Potential Deviation Factors Affecting Stereolithographic Surgical Guides

A Systematic Review

Sigcho López, Diego Antonio, MSc*; García, Iván, DDS; Da Silva Salomao, Gustavo, DDS; Cruz Laganá, Dalva, PhD§

doi: 10.1097/ID.0000000000000853
Meta-Analyses and Systematic Reviews
Free
SDC

Objective: The aim of this study was to review potential deviation factors in stereolithographic surgical guides for dental implantology, warnings, and limitations of the system.

Methods: An electronic search in databases EMBASE, the Cochrane Library, and PubMed were conducted by 3 researchers to collect information on the accuracy of static computer-guided implant placement to summarize and analyze the overall accuracy. The latter included a search for correlations between factors such as support (teeth/mucosa/bone), number of templates, use of fixation pins, jaw, template production, guiding system, and guided implant placement in articles related to guided surgery with stereolithographic static systems. Studies published between 2012 and 2017 were reviewed.

Results: From 761 identified articles, a total of 24 articles were reviewed, which included 2767 dental implants. Data from studies analysis had shown a mean deviation of 3.08 degrees in angular position, 1.14 at the entry point, and 1.46 at apex position. Involved deviation factors were related to planning, laboratory, and surgical phases.

Conclusion: Guided surgery may have a limited precision as technique, which surgeons need to be aware in the planning process. This review suggests some security measures in guided surgery process.

*PhD Student, University of São Paulo, São Paulo, Brazil.

PhD Student, Faculty of Dentistry, University of São Paulo, São Paulo, Brazil; Professor, Faculty of Dentistry, University of Ecuador, Ecuador, Brazil.

MSc Student, Faculty of Dentistry, University of São Paulo, São Paulo, Brazil.

§Professor, Faculty of Dentistry, University of São Paulo, São Paulo, Brazil.

Reprint requests and correspondence to: Diego Antonio Sigcho López, MSc, Central University of Ecuador, El Comercio E7-49 y el Tiempo, Quito, Ecuador, 170506 Phone: (593)984263660, Fax: +59322525238, E-mail: diegosigcho@usp.br

Currently, the application of computer-aided design and coordinated computer-aided manufacturing in implantology allows the virtual planning of positioning the dental implants, key in the longevity of the prosthesis, related to anatomical and prosthetic considerations.1

Surgical guide systems can be divided into static or dynamic. Static surgical guides are made in the laboratory through 3-dimensional printing (prototyping) known as stereolithographic (SLA) guides or perforations in templates made on the jaw models. They are called static because they do not allow modification of the planned position during the surgical procedure. Dynamic guides use a mechanical or optical system to transfer the virtual planning to the surgical field, displaying the process on a screen in real time.

Stereolithographic guides can be single or multiple, the multitype guides are generally used with no stabilization screws, changing the guide as the drill diameter increases and have showed similar results than single type guides. Single guides use physical stops for different diameter of drills.2

Several deviations have been described in literature and therefore the aim of this study was to determine what factors are involved in accuracy of SLA-guided surgery and how to prevent them.

Back to Top | Article Outline

Methods

This review was registered at the National Institute for Health Research PROSPERO, an international prospective register if systematic reviews,3 registration number CRD42017067806. To identify relevant studies, an electronic search was carried out according to PRISMA guidelines4 within PubMed, Cochrane, and Embase database of articles from 2012 to 2017. If a full text was not available, the electronically available data in the abstract were collected.

Back to Top | Article Outline

Eligibility Criteria

This review focus on total and partial edentulous patients, who were treated with endosseous titanium implants by computer-assisted stereolithographic surgical guides independent of implant number, length, diameter, position, or angulation, into healthy bone, either screw-retained or cement-retained prosthodontic rehabilitation, according to an immediate, early, or conventional loading protocol.

For inclusion in this study, articles reporting on accuracy, implant and prosthetic failures, biological or mechanical complications were selected.

Back to Top | Article Outline

Focus question

What factors are involved in the accuracy of stereolithographic guided surgery?

