Performance and short-term outcomes of three different transcatheter aortic valve replacement devices in patients with aortic stenosis

A single-center experience

Tzeng, Yun-Hsuana,b; Lee, Yung-Tsaic,d; Tsao, Tien-Pingb,c; Lee, Kuo-Chenb,c; Huang, Kuan-Chihc,e; Hsiung, Ming-Chonc; Wei, Jengb,c; Yin, Wei-Hsianc,f,*

Journal of the Chinese Medical Association: November 2019 - Volume 82 - Issue 11 - p 827–834
doi: 10.1097/JCMA.0000000000000187
Original Articles
Open

Background: Data on whether different transcatheter aortic valve replacement (TAVR) devices and delivery approaches can achieve equally favorable outcomes when performed by a single heart team are scarce. We sought to compare the performance and short-term outcomes of three different TAVR devices—self-expanding Medtronic CoreValve (MCV), mechanically expanded Lotus valve, and balloon-expandable Edwards SAPIEN XT (SXT)—for the treatment of severe aortic stenosis (AS) in a single large-volume center in Taiwan.

Methods: We retrospectively reviewed consecutive patients who underwent TAVR for the treatment of severe AS. Clinical outcomes were reported following Valve Academic Research Consortium 2 (VARC-2) criteria. The composite primary endpoint was combined all-cause mortality, myocardial infarction (MI), or disabling stroke within 180 days.

Results: A total of 231 patients (MCV n=112, Lotus n=18, and SXT n=101) were included. The device and procedural success rates were similar among all three TAVR devices. At 30 days, there was no significant difference in all-cause mortality, cardiovascular mortality, periprocedural MI, stroke, major vascular complications, life-threatening bleeding, acute kidney injury (AKI, stage 2/3), or VARC-2 composite early safety endpoints. There was no difference among groups in the rate of primary endpoint within 180 days. Lack of procedural success, presence of acute coronary occlusion during TAVR, and presence of AKI (stage 3) after TAVR were independent predictors of adverse outcomes.

Conclusion: TAVR using MCV, Lotus, or SXT was associated with similar 30- and 180-day clinical outcomes. The presence of periprocedural complications was one of the main determinants of short-term adverse outcomes.

aDivision of Medical Imaging, Health Management Center, Cheng Hsin General Hospital, Taipei, Taiwan, ROC

bFaculty of Medicine, National Defense Medical Center, Taipei, Taiwan, ROC

cHeart Center, Cheng Hsin General Hospital, Taipei, Taiwan, ROC

dInstitute of Microbiology and Immunology, School of Medicine, National Yang-Ming University, Taipei, Taiwan, ROC

eGraduate Institute of Clinical Medicine, National Taiwan University College of Medicine, Taipei, Taiwan, ROC

fFaculty of Medicine, School of Medicine, National Yang-Ming University, Taipei, Taiwan, ROC

Received May 19, 2019; accepted August 1, 2019.

Conflicts of interest: Prof. Wei-Hsian Yin and Dr. Yung-Tsai Lee are proctors of transcatheter aortic valve replacement (TAVR) devices of the Medtronic, Edwards, and Boston Scientific companies. Dr. Kuo-Chen Lee, Dr. Tien-Ping Tsao, and Prof. Jeng Wei are proctors of Edwards TAVR devices. The other authors declare that they have no conflicts of interest related to the subject matter or materials discussed in this article.

Address correspondence: Dr. Wei-Hsian Yin, Heart Center, Cheng Hsin General Hospital, 45, Zhenxing Street, Taipei 112, Taiwan, ROC. E-mail address: ch6548@chgh.org.tw (W.-H. Yin)

This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/4.0/).

Back to Top | Article Outline

1. INTRODUCTION

Transcatheter aortic valve replacement (TAVR) is a rapidly evolving technique providing robust treatment of severe aortic stenosis (AS) in high-risk patients.1–5 The main valve technologies available since 2010 have been the self-expanding Medtronic CoreValve (MCV) prosthesis (Medtronic Inc., Minneapolis, MN, USA), the mechanically expanded Lotus (Boston Scientific, Natick, MA, USA), and the balloon-expandable SAPIEN XT (SXT; Edwards Lifesciences, Irvine, CA, USA), which were approved by the Taiwan Food and Drug Administration for clinical use in 2012, 2015, and 2016, respectively. Whether these commonly used earlier generation TAVR prostheses are associated with different key limitations that can potentially impact patient outcomes still needs to be resolved.6,7 These limitations appear to be partially patient specific, but might also be specific to each TAVR system.

