Case report

Pupillary block due to reverse implantation of a sulcus intraocular lens

Zhang, Xuemin BS; Soni, Nishant MD; Alexander, Janet MD; Kalarn, Sachin MD; Saeedi, Osamah MD*

Author Information
Journal of Cataract and Refractive Surgery Online Case Reports: August 2016 - Volume 4 - Issue 3 - p 41-44
doi: 10.1016/j.jcro.2016.06.002
  • Free

Abstract

Pseudophakic pupillary block can occur after anterior chamber intraocular lens (IOL) implantation, as the IOL may act as a valve to occlude aqueous flow through the pupil.1,2 However, it can also occur after posterior chamber IOL implantation and may be related to several factors including alteration of the anterior chamber angle due to IOL placement in the ciliary sulcus,3 vitreous presentation in the anterior chamber due to disruption of zonular fibers intraoperatively,4 and exacerbation of a preexisting shallow anterior chamber. The onset can occur a few days to several months postoperatively.2 We present a case of pupillary block secondary to reverse implantation of an IOL in the ciliary sulcus, leading to chronic angle-closure glaucoma (ACG) and acute pupillary block.

CASE REPORT

An 88-year-old woman with a history of chronic ACG and complicated cataract extraction in the right eye presented to the glaucoma service with intermittent right-sided headache and severe right ocular pain, which had gradually worsened over 2 days.

Five years before presentation, the patient's cataract surgery had been complicated by a large posterior capsule tear, necessitating anterior vitrectomy and implantation of a 3-piece IOL in the ciliary sulcus. On the first postoperative day, the IOP in the right eye was 54 mm Hg and the IOL haptic had subluxated into the anterior chamber. The IOL was repositioned at the slitlamp with a 30-gauge needle. There had been no evidence of glaucoma prior to the surgery; apraclonidine 0.5%, brinzolamide 1.0%, oral acetazolamide, and prednisolone acetate were started postoperatively. The elevated IOP persisted despite discontinuing prednisolone acetate 1.0% eyedrops. Gonioscopy of the right eye revealed narrow anterior chamber angles. Chronic ACG was diagnosed and treated with chronic latanoprost therapy.

On presentation, the patient had been on latanoprost 0.005% daily in both eyes but had stopped the medication 2 days earlier. The corrected distance visual acuity (CDVA) was 20/25 in the right eye. Goldmann applanation tonometry measured 26 mm Hg in the right eye and 11 mm Hg in the fellow eye. A shallow anterior chamber with 360 degrees of iris-to-cornea touch was noted. Evaluation of the IOL revealed a slightly tilted and anteriorly displaced 3-piece IOL in the sulcus that appeared to be in a reverse-S configuration (Figure 1, A and B). The angle was notably closed on gonioscopy. Humphrey visual field testing showed a dense superior arcuate deficit. Optical coherence tomography (OCT) of the retinal nerve fiber layer revealed diffuse thinning. The fellow eye had a normal Humphrey visual field and OCT retinal nerve fiber layer. Dilated fundus examination revealed an asymmetric cup-to-disc ratio of 0.9 in the right eye and 0.4 in the left eye and was otherwise unremarkable.

Figure 1.
Figure 1.:
Slitlamp biomicroscopy. A: A tilted and anterior 3-piece IOL that appears to be in an S configuration in the sulcus and a narrow anterior chamber preoperatively before dilation. B: The eye preoperatively after dilation. C: A deep anterior chamber and repositioned IOL with tube shunt placement postoperatively before dilation.

Latanoprost 0.005% daily, brimonidine 0.2% and timolol 0.5% twice daily, atropine 1.0% daily, and oral acetazolamide were started. A week later the intraocular pressure (IOP) was elevated to 32 mm Hg; the CDVA had deteriorated to 20/40. Anterior segment ultrasound biomicroscopy (UBM) showed an incorrectly positioned 3-piece sulcus IOL that was vaulted anteriorly, leading to pupillary block (Figure 2, A) with a normal ciliary body position. Because of the advanced glaucoma, high IOP, and extensive peripheral anterior synechiae, incisional surgery was recommended.

Figure 2.
Figure 2.:
Right eye. A: Preoperative anterior segment UBM confirming 3-piece sulcus reversed IOL vaulted anteriorly, displacing the apposed iris forward. B: Postoperative anterior segment UBM showing resolution of the pupillary block, with IOL vaulted posteriorly and Baerveldt tube shunt correctly positioned in the anterior chamber.

