Phakic refractive lens implantation in high myopic patients

One-year results

Pallikaris, Ioannis G MD, PhDa,b; Kalyvianaki, Maria I MD*,a; Kymionis, George D MD, PhDa,b; Panagopoulou, Sophia I BSca

Author Information
Journal of Cataract & Refractive Surgery: June 2004 - Volume 30 - Issue 6 - p 1190-1197
doi: 10.1016/j.jcrs.2003.10.039
  • Free

Phakic intraocular lenses constitute an evolving technique in the field of refractive surgery for the correction of moderate to high refractive errors. In such cases, excimer laser treatment is limited by the amount of corneal tissue that can be removed safely.1,2 Furthermore, there is evidence that altering the shape of the cornea in attempted high corrections may result in poor quality of vision.3 The implantation of a phakic intraocular lens (IOL) does not affect the shape of the cornea. The technique has been proven to be stable and potentially reversible. In comparison with clear lens extraction, another treatment option for high refractive errors, phakic IOL implantation preserves accommodation and, therefore, is suitable for younger patients.

The PRL is a posterior chamber phakic refractive lens developed by Medennium Inc. and distributed by Ciba Vision Surgical.4 It is made of silicone with a high refractive index (1.46), which allows its ultrathin design. The PRL is not supported in the sulcus angle but “floats” in the posterior chamber over the crystalline lens and is made of hydrophobic material.4 Its centration is achieved by its self-centering design.

The purpose of this prospective study was to evaluate the efficacy and safety of PRL implantation in highly myopic eyes.

Patients and Methods

Thirty-four myopic eyes of 19 patients were treated with PRL implantation by the same surgeon (I.G.P.). Mean patient age was 29.0 years ± 7.9 (SD) (range 18 to 44 years). Each patient had been informed about the procedure, its risks, and its benefits and signed a consent form according to the Declaration of Helsinki. Exclusion criteria included age less than 18 years, previous intraocular surgery, anterior chamber depth less than 3 mm, glaucoma, or intraocular pressure (IOP) at initial measurement greater than 20 mm Hg, any sign of cataract, and any intraocular or systemic disease.

Preoperative examination included manifest and cycloplegic refractions, corneal topography, pachymetry, A-scan ultrasonography (Axis-II, Quantel Medical), slitlamp microscopy, pupil size measurement under scotopic conditions, white-to-white corneal diameter measurement with the use of a caliper, applanation tonometry, measurement of high-order aberrations with the WASCA analyzer (Carl Zeiss, Meditec), and dilated fundoscopy.

Mean preoperative spherical equivalent was −14.70 ± 2.65 diopters (D) (range −20.75 D to −10.50 D). Mean preoperative refractive cylinder was −2.02 D (range 0 to −5.50 D). Manifest refraction was performed over a soft contact lens in all eyes. The target postoperative refraction was emmetropia in all eyes. Preoperative uncorrected visual acuity (UCVA) was finger counting in all eyes; mean best corrected visual acuity (BCVA) was 0.70 ± 0.24 (range 0.10 to 1.00).

Lens power calculations were performed by Ciba Vision Surgical and were based on the preoperative cycloplegic spherical equivalent, the average keratometric power, the anterior chamber depth calculated with the use of A-scan ultrasonography, and the target postoperative refraction. The model of the myopic PRL implanted was based on the horizontal white-to-white diameter. Because this was more than 11.3 mm in all eyes, PRL101 was used in all cases.

Surgical Technique

One hour before surgery, cyclopentolate 1% and phenylephrine 5% were used every 15 minutes to dilate the pupil. Phakic refractive lenses were implanted under retrobulbar anesthesia through a 3.2 mm clear cornea temporal incision made with a diamond knife. The anterior chamber was then filled with a low-viscosity viscoelastic agent. At this step, the special loading block was filled with balanced salt solution and the PRL was placed on the recess with the special forceps. The lens was inserted through the main incision parallel to the iris. With the forceps or a manipulator, the haptics of the lens, one after the other were placed under the iris. An iridectomy was performed at 12 o'clock as peripherally as possible using the probe of a vitreotome.

Postoperative Period

At discharge, each patient was given 1 tablet of acetazolamide 250 mg. Antibiotic-steroid combination drops were prescribed for 2 weeks.