As a single approach may not fit all patients, the “tailored TAVR approach” has been proposed.8 Since centers with access to only one type of TAVR device may only be able to offer treatment to a limited portion of the eligible patient population, our TAVR team has striven to be skilled in multiple TAVR devices and delivery approaches so treatment can be offered to the vast majority of patients who are eligible for this procedure. However, data on whether multiple TAVR devices and delivery approaches performed by a single heart team can achieve equally favorable outcomes are scarce.9

Therefore, we evaluated the performance and outcomes of three different TAVR devices (MCV, Lotus, and SXT) for the treatment of severe AS at our center since 2013 to determine whether adverse events after TAVR were patient, device, and/or procedure specific.

Back to Top | Article Outline

2. METHODS

2.1. Study population and TAVR procedures

From March 2013 when the TAVR program was first introduced at Cheng-Hsin General Hospital to August 2017, 231 consecutive patients underwent TAVR at the institution. All patients with severe AS at high risk for conventional cardiac surgery were referred to the TAVR multidisciplinary team composed of interventional cardiologists, imaging cardiologists, cardiothoracic surgeons, radiologists, and anesthesiologists. The MCV device was used in 112 (48.5%) patients, the Lotus valve in 18 (7.8%) patients, and the SXT valve in 101 (43.7%) patients.

The MCV, Lotus, and SXT devices were introduced to our institution in 2013, 2015, and 2016, respectively. For all patients, the choice of access was based on pre-procedural imaging diagnostics, including computed tomography (CT) scan, angiography, and transesophageal and transthoracic echocardiography. In our institution, the default strategy for all patients is the transfemoral (TF) approach. If the first-choice TF access is not feasible due to diseased peripheral vessels, a trans-subclavian, transaortic, or transapical implantation would be considered.

All patients in the study population underwent implantation performed in a hybrid theater, and nearly all patients were treated under general anesthesia. TF TAVR was performed with the use of percutaneous closure devices or surgical cut down of the femoral artery in cases of vessel calcifications or severe obesity. In the trans-subclavian approach, the subclavian artery was dissected free for access through a 4- to 5-cm left infraclavicular incision. For transaortic access, an upper median mini-sternotomy was performed. Concurrent anterolateral mini-thoracotomy was performed in the fifth or sixth intercostal space to obtain straight access to the left ventricular apex in the transapical approach. In most cases, after balloon valvuloplasty during rapid ventricular pacing, valve deployment was performed under fluoroscopy.

After TAVR, all patients were monitored in the intensive care unit for at least 1 day, and heart rate monitoring was continued until discharge. All patients received treatment for platelet inhibition with aspirin 100 mg daily. Postoperatively, most patients were also prescribed clopidogrel 75 mg daily for 3 months. Patients with an indication for anticoagulant therapy received clopidogrel and warfarin or a direct oral anticoagulant without aspirin.

Back to Top | Article Outline

2.2. Follow-up and data collection

Echocardiography and clinical monitoring by the heart valve team were performed before and after the procedure for all patients. Follow-up included telephone interviews and office visits. Cases were censored at death or upon completion of 6 months of follow-up, whichever occurred first.

Prediction of patient operative mortality after TAVR was calculated, using the Society of Thoracic Surgeons’ (STS) Online Risk Calculator (http://riskcalc.sts.org/stswebriskcalc/calculate). Echocardiographic studies performed at baseline and at 30 days after TAVR were evaluated according to the criteria of the American Society of Echocardiography.10

According to the Valve Academic Research Consortium-2 (VARC-2) consensus document, device success was defined as (1) the absence of procedural mortality, (2) the correct positioning of a single prosthetic heart valve into the proper anatomical location, and (3) the intended performance of the prosthetic heart valve (no prosthesis–patient mismatch with mean aortic valve gradient <20 mmHg or peak velocity <3 m/s and no moderate or severe prosthetic valve regurgitation).15 Procedural success was defined as the achievement of successful deployment of the TAVR device and retrieval of the delivery system in the absence of mortality, conversion to surgical aortic valve replacement, or myocardial infarction (MI).

