Effect of supplementary implantation of a sulcus-fixated intraocular lens in patients with negative dysphotopsia

Makhotkina, Natalia Y. MD*; Dugrain, Vincent PhD; Purchase, Daniel EngD; Berendschot, Tos T.J.M. PhD; Nuijts, Rudy M.M.A. MD, PhD

Author Information
Journal of Cataract & Refractive Surgery: February 2018 - Volume 44 - Issue 2 - p 209-218
doi: 10.1016/j.jcrs.2017.11.013
  • Free

Abstract

The term “negative dysphotopsia” was originally introduced in 2000 by Davison,1 who described complaints of “a subjective darkness of a shadow, which can be arc-shaped, usually in the temporal field of vision” in patients after uneventful cataract surgery.

The symptoms are usually worse in photopic conditions2–5 and might be alleviated by pupil dilatation.2 Many case reports and clinical and laboratory studies have evaluated potential mechanisms of negative dysphotopsia over the past 15 years.6

The last most comprehensive laboratory study pointed out primary and secondary factors that increase the risk for negative dysphotopsia,7 and its results are partially supported by previous clinical studies (Table 12−5,8−16). The current opinion that negative dysphotopsia is a result of a complex interaction between intraocular lens (IOL) optics and an anatomically predisposed eye suggests the multifactorial nature of this phenomenon.3,7,17–19

Table 1
Table 1:
Causative factors of negative dysphotopsia identified with ray-tracing analysis by Holladay et al.7

The primary accepted treatment strategy is a combination of counseling and reassurance because most cases of negative dysphotopsia will likely abate. It has been suggested that negative dysphotopsia can abate because of neuroadaptation or early postoperative changes, such as opacification of the peripheral lens capsule.14,20–23 Severe persistent cases can be treated with IOL exchange with placement of a secondary IOL in the bag or in the ciliary sulcus, implantation of a supplementary IOL, reverse optic capture, or neodymium:YAG (Nd:YAG) anterior capsulectomy. Until recently, IOL exchange with implantation of a secondary IOL in the ciliary sulcus seemed to be most successful for the resolution of negative dysphotopsia, whereas other treatment modalities have had lower degrees of success.1,2,8–10,24–26

Patients with severe negative dysphotopsia at the University Eye Clinic Maastricht are treated with supplementary implantation of a sulcus-fixated IOL (Sulcoflex, Rayner Intraocular Lenses, Ltd.) in the ciliary sulcus, which is a safe procedure leading to improvement or resolution of negative dysphotopsia in 78% of cases.25 However, surgical outcomes cannot be predicted because the pathophysiological mechanism of the treatment effect of the sulcus-fixated IOL is still unclear.

In this study, we evaluated the relationship between clinical data and surgical outcomes of patients with bothersome negative dysphotopsia treated with implantation of the sulcus-fixated IOL to identify factors that contribute to the resolution or persistence of negative dysphotopsia after surgery. In addition, to account for the multifactorial nature of negative dysphotopsia, we assessed the combined effect of different factors using optical modeling of individual biometric data before and after implantation of the sulcus-fixated IOL. Optical simulation of a piggyback IOL suggested the role of multiple surface reflections in the resolution of negative dysphotopsia.23 However, this analysis was based on a general eye model and could not explain partial improvement or persistence of the symptoms in different cases. To our knowledge, the modeling of negative dysphotopsia in a systematic fashion with patient-specific data has not yet been performed.

Patients and methods

Clinical Data

Medical files of patients with severe negative dysphotopsia who had supplementary implantation of a sulcus-fixated IOL between March 2014 and June 2016 were retrospectively reviewed. The data were partially reported previously.25 All patients had uneventful phacoemulsification with IOL implantation in the capsular bag. Secondary implantations were performed by an experienced surgeon (R.N.) at the University Eye Clinic in Maastricht, the Netherlands. A supplementary IOL was implanted in the ciliary sulcus, anterior to the primary IOL–capsular bag complex. A sulcus-fixated IOL (Sulcoflex aspheric 653L in 9 eyes and Sulcoflex aspheric toric 653T in 1 eye) with powers varying from 0 to 0.5 diopter (D) was used in all cases. The use of retrospective clinical data is in line with the guidelines of the Medical Ethical Committee of Maastricht University Medical Center.

