LABORATORY SCIENCE

Intraocular lens far peripheral vision

image detail and negative dysphotopsia

Simpson, Michael J. PhD

Author Information
Journal of Cataract and Refractive Surgery: March 2020 - Volume 46 - Issue 3 - p 451-458
doi: 10.1097/j.jcrs.0000000000000103
  • Free

Hundreds of millions of patients have received intraocular lenses (IOLs) during cataract surgery, and there has rarely been any concern about the quality of vision in the far periphery. However, starting around the year 2000, patients occasionally reported bothersome dark shadows in their temporal peripheral visual field.1 Since then, there have been many publications about this type of negative dysphotopsia (ND), with the more recent ones summarizing the earlier studies.2–6 Initially, it was thought that these reports might be linked to specific IOLs that had a high refractive index, but further evaluation found that the phenomenon was also present with other lens materials. Various potential causes were proposed, such as a visual phenomenon from the surgical incision, light passing through the edge of the IOL, and effects from the capsulorhexis. Reports of cases that are particularly bothersome have been relatively rare, although 3% may still see shadows at 1 year postoperatively, and a recent study identified over 1% of about 6000 procedures as having ND.7,8

Something that was not perhaps fully discussed initially was a comparison between the phakic crystalline lens and the IOL. These are depicted in Figure 1 as an overlay of preoperative and postoperative optical coherence tomography images for the same eye, in which the crystalline lens and IOL profiles are also identified. Preoperatively, all the light that enters the pupil is directed toward the same general region of the retina, even if the image is defocused and aberrated. Postoperatively, however, the IOL is physically very different. Using ray tracing to evaluate an IOL in a model eye (Figure 2), it is found that at large visual angles, with a small pupil, the light no longer passes through an IOL at all because it is much smaller than the natural crystalline lens. The main image goes totally dark, and this is highly likely to be the primary cause of the shadow. This property of an optical system is called vignetting for a conventional lens, leading to a darkened periphery in early camera images that had simple lenses. The thing that makes it bothersome for an IOL patient may be the fact that light can also bypass the IOL and illuminate the retina further out. This can leave a distinct shadow within an illuminated region of the retina. The calculations also demonstrate that the shadow that is present with a small pupil is eliminated as the pupil opens up, which agrees with clinical reports. Evaluations of iris characteristics indicate that the typical gap between the posterior iris and the IOL is about 0.5 mm.9–11

Figure 1
Figure 1:
Overlay of preoperative and postoperative optical coherence tomography images for the same eye. Crystalline lens profile estimated using published data22 (IOL = intraocular lens).
Figure 2
Figure 2:
Input visual angles related to the retinal location and displayed approximately on a polar plot. The input visual angle is linearly related to the angle subtended at the second nodal point of the eye up to very large angles (IOL = intraocular lens).

The visual angle at which the shadow is seen has not been usually recorded objectively, but a typical example is where a hand raised to the side of the face like a military salute will remove the effect. This indicates that it is caused by light at very large visual angles, and a recent study gives a typical value of about 90 degrees.8 It is also always in the temporal direction, with the nose, eyebrow, and cheek blocking light from angles above about 60 degrees anyway. The large angle is consistent with the first reports of ND appearing about 20 years ago, when clearer peripheral capsules became more routine following the introduction of phacoemulsification, small incisions, and foldable IOLs with a sharp posterior edge. These improvements permitted light that had previously been obstructed or scattered when it missed the IOL to reach the retina directly.

It has only recently become clear quite how limited the scientific evaluation has been of the entire visual region of the far periphery.12 Perimetry is widely used to evaluate peripheral vision, but it is rarely used at large angles, and its purpose is primarily to assist in diagnosing specific diseases, rather than to characterize vision itself. Perimetry primarily relates input light sources to patient responses, and it is possible that the image on the retina during a perimetry test has never been modeled using ray tracing. This type of modeling is actually difficult to do because the image surface is highly curved. Even the latest ray-trace software that is used here does not include standard tools to evaluate this.