The composite primary endpoint of this study was major cardiac and cerebral adverse event (MACCE) in terms of combined all-cause death, nonfatal MI, and disabling stroke within 180 days following TAVR. Early 30-day safety endpoints included New York Heart Association (NYHA) functional class III/IV heart failure, life-threatening bleeding, acute kidney injury (AKI, stage 3), major vascular complication, paravalvular leakage, and need for permanent pacemaker implantation for complete heart block. AKI (stage 3) was defined as a ≥3.0-fold increase in serum creatinine (SCr) from baseline or increase to SCr ≥4.0 mg/dL (≥354 mmol/L) within 72 hours according to VARC-2 criteria.11

Back to Top | Article Outline

2.3. Statistical analysis

Univariate analyses were conducted to compare demographic, procedural, and outcome parameters of patients grouped by the type of TAVR device implanted. Continuous variables are expressed as mean ± SD and were compared using the Student’s t test or the Wilcoxon rank sum test. Categorical variables are presented as number and percent frequency and were compared using the Pearson’s chi-square test or the Fisher’s exact test.

For the survival analysis, patients were divided into two groups depending upon whether or not MACCE (primary endpoint) occurred by the 180-day follow-up. Univariate comparisons of clinical characteristics and laboratory measurements between the two groups were made with appropriate tests. In the multivariate Cox proportional hazards analyses, the independent predictors of MACCE at 180 days were determined using variables including device types and those variables associated with the MACCE in the univariate analysis.

A two-sided p-value of <0.05 was considered statistically significant for all analyses. All statistical analyses were carried out using commercially available software (IBM SPSS for Windows, version 22.0; IBM Corp., New York, NY, USA).

Back to Top | Article Outline

3. RESULTS

Baseline demographic and clinical characteristics among the three groups of patients according to TAVR device are summarized in Table 1. The groups were generally well matched, although patients in the MCV group tended to be older (mean ages of MCV, Lotus, and SXT groups were 80±8, 78±8, and 78±9 years, respectively; p=0.056). Peripheral artery disease was less frequent in the Lotus group (17%) than in the MCV group (29%) and the SXT group (40%) (p=0.068), and end-stage renal disease needing dialysis was less frequent in the MCV group (5%) than in the Lotus (11%) and SXT (13%) groups (p=0.078), although the differences were not significant. There was no significant difference in mean STS score or the proportion of patients with NYHA functional class III/IV heart failure at presentation among the three groups.

Table 1

Table 1

Although the type and size of the prosthesis to be used for each patient was chosen based on echocardiographic and CT findings, the baseline echocardiography and CT measurements showed no significant differences among the three groups, as shown in Table 2.

Table 2

Table 2

Technical aspects of the procedure and procedural outcomes are presented in Table 3. Valve sizes ranged from 23 to 31 mm for MCV, 23 to 27 mm for Lotus, and 23 to 29 mm for SXT. Compared with the other two groups, mean valve size used in the Lotus group was significantly smaller (p<0.001), and a greater proportion of the implanted devices were sized ≤26 mm (Lotus: 94%, MCV: 55%, and SXT: 84%; p<0.001). TAVR procedure was performed via TF (91%), trans-subclavian (4%), or transaortic (5%) approach with MCV. TF access was the only approach applied for Lotus valve implantation. The balloon-expandable transcatheter SXT valves were implanted via TF, transapical, or transaortic access in 85%, 10%, and 5% of patients, respectively. In addition, balloon valvuloplasty for pre-dilatation was less frequently required with Lotus implantation compared with MCV and SXT (Lotus: 50%, MCV: 71%, and SXT: 97%; p<0.001), whereas balloon valvuloplasty for post-dilatation was needed more frequently with SXT (SXT: 29%, MCV: 13%, and Lotus: 11%; p=0.008). The mean final implantation depth below annulus was significantly deeper in patients of the MCV group (MCV: 5.3±2.5 mm, Lotus: 1.9±1.1 mm, and SXT: 2.3±1.1 mm; p<0.001). Implantation depth is assessed routinely by fluoroscopy in our institution. The distance from the zero level to the ventricular edge of the TAVR frame was measured using Siemens software (Syngo.viaTMVB20A, Siemens Healthcare, Forchheim, Germany) for distance measurements. The bottom of the TAVR frame was first aligned with the axis of the device. Then, the distances between the lower edges of the TAVR frame and the non-coronary cusp and the left coronary cusp were acquired and averaged.

