The malfunction of vascular access accounts for considerable morbidity and mortality in hemodialysis (HD) patients. It is responsible for 17 to 25% of all hospitalizations among dialysis patients in the United States (1 , 2 ). Treatment of these problems requires invasive, expensive, and time-consuming procedures, which are associated with potential risk of complication. Failure of dialysis access can result from either inadequate blood flow on account of stenosis of the venous outflow tract or complete occlusion as a result of thrombosis (3 ). Decreased access flow is associated with an increased risk for access thrombosis. Access flow (Qa) <400 ml/min in native arteriovenous fistulas (AVF) and <800 ml/min in polytetrafluoroethylene (PTFE) grafts was demonstrated to be predictive of access failure in one ultrasonographic study of 477 AVF and 831 PTFE grafts (4 ). The methods of measuring Qa include dye dilution (5 ), constant infusion of technetium-99m (6 ), Doppler ultrasonography (7 ), electromagnetic flow meters (8 ), and magnetic resonance angiography (9 ). However, some of the above methods are invasive or highly technique dependent, and some require additional laboratory tests. Some of them are impractical and costly or cannot be performed during HD. In the past few years, several noninvasive techniques have been developed for calculating Qa, including ultrasound dilution (10 ), photo-optical sensor (11 ), conductivity dialysance (12 ), thermodilution (13 ), and ultrafiltration method (14 ).
Among these methods, the ultrasound dilution method (UDM), developed by Krivitski (10 ), has been widely accepted as a practical, reliable, and noninvasive method for Qa measurement. Besides, color Doppler ultrasound method additionally provides an anatomic picture of a dialysis access and its venous runoff. It not only is noninvasive but also can be performed serially. However, both of these methods have shortcomings that limit their application in access surveillance. UDM is time-consuming and requires dialysis blood line reversal. Duplex imaging is subject to operator-dependent error and is labor-intensive. A novel method (variable flow Doppler method) measures blood flow by using a changing Doppler signal as a function of blood flow in the access during dialysis (15 , 16 ). In comparison with the above two methods, the variable flow Doppler procedure does not have the aforementioned drawbacks. However, that technique requires additional equipment that is not easily available. Therefore, we developed a new technique—variable pump flow–based Doppler ultrasound method (VPFDUM)—that combines the advantages of the variable-flow Doppler procedure with additional imaging information of vascular access obtained by spectral analysis of Duplex Doppler study. This study compared Qa measurements by VPFDUM, ultrasound dilution method, and conventional Doppler ultrasound method (CDUM) in HD patients.
Materials and Methods
Patient Selection
Patients who were included in this study met the following criteria: (1 ) on maintenance HD for at least 6 mo at Taipei veterans general hospital and (2 ) present vascular access in use for more than 6 mo, without interventions within the last 3 mo. All patients were dialyzed three times weekly on standard bicarbonate dialysate bath (38 mEq/L HCO3 , 3.0 mEq/L Ca++ , 2.0 mEq/L K+ ), using the volumetric-controlled dialysis delivery system under constant dialysate flow at 500 ml/min. Patients were anticoagulated by means of systemic heparin, without change of the individual bolus or maintenance dose throughout the study. After obtaining informed consent from every patient who wanted to participate in this study, we measured Qa according to the following procedures.
Qa Measurement by UDM
Qa was measured during HD by UDM using the Transonic HD02 hemodialysis monitor (Transonic Systems, Inc., Ithaca, NY). The technique is widely used and validated extensively in the literature (10 ). In brief, the technique uses two ultrasound sensors attached to the two HD tubing lines, one to the arterial and another to the venous catheters, approximately 3 to 5 inches distant from the connection of tubing to dialysis needles. Initially, tubing lines are reversed, and ultrafiltration is turned off. The blood pump flow is set at different levels according to the protocol. A measured bolus of saline (10 ml) is injected into the venous catheter, resulting in changes in sound velocity that are measured by the transducers on the catheters. This change is then calculated by the Transonic software, giving the result of Qa (ml/min).
