Share this article on:

Emerging trends in minimally invasive haemodynamic monitoring and optimization of fluid therapy

Benington, Stevea; Ferris, Paulb; Nirmalan, Maheshc

European Journal of Anaesthesiology (EJA): November 2009 - Volume 26 - Issue 11 - p 893–905
doi: 10.1097/EJA.0b013e3283308e50
Review Articles

Background For decades the pulmonary artery catheter has been the mainstay of cardiac output monitoring in critically ill patients, and pressure-based indices of ventricular filling have been used to gauge fluid requirements with acknowledged limitations. In recent years, alternative technologies have become available which are minimally invasive, allow beat-to-beat cardiac output monitoring and permit assessment of fluid requirements by volumetric means and by allowing assessment of heart–lung interaction in mechanically ventilated patients.

Methods A qualitative review of the basic science behind the transpulmonary dilution technique used in the measurement of cardiac output, global end-diastolic volume and extravascular lung water; the basic science and validation of pulse contour analysis methods of real-time cardiac output monitoring; the application and limitations of these technologies to guide rational fluid therapy in surgical and critically ill patients.

Results Transpulmonary dilution techniques correlate well with pulmonary artery catheter-derived measurement of cardiac output. Volumetric measures of preload appear to be superior to central venous and pulmonary artery occlusion pressures. Dynamic indices of preload responsiveness such as stroke volume variation are more useful than static measures in mechanically ventilated patients.

Conclusion In fully mechanically ventilated patients, dynamic measurements of heart–lung interaction such as stroke volume variation are superior to static measures of preload in assessing whether a patient is volume-responsive (i.e. will increase stroke volume in response to a fluid challenge). For patients who are not fully mechanically ventilated, pulse contour analysis allows real-time assessment of increases in cardiac output in response to passive leg-raising.

aManchester Royal Infirmary, UK

bHope Hospital, UK

cManchester Royal Infirmary, Manchester, UK

Received 10 May, 2009

Revised 6 July, 2009

Accepted 6 July, 2009

Correspondence to Steve Benington, Department of Anaesthesia and Intensive Care Medicine, Manchester Royal Infirmary, Oxford Road, Manchester M13 9WL, UK E-mail:

Back to Top | Article Outline


Optimizing cardiovascular function in order to ensure adequate tissue oxygen delivery is a key objective in the care of critically ill patients. In order to achieve this safely, it is necessary to monitor left ventricular preload and cardiac output (CO) in a significant cohort of ICU patients. The pulmonary artery catheter (PAC) has been used extensively for this purpose for approximately four decades and has formerly been considered the gold standard measure of both left ventricular preload (pulmonary artery occlusion pressure, PAOP) and CO. However, routine right heart catheterization with a PAC is associated with a number of complications and has not been demonstrated to improve mortality or morbidity in a number of studies [1–4]. The National Heart, Lung and Blood Institute's Acute Respiratory Distress Syndrome (ARDS) Network has stated that ‘the pulmonary artery catheter should not be routinely used for the management of acute lung injury’, resulting in a dramatic reduction in the use of PACs in many institutions. This in turn is likely to erode the skill base and familiarity necessary for the safe use of the PAC.

In view of these changes, it has become necessary for every ICU to consider alternative strategies for haemodynamic monitoring. It is acknowledged that many options exist including oesophageal Doppler and transthoracic bioimpedance measurement, the former having acquired a strong evidence base of outcome data justifying its use in perioperative care [5]. Both methods have been expertly reviewed elsewhere [6,7], and will not be further discussed here. In this review, we consider in detail the scientific basis of two popular alternative strategies reliant on intra-arterial monitoring:

  1. The volumetric approach based on transpulmonary thermal dilution.
  2. Pulse contour analysis.

This review will focus on the underlying theory and the experimental and clinical evidence that would justify their use in current clinical practice.

Back to Top | Article Outline

A volume-based approach to measuring ventricular preload

Ventricular preload is a function of myocardial stretch and the most direct clinical correlate of this is the end-diastolic ventricular volume (EDV). The traditional approach of estimating EDV using pressure-based measurements such as central venous pressure (CVP) or PAOP has several limitations. Changes in CVP/PAOP have been shown to correlate poorly with changes in ventricular volume in numerous studies [8–10]. As an example, Kumar et al. [8] assessed the correlation between CVP/PAOP and EDV in response to a 3 l fluid challenge in healthy volunteers during spontaneous ventilation. Neither the initial CVP/PAOP nor the changes in these parameters (ΔCVP/ΔPAOP) were useful in predicting changes in end-diastolic volumes or stroke volume (SV) as measured by trans-oesophageal echocardiography (TOE). Such studies have emphasized the shortcomings of inferring ventricular volumes from pressure measurements and thereby stimulated interest in the use of volume-based estimation of ventricular preload. Although echocardiography can provide an accurate measure of ventricular and SVs, it is inconvenient for repeated use over a prolonged period and is highly operator-dependent. Therefore, devices that enable repeated bed-side assessment of intrathoracic blood volume (ITBV), global end-diastolic volume (GEDV) and extravascular lung water (EVLW) using transpulmonary indicator dilution technology are exciting and require further consideration.

Back to Top | Article Outline

Transpulmonary dilution techniques

The transpulmonary thermal-dye dilution technique

This technique was first developed in the 1970s and allowed the estimation of EVLW using simultaneously derived transpulmonary indocyanin green (IcG) and thermal dilution curves. Thermal-dye dilution technique requires the injection of cold IcG dye into the right atrium or superior vena cava with simultaneous measurement of the temperature and IcG dilution curves in the abdominal aorta. The volume of distribution for each of the indicators (dye and cold) is a function of CO and mean transit time (MTt). Although the ‘thermal bolus’ distributes throughout the entire fluid compartment (intrathoracic thermal volume, ITTV), the volume of distribution of IcG is limited to the intravascular compartment alone, because it is highly protein-bound. This gives an estimate of ITBV. EVLW is then derived as shown in equation [1]:

This approach to EVLW measurement has been validated in animal and human studies. It can reliably detect changes in EVLW of 20% or greater, and compares favourably with gold-standard gravimetric measurement [11]. Although accurate and effective [12–14], this technique is expensive, time consuming and too cumbersome for routine bedside use.

Back to Top | Article Outline

The single transpulmonary thermodilution technique

This technique allows derivation of ITTV, ITBV and EVLW using transpulmonary thermal dilution alone. A key concept is the derivation of pulmonary thermal volume (PTV) using the down slope time of the transpulmonary thermal dilution curve (DStT). The theoretical framework behind the use of DStT to determine PTV was originally described in 1951 by Newman et al. [15] and has been expertly reviewed more recently by Isakow and Schuster [16], to which the reader is referred for a more detailed account.

Newman et al. first considered a simple experimental model consisting of a single chamber with an inflow allowing injection of dye, and an outflow for repeated sampling and construction of a dilution curve (Fig. 1). In this model, a relationship between concentration at the sampling port and time after injection was derived based on the following assumptions: injection of dye into the chamber is instantaneous; mixing of dye in the chamber volume is complete and instantaneous; the flow rate of fluid through the chamber (Q) is constant; there is no loss of dye from the system; and after injection, only dye-free fluid flows into the system.

Fig. 1

Fig. 1

If a known quantity of dye is injected into the chamber of volume V, the rate of change in the number of units of dye remaining in the chamber at a given time is dependent on the concentration and the flow (Q) at that instant. Using this simplified model, Newman et al. demonstrated that it was possible to derive V from knowledge of Q and direct measurement of the gradient of the down slope portion of the dilution curve. In simple terms, subject to assumptions mentioned above, the volume of a compartment upstream of the sampling point may be derived from the gradient of the semi-logarithmic plot of the concentration–time curve (DStT) and the flow.

