Longer life spans and crystalline lens extractions in younger patients have increased the duration of intraocular lens (IOL) implantation in many patients. These changes are also increasing the number of patients with IOL dislocation, which can be treated by IOL extraction, vitrectomy, intrascleral IOL fixation, and suturing or a combination of these treatments. More recently, intrascleral fixation has become widely used for IOL fixation, and many reports have focused on the techniques and outcomes of this method.1–3 However, few reports have provided detailed information on IOL explantation, despite it being performed in most of these surgeries.4–7 Currently, there are 2 methods of IOL extraction: folding of the IOL into half with forceps, and cutting it into several pieces with scissors before removal.8 Both maneuvers must be performed within the narrow space of the anterior chamber and can cause corneal endothelial cell loss, iris injury, or anterior chamber hemorrhage.9–11 In addition, small debris that remains in the eye during IOL extraction can result in the loss of intraocular tissue, making IOL extraction a highly invasive procedure.
In recent years, cases of cataract surgery for refractive correction have increased. However, postoperative refractive error or dissatisfaction with multifocal IOL implantation outcomes can lead to a need for IOL extraction. Furthermore, when the purpose of primary surgery is refractive correction, good postoperative visual function can be achieved with a reoperation, which requires preservation of the lens capsule and a small incision for extracting the IOL to prevent induced astigmatism.
This study aimed to present the minimally invasive cartridge pull-through technique using a cartridge for IOL implantation and new forceps optimized for obtaining a secure grasp of the IOL for removal; the clinical application of the technique has also been presented. The cartridge used to extract the IOL is used off-label and was originally intended for the insertion of the IOL.
SURGICAL TECHNIQUE
The study and surgery were conducted in conformance with the tenets of the Declaration of Helsinki and was approved by the ethics committee of the Tane Memorial Eye Hospital. Written informed consent for participating in this study was obtained from all patients before they underwent the surgery.
In the preparation for IOL extraction, the anterior chamber is filled with an ophthalmic viscosurgical device (OVD) substance at room temperature. When replacing the IOL, it is removed from the lens capsule and pulled up on top of the iris. In the case of IOL dislocation, the vitreous body around the IOL is dissected using a vitreous cutter or an OVD substance; the IOL is then pulled up on top of the iris. To grasp the optics slightly to the right of the base of the haptics in the IOL, the IOL must be adjusted with a Sinskey hook to a position over the iris where it can be easily grasped with extraction forceps.
The cartridge pull-through technique uses a D1 cartridge (Hoya Corp.) and Fukuoka IOL extraction forceps (Handaya Co., Ltd.) (Figure 1). A sclerocorneal incision of 3.2 mm or wider was created; the cartridge lumen was filled with an OVD, after which the forceps were inserted until their tip was exposed from the cartridge end (Figure 2, A, a). The forceps were then inserted into the anterior chamber to grasp the optic slightly to the right of the base of the haptics (Figure 2, B, b). With the other hand, the cartridge tip was advanced slowly into the anterior chamber through the incision in the bevel down position and maintained steady (Figure 2, C, c), whereas the forceps grasping the IOL was retracted to draw the latter into the cartridge (Figure 2, D, d). The tip of the cartridge must be kept inside the anterior chamber until the IOL optic rolls up and into the barrel of the cartridge as it is pulled into the latter. The cartridge and forceps are then pulled out of the anterior chamber together to extract the IOL (Figure 2, E, e and F, f) (Video 1, available at https://links.lww.com/JRS/A380).
Figure 1.: Cartridge and forceps used in the cartridge pull-through technique. Top: D1 cartridge (Hoya Corp.). Bottom: Fukuoka IOL extraction forceps (Handaya Co. Ltd.)
Figure 2.: Cartridge pull-through technique procedure. A: The forceps are set so they pass through the cartridge, with the tip protruding from its end. B: The forceps held in the right hand are inserted into the anterior chamber to grasp the IOL optic. The optic is grasped slightly to the right of the base of the front loop. C: The tip of the cartridge is slowly pushed into the anterior chamber with the left hand through the incision and fixed into position. D: The IOL held with the forceps is pulled into the cartridge. E: The IOL can be extracted when approximately half of the optic is pulled into the cartridge to form a cylindrical shape by pulling out the cartridge and forceps together (pulling in the entire IOL is ideal). F: Successfully extracted IOL.