Drawings of rays passing through the pseudophakic eye in the periphery all show the type of bifurcation that is depicted in Figure 2, using several calculation methods.2,3,13,14 The image formed in this region has to be different from that of the phakic eye, although that is rarely noted (although image quality in the periphery is also extremely poor). These drawings have also been somewhat misleading, however, because although a single input beam splits into 2 when light starts missing the IOL, the light that is seen in the periphery no longer comes from the largest visual angles because it does not experience the focusing power of the IOL. The peripheral light hits the retina directly, and although it might appear to come from a larger angle because of its retinal location, the ray-trace calculations indicate that the maximum visual field angle is reduced with an IOL.

More recent evaluations calculated illumination on the retina from an extended bright source, rather than just single rays.15,16 These were presented as retinal images transferred back to the object space, by comparing retinal locations with object locations for the phakic eye using an initial estimate for the scaling. This method is now extended here to use an input text object, which evaluates the imaging properties from an enhanced perspective. A matching phakic eye model was also developed, as an initial attempt to compare the image properties between the phakic and pseudophakic eyes, with an improved estimate for the image scaling.

METHODS

Zemax OpticStudio optical design software (Zemax) was used to generate ray-trace data using an optical model for an average eye.3 This had a 2.5 mm diameter pupil (3.0 mm apparent pupil) decentered nasally by 0.25 mm. Angles were evaluated relative to the optical axis, and then 5 degrees was added to the input ray angles to account for the average foveola location. The Zemax ray-trace software does not have standard methods that facilitate this type of evaluation but additional capabilities were created for the modeling, and ray intersections with the retina were saved to a file. The following methods were used: (1) Special optical surfaces were created for both IOL surfaces using the user-defined surface capability, so that both rays passing through the IOL and rays missing the IOL were recorded with a sequential ray trace. Without this, all rays that miss the IOL are stopped. Earlier work had run 2 separate ray traces, but that prevented the rays reaching the edge of the IOL from being visible in a single plot.3,17 Any rays that hit the outer edge of the IOL were ignored, which simulates the effect of a rounded and frosted lens edge that widely scatters the light over a large region. (2) The model eye was set up so that it could be rotated about a vertical axis, and a Zemax macro was used to record the coordinates of ray intersections with the retina for object points at a 6 m distance. The macro takes advantage of the ray-aiming capability of Zemax to find the location of the iris boundary for rays from each object point and then launches a set of rays that are equally spaced in angle. A separate polarization routine was used to calculate the fraction of each refracted light ray that reached the retina, after Fresnel reflections were removed, and this was also saved. Both rays that pass through the IOL and rays that miss the IOL form part of the set of rays. (3) The retinal ray intersections were then imported into MATLAB for analysis and display.

For most of the evaluations here, the retinal locations were scaled directly to the object space using the angle subtended at the second nodal point of the pseudophakic eye (Figure 2). This is based on that angle being virtually identical to the input angle to beyond 70 degrees for the chief ray passing through the center of the pupil.3,15,18 This highly linear relationship starts to change at larger angles, but it is not known how accurate this is, and earlier publications have used the linear relationship for all angles to provide relative values.3,15,18 Aberrations will also alter the effective ray location for large angles for the light passing through the IOL. The use of the subtended angles permits the retinal image to be displayed as though it were on a quasipolar plot, in a similar manner to the way visual field results might be considered. This has been found to be very beneficial because it flattens out the image onto a plane surface for review, while also inverting the image. An input object that consists of a text chart is used, for the first time, in the analysis here, and this scaling method enables the input object and the resulting image to be compared.

The ray intensities also need to be weighted to approximate what a person might see. One adjustment comes from the total transmission values mentioned earlier. Another weighting factor attempts to compensate for the visual effect of a reduction in image intensity at very large angles because the pupil appears to be increasingly elliptical with increasing visual angles, resulting in the horizontal pupil opening getting progressively smaller. This has an effect on the transmitting area, which falls off as an approximate cosine function.15 The mechanism by which the eye compensates for the image getting dimmer in the periphery does not appear to have been researched, and to counteract it, each ray was weighted by 1/[cos(0.8 × retinal angle)].17 A gamma correction value of 0.5 was also used to enhance the visibility of lower intensity levels when they are displayed as an image. After each image was created, by adding weighted rays into a matrix of image points corresponding to horizontal and vertical visual angles, the overall intensity of each image was rescaled again to fill the full intensity range.