Newman et al. went on to describe the kinetics of dye flow through a three-compartment model, with a central largest compartment (analogous to the pulmonary circulation) and two smaller compartments on either side joined in series (analogous to the left and right heart). They demonstrated that DStT was dependent primarily on flow and volume of the larger central compartment, allowing pulmonary blood volume (PTV) to be estimated from the DStT as shown in the following equation [2].

Having ascertained ITTV and PTV, it is then possible to derive an accurate measure related to the volume of blood in the four cardiac chambers. This combined volume has been referred to as the GEDV and is given by the following equation [3]:

If one then assumes that ITBV is always a constant proportion of GEDV, it is possible to calculate ITBV from the GEDV value so derived. The precise numerical relationship between GEDV and ITBV was first defined by Sakka et al. [17] in a multicentre clinical study and further validated for common physiological perturbations such as acute lung injury and hypovolaemic shock by Nirmalan et al. [18,19]. These studies jointly support the derivation of ITBV using the transpulmonary thermal dilution curve alone using the following equation [4]:

The numbers derived using this approach are related to the volume of blood between the point of injection of the cold bolus (usually the superior vena cava) and the point of detection by the thermistor (usually the femoral or brachial arteries). Therefore, the terms ITBV and GEDV are misnomers in that they do not strictly refer to the volume of blood in the thoracic cavity or the four cardiac chambers, respectively. They are in effect hypothetical chambers that have been shown to be closely linked to the true central compartments of interest. This limitation should be born in mind when deciding the position of the sampling catheter.

A schematic diagram of the relevant volumes as estimated by the indicator dilution technology is shown in Fig. 2.

Fig. 2

Fig. 2

Back to Top | Article Outline

The derivation of cardiac output using transpulmonary indicator dilution

Accurate measure of CO is a prerequisite in the derivation of ITTV and PTV as described above. CO derived by applying the Stewart Hamilton equation to the transpulmonary indicator dilution curves is used in all commercially available monitors. Temperature (PiCCO; PULSION Medical Systems, Munich, Germany) and lithium (LidCO; LidCO Group PLC, London, UK) have become the most popular indicators. CO measured using transpulmonary dilution of either indicator has been shown to have an excellent correlation and agreement with PAC-derived CO measurements [20–24]. Compared with pulmonary artery thermodilution, transpulmonary dilution techniques involve dilution of the chosen indicator in a significantly greater volume of blood. Therefore, the resolution and accuracy of the dilution curve is greatly influenced by the background concentration of the chosen indicator. As lithium is not a naturally occurring ion in the body, transpulmonary lithium dilution curves might be expected to possess a better signal-to-noise ratio compared with transpulmonary thermal dilution curves. In spite of this potential limitation, transpulmonary thermal dilution has been used successfully to measure CO in normovolaemic as well as hypovolaemic persons in a variety of clinical and experimental conditions.

Back to Top | Article Outline

Value of volumetric assessment of cardiac preload

Several studies have evaluated the usefulness of GEDV and ITBV in determining ventricular preload and the adequacy of volume resuscitation. In general, changes in GEDV/ITBV better reflect changes in CO, SV and oxygen delivery than either ΔCVP or ΔPAOP [10,25–27]. In a study of 57 ICU patients with sepsis or septic shock, Sakka et al. [10] compared ITBV, CVP and PAOP as predictors of cardiac function as measured by PAC thermodilution. ITBV index was strongly correlated with stroke volume index (SVI, r = 0.66), whereas neither CVP nor PAOP demonstrated any significant correlation (r = 0.10 and 0.06, respectively). Changes in SVI were also more likely to be mirrored by changes in ITBV in this study. As both CO and GEDV are derived from the same thermodilution curve, a potential weakness in the use of GEDV is that any change in CO might lead to a corresponding change in GEDV due to mathematical coupling. In this context, a study by Michard et al. [28] is important that compared the effects of volume resuscitation and dobutamine infusion on SV and GEDV. This study found that although volume infusion increased both SV and GEDV, dobutamine infusion resulted in an increase in SV without a concurrent increase in GEDV. These results suggest that the positive correlation reported between SV and GEDV cannot be attributed to mathematical coupling alone.

The volumetric assessment of preload is not without drawbacks. As pointed out by Isakow and Schuster [16], the correlation coefficients between GEDV/ITBV and indices of myocardial performance are usually modest (r = 0.4–0.5) and hence of only of moderate strength. Although this may be a consequence of changes in other factors such as contractility and afterload that may vary during the study period, it is equally plausible that many of the assumptions related to indicator dilution technology do not hold true in a physiological system resulting in an inherent weakness in attempting to measure volumes using the above-mentioned principles. Further theoretical and experimental studies are required to explore the uncertainties arising from pulsatile (rather than constant) flow, the effect of transit time on indicator distribution, injection time (injection can never be an instantaneous process) and other relevant variables.

Pinsky and co-workers [29–31] have highlighted the importance of distinguishing between preload and preload responsiveness. ITBV/GEDV are static indicators of preload, and may prove to be less useful than dynamic measures of preload responsiveness such as systolic and pulse pressure variation (discussed later). Such measures have the advantage that they can be derived from a standard arterial line continuously in real time, whereas ITBV measurement is more invasive and noncontinuous.

Back to Top | Article Outline

Extravascular lung water

EVLW measurement provides an estimate of pulmonary oedema, not reliably detected clinically or by chest radiography. A value of more than 10 ml kg−1 is considered abnormally high, and is consistent with ARDS in the appropriate clinical context. High EVLW is associated with poor outcome; in a retrospective analysis of 373 ICU patients, an EVLW content of more than 15 ml kg−1 was associated with a mortality rate of 65%, approximately double that of patients with EVLW less than 10 ml kg−1 [32]. There is some evidence that using EVLW as an endpoint in treatment algorithms may improve outcome. In a trial of 101 critically ill ICU patients with PAC in situ, Mitchell et al. [33] randomized 52 patients to fluid management guided by bedside dual indicator EVLW measurement, whereas the remainder were managed based on PAC-derived pulmonary capillary wedge pressure. The EVLW group received significantly less fluid across the course of their illness (median cumulative fluid balance 754 vs. 1600 ml), and had a significantly shorter duration of mechanical ventilation and ICU stay.

As mentioned previously, dual indicator technology is cumbersome for routine bedside use; single indicator transpulmonary thermodilution (as used in the PiCCO system) allowing the derivation of EVLW is far more convenient. EVLW measurements obtained using the PiCCO system have been used as valid endpoints for a therapeutic trial in patients with acute lung injury and ARDS [34]. The authors were able to show significant differences in EVLW in response to a specific therapeutic intervention (intravenous salbutamol). A haemodynamic management protocol using EVLW is available from the PiCCO manufacturer (PULSION Medical Systems), although this has not been independently validated, and is not widely used in the critical care environment. The LiDCO plus system also has provision for determining EVLW during lithium calibration using the dual indicator principle as described previously.

EVLW is not without limitations as a measurement tool. Indicator dilution estimates of EVLW are inherently perfusion dependent. There is a slight overestimation of 5–10% in normal lungs [14,19] due to distribution of the cold bolus in extra-pulmonary tissues (e.g. heart muscle) leading to an overestimation of ITTV. With global or regional hypoperfusion, there is the potential for underestimation of ITTV and hence EVLW [14,16,17]. Although global hypoperfusion effects may be counterbalanced by the increased transit time (allowing more time for diffusion of the cold bolus within extra-pulmonary tissues), these effects are unpredictable, hindering the adoption of EVLW as a reliable endpoint for management protocols. Although the dual indicator technique has been compared favourably with gravimetric measurement [11], no published human studies have validated the accuracy of PiCCO-derived (single indicator) EVLW measurement.

A well publicized trial [35] from the ARDS Network group demonstrated improvements in pulmonary function with a conservative fluid strategy, in keeping with early observations on the value of EVLW measurement [12,13]. The association of worsening pulmonary function with a liberal fluid strategy is almost certainly due to interstitial or alveolar oedema, and although the ARDS Network group used the term ‘liberal’ for their control group fluid strategy, the fluid balance in this group (about >1000 ml/24 h) is in keeping with that of patients in previous ARDS Network studies [36,37]. Although fluid therapy will always remain important in critically ill patients, with increasing evidence of adverse pulmonary consequences due to liberal fluid strategies, detailed attention to EVLW and consequent rationalization of fluid therapy may help minimize these adverse consequences.

