Successful removal of a potentially lethal left atrial thrombus detected by transesophageal echocardiography following the removal of a left ventricular assist device inflow cannula

Chuang, Chia-Chun; Lee, Chien-Ching; Liao, Shu-Wei; Ting, Chien-Kun; Chu, Chi-Chun; Chan, Kwok-Hon; Lin, Su-Man*

Author Information
Journal of the Chinese Medical Association: October 2012 - Volume 75 - Issue 10 - p 543-546
doi: 10.1016/j.jcma.2012.07.010

    Abstract

    1. Introduction

    The left ventricular assist device (LVAD) is a battery-operated, mechanical pump-type device that is usually used as a so-called “bridge to transplant” for severe heart failure patients. However, there are a few problems related to LVAD usage, of which inflow and outlet cannula obstruction induced by thrombus formation1 is one of the most severe and frequently observed. LVAD thrombogenicity2 is mainly due to the anticoagulation-resistant surface thrombin activity harbored by LVAD Dacron grafts. This complication can lead to death if the thrombus causes obstruction in the flow between the concurrent LVAD and a failed heart and then circulatory collapse follows.

    Transesophageal echocardiography (TEE) is the most commonly used technique for the evaluation of obstructed LVAD conduits.3 Under TEE observation, pump flow and performance can be optimized by correct positioning of the cannula in LVAD recipients. Moreover, TEE may help in identifying detailed morphologic characteristics of an intracardiac mass lesion, such as a thrombus.4,5

    Here, we present the case of a patient with a lethal left atrial thrombus that was detected by TEE after the removal of the LVAD cannula. Under TEE monitoring, the LVAD was successfully removed, and the patient was uneventfully discharged later.

    2. Case report

    A 75-year-old man with a history of hypertension was admitted to the emergency department because of acute chest pain and shortness of breath. The patient went into shock when his blood pressure dropped to 76/48 mmHg. An emergency electrocardiographic examination revealed sinus rhythm with ST segment depression and T-wave inversion in leads V3–6. Serum analysis revealed elevated levels of cardiac enzymes [creatine kinase (CK), 1245 U/L; MB isoenzyme of CK, 101 U/L; troponin-I, 39 ng/mL]. Emergent coronary angiography was performed, revealing severe coronary artery disease with triple vessel disease [LAD-P total occlusion, LCX-P 90% stenosis, -D 70% stenosis, - OM3 95 % stenosis, RCA-P to -M segmental stenosis, with collaterals to LAD], apical aneurysm, acute myocardial infarction, and poor left ventricular (LV) function. The LV ejection fraction was extremely low, about 13%. Because the patient had severe stenosis of the coronary artery, stent insertion was considered relatively difficult and risky. Surgical intervention was suggested, and a cardiovascular surgeon was consulted.

    The LV ejection fraction improved to 29% after several days of medical treatment and intra-aortic pump support. However, transthoracic echocardiography (TTE) revealed moderate mitral regurgitation, which resulted from the dilated mitral annulus with a 10 cm × 8-cm LV aneurysm present. Considering the patient's poor heart function and that there were already collaterals from the RCA to the LAD, the surgeon decided to revascularize only the LCX to shorten the operation time for to get the most benefit for this patient. Coronary artery bypass graft (CABG) surgery [gastroepiploic artery to the first obtuse marginal branch], left ventricular aneurysmectomy, and mitral valve annuloplasty (28 mm Edwards IMR ETlogix annuloplasty ring; Edwards, Irvine, CA, USA) were scheduled. Because of poor LV function and the high dose of inotropic agents administered during cardiopulmonary bypass removal, an LVAD (Levitronix CentriMag Blood Pump; Levitronix, Waltham, Mass, USA) was chosen for circulatory support instead of extracorporeal membrane oxygenation because of having fewer complications and better outcome.6 The LVAD was implanted with inflow from the left atrium and outflow to the ascending aorta; the flow was maintained at approximately 3.6 L/min. However, the LVAD flow occasionally dropped to 1.51 L/min, and TTE revealed a mass over the inflow cannula 3 weeks after implantation (Fig. 1). The patient received thrombolytic therapy, but it was ineffective, and therefore the LVAD had to be removed. Before the procedure, TEE showed an organized thrombus of diameter 3.52 cm × 1.71 cm encircling the LVAD inflow cannula in the left atrium (Fig. 2). The entire procedure of LVAD removal under general anesthesia was performed under real-time TEE monitoring. A lethal thrombus was detected in the left atrium after the LVAD cannula was removed (Fig. 3). Cardiopulmonary bypass was performed immediately, and the thrombus was successfully removed by opening the left atrial chamber. TEE showed no residual thrombus in the left atrium(Fig. 4).