Back to Top | Article Outline

Search Strategy

The following search terms were used: “surgery, computer-assisted” [MeSH Terms] OR (“surgery” [All Fields] AND “computer-assisted”[All Fields]) OR “computer-assisted surgery” [All Fields] OR (“computer” [All Fields] AND “assisted”[All Fields] AND “surgery”[All Fields]) OR “computer assisted surgery” [All Fields] AND “dental implants” [MeSH Terms] OR (“dental” [All Fields] AND “implants” [All Fields]) OR “dental implants” [All Fields] OR (“dental” [All Fields] AND “implant” [All Fields]) OR “dental implant” [All Fields]) AND (“2007/05/23” [PDat]).

Back to Top | Article Outline

Studies Selection and Data Extraction

Inclusion criteria for articles were as follows: clinical studies related to guided surgery with stereolithographic static systems in patients; prospective, retrospective, randomized clinical trials, cohort, transversal and longitudinal studies about accuracy; clinical studies including at least 5 human subjects, which outcomes on angular, entry point, or apical position were specified.

Exclusion criteria were as follows: studies about zygomatic implants, postextraction sockets, short implants, bone grafts, sinus lift, clinical cases, animals studies, in vitro, laboratory studies, reviews of literature, craniomaxillofacial reconstructions, oncological, cleft lip-palatal, syndromes or disease patients related studies.

Two reviewers analyzed the studies independently and discussed with a third reviewer for final selection.

Back to Top | Article Outline

Risk Analysis of Bias

The quality analysis tool for the studies was QUADAS 2 (Quality assessment tool for diagnostic accuracy studies) from Bristol University, which evaluated risk of bias and applicability for patient selection, index test, reference standard, flow period, and timing.

Back to Top | Article Outline

Analysis Risk of Bias

As we can see in Figure 1, positive results were found in relation to patients selection; this could be due to most of authors agree in the basic patients conditions for guided surgery treatment. But quite opposite findings were observed in relation to index tests, as conditions and measurements differ from the studies, which avoid a meta-analysis. In general, the risk of bias was indicated as medium level.

Fig. 1

Fig. 1

Back to Top | Article Outline

Results

Flow Diagram

The initial electronic database search yielded 761 titles which were independently screened resulting in the consideration of 321 publications. Abstracts were then reviewed and the remaining 51 studies were analyzed in detail, resulting in the exclusion of 25 articles (Fig. 2). Three reviewers agreed on the classification of 24 of the studies, with an estimated kappa test of 0.75 (Table 1). Most of protocols used retention pins in the guide and flapless technique. Dental, mucosal, and bone supported guides were included, as well as multiguides without retention pins. The variability of protocols and methods used in the studies did not allow us to perform a meta-analysis; therefore, we use all data as a narrative review.

Fig. 2

Fig. 2

Table 1

Table 1

Back to Top | Article Outline

Studies' Characteristics

The implant position on the dental arch, the use of fixation pins, and the type of surgery (with local or general anesthesia) showed no statistical significant difference for any variable.5

Cassetta et al6 found that guide fixation (fixed vs movable), supporting arch (maxilla vs mandible), and bone density also influenced intrinsic errors, during an assessment of the clinical relevance of potential mechanical errors in stereolithographic surgical templates.

Back to Top | Article Outline

Cautions

Vercruyssen et al7 have described some precautions as follows: make sure optimal fit of denture; in case of total edentulous patients, the denture can be relined using a soft reline, for this matter. Register the position of the denture in relation to the bone according to double scan procedure, with radiopaque markers of at least of 2 mm diameter.

Check the scan for movement artifacts and absence of space/air between denture and underlying soft tissues to be certain of proper fit. Cone beam computed tomography scan setting of 120 kVp, pulses of 3 to 8 mA, 8 cm scan height, and exposure time of 20 seconds reconstructed with an 0.3-mm isotropic voxel size was recommended. Regarding the radiographic techniques used, no significant differences were found.5

Inadequate mouth opening and high tension in the lips or cheek can make treatment more difficult and reduces the accuracy of the implant position.8 Especially in the canine tooth region, during drilling, the angled drill head had to be placed toward the cheeks, with the drill tip angulated toward the palate.5 Inaccuracies are mainly explained due to gaps between the drill, drill guide, and sleeve in the surgical template.9