The appropriateness of EVLW as an endpoint for fluid titration will obviously vary with the clinical circumstances; running an ICU patient with ARDS ‘dry’ by titrating against EVLW might improve outcome, but would be inappropriate in the perioperative management of a patient undergoing major abdominal surgery.

Back to Top | Article Outline

Stroke volume derivation from arterial pulse wave analysis

The existence of a relationship between the arterial waveform and CO has been recognized for over a century [38]. The ability to accurately determine SV on a beat-to-beat basis has obvious attractions. In addition to CO monitoring, this technology offers the prospect of an answer to the key question: ‘will such patients increase their CO in response to a fluid challenge?’ This has been termed preload responsiveness and is discussed further below. The various commercially available arterial pulse wave analysis (APWA) systems derive CO from the arterial waveform using different methods.

Back to Top | Article Outline

Pulse contour analysis: PiCCO

The PiCCO (Pulse Contour Continuous Cardiac Output) system relies on the central tenet that flow is proportional to the driving pressure divided by resistance. If pressure is measured directly and the resistance is known or assumed, the flow (CO) can then be calculated. A measured pressure–time waveform can be transformed into a flow-time waveform by mathematical modelling of the resistance to flow provided by the vascular tree [39,40]. This complex process is summarized diagrammatically in Fig. 3 (modified from Cholley et al. [40]).

Fig. 3

Fig. 3

The above model incorporates resistance, impedance and capacitance and allows a flow–time curve to be approximated from a given arterial trace. Other complex factors also affect the relationship between the flow and pressure waveforms in vivo, including wave reflectance phenomena and variation in resistance of the vascular tree. Taking such factors into account allows further refinement of this model.

Back to Top | Article Outline

Pulse power analysis: PulseCO

The PulseCO algorithm employed in the LiDCO monitoring system derives CO through a different theoretical basis to PiCCO, using a three-step approach [41–43]:

  1. A pressure–volume transformation is first performed on the entire arterial pressure waveform. This pressure–volume conversion makes assumptions about the compliance of the arterial tree, which varies throughout the cardiac cycle, being reduced at higher pressures such that expansion of the vascular tree is less for a given increase in pressure.
  2. A continuous curve of arterial volume change throughout the cardiac cycle is thus produced. This is used as the basis for calculation of a nominal SV, a process involving application of the mathematical technique of autocorrelation.
  3. Nominal SV is converted to an actual SV by calibration with the lithium indicator dilution technique.

Both the PICCO and PulseCO systems require calibration initially and at intervals to maintain accuracy. Although pulmonary artery catheterization can be employed [44–46], transpulmonary dilution has become the standard as outlined above. PICCO uses a thermal bolus and requires central venous access and femoral arterial sampling, whereas PulseCO uses lithium dilution and can be performed with a simple intravenous cannula and standard arterial line. Both transpulmonary techniques have been extensively validated as referenced above. With either method, measured CO is factored into the algorithm giving a picture of the individual circulatory characteristics of the patient. A continuous drift in the accuracy of these systems occurs over time, necessitating periodic recalibration, especially if large changes in systemic vascular resistance (SVR) are anticipated [47].

Back to Top | Article Outline


The FloTrac/Vigileo system (Edwards Life Sciences, California, USA) measures the pulsatility of the arterial waveform by calculating the standard deviation of the arterial pressure wave over a 20-s period. This is multiplied by a constant quantifying arterial compliance and vascular resistance based on patient demographic data (age, sex, height and weight). The system constantly ‘fine tunes’ according to the character of the arterial waveform and autocalibrates every minute [48]. This system is not calibrated by direct CO measurement, as are the other systems, calling into question its accuracy; autocalibration is based on experimentally determined cadaver data.

Back to Top | Article Outline

Value of arterial pulse wave analysis in assessing cardiac output

The value of APWA lies in its ability to provide a continuous measure of CO rather than the snapshots provided by dilution techniques. APWA has been extensively validated with reference to existing methods of CO monitoring, primarily PAC-derived thermodilution. Although the latter has historically been considered the gold standard in clinical practice, it has also long been recognized that numerous sources of error are inherent to the technique. Errors of more than 20% for single measurements and nearly 13% for triplicate measurements can be expected [49], and new technologies must be judged against this standard. Comparison between methods of CO monitoring is expressed in terms of bias and limits of agreement [50]. Limits of agreement of ±30% are generally considered acceptable for a new method of CO measurement [51].

Numerous studies have examined the correlation between APWA-derived and PAC-derived CO measurements in cardiac surgical patients, both intra-operatively [32,41,52,53] and postoperatively [54–58]. In a series of 24 postoperative cardiac surgical patients, Goedje et al. [59] measured CO over the first 24 h by both PAC thermodilution and PiCCO, finding minimal bias (0.07 ± 1.4 l min−1) between the two techniques, and good correlation even during periods of substantial haemodynamic instability. In a series of 10 patients undergoing cardiac surgery, PulseCO-derived CO measurements showed a strong correlation with PAC thermodilution (r 2 = 0.88) [41]. During this study, the PulseCO system tracked phenylephrine-induced alterations in SVR accurately. Good agreement has also been shown in other patient groups (including hepatic surgical [60], lung transplant [61] and paediatric [62] patients) and using other comparators (including oesophageal Doppler [63] and lithium dilution [64]).

Not all studies have shown a good correlation with reference methods of CO measurement. Yamashita et al. [65] compared PulseCO with PAC thermodilution in 23 patients undergoing off-pump coronary artery bypass grafting and found only a weak correlation (r 2 = 0.55), with greater bias than comparable studies. The authors raised the possibility that the rapidly changing arterial compliance characteristics during surgery and pharmacological manipulation might cause drift requiring more frequent recalibration.

The FloTrac/Vigileo system does not require external calibration in contrast to rival commercially available APWA systems discussed above. Although isolated case reports suggest that it may be of some use in guiding rational fluid resuscitation [66], the majority of studies have found poor limits of agreement with reference methods of CO measurement. In a study of CO measurement in 40 cardiac surgical patients, Mayer et al. [67] found only moderate agreement with PAC-derived values (r = 0.53), and an average error of 46%. Biais et al. [68] found poor agreement with PAC-derived CO measurement in patients undergoing liver transplantation (43% error), especially in conditions of low SVR.

The FloTrac algorithm has been refined since its introduction and in its current form (version 01.10) autocalibrates every minute rather than just on initial set-up. It is clearly important to be aware of which version was used when comparing validation studies (by contrast, the LiDCO/PulseCO algorithm remains unchanged since its introduction). Some studies using the latest FloTrac version have shown narrower limits of agreement with reference CO measurement. Mayer et al. [69] found an acceptable margin of error (24.6%) in cardiac surgical patients both intra-operatively and postoperatively compared with simultaneous PAC-derived pulmonary artery thermodilution. In a similar study, Prasser et al. [70] found a clinically acceptable percentage error of 26.9%. Other studies using this latest generation of software have not shown such favourable limits of agreement, however. In cirrhotic patients undergoing liver transplantation, Biancofiore et al. [71] found a percentage error of 54% between FloTrac and PCA-derived CO measurement, with the greatest degree of bias occurring under conditions of low SVR. This result agreed with the findings of Sakka et al. [72] in an earlier study of septic ICU patients (using the previous software version), casting doubt on the reliability of the device in low-resistance states. In general, most studies show the percentage error of the FloTrac/Vigileo system to exceed the 30% limit of acceptability for clinical use, as summarized elegantly by Biancofiore et al. [71] in the study quoted above.