Patient evaluations were performed preoperatively and 1 month after surgery. The evaluations included uncorrected and corrected distance visual acuities (Snellen), manifest refraction, Goldmann applanation tonometry, slitlamp evaluation, Scheimpflug photography (Pentacam, Oculus Optikgeräte GmbH), anterior segment optical coherence tomography (AS-OCT) (SS-1000, Tomey Corp.), partial coherence interferometry (PCI) (IOLMaster, Carl Zeiss Meditec AG), and kinetic perimetry (Goldmann perimeter, Haag-Streit AG). Kinetic perimetry was performed in photopic conditions with stimuli of maximum size and intensity (V4e) to evaluate the extension of the peripheral visual field and/or a presence of the relative scotoma as it was previously described.11 Data from Scheimpflug photography, AS-OCT, and kinetic perimetry were used in the current study. The chord μ (ie, approximation of angle κ, the angle between the apex [visual axis] and the pupil center) was assessed using the Holladay report on the Scheimpflug photography device.A

Optical Modeling of Light Distribution on the Retina

Biometric measurements of the patients were used for the construction of ray-tracing models. Anterior chamber depth, cornea–iris distance, thickness of the iris, thickness of the primary and supplementary IOLs, and distance between primary and supplementary IOLs were measured with AS-OCT. Because the AS-OCT software uses the refractive index of the aqueous humor of 1.343 for calculation of the IOL thicknesses and thus overestimates it, the measured IOL thicknesses were adjusted for the differences in the refractive indexes between aqueous humor and the IOL. The axial length (AL) of the eyes was measured with PCI.

Patient-specific optical models of the eyes were built in Zemax Optic Studio Optical Simulation softwareB in nonsequential ray-tracing mode. The light hitting the patient’s cornea was modeled as a 3-dimension (3-D) Ganzfeld source (described previously21), which is a good model for normal (homogeneous) lighted surroundings. Between 1.7 million and 2.3 million rays were traced for each patient. The geometry of the eyes was built on patient-specific biometry data as specified in Table 2.27

Table 2
Table 2:
Parameters used in the optical model.

Because data on the optical design of the aspheric primary IOLs were not available, these were modeled as standard equi-biconvex spherical IOLs with a square edge that was at least 0.2 mm thick (Figure 1). For IOLs of low refractive index and/or large power, this leads to a clear optic diameter below the value of the primary IOL radial size. The geometry of the sulcus-fixated secondary IOL was based on a technical drawing that was provided by the manufacturer. The haptics were not considered for the primary and the secondary IOLs. In the optical model, the IOLs are free standing in aqueous humor with no mechanical support.

Figure 1
Figure 1:
Geometry of a standard spherical equi-biconvex IOL with a standard edge.

The optical model was a spherical equivalent (SE) model of the eye: corneal astigmatism was not considered and a single curvature was included in the model, taken as the average of flat keratometry and steep keratometry. When an aspheric toric sulcus-fixated IOL was present, it was modeled as a spherical IOL of power equal to the SE of the toric IOL.

The retina was modeled by a spherical nonsequential detector of homogeneous sensitivity. The capsular bag was modeled with a shield, blocking any rays that missed the primary IOL. The 3-D Ganzfeld source was built as a spherical source of 1 m diameter emitting rays at each point of its surface following a Lambertian law. Chromaticity and ray splitting were not considered. For each eye, for 4 iris apertures, and for preoperative and postoperative cases, the full number of rays were traced and the light irradiance was calculated at each point of the detector. The data were then extracted in terms of irradiance per angle on retina (Figure 2). Because the eye model has a complete symmetry under rotation around the optical axis, data were averaged in a cylindrical fashion allowing for a significant reduction of the noise that originates from the finite number of rays.

Figure 2
Figure 2:
Example of the irradiance results obtained for the eye of patient 8. A: Defined angle on retina. B and D: Results of the ray tracing for patient 8 (complete resolution of negative dysphotopsia) and patient 2 (no resolution of negative dysphotopsia), shown as the irradiance as a function of the angle on the retina for both the preoperative and postoperative case and for 4 different iris apertures. C and E: Irradiance as a function of the visual angle.