In addition to the rays being weighted appropriately, simulated images of a text chart were also created (Figure 3). The point spread function (PSF), which is the image of a point source, varies strongly with a radial location, and this is illustrated in Figure 4, B for a very sparse set of object points in the horizontal direction. Most imaging systems are isoplanatic, where the PSF shows very little variation from point to point, and usually, Fourier transform methods are used to generate a simulated image, using a constant PSF. Even where the image is more complex, it can be broken down into subregions that are isoplanatic for calculation. This particular optical system has a highly variable PSF, however, and the image here was generated directly using individual object points. Each ray was weighted by the object intensity, in addition to the other factors, before being added to the final image. Because of the time involved with this, the main images were generated using just the horizontal set of PSF values for all the vertical points in the image because the changes to the PSF were very modest for the other locations. The images are all monochromatic.

Figure 3
Figure 3:
Approximate location of the illuminated peripheral text object displayed on a polar plot.
Figure 4
Figure 4:
PSF images for a few widely spaced object points using linear scaling for a phakic and a pseudophakic eye. With the pseudophakic eye, light that misses the intraocular lens creates a separate arc as part of the PSF (PSF = point spread function).

Separately, an exploration was undertaken to see whether the angular scaling could be improved, and a phakic eye model for a 70-year-old eye was created with a gradient index lens. Earlier evaluations of the scaling for a phakic eye had used a gradient index lens in a widely used model for a young eye, although the model had not been created for use at large angles.3 There are several other phakic eye models in the literature, but even when they are described as being wide-angle models for the eye, they are typically only used up to perhaps 40 degrees. None of the earlier work seems to go even to 70 degrees, let alone to the limit of temporal vision that is generally assumed to be approaching 110 degrees. The limiting visual angle also appears to have seen little research. In particular, the variation with age does not appear to be known, and it is not clear how large the limiting visual angle is for a phakic 70-year-old eye. The modeling here used a user-defined surface from the study by Akram et al., based on the work by Bahrami and Goncharov, and it was further modified to have an aspheric anterior surface (Figure 5).19,20 A 70-year-old eye was modeled, and the chief ray was evaluated with increasing angles.21,22 This largely verified the earlier simpler calculations, with a linear relationship to very large input angles, followed by a droop in the curve. This new scaling was then used to generate images comparing the phakic and pseudophakic eyes directly, this time using the actual calculated PSF values for each horizontal and vertical object point.

Figure 5
Figure 5:
Ray trace of a 70-year-old phakic eye with a gradient index lens. The axial and radial refractive index values are sketched in the inset.

Sketches were also created illustrating the characteristics of positive dysphotopsia, using a polar plot representing an object space, to assist with the Discussion section.23

RESULTS

The main simulated image in the periphery from this calculation method for a 2.5 mm diameter pupil is given in Figure 6. At a very large angle, the main image that is formed by the IOL becomes dark, which is where the vignetting is so prominent that it truncates the focused image. Light can also miss the IOL, however, and that creates the additional image region further out, with a dark region in-between that would appear as a shadow. This image also illustrates an additional characteristic of the overall optical system, however, because the outer image actually comes from a lower visual angle, and part of the object is actually seen twice. This information is already included in ray-trace plots like the depiction in Figure 2, in which a single input ray is split at the edge of the IOL, but the text image characterizes this in a clearer manner. It can also be seen that the outer region is larger because it has not experienced the additional power of the IOL, and also that the 2 regions have different aberrations. Text would not normally be clearly resolved at these very large angles, but the overall optical effects are illustrated by these imaging simulations.

Figure 6
Figure 6:
Simulated peripheral image (IOL = intraocular lens).

Additional characteristics are depicted in Figure 7. In Figure 7, A, the eye rotates about a vertical axis, and it can be seen that the K and D move from the outer region to the central region as the eye rotates, but the dark shadow stays in the same place. With the eye steady and the pupil opening, however, the shadow is rapidly dissipated with very small pupil diameter changes (Figure 7, B).

Figure 7
Figure 7:
A: The effect of rotating the eye with a 2.5 mm pupil. B: The effect of increasing the pupil diameter.