    Fig. 1
    Fig. 1:
    Preoperative transthoracic echocardiography revealed a suspected thrombus encircling the left atrial tubing of the left ventricular assist device (white arrow).
    Fig. 2
    Fig. 2:
    Transesophageal echocardiography before the removal of the left ventricular assist device (LVAD) inflow cannula revealed a thrombus of diameter 3.52 × 1.71 cm (arrows) attached to the LVAD inflow cannula in the left atrium in the mid-esophageal four chamber view.
    Fig. 3
    Fig. 3:
    Transesophageal echocardiography after the removal of the left ventricular assist device (LVAD) inflow cannula revealed (A) a retained thrombus after the removal of the LVAD tube; (B) the tubular thrombus at 80°.
    Fig. 4
    Fig. 4:
    Resected thrombus; (A) the tubular thrombus that was resected; (B) the cross section of the thrombus.

    3. Discussion

    LVAD is increasingly used as a bridge to cardiac transplantation,7,8 for destination therapy in patients with end-stage heart failure and for postcardiotomy cardiogenic shock occurring in 2% to 6% of all patients undergoing CABG or valvular surgery.9 However, LVAD usage often poses complex hematologic challenges, including thromboembolism, which can lead to death.1 TTE or TEE can be used for thrombus detection,10 and further management including thrombolytic therapy or surgical intervention can be carried out.

    LVAD recipients often have increased levels of platelet release and thrombin activity during device implantation. LVAD Dacron grafts harbor anticoagulation-resistant surface thrombin activity, which may play an important role in LVAD thrombogenicity.2 There is evidence of increased thrombin generation and fibrinolysis in LVAD recipients, despite a normal prothrombin time, activated partial thromboplastin time, and platelet count.1 Thus, thromboembolism is a significant problem in LVAD implantation; it occurs in up to 35% of patients with adverse sequelae.11 In an expected short-term-use LVAD system, continuous intravenous heparin will be prescribed to keep activated clotting time between 200 and 250 seconds. Even though thromboembolism is a significant risk, and Symbol may be related to the relationship between cannula size and flow rate, more importantly, it is related to the concurrent activation of the coagulation and fibrinolytic systems.12

    Symbol
    Symbol

    In our case, we chose the left atrium to be the inflow site of the LVAD because the operator could instantly establish an effective circulation system and also because the purpose was for short-term assistance of LV function. The inflow site and the native heart function can affect the flow rate; Tevaearai et al13 concluded that drainage through the left atrium (LA) or the LV was similar when CVP was set at 8 mmHg, and increasing CVP to 14 mmHg allowed for better drainage through the LV cannula. Due to easier collapse of LA and there Symbol being lower flow through the LVAD inflow cannula from the LA compared to from the LV, the risk of thromboembolism seems higher in the LVAD recipients with the inflow from the LA. That may be one reason for our patient receiving heparin therapy following LVAD implantation; there was still a thromboembolic event occurring. Some studies report that thrombolytic therapy has been used to successfully treat thrombi in LVAD recipients,14,15 while this therapy has been found to be ineffective in some other studies.16 Similarly, our patient received heparin for prophylaxis and thrombolytic therapy following thrombus detection, but the result was unsatisfactory.