Patients were examined on the first postoperative day, at 1 week and at 1, 3, 6, 9, and 12 months. After the first postoperative day, the examination included UCVA, BCVA, manifest refraction, corneal topography, slit-lamp microscopy, tonometry, and wavefront aberrometry. At 6 and 12 months, the examination also included gonioscopy and dilated fundoscopy.

Statistical Analysis

Group differences for continuous variables were tested using the unpaired and paired Student t test. Results are presented as mean ± SD. A P value less than .05 was regarded as statistically significant.

Results

Mean follow-up after PRL implantation was 17.17 ± 3.76 months (range 12 to 24 months). A summary of patient data is presented in Table 1.

Table 1
Table 1:
Summary of patients' preoperative data.

Efficacy

The mean UCVA significantly improved from counting fingers preoperatively to 0.62 ± 0.28 (range 0.08 to 1.20) at the last follow-up examination (P<.001) (Figure 1, A). Of the 34 eyes, all eyes experienced 1- to 12-line gain. The mean difference between preoperative and postoperative UCVA was a 6.2-line gain (range 1- to 12-line gain).

Figure 1.
Figure 1.:
(Pallikaris) Changes in mean UCVA (A) and BCVA (B) (decimal scale) during the follow-up period. The error bars indicate 95% confidence intervals for the means.

Safety

The mean BCVA significantly improved from 0.70 ± 0.24 (range 0.10 to 1.00) to 0.85 ± 0.24 (range 0.10 to 1.20) (P<.001) (Figure 1, B). Of the 34 eyes, 1 eye lost 2 lines of preoperative BCVA, 8 maintained pre-PRL BCVA, and the rest (25 eyes) experienced a 1- to 5-line gain (Figure 2). Mean difference between pre-PRL and last follow-up after PRL was a gain of 1.5 ± 1.5 lines (range, loss of 2 to gain of 5 lines).

Figure 2.
Figure 2.:
(Pallikaris) Changes in BCVA (lines in decimal scale) between preoperative and the last postoperative follow-up.

Predictability

Preoperative and last follow-up mean values for spherical equivalent refraction revealed a statistically significant reduction (P<.001) from −14.70 ± 2.65 D (range −20.75 to −10.50 D) to −0.61 ± 0.89 D (range −2.25 to 1.00 D) (P<.001) with a mean reduction value of 14.08 ± 2.72 D (range −10.00 to −19.50 D) at the last follow-up (Figure 3). The mean difference between the intended and achieved correction at the last follow-up examination was −0.55 ± 0.86 D (range −2.25 to 1.00 D). Twenty-seven eyes (79%) and 15 eyes (44%) were within ±1.00 D and ±0.50 D of target refraction, respectively (Figure 4).

Figure 3.
Figure 3.:
(Pallikaris) Changes of mean spherical equivalent refraction after PRL implantation. The error bars indicate 95% confidence intervals for the means.
Figure 4.
Figure 4.:
(Pallikaris) Scattergram between achieved and intended spherical equivalent refractive change after PRL implantation. The diagonal lines show equality and over- and undercorrection by 1 D.

IOP Measurements

A statistically significant increase in preoperative IOP measurements was found after 1-month follow-up (pre-PRL, 15.29 ± 1.84 mm Hg; 1 month, 17.24 ± 5.44 [P = .037]), which returned to preoperative levels at 3 months (6 eyes were corticosteroid responders) (Figure 5).

Figure 5.
Figure 5.:
(Pallikaris) Changes in IOP after PRL implantation. The error bars indicate 95% confidence intervals for the means.