Following this, a conversion of the results from retinal space to visual space was carried out. At first, the relationship between the visual angle (eccentricity) and the angle on the retina was established by tracing rays from a collimated source placed at a variable eccentricity and by identifying the coordinate of the center of the light spot created on the retina. Thereafter, the irradiance results were converted to the visual space. The contribution of the rays hitting the retina before the capsular bag block was prevented.

Analysis of Ray-Tracing Data

The irradiance of light at the retina was evaluated as a function of degrees of eccentricity relative to the centrum of visual field and simulated with 4 iris apertures as follows: 1.5 mm, 2.5 mm, 3.5 mm, and 5.0 mm. For better comparison and to compensate for a decrease in light irradiance with small pupils, the ratio was calculated between the mean irradiance in the peripheral retinal region (60 to 70 degrees of eccentricity) and the mean irradiance in a more central region of the retina (20 to 30 degrees of eccentricity), both for preoperative and postoperative data (Table S1, available at http://jcrsjournal.org). The postoperative ratio was divided by the preoperative ratio to assess for possible relative improvement in perceived light irradiances at eccentricities between 60 degrees and 70 degrees after sulcus-fixated IOL implantation.

Statistical Analysis

Data were collected using Excel software (Microsoft Office 2010, Microsoft Corp.) and analyzed using SPSS software (version 23.0, IBM Corp.). The means ± SD were calculated for continuous variables. The statistical difference between the means of preoperative and postoperative parameters in patients with and without improvement of negative dysphotopsia was assessed with the independent-samples t test. The paired-samples t test was used to evaluate statistical differences between the simulated irradiance changes before and after sulcus-fixated IOL implantation. The statistical difference between the means of simulated relative irradiance changes at the periphery in patients with partial, complete, and no resolution of negative dysphotopsia was assessed with analysis of variance; P values less than 0.05 were considered statistically significant.

Results

Clinical Data

A sulcus-fixated IOL was implanted in 8 patients (10 eyes) with negative dysphotopsia (Tables 3A and 3B). The mean age of the patients was 67.6 ± 3.2 years. These patients reported a bothersome black crescent, a black edge, or a blurred temporal spot. The symptoms developed after cataract surgery and increased in photopic conditions. After sulcus-fixated IOL implantation, negative dysphotopsia resolved completely in 6 eyes, partially in 2 eyes, and persisted in 2 eyes.

Table 3A
Table 3A:
Patient characteristics and course of negative dysphotopsia after sulcus-fixated IOL implantation.
Table 3B
Table 3B:
Supplementary implantation of sulcus-fixated IOL.

In 5 eyes, kinetic perimetry was performed before and after implantation of the sulcus-fixated IOL (Figure 3). Before surgery, 2 patients (3 eyes) reported the constriction of the inferior–temporal or temporal visual field by a black shadow and 2 patients (2 eyes) reported a blurred gray temporal spot. After sulcus-fixated IOL implantation, the extension of the visual field increased in patients reporting black shadows (Case 4 and 8) and both patients also reported subjective improvement of negative dysphotopsia. In Case 3, a relative scotoma was mapped during perimetry. After supplementary implantation of the sulcus-fixated IOL, the relative scotoma increased in size concurrently with increased symptoms of negative dysphotopsia. In Case 5, a relative scotoma found before surgery completely resolved concurrently with resolution of the negative dysphotopsia after sulcus-fixated IOL implantation.

Figure 3
Figure 3:
The V4e isopters of patients with negative dysphotopsia before (gray line) and after (blue line) sulcus-fixated IOL implantation. (The preoperative visual fields of these cases were reported previously.11)

There were no statistically significant differences in age, anterior chamber dimensions, AL, IOL power, and photopic pupil diameters between patients with (partial) resolution and persistence of negative dysphotopsia (Table 4). There was a difference in the chord μ between patients assessed using the Holladay report of the Scheimpflug photography device. Two patients with no resolution of negative dysphotopsia had a larger chord μ of 0.66 mm (Case 2) and 0.31 mm (Case 3) after surgery than patients with complete or partial resolution of negative dysphotopsia (range 0.14 to 0.24 mm). However, the preoperative data of chord μ were not available in all cases and therefore, it was not reported. In 2 patients (Case 5, left eye and Case 6, left eye) the optic–haptic junction of the primary IOL and sulcus-fixated IOL was oriented superior-temporal at 1 o’clock and 7 o’clock and in 1 patient (Case 8), the optic–haptic junction of both the primary and sulcus-fixated IOLs was oriented horizontally at 3 o’clock and 9 o’clock. The data of haptic junction of the IOLs’ orientation were not available for other cases.