Although APWA has been extensively validated as a CO measurement tool, few studies have investigated its outcome. A systematic review [73] recently failed to demonstrate a survival advantage with the use of the PAC, which might be in part be explained by the morbidity associated with the use of such invasive monitoring. It is tempting to suppose that APWA might avoid such morbidity and improve the risk: benefit ratio of CO monitoring, improving patient outcome. The PAC/PiCCO Use and Likelihood of Success Evaluation (PULSE) study prospectively followed 331 ICU patients requiring CO monitoring by either PAC or PICCO. Use of PiCCO was preferred in septic shock, whereas PAC was preferred in patients with cardiac morbidity. No difference in outcome (ventilator free days, ICU-free days or hospital mortality) was found between the two groups after correction for demographic and clinical differences [74].

Back to Top | Article Outline

Limitations of arterial pressure waveform analysis

All forms of APWA require some degree of invasive vascular access. The LiDCO system requires only standard arterial and peripheral venous cannulae. The PiCCO system requires a specialized femoral arterial cannula as well as a central venous catheter. Although it has been demonstrated to work with a radial artery waveform [75], calibration of the PiCCO system by transpulmonary thermodilution has proved to be unreliable using a peripheral (radial) arterial cannula [76]. The FloTrac/Vigileo system requires only a specialized transducer and cable spliced into the patient's existing arterial pressure monitoring set-up.

Both PiCCO and LiDCO employ a transpulmonary technique for system calibration. Transpulmonary thermodilution can be inaccurate in the presence of intrathoracic haemorrhage, or intracardiac shunts. Lithium dilution is unsuitable in patients taking lithium salts (overestimating CO) and can be inaccurate in the presence of certain muscle relaxants including atracurium. Although the FloTrac/Vigileo system does not require such external calibration, this has raised concerns about its accuracy by some commentators.

APWA measurement of CO may be less accurate with significant variation in SVR (assumed to be constant for the purposes of converting an arterial pressure into CO). Although this has been borne out by some studies [50,77], others have demonstrated good performance in the face of changing SVR following calibration [78,79]. Arrhythmias reduce accuracy, as CO is calculated on a beat-to-beat basis, and significant fluctuation may render data meaningless; if the heart rate is reasonably constant, this is less of a problem.

Back to Top | Article Outline

Preload responsiveness

A key objective of critical care is to deliver optimal fluid resuscitation. Fluid requirements will vary with the clinical situation; excessive fluid administration is associated with increased morbidity and mortality in patients with acute lung injury [35], although under-resuscitation in the first 6 h increases mortality in patients with septic shock [80]. Although the Surviving Sepsis Campaign guidelines advocate a CVP-based approach to fluid resuscitation [81], static measures of cardiac preload have repeatedly been shown to poorly reflect fluid requirements. Indices such as CVP, PAOP, and ventricular end-diastolic volumes measured by echocardiography have been repeatedly shown to correlate poorly with increase in SV in response to a fluid challenge [8,82–84]. Dynamic markers of fluid responsiveness based on heart–lung interactions across the respiratory cycle have shown promise as better discriminators. The goal of such methods is to determine the position of the patient's heart on the Starling curve, as illustrated in Fig. 4.

Fig. 4

Fig. 4

Back to Top | Article Outline

Systolic and pulse pressure variation

Mechanical ventilation induces cyclical changes in systolic and pulse pressures, referred to as systolic pressure variation (SPV) and pulse pressure variation (PPV) [85,86]. Many clinicians use this ‘swing on the arterial line’ as the impetus for a fluid challenge. These parameters are themselves surrogates for SV variation (SVV), the physiology of which has been expertly reviewed elsewhere [87,88], and is summarized below.

During mechanical ventilation, inspiration raises intrathoracic pressure, causing a reduction in right ventricular preload and an increase in afterload. Over the course of a few heartbeats, the consequent reduction in right ventricular SV impacts on left ventricular filling, as the right ventricular and left ventricular are in series. This reduction in left ventricular preload (and hence CO) coincides temporally with the expiratory phase of mechanical ventilation. Although a degree of variation across the respiratory cycle is normal, this reduction is magnified in hypovolaemic states for the following reasons:

  1. In low-volume states, the vena cavae are more collapsible and, therefore, right ventricular preload is more compromised by the transmitted inspiratory increase in intrathoracic pressure.
  2. The transmission of pressure through the RA (more collapsible in low-volume states) also reduces right heart filling.
  3. In hypovolaemia, the ventricles operate on a steeper portion of the Frank–Starling curve, hence a more marked change in SV for a given change in preload.

The magnitude of SVV (and SPV/PPV) broadly correlates to the degree of preload responsiveness, a high SVV predicting a larger increase in SV in response to a fluid challenge. SPV and PPV are not entirely interchangeable: SPV is influenced by transmitted changes in pleural pressure, meaning that increasing inspiratory pressures may falsely accentuate SPV. PPV is not influenced in the same way, as inspiratory pressure changes are transmitted to both systolic and diastolic components of the arterial waveform, and pulse pressure remains unchanged. Hence, PPV theoretically reflects changes in CO more accurately, and may be a better index of preload responsiveness. A refinement of SPV is delta Down (ΔDown), defined as the percentage fall in arterial pressure from its maximum value to that at the end of expiration; this measurement requires a short end-expiratory pause. The heart–lung interaction and its relation to mechanical ventilation is summarized in Fig. 5.

Fig. 5

Fig. 5

Back to Top | Article Outline

Stroke volume variation and pulse contour analysis

Much of the literature on heart–lung interaction has focused on SPV and PPV as surrogates for SVV, as formerly no reliable beat-to-beat SV indicator was available. In animal studies, venesection of 30% of circulating volume in ventilated canine and porcine models has demonstrated a strong correlation between the degree of blood loss and the increase in SPV [89,90]. The same correlation was not observed with static markers such as CVP. Pizov et al. [91] rendered ventilated dogs hypotensive either by controlled haemorrhage or sodium nitroprusside infusion, demonstrating that PAOP and CVP were similarly reduced in both groups, whereas SPV and ΔDown were significantly increased only in the haemorrhagic group, distinguishing between preload and vasodilatation as a cause of hypotension. This would be a useful distinction in anaesthetized patients, in whom drug-induced vasodilatation or hypovolaemia of various causes might be responsible for hypotension. In patients with septic shock, Tavernier et al. [92] found that a ΔDown of more than 5 mmHg was a better predictor of a 15% increase in SV in response to a 500 ml fluid challenge than either PAOP or LVEDA. In a similar patient population, Michard et al. [93] found that a 13% variation in PPV predicted a response [defined as a 15% increase in cardiac index (CI)] to a 500 ml fluid challenge with high sensitivity (94%) and specificity (96%).

The advent of APWA has allowed the study of real-time SV variation monitoring. In postoperative cardiac surgical patients, SVV measured by PiCCO has been shown to correlate well with SPV and to predict fluid responsiveness, being superior in this regard to CVP and PAOP [94,95]. In a more recent study [96], a SVV of more than 10% measured by FloTrac/Vigileo in cardiac surgical patients predicted a 15% increase in CO in response to a 500 ml fluid challenge with a sensitivity of 82% and a specificity of 87%. In neurosurgical patients, Berkenstadt et al. [97] found that a SVV of 9.5% or more predicted a SV increase of more than 5% in response to a standard fluid challenge with a sensitivity of 79% and a specificity of 93%. In the same study, CVP was found to be of no better predictive value than random chance.

Although much of the work on SVV has been conducted on elective surgical patients, of perhaps greater interest is the reliability of SVV in physiologically abnormal situations, where greater benefit from appropriate fluid management may be expected. In a canine model, the predictive value of SVV was preserved even in extreme haemorrhage [98]. In septic shock patients, Marx et al. [99] found a good correlation between SVV and increase in CI in response to a standard fluid challenge. Reuter et al. [100] found SVV predicted fluid response equally well in both patients with reduced left ventricular function (ejection fraction <35%) and normal cardiac function.

