Laser in situ keratomileusis (LASIK ) is one of the most commonly performed refractive surgeries in the world today as it reduces spectacle dependence and has a positive impact on an individual’s quality of life.[1 2 ] Although LASIK has proven to be safe, effective, and predictable, complications could still occur[3 ] and ectasia has been reported as a serious complication post-refractive surgery.[4 ] In the early stages, corneal ectasia can be managed with corneal collagen cross-linking, intrastromal corneal rings, and specialty contact lenses such as Rose K scleral lenses, however, the advanced cases require lamellar or partial thickness keratoplasty.[5 6 ] Thus, a better refractive screening strategy is warranted to prevent the occurrence of ectasia and the further implication of ectasia.
With refractive surgeries, alteration in corneal biomechanical parameters is unavoidable, hence, corneal biomechanical assessment has become a potential tool in screening post-refractive surgical ectasia. With the advancement in technology, the in vivo corneal biomechanical assessment in a clinical setting is possible. Previous studies have hypothesized that corneal biomechanical changes precede tomographical and topographical changes.[7 ] Thus, it is important to study corneal biomechanical parameters to predict ectasia. The recently designed Corneal Visualization Scheimpflug Technology (Corvis ST) (Oculus Optikgeräte GmbH, Wetzlar, Germany) has shown to have good repeatability and reproducibility in measuring dynamic corneal responses[8 ] and metrics of Corvis ST such as Corvis Biomechanical Index (CBI) and tomographical biomechanical index (TBI) have a better diagnostic ability to separate ectatic eye from the normal eye.[9 10 11 ] CBI has been developed for keratoconus screening and includes various dynamic corneal response parameters such as deformation amplitude ratio at 1 and 2 mm, applanation 1 velocity, the standard deviation of deformation amplitude at the highest concavity (HC), Ambrósio’s Relational Thickness to the horizontal profile, and stiffness parameter.[9 ] On the other hand, the TBI uses combined data from Pentacam HR and Corvis ST to screen ectatic corneal diseases.
The newly introduced metric such as corneal stress-strain index (SSI) describes the whole SSI curve of the corneal tissue. The SSI measures in vivo the material stiffness, and thus, helps facilitate the optimization of several corneal refractive treatments and management procedures, especially enhancement procedures. The SSI algorithm predicts the corneal behavior using the least square method and finite element models which simulate the effects of IOP and the Corvis ST air puff.[12 ] This algorithm has been validated previously in normal eyes,[12 ] however, it can be useful for clinical documentation of corneal biomechanical changes post-C3R.[13 ] In addition, it can also be used to screen the patients who are at high risk or are susceptible to developing/progression of post-refractive surgical ectasia such as LASIK . The previous studies have reported changes in the conventional corneal biomechanical parameters post-refractive surgeries.[14 ] However, no studies have reported post-refractive surgery changes in novel and more promising metrics of corneal biomechanics such as SSI. Hence, our aim is to report short-term changes in corneal SSI and other corneal biomechanical parameters post-LASIK surgeries (Microkeratome Blade and Femtosecond Bladeless LASIK ).
Methods
This was a retrospective study conducted at a tertiary eye care center wherein subjects who have undergone refractive surgery (LASIK ) between July 2019 and December 2020 were enrolled. The subjects were further divided into two groups depending on the technique used to create the flap.
Microkeratome Blade LASIK (Blade LASIK )
Femtosecond Bladeless LASIK (Femto bladeless LASIK ).
The study was reviewed and approved by the Institutional Ethics Committee (EC reg. details: ECR/1088/Inst/MH/2018; EC approval ref. no. 2020/01: 15th Nov 2020) and was conducted in accordance with the tenets of the Declaration of Helsinki. The medical records of these patients were reviewed and analyzed. Patients between the age group of 20 and 40 years, best-corrected visual acuity of 20/20, IOP <22 mmHg, pre-LASIK pachymetry >500 microns, corneal astigmatism ≤3.00 D were enrolled. Corneal astigmatism >3.0 D is uncommon, acquired, and generally associated with ocular comorbidity. Thus, in order to make the research participants’ group coherent, corneal astigmatism equal to or less than 3.00 D was included. Subjects with a previous history of refractive surgery, any ocular or systemic disease, corneal astigmatism >3.00 D, poor-quality scans, intraoperative complications, and any missing data were excluded.
