 10 mm with adequate electronic equilibrium to measure the dose. Before each measurement, the in vivo diodes were calibrated to measure accurately 1 monitor unit (MU) equal to 1 cGy at 100 cm SSD. Each diode was calibrated separately with gantry angle 0°, collimator angle 0°, 6 MV energy, and 10 cm × 10 cm field size at 98.5 cm SSD. During calibration, 10 cm thickness of polymethyl methacrylate phantom was placed below the diodes to account for backscatter and the diodes were placed perpendicular to the central axis of the beam. The calibration factor for each diode is stored separately in the software system. During the dose measurement, the calibration factor was included by the software for that particular diode.
Manual field-in-field total body irradiation delivery
The RFP was placed on the TBI couch with the knee bent position at 385 cm SSD. The length of the RFP, umbilicus (isocenter) to skull, umbilicus (isocenter) to feet end distances were measured. The machine isocenter to umbilicus distance, machine isocenter to RFP Knee distance, machine isocenter to RFP ankle distance and machine isocenter to RFP skull distance were measured. The separation was measured using caliper at different places in the RFP such as skull, umbilicus, and calf. Using the above separation, the midplane depth (MPD) was calculated for skull, umbilicus and calf regions. Figure 2 shows the number of sub-fields used in the MFIF-TBI technique. There are three fields – one primary field and two subfields. The different subfield size was opened for different regions using the MLC and jaws.
Figure 2: The number of sub-fields used in the MFIF-TBI technique. MFIF-TBI: Manual field in field-Total body irradiation
To calculate the projected field size at extended SSD 385 cm, the required field size at standard SSD was calculated using the following formula:
where,
rd is the projected Field size at the extended SSD,
r is the field size at the standard SSD 100 cm,
f is the standard SSD,
d is the distance from isocenter to the patient surface at extended SSD.
The treatment MU was calculated for three different fields using the MPD of skull, calf, and umbilicus. The primary open field for the entire length of the patient and MU was calculated using MPD of skull. The primary field treatment MU (PFT MU) was calculated using the following formula.
where,
output is the dose in cGy/MU at the depth of maximum dose for a field size of 40 × 40 cm2 and PDD is the percentage depth dose at mid-plane depth at the skull level.
The first subfield cranial border starts at lower neck and its caudal border ends at covering full calf region. The field size length of the patient from isocenter (umbilicus) to lower neck distance and isocenter (umbilicus) to the end of the calf region was measured. To project the field size at extended SSD, the standard field size was calculated using the formula (1). The treatment MU for first subfield was calculated using MPD of calf. The first subfield treatment MU (FST MU) was calculated using the following formula.
where,
PDD is the percentage depth dose at calf depth.
PFT MU is the primary field treatment MU calculated with skull depth.
The second subfield cranial border starts at lower neck and its caudal border ends at the thigh. The field size length of the patient from isocenter (umbilicus) to lower neck distance and isocenter (umbilicus) to end of the thigh region was measured. To project the field size at extended SSD, the standard field size was calculated using the formula (1). The treatment MU for the second subfield was calculated using MPD of umbilicus. The second subfield treatment MU (SST MU) was calculated using the following formula.
where,
PDD is the percentage depth dose at umbilicus depth.
PFT MU is the primary field treatment MU calculated with skull depth.
FST MU is the first subfield treatment MU calculated with calf depth.
The RFP was irradiated with the calculated MU for different field sizes by MFIF-TBI technique. The prescribed dose for this study was 200 cGy in 1 fraction with each side receiving 100 cGy to the midplane of the RFP. The in vivo diodes were placed on the surface of different regions of RFP such as skull, umbilicus, thigh, calf, knee, neck, and ankle to measure the irradiated dose at Dmax. Using the measured Dmax dose the midplane dose for different regions was calculated.
