Early Recognition of Raised Intracranial Pressure in Craniosynostosis Using Optical Coherence Tomography

Rufai, Sohaib R. BMBS, MRes∗,†; Jeelani, Noor ul Owase MBA, FRCS‡,§; McLean, Rebecca J. MSc, PhD

Author Information
doi: 10.1097/SCS.0000000000006771

Abstract

Recognition of raised intracranial pressure (ICP) is crucial in children with craniosynostosis to enable timely surgical intervention. Raised intracranial pressure can cause visual impairment and death if untreated.1 Prevalence of craniosynostosis is 3.1 to 6.4 in 10,000 live births and rising.2 Nonsyndromic cases account for 65% of craniosynostosis and can be uni or multisutural - raised ICP affects 13% and 50% to 75% of such cases, respectively.3,4 Twenty-one percent of cases are syndromic, which are associated with raised ICP in 30% to 40%5,6 of syndromic cases; Apert7 (71%), Crouzon8 (65%) and Pfeiffer8 (60%) syndromes are more commonly associated.

Estimating ICP in children is difficult. Direct intraparenchymal measurement represents the gold standard, but involves hospital admission and carries risks of infection, bleeding, and device failure.6 Noninvasive methods such as ophthalmoscopy,9 transorbital ultrasound,10 and radiography11 deliver inadequate sensitivity to be used for screening. An ideal solution would be a safe, noninvasive, and quantitative measure of ICP that can be repeated over time with good sensitivity and specificity.

Raised ICP can cause optic nerve swelling and secondary retinal changes. Detecting these changes early could help prevent visual damage. Optical coherence tomography (OCT) is a noninvasive imaging method that objectively generates ultra-high resolution 3-dimensional scans of the optic nerve and retina, in vivo, within seconds and with excellent repeatability.12 The University of Leicester Ulverscroft Eye Unit has extensive experience in using handheld OCT – an adapted device suitable for children. Our unit has described the normal development of the optic nerve13 and retina14 in children. Our unit has also investigated a wide range of conditions using handheld OCT including nystagmus,15 foveal hypoplasia,16 optic nerve hypoplasia,17 glaucoma,18 microcephaly,19 retinopathy of prematurity,20 retinoblastoma,21 and others.

Like craniosynostosis, cerebral malaria can also cause raised ICP. Our unit recently conducted a study in Queen Elizabeth Central Hospital, Blantyre, Malawi, in children with cerebral malaria (n = 35; age range: 7–131 months; mean age 56.1 months), demonstrating that handheld OCT can detect changes associated with raised ICP.22 Cerebrospinal fluid opening pressure was positively correlated with optic nerve parameters, including optic nerve head rim volume (r = 0.582; P = 0.0002), rim area (r = 0.504; P = 0.002) and superior peripapillary retinal nerve fiber layer (RNFL) thickness (r = 0.479; P = 0.004). Cerebrospinal fluid opening pressure was negatively correlated with cup parameters, including cup/disc ratio (r = −0.498; P = 0.002) and cup depth (r = −0.432; P = 0.01).

The role of OCT in craniosynostosis has not yet been rigorously assessed by systematic review, nor has it been studied yet by our unit. The aim of this systematic review is to assess the available evidence supporting the role of OCT in managing craniosynostosis. Specifically, the sensitivity and specificity of OCT in the detection of raised ICP in craniosynostosis will be assessed.

OBJECTIVES

The primary objective of this systematic review is to assess the sensitivity and specificity of OCT in the detection of raised ICP in children with craniosynostosis.

The secondary objectives are to explore the OCT parameters used per study and ICP ranges determined as normal and raised.

METHODS

Search Strategy

We entered medical subject headings terms for "raised intracranial pressure" and "optical coherence tomography" into the following search platforms: Cochrane Central Register of Controlled Trials (CENTRAL) (including the Cochrane Eyes and Vision Group (CEVG) Trials Register), Ovid MEDLINE (1946 to present), Ovid MEDLINE Inprocess and Other Nonindexed Citations, EMBASE Classic + Embase (1947 to present) and PubMed (1948 to present). Appendix 1 includes full details of keywords and medical subject headings terms used, http://links.lww.com/SCS/B771. We uploaded search results into EndNote X10 (Thomson Reuters, New York, NY) reference management software. No date or language restrictions were stipulated. We included Level IV evidence and above, that is, case series, case-control studies, cohort studies, randomized controlled trials and systematic reviews, as defined by the Oxford Centre for Evidence-based Medicine.23

Optical coherence tomography studies of children, defined as being aged 0 to 16 years, diagnosed with forms of craniosynostosis and raised ICP were included. Exclusion criteria are as follows:

  • 1. studies with participants over 16 years old;
  • 2. studies not pertaining to children with craniosynostosis;
  • 3. studies not pertaining to diagnosis of raised ICP.