The extraction was performed according to the above-outlined method for IOL exchange by preserving the lens capsule or when a dislocated IOL was close to the iris. A dropped IOL lying on the retinal surface can be lifted up to the back of the iris with vitreous forceps after vitrectomy and can be extracted directly from the vitreous cavity without first pulling it up on top of the iris (Figure 3) (Video 2, available at https://links.lww.com/JRS/A381). The IOL, which is lifted to the back of the iris using vitreous forceps, was rarely positioned for grasping with the extraction forceps; the IOL optics are adjusted several times using a set of forceps in each hand. This adjustment was repeated until the IOL was at the ideal position for grasping with extraction forceps. The procedure should be performed with intraocular irrigation through an infusion port if there is a risk for globe instability.
Figure 3.: Direct extraction of the IOL from the vitreous cavity. Fallen IOLs can be extracted directly from the vitreous cavity without first lifting the IOL onto the iris.
The following information and points of caution should be considered when performing the cartridge pull-through technique:
A corneal incision can be made if the extracted IOL is confirmed to be a foldable IOL; however, if the dislocated IOL type is unknown preoperatively, a scleral tunnel incision should be made in case the IOL is intraoperatively determined to be composed of poly(methyl methacrylate). In addition, if the internal incision is close to the iris, the latter may be caught in the cartridge; therefore, an internal incision must be created away from the iris.
The D1 cartridge is suitable for implantation of IOL with an optic diameter of 6.5 mm. The recommended lengths of incision are 3.2 mm for corneal and 3.0 mm for sclerocorneal incisions. An incision wound of this size allows extraction of a foldable 6.0 mm optic diameter IOL to pass through the cartridge. However, the recommended incision wound size is a tight fit for simultaneous passage of the forceps grasping the optic and loop through the cartridge for IOL extraction; this explains the recommended sclerocorneal incision of 3.2 mm or wider.
When the extracted IOL is expected to be of high power or low refractive index and thickness, the cartridge should be warmed in a warming cabinet before use to promote stretchability; in addition, incisions should be made on the cartridge tip to enlarge the opening diameter.
Various OVDs can be used. As a cooled OVD will harden the IOL, rendering it difficult to extract, OVDs should be warmed to room temperature. The cartridge needs to be filled with OVD for the procedure because cartridges inserted into the anterior chamber without OVD will result in iris incarceration in the cartridge lumen. Furthermore, an OVD lubricates the cartridge lumen.
A nonfoldable IOL made of hard materials such as poly(methyl methacrylate) cannot be extracted using this method.
Retinal detachment may occur when the vitreous body is tangled on the IOL and pulled out together with it during IOL extraction; therefore, an OVD and vitreous cutter are required for dissecting the vitreous body before extraction.
Before performing this procedure, it is important to understand the movement of the IOL as it is pulled into the cartridge with forceps (Video 3, available at https://links.lww.com/JRS/A382). If the cartridge is beveled down, the IOL optics will fold and bulge out. Furthermore, the behavior of the haptics varies depending on the grasped portion of the IOL optic (Figure 4). If the grasped portion of the optic is slightly to the right of the base of the haptic, the nearest haptic is immediately pulled into the cartridge along with forceps. The furthest haptics actually turn up and toward the cornea as the IOL is being pulled into the cartridge. The barrel of the cartridge needs to be rotated so that the leading haptic stays away from the cornea. The behavior of only the one leading haptic may be checked to make the extraction easy to control (Figure 4, A, a). If the intermediate optic of both haptics has been grasped, caution is necessary; both haptics will rise to the vitreous side. If the IOL is removed when the haptics are still outside the cartridge, this may damage the iris (or lens capsule in case of IOL replacement). In addition, as it cannot be confirmed whether the loop is hidden behind the cartridge, it is difficult to decide whether the maneuver is safe. Therefore, grasping the optic at the midpoint is incorrect and can lead to iris (or capsule) trauma; it cannot be recommended (Figure 4, B). Caution is necessary during extraction, and the procedure must be performed slowly while checking the behavior of the IOL pulled into the cartridge.