Surgical procedure
Both Microkeratome Blade and Femtosecond Laser-Assisted Bladeless LASIK surgery were performed by a single surgeon (VK).
Blade LASIK
Blade LASIK was done using Nidek EC 5000 CXII using Optimized Aspheric Treatments (OATz). All flaps were created using the Sub-Bowman Keratomileusis (SBK) automated disposable microkeratome with a 90 mm head (Moria, Antony, France).
Femto Bladeless LASIK
Femto LASIK was done using the Ziemer classic (Ziemer Ophthalmic Systems). The flap was created with the Zeimer Femtosecond Laser (Ziemer Ophthalmic Systems), targeting a flap thickness of 90 microns.
Outcome variables
Demographics
Parameters such as age, gender, types of refractive surgery, and scan quality were noted. Also, refractive and ocular parameters such as Best corrected visual acuity (BCVA), refraction, IOP, pachymetry, and keratometry before LASIK were noted.
Corneal biomechanics parameter assessment
The corneal biomechanical properties were analyzed using Corvis ST (OCULUS Optikgeräte GmbH; Wetzlar, Germany). Corvis® ST records the reaction of the cornea to a defined air pulse using a high-speed Scheimpflug camera. The camera takes over 4,300 images per second and 576 points per image. The IOP and corneal thickness can be measured with great precision on the basis of the Scheimpflug images. We noted corneal biomechanical parameters including SSI that were assessed using Corvis ST at pre-LASIK and 1-month post-operatively. Table 1 describes corneal biomechanical parameters in detail.
Table 1: The corneal biomechanical parameters, their abbreviations and brief description
Statistical analysis
The data were entered into the Microsoft Excel spreadsheet (Microsoft Corporation). Further, Minitab 17 statistical software (Minitab LLC, State University, PA, USA) was used for statistical analysis. Pre- and post-surgery corneal biomechanical metrics were compared using the Paired T-test (Two-tailed, a <0.05) for each group separately.
Results
Out of the 202 eyes of 101 patients who had undergone LASIK refractive surgery, 119 eyes were excluded due to missing data whereas 2 eyes had LASIK enhancement and 2 eyes had astigmatism >3.00 D and were excluded. Finally, 79 eyes were included out of which 43 eyes had undergone Blade LASIK (Group I) whereas 36 eyes had undergone Femto LASIK (Group II). The pre-surgery mean ± SD age, spherical equivalent refractive error, IOP, keratometry, and pachymetry of Groups I and II have been described in Table 2 .
Table 2: Describes pre-surgery mean±SD age, spherical equivalent refractive error, IOP, keratometry and pachymetry of Groups I and II
Corneal biomechanical parameters and SSI
Out of 33, 29, and 26 corneal biomechanical parameters changed significantly post-Blade LASIK and Femto Bladeless LASIK , respectively.
At applanation 1
A1 time, A1 deformation amplitude, A1 deflection length, A1 deflection amplitude, and A1 deflection area were reduced significantly post-LASIK in both Blade (P < 0.001) and Femto bladeless groups (P < 0.001) whereas the A velocity and A delta arc length increased significantly post-LASIK in both the groups (P < 0.001).
At Applanation 2
In the blade group, post-LASIK significant changes were noted in A2 time, A2 velocity, A2 deflection amplitude, A2 deflection area, A2 delta arc length (P < 0.001) whereas A2 deformation amplitude and A2 deflection length remained unchanged. On the other hand, in the Femto bladeless group, significant changes were noted in A2 time (P = 0.003), A2 deflection length (P = 0.04), and A2 delta arc length (P < 0.001) post-surgery, however, the other parameters at A2 applanation remained unchanged.
At the highest concavity
Radius, HC deformation amplitude, HC deflection amplitude, peak distance, HC deflection area, and HC delta arc length significantly changed post-LASIK in both the groups (P < 0.001), however, no changes were noted in HC time in Blade LASIK (P = 0.32) and Femto groups (P = 0.15). Also, HC deflection length was significantly increased in the blade group (P = 0.003) whereas a non-significant increase was noted in the Femto group (P = 0.18).
Maximum parameters
Maximum deformation, maximum deflection, maximum delta arc length, and maximum inverse radius were found to be significantly increased in post-LASIK in both groups (P < 0.001).