Table 3

Table 3

Ten (9%) patients with MCV and four (4%) patients with SXT required implantation of a second valve due to initial implant embolization to the ascending aorta or malpositioning (p=NS; non-significant). Moderate or more severe paravalvular leakage after TAVR procedure was found in four (4%) patients with MCV, one (6%) patient with Lotus, and two (2%) patients with SXT (p=NS). Two patients in each group had a post-procedural transvalvular gradient of >20 mmHg (p=NS). Overall, the device success rate was 86% for MCV, 83% for Lotus, and 90% for SXT (p=NS).

Rates of major intraoperative complications, including emergency conversion to open-heart surgery, annular or left ventricular rupture, coronary occlusion, or need for emergent hemodynamic support, were similar among the three groups, and the procedural success rate was 99% for MCV, 100% for Lotus, and 98% for SXT (p=NS). Mean procedure and fluoroscopic times were also similar among the three groups. However, the MCV group received lower mean contrast volume (MCV: 112.1±40.2 mL, Lotus: 132.7±58.4 mL, and SXT: 142.1±59.2 mL; p<0.001).

Table 4 shows the hemodynamic performance of the three TAVR devices. A significant reduction in prosthetic valvular pressure gradient (PG) and increase in prosthetic aortic valve area (AVA) at 30 days were observed in all patients who underwent successful TAVR. However, in patients receiving a Lotus valve, a significantly higher mean transaortic valve PG (Lotus: 14.0±6.7 mmHg, MCV: 8.1±3.5 mmHg, and SXT: 8.2±3.6 mmHg; p<0.001) and smaller AVA (Lotus: 1.7±0.2 cm2, MCV: 1.8±0.3 cm2, and SXT: 2.0±0.3 cm2; p<0.001) were observed. Follow-up echocardiography revealed significant improvements in left ventricular ejection fraction and right ventricular systolic pressure at 30 days post TAVR in nearly all patients with no significant difference among groups. Incidence of moderate or severe aortic regurgitation (AR) was low overall and not statistically different among the three devices used (MCV: 7%, Lotus: 6%, and SXT: 3%; p=0.352).

Table 4

Table 4

Clinical outcomes of the study cohort are shown in Table 5. Mean duration of stay in the intensive care unit did not differ among the three groups. Significant improvement in NYHA functional class was observed in all groups. At 30 days post TAVR, there were no significant difference in rates of all-cause mortality (MCV: 3%, Lotus: 0%, and SXT: 3%; p=0.605), cardiovascular mortality (MCV: 2%, Lotus: 0%, and SXT: 3%; p=0.564), periprocedural MI (0% in all groups; p=NS), stroke (MCV: 2%, Lotus: 0%, and SXT: 1%; p=0.691), major or life-threatening bleeding (MCV: 2%, Lotus: 0%, and SXT: 5%; p=0.239), major vascular complications (MCV: 11%, Lotus: 17%, and SXT: 5%; p=0.388), or AKI (stage 3) (MCV: 9%, Lotus: 0%, and SXT: 3%; p=0.130). Notably, patients in the Lotus group had a significantly higher rate of need for a permanent pacemaker (Lotus: 22%, MCV: 13%, and SXT: 3%; p=0.004).

Table 5

Table 5

At 180 days, no significant difference among groups in the rate of MACCE, including all-cause death, cardiovascular death, recurrent nonfatal MI, and stroke, was found. Multivariate analysis identified the lack of procedural success (p=0.023), presence of AKI (stage 3; p=0.037), and presence of coronary obstruction (p=0.05) to be independently associated with the primary composite endpoint within 180 days after TAVR (Table 6).