Table 4
Table 4:
Preoperative and postoperative characteristics of patients with and without improvement in negative dysphotopsia after sulcus-fixated IOL implantation.

Optical Modeling

The irradiances of light between 60 degrees and 70 degrees of visual field were smaller than between 20 to 30 degrees of visual field (Figure 4 and Table S1, available at http://jcrsjournal.org). Before sulcus-fixated IOL implantation, there was a marked decrease in the light irradiance at the periphery with a 1.5 mm pupil diameter that was less obvious with larger pupil diameters (Figure 4, A). After sulcus-fixated IOL implantation, ray-tracing simulation showed a significant increase in light irradiance at the periphery with 1.5 mm pupil diameter (mean difference in irradiance ratios of 0.22, P < .05) (Figures 4, B, and 5). This increase was greater for patients in which the negative dysphotopsia resolved completely than those with partial resolution of negative dysphotopsia or persistent negative dysphotopsia (Figure 6), although not significant (P = .065). There was no improvement in patients with larger pupils. In addition, there was no consistent relationship between light irradiance at the periphery and anterior or posterior chamber dimensions, AL, or IOL power (data not shown).

Figure 4
Figure 4:
Preoperative (A) and postoperative (B) light irradiance decrease in the peripheral retinal region (60 to 70 degrees of eccentricity) relative to the mean irradiance in a more central region of the retina (20 to 30 degrees of eccentricity) simulated with 1.5 mm, 2.5 mm, 3.5 mm, and 5.0 mm iris aperture.
Figure 5
Figure 5:
Relative irradiance improvement in the peripheral retinal region calculated by dividing the postoperative irradiance ratio by the preoperative irradiance ratio (ND = negative dysphotopsia).
Figure 6
Figure 6:
Relative light irradiance improvement in the peripheral retinal region after surgery simulated with 1.5 mm iris aperture in eyes with complete (n = 6), partial (n = 2), and no resolution (n = 2) of negative dysphotopsia (ND = negative dysphotopsia).

Discussion

The development of negative dysphotopsia after cataract surgery as well as the chance of its resolution, either spontaneously or after a variety of treatments, cannot yet be predicted. In addition, even after extensive research, the pathophysiologic mechanisms underlying negative dysphotopsia are still unclear. Various IOL-related and eye-related factors have been suggested to play a role in the pathogenesis of negative dysphotopsia1–3,7–17,20–22,28 and therefore, negative dysphotopsia is considered at present to be a multifactorial phenomenon caused by an interaction between IOL optics and a predisposed eye.1–3,18,19 This conception is also supported by the current study in which optical modeling that combines all biometry results of a given patient seems to provide some insight into the course of negative dysphotopsia.

In this study, we did not find a relationship between the course of negative dysphotopsia after implantation of a sulcus-fixated IOL and the characteristics of the IOLs or patients such as age, IOL power, AL, anterior and posterior chamber dimensions, or photopic pupil diameter. In our previous study,11 we reported that negative dysphotopsia patients, had significantly shorter eyes than asymptomatic controls (22.7 ± 0.7 mm versus 24.4 ± 0.93 mm) and were implanted with IOLs of higher power (23.8 ± 2.2 D versus 19.5 ± 2.5 D). Although the eyes with negative dysphotopsia analyzed in the current study had a similar mean AL of 22.7 ± 0.7 mm and a mean IOL power of 23.7 ± 2.5 D, these parameters were not statistically different in patients with and without resolution of negative dysphotopsia after surgery. However, an equi-convex design in combination with a higher dioptric power as well as square edges and acrylic material of the IOL might have predisposed our patients to the development of negative dysphotopsia.7

Although negative dysphotopsia symptoms were worse in photopic conditions, we did not find a statistically significant relationship between the course of negative dysphotopsia and photopic pupil diameters before or after sulcus-fixated IOL implantation. Although worsening of negative dysphotopsia in photopic conditions is acknowledged by previous research,2,3,5,6 negative dysphotopsia might either improve2 or persist in mydriasis (Case 3 of this study). To our knowledge, no systematic intraindividual evaluation of negative dysphotopsia symptoms with different pupil diameters and in different light conditions has yet been reported.