Figure 8 gives images that compare a phakic and pseudophakic 70-year-old eye, using the revised scaling that is not linear at very large visual angles, and where each object point is used with its own PSF (rather than using the horizontal PSFs for all image locations). This indicates that the total limit for the image of the phakic eye is about 105 degrees, which is in agreement with general thinking about the eye, although it is not clear that there are any published values at all for the older eye. With the pseudophakic eye, the magnification is slightly different because although the IOL power is chosen so that the image is in focus on the retina, the effective axial location where the IOL power has an effect is different from that of the phakic eye. The main focused image only extends to about 95 degrees because the light is vignetted at the IOL. The displayed image is extended out to 120 degrees here to make the D clearly visible, although presumably vision does not normally extend to this angle. This is really scaling the retina as though the eye was a phakic eye, and details the retinal behavior in this region are not really known.

Figure 8
Figure 8:
Comparison between images for phakic and pseudophakic 70-year-old eyes, using scaling for the phakic eye relative to the second nodal point for both eyes.

Figure 9 adds additional schematic images related to positive dysphotopsia, using the framework of a polar plot in object space that was introduced previously. This is used in a discussion later about what might be bothersome for a patient experiencing ND, and although there are general discussions about this in the literature, it is not clear that the motion characteristics of the image are fully described anywhere. Positive dysphotopsia, which is characterized by an arc or flash of light near the fovea, is seen with a large pupil at night (in comparison with ND, which is seen with a small pupil under bright illumination). In Figure 9 (center), if there is only a single-point light source at about 30 degrees of the visual angle, and the IOL has a perfectly cylindrical outer edge, ray tracing indicates that all the light that hits the edge near the horizontal point will be totally internally reflected in the general direction of the fovea. The illuminated region looks like an arc. If the eye rotates to look at the light (Figure 9 (left)), the light does not hit the lens edge, so there is no secondary image. If the eye rotates the other way, an arc of light will suddenly appear near the fovea when light starts to be reflected from the edge, but it appears to move away from the rotating eye because it is being seen in reflection, and it moves at twice the speed of the eye rotation. Also, as the angle increases, more of the edge is receiving the incident beam as it moves toward more normal incidence on the edge within the lens. However, light also gets increasingly transmitted through the edge rather than being reflected when it is beyond the critical angle, and the image rapidly disappears.

Figure 9
Figure 9:
Schematic representation of positive dysphotopsia with an intraocular lens.

DISCUSSION

Text is never resolved clearly in the far periphery, but the simulated text images that have been calculated here provide clarification about what happens to light reaching the retina in this visual region. With a small pupil, the main focused image can go totally dark at large visual angles, and this is very likely to be the primary cause of ND. However, rather than being an unexpected artifact caused by specific IOLs or surgical techniques, it is a fundamental property of the pseudophakic eye. This visual region has never really been evaluated as an imaging system before, for either phakic or pseudophakic eyes, because patients typically do not complain about vision in the far periphery, and it is not a specific field of study. Even if nobody was complaining about dark shadows in the periphery, the analysis raises questions about what is actually seen there. The topic has arisen because of IOLs, but similar questions apply to the phakic eye.

The initial ray-trace modeling showed where the light rays went, this was extended later to illustrate peripheral retinal illumination with an extended source, and the simulated text charts included here now provide additional clarification (Figure 7).3,15 The images show that the shadow can be very prominent with small pupils, but that it rapidly gets fainter as the pupil opens up and the peripheral retina is flooded with light that misses the IOL. This is consistent with the general clinical understanding about ND, where it is most noticeable with small pupils.6,10,24 The text images also show that the first letter in the region beyond the shadow is the same as the last letter in the primary image, so the object is being seen twice. This is the same type of information that the bifurcated beam provides in ray-trace drawings, but it is easier to visualize (Figure 2). The evaluations demonstrate a clear mechanism for how a shadow is created in the periphery, although at the moment there is not enough detail in clinical reports of ND to definitively confirm the cause (and there may also be other shadow-like phenomena that have different causes).

One particular difficulty with ray-trace modeling is the scaling of the images. This is partly because the pseudophakic eye can have a double image in the periphery, but also because the retina is highly curved, and the physical location of image regions is not usually mapped back to visual angles. The image location on the retina has to be related back to what is perceived to be the corresponding visual angle. Using linear scaling extrapolated from lower angles, the ray-trace modeling indicates that the shadow is at large angles of 80 to 90 degrees, but the angle is not exactly known.3 A new attempt here to model the older phakic eye indicates that there is a nonlinearity at larger visual angles, but there appear to be no studies that have evaluated this relationship.