Qa Measurement by CDUM
Color Doppler ultrasonography was carried out with commercially available equipment (Model SSA 340A; Toshiba, Tokyo, Japan). The procedure was performed during HD with the patient in a supine position. A 7.5-MHz linear array transducer was placed on the skin, and the angle between the ultrasound beam and the vessel was optimized at 60 degrees. Velocity and Qa were determined by spectral analysis of flow at the site approximately 3 cm upstream from the placement of arterial catheter. The access flow was calculated according to Zierler et al. (17 ): Access flow (ml/min)=time-averaged maximum velocity (cm/s) × radius2 (cm2 ) × π × 60. All examinations were performed and interpreted under the direction of the same ultrasound expert (H.J.C.).
Qa Measurement by VPFDUM
Figure 1 shows the schematic of a dialysis vascular access with flow direction shown by arrows and the probe placement shown by the box. The vascular access flow is represented by Qa, and the pump flow is represented by Qb. As we know, Qa=A × v, where A is the cross-sectional area of the dialysis access and v is the average blood velocity in the vascular access. The access flow detected at transducer site 1 of the vascular access under Qb at 0 is represented by Qa0 (static Qa). The access flow (Qa) may be changed under different Qb. We assumed that the different Qa [Qa(n)] might be correlated with Qa0 (static Qa) and the different pump flow [Qb(n)] as follows:
The abbreviation “v(n)” represents the velocity detected by the ultrasound probe at transducer site 1 under different pump flow [Qb(n)]. The flow detected by the ultrasound probe at transducer site 1 is represented by Qa1 (under Qb1) or by Qa2 (under Qb2). While the probe is fixed at the same detection angle and placing site during these measurements, the cross-sectional area of dialysis access (represented by A) is the same. We can obtain the following equations:
Qa0 represents the access flow when Qb is 0 (static access flow), Qb represents the pump flow, Qa1 represents the Qa when Qb is set at Qb1, Qa2 represents the Qa at Qb2, v1 represents the time-averaged velocity (TAV) detected at transducer site 1 under Qb1, v2 represents the TAV detected at transducer site 1 under Qb2, and k is the constant related to the changed Qb and Qa0.
When equation (1) is divided by equation (2), we can obtain k as it is shown in equation (3).
If we move the probe at the site downstream from the arterial needle (transducer site 2), then the flow detected by the probe at this site will be [Qa − Qb]. The flow detected by the ultrasound probe at transducer site 2 will be [Qa1 − Qb1] under Qb1 and [Qa2 − Qb2] under Qb2. While the probe is fixed at the same detection angle and the same placing site during these measurements, the cross-sectional area of dialysis access (represented by A′) will be the same. We can obtain the following equations:
v3 represents the TAV detected at transducer site 2 at Qb1, and v4 represents the TAV detected at transducer site 2 at Qb2.
If equation (4) is divided by equation (5),
then we can obtain Qa0 by equation (6) as follows:
From equation (3), we have obtained that k=(v1 − v2)/(v1 × Qb2 − v2 × Qb1)
When we substitute “(v1 − v2)/(v1 × Qb2 − v2 × Qb1)” for “k” in equation (6), we can obtain equation (7) as follows:
As we assumed that the different Qa[Qa(n)] may be correlated with Qa and Qb in the above-mentioned equation (0): Qa(n)=Qa0 −k × Qb(n) × Qa0
When we substitute Qb1 for Qb (n) in equation (8), we can get equation (9):
When we substitute Qb2 for Qb (n) in equation (8), we can get equation (10):
Finally, Y.-C.S. wrote a computerized formula under the structure of the Excel program. After we keyed in the data of v1, v2, v3, v4, Qb1, and Qb2, we could obtain Qa1 and Qa2 right away.