According to a recent study,7 angle κ can be considered as 1 of the primary causative factors and a chord length κ (approximation of angle κ) longer than 0.44 mm can be associated with a higher risk for negative dysphotopsia. All but 1 of our patients had a chord length κ shorter than 0.44 mm. Of note, the 2 patients with a longer chord length κ had no resolution of negative dysphotopsia after surgery. Chord length κ was the only biometric factor that was associated with the course of negative dysphotopsia in this study.

It has been proposed that a distance between an IOL and a posterior iris surface larger than 0.5 mm might predispose some patients to negative dysphotopsia and that reduction of this distance might lead to resolution of symptoms.3,16 In contrast, the last optical simulation analysis by Holladay and Simpson7 suggested a smaller axial distance of the IOL behind the iris as the primary causative factor. In this study, the iris–IOL distance measured with AS-OCT before surgery was less than 0.5 mm in 2 eyes and larger than 0.5 mm in 8 eyes, and after surgery the sulcus-fixated IOL was adjacent to the posterior iris surface in all cases. We did not find a significant relationship between the iris–IOL distance and the course of negative dysphotopsia after sulcus-fixated IOL implantation. Moreover, in 2 eyes, negative dysphotopsia persisted despite a shallow posterior chamber after sulcus-fixated IOL implantation.

Recently, an inferior–temporal location of the optic–haptic junction of the IOL was shown to be associated with a 2.3-fold decrease in the risk for transient negative dysphotopsia at 1 day after surgery compared with a vertical location of the optic–haptic junction.12 However, this association was absent at 1 week and 1 month after surgery and thus was not considered to be related to persistent negative dysphotopsia.12 In our study, the optic–haptic junction was oriented superior-temporally in 2 cases and horizontally in 1 case with persistent negative dysphotopsia, which was completely resolved after implantation of the sulcus-fixated IOL. Because the optic–haptic junction of the sulcus-fixated IOL was positioned above that of the primary IOL in these cases, it is unlikely that the position of the optic–haptic junction of sulcus-fixated IOL could lead to the resolution of negative dysphotopsia.

In a previous study,11 we found that patients with negative dysphotopsia compared with asymptomatic pseudophakic patients had either constriction of their inferior temporal visual field or a relative scotoma between 50 degrees and 90 degrees of the temporal visual fields mapped by kinetic perimetry. Five eyes from the previous study were treated with implantation of the sulcus-fixated IOL. Visual field changes, such as an increase in the constricted visual field and a resolution or increase in the relative scotoma after sulcus-fixated IOL implantation, matched the course of negative dysphotopsia after surgery. This confirms our statement that kinetic perimetry can support the diagnosis of negative dysphotopsia and should be used for the objective evaluation of the course of negative dysphotopsia.

To account for the multifactorial nature of negative dysphotopsia, we combined IOL-related parameters with clinical data in patient-specific eye models before and after sulcus-fixated IOL implantation. In our analysis, we focused on the simulated light irradiance between 60 degrees and 70 degrees of visual field because changes were noticed in this region before and after surgery. Our approach showed that all simulated cases had a marked decrease in light irradiance at the periphery before sulcus-fixated IOL implantation that was especially prominent with the smallest pupil diameter of 1.5 mm. Implantation of the sulcus-fixated IOL led to a significant increase (P < .05) in the simulated light irradiance at the periphery with the 1.5 mm pupil diameter, which was smaller although not significant (P = .065) in cases in which negative dysphotopsia persisted after surgery compared with cases in which an improvement or a complete resolution of the symptoms occurred. Holladay and Simpson7 suggested that implantation of a piggyback IOL can shift the rays refracted by the IOL and thereby move the shadow toward a more peripheral part of the visual field. The shift of light irradiance caused by introduction of the plano surfaces of the sulcus-fixated IOL in the optical system in combination with the effect of the larger optic diameter (6.5 mm) of this IOL7 might lead to the postoperative increase in light irradiance in our study.