Wavefront Aberrations

Wavefront aberrations were assessed in 15 eyes (44.1%). Total high-order root-mean-square (RMS) was evaluated for the same pupil diameters (5 mm, 3 mm) pre-PRL implantation and 1-year postoperatively. There was an increase in total high-order aberrations (third and fourth) but not a statistically significant one (pre-PRL RMS: 0.18 μm ± 0.08 μm [range 0.09 μm to 0.38 μm]; post-PRL: 0.21 μm ± 0.08 μm [range 0.05 μm to 0.38 μm]) (P = .12) for a pupil diameter of 5 mm. Total high-order aberrations in 3 mm pupil diameter did not change significantly (pre-PRL: RMS 0.035 ± 0.016; post-PRL RMS: 0.045 ± 0.018) (P = .08). The spherical aberration (Z4-0 Zernike coefficient) (Table 2) in 5 mm pupils was significantly decreased 1 year postoperatively (pre-PRL, 0.05 ± 0.04; post-PRL: 0.008 ± 0.05) (P = .012). More specifically, Zernike coefficients pre- and post-PRL implantation are shown in Table 2. Modulation transfer function (MTF) before and after surgery was computed for each eye from the corresponding wave aberration, for 5-mm pupil (Figure 6) and ignoring apodization imposed the Stiles-Crawford effect. Contribution of tilt, defocus, and astigmatism were cancelled.5 There was a small contrast sensitivity loss post-PRL implantation. For example, the MTF for 20 cycles/degree decreased by a factor of 1.3.

Table 2
Table 2:
Zernike coefficients (mean values, OSA notation) of 15 eyes before and after PRL implantation (pupil = 5 mm).
Figure 6.
Figure 6.:
(Pallikaris) Average MTF before and after PRL implantation computed from the wave aberration for 5 mm pupil diameter.

Adverse Effects and Their Management

During surgical iridectomy with the probe of a vitreotome, 3 eyes experienced damage of the anterior capsule of the crystalline lens. These eyes were examined very closely, and it was noticed that the opacification remained focal behind the iridectomy and did not progress to cataract in the visual axis (Figure 7). Another eye presented focal anterior capsule opacification on the first postoperative day probably because of surgical contact with the crystalline lens. One year after the PRL implantation, the opacification had not progressed or caused any BCVA loss (Figure 8).

Figure 7.
Figure 7.:
(Pallikaris) Slitlamp photograph of patient 3 months after PRL implantation shows focal opacity behind iridectomy because of damage to the anterior capsule.
Figure 8.
Figure 8.:
(Pallikaris) Slitlamp photograph shows anterior subcapsular opacification 1 year post-PRL implantation.

Postoperative Complications

Eight eyes experienced IOP higher than 20 mm Hg during the first postoperative month. Six eyes were corticosteroid responders. Intraocular pressure returned to normal levels after discontinuation of steroid drops. The other 2 eyes of the same patient had a resistant increase of IOP with open anterior chamber angle, no pigment dispersion, and patent iridectomies in both eyes. Visual field test was performed 1 month postoperatively and revealed large glaucomatous defects. Intraocular pressure could not be controlled with combination of topical medication (20 mm Hg to 30 mm Hg). Because of the visual field analysis and the lack of other symptoms in relation to the implant, the patient is considered to have had pre-existing, undiagnosed glaucoma. He refused the removal of the implants and underwent successful trabeculectomies in his left and right eye 1 and 2 months after the PRL implantation, respectively. Intraocular pressure was checked closely and was under 16 mm Hg in both eyes in every postoperative examination. Visual fields remained stable 1 year after the trabeculectomies.

Six patients (28.5%) complained of glare and halo at night. These symptoms decreased 6 months after PRL implantation. Five of these patients had pupils greater than 7 mm; the pupil of the other patient was 6 mm. Halo and glare are attributed to the fact that the optic zone of the PRLs used in this study was 5 mm, which is too small in comparison to these patients' scotopic pupil size.

Discussion

The implantation of an IOL in phakic eyes is indicated for the surgical treatment of high ametropia. Laser in situ keratomileusis (LASIK) has been used to treat high levels of refractive errors, but its predictability and stability decrease with the amount of the attempted correction.1 Large ablation depths also predispose the cornea to the risk for ectasia, which makes surgeons more conservative with the amount of laser corrections.2 The implantation of an IOL in a phakic eye is a theoretically reversible and stable technique, whereas clear lens extraction is more invasive and results in the loss of accommodation.6–8

Anterior chamber lenses supported in the anterior chamber angle have the advantage of a comparatively simple surgical technique. The complications that might follow the implantation of an anterior chamber lens are damage to the corneal endothelium, mostly during the first year after implantation; pupil ovalization with iris atrophy; anterior uveitis; and elevation of IOP.9–11 Iris-fixated lenses require a more sophisticated surgical technique.12 Although they may have a good refractive outcome13 and are considered safer for the corneal endothelium14 because they are not fixated in the angle, they also may result in several complications such as localized iris ischemia.15