To calculate the midplane dose the following formula was used:[ 24 ]
This formula was modified as following
Midplane dose of given region
Using the above formula the midplane dose of different regions was calculated. The calculated midplane dose was compared with the planned midplane dose at different regions. Similarly, the couch was rotated to irradiate opposite side of the RFP with the calculated MU and field size. To show diode measurement accuracy, the same was repeated 5 times and the diode readings were noted. To measure the reproducibility of MFIF measurements using RFP, each rice flour bags were removed and the RFP bags were repositioned and realigned along with diodes, and the measurements was repeated. The entire process was repeated twice.
Compensator-based total body irradiation delivery
The RFP was placed on the TBI couch with the knee bent position at 385 cm SSD. The separation was measured using caliper at different places in the RFP such as skull, neck, umbilicus, calf, chest, knee, shoulder, thigh, and ankle. Using the above separation, the MPD was calculated for all the regions. Using the MPD of the different regions the required Perspex compensator thickness was calculated to compensate the irregular thickness of the patient separation in different regions such as skull, neck, knee, calf, and ankle.[ 11 , 25 ] Calculated compensator thickness was positioned at the appropriate places. Figure 3 shows the CB TBI treatment setup. The treatment MU for CB-TBI was calculated using the MPD of umbilicus region.
Figure 3: The compensator-based TBI treatment setup. TBI: Total body irradiation
where,
output is the dose in cGy/MU at depth of Dmax for a field size of 40 cm × 40 cm and PDD is the percentage depth dose at the umbilicus level.
The prescribed dose for CB-TBI study was 200 cGy in 1 fraction. Each side of the RFP was irradiated with the calculated MU for 100 cGy to the midplane of the RFP and the Dmax dose was measured at different regions using in vivo diodes. Using the measured Dmax dose, the mid-plane dose for different regions was calculated for CB-TBI technique using the formula (5). The calculated mid-plane dose was compared with the planned mid-plane dose at different region. Similarly, the couch was rotated to irradiate opposite side of the phantom with CB-TBI technique.
Open field total body irradiation delivery
The RFP was placed on the TBI couch with the knee bent position at 385 cm SSD. The separation was measured using caliper at different places in the RFP such as skull, neck, umbilicus, calf, chest, knee, shoulder, thigh, and ankle. Using the above separation, the MPD was calculated for the umbilicus region. Using the MPD of umbilicus region the treatment MU for open field TBI was calculated with the formula (6). The prescribed dose for open field TBI study was 200 cGy in 1 fraction. Each side of the RFP was irradiated with the calculated MU for 100 cGy to the midplane of the RFP and the Dmax dose was measured at different regions using in vivo diodes. Using the measured Dmax dose the mid-plane dose for different regions was calculated for open field TBI technique using the formula (5). The calculated mid-plane dose was compared with the planned mid-plane dose for different region. Similarly, the couch was rotated to irradiate opposite side of the phantom with open field TBI technique.
RESULTS
Manual field-in-field total body irradiation results
The length of the RFP with knee bent position, umbilicus (isocenter) to skull, umbilicus (isocenter) to feet end distances were verified with the measured distance. The machine isocenter to umbilicus, machine isocenter to RFP knee, machine isocenter to RFP ankle distance, and machine isocenter to RFP skull distances were verified with the measured distance. Verification of these distances will reduce the positional errors and MLC and jaw accuracy in the MFIF technique. The time taken for RFP position and verification of distance was 6 min for each side. The field sizes of the three subfields of MFIF TBI were measured on RFP at extended SSD and the actual field sizes at standard SSD were calculated using the formula (1). Table 1 shows the standard field size at isocenter and calculated field size (projected field size) at extended SSD. The RFP was irradiated with the calculated MU for different subfields. Table 2 shows the diode measurement and percentage of deviation for MFIF technique. The deviations from the prescribed dose for umbilicus, skull, neck, thigh, knee, calf, and ankle were − 2.22%, −1.64%, +8.72%, −0.28%, −1.26%, −1.85%, −0.99%, respectively, for MFIF TBI. Figure 4 shows the percentage of deviation for different regions of the RFP in MFIF technique. The total time taken for MFIF TBI delivery was 30 min. The same measurements were repeated five times to show the accuracy of diode measurements. The accuracy of diode measurements was within ±2% deviation. Table 3 shows the accuracy of diode measurement repeated over five times with the MFIF TBI technique. Figure 5 shows the diode measurement uncertainty error bar obtained by repeating the irradiation with MFIF-TBI technique. To measure the uncertainty of the measurement, the RFP was repositioned two times. The percentage of deviation between each repositioning was within ± 2% deviation. Table 4 shows the measurement uncertainty after repositioning of RFP in the MFIF TBI technique. Figure 6 shows the measurement uncertainty error bar obtained by repositioning the RFP in the MFIF-TBI technique.