Table 6

Table 6

Back to Top | Article Outline

4. DISCUSSION

The main findings of our study are: (1) the three types of TAVR devices studied demonstrated favorable and comparable safety and efficacy outcomes at 30 and 180 days in this real-world cohort; (2) the Lotus device was associated with higher post-TAVR transaortic valve PG and greater need for permanent pacemaker implantation; and (3) the presence of acute coronary occlusion during TAVR, AKI, and persistent refractory heart failure after TAVR were independent predictors of adverse outcomes at 180 days.

In line with the results of the CHOICE trial, device success rate was somewhat lower with MCV at 86% than with the other two devices in this study, which can be mainly attributed to higher rates of embolization/migration and moderate or severe AR.12,13 Compared with SXT, which was associated with a device success rate of 90%, device success rate was also lower with the Lotus valve at 83%, which is even lower than that reported in the REPRISE III trial.14 This was mainly driven by a higher rate of patient–prosthesis mismatch. Comparable to the findings of the REPRISE III study, we found a similar higher mean post-TAVR transaortic valve PG of 12 mmHg and smaller mean AVA of 1.59 cm2 with the Lotus valve compared to the other devices.14 In the present study, despite well-matched baseline annular dimensions, a significantly larger number of small prostheses were inserted in the Lotus cohort in following with the manufacturer’s recommendations of less oversizing with this device, possibly relating to the valve design. It is possible that the smaller average device size contributed to a higher mean gradient. However, severe patient–prosthesis mismatch after TAVR is rare and can be predicted by larger body surface area and does not seem to affect midterm mortality or composite clinical outcome. Larger studies are needed to identify different independent predictors of PPM and to elucidate its impact in terms of device durability and long-term clinical efficacy. Moreover, the lower device success rates with MCV and Lotus valves in our study may have occurred by chance, as our study was underpowered for that event, but it could also have been related to the learning curve specific to those prostheses. Previous studies have shown that MCV implantation, having the advantage of supra-annular valve positioning, led to a higher AVA and lower MPG in comparison to SXT.12,13 In our study, however, MCV was associated with a smaller AVA, possibly related to deeper implantation, less frequent pre-dilatation, and slightly smaller baseline annulus diameter.

It is worth noting that the necessity for permanent pacemaker implantation was significantly higher with MCV (13%) and Lotus valve (22%) than with STX (3%) in this study, as has been shown in many previous publications.5,12,14 In the head-to-head comparison REPRISE III study, the use of Lotus was associated with low rates of death, stroke, and paravalvular leakage, but the rate of pacemaker implantation at 30 days was nearly doubled in Lotus-treated vs MCV-treated patients (35.5% vs 19.6%; p<0.001).14 It has been well shown that pacemaker implantation leads to greater ventricular dysfunction and increased hospitalization costs, despite having no impact on mortality in TAVR patients.15 Hence, even though the Lotus device is expected to return to market as the newer-generation Lotus Edge, whether the higher post-TAVR PG and pacemaker implantation rate would compromise long-term outcomes remains uncertain. As for MCV, the higher pacemaker implantation rate may be related to deeper implantation of the device. Fortunately, the next generation of the MCV device, the Evolut-R valve, allows retrieval and repositioning. Recently launched in Taiwan, the new device is expected to result in a lower pacemaker implantation rate compared with the earlier MCV valve. The Evolut-R valve may also improve device success rate through reduction in embolization/migration and moderate or severe AR.

Among uncommon but clinically relevant complications following the TAVR procedure, acute coronary obstruction during valve implantation or delayed coronary obstruction occurring in the early or late (≥60 days) post-procedural phase may occur.16,17 In the present study, the development of coronary obstruction within 30 days was one of the main determinants of adverse outcome at 180 days. In this regard, the possibility of coronary obstruction should be considered during the prosthesis selection process. For example, the Evolut-PRO valve has a longer frame that extends beyond the coronary ostia, thereby likely lowering risk of coronary obstruction.18 On the other hand, valves with a larger stent cell size and shorter frame may facilitate future access to the coronary orifices.19 Moreover, certain newer-generation TAVR devices are fixed in place via direct anchoring to either calcified native leaflets or surgical valve leaflets, which mitigates the risk of future valve tissue prolapse and coronary obstruction.20 In addition, to optimize future coronary reaccess, implantation depth is critical, especially if the ostia is <10 mm.18