A 2-stage screening process was employed by 2 screeners (SRR and RJM). First, titles and abstracts were screened, followed by full papers. Screening questions are included in Appendix 2, http://links.lww.com/SCS/B771. The references of included papers were reviewed and investigators contacted to identify published and unpublished works. Unpublished works identified and recommended by experts were included if applicable. The systematic review protocol was registered with PROSPERO (CRD42019154254).

Outcome Measures

The primary outcome measure was sensitivity and specificity for OCT parameters in the detection of raised ICP. The OCT parameters could include any combination of the following:

  • Optic nerve parameters: cup depth, cup width, disc width, and cup to disc ratio;
  • Rim parameters: RNFL thickness, rim volume, Bruch's membrane opening-minimum rim width, Bruch's membrane orientation, full peripapillary analysis;
  • Retinal parameters: macular and perimacular retinal thickness, foveal pit width, depth and area, plus segmentation of all retinal layers.

The secondary outcome measures were as follows:

  • Other OCT parameters not listed above;
  • ICP range determined as normal;
  • ICP range determined as abnormal, that is, low or raised;
  • Other surrogate estimates of ICP.
  • Quality of life and adverse events were recorded and reported where applicable.

Data Collection and Analysis

Two-stage screening was performed by 2 screeners (SRR and RJM) whereby titles and abstracts were screened first, followed by full papers. Screening questions are included in Appendix 2, http://links.lww.com/SCS/B771. Data was extracted using a data extraction tool included in Appendix 3, http://links.lww.com/SCS/B771, adapted from the Cochrane Collaboration.24 Quality assessment was performed using the National Institutes of Health (NIH) Quality Assessment Tool for Observational Cohort and Cross-Sectional Studies,25 included in Appendix 4, http://links.lww.com/SCS/B771.

RESULTS

Figure 1 is a PRISMA flow diagram summarizing the screening process. Of 318 records identified, three papers were included for review. Supplementary Digital Content, Table 1, http://links.lww.com/SCS/B587 displays the main characteristics of the included studies. Appendix 5, http://links.lww.com/SCS/B771 is a table of excluded studies with reasons.

FIGURE 1
FIGURE 1:
PRISMA study inclusion flow diagram.

Quality Assessment

As per the NIH Quality Assessment Tool for Observational Cohort and Cross-Sectional Studies,25 the internal validity of the study by Swanson et al26 was deemed as fair, whereas that of the studies by Dagi et al27 and Driessen et al28 was deemed as poor. Hence, the risk of bias for the study by Swanson et al26 was deemed as moderate, whereas that of the studies by Dagi et al27 and Driessen et al28 was deemed as high. This is largely due to the fact that Swanson et al26 compared OCT parameters to direct ICP measurements, whereas other studies compared OCT parameters to fundus examination – a subjective estimate of ICP.

Primary Outcome Measures

Optical coherence tomography was effective in detecting raised ICP or evidence thereof in all included studies. However, only Swanson et al26 compared OCT parameters to direct ICP measurements, wherein they considered ICP greater than 15 mmHg as raised. On the other hand, Dagi et al27 and Driessen et al28 compared OCT parameters to papilledema detected on fundus examination. Spectral-domain OCT devices used in these studies included Spectralis27,28 (axial resolution 3.9 μm; Heidelberg Engineering, Heidelberg, Germany) and iVue (axial resolution 5 μm; Haag-Streit, Wedel, Germany).26 No study used handheld OCT. A qualitative review was performed without meta-analysis due to the absence of randomized controlled trials required for the latter. Supplementary Digital Content, Table 2, http://links.lww.com/SCS/B587 summarizes the OCT outcome measures of all included studies.

Secondary Outcome Measures

All included studies explored the association of secondary outcome measures with the gold standard, as displayed in Supplementary Digital Content, Table 3, http://links.lww.com/SCS/B587. Some of the alternative methods had either good sensitivity or good specificity, but none of the methods demonstrated as strong a combination of sensitivity and specificity as the OCT outcomes in Supplementary Digital Content, Table 2, http://links.lww.com/SCS/B587.

Quality of Life Outcomes and Adverse Events

No included study reported quality of life outcomes or adverse events.