Back to Top | Article Outline

Potential Deviation Factors

Presurgical planning

Not all patients are candidates for guided surgery without flap detachment. It is important during planning to determine if the surgical guide will be bone, mucosal, or dental supported. The mouth opening should be evaluated as well as the length of the drills. In general, when using the surgical guide, a minimum of 35 mm of oral space is required for the surgeon to position the drills correctly in the holes of the guide: 20 mm are necessary for drill bits, 10 mm of distance between the guidewire and the cervical end of the implant, and the 5 mm thickness of the contra-angle head.10 Based on the amount of available bone and considering the possibility of fenestration,11 it is recommended a safety distance of 2 mm from anatomical structures.12

The patient's movements during scanning may affect the deviation in implant angulation. Currently, the new tomographers have reduced scanning times and these movement errors.13 Surgical planning may change in case of complications as template fractures during the surgery, limited primary implant stability, need for additional grafting procedures, and prosthetic misfit.14

To avoid errors related to mucosal thickness, it is recommended to use a stable pressure during bite registration, during scanning, and in surgery.13 The amount of soft tissue and its mobility may hinder the accuracy of the guide fixation.15 The increase of 1 mm of mucosa increases buccolingual deviation by 0.4 mm.16

The observed deviations in the posterior region have been related to the limitation of mouth opening.17 In this case, it may be difficult to drill long implants and the insertion of the drill bits. The length of the implant and the depth into the bone perforation has been associated to deviations.18

Precision increases when the surgical guides are fixed with screws tightly in the jaws. At least 3 fixation screws are recommended in the anterior and 2 posterior regions.18 In case of not been possible to fix the guide with screws, it is suggested to use multiple guides avoiding micro-movements.18 Kalt and Gehrke19 report higher precision when using double guide sleeve during drilling with the first drill, the lance. Long, wide, and strong metal sleeves would give greater stability to guide.

The inappropriate use of the guides and mismatch of the drill with the guide cylinder negatively influences the precision of the implants. In addition, the prosthesis must be of sufficient thickness to allow the segmentation of the tomographic images. If it is too thin, it does not differentiate the prosthesis edges from the air spaces.6

Back to Top | Article Outline

Surgical

Confirm that the guide is well positioned and stable in the jaw. Look for areas of tension or possible interference in the adaptation of the guide. Difficulties may arise in positioning the surgical guide.20 After anesthetic infiltration, wait around 10 minutes and massage the infiltrated area to prevent edema from interfering with the adaptation of the guide. During the infiltration of anesthesia, the gingiva undergoes local volume increase.21

Be aware of micro-movements that may arise during drilling, especially in cases of free end and total edentulous. It is important to fix and support the guide, and it is even possible to fix the guide with temporary implants.22 Keep the drill parallel and centered relative to the guide cylinder.23

Torques larger than 50 N were related to guide fracture.20 However, avoid excessive pressure, in this case, remove the guide for the last millimeters of insertion of the implant. The fracture of the surgical guide can be avoided by increasing the space between the teeth or implants and the guide cylinder, maintaining the required thickness of the material.6

Even when all perforations are made with a surgical guide, the deviation can occur when the implant is inserted manually at the end of the procedure.24 The depth of the implant placed with guided surgery can also vary from the planning, depending on the system employed.22

The vertical dimension depends on the level of the bone crest, but in aesthetic cases, it is advisable to detach the flap for visual control.22 Walton et al25 have suggested that professional training and experience influences the accuracy of guided surgery.

Patient movements during surgery, limited visualization of the surgical site,7 and overfitting the implant during insertion may also change the position of the guide.21

Guided surgery can increase bone temperature during drilling because the surgical guide does not allow direct irrigation through it compared to the conventional procedure.26

Among the surgical complications described, the most common is the lack of primary stability, related to low torque in type IV bones and implants placed in the maxilla.26 Difficulty in positioning the surgical guide19 and fenestration of the bone10 and surgical planning change around 2%.13

Back to Top | Article Outline

Discussion

Until 2011, limited information about accuracy of implant placement of the maxilla of fully edentulous patients was given.16 This study includes articles from 2012 to 2017, which reduces the risk of bias obtaining more homogenic data.