In our series, the presence of stage 3 AKI was a significant predictor of adverse events. The etiology of post-TAVR AKI is multifactorial, but the principle procedural issues are contrast-induced nephropathy and renal hypoperfusion secondary to intra-procedural hypotension.21 Although the mean amounts of contrast used in our study were similar to those reported in previous publications, there is still room for improvement. For those patients with preexisting chronic renal impairment, the use of a self-expandable or mechanically expandable valve that features retrievability and repositionability under the guidance of echocardiography or fluoroscopy–echocardiography fusion imaging may allow more accurate valve placement without the need for contrast aortography.22

Back to Top | Article Outline

4.1. Study limitations

First, although the three groups of patients according to prosthesis used were similar in terms of comorbidities and pre-procedural risk, this was not a randomized trial; therefore, the study is subject to selection bias and unmeasured confounders. Second, it was performed at a single center including a relatively small number of patients. Third, as follow-up was restricted to 180 days, further research on longer-term outcomes of these three devices is warranted.

In conclusion, our data clearly demonstrated that there were no differences in VARC-2 combined early safety endpoints at 30 days or the primary endpoint at 180 days among patients receiving MCV, Lotus, or SXT devices. Results suggest that once a heart team achieves technical proficiency in the implantation of a specific device, post-TAVR adverse events seem to be mainly patient specific. This study compared the technical success rate, hemodynamic performance, and VARC-2 outcomes of three different TAVR devices in a single center in Asia. The findings could serve as valuable benchmarks for guiding patient selection for specific TAVR devices and assessing the performance of newer-generation TAVR devices in the future.