Table 1: The standard field size at isocenter and calculated field size (projected field size) at extended source to surface distance
Table 2: The diode measurement and percentage of deviation for the manual field in field technique
Figure 4: The percentage of deviation for different regions of the RFP with respect to different techniques in bilateral TBI. TBI: Total body irradiation. RFP: Rice flour phantom
Table 3: The accuracy of diode measurement repeated over five times with the manual field in field total body irradiation technique
Figure 5: The diode measurement uncertainty error bar obtained by repeating irradiation with MFIF-TBI technique. MFIF-TBI: Manual field-in-field-Total body irradiation
Table 4: The measurement uncertainty after repositioning of rice flour phantom in the manual field in field-total body irradiation technique
Figure 6: The measurement uncertainty error bar obtained by repositioning the RFP in the MFIF- TBI technique. MFIF-TBI: Manual field in field-Total body irradiation, RFP: Rice flour phantom
Compensator-based-total body irradiation results
The initial measurements of RFP were taken and verified similarly to the MFIF-TBI. Verification of these distances will reduce the positional error of compensator placing in the CB-TBI technique. Based on the separation of different regions, the Perspex compensator thickness was calculated.[ 11 ] The calculated compensator thickness was placed at the appropriate places. Time taken for this CB-TBI treatment positioning including initial measurements, verification of distance and compensator positioning was 10 min for each side. Table 5 shows the compensator thickness, diode measurement, and percentage of deviation for CB-TBI technique. The deviations from the prescribed dose for umbilicus, skull, neck, thigh, knee, calf, and ankle were -1.86, +1.92, +0.50, +0.09, +1.26, +1.69, -1.32, respectively, for CB-TBI. Figure 4 shows the percentage of deviation for different regions of the RFP in CB-TBI technique. The total time taken for CB-TBI delivery was 45 min.
Table 5: The compensator thickness, diode measurement and percentage of deviation for calculated based-total body irradiation technique
Open field total body irradiation results
The initial measurements of RFP were taken and verified similarly to the MFIF-TBI. The time taken for open field-TBI treatment positioning including initial measurements, verification of distance, and beam spoiler positioning was 6 min. The treatment MU was calculated using the MPD of umbilicus region. The calculated treatment MU was delivered only with beam spoiler. Table 6 shows the diode measurement reading for different regions and percentage of deviation for the open field-TBI technique. The deviations from the prescribed dose for umbilicus, skull, neck, thigh, knee, calf, and ankle were − 2.01%, +13.76%, +25.30%, −0.72%, +11.38%, +8.7%, +14.04%, respectively, for Open field TBI. Figure 4 shows the percentage of deviation for different regions of the RFP in open filed-TBI technique. The total time taken for open field TBI delivery was 25 min.
Table 6: The diode measurement reading for different regions and percentage of deviation for open field-total body irradiation technique
DISCUSSION
TBI plays a major role as a conditioning regimen in stem cell transplantation. The target volume for TBI treatment is the whole body. Due to the field size limitation, the extended SSD treatment was introduced for TBI treatment. Using extended SSD, TBI treatment can be delivered conventionally with AP/PA TBI technique and bilateral TBI technique. Various bilateral TBI techniques were studied and different techniques were implemented using RFP and these were dosimetrically verified.