The peripheral double image is also a particular problem for perimetry, where the patient responds only that they have seen a stimulus and not where that stimulus was seen. The pupil diameter is also typically not controlled, or even measured, during perimetry, and even for perimetry studies related to dark shadows, no pupil diameters are provided. There are now several reports of differences in perimetry for patients who see dark shadows, but it is not clear how closely the features that are measured correspond to the shadows that the patient perceives to be bothersome.5,25,26 One difference between patients is also the lens style that is used, and in many studies, the IOLs are evaluated as a single group. However, it has recently been noted that IOLs with a lower refractive index typically have a reduced optical diameter on 1 or both surfaces, and differences like this may affect image luminance at different angles.18

The actual situation is probably that there are nuances in the overall visual field of the pseudophakic eye that have just never been explored before, and these probably vary with the pupil diameter and lens style. A recent study reports that most patients with ND feel that the ND-shadow is far in the periphery, but additional shadow-like effects are also recorded for those patients at much lower visual angles.8 The incidence of ND reports in that study is about 1.3%. At the recent symposium on negative dysphotopsia at the ESCRS in Paris, 2019, it was also reported that shadows were measured objectively for patients for many different visual angles (“Quantifying negative dysphotopsia,” S. Palkovits). The specific cause of these shadows was not modeled, but this must be a different type of shadow to the vignetting described here at large angles.

The various evaluations and discussions in the main references about ND indicate that the bothersome shadows are only seen with small pupils. The calculations here can be used with that information to actually define a very clear hypothesis for the primary cause of dark shadows in the far periphery. The hypothesis might be stated as follows: At very large visual angles, light cannot pass through the IOL when the pupil is small, and the main image goes dark. There do not appear to be standard measurement methods that can be used to evaluate this hypothesis because this visual region is usually measured only with perimetry, which is generally set up so that the effects of things such as refractive error and pupil diameter do not affect the results. What is really needed is a test in which the patient actually sees a shadow, and the visual angles relating to it are recorded, along with the pupil diameter. The lack of control over the pupil diameter is a particular problem with conventional testing.

A description of positive dysphotopsia has also been included here because lens edge reflections are known to be bothersome, and the use of an input visual angle plot provides a new format for discussing the characteristics. There are various discussions in the literature about this topic, but little that actually fully describes what might make this very annoying for a patient. The discussion in the Results section relating to Figure 9 indicates that light entering the eye at modest visual angles (for example, at about 30 degrees) can reflect an arc of light near the fovea, and that this arc can seem to appear and disappear as the eye moves. The bright arc does not appear to be linked to any light source that is in the visual field, and the patient is not normally able to fixate on it. This is probably what makes it particularly bothersome, although evaluations of positive dysphotopsia have not typically related detailed laboratory measurements of specific IOL styles to clinical observations. Discussions have been of a more general nature, without stating a clear hypothesis that was evaluated in detail.

A similar situation exists now for ND, in which it is known that peripheral shadows are a problem, but it has not been explained why they are bothersome. Vision in the far periphery is particularly sensitive to motion, and the ray-trace calculations indicate 2 potential mechanisms that may attract attention to the region of the shadow. One possibility is that a very small change in the pupil diameter can cause a very large change in the darkness of the shadow, and if the pupil changes rapidly, it may be perceived as being movement in the periphery. A second possibility is that because the light that misses the IOL has a different magnification to the primary image, with the image detail appearing twice, that may also lead to a sense of unusual motion in the far periphery. These 2 mechanisms may also combine, because the eye and the iris are typically always moving, and the patient may also move physically through an environment that has strong visual stimuli. In comparison, the eye is never bothered by the blind spot created by the optic nerve, although this is always in the same location on the retina, in a similar manner to the dark shadow being primarily at a fixed location. However, the retinal region near the blind spot never sees at all, but the peripheral retinal region evaluated here is always sensitive, and for larger pupil diameters, it is always illuminated. There are no publications that specifically mention effects like these, but a second part of the hypothesis might be: Light missing the IOL illuminates the peripheral retina directly, creating the appearance of a shadow with small pupils, while also repeating image detail with a different magnification. Rapid changes in the pupil diameter, or motion of the eye when there is strong object detail, may lead to a sudden change in the appearance of the visual field in the periphery, making it bothersome for the patient.