An experienced anterior segment surgeon (O.S.) performed uneventful IOL repositioning, goniosynechialysis, and placement of a Baerveldt 350 tube shunt with corneal patch graft (Figure 3). The latter was achieved by placing a 4-0 polypropylene (Prolene) suture in the lumen of the Baerveldt tube shunt, placing the Baerveldt tube shunt and securing it under the superior and medial rectus muscles, and then tying off the tube before proceeding to the IOL repositioning. A clear corneal incision was made, and an ophthalmic viscosurgical device (OVD) was used to deepen the anterior chamber. Iris hooks were used to expand the pupil. The IOL was noted to be in the sulcus and mobile. An OVD cannula was placed under the IOL, and OVD was injected to push the temporal aspect of the IOL anteriorly to flip the IOL on its vertical axis using a Sinskey hook. After the haptics were positioned appropriately in the sulcus, anterior synechialysis was performed. The tube shunt was subsequently placed in the eye and tied off. A corneal patch graft was placed, and conjunctiva was reapproximated.

Figure 3.
Figure 3.:
Intraoperative photographs of IOL repositioning surgery. A: The incorrectly positioned IOL at the beginning of surgery with haptic in the S configuration (arrow). B: Repositioning the IOL by flipping it on its horizontal axis using an OVD. C: The IOL after repositioning, with haptic in the correct configuration (arrow).

Postoperatively, the eye pain diminished significantly and the anterior chamber deepened (Figure 1, C). The IOP ranged from 6 to 33 mm Hg after surgery before the intubating suture was removed (rip cord). After the suture was removed at 2 months, the IOP stabilized and ranged between 9 mm Hg and 13 mm Hg on brimonidine 0.2% and timolol 0.5% twice daily; the CDVA was 20/30. At 4 months, repeat anterior segment UBM revealed a posteriorly vaulted IOL (Figure 2, B). ImageJ software was used to measure the anterior chamber depth (ACD), angle opening distance, and the trabecular–iris angle in the preoperative and postoperative anterior segment UBM images. The ACD increased from 1.74 mm to 2.83 mm. The angle opening distance at 500 μm increased from 0.36 mm to 0.65 mm nasally and from 0.24 mm to 0.91 mm temporally. The trabecular–iris angle increased from 19 to 36 degrees nasally and from 29 to 46 degrees temporally.

DISCUSSION

Pupillary block is an infrequent complication of posterior chamber IOL implantation. It can be caused by 360 degrees of posterior synechiae between the IOL and iris or, if the IOL is placed in the ciliary sulcus, by alteration of the anatomy of the anterior chamber angle.4 In the case of reversed IOL orientation in the sulcus, the anterior vault of a 3-piece IOL can cause the optic to be apposed to the posterior surface of the iris, leading to pupillary block. Although accidental reversal of the IOL is not a commonly reported cause of pseudophakic pupillary block, incorrect positioning of the IOL in the sulcus has been noted to occur in 1.3% of cases, even in experienced hands.5 All these situations can result in anterior chamber shallowing, IOP elevation, and subsequent ACG.

Our patient is unique in that she presented with acute chronic pupillary block glaucoma 5.5 years after the initial sulcus placement of the IOL. Similar to previous reports of pseudophakic pupillary block after sulcus IOL reversal, anterior displacement of the IOL optic was noted in the early postoperative period and was relieved by repositioning at the slitlamp. Although this technique has been reported,6 our case shows that in the long term, this approach may not reduce the risk for chronic glaucomatous damage. Laser peripheral iridotomy has been attempted in this setting, but based on the anterior segment UBM in our case, we believed the vaulted IOL would continue to push the iris diaphragm anteriorly and cause further peripheral anterior synechiae. Therefore, we recommend surgical repositioning of the IOL in the ciliary sulcus as soon as possible after reverse IOL implantation is recognized.

Anterior segment UBM is useful in the evaluation of pupillary block, plateau iris, IOL positioning, ocular trauma, and other ocular diseases involving the anterior segment.7 The technique has been used to determine IOL positioning and to elucidate the cause of reverse pupillary block.8–10 In a case study by Sathish et al.,10 anterior segment UBM helped diagnose iris bombe as the cause of pseudophakic pupillary block glaucoma 3 years after implantation of a 3-piece IOL in the bag. In our case, although the configuration of the haptic provided a clue to the etiology of pupillary block, anterior segment UBM confirmed that the IOL was causing anterior bowing of the iris and excluded other causes such as aqueous misdirection syndrome,11 bilateral ring cysts of the ciliary body,8 late-onset secondary pigmentary glaucoma,9 a thick Soemmerring ring, and anterior displacement of the ciliary body.12

Aydin et al.13 describe a case in which reverse implantation of a vaulted IOL in the ciliary sulcus led to pupillary block, the sequelae of which were resolved with neodymium:YAG laser peripheral iridotomy. Harsum and Low6 report the case of a patient who developed pupillary block after dilation was performed 4 weeks after reverse IOL implantation. It was resolved by repositioning the IOL at the slitlamp. Prior case reports have not shown the long-term follow-up of patients who had such temporizing measures.