Back to Top | Article Outline

Limitations of heart–lung interaction as predictors of fluid responsiveness

Perhaps the biggest limitation of SVV and PPV is its restriction to mechanically ventilated patients with no spontaneous breathing activity. Due to the completely different heart–lung interaction at play in the spontaneously breathing patient, this work is not generalizable to patients making spontaneous or pressure-supported respiratory efforts, and is, therefore, of limited use in perhaps the majority of ICU patients.

Lower tidal volumes reduce the magnitude of SVV. Many studies validating SVV/SPV have used supranormal tidal volumes of 10–15 ml kg−1, which have been shown to be deleterious in the setting of ARDS [36]. Using volumes of 10 ml kg−1, Wiesenack et al. [101] were unable to demonstrate a correlation between SVV and fluid response in cardiac surgical patients. De Backer et al. [102] found that with tidal volumes less than 8 ml kg−1, PPV was a much poorer predictor of a response to fluid challenge in ICU patients. The influence of tidal volume on the interpretation of SVV has been neatly illustrated by Renner et al. [103], who studied SVV in anaesthetized pigs rendered first hypovolaemic by venesection then hypervolaemic by volume replacement. At low tidal volumes (5 ml kg−1), SVV did not change significantly between different loading conditions. Although the use of higher tidal volumes (10 and 15 ml kg−1) allowed differentiation between different volume states, SVV at 15 ml kg−1 was elevated to a degree suggesting the need for fluid even in the hypervolaemic phase of the study. It seems, therefore, that low tidal volumes may reduce the sensitivity of SVV as a predictor of the need for fluid, whereas high tidal volumes may overestimate fluid requirements.

The Respiratory Systolic Variation Test (RSVT) may offer a solution to this problem. First described by Perel et al. [104], this test entails delivery of three consecutive mechanical breaths of 10, 20 and 30 cmH2O, causing a progressively increasing ‘swing’ in the arterial trace. The minimum systolic blood pressure during each breath is measured from the arterial pressure waveform, and the slope of the line of best fit for these three points is calculated (Fig. 6). A steeper gradient corresponds to an ‘emptier’ circulation and predicts a greater response to fluid challenge. Preisman et al. [105] evaluated the RSVT in a group of cardiac surgical patients, comparing its ability to predict preload response with a variety of indices including CVP, left ventricular end diastolic area (LVEDA), ITBV, SVV and PPV. RSVT (alongside PPV) was found to be the most sensitive and specific predictor of a response to a fluid challenge as assessed by transoesophageal echocardiography.

Fig. 6

Fig. 6

SVV is influenced by other factors besides preload, summarized by Kubitz and Reuter [106]. These include positive end-expiratory pressure (PEEP), chest and lung compliance, heart rate, arrhythmias and ventricular function. Although there is a collective weight of evidence to support SVV as a guide to volume responsiveness, there is considerable heterogeneity between studies with respect to patient group, fluid type, amount and speed of administration and criteria for a positive response to a fluid challenge.

Finally, although APWA has been shown to be accurate in steady-state conditions, it has not been extensively validated as accurate on a beat-to-beat basis, leading Pinsky [107] to urge caution when using APWA-derived SVV as a basis for clinical decision-making. This has been partly addressed by a recent study in anaesthetized pigs in which PiCCO-derived SVV showed excellent agreement with the gold standard technique of aortic ultrasonic flow probe SVV measurement [108].

Back to Top | Article Outline

Predicting fluid responsiveness in spontaneously breathing patients

The mechanics of heart–lung interaction are different in spontaneously breathing patients from those who are fully mechanically ventilated. In the spontaneously breathing patient, inspiration reduces intrathoracic pressure, with a consequent increase in right ventricular preload and a more modest increase in afterload. This effectively increases right ventricular SV, which, in turn, increases left ventricular SV a few heartbeats later (coinciding with expiration). In a literature review, Coudray et al. [109] found that an average of 37.5% of patients in studies of indices of preload responsiveness (such as those outlined above) exhibited spontaneous breathing activity. Few studies have differentiated between mechanically ventilated and spontaneously breathing patients in their analysis.

Of those studies specifically considering the spontaneously breathing patient, the evidence is unconvincing. SVV was shown not to predict fluid responsiveness in a series of 30 ITU patients with septic shock receiving pressure support ventilation [110]. Heenen et al. [111] conducted a study of 21 critically ill patients receiving either pressure-supported ventilation or breathing through facemask. In this study, PPV failed to predict increase in CO in response to a fluid challenge and was inferior to static indices (right atrial pressure, PAOP). In a recent series of 32 haemodynamically unstable ITU patients with spontaneous breathing activity, PPV of more than 12% was a specific (92%) but not sensitive (63%) predictor of response to a 500 ml fluid bolus [112].

Unfortunately, perhaps the majority of patients in the intensive care setting are on some form of pressure-supported ventilation, which limits the application of SVV to a small subset of patients. However, those patients who are fully mechanically ventilated are often the most unwell, and, therefore, have the most to gain through rational fluid management and optimization of CO.

Back to Top | Article Outline

Passive leg raising

Passive leg raising (PLR) has been proposed as an alternative means of predicting response to a fluid challenge which may be of greater value in spontaneously breathing patients. Elevation of the patients' legs to 45° effectively autotransfuses around 300 ml of whole blood into the central circulation [113]. This allows the effect of a fluid bolus to be assessed with the advantage of full reversibility, thus avoiding unwanted effects of extracellular fluid accumulation and cardiac overload that may accompany a speculative volume challenge. The response to PLR has been shown to be a sensitive predictor of response to a fluid challenge in mechanically ventilated patients. Boulain et al. [114] studied PLR in 15 ICU patients with acute circulatory failure and showed that a 4-min PLR caused a mean increase in SV (measured by PAC) of 4 ml, and an increase of 4 mmHg in the radial artery pulse pressure (PPrad). A good correlation was seen between the degree of increase in PPrad in response to SLR and the subsequent increase in SV in response to a 300 ml fluid challenge (r = 0.84).

Monnet et al. [115] investigated PLR in 71 critically ill patients, 31 of whom displayed spontaneous breathing activity. The effect of PLR was assessed by oesophageal Doppler (OD) measurement of aortic blood flow and by PPV measurement. An increase of more than 10% in OD-measured aortic blood flow in response to PLR was found to be an excellent predictor of subsequent response to fluid challenge (sensitivity 97%, specificity 94%). Although the excellent predictive value of OD measurement was retained in the spontaneously breathing subgroup, PPV failed to discriminate in these patients.

APWA-derived CO measurement may also be of value in PLR assessment of volume responsiveness. Ridel et al. [116] demonstrated that in patients with spontaneous breathing activity, an increase of more than 12% in the PiCCO-measured CI in response to PLR discriminated between responders and nonresponders to a subsequent fluid challenge with 70% sensitivity and 92% specificity.

Back to Top | Article Outline


Increasingly, the pulmonary artery catheter has come to inhabit a shrinking corner of an expanding marketplace. Numerous alternative means of measuring CO are now available to a clinician, although none ‘ticks all the boxes’, hence the multiplicity of devices. There is little doubt that ITBV and GEDV provide better estimates of cardiac preload than traditional markers such as CVP and PAOP. Their measurement also allows the derivation of EVLW, with the promise of better outcomes through accurate titration of intravenous fluid to the patient's requirements. However, such measurements are not provided in real time and are static rather than dynamic. In most clinical circumstances, what is required is a prediction of preload responsiveness. Pulse contour analysis offers the promise of real-time CO measurement coupled with such functional haemodynamic monitoring. Although questions remain about the ability of this technology to reproducibly predict volume responsiveness with lung protective ventilation (i.e. limited tidal volumes) and expectations of improved patient outcome remain unproven, we may soon inherit a future in which all intensively monitored patients have CO and SV variation monitoring as a standard function of arterial line insertion in theatre or in the ICU.