In 1986, Fyodorov and coauthors16 designed a posterior chamber IOL, which was made of silicone. This lens underwent improvements in its design and passed through 3 generations until the PRL implanted in this study was produced.17

In our study, comparison of UCVA and BCVA before and after PRL implantation demonstrates the efficacy and safety of this posterior chamber lens. Seventy-nine percent of myopic eyes were within ±1.00 D of target refraction, whereas 44% were within ±0.50 D of target refraction (Figure 4). To eliminate the effects of magnification, optical distortion, and vertex power imposed by spectacle lenses, we measured the preoperative contact lens BCVA and compared this with the BCVA after the PRL implantation. Seventy-three percent of myopic eyes gained 1 to 5 lines of BCVA postoperatively. These results are even better compared with those of other posterior chamber lenses.18–22

Few short-term complications were observed, such as IOP increase during the first postoperative day because of residual viscoelastic23 and during the first month because of corticosteroid response. Corticosteroids were used to prevent postsurgery inflammation, but it might be useful to use nonsteroidal anti-inflammatory drops instead to avoid this steroid side effect. Because large, surgical iridectomies were performed, no eye presented angle closure or pupillary block. After proper medication and discontinuation of steroid drops, IOP returned to normal levels, respectively, to the IOP before the operation. This shows that the presence of a PRL inside a myopic eye does not cause a long-term increase of IOP because it is also reported for other posterior chamber lenses.23

The main short- or long-term risk from implantation of a posterior chamber lens is cataract formation because of contact between the phakic lens and the crystalline lens or because of metabolic disturbances in the latter.21,24 In our study, we noticed 1 case of localized anterior capsule opacity on the first postoperative day. This could have been caused by surgical trauma. At all the postoperative examinations, there was no contact between the implant and the crystalline lens. The opacity had not progressed 15 months after PRL implantation and had not caused any visual acuity loss.

Another possible complication following the implantation of a posterior chamber lens is pigment dispersion because of irritation of the posterior surface of the iris by the anterior surface of the implant.25 No pigment dispersion was noticed in any of the eyes in this study.

Glare and halo at night were mentioned by 6 patients (28.5%) and decreased during follow-up. Because the optic zone of PRL is definite, patients with pupils greater than 7 mm in scotopic conditions should be informed of the possibility of glare and halo at night. However, PRL implantation cannot be totally excluded in these patients because not all patients with large pupils experienced these night phenomena, whereas those who did were quite satisfied with their vision after PRL implantation and considered these symptoms insignificant.

Higher-order aberrations of the 15 eyes measured at pupil of 5 mm remained almost unchanged after the operation. The decrease in spherical aberration after PRL implantation could be a benefit for mesopic vision. In myopic eyes, the MTF, which refers to the retinal image quality, was decreased 1 year post-PRL implantation. Moreno-Barriuso and coauthors26 reported that in post-LASIK eyes there was an increase in spherical aberration, which decreased retinal image quality (MTF). Higher-order aberrations of more eyes need to be evaluated before and after PRL implantation before we can draw conclusions about the effect of this lens on the quality of vision.

In conclusion, PRL implantation in highly myopic eyes seems to be a safe, effective, and minimally invasive technique without serious intra- or postoperative complications. However, further follow-up and additional patients must be reviewed to draw final conclusions about the efficacy and safety of this new posterior chamber PRL.