Vercruyssen et al27 evaluating computer-assisted surgery in totally edentulous patients with surgical guides supported on mucosal or bone, found a mean vertical deviation of 0.9 ± 0.8 mm (range: 0.0–3.7 mm) and 0.9 ± 0.6 mm (range: 0.0–2.9 mm) in a horizontal direction. The overall mean deviation for the guided surgery in mesio-distal direction was 0.6 ± 0.5 mm (range: 0.0–2.5 mm) and 0.5 ± 0.5 mm (range: 0.0–2.9 mm) in buccolingual direction. These results are within the values found in this study, which are less than reported by Tahmaseb et al12 in a meta-analysis of the accuracy of 1.12 mm (maximum of 4.5 mm) at the entry point measured in 1530 implants and 1.39 mm at the apex (maximum of 7.1 mm), 3.5 degrees (95% CI 3.0–4.1), with a maximum of 21.2; similar values were found in meta-analysis of Van Assche et al28 Pettersson et al12 investigated 25 fully edentulous patients in 15 patients receiving implant in their upper jaw, 5 or 6 implants were placed, showing a mean 3-dimensional deviation for “implant tip” of 1.05 mm (range: 0.25–2.63 mm), a mean deviation for “shoulder” of 0.80 mm (range: 0.10–2.68 mm), a mean “angular deviation” of 2.31 degrees (range: 0.24–6.96 degrees), and a mean “depth deviation” of −0.06 mm (range: −1.65 to 2.05 mm). Higher deviations were seen at the implant apex, as major control of the drill guide is at the entry point, leaving a relative small gap for the tip movement.

Due to improvement in technology and techniques, we found less mean deviations in a review of publications from 2012 to 2017 comparing to publications from 1996 to 2011 as reviewed by Van Assche et al.28

Fewer deviations were observed in cases of unitary implants, with mesial and distal dental support, a mean of 0.21 ± 0.16 mm at the cervical level, 0.32 ± 0.34 mm at the apex, and 1.35 ± 1.11 degrees. The deviation is greater when the support is unilateral or in free end.29

The posterior edentulous spaces have a negative influence on guided surgery, due to the support of the guide at one end only. However, an important factor in the accuracy of guided surgery is the fixation and support of the guide. One solution is to fix the guide with temporal implants.22

The depth of the implant placed with guided surgery can also vary from the planning, depending on the system employed. Naziri et al22 investigated different systems of guided surgery in partial edentulous patients, placing a total of 246 implants, of which 2 implants were 3 mm under planning. Comparing guided surgery systems showed that the accuracy of implant placement decreases with increasing implant length. In this same study, the greatest deviation observed in implant placement was 16.6 degrees. This considerable deviation was due to the first 2 perforations made with the surgical guide and the last without it, free hand.

Implant angulation can be affected by the difficulty of drilling, especially in long implants, by positioning long drills in the limited opening of the mouth. This is evidenced by the fact that significant deviations related to implant length in the distal mesio sense have been observed. Implants smaller than 11 mm were recommended for guided surgery and with minimum perforations possible.22

The friction of the drill with the bone and the surgical guide may interfere with the accuracy of the guide. The use of titanium sleeves to direct the drill during drilling increases guide deflections. Thus, it can also influence bone strength in drill orientation during drilling.14

Laederach et al29 observed significant differences when comparing different systems of stereolithographic guides, suggesting that the system and the protocol used influence the accuracy of the implants. The shape of the tip of the drill facilitates the sliding during the first millimeters of the drilling, acquiring stability with the sinking and greater contact with the bone. The use of plexiglas increases the error due to material resistance and the use of parallel cylinders is better than tapered cylinders.

Regardless of the tomographic equipment, the scanner protocol is more important than the device itself. To improve accuracy, a high-definition protocol is required.30

Vercruyssen et al31 reported deviations from the surgical guide in guided surgery of 1.4 mm with a range of 0.3 to 3.7 at cervical level, at the apical level was 1.6 mm (0.2–3.7), and in inclinations was 3 degrees (0.2–16 degrees). Similar results were found in this review (Table 2); however, it is possible that even with the technological developments and the use of more precise equipment, the stereolithography technique has reached a limit of association of factors specific to the technique.