Vinciguerra screening parameters and SSI
Post-LASIK , DA ratio max at 2 mm, and integrated radius increased significantly and Ambrósio’s Relational Thickness to the horizontal profile (ARTh), Biomechanically corrected Intraocular Pressure (bIOP), and the stiffness parameter at A1 was reduced significantly (Paired T-test, P < 0.001), however, no changes were noted in SSI in both the groups when compared with the pre-surgery data. Table 3 describes the mean ± SD corneal biomechanical parameters pre- and post-refractive surgeries.
Table 3: Describes the mean±SD corneal biomechanical parameters pre- and post-refractive surgeries
Discussion
The alteration in the corneal biomechanical parameters post-ocular surgeries such as cataracts and various types of refractive surgeries have been well-described and reported in the literature previously.[15 16 17 18 19 ] Various instruments such as ORA Corvis ST, supersonic shear imaging, and optical coherence tomography have been used to assess the corneal biomechanical changes.[7 ] Instruments such as Corvis ST are still evolving and also record new metric that is SSI. In the present study, we have reported short-term changes in SSI along with corneal biomechanical parameters post-refractive surgery such as Blade LASIK and Femto Bladeless LASIK . We noted changes in various corneal biomechanical parameters at the first applanation at the highest concavity and in the Vinciguerra screening parameters, however, there was no statistically significant change in the SSI.
In the present study, the parameters at the second applanation such as A2 velocity, deflection amplitude, and deflection areas were affected post-surgery in the blade group, however, no difference was noted in the Femto LASIK group. Previous studies have also reported significant changes in the corneal biomechanical parameters post-refractive surgeries such as Blade LASIK , Femto LASIK , and SMILE.[20 21 ]
New Corvis ST parameters such as stiffness parameter at first applanation (SPA1), DA ratio max at 2 mm, Ambrósio’s relational thickness horizontal (ARTh), integrated radius, and CBI are significantly different in LASIK patients as compared to normal subjects.[22 ] In the present study, we noted a reduction in the stiffness parameter at A1 post-LASIK surgeries in both groups. In the blade group, the stiffness parameter reduced to 92.52 from 118.08 units whereas, in the bladeless group, it reduced to 87.29 from 114.30 units. Also, the DA ratio Max at 2 mm was found to be significantly increased post-surgery in both groups. The DA ratio describes the corneal resistance to deformation where a higher value is associated with lower resistance to deformation. Thus, the increase in the DA ratio values post-refractive surgery suggests that the cornea becomes softer post-refractive surgery, however, in the present study the increase in the DA ratio was similar in both groups. The DA ratio was found to increase by 0.7 and 0.8 units in post-Blade LASIK and Femto bladeless LASIK groups, respectively.
CBI is a composite biomechanical metric that distinguishes a normal cornea from an abnormal cornea and helps predict ectasia. CBI up to 0.2 is considered as normal, from 0.21 to 0.49 is subnormal, and >0.5 is considered as abnormal. Previous studies have reported an increase in CBI in eyes with ectasia.[13 23 ] However, the newer software allows an automatic assessment of the biomechanical stability post-laser vision correction using a new metric called CBI- Laser Vision Correction (LVC). CBI-LVC estimates the risk for ectasia after laser vision correction. This information helps to make a clinical decision about the re-treatment or further laser touchups. A recent study has shown that the CBI-LVC is highly sensitive and specific in distinguishing stable from ectatic post-LVC eyes.[24 ] The mean ± SD pre-LASIK CBI was 0.29 ± 0.22 and 0.31 ± 0.17 in the Blade and Femto group, respectively, whereas the mean ± SD post-LVC-CBI was 0.07 ± 0.19 and 0.09 ± 0.23 in the Blade and Femto group, respectively, which are within normal ranges suggesting stable corneal biomechanics post-LVC, and thus, no further requirement of treatment. Since both CBI and LVC-CBI follow a different scale, we did not compare pre-op CBI and post-LVC-CBI.