Study Procedure for Access Flow Measurement
Protocol for the HD session when initial measurement of access flow is performed
Reversal of arterial and venous lines; for measurement of Qa by UDM twice according to the following sequence:
under Qb of 250 ml/min
under Qb of 150 ml/min
Normal position of arterial and venous lines:
Measurement of Qa by CDUM twice according to the following sequence:
at transducer site 1 under Qb of 250 ml/min (Qa1=v1 × cross-sectional area)
at transducer site 1 under Qb of 150 ml/min (Qa2=v2 × cross-sectional area)
Measurement of Qa by VPFDUM twice according to the following sequence:
using the same v1 obtained by CDUM at transducer site 1 under Qb of 250 ml/min
using the same v2 obtained by CDUM at transducer site 1 under Qb of 150 ml/min
obtaining v3 at transducer site 2 under Qb of 250 ml/min
obtaining v4 at transducer site 2 under Qb of 150 ml/min
Protocol for measurement of access flow in the two subsequent HD sessions: Measurement of Qa by VPFDUM once for each dialysis session according to the above-mentioned sequence. Every measurement of Qa was performed within the first 90 min after the initiation of each HD treatment.
Statistical Analyses
Intermeasurement variation for each patient was expressed as a percentage of the coefficient of variation (CV). Percentage of the CV is defined as the SD in Qa divided by mean Qa multiplied by 100 for each patient. Then the mean of the CV of each patient could be calculated. The correlation between repeated measurements or those by different methods was sought using linear regression analysis. The correlation coefficient of this regression line was measured to give an index of measurement reliability. Measurements with r 2 > 0.9 were considered reliable. Intraclass correlation coefficient (r i ) was used to test (1 ) the reproducibility of Qa measurements by one of the three above methods and (2 ) the agreement of Qa measurements between different methods. Unassisted survival time of vascular access (which was defined as the time interval from the date of measurement of Qa to the date of thrombosis) was determined by product-limit life-table distributions and compared using a log-rank test. All tests of significance were two-sided, and differences were considered statistically significant at P < 0.05. SPSS version 11.0 was used for all analyses.
Current guidelines for the combined graphical/statistical interpretation of method-comparison studies include a Bland-Altman plot combined with calculation of the 2 SD limits of the differences (the so-called 95% limits of agreement) between these methods (18 ). The x axis shows the mean of the results of the two methods, whereas the y axis represents the absolute difference between the two methods.
Results
Patient Characteristics
We initially tried to measure Qa in 73 maintenance HD patients by the above three methods. VPFDUM and CDUM can be completed smoothly in all 73 patients. However, UDM failed in 18 patients because blood could not be withdrawn smoothly with reversed lines even under Qb as low as 150 ml/min. Consequently, all three methods could be performed successfully in only 55 patients. Distribution of the clinical characteristics and causes of renal disease in each group is listed in Table 1 .
Measurements of Qa by VPFDUM versus UDM or CDUM
The result of Qa measurement by three different methods is shown in Table 2 . When we performed repeated measurements of Qa by each one of the three methods, the mean value of Qa by VPFDUM (870.8 ± 412.0 ml/min) was close to that by UDM (868.6 ± 417.9 ml/min) but higher than that by CDUM (either of the above values versus 685.1 ± 303.6 ml/min; P < 0.005). The mean values of the CV were similar by VPFDUM (1.6%) and UDM (1.4%) but lower than that by CDUM (either of the above values versus 6.8%; P < 0.01). The correlation coefficient and intraclass correlation coefficient of the repeated Qa measurements by VPFDUM (0.985 and 0.993; P < 0.001) were also similar to that by UDM (0.992 and 0.995; P < 0.001) but slightly higher than that by CDUM (0.917 and 0.948; P < 0.005). The correlation between repeated measurements of Qa by VPFDUM, with a very good reproducibility (r =0.98, second measurement of Qa=1.02 × first measurement of Qa − 9.02; P < 0.0001) is shown in Figure 2 . We also measured the Qa in 16 of these patients by VPFDUM during the two subsequent HD sessions. The intraclass correlation coefficient was 0.997 (95% confidence interval from 0.994 to 0.999). This result indicated that the reproducibility and the reliability of this technique were very good.