    Symbol
    Symbol

    Although TTE revealed the lesion before the operation, the detailed relationship between the mass lesion, left atrium, and LVAD inflow cannula was unclear. Instead, TEE can provide more information on the nature of a lesion.5 In a study by Srichai,17 TEE showed a higher sensitivity (40% ± 14%) than TTE (23% ± 12%) for thrombus detection in 160 patients with ischemic heart disease and thrombus formation. Many studies have proven the greater efficiency of TEE compared to than TTE in diagnosing cardiovascular diseases4,5,18; TEE allows identification of detailed morphologic characteristics of intracardiac mass lesions. In our case, TEE monitoring during the whole procedure was especially effective because TEE access is closest to the LA.

    Thus, as TEE is a powerful tool for heart evaluation, some studies have advocated the efficacy of TEE monitoring during LVAD removal.19 For LVAD removal, TEE is effective not only in evaluating ventricular function but also for detecting problems such as a residual thrombus following LVAD decannulation.19 One study with four cases of LVAD malfunction20 concluded that TEE played a pivotal role in the clinical management of LVAD failure. TEE can be used not only to assess the patency and position of the LVAD cannula but also to examine the source of the thromboembolic material. In our case, although the LVAD removal procedure did not initially require cardiopulmonary bypass, we decided to use TEE monitoring for evaluating the ventricular function and the nature of the thrombus. This led to the accidental detection of the thrombus in the left atrium retained after LVAD removal, which could have led to a lethal complication.

    According to the Practice Guidelines for Perioperative Transesophageal Echocardiography (Anesthesiology 2010; 112:1–1.), TEE monitoring is strongly recommended during modern open-heart surgery, including LVAD implantation. However, TEE is not routinely used during the procedures of cannulation removal of circulatory assist device, such as intraaortic balloon, extracorporeal membrane oxygenation device, or ventricular assist device. These procedures are considered relatively simple, and some even can be performed at bedside. There could be a catastrophe if thrombosis formed around the cannula and residual thrombosis was left in the heart after cannula removal. Therefore TEE monitoring in these procedures becomes more important. In our case, opening the left atrium was not in the surgical plan of LVAD removal in the beginning. If we had not used TEE as a monitoring modality, the residual thrombus would not have been detected and lethal complication might have happened. Furthermore, the inflow cannula was inserted in the LA, which could be estimated more precisely by TEE, especially when there were problems such as decreased inflow rate or mechanical obstruction. Through this experience, we strongly recommend that TEE should be routinely used in LVAD removal, especially in cases with thrombus formation, and also should be considered in the perioperative period when LVAD malfunction happens.