DISCUSSION

This systematic review has highlighted a lack of quality evidence for the role of OCT in detecting raised ICP in craniosynostosis. Only 3 studies were identified and included, 2 of which were deemed as poor quality according to the NIH Quality Assessment Tool for Observational Cohort and Cross-sectional Studies.25 This could be because conventional OCT, as used in these studies, is very challenging in conscious children and infants.

Optical Coherence Tomography as a Screening Tool

All included studies reported better association between OCT and ICP (or estimates by fundus examination) as compared to other conventional clinical methods, including clinical history, visual acuity, visual field testing, radiological examination, and fundus examination (where ICP measurement was used as the gold standard). Spectral-domain OCT devices used in these studies included Spectralis27,28 (axial resolution 3.9 μm; Heidelberg Engineering, Heidelberg, Germany) and iVue (axial resolution 5 μm; Haag-Streit, Wedel, Germany).26 The former was used in older, conscious patients27,28 whilst the latter was used in younger children under general anaesthesia.26

With regards to feasibility, most children were cooperative with OCT examination and successfully imaged on first attempt. Driessen et al28 reported that OCT imaging was unsuccessful in 6 patients (13.6%), whilst Dagi et al27 reported that 11 children (16.9%) were excluded due to limited cooperation, severe nystagmus, poor scan quality or retinal degeneration – the full breakdown for reasons for exclusion was not provided. Swanson et al26 successfully imaged all included patients, albeit under general anesthesia and hence were nonreliant on patient cooperation. In reality, if OCT is initially unsuccessful in a child, it is common practice to offer further appointments for reattempt. Furthermore, our unit has extensive experience with the more recently developed handheld OCT, where we have demonstrated excellent feasibility in awake infants.12,13,15,16 Handheld OCT may, therefore, represent a good solution in this setting, but a high quality prospective study is required to clarify this.

Swanson et al26 and Dagi et al27 reported better sensitivity and specificity for OCT as compared to other clinical methods, but Driessen et al28 did not perform sensitivity and specificity analyses. Rather, Driessen et al28 demonstrated that funduscopic anomaly showed a significantly increased total retinal thickness as compared to patients with normal fundus examination (Mann Whitney U test, P < 0.001). Crucially, all included studies highlighted the objective and quantitative nature of OCT for evaluating optic nerve and/or retinal parameters, in contrast to the subjective nature of clinical history, fundus examination and visual field testing.

There was no standardization of OCT parameters used across the included studies. Dagi et al27 used RNFL thickness as their sole OCT parameter, whereas Driessen et al28 used total retinal thickness as their sole parameter. Swanson et al26 used a combination of maximal RNFL thickness, maximal retinal thickness, and maximal anterior retinal projection. Further studies are required to establish which parameters should be used in the assessment of raised ICP in craniosynostosis.

Other Clinical Screening Methods

Clinical history, visual acuity, visual field testing, radiological examination, and fundus examination were all poor screening methods for raised ICP in the included studies. Although specificity for raised ICP was good in all but historical evidence in the study by Dagi et al,27 sensitivity was too low to qualify any of these methods as screening tools. This is consistent with previous literature.29 Interestingly, Dagi et al27 reported that historical evidence of raised ICP had good sensitivity but poor specificity.

Such investigations may represent useful adjuncts to overall clinical assessment. For instance, Swanson et al26 found that papilledema found on fundus examination was 100% specific (95% confidence interval 85%–100%) for raised ICP and hence should prompt urgent referral for neurosurgical intervention on these grounds alone. Abnormal signs on radiological examination also demonstrated good specificity and represent an objective investigation.26 Visual acuity and visual field testing are subjective and difficult for young children with craniosynostosis, particularly those with associated cognitive impairment.

Transorbital ultrasound has also been evaluated as a potential screening method.30,31 Raised ICP can increase optic nerve sheath diameter, hence elevation above age-controlled diameters may predict raised ICP. However, Driessen et al10 demonstrated that transorbital ultrasound delivers sensitivity of only 11% for detecting raised ICP in children with craniosynostosis, excluding this method as a potential screening tool. Visual evoked potentials testing involves measuring the amplitude and latency time of the averaged encephalographic response to a visual stimulus; reduction of amplitude or prolongation of the latency period represent axonal injury and correlate with elevated ICP.4,29 However, visual evoked potentials testing demands good cooperation and displays high variability in normal subjects.32

Intracranial Pressure Measurement

Swanson et al26 employed an intraoperative single ICP measurement protocol consistent with other published methods, but lacked the 24- to 48-hour duration or surveillance for multiple B-type waves reported in other protocols, which may represent the gold standard.1,33,34 They attempted to mitigate confounding perioperative factors. However, children with craniosynostosis and associated upper airway obstruction may have increased ICP during sleep,35 which will not be detected during general anesthesia as the child is intubated. Hence, no included study has used the current gold standard measure for ICP.