Back to Top | Article Outline

REFERENCES

1. Kapadia SR, Leon MB, Makkar RR, Tuzcu EM, Svensson LG, Kodali S, et al; PARTNER trial investigators5-year outcomes of transcatheter aortic valve replacement compared with standard treatment for patients with inoperable aortic stenosis (PARTNER 1): a randomised controlled trial.Lancet20153852485–91
2. Leon MB, Smith CR, Mack M, Miller DC, Moses JW, Svensson LG, et al; PARTNER Trial InvestigatorsTranscatheter aortic-valve implantation for aortic stenosis in patients who cannot undergo surgery.N Engl J Med20103631597–607
3. Popma JJ, Adams DH, Reardon MJ, Yakubov SJ, Kleiman NS, Heimansohn D, et al; CoreValve United States Clinical InvestigatorsTranscatheter aortic valve replacement using a self-expanding bioprosthesis in patients with severe aortic stenosis at extreme risk for surgery.J Am Coll Cardiol2014631972–81
4. Smith CR, Leon MB, Mack MJ, Miller DC, Moses JW, Svensson LG, et al; PARTNER Trial InvestigatorsTranscatheter versus surgical aortic-valve replacement in high-risk patients.N Engl J Med20113642187–98
5. Adams DH, Popma JJ, Reardon MJ, Yakubov SJ, Coselli JS, Deeb GM, et al; U.S. CoreValve Clinical InvestigatorsTranscatheter aortic-valve replacement with a self-expanding prosthesis.N Engl J Med20143701790–8
6. Pilgrim T, Windecker S. Newer-generation devices for transcatheter aortic valve replacement: resolving the limitations of first-generation valves?JACC Cardiovasc Interv20169373–5
7. Chamandi C, Puri R, Rodriguez-Gabella T, Rodés-Cabau J. Latest-generation transcatheter aortic valve replacement devices and procedures.Can J Cardiol2017331082–90
8. Bedogni F, Frigiola A, Ranucci M, Brambilla N, Montone RA, Agnifili M, et al. Transcatheter aortic valve replacement-state of the art and a glimpse to the future: ‘the tailored approach’.Eur Heart J Suppl201618Suppl EE86–95
9. Bjursten H, Nozohoor S, Johansson M, Zindovic I, Appel CF, Sjögren J, et al. The safety of introducing a new generation TAVR device: one departments experience from introducing a second generation repositionable TAVR.BMC Cardiovasc Disord20171725
10. Zoghbi WA, Chambers JB, Dumesnil JG, Foster E, Gottdiener JS, Grayburn PA, et al; American Society of Echocardiography’s Guidelines and Standards Committee; Task Force on Prosthetic Valves; American College of Cardiology Cardiovascular Imaging Committee; Cardiac Imaging Committee of the American Heart Association; European Association of Echocardiography; European Society of Cardiology; Japanese Society of Echocardiography; Canadian Society of Echocardiography; American College of Cardiology Foundation; American Heart Association; European Association of Echocardiography; European Society of Cardiology; Japanese Society of Echocardiography; Canadian Society of EchocardiographyRecommendations for evaluation of prosthetic valves with echocardiography and Doppler ultrasound: a report from the American society of echocardiography’s guidelines and standards committee and the task force on prosthetic valves, developed in conjunction with the American college of cardiology cardiovascular imaging committee, cardiac imaging committee of the American heart association, the European association of echocardiography, a registered branch of the European society of cardiology, the Japanese society of echocardiography and the Canadian society of echocardiography, endorsed by the American college of cardiology foundation, American heart association, European association of echocardiography, a registered branch of the European society of cardiology, the Japanese society of echocardiography, and Canadian society of echocardiography.J Am Soc Echocardiogr200922975–1014quiz 1082–4
11. Kappetein AP, Head SJ, Généreux P, Piazza N, van Mieghem NM, Blackstone EH, et al. Updated standardized endpoint definitions for transcatheter aortic valve implantation: the valve academic research consortium-2 consensus document.Eur Heart J2012332403–18
12. Abdel-Wahab M, Mehilli J, Frerker C, Neumann FJ, Kurz T, Tölg R, et al; CHOICE investigatorsComparison of balloon-expandable vs self-expandable valves in patients undergoing transcatheter aortic valve replacement: the CHOICE randomized clinical trial.JAMA20143111503–14
13. Abdel-Wahab M, Neumann FJ, Mehilli J, Frerker C, Richardt D, Landt M, et al; CHOICE Investigators1-year outcomes after transcatheter aortic valve replacement with balloon-expandable versus self-expandable valves: results from the CHOICE randomized clinical trial.J Am Coll Cardiol201566791–800
14. Feldman TE, Reardon MJ, Rajagopal V, Makkar RR, Bajwa TK, Kleiman NS, et al. Effect of mechanically expanded vs self-expanding transcatheter aortic valve replacement on mortality and major adverse clinical events in high-risk patients with aortic stenosis: the REPRISE III randomized clinical trial.JAMA201831927–37
15. Urena M, Webb JG, Tamburino C, Muñoz-García AJ, Cheema A, Dager AE, et al. Permanent pacemaker implantation after transcatheter aortic valve implantation: impact on late clinical outcomes and left ventricular function.Circulation20141291233–43
16. Sultan I, Siki M, Wallen T, Szeto W, Vallabhajosyula P. Management of coronary obstruction following transcatheter aortic valve replacement.J Card Surg201732777–81
17. Jabbour RJ, Tanaka A, Finkelstein A, Mack M, Tamburino C, Van Mieghem N, et al. Delayed coronary obstruction after transcatheter aortic valve replacement.J Am Coll Cardiol2018711513–24
18. Yudi MB, Sharma SK, Tang GHL, Kini A. Coronary angiography and percutaneous coronary intervention after transcatheter aortic valve replacement.J Am Coll Cardiol2018711360–78
19. Søndergaard L, De Backer O. Transcatheter aortic valve implantation: don’t forget the coronary arteries!Eurointervention201814147–9
20. Jabbour RJ, Tanaka A, Colombo A, Latib A. Delayed coronary occlusion after transcatheter aortic valve implantation: implications for new transcatheter heart valve design and patient management.Interv Cardiol201813137–9
21. Scherner M, Wahlers T. Acute kidney injury after transcatheter aortic valve implantation.J Thorac Dis201571527–35
22. Eskandari M, Aldalati O, Byrne J, Dworakowski R, Wendler O, Alcock E, et al. Zero contrast transfemoral transcatheter aortic valve replacement using fluoroscopy-echocardiography fusion imaging.Am J Cardiol20161171861–2
Keywords:

Balloon-expandable valve; Mechanically expanded valve; Outcomes; Self-expanding valve; Transcatheter aortic valve replacement