Comparison of ocular biometry and intraocular lens power using a new biometer and a standard biometer

Srivannaboon, Sabong MD; Chirapapaisan, Chareenun MD*; Chonpimai, Pratuangsri BS; Koodkaew, Sunisa BS

Author Information
Journal of Cataract & Refractive Surgery: May 2014 - Volume 40 - Issue 5 - p 709-715
doi: 10.1016/j.jcrs.2013.09.020
  • Free

Abstract

Ocular biometry is essential for intraocular lens (IOL) power calculation in cataract surgery. The biometric variables that are used for IOL calculation depend on the chosen IOL formula. Current third-generation and fourth-generation IOL formulas require the basic variables of keratometry (K) values and axial length (AL).1,2 Some formulas may require additional variables, such as anterior chamber depth (ACD) or horizontal white-to-white (WTW) corneal diameter.3 However, the accurate measurement of variables is crucial.

Unlike ultrasound (US) biometry, modern optical biometry devices measure several variables. Moreover, the built-in software in these devices provides more accurate IOL power calculation and multiple choices of IOL formulas.4,5 The IOLMaster (Carl Zeiss Meditec AG) is considered to be one of the first standard modern optical biometry devices.6 It uses the principle of partial coherence interferometry (PCI) to obtain the AL with high precision.7–9 The device also measures the K values, ACD, and WTW diameter.10 Several reports have shown the accuracy of the the newest version of IOLMaster (IOLMaster 500) for IOL calculation in routine and complicated cataract cases.8,11–13

Several new optical biometry devices can also perform ocular biometry and IOL power calculation.14 Among them, the AL-Scan (Nidek Co., Ltd.) is one of the most recently released on the market. The device measures 6 variables, including the K value, AL, ACD, WTW diameter, pupil size, and central corneal thickness (CCT). The IOL power is generated by several formulas built into the device. To our knowledge, no study has evaluated the repeatability, reproducibility, and accuracy of this device.

The purpose of this study was to evaluate the repeatability, reproducibility of variables (eg, K, AL, ACD, horizontal WTW corneal diameter), and IOL power measured using the AL-Scan device and compare the results with those obtained with the IOLMaster 500 device.

Patients and methods

This prospective comparative study recruited cataract patients from Siriraj Eye Clinic, Bangkok, Thailand. Institutional ethic committee approval and patient informed consent were obtained. Eyes with ocular disease other than cataract were excluded.

The AL-Scan optical biometer (new biometer) measures K values using double-mire rings projected onto the cornea at the 2.4 mm zone and 3.3 mm zone. The Scheimpflug imaging technique is applied for CCT and ACD measurements. The AL is acquired by PCI.

The IOLMaster 500 optical biometer (standard biometer) measures K values using 6 spots of light projected onto the cornea at the central 2.5 mm zone. The ACD is measured by lateral slit illumination, and the AL is acquired by the PCI method.

The first operator took 3 consecutive measurements with each device (intraoperator repeatability). The patients were asked to close their eyes between each measurement. Then, the second operator took another measurement (interoperator reproducibility). Four major parameters (K value, AL, ACD, WTW) were recorded. The IOL power by the Holladay 1 formula from both devices was calculated using the IOL constant provided by the User Group for Laser Interference Biometry (ULIB) web site.A

Statistical analysis was performed using SPSS software (version 19.0, SPSS, Inc.). The data normality was checked with the Shapiro-Wilk test. The intraoperator repeatability was calculated using the 3 sets of the measurements taken by the first operator. For interoperator reproducibility, the first measurement of the first operator was selected and compared with the measurement of the second operator. The agreement between the 2 devices was also evaluated using the first measurement of the first operator for each device. An intraclass correlation coefficient (ICC) of more than 0.8 was acceptable as a high value.15 The Pearson correlation was used to determine the relationship between all parameters. A statistically significant difference was defined as P<.05. Bland-Altman plots were used to assess the agreement of measurements between the 2 devices.

Results

The study recruited 137 eyes of 81 patients. The mean age of patients was 65.28 years ± 10.56 (SD). Table 1 shows the values of 4 ocular parameters (K, AL, ACD, WTW) and the calculated IOL power obtained with the new biometer and the standard biometer. The mean value of all measurements, except the WTW distance, and the calculated IOL power showed a good correlation between the 2 devices.

Table 1
Table 1:
Parameters obtained by the 2 optical biometers.