Strengths and Limitations

To our knowledge, this study represents the first systematic review of the role of OCT in detecting raised ICP in craniosynostosis. A rigorous search strategy was employed, screening was executed by 2 screeners (SRR and RJM) as per Cochrane Collaboration guidance and the data extraction sheet was adapted from the Cochrane Collaboration, with no language restrictions.24 Both screeners returned the same studies for inclusion, obviating the need for arbitration.

This study has a number of limitations. No randomized controlled trials were returned by the systematic search, rather only Level IV evidence was identified, as defined by the Oxford Centre for Evidence-based Medicine.23 Hence, it was not possible to perform conventional meta-analysis with a consistent comparator across the studies. Out of the 3 included studies, 2 used fundus examination as the gold standard, which is demonstrably an unreliable indicator of raised ICP due to poor sensitivity, hence both were deemed to be at a high risk of bias. Publication bias assessment was not possible in the available studies.

CONCLUSIONS

This systematic review has revealed a lack of quality evidence for the role of OCT in detecting raised ICP in craniosynostosis. High quality evidence is required to validate and/or recommend OCT as an effective screening tool in craniosynostosis, as has been demonstrated in other studies of pediatric conditions affecting the eye. Further research is required to ascertain which OCT parameters are most appropriate for identifying raised ICP.

Acknowledgments

The authors thank Professor Irene Gottlob and Dr Frank Proudlock for their support and contributions towards this systematic review.

