Abstracts and Programme: EUROANAESTHESIA 2011: The European Anaesthesiology Congress: Monitoring: Equipment and Computers

Effect of intermitent noninvasive blood pressure monitoring on the continuous noninvasive hemoglobin monitoring by pulse CO-oximetry (SpHb) with the masimo radical-7TMhemoglobin monitor

3AP1-5

Balaguer, J.; Soliveres, J.; Estruch, M.; Sánchez, A.; Hernández, M. J.; Solaz, C.

Author Information
European Journal of Anaesthesiology: June 2011 - Volume 28 - Issue - p 27
  • Free

Background and Goal of Study: SpO2 monitorization is known to be altered by non invasive blood pressure determiantion (NIBP). Althrough routine pulse oxymetry (SpO2) and NIBP cuff in the same arm is not recommended, the alteration on SpO2 and SpHb monitoring when NIBP is measured in the same arm are not described with the new Radical 7(TM) continuous SpHb (Masimo, USA). Our objective is to know wether NIBP and SpHb and SpO2 can be determined on the same arm.

Materials and Methods: After local ethics committee approval, written informed consent was obtained from 15 ASA I-III surgery scheduled patients. After standard monitoring, SpHb sensor and NIBP cuff were placed in the same arm patient. Patients with hemodynamic instability, severe vascular abnormalities or unable to acces to the upper extremity were excluded. After SpHb measurement stabilization, NIBP cuff was inflated every 10 minutes (a minimum of 4 measurements for each patient). Time between starting NIBP cuff inflation and SpO2 and spHb signal loss was recorded (measurement alteration time). Time between SpO2 and SpHb signal loss and SpO2 and SpHb baseline recovery was recorded (failure interval SpO2 and SpHb). Student's t test was used for statistical analysis. A p< 0,05 was considered significant.

Results and Discussion: No patients were excluded. 67 measurements were collected. Three measurements were excluded (one because of twice NIBP cuff inflation, two because of inflation time not recorded).

Demographic data ar shown in Table 1.Measurement alteration time and failure interval are shown in Table 2.

[Table 1. Data show number or mean (±SD)]
[Table 2. Data show mean (±SD) (*)p<0,05]

Twenty-five of the sixty-four measurements showed no alteration of the SpO2 and SpHb measurement (measurement alteration time and failure interval of SpO2 and SpHb equal to zero seconds).

Conclusion(s): SpO2 monitoring could be placed on the same arm as NIBP, althrough SpHb monitoring should not because of the much longer failure time.