A recent ray-tracing analysis of Holladay and Simpson,7 based on general pseudophakic eye models, showed a 2.6-degree shadow between 85.7 degrees and 88.3 degrees of retinal field angle, which was formed between the rays refracted by the IOL and the rays missing the IOL, that would potentially result in a shadow near 90 degrees of the visual field. Our patient-specific ray-tracing modeling showed a decrease in irradiance despite the blockage of rays missing the IOL. Moreover, this decrease occurred more centrally than the shadow described by Holladay and Simpson,7 and is in accordance with our clinical perimetric data.

In the present study, we did not find a clear correlation between simulated decreased light irradiance at the periphery and biometric data of the simulated cases. This might be related to the small number of cases with persistence of negative dysphotopsia after surgery and the limitations of our modeling. Only 10 eyes were simulated in this study, and in 5 eyes, the results were supported by Goldmann kinetic perimetry for which variable visual field changes (increase of the extension of inferior temporal visual field, resolution, or increase of the relative scotoma) were noted. Peripheral visual field evaluation should be included in routine examinations of patients with negative dysphotopsia to establish a specific region of interest in which shadows can be detected.29 A more extensive combination of clinical and laboratory research is required to predict and understand the effect of sulcus-fixated IOLs as well as the other negative dysphotopsia treatment modalities on a patient-by-patient basis.

All patients were implanted with square-edged IOLs: Acrysof SN60WF (Alcon Laboratories, Inc.), Tecnis ZCB00, or Tecnis PCB00 (both Abbott Medical Optics, Inc.) (Table 3A). However, a simplified model of the primary IOL (spherical IOL with a standard sharp edge) was used in all cases. Because the data about transparency of the anterior and peripheral lens capsule were not documented in all cases, we did not simulate the effect of the capsulorhexis. Therefore, our results cannot be compared with the modeling of Hong et al.,23 where the interaction between anterior capsulorhexis and IOL was proposed to cause negative dysphotopsia by creating shadows on the retina. Although various treatment strategies directed to the IOL–capsule bag interface have been suggested, there are very limited data on the success rates. Reverse optic capture resulted in complete resolution of negative dysphotopsia in 3 cases,2 nasal IOL optic truncation in complete resolution in 1 case,30 and Nd:YAG anterior capsulotomy in complete resolution of negative dysphotopsia in 4 cases, in partial resolution in 2 cases, and no resolution in 1 case.8,9 The potential role of the IOL–capsule bag interface was further highlighted by a successful preventive role of primary reverse optic capture in patients with negative dysphotopsia in the first eye who were scheduled for cataract surgery in the fellow eye.2,31 An additional modeling with patient-specific data (including visual field evaluation and assessment of capsular bag transparency anterior and peripheral to the IOL optic) might help to further evaluate the role of capsulorhexis as well as the rays missing the IOL in the pathogenesis of negative dysphotopsia. However, a clinical technique should be developed first for the assessment of capsular bag transparency in vivo.

Simulation of the piggyback IOL by Hong et al.23 showed multiple surface reflections that filled the shadow created by the IOL–capsulorhexis interface. In the current study, ray splitting was not considered. This assumption is not thought to affect the results of the modeling as most of the light energy is carried by the main ray and our analysis relies on irradiance distribution, that is, energy per unit of time and area (the energy reflection coefficient is 0.3% for an IOL of index 1.5 placed in the eye). Thus, the sole multiple reflections included in the model were multiple total internal reflections. These reflections occur inside the sulcus-fixated IOL and were observed only for extreme angles and pupil diameters; these rays were found to consistently fall outside the range of the visual field and were therefore disregarded as a contributor to the appearance or resolution of negative dysphotopsia.