Table 2

Table 2

Also, D'Haese et al1 investigated the accuracy of implant placement in the maxilla of 13 fully edentulous patients. A total of 78 implants were placed using mucosa-supported surgical templates. Accuracy evaluation showed a mean tip deviation of 1.13 mm (range: 0.32–3.01 mm), a mean shoulder deviation of 0.91 mm (range: 0.29–2.45 mm), and a mean angular deviation of 2.60 degrees (range: 0.16–8.86 degrees).

Back to Top | Article Outline

Conclusion

The increased accuracy and benefits of implants drilling that guided surgery represents for the professional and patient is indisputable. However, surgical guides are not yet free of risks and complications; even with the improvement of techniques and instruments, the deviations of the implants remain a challenge for the surgeon. It is important to keep in mind the factors related to deviations and risks to take preventive measures.

Back to Top | Article Outline

Roles/Contributions by Authors

D. A. S. López: Principal researcher, corresponding author. I. García: Research collaborator. G. Da S. Salomao: Research collaborator. D. C. Laganá: Supervisor, research collaborator, reviewer, and editor.

Back to Top | Article Outline

Disclosure

The authors claim to have no financial interest, either directly or indirectly, in the products or information listed in the article.

Back to Top | Article Outline

References

1. D'Haese J, Van De Velde T, Elaut L, et al. A prospective study on the accuracy of mucosally supported stereolithographic surgical guides in fully edentulous maxillae. Clin Implant Dent Relat Res. 2012;14:293–303.
2. Cassetta M, Di Mambro A, Giansanti M, et al. The intrinsic error of a stereolithographic surgical template in implant guided surgery. Int J Oral Maxillofac Surg. 2013;42:264–275.
3. Chien PF, Khan KS, Siassakos D. Registration of systematic review: PROSPERO. BJOG. 2012;119:903–905.
4. Moher D, Liberati A, Tetzlaff J, et al; The PRISMA Group. Preferred reporting items for systematic reviews and meta-analyses: The PRISMA statement. Ann Intern Med. 2009;151:264–269.
5. Verhamme LM, Meijer GJ, Boumans T, et al. A clinically relevant accuracy study of computer-planned implant placement in the edentulous maxilla using mucosa-supported surgical templates. Clin Implant Dent Relat Res. 2015;17:343–352.
6. Cassetta M, Di Mambro A, Giansanti M, et al. How does an error in positioning the template affect the accuracy of implants inserted using a single fixed mucosa-supported stereolithographic surgical guide? Int J Oral Maxillofac Surg. 2014;43:85–92.
7. Vercruyssen M, Coucke W, Naert I, et al. Depth and lateral deviations in guided implant surgery: An RCT comparing guided surgery with mental navigation or the use of a pilot-drill template. Clin Oral Implants Res. 2015;26:1315–1320.
8. Arataki T, Furuya Y, Ito T, et al. Pre-operative drilling simulation method for dental implant treatment. Bull Tokyo Dent Coll. 2007;48:27–35.
9. Verhamme LM, Meijer GJ, Boumans T, et al. A clinically relevant validation method for implant placement after virtual planning. Clin Oral Implants Res. 2013;24:1265–1272.
10. Sun Y, Luebbers HT, Agbaje JO, et al. Accuracy of dental implant placement using CBCT-derived mucosa-supported stereolithographic template. Clin Implant Dent Relat Res. 2015;17:862–870.
11. Pomares C, Pomares Puig C. A retrospective study of edentulous patients rehabilitated according to the “all-on-four” or the “all-on-six” immediate function concept using flapless computer-guided implant surgery. Eur J Oral Implantol [internet]. 2010;3:155–163.
12. Beretta M, Poli PP, Maiorana C. Accuracy of computer-aided template-guided oral implant placement: A prospective clinical study. J Periodontal Implant Sci. 2014;44:184–193.