There are several recent publications that discuss methods for treating ND, and these include alternative theories for their cause.2–4,6,24,27 One treatment method is reverse optic capture, with discussions also about how the anterior capsule may be involved. This method may also move the IOL closer to the iris, which would be expected to alter the visual phenomena anyway. Another treatment method is to use a secondary piggyback IOL in the sulcus, which is thought to be 70% effective.26,28 A ray-trace evaluation of thick silicone piggyback IOLs found that the iris was moved forward, which altered the shadow, although this was only effective in 2 of 3 patients.16 Other piggyback IOLs are thinner, but alleviation of ND symptoms was similar.26 The haptic orientation may also have an effect, and neuroadaptation, and the effect of the second eye, have also been considered to affect the perception of ND.27,29

The ray-trace modeling also raises questions about the total field of view of the pseudophakic eye, and there do not appear to have been any evaluations of this. It has probably always been assumed that a patient with an IOL sees things in a similar manner to a phakic person, but just a glance at Figure 1 would question this. However, IOL surgery developed over a period of decades without the benefit of images like this, and patients never really complained. The ray-trace calculations indicate that the limiting field of view is smaller for the pseudophakic eye. The evaluation here only considers a specific eye, and there are large variations in eye parameters across the population, so there may be a large number of expected outcomes because of variations in such things as corneal power, axial length, pupil diameter, IOL centration, IOL style, and the separation between the iris and the IOL.

Overall, the earlier publications, and the more recent ray-trace evaluations, support a single theory about ND in the far periphery. Both the thickness and diameter of the IOL are very much smaller than the same dimensions for a crystalline lens. With a small pupil, the main image is no longer focused by the IOL, and it goes dark at large angles. Light can also miss the IOL, and this can provide illumination to the far peripheral retina, but this does not come from visual angles that correspond with phakic vision. The combination of these effects can lead to a region whose darkness varies dramatically with small pupil diameter changes, with dark regions being perceived as a shadow. Any motion of image detail in the very far periphery will also be different from the motion of the more central image because it is not focused by the IOL. A systematic evaluation of all the factors that affect this visual region should be able to confirm the cause.

WHAT WAS KNOWN

  • The peripheral image formed by an intraocular lens (IOL) is limited by the lens diameter with a small pupil, but light missing the IOL can also illuminate more peripherally, leaving a dark shadow-like region.

WHAT THIS PAPER ADDS

  • Simulated images confirm the shadow and also demonstrate that the peripheral light comes from a lower visual angle and that it is very sensitive to the pupil diameter.
  • Far peripheral imaging with the pseudophakic eye is very different from that of the natural eye, whether or not dark shadows are reported.

Prof. Nadeem Akram provided the Zemax gradient index model and Dr. Maria Muzyka-Woźniak provided optical coherence tomography images.