Our patient required low IOP to manage advanced glaucoma. Surgery included placement of a Baerveldt 350 tube shunt with corneal patch graft in addition to goniosynechialysis and IOL repositioning. This treatment resolved the pupillary block and successfully controlled the patient's IOP. All anterior segment UBM measurements, especially ACD, angle opening distance at 500 μm, and trabecular–iris angle, revealed significant improvement postoperatively. In normal subjects, mean anterior segment UBM measurements for ACD is 3.13 mm ± .37 (SD).14 Although our patient's ACD was 1.74 mm during the episode of pupillary block, it increased to 2.83 mm following the procedure. The opening of the trabecular–iris angle increased 17 degrees both nasally and temporally. Thus, we conclude that the reverse implantation of the IOL led to both chronic ACG and acute pupillary block.

The aim of our case report is to remind surgeons to assess the position of the IOL in a case of pseudophakic pupillary block or when the IOP is persistently high postoperatively. Anterior segment UBM is a reliable method to assess IOL positioning in such cases. Temporizing measures may resolve acute pupillary block in such cases, but chronic glaucomatous changes can still occur. Early repositioning of the IOL could prevent progression of ACG and pupillary block.

REFERENCES

1. Downing JE. Ten-year follow up comparing anterior and posterior chamber intraocular lens implants. Ophthalmic Surg 1992; 23:308-315.
2. Samples JR, Bellows AR, Rosenquist RC, Hutchinson BT, Fine IH, Van Buskirk EM. Pupillary block with posterior chamber intraocular lenses. Arch Ophthalmol 1987; 105:335-337.
3. Gaton DD, Mimouni K, Lusky M, Ehrlich R, Weinberger D. Pupillary block following posterior chamber lens implantation in adults. Br J Ophthalmol 2003; 87:1109-1111. Available at: http://www.ncbi.nlm.nih.gov/pmc/articles/PMC1771845/pdf/bjo08701109.pdf. Accessed March 21, 2016.
4. Willis DA, Stewart RH, Kimbrough RL. Pupillary block associated with posterior chamber lenses. Ophthalmic Surg 1985; 16:108-109.
5. Halpern BL, Gallagher SP. Refractive error consequences of reversed-optic AMO SI-40NB intraocular lens. Ophthalmology 1999; 106:901-903.
6. Harsum S, Low S. Reversed vaulted AcrySof intraocular lens presenting as pupillary block. [letter]. Eye 2009; 23:1880-1882. Available at: http://www.nature.com/eye/journal/v23/n9/pdf/eye2008343a.pdf. Accessed March 21, 2016.
7. Ishikawa H, Schuman JS. Anterior segment imaging: ultrasound biomicroscopy. Ophthalmol Clin North Am 2004; 17:7-20. Available at: http://www.ncbi.nlm.nih.gov/pmc/articles/PMC1978090/pdf/nihms24574.pdf. Accessed March 21, 2016.
8. Seki M, Fukuchi T, Yoshino T, Ueda J, Hasebe H, Ueki S, Oyama T, Fukushima A, Abe H. Secondary glaucoma associated with bilateral complete ring cysts of the ciliary body. J Glaucoma 2014; 23:477-481.
9. Chang SHL, Wu W-C, Wu S-C. Late-onset secondary pigmentary glaucoma following foldable intraocular lenses implantation in the ciliary sulcus: a long-term follow-up study. BMC Ophthalmol 2013; 13:22. Available at: http://www.biomedcentral.com/content/pdf/1471-2415-13-22.pdf. Accessed March 21, 2016.
10. Sathish S, MacKinnon JR, Atta HR. Role of ultrasound biomicroscopy in managing pseudophakic pupillary block glaucoma. J Cataract Refract Surg 2000; 26:1836-1838.
11. Moinul P, Hutnik CML. Aqueous misdirection syndrome: an interesting case presentation. Clin Ophthalmol 2015; 9:183-186. Available at: http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4322885/pdf/opth-9-183.pdf. Accessed March 21, 2016.
12. Esmenjaud E, Rebollo O. Glaucome chronique par fermeture de l'angle chez le pseudophake avec anneau de Soemmering et antéroposition du corps ciliaire: à propos d'un cas. [Chronic angle-closure glaucoma in a pseudophakic eye with Soemmering's ring and plateau iris: case report]. J Fr Ophtalmol 2013; 36:455-460.
13. Aydin B, Erdurmuş M, Yağcı R, Karadağ R, Hepsen İ. Pupillary block due to reverse implantation of angulated posterior chamber intraocular lens. Glo-Kat 2008; 3:208-210. Available at: http://www.glaucomacataractjournal.com/pdf/PDF_166.pdf. Accessed March 21, 2016.
14. Pavlin CJ, Harasiewicz K, Foster FS. Ultrasound biomicroscopy of anterior segment structures in normal and glaucomatous eyes. Am J Ophthalmol 1992; 113:381-389.
© 2016 by Lippincott Williams & Wilkins, Inc.