The patient separation will not be uniform in bilateral TBI treatment from the lateral view. Due to nonuniform separation, dose received in different regions will not be uniform. To deliver a uniform dose to the nonuniform patient body, FIF technique and compensator was used in the recent studies.[ 18 , 20 ] A novel TBI treatment technique with simple and possible method using MFIF-TBI technique was implemented. Some centers have opted for open field TBI technique due to various reasons, one of which can be a the limitation of dose calculation in TPS for extended SSD as seen with Monaco TPS (version 5.51.10). In this study, we have performed the manual FIF-TBI without TPS, CB-TBI, and open-filed TBI. For MFIF-TBI, we calculated the standard field size needed to project the measured field size at extended SSD for different subfields. The treatment MU was calculated based on the separation of different MPD and the treatment was executed. The dose deviation in the different regions was well within acceptable limits using the MFIF-TBI technique. After repositioning the RFP, measurement uncertainty showed the percentage of deviation to be very less when compared with the initial measurements. The overall treatment time for MFIF-TBI was less compared with CB-TBI technique. The CB-TBI was performed after placing the compensator in the appropriate regions.[ 11 ] The results showed that the deviation was within the acceptable limit of ± 10% for CB-TBI as per the AAPM report no 17.[ 21 ] The advantages of treating MFIF-TBI technique are less patient set up time, calculation of compensator thickness and binding the compensator for each patient can be avoided, placement of compensator in the correct regions, and achieving dose deviation within the limit without TPS. The open field TBI treatment was executed without using compensator and FIF technique. The results were compared with planned dose at plane of the RFP. The percentage of deviation for open field TBI was not within ± 10% for skull, neck, ankle, and knee regions due to variation in separation of the human anatomy in the lateral view. The dose was prescribed to a point in the umbilicus region along the midplane of the body. While the umbilicus region receives 100% dose, it will vary for the other regions according to their separation. Therefore the dose deposited will not be uniform along the length of the body in open field TBI. The percentage of deviation for different regions of the RFP with respect to different technique in bilateral TBI. This study shows that without using TPS and compensators, homogeneous dose can be achieved with the help of MLC and jaw using MFIF-TBI technique.[ 19 ]
The disadvantage of this technique is the image verification of patient position. Imaging is not possible with extended SSD-based techniques. The accuracy of dose distribution in various organs and normal structures cannot be determined with this type of treatment technique. The lung shielding is an important part of the extended SSD TBI treatment technique. Whenever lung dose increases beyond the tolerance limit, the lung toxicity will also increase. Pandu et al . showed that the lung compensator thickness can be calculated to reduce the lung dose within the tolerance limit.[ 11 ] The same lung compensators calculation can be used in MFIF-TBI technique.[ 11 ] The dose received by different organs can be known with FIF TBI treatment delivery planned using TPS.[ 20 ] The dose distribution can be improved with the help of a TPS using FIF-TBI. Lately, to deliver a homogeneous dose, different techniques were tried to treat the TBI patients. Using VMAT and IMRT technology uniform dose can be achieved and other normal structure doses can also be reduced.[ 26–28 ] Limitation of using the standard SSD VMAT technique includes multiple overlapping of arcs, three to four isocenters, long duration of contouring, treatment planning, and treatment delivery (1.5 h or nearly 2 h during the first fraction) when compared with MFIF technique.[ 28 ] If the immobilization is not properly used, the patient will move during the treatment due to longer treatment time which will impact the dose distribution. The dose rate used to treat VMAT, IMRT, and tomotherapy technique is high compared to extended SSD Technique. Using low-dose rate for TBI treatment will reduce the serious complication and better patient outcome.[ 21 ] The AAPM report no 17 guidelines state that in vivo dosimetry is recommended for all the TBI patients to measure the accurate dose delivered. The in vivo dosimetry can be performed with MOSFET and diodes.[ 17 , 21 , 29 ] In this study, we used in vivo diodes for accurate dose measurement.