13. Pettersson A, Komiyama A, Hultin M, et al. Accuracy of virtually planned and template guided implant surgery on edentate patients. Clin Implant Dent Relat Res. 2012;14:527–537.
14. Tahmaseb A, Wismeijer D, Coucke W, et al. Computer technology applications in surgical implant dentistry: A systematic review. Int J Oral Maxillofac Implants. 2014;29(suppl):25–42.
15. Horwitz J, Zuabi O, Machtei EE. Accuracy of a computerized tomography-guided template-assisted implant placement system: An in vitro study. Clin Oral Implants Res. 2009;20:1156–1162.
16. Vasak C, Watzak G, Gahleitner A, et al. Computed tomography-based evaluation of template (NobelGuide)-guided implant positions: A prospective radiological study. Clin Oral Implants Res. 2011;22:1157–1163.
17. Valente F, Schiroli G, Sbrenna A. Accuracy of computer-aided oral implant surgery: A clinical and radiographic study. Int J Oral Maxillofac Implants. 2009;24:234–242.
18. Cassetta M, Di Mambro A, Giansanti M, et al. Is it possible to improve the accuracy of implants inserted with a stereolithographic surgical guide by reducing the tolerance between mechanical components? Int J Oral Maxillofac Surg. 2013;42:887–890.
19. Kalt G, Gehrke P. Transfer precision of three-dimensional implant planning with CT assisted offline navigation. Int J Comput Dent. 2008;11:213–225.
20. Gillot L, Noharet R, Cannas B. Guided surgery and presurgical prosthesis: Preliminary results of 33 fully edentulous maxillae treated in accordance with the NobelGuide protocol. Clin Implant Dent Relat Res. 2010;12(suppl 1):e104–13.
21. Balshi SF, Wolfinger GJ, Balshi TJ. Surgical planning and prosthesis construction using computed tomography, CAD/CAM technology, and the Internet for immediate loading of dental implants. J Esthet Restor Dent. 2006;18:312–323; discussion 324–325.
22. Naziri E, Schramm A, Wilde F. Accuracy of computer-assisted implant placement with insertion templates. GMS Interdiscip Plast Reconstr Surg DGPW. 2016;5:Doc15.
23. Koop R, Vercruyssen M, Vermeulen K, et al. Tolerance within the sleeve inserts of different surgical guides for guided implant surgery. Clin Oral Implants Res. 2013;24:630–634.
24. Park C, Raigrodski AJ, Rosen J, et al. Accuracy of implant placement using precision surgical guides with varying occlusogingival heights: An in vitro study. J Prosthet Dent. 2009;101:372–381.
25. Walton JN, Huizinga SC, Peck CC. Implant angulation: A measurement technique, implant overdenture maintenance, and the influence of surgical experience. Int J Prosthodont. 2001;14:523–530.
26. Di Giacomo GA, da Silva JV, da Silva AM, et al. Accuracy and complications of computer-designed selective laser sintering surgical guides for flapless dental implant placement and immediate definitive prosthesis installation. J Periodontol. 2012;83:410–419.
27. Vercruyssen M, Laleman I, Jacobs R, et al. Computer-supported implant planning and guided surgery: A narrative review. Clin Oral Implants Res. 2015;26(suppl 11):69–76.
28. Van Assche N, Vercruyssen M, Coucke W, et al. Accuracy of computer-aided implant placement. Clin Oral Implants Res. 2012;23(suppl 6):112–123.
29. Laederach V, Mukaddam K, Payer M, et al. Deviations of different systems for guided implant surgery. Clin Oral Implants Res. 2017;28:1147–1151.
30. Vercruyssen M, Cox C, Coucke W, et al. A randomized clinical trial comparing guided implant surgery (bone- or mucosa-supported) with mental navigation or the use of a pilot-drill template. J Clin Periodontol. 2014;41:717–723.
31. Vercruyssen M, Hultin M, Van Assche N, et al. Guided surgery: Accuracy and efficacy. Periodontol 2000. 2014;66:228–246.
Keywords:

surgery; computer-assisted; dental implants; ambulatory surgical procedures; tomography; x-ray computed

Copyright © 2019 Wolters Kluwer Health, Inc. All rights reserved.