Repeatability

The intraoperator repeatability of both devices was very good. The ICC of the new biometer and the standard biometer ranged from 0.94 to 1.00 (Table 2). For ACD, the new biometer produced a much lower mean difference with a small standard deviation (SD) than the standard biometer. In contrast, the WTW measurement of the new biometer had a slightly lower ICC and wider SD of the mean difference than the standard biometer.

Table 2
Table 2:
Repeatability.

Reproducibility

The interoperator reproducibility of both devices was good, which yielded a high ICC value for all parameters (Table 3). The new biometer produced slightly higher ICC values than the standard biometer for most parameters, especially for the ACD measurement. Moreover, the mean difference and SD of the ACD measurement by the new biometer was much smaller than by the standard biometer. However, the new biometer provided a lower ICC value than the standard biometer for WTW.

Table 3
Table 3:
Reproducibility.

Agreement

The agreement of the ocular parameters derived from both biometry devices was high except for the WTW distance (Table 4). The mean difference and SD of the WTW parameter had the highest values of all parameters. However, the Bland-Altman plots showed that the limits of agreement (LoA) between the 2 biometers remained narrow over all parameters (Figures 1 to 5).

Figure 1
Figure 1:
Agreement of K measurements between biometers. The solid lines show the mean difference and 95% LoA for K at the 2.4 mm zone. The dashed lines show the mean difference and 95% LoA for K at the 3.3 mm zone (K = keratometry).
Figure 2
Figure 2:
Agreement of AL measurements between biometers. The lines show the mean difference and 95% LoA (AL = axial length).
Figure 3
Figure 3:
Agreement of ACD measurements between biometers. The lines show the mean difference and 95% LoA (ACD = anterior chamber depth).
Figure 4
Figure 4:
Agreement of WTW corneal diameter measurements between biometers. The lines show the mean difference and 95% LoA (WTW = white-to-white corneal diameter).
Figure 5
Figure 5:
Agreement of IOL power measurements between biometers. The lines show the mean difference and 95% LoA (IOL = intraocular lens).
Table 4
Table 4:
Agreement of each parameter between the new biometer and standard biometer.

Discussion

This study found that the AL-Scan biometer (new biometer) and the IOLMaster 500 biometer (standard biometer) had high repeatability and reproducibility for all tested parameters. The agreement between them was also very high except for the WTW parameter. The WTW diameter measured with the new biometer was significantly smaller than that measured with the standard biometer, and there was low correlation between the measurements.

Today, ophthalmologists are gaining more accessibility to optical biometry devices. In a study of cataract surgery practice patterns in the United States Veterans Health Administration,16 83% of the respondents said they used PCI devices for preoperative biometry. At present, several optical biometry devices are used in clinical practice. However, different types of optical biometry devices may yield different IOL power results. Most IOL power calculations based on PCI technology use the IOL constant provided by the ULIB web site.A The web site currently provides the optimized IOL constant for 2 optical biometers only; that is, the IOLMaster and the Lenstar (Haag-Streit AG). A new device that uses the ULIB constant must be verified against the standard biometer. Therefore, we chose to validate the AL-Scan biometer using the IOLMaster 500 biometer. The parameters in this study (K, AL, ACD, WTW) were chosen for comparison because they are required by current third-generation and fourth-generation IOL formulas.2,3,17 The new biometer measures the K value at 2 zone diameters (2.4 mm and 3.3 mm), while the standard biometer measures it at 2.5 mm only. The other parameters are measured similarly by the 2 devices.

In this study, all parameter measurements, except the WTW, were comparable between devices. The WTW from the IOLMaster device has been shown to be precise.18,19 There had been no study of the WTW diameter measured by the AL-Scan device. The mean difference in the WTW measured by the 2 devices was 0.51 mm. The agreement and correlation between them was also low (ICC = 0.448 and r = 0.445, respectively). The explanation for this finding could be the difference in the algorithm for edge detection around iris image between the 2 devices. Furthermore, the dissimilarity in the light source for image acquisition between the devices could have a role. The AL-Scan device uses a green light source (wavelength 525 nm) to measure the WTW, whereas the IOLMaster device uses an infrared light source (wavelength 880 nm).B,C The infrared light source has been used in ophthalmic devices for eye tracking with high accuracy for decades.20–22 The infrared light might be a better choice for the WTW measurement. Even though the repeatability and reproducibility of the WTW measured from the AL-Scan device were quite high (ICC 0.945 and ICC 0.873, respectively), they were lower than those with the IOLMaster 500 device (ICC 0.974 and ICC 0.946, respectively). Moreover, the SDs of the measurement difference in repeatability and reproducibility of the WTW measurement were higher for the AL-Scan device (0.58 and 0.50, respectively) than for the IOLMaster 500 device (0.17 and 0.20, respectively). Therefore, the WTW measurement from the AL-Scan device should be used with caution.