REFERENCES

1. Renier D, Sainte-Rose C, Marchac D, et al. Intracranial pressure in craniostenosis. J Neurosurg 1982; 57:370–377.
2. Cornelissen M, Ottelander BD, Rizopoulos D, et al. Increase of prevalence of craniosynostosis. J Craniomaxillofac Surg 2016; 44:1273–1279.
3. Wilkie AOM, Johnson D, Wall SA. Clinical genetics of craniosynostosis. Curr Opin Pediatr 2017; 29:622–628.
4. Thompson DA, Liasis A, Hardy S, et al. Prevalence of abnormal pattern reversal visual evoked potentials in craniosynostosis. Plast Reconstr Surg 2006; 118:184–192.
5. Bannink N, Nout E, Wolvius EB, et al. Obstructive sleep apnea in children with syndromic craniosynostosis: long-term respiratory outcome of midface advancement. Int J Oral Maxillofac Surg 2010; 39:115–121.
6. Tamburrini G, Caldarelli M, Massimi L, et al. Intracranial pressure monitoring in children with single suture and complex craniosynostosis: a review. Childs Nerv Syst 2005; 21:913–921.
7. Taylor WJ, Hayward RD, Lasjaunias P, et al. Enigma of raised intracranial pressure in patients with complex craniosynostosis: the role of abnormal intracranial venous drainage. J Neurosurg 2001; 94:377–385.
8. Thompson DN, Harkness W, Jones B, et al. Subdural intracranial pressure monitoring in craniosynostosis: its role in surgical management. Childs Nerv Syst 1995; 11:269–275.
9. Tuite GF, Chong WK, Evanson J, et al. The effectiveness of papilledema as an indicator of raised intracranial pressure in children with craniosynostosis. Neurosurgery 1996; 38:272–278.
10. Driessen C, Bannink N, Lequin M, et al. Are ultrasonography measurements of optic nerve sheath diameter an alternative to funduscopy in children with syndromic craniosynostosis? J Neurosurg Pediatr 2011; 8:329–334.
11. Tuite GF, Evanson J, Chong WK, et al. The beaten copper cranium: a correlation between intracranial pressure, cranial radiographs, and computed tomographic scans in children with craniosynostosis. Neurosurgery 1996; 39:691–699.
12. Lee H, Proudlock FA, Gottlob I. Pediatric optical coherence tomography in clinical practice-recent progress. Invest Ophthalmol Vis Sci 2016; 57:OCT69–OCT79.
13. Patel A, Purohit R, Lee H, et al. Optic nerve head development in healthy infants and children using handheld spectral-domain optical coherence tomography. Ophthalmology 2016; 123:2147–2157.
14. Lee H, Purohit R, Patel A, et al. In vivo foveal development using optical coherence tomography. Invest Ophthalmol Vis Sci 2015; 56:4537–4545.
15. Lee H, Proudlock F, Gottlob I. Is handheld optical coherence tomography reliable in infants and young children with and without nystagmus? Invest Ophthalmol Vis Sci 2013; 54:8152–8159.
16. Rufai SR, Thomas MG, Purohit R, et al. Can structural grading of foveal hypoplasia predict future vision in infantile nystagmus? A longitudinal study. Ophthalmology 2020; 127:492–500.
17. Pilat A, Sibley D, McLean RJ, et al. High-resolution imaging of the optic nerve and retina in optic nerve hypoplasia. Ophthalmology 2015; 122:1330–1339.
18. Pilat AV, Proudlock FA, Shah S, et al. Assessment of the anterior segment of patients with primary congenital glaucoma using handheld optical coherence tomography. Eye (Lond) 2019; 33:1232–1239.
19. Papageorgiou E, Pilat A, Proudlock F, et al. Retinal and optic nerve changes in microcephaly: an optical coherence tomography study. Neurology 2018; 91:e571–e585.
20. Anwar S, Nath M, Patel A, et al. Potential utility of foveal morphology in preterm infants measured using hand-held optical coherence tomography in retinopathy of prematurity screening. Retina 2019; [published online ahead of print July 24, 2019]. doi:10.1097/IAE.0000000000002622.
21. Soliman SE, Vanden Hoven C, MacKeen LD, et al. Optical coherence tomography-guided decisions in retinoblastoma management. Ophthalmology 2017; 124:859–872.
22. Tu Z, Sheth V, Gormley J, et al. Hand-held optical coherence tomography – a new biomarker of brain swelling in cerebral malaria. Invest Ophthalmol Vis Sci 2018; 59:612.
23. Oxford Centre for Evidence-based Medicine. Levels of Evidence. 2009. Available at: https://www.cebm.net/2009/06/oxford-centre-evidence-based-medicine-levels-evidence-march-2009/ Accessed May 6, 2020
24. Cochrane. Data extraction forms. 2019. Available at: https://dplp.cochrane.org/data-extraction-forms Accessed May 6, 2020
25. National Institutes of Health. Study Quality Assessment Tools. 2019. Available at: https://www.nhlbi.nih.gov/health-topics/study-quality-assessment-tools Accessed May 6, 2020
26. Swanson JW, Aleman TS, Xu W, et al. Evaluation of optical coherence tomography to detect elevated intracranial pressure in children. JAMA Ophthalmol 2017; 135:320–328.
27. Dagi LR, Tiedemann LM, Heidary G, et al. Using spectral-domain optical coherence tomography to detect optic neuropathy in patients with craniosynostosis. J AAPOS 2014; 18:543–549.
28. Driessen C, Eveleens J, Bleyen I, et al. Optical coherence tomography: a quantitative tool to screen for papilledema in craniosynostosis. Childs Nerv Syst 2014; 30:1067–1073.
29. Derderian C, Seaward J. Syndromic craniosynostosis. Semin Plast Surg 2012; 26:64–75.
30. Wiegand C, Richards P. Measurement of intracranial pressure in children: a critical review of current methods. Dev Med Child Neurol 2007; 49:935–941.
31. Newman WD, Hollman AS, Dutton GN, et al. Measurement of optic nerve sheath diameter by ultrasound: a means of detecting acute raised intracranial pressure in hydrocephalus. Br J Ophthalmol 2002; 86:1109–1113.
32. Andersson L, Sjölund J, Nilsson J. Flash visual evoked potentials are unreliable as markers of ICP due to high variability in normal subjects. Acta Neurochir (Wien) 2012; 154:121–127.
33. Wall SA, Thomas GP, Johnson D, et al. The preoperative incidence of raised intracranial pressure in nonsyndromic sagittal craniosynostosis is underestimated in the literature. J Neurosurg Pediatr 2014; 14:674–681.
34. Gault DT, Renier D, Marchac D, et al. Intracranial pressure and intracranial volume in children with craniosynostosis. Plast Reconstr Surg 1992; 90:377–381.
35. Gonsalez S, Hayward R, Jones B, et al. Upper airway obstruction and raised intracranial pressure in children with craniosynostosis. Eur Respir J 1997; 10:367–375.
Keywords:

Craniosynostosis; intracranial hypertension; intracranial pressure; OCT; optical coherence tomography; systematic review