The Effect of Contact Lens Induced Corneal Edema on Goldmann Applanation Tonometry Measurements

Hamilton, Kirsten E. B.Optom(Hons); Pye, David C. M.Optom; Hali, Amir B.Optom; Lin, Cassie B.Optom; Kam, Phillip B.Optom; Ngyuen, Trinh B.Optom

doi: 10.1097/01.ijg.0000212277.95971.be
Original Studies
Buy

Aim To determine the effect of small increases in corneal hydration on the accuracy of Goldmann applanation tonometry estimates of intraocular pressure (IOP).

Materials and Methods Twenty-five young healthy subjects presented on 3 separate days approximately 1 week apart. On 2 visits, subjects were required to wear a hydrogel contact lens with either a center thickness of 0.3 and 0.7 mm (HEMA 38% water content, parallel surface curve) in 1 eye only under closed-eye conditions for 2 hours to induce corneal swelling. The third visit acted as a control. IOP, corneal thickness, and corneal curvature were measured in both eyes before and after contact lens wear on all visits.

Results There was a statistically significant increase in corneal thickness of 40.2±14.4 μm (P<0.001) and 41.9±16.4 μm (P<0.001) after wearing the 0.3 and 0.7 mm thick contact lenses, respectively (2-tailed paired t test). There was an increase in IOP of 2.8±2.2 mm Hg (P<0.001) after wearing the 0.3 mm thick contact lens, and a statistically insignificant difference of 1.3±3.0 mm Hg (P=0.058) after wearing the 0.7 mm thick contact lens (2-tailed paired t test). There was a statistically significant Pearson correlation between the change in corneal thickness and the change in IOP after lens wear (r=0.500, P<0.001, 0.3 mm lens and r=0.399, P<0.001, 0.7 mm lens). The corneal hydration-induced measurement error was 0.46 mm Hg per 10 μm change in corneal thickness (0.3 mm lens) and 0.35 mm Hg per 10 μm change in corneal thickness (0.7 mm lens).

Conclusion A small increase in corneal hydration and thickness may cause a clinically significant overestimation of IOP when measured using Goldmann applanation tonometry.

School of Optometry and Vision Science, University of New South Wales, Sydney, Australia

Reprints: David C. Pye, School of Optometry and Vision Science, University of New South Wales, Level 1, North Wing, Rupert Myers Building, Sydney NSW 2052, Australia (e-mail: d.pye@unsw.edu.au).

Received for publication March 6, 2006; accepted July 19, 2006

Sources of support: None.

Prior publication: The information contained in this article was published in part at the ARVO meeting in Florida, 2005. Hamilton KE, Pye DC, Hali A, Kam P, Lin C, Nguyen T. Corneal oedema can cause an overestimation error in Goldmann tonometry. IOVS 2005;46, E-abstract no. 4856.

Conflict of interest, financial disclosure: No author has any professional or personal relationships that may result in a conflict of interest.

© 2007 Lippincott Williams & Wilkins, Inc.