The ACD measurement by both devices in this study was similar. However, the new biometer had slightly better repeatability and reproducibility. The mean difference and its SD were lower for the new biometer than for the standard biometer (repeatability 0.003 ± 0.014 versus 0.029 ± 0.106; reproducibility 0.004 ± 0.01 versus −0.022 ± 0.10). The LoA range was also lower for the new biometer. This finding could be the result of the Scheimpflug image principle used for the ACD measurement by the new biometer; the standard biometer uses a scanning-slit image. The measurement from the Scheimpflug image has been shown to have better repeatability than slit imaging, US biomicroscopy, and magnetic resonance imaging.23,24 However, the ACD measured using a different method was considerably below a clinically significant level.

The IOL power (Holladay 1 formula) derived by both devices was quite similar, with good agreement and correlation. The mean difference was only 0.26 diopter (D), less than the increment in the IOL power step (0.50 D). Although both devices did not have a clinically significant impact on IOL power, the chosen IOL formula should be considered. If the WTW measurement is required in a formula, such as in the Holladay 2, the result might be different.

In the Bland-Altman plots, a few cases were not within the LoA. But in most cases, the difference between the 2 devices was considered not clinically significant. The maximum difference in K value was 0.48 D, in AL was 0.07 mm, in ACD was 0.3 mm, in WTW was 2.3 mm, and in IOL power was 0.78 D. In cases in which the difference in IOL power was more than 0.50 D, it was mainly the result of the difference in the K value.

This study was not intended to compare penetration through a dense cataract between the 2 devices. However, the 2 devices performed very similarly. In eyes in which the AL cannot be obtained by the IOLMaster 500 device, it cannot be obtained by the AL-Scan device either.

A limitation of this study might be both eyes of the some patients were enrolled. In this study, we recruited both eyes of 56 patients (112 eyes) and 1 eye of 25 patients. A previous study25 discussed the bias effects of collecting the data from both eyes and comparing them with data from 1 eye. However, the aim of our study was to compare the performance of the 2 devices; thus, the data obtained from 1 eye or 2 eyes should not significantly affect the accuracy of either biometer.

In conclusion, the AL-Scan biometer performed very similarly to the IOLMaster 500 biometer. They both showed very high repeatability and reproducibility and were in good agreement with each other. The WTW measurement from the AL-Scan biometer should be further verified.

What Was Known

  • The standard device in the study is one of the most accurate PCI optical biometers and had high repeatability and reproducibility.

What This Paper Adds

  • The repeatability and reproducibility of the new PCI optical biometer was high.
  • The accuracy of the new biometer was comparable to that of the standard biometer. In most cases, the mean ocular parameter values, repeatability, and reproducibility were similar between the 2 biometers.
  • The WTW measurements were different between the 2 biometers. The results have to be verified. Therefore, the WTW distance derived from the new biometer should be used with caution.