References

1. Pérez-Santonja JJ, Bellot J, Claramonte P, et al. Laser in situ keratomileusis to correct high myopia. J Cataract Refract Surg 1997; 23:372-385
2. Pallikaris IG, Kymionis GD, Astyrakakis NI. Corneal ectasia induced by laser in situ keratomileusis. J Cataract Refract Surg 2001; 27:1796-1802
3. Applegate RA, Howland HC. Refractive surgery, optical aberrations, and visual performance. J Refract Surg 1997; 13:295-299
4. Hoyos JE, Dementiev DD, Cigales M, et al. Phakic refractive lens experience in Spain. J Cataract Refract Surg 2002; 28:1939-1946
5. Thibos LN, Applegate RA, Schwiegerling JT, Webb RH. Standards for reporting the optical aberrations of eyes. OSA Trends in Optics and Photonics. Vision Science and its Applications. Washington, DC, Optical Society of America 2000; 35:232-244
6. Goldberg MF. Clear lens extraction for axial myopia; an appraisal. Ophthalmology 1987; 94:571-582
7. Lyle WA, Jin GJC. Clear lens extraction for the correction of high refractive error. J Cataract Refract Surg 1994; 20:273-276
8. Lee KH, Lee JH. Long-term results of clear lens extraction for severe myopia. J Cataract Refract Surg 1996; 22:1411-1415
9. Alió JL, de la Hoz F, Pérez-Santonja JJ, et al. Phakic anterior chamber lenses for the correction of myopia; a 7-year cumulative analysis of complications in 263 cases. Ophthalmology 1999; 106:458-466
10. Mimouni F, Colin J, Koffi V, Bonnet P. Damage to the corneal endothelium from anterior chamber intraocular lenses in phakic myopic eyes. Refract Corneal Surg 1991; 7:277-281
11. Pérez-Santonja JJ, Iradier MT, Sanz-Iglesias L, et al. Endothelial changes in phakic eyes with anterior chamber intraocular lenses to correct high myopia. J Cataract Refract Surg 1996; 22:1017-1022
12. Menezo JL, Cisneros A, Hueso JR, Harto M. Long-term results of surgical treatment of high myopia with Worst-Fechner intraocular lenses. J Cataract Refract Surg 1995; 21:93-98
13. Budo C, Hessloehl JC, Izak M, et al. Multicenter study of the Artisan phakic intraocular lens. J Cataract Refract Surg 2000; 26:1163-1171
14. Maloney RK, Nguyen LH, John ME. Artisan phakic intraocular lens for myopia; short-term results of a prospective, multicenter study; the Artisan Lens Study Group. Ophthalmology 2002; 109:1631-1641
15. Menezo JL, Cisneros AL, Rotriquez-Salvador V. Endothelial study of iris-claw phakic lens: four year follow-up. J Cataract Refract Surg 1998; 24:1039-1049
16. Fyodorov SN, Zuyev VK, Aznabayev BM. Intraocular correction of high myopia with negative posterior chamber lens [Russian]. Oftalmokhirurgiia 1991; 3:57-58
17. Dementiev DD, Hoffer KJ, Sborgia G, et al. Phakic refractive lenses (PRLs). In: Lovisolo CF, Pesando PM. The Implantable Contact Lens (ICL). Regione S. Giovanni, Ed Fabiano; 1999; 391
18. Assetto V, Benedetti S, Pesando P. Collamer intraocular contact lens to correct high myopia. J Cataract Refract Surg 1996; 22:551-556
19. Zaldivar R, Davidorf JM, Oscherow S. Posterior chamber phakic intraocular lens for myopia of −8 to −19 diopters. J Refract Surg 1998; 14:294-305
20. Rosen E, Gore C. Staar Collamer posterior chamber phakic intraocular lens to correct myopia and hyperopia. J Cataract Refract Surg 1998; 24:596-606
21. Arne JL, Lesueur LC. Phakic posterior chamber lenses for high myopia: functional and anatomical outcomes. J Cataract Refract Surg 2000; 26:369-374
22. Pesando PM, Ghiringhello MP, Tagliavacche P. Posterior chamber collamer phakic intraocular lens for myopia and hyperopia. J Refract Surg 1999; 15:415-423
23. Jimenez-Alfaro I, Bençítez del Castillo JM, Garcçía-Feijoo J, et al. Safety of posterior chamber phakic intraocular lenses for the correction of high myopia-anterior segment changes after posterior chamber phakic intraocular lens implantation. Ophthalmology 2001; 108:90-99; discussion by SM MacRae, 99
24. Fink AM, Gore C, Rosen E. Cataract development after implantation of the Staar collamer posterior chamber phakic lens. J Cataract Refract Surg 1999; 25:1278-1282
25. Brandt JD, Mockovac ME, Chayet A. Pigmentary dispersion syndrome induced by a posterior chamber phakic refractive lens. Am J Ophthalmol 2001; 131:260-263
26. Moreno-Barriuso E, Lloves JM, Marcos S, et al. Ocular aberrations before and after myopic corneal refractive surgery: LASIK-induced changes measured with laser ray tracing. Invest Ophthalmol Vis Sci 2001; 42:1396-1403
© 2004 by Lippincott Williams & Wilkins, Inc.