In conclusion, from all individual biometry results, only angle κ showed a relationship with the course of negative dysphotopsia after sulcus-fixated IOL implantation. Ray-tracing modeling that combines individual biometry data showed an increase in the light irradiance at the periphery after sulcus-fixated IOL implantations that was also associated with the course of negative dysphotopsia. More cases as well as improvements in the model are required to understand anatomical or IOL-related factors that might explain this association. Possible improvements to the model include (1) a more realistic model of the capsular bag combined with developing a clinical diagnostic method; (2) a more accurate model of the IOL with regard to the optic, edge, and haptic design, including the effects of ray splitting; (3) building a non-rotationally symmetric eye model that considers patient-specific angle κ; and (4) simulation of the effect of higher-order aberrations. Most importantly, the model can be used to assess in greater detail the reasons for the change in irradiance patterns before implantation and after implantation of a secondary IOL, and gain a better understanding of the reasons for the difference between patients. This might open the path to predict the effect of implantation of a sulcus-fixated IOL in treating negative dysphotopsia symptoms on a patient-by-patient basis.

What Was Known

  • Negative dysphotopsia is a common side effect of cataract surgery that is probably caused by blockage of light at the peripheral nasal retina as a result of the interaction between the optics of the IOL and the anatomically predisposed eye.
  • Sulcus-fixated IOL implantation in the ciliary sulcus can effectively treat severe negative dysphotopsia in about 80% of cases, although the mechanism of its effect is currently unknown.

What This Paper Adds

  • The course of negative dysphotopsia after sulcus-fixated IOL implantation was not related to the patient’s age or individual biometry results such as AL, IOL power, photopic pupil diameter, or dimensions of anterior and posterior eye chambers.
  • Ray-tracing models showed a decrease in light irradiance at the periphery relative to the center of visual field in eyes, which increased significantly after sulcus-fixated IOL implantation.
  • Combining individual biometric parameters and ray-tracing optical modeling showed a promising association with the course of negative dysphotopsia after sulcus-fixated IOL implantation and might be helpful in explaining the multifactorial nature of negative dysphotopsia.