    References

    1. Spanier T, Oz M, Levin H, Weinberg A, Stamatis K, Stern D, et al. Activation of coagulation and fibrinolytic pathways in patients with left ventricular assist devices. J Thorac Cardiovasc Surg. 1996;112:1090-1097.
    2. Wagner WR, Johnson PC, Heil BV, Thompson KA, Kormos RL, Griffith BP. Thrombin activity resides on LVAD Dacron inflow and outflow grafts. ASAIO J. 1992;38:M634-M637.
    3. Ferns J, Dowling R, Bhat G. Evaluation of a patient with left ventricular assist device dysfunction. ASAIO J. 2001;47:696-698.
    4. Shyu KG, Chen JJ, Cheng JJ, Hwang JJ, Kuan P, Lien WP. Comparison of transthoracic and transesophageal echocardiography in the diagnosis of intracardiac tumors in adults. J Clin Ultrasound. 1994;22:381-389.
    5. Mugge A, Daniel WG, Haverich A, Lichtlen PR. Diagnosis of noninfective cardiac mass lesions by two-dimensional echocardiography: comparison of the transthoracic and transesophageal approaches. Circulation. 1991;83:70-78.
    6. Klotz S, Rukosujew A, Welp H, Schmid C, Tjan TD. Primary extracorporeal membrane oxygenation versus primary ventricular assist device implantation in low cardiac output syndrome following cardiac operation. Artif Organs. 2007;31:390-394.
    7. Camp D. The left ventricular assist device (LVAD): a bridge to heart transplantation. Crit Care Nurs Clin North Am. 2000;12:61-68.
    8. Wang SS, Ko WJ, Chen YS, Hsu RB, Chou NK, Chu SH. Mechanical bridge with extracorporeal membrane oxygenation and ventricular assist device to heart transplantation. Artif Organs. 2001;25:599-602.
    9. DeRose JJ Jr, Umana JP, Argenziano M, Catanese KA, Levin HR, Sun BC, et al. Improved results for postcardiotomy cardiogenic shock with the use of implantable left ventricular assist devices. Ann Thorac Surg. 1997;64:1757-1762.
    10. Scalia GM, McCarthy PM, Savage RM, Smedira NG, Thomas JD. Clinical utility of echocardiography in the management of implantable ventricular assist devices. J Am Soc Echocardiogr. 2000;13:754-763.
    11. Apostolidou I, Sweeney MF, Missov E, Joyce LD, John R, Prielipp RC. Acute left atrial thrombus after recombinant factor VIIa administration during left ventricular assist device implantation in a patient with heparin-induced thrombocytopenia. Anesth Analg. 2008;106:404-408.
    12. Gross DR. Concerning thromboembolism associated with left ventricular assist devices. Cardiovasc Res. 1999;42:45-47.
    13. Tevaearai HT, Mueller XM, Jegger D, Horisberger J, Von Segesser L. Atrial, ventricular, or both cannulation sites to optimize left ventricular assistance? ASAIO J. 2001;47:261-265.
    14. Tschirkov A, Nikolov D, Tasheva I, Papantchev V. Successful fibrinolysis after acute left ventricular assist device thrombosis. J Heart Lung Transplant. 2007;26:553-555.
    15. Delgado R 3rd, Frazier OH, Myers TJ, Gregoric ID, Robertson K, Shah NA, et al. Direct thrombolytic therapy for intraventricular thrombosis in patients with the Jarvik 2000 left ventricular assist device. J Heart Lung Transplant. 2005;24:231-233.
    16. Zotz RJ, Pinnau U, Genth S, Erbel R, Meyer J. Left atrial thrombi despite anticoagulant and antiplatelet therapy. Clin Cardiol. 1994;17:375-382.
    17. Srichai MB, Junor C, Rodriguez LL, Stillman AE, Grimm RA, Lieber ML, et al. Clinical, imaging, and pathological characteristics of left ventricular thrombus: a comparison of contrast-enhanced magnetic resonance imaging, transthoracic echocardiography, and transesophageal echocardiography with surgical or pathological validation. Am Heart J. 2006;152:75-84.
    18. Kuznetsova LM, Rykunov IE, Sandrikov VA. Comparison of transesophageal and transthoracic echocardiography in cardiac surgical care. Vestn Ross Akad Med Nauk. 1997;11:42-46.
    19. Lanquetot H, Catoire P, Corbi P, Charriere JM, Baudoin D, Fusciardi J. Detection of left atrial thrombus by transesophageal echocardiography after left ventricular assist device decannulation. J Cardiothorac Vasc Anesth. 1998;12:435-436.
    20. Catena E, Milazzo F, Montorsi E, Bruschi G, Cannata A, Russo C, et al. Left ventricular support by axial flow pump: the echocardiographic approach to device malfunction. J Am Soc Echocardiogr. 2005;18:1422.
    Keywords:

    left ventricular assist device; thrombus; transesophageal echocardiography

    © 2012 by Lippincott Williams & Wilkins, Inc.