The corneal deformation parameters in vivo are affected by IOP and Central corneal thickness (CCT),[25 ] however, SSI is independent of CCT and IOP and positively associated with age.[12 ] Using the bIOP algorithm, tangent modulus, which is a measure of material stiffness, can be determined under any IOP. Also, the stiffness parameter at HC (SP-HC) is more strongly associated with corneal stiffness than IOP, and thus, is used in the algorithm to estimate the SSI. In the present study, despite the changes in the other corneal deformation parameters, we did not find any significant changes in post-LASIK SSI which is a better predictor of strengthening. Previous research by Lopes et al .[26 ] has reported a significant increase in SSI to 0.87 post-Corneal cross-linking (CXL) from 0.78 pre-CXL. SSI values >1.0 suggest a stiffer cornea and <1.0 suggest a softer cornea. To the best of our knowledge, this is the first study which describes the SSI post-Blade and Femto Bladeless LASIK . In the Blade group, the SSI decreased from 0.86 to 0.84 and in the Femto group, it reduced from 0.80 to 0.79. Though the reduction in the SSI was not statistically significant, other individual biomechanical parameters showed significant biomechanical changes pre- and post-refractive surgeries.
The strain is dependent on various factors such as refractive procedure (LASIK vs. PRK), preoperative refractive correction, flap thickness, and corneal geometries.[27 ] In the present study, we did not find any association between preoperative refractive corrections with SSI (Pearson Correlation Blade LASIK : r = 0.25, P = 0.09; Femto LASIK : r = 0.07, P = 0.67). SSI algorithm is suitable for corneas with normal topography. It is noteworthy that all patients enrolled in the study had normal corneal geometries. Also, the strain-stress relation of the cornea is non-linear[28 29 ] which means that the tissue has tangent modulus and it increases with stress or pressure. This suggests that the weakening in the corneal biomechanics is not changing linearly with the laser ablation or percentage of tissue alteration. Thus, it is important to evaluate factors such as the affection of the corneal collagen fibers arrangement during refractive surgeries that are directly responsible for corneal strengthening.
Previous studies have reported that various corneal biomechanical parameters such as Corneal hysteresis (CH) and corneal resistance factor (CRF) by ORA and deformation parameters by Corvis ST were based on geometrical effects and IOP and did not provide a measure of material behavior. On the other hand, SSI describes intrinsic material stiffness. The SSI numerical model includes the effect of the ciliary muscle in simulating the corneal biomechanical response to IOP and air puff whereas the iris, lens, and retina effects are not accounted for due to their much lower stiffness relative to the cornea.
Corneal biomechanical weakening increases with an increase in the flap thickness and ablation depth.[30 31 32 33 ] A thin flap (SBK) allows a thicker residual stromal bed, less damage to the nerves, and induces less biomechanical weakening of the cornea. Thus, it is more relevant to have a LASIK procedure which is tissue saving and inducing the least amount of corneal weakening. Thus, the present study evaluated the corneal biomechanical changes with the thinnest flap (90 microns). Although it has been reported that microkeratome tends to over-cut the flap by 10–15 microns on average.[30 34 35 ] With this consideration, the present study could be useful to extrapolate biomechanical changes in the eyes with the 110-microns flap surgeries. Also, with the advent of Femtosecond procedures opting for thinner flaps and caps being a trend, the results of the present study would be applicable to clinical practice.
The refractive error range is not similar in both the groups (the mean ± SD spherical equivalent refractive error for Blade and Femto LASIK group is −4.07 ± 1.52 and −5.34 ± 3.05 years, respectively). The primary aim of the paper is to compare the LASIK -induced changes in SSI which are independent of the refractive error. Also, the other biomechanical parameter depends on the amount of refractive error treated and laser refractive correction decreases the biomechanical strength. However, the present study compares the pre-versus post-biomechanical changes in each group separately and not making any head-to-head comparison of biomechanical changes between post-Blade LASIK versus post-Femto LASIK .
Biomechanical changes post-LASIK over different timelines have been studies previously.[32 16 36 ] Understanding the earliest time-point for post-LASIK biomechanical changes would help us to know the impact of weakening due to the procedure before the major reparative process sets in. Immediate 1-month changes were studied to understand the biomechanical changes and to understand each biomechanical parameter individually. This would help us to enhance our understanding of the behavior of the newer parameter (SSI) as well as other biomechanical parameters in response to LASIK . It has been reported that corneal biomechanical changes occur as early as 1 week to 10 days post-surgery.[37 ] Here, we report short-term results at 1-month follow-up as studied previously.[16 14 ] Nevertheless, the ultra-structural healing after LASIK takes at least 3 months, however, due to retrospective design. A Smaller sample size is the limitation of our study.