References

1. Davison JA. Positive and negative dysphotopsia in patients with acrylic intraocular lenses. J Cataract Refract Surg. 2000;26:1346-1355.
2. Masket S, Fram NR. Pseudophakic negative dysphotopsia: surgical management and new theory of etiology. J Cataract Refract Surg. 2011;37:1199-1207.
3. Osher RH. Negative dysphotopsia: long-term study and possible explanation for transient symptoms. J Cataract Refract Surg. 2008;34:1699-1707.
4. Stulting RD. Cataract surgical problem. In: Masket S, editor. Consultation section, 31. J Cataract Refract Surg; 2005. p. 651-652.
5. Trattler WB, Whitsett JC, Simone PA. Negative dysphotopsia after intraocular lens implantation irrespective of design and material. J Cataract Refract Surg. 2005;31:841-845.
6. Henderson BA, Geneva II. Negative dysphotopsia: a perfect storm. J Cataract Refract Surg. 2015;41:2291-2312.
7. Holladay JT, Simpson MJ. Negative dysphotopsia: causes and rationale for prevention and treatment. J Cataract Refract Surg. 2017;43:263-275.
8. Cooke DL, Kasko S, Platt LO. Resolution of negative dysphotopsia after laser anterior capsulotomy. J Cataract Refract Surg. 2013;39:1107-1109.
9. Folden DV. Neodymium:YAG laser anterior capsulectomy: surgical option in the management of negative dysphotopsia. J Cataract Refract Surg. 2013;39:1110-1115.
10. Davison JA. Clinical performance of Alcon SA30AL and SA60AT single-piece acrylic intraocular lenses. J Cataract Refract Surg. 2002;28:1112-1123.
11. Makhotkina NY, Berendschot TTJM, Nuijts RMMA. Objective evaluation of negative dysphotopsia with Goldmann kinetic perimetry. J Cataract Refract Surg. 2016;42:1626-1633.
12. Henderson BA, Yi DH, Constantine JB, Geneva II. New preventative approach for negative dysphotopsia. J Cataract Refract Surg. 2016;42:1449-1455.
13. Narváez J, Banning CS, Stulting RD. Negative dysphotopsia associated with implantation of the Z9000 intraocular lens. J Cataract Refract Surg. 2005;31:846-847.
14. Olson RJ. Cataract surgical problem. In: Masket S, editor. Consultation section, 31. J Cataract Refract Surg; 2005. p. 653-654.
15. Fishkind WJ. Cataract surgical problem. In: Masket S, editor. Consultation section, 32. J Cataract Refract Surg; 2006. p. 908-909.
16. Vámosi P, Csákány B, Németh J. Intraocular lens exchange in patients with negative dysphotopsia symptoms. J Cataract Refract Surg. 2010;36:418-424.
17. Masket S, Fram N. Etiology of negative dysphotopsia [letter]. J Cataract Refract Surg. 2013;39:485-486.
18. Radford SW, Carlsson AM, Barrett GD. Comparison of pseudophakic dysphotopsia with Akreos Adapt and SN60-AT intraocular lenses. J Cataract Refract Surg. 2007;33:88-93.
19. Simpson MJ. Double image in far peripheral vision of pseudophakic eye as source of negative dysphotopsia. J Opt Soc Am A Opt Image Sci Vis. 2014;31:2642-2649.
20. Holladay JT. Reply to letter by S Masket and N Fram. Etiology of negative dysphotopsia. J Cataract Refract Surg. 2013;39:486.e1-486.e4.
21. Holladay JT, Zhao H, Reisin CR. (2012). Negative dysphotopsia: the enigmatic penumbra. J Cataract Refract Surg, 38, 1251-1265, Available at: http://www.hicsoap.com/publications/107%20Neg%20Dys%20FINAL%20PUB%20JCRS%20July%202012_38_1251-1265.pdf Accessed 16-12-2017
22. Arnold PN. Cataract surgical problem. In: Masket S, editor. Consultation section, 31. J Cataract Refract Surg; 2005. p. 1487-1488.
23. Hong X, Liu Y, Karakelle M, Masket S, Fram NR. Ray-tracing optical modeling of negative dysphotopsia. J Biomed Opt. 2011;16(12):125001.
24. Burke TR, Benjamin L. Sulcus-fixated intraocular lens implantation for the management of negative dysphotopsia. J Cataract Refract Surg. 2014;40:1469-1472.
25. Makhotkina NY, Berendschot TTJM, Beckers HJM, Nuijts RMMA. Treatment of negative dysphotopsia with supplementary implantation of a sulcus-fixated intraocular lens. Graefes Arch Clin Exp Ophthalmol. 2015;253:973-977.
26. Weinstein A. Surgical experience with pseudophakic negative dysphotopsia [letter]. J Cataract Refract Surg. 2012;38:561. reply by S Masket, 561.
27. Gross H, Blechinger F, Achtner B. 2008. Human Eye. In: Gross H, Blechinger F, Achtner B, editors., Handbook of Optical Systems, volume 4: Survey of Optical Instruments, Wiley-VCH Verlag GmbH & Co. KGaA, Weinheim, Germany, chapter 36.
28. Franchini A. Cataract surgical problem. In: Masket S, editor. Consultation section, 31. J Cataract Refract Surg; 2005. p. 656.
29. Simpson MJ. Mini-review: far peripheral vision. Vision Res. 2017;140:96-105.
30. Alapati NM, Harocopos GJ, Sheybani A. (2016). In-the-bag nasal intraocular lens optic truncation for treatment of negative dysphotopsia. J Cataract Refract Surg, 42, 1702-1706, Available at: http://www.jcrsjournal.org/article/S0886-3350(16)30454-0/pdf Accessed 16-12-2017
31. Narang P, Narang S. Primary reverse optic capture with implantation of capsular tension ring to prevent pseudophakic negative dysphotopsia. J Cataract Refract Surg. 2015;41:891-892.

Supplementary data

Figure
Figure:
No Caption available.

Supplementary data

Supplemental material available at www.jcrsjournal.org.

Disclosures

Drs. Dugrain and Purchase are employees of Rayner Intraocular Lenses, Ltd. Dr. Nuijts is a consultant to Alcon Laboratories, Inc. ASICO LLC, and Théa Pharma GmbH. He received study grants from Abbott Medical Optics, Inc., Alcon Laboratories, Inc., Gebauer Co., Humanoptics AG, Oculentis GmbH, and Ophtec BV. He received lecture fees from Alcon Laboratories, Inc. None of the other authors has a financial or proprietary interest in any material or method mentioned.