In comparison with the correlation of Qa measurements between different methods (Table 2 ), higher regression coefficient (0.93 versus 0.63) and higher intraclass correlation coefficient (0.994 versus 0.856) were noted between VPFDUM and UDM than that between CDUM and UDM. The intraclass correlation coefficient not only between VPFDUM and UDM but also between CDUM and UDM was higher for patients with Qa <1200 ml/min than that for those with Qa >1200 ml/min (Table 2 ). The relationship between measurements of Qa by VPFDUM and UDM is shown in Figure 3 , with a good correlation (r =0.99, measurement of Qa by VPFDUM=0.933 × measurement of Qa by UDM + 55.2; P < 0.0001).
As illustrated in Figure 4 , the Bland-Altman plot reveals the differences in measurements of UDM and VPFDUM versus mean absolute values of these measurements. The mean difference (−3.5 ml/min) did not depend on the absolute value of Qa for most Qa measurements.
Access Survival
We evaluated the effect of Qa measured by VPFDUM on vascular access survival in all of the initially included 73 HD patients in this study. The unassisted patency of vascular access at 6 mo was significantly poorer in patients with Qa <500 ml/min than those with Qa >500 ml/min (13.6% [3 of 22] versus 92.2% [47 of 51]; P < 0.0001; Figure 5 ).
Discussion
Access flow is an important parameter for predicting the prognosis of vascular access. According to Kidney Disease Outcomes Quality Initiative guidelines for vascular access, HD patients with Qa <600 ml/min or Qa <1000 ml/min that has decreased by >25% over 4 mo should be referred for fistulogram (the same threshold for both AVF and graft) (19 ). Canadian clinical practice guidelines recommend performing angiography when Qa is <500 ml/min in native arteriovenous fistulas (AVF), which was supported by the study result of Tonelli et al. (20 ). UDM, invented by Krivitski et al. (10 , 21 ), has been proved to be an accurate and reliable procedure in Qa measurement, but this technique may fail in some patients. The following clinical conditions may limit the application of UDM in the measurement of Qa in HD patients: (1 ) accessory veins may exist between two needles; (2 ) the length of AVF in some patients may be too short to allow puncture of two needles at the same side of the elbow joint; and (3) some patients have malfunctioning or fragile native AVF, so their blood cannot be withdrawn, even under the setting of Qb <200 ml/min with reversed lines. This method also carries some disadvantages: (1 ) reversing lines will increase the workload of nurses in a busy HD center, and (2 ) infusion of saline may exacerbate the difficulty in attaining ultrafiltration goal in patients with excessive interdialytic body weight gain or preexistent fluid overload. Hence, we would like to develop a new method for measurement of access blood flow without the above-mentioned shortcomings.
Doppler ultrasound method has been used widely to measure Qa; however, our study showed that the accuracy and the reproducibility of CDUM were not as good as that for VPFDUM. This difference may derive from the following reasons. First, the Doppler signal beam angle is an important source of error in flow measurement; accurate angle is necessary for determination of the velocity required for duplex volume flow calculation by CDUM. However, the calculation of Qa will be insensitive to this error by VPFDUM if the Doppler beam angle remains fixed during measurement. Second, a large error in Qa calculations may result from a small change in cross-sectional diameter measurement because area varies with the square of the diameter during study by CDUM. Because VPFDUM does not require calculation of the cross-sectional area, this potential source of error can be avoided by the new technique. We also tried to avoid the influence of turbulence by the following 2 steps: (1 ) the sampling gate was set at the central part of the vascular lumen to avoid the multidirectional turbulence near the edges of the vessel wall, and (2 ) the transducer was placed at the site at least 2 cm away from the anastomotic site of the AVF or the stenotic lesion to avoid the turbulence caused by considerable variation of pressure gradient or flow separation zone as a result of different shear stresses.
Although VPFDUM has some advantages over UDM and CDUM and is without their disadvantages, the equations of this study seemed to be more complicated. Originally, we assumed that Qa would not change significantly under different settings of Qb. However, we found that the data of Qa (measured by UDM) would change significantly under different settings of Qb. In 85.5% of these patients (47 of 55), Qa would be increased if Qb were decreased from 250 to 150 ml/min. The mean value of the incremental Qa was 105.3 ml/min (95% confidence interval 77.8 to 132.7 ml/min; P < 0.001), and the increased percentage was 12.1%. Reversely, Qa would be decreased if Qb were increased from 150 to 250 ml/min. Seeing that different Qb will change Qa, we have to take this relationship into consideration. Therefore, we placed the ultrasound probe on transducer site 1 (Figure 1 ) to detect the possible change in Qa, the degree of which might be influenced by two factors: the static access flow (Qa0) and the Qb. The final result of good correlation between Qa measurements by VPFDUM and UDM (r 2 =0.98, P < 0.0001) also supported that our initial assumption “Qa(n)=Qa − k × Qb(n) × Qa” should be reasonable.