Back to Top | Article Outline


1 The National Heart, Lung, and Blood Institute Acute Respiratory Distress Syndrome (ARDS) Clinical Trials Network. Pulmonary-Artery versus Central Venous Catheter to Guide Treatment of Acute Lung Injury. N Eng J Med 2006; 354:2213–2224.
2 Harvey S, Harrison DA, Singer M, et al. Assessment of the clinical effectiveness of pulmonary artery catheters in management of patients in intensive care (PAC-Man): a randomised controlled trial. Lancet 2005; 366:472–477.
3 Rhodes A, Cusack RJ, Newman PJ, et al. A randomised, controlled trial of the pulmonary artery catheter in critically ill patients. Intensive Care Med 2002; 28:256–264.
4 Richard C, Warszawski J, Anguel N, et al. Early use of the pulmonary artery catheter and outcomes in patients with shock and acute respiratory distress syndrome: a randomized controlled trial. JAMA 2003; 290:2713–2720.
5 Abbas SM, Hill AG. Systematic review of the literature for the use of oesophageal Doppler monitor for fluid replacement in major abdominal surgery. Anaesthesia 2008; 63:44–51.
6 Singer M. Oesophageal Doppler. Curr Opin Crit Care 2009; 15:244–248.
7 Bayram M, Yancy CW. Transthoracic impedance cardiography: a noninvasive method of hemodynamic assessment. Heart Fail Clin 2009; 5:161–168.
8 Kumar A, Anel R, Bunnell E, et al. Pulmonary artery occlusion pressure and central venous pressure fail to predict ventricular filling volume, cardiac performance, or the response to volume infusion in normal subjects. Crit Care Med 2004; 32:691–699.
9 Della RG, Costa MG, Coccia C, et al. Preload and haemodynamic assessment during liver transplantation: a comparison between the pulmonary artery catheter and transpulmonary indicator dilution techniques. Eur J Anaesthesiol 2002; 19:868–875.
10 Sakka SG, Bredle DL, Reinhart K, Meier-Hellmann A. Comparison between intrathoracic blood volume and cardiac filling pressures in the early phase of hemodynamic instability of patients with sepsis or septic shock. J Crit Care 1999; 14:78–83.
11 Mihm FG, Feeley TW, Jamieson SW. Thermal dye double indicator dilution measurement of lung water in man: comparison with gravimetric measurements. Thorax 1987; 42:72–76.
12 Bock JC, Lewis FR. Clinical relevance of lung water measurement with the thermal-dye dilution technique. J Surg Res 1990; 48:254–265.
13 Lewis FR, Elings VB, Hill SL, Christensen JM. The measurement of extravascular lung water by thermal-green dye indicator dilution. Ann N Y Acad Sci 1982; 384:394–410.
14 Mihm FG, Feeley TW, Rosenthal MH, Lewis F. Measurement of extravascular lung water in dogs using the thermal-green dye indicator dilution method. Anesthesiology 1982; 57:116–122.
15 Newman EV, Merrell M, Genecin A, et al. The dye dilution method for describing the central circulation: an analysis of factors shaping the time–concentration curves. Circulation 1951; 4:735–746.
16 Isakow W, Schuster DP. Extravascular lung water measurements and hemodynamic monitoring in the critically ill: bedside alternatives to the pulmonary artery catheter. Am J Physiol Lung Cell Mol Physiol 2006; 291:L1118–L1131.
17 Sakka SG, Rühl CC, Pfeiffer UJ, et al. Assessment of cardiac preload and extravascular lung water by single transpulmonary thermodilution. Intensive Care Med 2000; 26:180–187.
18 Nirmalan M, Willard TM, Edwards DJ, et al. Estimation of errors in determining intrathoracic blood volume using the single transpulmonary thermal dilution technique in hypovolemic shock. Anesthesiology 2005; 103:805–812.
19 Nirmalan M, Niranjan M, Willard T, et al. Estimation of errors in determining intrathoracic blood volume using thermal dilution in pigs with acute lung injury and haemorrhage. Br J Anaesth 2004; 93:546–551.
20 Linton RA, Jonas MM, Tibby SM, et al. Cardiac output measured by lithium dilution and transpulmonary thermodilution in patients in a paediatric intensive care unit. Intensive Care Med 2000; 26:1507–1511.
21 Linton RA, Band DM, Haire KM. A new method of measuring cardiac output in man using lithium dilution. Br J Anaesth 1993; 71:262–266.
22 Linton R, Band D, O'Brien T, et al. Lithium dilution cardiac output measurement: a comparison with thermodilution. Crit Care Med 1997; 25:1796–1800.
23 Friedman Z, Berkenstadt H, Margalit N, et al. Cardiac output assessed by arterial thermodilution during exsanguination and fluid resuscitation: experimental validation against a reference technique. Eur J Anaesthesiol 2002; 19:337–340.
24 Sakka SG, Reinhart K, Meier-Hellmann A. Comparison of pulmonary artery and arterial thermodilution cardiac output in critically ill patients. Intensive Care Med 1999; 25:843–846.
25 Huber W, Umgelter A, Reindl W, et al. Volume assessment in patients with necrotizing pancreatitis: a comparison of intrathoracic blood volume index, central venous pressure, and hematocrit, and their correlation to cardiac index and extravascular lung water index. Crit Care Med 2008; 36:2348–2354.
26 Gödje O, Peyerl M, Seebauer T, et al. Central venous pressure, pulmonary capillary wedge pressure and intrathoracic blood volumes as preload indicators in cardiac surgery patients. Eur J Cardiothorac Surg 1998; 13:533–539.
27 Della Rocca G, Costa MG, Coccia C, et al. Preload index: pulmonary artery occlusion pressure versus intrathoracic blood volume monitoring during lung transplantation. Anesth Analg 2002; 95:835–843.
28 Michard F, Alaya S, Zarka V, et al. Global end-diastolic volume as an indicator of cardiac preload in patients with septic shock. Chest 2003; 124:1900–1908.
29 Pinsky MR, Payen D. Functional hemodynamic monitoring. Crit Care 2005; 9:566–572.
30 Pinsky MR, Teboul JL. Assessment of indices of preload and volume responsiveness. Curr Opin Crit Care 2005; 11:235–239.
31 Pinsky MR. Hemodynamic monitoring in the intensive care unit. Clin Chest Med 2003; 24:549–560.
32 Sakka SG, Klein M, Reinhart K, et al. Prognostic value of extravascular lung water in critically ill patients. Chest 2002; 122:2080–2086.
33 Mitchell JP, Schuller D, Calandrino FS, Schuster DP. Improved outcome based on fluid management in critically ill patients requiring pulmonary artery catheterization. Am Rev Respir Dis 1992; 145:990–998.
34 Perkins GD, McAuley DF, Thickett DR, Gao F. The beta-agonist lung injury trial (BALTI): a randomized placebo-controlled clinical trial. Am J Respir Crit Care Med 2006; 173:281–287.
35 National Heart, Lung, and Blood Institute Acute Respiratory Distress Syndrome (ARDS) Clinical Trials Network, Wiedemann HP, Wheeler AP, Bernard GR, et al. Comparison of two fluid-management strategies in acute lung injury. N Engl J Med 2006; 354:2564–2575.
36 The Acute Respiratory Distress Syndrome Network. Ventilation with lower tidal volumes as compared with traditional tidal volumes for acute lung injury and the acute respiratory distress syndrome. N Engl J Med 2000; 342:1301–1308.
37 Brower RG, Lanken PN, MacIntyre N, et al. Higher versus lower positive end-expiratory pressures in patients with the acute respiratory distress syndrome. N Engl J Med 2004; 351:327–336.
38 Erlanger J, Hooker D. An experimental study of blood pressure and of pulse pressure in man. John Hopkins Hosp Rep 1904; 12:145–378.
39 Wesseling KH, Jansen JR, Settels JJ, Schreuder JJ. Computation of aortic flow from pressure in humans using a nonlinear, three-element model. J Appl Physiol 1993; 74:2566–2573.