Figure 4.: Haptics behavior depending on the location of grasping the IOL optic. A: If the optic is grasped slightly to the right of the haptics base, the furthest haptics actually turn up and toward the cornea. The barrel of the cartridge needs to be rotated so that the leading haptic stays away from the cornea. B: If the optic is grasped in the midpoint between both haptics, they will rise to the vitreous side. If the IOL is removed while haptics are still outside the cartridge, the iris (or lens capsule in case of IOL replacement) may be damaged. In addition, it cannot be confirmed whether the loop is hidden behind the cartridge.
RESULTS
A total of 16 eyes with IOL dislocation that underwent IOL extraction + vitrectomy + intrascleral IOL fixation with a D1 cartridge and Fukuoka IOL extraction forceps between July 2019 and March 2020 at our hospital were included. Data regarding the patient background, preoperative and postoperative corrected distance visual acuity, preoperative and postoperative corneal endothelial cell density, and type of the extracted IOLs are summarized in Table 1. There were no patients whose visual acuity worsened postoperatively; the IOL could be extracted with this technique in all patients. The extracted IOLs were 3-piece IOLs in 10 eyes and 1-piece IOL in 6 eyes. There were no complications such as iris injury, corneal endothelial cell loss, anterior chamber hemorrhage, or postoperative endophthalmitis.
Table 1. -
Summary of the 16 Cases.
Age (y) |
Sex |
CDVA (logMAR) |
ECD (cells/mm2) |
ECL (%) |
Extracted IOL |
Preop |
Postop |
Preop |
Postop |
Shape |
Model/power |
74 |
M |
0.046 |
−0.079 |
1836 |
2154 |
−17.32 |
3 piece |
Foldable IOL
a
|
75 |
F |
0.046 |
0.000 |
2528 |
2436 |
3.64 |
3 piece |
MA30BA
b
/+15.0 D |
59 |
M |
0.046 |
−0.079 |
2042 |
1579 |
22.67 |
3 piece |
ZA9003
c
/+21.5 D |
63 |
M |
0.398 |
0.046 |
2781 |
2830 |
−1.76 |
1 piece |
SN6AD
b
/unknown |
42 |
M |
0.000 |
−0.079 |
1692 |
1522 |
10.05 |
3 piece |
ZA9003
c
/+21.0 D |
53 |
M |
0.000 |
−0.176 |
2710 |
2687 |
0.85 |
3 piece |
ZA9003
c
/+17.0 D |
49 |
M |
0.000 |
0.000 |
2754 |
2682 |
2.61 |
1 piece |
PCB00V
c
/+17.0 D |
61 |
M |
−0.079 |
−0.176 |
2285 |
2260 |
1.09 |
3 piece |
AR40
c
/+20.0 D |
57 |
F |
−0.079 |
−0.079 |
2118 |
2174 |
−2.64 |
1 piece |
SA60AT
b
/unknown |
65 |
M |
0.155 |
0.155 |
2927 |
2764 |
5.57 |
1 piece |
NS-60YG
d
/unknown |
66 |
M |
0.000 |
−0.079 |
2860 |
2712 |
5.17 |
3 piece |
PY60R
e
/+14.5 D |
53 |
M |
−0.176 |
−0.301 |
2401 |
2206 |
8.12 |
3 piece |
MA60BM
b
/+17.0 D |
66 |
M |
0.000 |
−0.079 |
2962 |
2712 |
8.44 |
1 piece |
SZ-1
d
/+21.50 D |
76 |
M |
0.000 |
0.000 |
2884 |
2578 |
10.61 |
3 piece |
Foldable IOL
a
|
55 |
M |
0.000 |
−0.079 |
2645 |
2483 |
6.12 |
3 piece |
Foldable IOL
a
|
62 |
M |
0.000 |
−0.079 |
2495 |
2521 |
−1.04 |
1 piece |
PCB00V
c
/+17.5 D |
61 ± 9.5
f
|
|
0.022 ± 0.12
f
|
−0.068 ± 0.10
f
|
2495 ± 398
f
|
2394 ± 395
f
|
3.9 ± 8.4
f
|
|
|
ECD = endothelial cell density; ECL = endothelial cell loss; preop = preoperative; postop = postoperative
aDetails unknown
bAlcon Laboratories, Inc.