CONCLUSION
This study concludes that novel MFIF-TBI treatment delivery is an effective method in terms of less treatment time, proper in vivo dosimetry, easy to implement, and able to maintain uniform dose distribution under tolerance limit without TPS and compensator. The advantages of this technique are no TPS is required, can avoid the laborious process of making compensators. Furthermore, study results ensured that both the MFIF and CB-TBI method can be implemented for TBI treatment. Using VMAT and IMRT technology, the uniform dose can be achieved and other normal structure doses can also be reduced. The disadvantage of this type of technique includes patient immobilization, field junction between the different isocenters and increase in time taken for contouring, planning, and treatment execution when compared with MFIF technique. The center which does not have the facility to plan patient treatment with TPS can use MFIF-TBI technique as the conditioning regime for stem cell transplant.
Financial support and sponsorship
Nil.
Conflicts of interest
There are no conflicts of interest.
REFERENCES
1. Sabloff M, Tisseverasinghe S, Babadagli ME, Samant R. Total body irradiation for hematopoietic stem cell transplantation:What can we agree on?. Curr Oncol 2021;28:903–17.
2. Wong JY, Filippi AR, Dabaja BS, Yahalom J, Specht L. Total body irradiation:Guidelines from the International Lymphoma Radiation Oncology Group (ILROG). Int J Radiat Oncol Biol Phys 2018;101:521–9.
3. Gyurkocza B, Sandmaier BM. Conditioning regimens for hematopoietic cell transplantation:One size does not fit all. Blood 2014;124:344–53.
4. Hoeben BA, Wong JY, Fog LS, Losert C, Filippi AR, Bentzen SM, et al. Total body irradiation in haematopoietic stem cell transplantation for paediatric acute lymphoblastic leukaemia:Review of the literature and future directions. Front Pediatr 2021;9:774348.
5. Mikell JL, Waller EK, Switchenko JM, Rangaraju S, Ali Z, Graiser M, et al. Similar survival for patients undergoing reduced-intensity total body irradiation (TBI) versus myeloablative TBI as conditioning for allogeneic transplant in acute leukemia. Int J Radiat Oncol Biol Phys 2014;89:360–9.
6. Ben Hassine K, Powys M, Svec P, Pozdechova M, Versluys B, Ansari M, et al. Total body irradiation forever? Optimising chemotherapeutic options for irradiation-free conditioning for paediatric acute lymphoblastic leukaemia. Front Pediatr 2021;9:775485.
7. Hahn T, Wall D, Camitta B, Davies S, Dillon H, Gaynon P, et al. The role of cytotoxic therapy with hematopoietic stem cell transplantation in the therapy of acute lymphoblastic leukemia in children:An evidence-based review. Biol Blood Marrow Transplant 2005;11:823–61.
8. Tracey J, Zhang MJ, Thiel E, Sobocinski KA, Eapen M. Transplantation conditioning regimens and outcomes after allogeneic hematopoietic cell transplantation in children and adolescents with acute lymphoblastic leukemia. Biol Blood Marrow Transplant 2013;19:255–9.
9. Pearlman R, Hanna R, Burmeister J, Abrams J, Dominello M. Adverse effects of total body irradiation:A two-decade, single institution analysis. Adv Radiat Oncol 2021;6:100723.
10. Thomas O, Mahé M, Campion L, Bourdin S, Milpied N, Brunet G, et al. Long-term complications of total body irradiation in adults. Int J Radiat Oncol Biol Phys 2001;49:125–31.
11. Pandu B, Khanna D, Mohandass P, Ninan H, Elavarasan R, Jacob S, et al. Evaluation of surface dose and commissioning of compensator-based total body irradiation. J Med Phys 2022;47:173–80.
12. Papiez L, Montebello J, DesRosiers C, Papiez E. The clinical application of dynamic shielding and imaging in moving table total body irradiation. Radiother Oncol 1999;51:219–24.