In 1984, van Gemert et al. (22 ) proposed a two-needle extracorporeal dialysis circuit model to demonstrate the possible relationship between Qb and other hemodynamic parameters. He concluded that Qb has no instantaneous influence on vascular hemodynamics if Qb is less than Qa. At higher Qa, the BP decreased and the increase in cardiac output was lower than the increase in Qa. However, he did not mention the relationship between Qb and Qa. Further studies should be focused on the relationship between the alteration of some hemodynamic parameters, such as cardiac output, BP, and total peripheral resistance.
In addition to having a good correlation with the currently available methods of flow measurement, a good method should predict access dysfunction reliably and early. For this reason, we evaluated the effect of Qa measurement by VPFDUM on the unassisted patency of vascular access. Because the interventional criteria for AVF was Qa <500 ml/min according to Canadian clinical practice guidelines and Tonelli et al. (20 ), we also divided our patients into two groups at the value of Qa of 500 ml/min. Patients with lower Qa had significantly poorer patency of vascular access at 6 mo than the higher group did. Therefore, our method can predict effectively the prognosis of vascular access in advance.
In conclusion, our study proved that this novel technique, VPFDUM, not only is as good as ultrasound dilution method but also has several additional advantages, being an accurate, noninvasive, and reliable procedure for measuring the vascular access flow and predicting its prognosis in HD patients.
Figure 1: The schematic of a dialysis vascular access with flow direction shown by arrows and the probe placement shown by the box. Qa, vascular access flow; Qb, pump flow. Qa is the flow detected at transducer site 1; (Qa − Qb) is the flow rate detected at transducer site 2.
Figure 2: Reproducibility of measurements of Qa by VPFDUM (n =55). Correlation line is shown (r =0.985; P < 0.0001).
Figure 3: Correlation of the measurements of Qa by variable pump flow–based Doppler ultrasound method (VPFDUM) and ultrasound dilution method (UDM) with regression line. Correlation coefficient (r ) is 0.99 for all comparison measurements (n =55).
Figure 4: Bland-Altman plot of differences in measurements of UDM and VPFDUM (y axis) versus mean absolute values of these measurements (x axis). The mean difference did not depend on the absolute value of flow volume for most Qa measurements.
Figure 5: Unassisted patency of arteriovenous fistula comparing patients with Qa >500 ml/min with those with Qa <500 ml/min; P < 0.0001 for the comparison between the two survival curves. Numbers below the figure represent patients at risk at each time point in the two groups.
Table 1: Patient characteristics and causes of renal diseasea
Table 2: Statistical analysis of the data of Qa measured by three different methodsa
Acknowledgments
This work was supported by a grant (2004-A042) from Taipei Veterans General Hospital in Taiwan.
This abstract was accepted as a poster presentation at the XLI ERA-EDTA Congress; Lisbon, Portugal; May 15–18, 2004.
Published online ahead of print. Publication date available at www.jasn.org .