cJohnson & Johnson Vision
dNidek Co., Ltd.
eHoya Corp.
fMean ± SD
DISCUSSION
In this study, we described the cartridge pull-through technique, which was developed for IOL explantation, and the results of the procedure in 16 patients. The technique uses a cartridge for IOL implantation and forceps optimized for securely grasping the IOL for its removal. Bhaumik and Mitra used a cartridge (Alcon Laboratories, Inc.), wire snare, and reported that minimally invasive IOL extraction was possible through a small 3.5 mm corneal or limbal wound.4 However, the snare must be custom-made from a wire, and only 1-piece IOLs can be extracted with this method. The twist-and-out explantation technique proposed by Pandit et al. also allows removal of a foldable IOL through a 2.2 mm incision.7 However, this method requires rolling of the IOL that is twisted around the forceps in the narrow anterior chamber. As the IOL tries to spread to its original state, we believe that it may be difficult to keep the IOL rolled.
Our hospital has been performing IOL extraction using the cartridge pull-through technique since June 2018. Early in the development of this technique, the D1 cartridge and Zaldivar-Kraff ICL PACman forceps (ASICO LLC) for surgery of implantable collamer lens (STAAR Surgical AG) were used. Extraction of many 1-piece and 3-piece IOLs was possible with these devices. However, the jaws of the PACman forceps were too wide to glide smoothly into the cartridge lumen and too short to provide sufficient grasping power. As a result, thick IOLs could not be removed occasionally, and hydrophilic IOLs were too slippery to be extracted. This led to the development of Fukuoka IOL extraction forceps, which have a narrow and extended tip with hollow jaws to release pressure and are serrated for improved grip. These modifications have enabled stable grasping across the center of the IOL optic.
Before clinical use of the new forceps, an IOL extraction experiment was performed in water at the same temperature (29 to 32 degrees) as that of the anterior chamber water. Three of each type of the following IOLs were extracted: MA60AC (Alcon Laboratories, Inc.), SN60WF (Alcon Laboratories, Inc.), ZCB00 (Johnson & Johnson Vision), KS-SP (Nidek Co., Ltd./STAAR Surgical AG), and Lentis Mplus (Oculentis GmbH). As these were practice IOLs, the power of each IOL were unknown. The experimental results showed that all 3 Lentis Mplus IOLs could not be extracted; however, all other IOLs were successfully extracted. Since it is not possible to perform the experiment using IOLs at every IOL power, the range of frequency of successful extraction is unknown. In actual clinical practice, all 16 foldable IOLs used in this study were successfully extracted, and most of the frequently used IOLs can be extracted by this method. Furthermore, although there has been no opportunity in clinical practice to extract hydrophilic IOLs, an experiment was performed where a hydrophilic IOL was extracted in the same manner as described. Although the Lentis Mplus IOL could be grasped, it could not be pulled into the cartridge because of its large volume as it is of the plate type. The Micro F (FineVision) could be firmly grasped and extracted easily. Switching from the PACman to Fukuoka forceps has made it easier to perform various IOL extractions.
There were 2 cases of dropped IOL on the retinal surface among the 16 eyes. In these 2 patients, the dropped IOLs were extracted directly from the vitreous cavity. In our IOL explantation performed using the conventional method, dropped IOLs were first placed on the iris before extraction, but pulling up onto the iris failed in some cases, resulting in multiple drops in this step. Repeated attempts to retrieve a dropped IOL may damage the retina. Thus, direct IOL extraction from the vitreous cavity according to this method provides the additional benefits of protecting the anterior chamber tissue and suppressing retinal invasion.
No serious complications were noted in this study. The reason for this is that the use of a cartridge makes the incision smaller, and the pressure difference between the anterior chamber and the outside is small; this makes it difficult for the iris and vitreous to escape during IOL removal. In addition, it is possible to further reduce the wear of endothelial cells and iris by devising the angle of the wound opening and the direction of the bevel of the cartridge. However, since the cartridge being used is not made specifically for IOL extraction, the entire IOL cannot be pulled in; the haptics, therefore, remain outside the cartridge. As such, there is a need to be careful to not to damage the cornea or iris.