References

1. Lee AC, Qazi MA, Pepose JS. Biometry and intraocular lens power calculation. Curr Opin Ophthalmol. 2008;19:13-17.
2. Joo J, Whang W-J, Oh T-H, Kang K-D, Kim H-S, Moon J-I. Accuracy of intraocular lens power calculation formulas in primary angle closure glaucoma. Korean J Ophthalmol. 25, 2011, p. 375-379, Available at http://www.ncbi.nlm.nih.gov/pmc/articles/PMC3223703/pdf/kjo-25-375.pdf. Accessed November 28, 2013.
3. Terzi E, Wang L, Kohnen T. Accuracy of modern intraocular lens power calculation formulas in refractive lens exchange for high myopia and high hyperopia. J Cataract Refract Surg. 2009;35:1181-1189.
4. Wang J-K, Chang S-W. Optical biometry intraocular lens power calculation using different formulas in patients with different axial lengths. Int J Ophthalmol. 6, 2013, p. 150-154, Available at http://www.ncbi.nlm.nih.gov/pmc/articles/PMC3633751/pdf/ijo-06-02-150.pdf. Accessed November 28, 2013.
5. Rabsilber TM, Jepsen C, Auffarth GU, Holzer MP. Intraocular lens power calculation: clinical comparison of 2 optical biometry devices. J Cataract Refract Surg. 2010;36:230-234.
6. Sheng H, Bottjer CA, Bullimore MA. Ocular component measurement using the Zeiss IOLMaster. Optom Vis Sci. 81, 2004, p. 27-34, Available at http://journals.lww.com/optvissci/Fulltext/2004/01000/The_Contact_Lens_and_Myopia_Progression__CLAMP_.7.aspx. Accessed November 28, 2013.
7. Vogel A, Dick HB, Krummenauer F. Reproducibility of optical biometry using partial coherence interferometry: intraobserver and interobserver reliability. J Cataract Refract Surg. 2001;27:1961-1968.
8. Kunavisarut P, Poopattanakul P, Intarated C, Pathanapitoon K. Accuracy and reliability of IOL master and A-scan immersion biometry in silicone oil-filled eyes. Eye. 26, 2012, p. 1344-1348, Available at http://www.nature.com/eye/journal/v26/n10/pdf/eye2012163a.pdf. Accessed November 28, 2013.
9. Fontes BM, Fontes BM, Castro E. Intraocular lens power calculation by measuring axial length with partial optical coherence and ultrasonic biometry. Arq Bras Oftalmol. 74, 2011, p. 166-170, Available at http://www.scielo.br/pdf/abo/v74n3/04.pdf. Accessed November 28, 2013.
10. Fotedar R, Wang JJ, Burlutsky G, Morgan IG, Rose K, Wong TY, Mitchell P. Distribution of axial length and ocular biometry measured using partial coherence laser interferometry (IOL Master) in an older white population. Ophthalmology. 2010;117:417-423.
11. Hsieh Y-T, Wang I-J. Intraocular lens power measured by partial coherence interferometry. Optom Vis Sci. 2012;89:1697-1701.
12. Roessler GF, Dietlein TS, Plange N, Roepke A-K, Dinslage S, Walter P, Mazinani BA. Accuracy of intraocular lens power calculation using partial coherence interferometry in patients with high myopia. Ophthalmic Physiol Opt. 2012;32:228-233.
13. Olsen T. Improved accuracy of intraocular lens power calculation with the Zeiss IOLMaster. Acta Ophthalmol Scand. 85, 2007, p. 84-87, Available at http://onlinelibrary.wiley.com/doi/10.1111/j.1600-0420.2006.00774.x/pdf. Accessed November 28, 2013.
14. Chen Y-A, Hirnschall N, Findl O. Evaluation of 2 new optical biometry devices and comparison with the current gold standard biometer. J Cataract Refract Surg. 2011;37:513-517.
15. Rosner B. Fundamentals of Biostatistics. 5th ed. 2000, Duxbury Press, Pacific Grove, CA, p. 562-563
16. Greenberg PB, Havnaer A, Oetting TA, Garcia-Ferrer FJ. Cataract surgery practice patterns in the United States Veterans Health Administration. J Cataract Refract Surg. 2012;38:705-709.
17. Hoffer KJ. Clinical results using the Holladay 2 intraocular lens power formula. J Cataract Refract Surg. 2000;26:1233-1237.
18. Baumeister M, Terzi E, Ekici Y, Kohnen T. Comparison of manual and automated methods to determine horizontal corneal diameter. J Cataract Refract Surg. 2004;30:374-380.
19. Kiraly L, Duncker G. Biometrie des vorderen Augensegmentes zur Implantation phaker Vorderkammerlinsen. Ein Vergleich aktueller Messgeräte [Biometry of the anterior eye segment for implantation of phakic anterior chamber lenses. A comparison of current measurement devices]. Ophthalmologe. 2012;109:242-249.
20. Okuyama F, Tokoro T, Fujieda M. Eye-tracking infra-red optometer. Ophthalmic Physiol Opt. 1990;10:291-299.
21. Pruehsner W, Enderle JD. Eye tracker. Biomed Sci Instrum. 1999;35:235-240.
22. Fassi A, Riboldi M, Forlani CF, Baroni G. Optical eye tracking system for noninvasive and automatic monitoring of eye position and movements in radiotherapy treatments of ocular tumors. Appl Opt. 2012;51:2441-2450.
23. Wolffsohn JS, Davies LN. Advances in anterior segment imaging. Curr Opin Ophthalmol. 2007;18:32-38.
24. Koretz JF, Strenk SA, Strenk LM, Semmlow JL. Scheimpflug and high-resolution magnetic resonance imaging of the anterior segment: a comparative study. J Opt Soc Am A Opt Image Sci Vis. 2004;21:346-354.
25. Armstrong RA. Statistical guidelines for the analysis of data obtained from one or both eyes. Ophthalmic Physiol Opt. 33, 2013, p. 7-14, Available at http://onlinelibrary.wiley.com/doi/10.1111/opo.12009/pdf. Accessed November 28, 2013.