Conclusion
To conclude, the reduction in SSI was not statistically significant. Other biomechanical parameters showed significant changes post-LASIK surgeries in both groups. However, a long-term study with a larger sample size would be required to understand the changes and stability in SSI and other corneal biomechanical parameters post-refractive surgery.
Financial support and sponsorship
Nil.
Conflicts of interest
There are no conflicts of interest.
References
1. Pesudovs K, Garamendi E, Elliott DB The Quality of life impact of refractive correction (QIRC) questionnaire:Development and validation Optom Vis Sci 2004 81 769 77
2. Pesudovs K, Garamendi E, Elliott DB A quality of life comparison of people wearing spectacles or contact lenses or having undergone refractive surgery J Refract Surg 2006 22 19 27
3. Bohac M, Koncarevic M, Pasalic A, Biscevic A, Merlak M, Gabric N, et al. Incidence and clinical characteristics of post-
LASIK ectasia:A review of over 30,000
LASIK cases Semin Ophthalmol 2018 33 869 77
4. Wolle M, Randleman JB, Woodward M Complications of refractive surgery:Ectasia after refractive surgery Int Ophthalmol Clin 2016 56 129 41
5. Siganos CS, Kymionis GD, Astyrakakis N, Pallikaris IG Management of corneal ectasia after laser
in situ keratomileusis with intacs J Refract Surg 2002 18 43 6
6. Bromley JG, Randleman JB Treatment strategies for corneal ectasia Curr Opin Ophthalmol 2010 21 255 8
7. Roy AS, Shetty R, Kummelil MK Keratoconus:A biomechanical perspective on loss of corneal stiffness Indian J Ophthalmol 2013 61 392 3
8. Lopes BT, Roberts CJ, Elsheikh A, Vinciguerra R, Vinciguerra P, Reisdorf S, et al. Repeatability and reproducibility of intraocular pressure and dynamic corneal response parameters assessed by the Corvis ST J Ophthalmol 2017 2017 8515742
9. Vinciguerra R, Ambrósio R Jr, Elsheikh A, Roberts CJ, Lopes B, Morenghi E, et al. Detection of keratoconus with a new biomechanical index J Refract Surg 2016 32 803 10
10. Ambrósio R Jr, Lopes BT, Faria-Correia F, Salomão MQ, Bühren J, Roberts CJ, et al. Integration of scheimpflug-based corneal tomography and biomechanical assessments for enhancing ectasia detection J Refract Surg 2017 33 434 43
11. Yang K, Xu L, Fan Q, Zhao D, Ren S Repeatability and comparison of new Corvis ST parameters in normal and keratoconus eyes Sci Rep 2019 9 15379
12. Eliasy A, Chen K-J, Vinciguerra R, Lopes BT, Abass A, Vinciguerra P, et al. Determination of corneal biomechanical behavior
in vivo for healthy eyes using Corvis ST tonometry:Stress-strain index Front Bioeng Biotechnol 2019 7 105
13. De Stefano VS, Dupps WJ Biomechanical diagnostics of the cornea Int Ophthalmol Clin 2017 57 75 86
14. Osman IM, Helaly HA, Abdalla M, Shousha MA Corneal biomechanical changes in eyes with small-incision lenticule extraction and laser-assisted
in situ keratomileusis BMC Ophthalmol 2016 16 123
15. Zhang J, Zheng L, Zhao X, Xu Y, Chen S Corneal biomechanics after small-incision lenticule extraction versus Q-value - Guided femtosecond laser-assisted
in situ keratomileusis J Curr Ophthalmol 2016 28 181 7
16. Chen MC, Lee N, Bourla N, Hamilton DR Corneal biomechanical measurements before and after laser
in situ keratomileusis J Cataract Refract Surg 2008 34 1886 91
17. Kato Y, Nakakura S, Asaoka R, Matsuya K, Fujio Y, Kiuchi Y, et al. Cataract surgery causes biomechanical alterations to the eye detectable by Corvis ST tonometry PLoS One 2017 12 e0171941
18. Hirasawa K, Nakakura S, Nakao Y, Fujino Y, Matsuura M, Murata H, et al. Changes in corneal biomechanics and intraocular pressure following cataract surgery Am J Ophthalmol 2018 195 26 35