40 Cholley BP, Shroff SG, Sandelski J, et al. Differential effects of chronic oral antihypertensive therapies on systemic arterial circulation and ventricular energetics in African-American patients. Circulation 1995; 91:1052–1062.
41 Linton NW, Linton RA. Estimation of changes in cardiac output from the arterial blood pressure waveform in the upper limb. Br J Anaesth 2001; 86:486–496.
42 Jonas M, Hett D, Morgan J. Real time, continuous monitoring of cardiac output and oxygen delivery. Int J Intensive Care 2002; 9:33–42.
43 Jonas MM, Tanser SJ. Lithium dilution measurement of cardiac output and arterial pulse waveform analysis: an indicator dilution calibrated beat-by-beat system for continuous estimation of cardiac output. Curr Opin Crit Care 2002; 8:257–261.
44 Irlbeck M, Forst H, Briegel J, et al. Continuous measurement of cardiac output with pulse contour analysis. Anaesthesist 1995; 44:493–500.
45 Tannenbaum GA, Mathews D, Weissman C. Pulse contour cardiac output in surgical intensive care unit patients. J Clin Anesth 1993; 5:471–478.
46 Weissman C, Ornstein EJ, Young WL. Arterial pulse contour analysis trending of cardiac output: hemodynamic manipulations during cerebral arteriovenous malformation resection. J Clin Monit 1993; 9:347–353.
47 Rodig G, Prasser C, Keyl C, et al. Continuous cardiac output measurement: pulse contour analysis vs thermodilution technique in cardiac surgical patients. Br J Anaesth 1999; 82:525–530.
48 Manecke GR, Auger WR. Cardiac output determination from the arterial pressure wave: clinical testing of a novel algorithm that does not require calibration. J Cardiothorac Vasc Anesth 2007; 21:3–7.
49 Stetz CW, Miller RG, Kelly GE, Raffin TA. Reliability of the thermodilution method in the determination of cardiac output in clinical practice. Am Rev Respir Dis 1982; 126:1001–1004.
50 Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986; 1:307–310.
51 Critchley LA, Critchley JA. A meta-analysis of studies using bias and precision statistics to compare cardiac output measurement techniques. J Clin Monit Comput 1999; 15:85–91.
52 Godje O, Thiel C, Lamm P, et al. Less invasive, continuous hemodynamic monitoring during minimally invasive coronary surgery. Ann Thorac Surg 1999; 68:1532–1536.
53 Mielck F, Buhre W, Hanekop G, et al. Comparison of continuous cardiac output measurements in patients after cardiac surgery. J Cardiothorac Vasc Anesth 2003; 17:211–216.
54 Zollner C, Haller M, Weis M, et al. Beat-to-beat measurement of cardiac output by intravascular pulse contour analysis: a prospective criterion standard study in patients after cardiac surgery. J Cardiothorac Vasc Anesth 2000; 14:125–129.
55 Felbinger TW, Reuter DA, Eltzschig HK, et al. Cardiac index measurements during rapid preload changes: a comparison of pulmonary artery thermodilution with arterial pulse contour analysis. J Clin Anesth 2005; 17:241–248.
56 Ostergaard M, Nielsen J, Rasmussen JP, Berthelsen PG. Cardiac output: pulse contour analysis vs. pulmonary artery thermodilution. Acta Anaesthesiol Scand 2006; 50:1044–1049.
57 Godje O, Hoke K, Goetz AE, et al. Reliability of a new algorithm for continuous cardiac output determination by pulse-contour analysis during hemodynamic instability. Crit Care Med 2002; 30:52–58.
58 Hamilton TT, Huber LM, Jessen ME. PulseCO: a less-invasive method to monitor cardiac output from arterial pressure after cardiac surgery. Ann Thorac Surg 2002; 74:S1408–S1412.
59 Goedje O, Hoeke K, Lichtwarck-Aschoff M, et al. Continuous cardiac output by femoral arterial thermodilution calibrated pulse contour analysis: comparison with pulmonary arterial thermodilution. Crit Care Med 1999; 27:2407–2412.
60 Della RG, Costa MG, Pompei L, et al. Continuous and intermittent cardiac output measurement: pulmonary artery catheter versus aortic transpulmonary technique. Br J Anaesth 2002; 88:350–356.
61 Della RG, Costa MG, Coccia C, et al. Cardiac output monitoring: aortic transpulmonary thermodilution and pulse contour analysis agree with standard thermodilution methods in patients undergoing lung transplantation. Can J Anaesth 2003; 50:707–711.
62 Fakler U, Pauli C, Balling G, et al. Cardiac index monitoring by pulse contour analysis and thermodilution after pediatric cardiac surgery. J Thorac Cardiovasc Surg 2007; 133:224–228.
63 Bein B, Worthmann F, Tonner PH, et al. Comparison of esophageal Doppler, pulse contour analysis, and real-time pulmonary artery thermodilution for the continuous measurement of cardiac output. J Cardiothorac Vasc Anesth 2004; 18:185–189.
64 Pittman J, Bar-Yosef S, SumPing J, et al. Continuous cardiac output monitoring with pulse contour analysis: a comparison with lithium indicator dilution cardiac output measurement. Crit Care Med 2005; 33:2015–2021.
65 Yamashita K, Nishiyama T, Yokoyama T, et al. Cardiac output by PulseCO is not interchangeable with thermodilution in patients undergoing OPCAB. Can J Anaesth 2005; 52:530–534.
66 Reid RD, Jayamaha J. The use of a cardiac output monitor to guide the initial fluid resuscitation in a patient with burns. Emerg Med J 2007; 24:e32.
67 Mayer J, Boldt J, Schollhorn T, et al. Semi-invasive monitoring of cardiac output by a new device using arterial pressure waveform analysis: a comparison with intermittent pulmonary artery thermodilution in patients undergoing cardiac surgery. Br J Anaesth 2007; 98:176–182.
68 Biais M, Nouette-Gaulain K, Cottenceau V, et al. Cardiac output measurement in patients undergoing liver transplantation: pulmonary artery catheter versus uncalibrated arterial pressure waveform analysis. Anesth Analg 2008; 106:1480–1486.
69 Mayer J, Boldt J, Wolf MW, et al. Cardiac output derived from arterial pressure waveform analysis in patients undergoing cardiac surgery: validity of a second generation device. Anesth Analg 2008; 106:867–872.
70 Prasser C, Trabold B, Schwab A, et al. Evaluation of an improved algorithm for arterial pressure-based cardiac output assessment without external calibration. Intensive Care Med 2007; 33:2223–2225.
71 Biancofiore G, Critchley LAH, Lee A, et al. Evaluation of an uncalibrated arterial pulse contour cardiac output monitoring system in cirrhotic patients undergoing liver surgery. Br J Anaesth 2009; 102:47–54.
72 Sakka SG, Kozieras J, Thuemer O, et al. Measurement of cardiac output: a comparison between transpulmonary thermodilution and uncalibrated pulse contour analysis. Br J Anaesth 2007; 99:337–342.
73 Harvey S, Young D, Brampton W, et al. Pulmonary artery catheters for adult patients in intensive care. Cochrane Database Syst Rev 2006:CD003408.
74 Uchino S, Bellomo R, Morimatsu H, et al. Pulmonary artery catheter versus pulse contour analysis: a prospective epidemiological study. Crit Care 2006; 10:R174.
75 de Wilde RB, Breukers RB, van den Berg PC, Jansen JR. Monitoring cardiac output using the femoral and radial arterial pressure waveform. Anaesthesia 2006; 61:743–746.
76 Orme RM, Pigott DW. Monitoring cardiac output from the radial artery pressure waveform. Anaesthesia 2007; 62:90–91.
77 Sander M, von HC, Foer A, et al. Pulse contour analysis after normothermic cardiopulmonary bypass in cardiac surgery patients. Crit Care 2005; 9:R729–R734.