Unfortunately, after developing this technique, the D1 cartridge (outer diameter 2.25 mm) has been discontinued and is currently difficult to obtain. Next to the D1, the company's largest cartridges are C1 (for VA-60BB insertion) and C7 (for VA-70AD insertion); the tips of both cartridges have the same shape (outer diameter 2.13 mm). The recommended incisions are 3.0 mm in the cornea and 2.8 mm in the sclerocornea. In the aforementioned underwater experiment, using Fukuoka IOL extraction forceps and C1 and C7 cartridges, the same results were obtained as with the D1 cartridge. Currently, we are using commercially available cartridges for the procedure, but the development of a specialized extraction tool would be ideal for further fine-tuning of this technique.
In summary, the cartridge pull-through technique cannot be applied to extraction of all IOLs. However, it allows for effective foldable IOL extraction through an innovative minimally invasive procedure, with the potential of changing the future of IOL explantation.WHAT WAS KNOWN
- Longer life spans and earlier ages for crystalline lens extraction have increased the duration of IOL implantation in many patients. There is an increase in the number of patients with IOL dislocation and the frequency of cases that require IOL extraction because of longer lifespans and earlier age for crystalline lens extraction.
- Postoperative refractive error or dissatisfaction with multifocal IOL implantation outcomes can lead to a need for IOL extraction.
- IOL extraction requires folding the IOL in half with forceps or cutting it into several pieces with scissors before removal. Both maneuvers must be performed within the narrow space of the anterior chamber and can cause complications such as iris or corneal injury.
WHAT THIS PAPER ADDS
- Pulling the IOL into the cartridge with forceps through a small wound to minimizes intra-anterior chamber maneuvers, offering a simple solution for IOL extraction.
- This method allows for direct extraction of the IOL dropped into the vitreous cavity without placement onto the iris and is minimally invasive to tissues surrounding the anterior chamber.
Acknowledgments
Ryotaro Tano, MD, and Editage (www.editage.com) for English language editing.
REFERENCES
1. Yamane S, Inoue M, Arakawa A, Kadonosono K. Sutureless 27-gauge needle-guided intrascleral intraocular lens implantation with lamellar scleral dissection. Ophthalmology 2014;121:61–66
2. Yamane S, Sato S, Maruyama-Inoue M, Kadonosono K. Flanged intrascleral intraocular lens fixation with double-needle technique. Ophthalmology 2017;124:1136–1142
3. Ohta T, Toshida H, Murakami A. Simplified and safe method of sutureless intrascleral posterior chamber intraocular lens fixation: Y-fixation technique. J Cataract Refract Surg 2014;40:2–7
4. Bhaumik A, Mitra S. A simple technique of intraocular lenses explanation for single-piece foldable lenses. Indian J Ophthalmol 2017;65:1428–1430
5. Lee MH, Webster DL. Intraocular lens exchange-removing the optic intact. Int J Ophthalmol 2016;9:925–928
6. Henderson BA, Yang EB. Intraocular lens explantation technique for one-piece acrylic lenses. J Refract Surg 2012;28:499–502
7. Pandit RT, Devgan U, Chapman JM Jr. Twist and out intraocular lens removal. J Cataract Refract Surg 2020;46:1072–1074
8. Yu AKF. Complications and clinical outcomes of intraocular lens exchange in patients with calcified hydrogel lenses. J Cataract Refract Surg 2002;28:1217–1222
9. Hayashi K, Hirata A, Hayashi H. Possible predisposing factors for in-the-bag and out-of-the-bag intraocular lens dislocation and outcomes of intraocular lens exchange surgery Ophthalmology 2007;114:969–975
10. Theoulakis PE, Brinkmann CK, Petropoulos IK, Gatzogias MI, Katsimpris JM. Hydrogel intraocular lens exchange: five-year experience. Clinical Trial 2009;226;254–257
11. Gashau AJ, Anand A, Chawdhary S. Hydrophilic acrylic intraocular lens exchange: five-year experience. J Cataract Refract Surg 2006;32:1340–1344