19. Pepose JS, Feigenbaum SK, Qazi MA, Sanderson JP, Roberts CJ Changes in corneal biomechanics and intraocular pressure following
LASIK using static, dynamic, and noncontact tonometry Am J Ophthalmol 2007 143 39 48
20. Sefat SMM, Wiltfang R, Bechmann M, Mayer WJ, Kampik A, Kook D Evaluation of changes in human corneas after femtosecond laser-assisted
LASIK and Small-incision lenticule extraction (SMILE) using noncontact tonometry and ultra-high-speed camera (Corvis ST) Curr Eye Res 2016 41 917 22
21. Frings A, Linke SJ, Bauer EL, Druchkiv V, Katz T, Steinberg J Effects of Laser
in situ keratomileusis (
LASIK ) on corneal biomechanical measurements with the Corvis ST tonometer Clin Ophthalmol 2015 9 305 11
22. Yang K, Xu L, Fan Q, Gu Y, Song P, Zhang B, et al. Evaluation of new Corvis ST parameters in normal, Post-
LASIK , Post-
LASIK keratectasia and keratoconus eyes Sci Rep 2020 10 5676
23. Belin MW, Ambrósio R Enhanced screening for ectasia risk prior to laser vision correction Int J Keratoconus Ectatic Corneal Dis 2017 6 23 33
24. Vinciguerra R, Ambrósio R Jr, Elsheikh A, Hafezi F, Yong Kang DS, Kermani O, et al. Detection of Post-Laser Vision Correction Ectasia with a new Combined Biomechanical Index J Cataract Refract Surg 2021 02 24 doi:10.1097/j.jcrs.0000000000000629 Epub ahead of print 33769761
25. Kenia VP, Kenia RV, Pirdankar OH Association between corneal biomechanical parameters and myopic refractive errors in young Indian individuals, Taiwan Taiwan J Ophthalmol 2020 10 45 53
26. Lopes B, Elsheikh A Development and validation of a material stiffness parameter based on the Corvis ST Discover a New World:Oculus Special Supplement 2019 September 7 8
27. Dupps WJ, Santhiago MR Structural relationships in post-refractive surgery ectasia:What have we learned? J Cataract Refract Surg 2019 45 391 3
28. Elsheikh A, Wang D, Brown M, Rama P, Campanelli M, Pye D Assessment of corneal biomechanical properties and their variation with age Curr Eye Res 2007 32 11 9
29. Ethier CR, Johnson M, Ruberti J Ocular biomechanics and biotransport Annu Rev Biomed Eng 2004 6 249 73
30. Goussous IA, El-Agha M-S, Awadein A, Hosny MH, Ghaith AA, Khattab AL The effect of flap thickness on corneal biomechanics after myopic laser
in situ keratomileusis using the M-2 microkeratome Clin Ophthalmol 2017 11 2065 71
31. Knox Cartwright NE, Tyrer JR, Jaycock PD, Marshall J Effects of variation in depth and side cut angulations in
LASIK and thin-flap
LASIK using a femtosecond laser:A biomechanical study J Refract Surg 2012 28 419 25
32. Guo H, Hosseini-Moghaddam SM, Hodge W Corneal biomechanical properties after SMILE versus FLEX,
LASIK , LASEK, or PRK:A systematic review and meta-analysis BMC Ophthalmol 2019 19 167
33. Medeiros FW, Sinha-Roy A, Alves MR, Dupps WJ Jr Biomechanical corneal changes induced by different flap thickness created by femtosecond laser Clinics 2011 66 1067 71
34. Talamo JH, Meltzer J, Gardner J Reproducibility of flap thickness with intraLase FS and Moria LSK-1 and M2 microkeratomes J Refract Surg 2006 22 556 61
35. Muallem MS, Yoo SY, Romano AC, Schiffman JC, Culbertson WW Corneal flap thickness in laser
in situ keratomileusis using the Moria M2 microkeratome J Cataract Refract Surg 2004 30 1902 8
36. Wang D, Liu M, Chen Y, Zhang X, Xu Y, Wang J, et al. Differences in the corneal biomechanical changes after SMILE and
LASIK J Refract Surg 2014 30 702 7
37. Chen S, Chen D, Wang J, Lu F, Wang Q, Qu J Changes in ocular response analyzer parameters after
LASIK J Refract Surg 2010 26 279 88