78 Jellema WT, Wesseling KH, Groeneveld AB, et al. Continuous cardiac output in septic shock by simulating a model of the aortic input impedance: a comparison with bolus injection thermodilution. Anesthesiology 1999; 90:1317–1328.
79 Hamzaoui O, Monnet X, Richard C, et al. Effects of changes in vascular tone on the agreement between pulse contour and transpulmonary thermodilution cardiac output measurements within an up to 6-h calibration-free period. Crit Care Med 2008; 36:434–440.
80 Rivers E, Nguyen B, Havstad S, et al, Early Goal-Directed Therapy Collaborative Group. Early goal-directed therapy in the treatment of severe sepsis and septic shock. N Engl J Med 2001; 345:1368–1377.
81 Dellinger RP, Levy MM, Carlet JM, et al. Surviving Sepsis Campaign: international guidelines for management of severe sepsis and septic shock. Crit Care Med 2008; 36:296–327.
82 Reuter DA, Goepfert MS, Goresch T, et al. Assessing fluid responsiveness during open chest conditions. Br J Anaesth 2005; 94:318–323.
83 Michard F, Teboul JL. Predicting fluid-responsiveness in ICU patients: a critical analysis of the evidence. Chest 2002; 121:2000–2008.
84 Bendjelid K, Romand JA. Fluid responsiveness in mechanically ventilated patients: a review of indices used in intensive care. Intensive Care Med 2003; 29:352–360.
85 Morgan BC, Martin WE, Hornbein TF, et al. Hemodynamic effects of intermittent positive pressure respiration. Anesthesiology 1966; 27:584–590.
86 Jardin F, Farcot JC, Gueret P, et al. Cyclic changes in arterial pulse during respiratory support. Circulation 1983; 68:266–274.
87 Michard F. Changes in arterial pressure during mechanical ventilation. Anesthesiology 2005; 103:419–428.
88 Magder S. Clinical usefulness of respiratory variations in arterial pressure. Am J Respir Crit Care Med 2004; 169:151–155.
89 Perel A, Pizov R, Cotev S. Systolic blood pressure variation is a sensitive indicator of hypovolemia in ventilated dogs subjected to graded hemorrhage. Anesthesiology 1987; 67:498–502.
90 Preisman S, DiSegni E, Vered Z, Perel A. Left ventricular preload and function during graded haemorrhage and retranfusion in pigs: analysis of arterial pressure waveform and correlation with echocardiography. Br J Anaesth 2002; 88:716–718.
91 Pizov R, Ya'ari Y, Perel A. Systolic pressure variation is greater during hemorrhage than during sodium nitroprusside-induced hypotension in ventilated dogs. Anesth Analg 1988; 67:170–174.
92 Tavernier B, Makhotine O, Lebuffe G, et al. Systolic pressure variation as a guide to fluid therapy in patients with sepsis-induced hypotension. Anesthesiology 1998; 89:1313–1321.
93 Michard F, Boussat S, Chemla D, et al. Relation between respiratory changes in arterial pulse pressure and fluid responsiveness in septic patients with acute circulatory failure. Am J Respir Crit Care Med 2000; 162:134–138.
94 Reuter DA, Felbinger TW, Kilger E, et al. Optimizing fluid therapy in mechanically ventilated patients after cardiac surgery by on-line monitoring of left ventricular stroke volume variations: comparison with aortic systolic pressure variations. Br J Anaesth 2002; 88:124–126.
95 Rex S, Brose S, Metzelder S, et al. Prediction of fluid responsiveness in patients during cardiac surgery. Br J Anaesth 2004; 93:782–788.
96 Cannesson M, Musard H, Desebbe O, et al. The ability of stroke volume variations obtained with Flotrac/Vigileo™ system to monitor fluid responsiveness in mechanically ventilated patients. Anesth Analg 2009; 108:513–517.
97 Berkenstadt H, Margalit N, Hadani M, Friedman Z, et al. Stroke volume variation as a predictor of fluid responsiveness in patients undergoing brain surgery. Anesth Analg 2001; 92:984–989.
98 Berkenstadt H, Friedman Z, Preisman S, et al. Pulse pressure and stroke volume variations during severe haemorrhage in ventilated dogs. Br J Anaesth 2005; 94:721–726.
99 Marx G, Cope T, McCrossan L, et al. Assessing fluid responsiveness by stroke volume variation in mechanically ventilated patients with severe sepsis. Eur J Anaesthesiol 2004; 21:132–138.
100 Reuter DA, Kirchner A, Felbinger TW, et al. Usefulness of left ventricular stroke volume variation to assess fluid responsiveness in patients with reduced cardiac function. Crit Care Med 2003; 31:1399–1404.
101 Wiesenack C, Prasser C, Rodig G, Keyl C. Stroke volume variation as an indicator of fluid responsiveness using pulse contour analysis in mechanically ventilated patients. Anesth Analg 2003; 96:1254–1257.
102 De Backer D, Heenen S, Piagnerelli M, et al. Pulse pressure variations to predict fluid responsiveness: influence of tidal volume. Intensive Care Med 2005; 31:517–523.
103 Renner J, Cavus E, Meybohm P, et al. Stroke volume variation during hemorrhage and after fluid loading: impact of different tidal volumes. Acta Anaesthesiol Scand 2007; 51:538–544.
104 Perel A, Minkovich L, Preisman S, et al. Assessing fluid responsiveness by a standardized ventilatory maneuver: the respiratory systolic variation test. Anesth Analg 2005; 100:942–945.
105 Preisman S, Kogan S, Berkenstadt H, Perel A. Predicting fluid responsiveness in patients undergoing cardiac surgery: functional haemodynamic parameters including the Respiratory Systolic Variation Test and static preload indicators. Br J Anaesth 2005; 95:746–755.
106 Kubitz JC, Reuter DA. Using heart–lung interactions for functional haemodynamic monitoring: important factors beyond preload. In: Vincent JL, editor. Yearbook of intensive care and emergency medicine. Berlin, Heidelberg: Springer-Verlag; 2007. pp. 511–520.
107 Pinsky MR. Probing the limits of arterial pulse contour analysis to predict preload responsiveness. Anesth Analg 2003; 96:1245–1247.
108 Kubitz JC, Annecke T, Forkl S. Validation of pulse contour derived stroke volume variation during modifications of cardiac afterload. Br J Anaesth 2007; 98:591–597.
109 Coudray A, Romand JA, Treggiari M, Bendjelid K. Fluid responsiveness in spontaneously breathing patients: a review of indexes used in intensive care. Crit Care Med 2005; 33:2757–2762.
110 Perner A, Faber T. Stroke volume variation does not predict fluid responsiveness in patients with septic shock on pressure support ventilation. Acta Anaesthesiol Scand 2006; 50:1068–1073.
111 Heenen S, De Backer D, Vincent JL. How can the response to volume expansion in patients with spontaneous respiratory movements be predicted? Crit Care 2006; 10:R102.
112 Soubrier S, Saulnier F, Hubert H, et al. Can dynamic indicators help the prediction of fluid responsiveness in spontaneously breathing critically ill patients? Intensive Care Med 2007; 33:1117–1124.
113 Rutlen DL, Wackers FJT, Zaret BL. Radionuclide assessment of peripheral intravascular capacity: a technique to measure intravascular volumes changes in the capacitance circulation in man. Circulation 1981; 64:146–152.
114 Boulain T, Achard J, Teboul J, et al. Changes in BP induced by passive leg raising predict response to fluid loading in critically ill patients. Chest 2002; 121:1245–1252.
115 Monnet X, Rienzo M, Osman D, et al. Passive leg raising predicts fluid responsiveness in the critically ill. Crit Care Med 2006; 34:1402–1407.
116 Ridel C, Lamia B, Monnet X, et al. Passive leg raising and fluid responsiveness during spontaneous breathing: pulse contour evaluation. Intensive Care Med 2007; 32:S81 (abstract).

analysis; cardiac output; history; methods; methods; monitoring; physiologic

© 2009 European Society of Anaesthesiology