REFERENCES

1. Davison JA. Positive and negative dysphotopsia in patients with acrylic intraocular lenses. J Cataract Refract Surg 2000;26:1346–1355
2. Holladay JT, Zhao H, Reisin CR. Negative dysphotopsia: the enigmatic penumbra. J Cataract Refract Surg 2012;38:1251–1265
3. Holladay JT, Simpson MJ. Negative dysphotopsia: causes and rationale for prevention and treatment. J Cataract Refract Surg 2017;43:263–275
4. Masket S, Fram NR, Cho A, Park I, Pham D. Surgical management of negative dysphotopsia. J Cataract Refract Surg 2018;44:91–97
5. Makhotkina NY, Berendschot TTJM, Nuijts RMMA. Objective evaluation of negative dysphotopsia with Goldmann kinetic perimetry. J Cataract Refract Surg 2016;42:1626–1633
6. Henderson BA, Geneva II. Negative dysphotopsia: a perfect storm. J Cataract Refract Surg 2015;41:2291–2312
7. Makhotkina NY, Nijkamp MD, Berendschot TTJM, van den Borne B, Nuijts RMMA. Effect of active evaluation on the detection of negative dysphotopsia after sequential cataract surgery : discrepancy between incidences of unsolicited and solicited complaints. Acta Ophthalmol 2017;96:81–87
8. Wenzel M, Langenbucher A, Eppig T. Causes, diagnosis and therapy of negative dysphotopsia [in German]. Klinische Monatsblatter fur Augenheilkunde 2019;236:767–776
9. Simpson MJ, Muzyka-Woźniak M. Iris characteristics affecting far peripheral vision and negative dysphotopsia. J Cataract Refract Surg 2018;44:459–465
10. Vámosi P, Csákány B, Németh J. Intraocular lens exchange in patients with negative dysphotopsia symptoms. J Cataract Refract Surg 2010;36:418–424
11. Pereira FAS, Cronemberger S. Ultrasound biomicroscopic study of anterior segment changes after phacoemulsification and foldable intraocular lens implantation. Ophthalmology 2003;110:1799–1806
12. Simpson MJ. Mini-review: far peripheral vision. Vision Res 2017;140:96–105
13. Coroneo M, Pham T, Kwok L. Off-axis edge glare in pseudophakic dysphotopsia. J Cataract Refract Surg 2003;29:1969–1973
14. Hong X, Liu Y, Karakelle M, Masket S, Fram NR. Ray-tracing optical modeling of negative dysphotopsia. J Biomed Opt 2011;16:125001
15. Simpson MJ. Simulated images of intraocular lens negative dysphotopsia and visual phenomena. J Opt Soc Am A 2019;36:B44–B51
16. Erie JC, Simpson MJ, Bandhauer MH. Effect of a sulcus-fixated piggyback IOL on negative dysphotopsia: a ray-trace analysis. J Cataract Refract Surg 2019;45:443–450
17. Simpson MJ. Mini-review: far peripheral vision. Vis Res 2017;140:96–105
18. Erie JC, Simpson MJ, Bandhauer MH. Effect of a sulcus-fixated piggyback intraocular lens on negative dysphotopsia: ray-tracing analysis. J Cataract Refract Surg 2019;45:443–450
19. Akram MNA, Baraas RC, Baskaran K. A wide-field emmetropic human eye model based on ocular wavefront measurements and geometry-independent gradient index lens. J Opt Soc Am A 2018;35:1954–1967
20. Bahrami M, Goncharov AV. Geometry-invariant GRIN lens: finite ray tracing. Opt Express 2014;22:27797–27810
21. Moffat B, Atchison DA, Pope J. Age-related changes in refractive index distribution and power of the human lens as measured by magnetic resonance micro-imaging in vitro. Vis Res 2002;42:1683–1693
22. Atchison DA, Markwell EL, Pope JM, Swann PG. Age-related changes in optical and biometric characteristics of emmetropic eyes. J Vis 2008;8:1–20
23. Holladay JT, Lang A, Portney V. Analysis of edge glare phenomena in intraocular lens edge designs. J Cataract Refract Surg 1999;25:748–752
24. Masket S, Fram N. Etiology of negative dysphotopsia. J Cataract Refract Surg 2013;39:485–486
25. Makhotkina NY, Berendschot TTJM, Beckers HJM, Nuijts RMMA. Treatment of negative dysphotopsia with supplementary implantation of a sulcus-fixated intraocular lens. Graefes Arch Clin Exp Ophthalmol 2015;253:973–977
26. Makhotkina NY, Dugrain V, Purchase D, Berendschot TTJM, Nuijts RMMA. Effect of supplementary implantation of a sulcus-fixated intraocular lens in patients with negative dysphotopsia. J Cataract Refract Surg 2018;44:209–218
27. Masket S, Rupnik Z, Fram NR. Neuroadaptive changes in negative dysphotopsia during contralateral eye occlusion. J Cataract Refract Surg 2019;45:242–243
28. Masket S, Fram NR. Sulcus-fixated IOLs for negative dysphotopsia. J Cataract Refract Surg 2015;41:478
29. Erie JC, Simpson MJ, Bandhauer MH. Influence of the intraocular lens optic–haptic junction on illumination of the peripheral retina and negative dysphotopsia. J Cataract Refract Surg 2019;45:1335–1339
© 2020 Published by Wolters Kluwer on behalf of ASCRS and ESCRS