13. Chui CS, Fontenla DP, Mullokandov E, Kapulsky A, Lo YC, Lo CJ. Total body irradiation with an arc and a gravity-oriented compensator. Int J Radiat Oncol Biol Phys 1997;39:1191–5.
14. Chen HH, Wu J, Chuang KS, Lin JF, Lee JC, Lin JC. Total body irradiation with step translation and dynamic field matching. Biomed Res Int 2013;2013:216034.
15. Ganapathy K, Kurup PG, Murali V, Muthukumaran M, Bhuvaneshwari N, Velmurugan J. Patient dose analysis in total body irradiation through
in vivo dosimetry. J Med Phys 2012;37:214–8.
16. Lancaster CM, Crosbie JC, Davis SR.
In-vivo dosimetry from total body irradiation patients (2000-2006):Results and analysis. Australas Phys Eng Sci Med 2008;31:191–5.
17. Mangili P, Fiorino C, Rosso A, Cattaneo GM, Parisi R, Villa E, et al.
In-vivo dosimetry by diode semiconductors in combination with portal films during TBI:Reporting a 5-year clinical experience. Radiother Oncol 1999;52:269–76.
18. van Leeuwen RG, Verwegen D, van Kollenburg PG, Swinkels M, van der Maazen RW. Early clinical experience with a total body irradiation technique using field-in-field beams and on-line image guidance. Phys Imaging Radiat Oncol 2020;16:12–7.
19. Yamamoto H, Sugimoto S, Furuya T, Hara N, Fukata K, Kojima H, et al. Anew method of total body irradiation using a field-in-field technique. Int J Radiat Oncol 2012;84:S852–3.
20. Onal C, Sonmez A, Arslan G, Sonmez S, Efe E, Oymak E. Evaluation of field-in-field technique for total body irradiation. Int J Radiat Oncol Biol Phys 2012;83:1641–8.
21. Dyk J. The physical aspects of total and half body photon irradiation. s AAPM report no 17. American Association of Physicists in Medicine. Med Phys New York American Institute of Physics 1986;17:22–4.
22. Vrtar M, Kovacević N. A model of
in vivo dosimetry and quality assurance analysis of total body irradiation in Zagreb. Acta Med Croatica 1998;52:15–26.
23. Ravichandran R, Binukumar JP, Davis CA, Sivakumar SS, Krishnamurthy K, Mandhari ZA, et al. Beam configuration and physical parameters of clinical high energy photon beam for total body irradiation (TBI). Phys Med 2011;27:163–8.
24. Faiz M, Khan JP. The Physics of Radiation Therapy 5th ed Philadelphia, PA Lippincott Williams &Wilkins/Wolters Kluwer 2014 144–5.
25. Ravichandran R, Binukumar JP, Davis CA, Al Rahbi Z, Balakrishnan R, Al Mandhari Z. Total body irradiation (TBI):Preliminary experience on clinical implementation. J Med Phys 2013;38:210–1.
26. Tas B, Durmus IF, Okumus A, Uzel OE, Gokce M, Goksoy HS, et al. Total-body irradiation using linac-based volumetric modulated arc therapy:Its clinical accuracy, feasibility and reliability. Radiother Oncol 2018;129:527–33.
27. Symons K, Morrison C, Parry J, Woodings S, Zissiadis Y. Volumetric modulated arc therapy for total body irradiation:A feasibility study using pinnacle (3) treatment planning system and Elekta Agility™linac. J Appl Clin Med Phys 2018;19:103–10.
28. Springer A, Hammer J, Winkler E, Track C, Huppert R, Böhm A, et al. Total body irradiation with volumetric modulated arc therapy:Dosimetric data and first clinical experience. Radiat Oncol 2016;11:46.
29. Eaton DJ, Warry AJ, Trimble RE, Vilarino-Varela MJ, Collis CH. Benefits of online
in vivo dosimetry for single-fraction total body irradiation. Med Dosim 2014;39:354–9.