References
1. Feldman HI, Kobrin S, Wasserstein A: Hemodialysis vascular access morbidity. J Am Soc Nephrol 7: 523–535, 1996
2. Smits JH, van der LJ, Hagen EC, Modderkolk-Cammeraat EC, Feith GW, Koomans HA, van den Dorpel MA, Blankestijn PJ: Graft surveillance: Venous pressure, access flow, or the combination? Kidney Int 59: 1551–1558, 2001
3. Paun M, Beach K, Ahmad S, Hickman R, Meissner M, Plett C, Strandness DE Jr: New ultrasound approaches to dialysis access monitoring. Am J Kidney Dis 35: 477–481, 2000
4. Miranda C, Sands J: Flow volume as a predictor of hemodialysis access failure. J Vasc Technol 22: 73–76, 1998
5. Gothin J, Lindstedt E, Olin T: A dye-dilution method for the determination of blood flow in cimino-brescia arteriovenous fistulae. Invest Urol 15: 167–168, 1977
6. O’Regan S, Lemaitre P, Kaye M: Hemodynamic studies in patients with expanded polytetrafluoroethylene (PTFE) forearm grafts. Clin Nephrol 10: 96–100, 1978
7. Koksoy C, Kuzu A, Erden I, Turkcapar AG, Duzgun I, Anadol E: Predictive value of colour Doppler ultrasonography in detecting failure of vascular access grafts. Br J Surg 82: 50–52, 1995
8. Anderson CB, Etheredge EE, Harter HR, Codd JE, Graff RJ, Newton WT: Blood flow measurements in arteriovenous dialysis fistulas. Surgery 81: 459–461, 1977
9. Oudenhoven LF, Pattynama PM, de Roos A, Seeverens HJ, Rebergen SA, Chang PC: Magnetic resonance, a new method for measuring blood flow in hemodialysis fistulae. Kidney Int 45: 884–889, 1994
10. Krivitski NM: Theory and validation of access flow measurement by dilution technique during hemodialysis. Kidney Int 48: 244–250, 1995
11. Steuer RR, Miller DR, Zhang S, Bell DA, Leypoldt JK: Noninvasive transcutaneous determination of access blood flow rate. Kidney Int 60: 284–291, 2001
12. Gotch FA, Buyaki R, Panlilio F, Folden T: Measurement of blood access flow rate during hemodialysis from conductivity dialysance. ASAIO J 45: 139–146, 1999
13. Schneditz D, Wang E, Levin NW: Validation of haemodialysis recirculation and access blood flow measured by thermodilution. Nephrol Dial Transplant 14: 376–383, 1999
14. Yarar D, Cheung AK, Sakiewicz P, Lindsay RM, Paganini EP, Steuer RR, Leypoldt JK: Ultrafiltration method for measuring vascular access flow rates during hemodialysis. Kidney Int 56: 1129–1135, 1999
15. Weitzel WF, Rubin JM, Swartz RD, Woltmann DJ, Messana JM: Variable flow Doppler for hemodialysis access evaluation: Theory and clinical feasibility. ASAIO J 46: 65–69, 2000
16. Weitzel WF, Rubin JM, Leavey SF, Swartz RD, Dhingra RK, Messana JM: Analysis of variable flow Doppler hemodialysis access flow measurements and comparison with ultrasound dilution. Am J Kidney Dis 38: 935–940, 2001
17. Zierler BK, Kirkman TR, Kraiss LW, Reiss WG, Horn JR, Bauer LA, Clowes AW, Kohler TR: Accuracy of duplex scanning for measurement of arterial volume flow. J Vasc Surg 16: 520–526, 1992
18. Dewitte K, Fierens C, Stockl D, Thienpont LM: Application of the Bland-Altman plot for interpretation of method-comparison studies: A critical investigation of its practice. Clin Chem 48: 799–801, 2002
19. National Kidney Foundation: Kidney Disease Outcomes Quality Initiative (NKF K/DOQI) clinical practice guidelines for vascular access. Am J Kidney Dis 37: S137–S181, 2001
20. Tonelli M, Jhangri GS, Hirsch DJ, Marryatt J, Mossop P, Wile C, Jindal KK: Best threshold for diagnosis of stenosis or thrombosis within six months of access flow measurement in arteriovenous fistulae. J Am Soc Nephrol 14: 3264–3269, 2003
21. Depner TA, Krivitski NM: Clinical measurement of blood flow in hemodialysis access fistulae and grafts by ultrasound dilution. ASAIO J 41: M745–M749, 1995
22. Van Gemert MJ, Bruyninckx CM, Baggen MJ: Shunt haemodynamics and extracorporeal dialysis: An electrical resistance network analysis. Phys Med Biol 29: 219–235, 1984