Secondary Logo

Journal Logo

Machine perfusion in organ transplantation: a tool for ex-vivo graft conditioning with mesenchymal stem cells?

Van Raemdonck, Dirka,b; Neyrinck, Arnec; Rega, Filipd; Devos, Timothye; Pirenne, Jacquesf

Current Opinion in Organ Transplantation: February 2013 - Volume 18 - Issue 1 - p 24–33
doi: 10.1097/MOT.0b013e32835c494f
STEM CELL TRANSPLANTATION: Edited by Elke Eggenhofer
Free

Purpose of review Machine perfusion has emerged as a tool to evaluate pretransplant graft function more objectively during preservation. Machine perfusion also offers the possibility to recondition questionable organs and to ‘immunomodulate’ allografts ex vivo. This article aims to review the current knowledge on machine perfusion of the various solid thoracic and abdominal organs, and to discuss the new possibility of conditioning and treating grafts with mesenchymal stem cells (MSCs) during machine perfusion.

Recent findings Different methods of machine perfusion have been described varying among organs in temperature and composition of perfusate. Commercial devices have recently become available for machine perfusion of all organs, with the largest clinical experience acquired in kidney and lung transplantation. Clinical studies are ongoing for liver, heart, and pancreas. MSC therapy in organ transplantation is now emerging with clinical studies set up to investigate its potential to attenuate ischemia/reperfusion injury (innate immunity) and to downregulate the alloimmune response (adaptive immunity) and promote engraftment after transplantation. We hypothesize that delivery of MSCs directly into the machine perfusion circuit may provide a unique opportunity to treat and immunomodulate organs prior to transplantation. To our knowledge, no study on ex-vivo delivery of MSCs during machine perfusion has been reported.

Summary Machine perfusion of solid organs has regained much attention during the last decade. It provides a new promising tool that may allow pretransplant ex-vivo assessment, preservation, repair, and conditioning of grafts. Experimental research and clinical trials testing the administration of MSCs during machine perfusion are warranted to explore the potential benefit and mechanisms of this approach.

aDivision of Experimental Thoracic Surgery, KU Leuven

bChair of the Transplant Centre, University Hospitals Leuven

cDivision of Anesthesiology and Algology

dDivision of Experimental Cardiac Surgery

eLaboratory for Experimental Transplantation

fLaboratory of Abdominal Transplant Surgery, KU Leuven, Leuven, Belgium

Correspondence to Dirk E.M. Van Raemdonck, Department of Thoracic Surgery, University Hospital Gasthuisberg, Herestraat 49, B-3000 Leuven, Belgium. Tel: +32 16 34 68 23; fax: +32 16 34 68 24; e-mail: dirk.vanraemdonck@uzleuven.be;web:www.dirkvanraemdonck.be

Back to Top | Article Outline

INTRODUCTION

Solid-organ transplantation has become a standard life-saving therapy in selected patients suffering from end-stage organ failure. In addition, organ transplantation remarkably ameliorates the quality of life. The medical community has come a long way in organ transplantation to where we are today [1]. The application of this ultimate treatment modality is only limited by the number of ‘acceptable’ organ donors and ‘transplantable’ grafts.

Static or simple cold storage (SCS) with conventional preservation solutions was designed at a time when donors were ‘ideal’ and preservation periods were short. In an era of donor shortage, increased use of suboptimal grafts, and organ exchange across sometimes distant geographical areas, SCS has reached its limits [2]. In the last decade, old techniques dating from the early days of transplantation have reemerged whereby organs are continuously and dynamically perfused instead of being statically cold stored during their preservation. This technology has been substantially refined and new portable machines have been developed and introduced in the clinical arena to facilitate ex-vivo machine perfusion of solid organs. The purpose of this review is to summarize the current experience with machine perfusion of thoracic and abdominal organs, to give an overview of the currently available devices, and to discuss their clinical applications. In addition, the attractive but yet to be explored possibility of treating and conditioning solid organs ex vivo during machine perfusion with mesenchymal stem cells (MSCs) will be discussed.

Box 1

Box 1

Back to Top | Article Outline

MACHINE PERFUSION: CONCEPT AND METHODS

Machine perfusion consists of creating a flow through the organ generated by a pump in a circuit to recirculate a preservative solution at various temperatures through the vasculature. This continuous perfusion permits better penetration of preservation solutions within the organ, delivery of oxygen and nutrients to the parenchyma (in case the perfusate is oxygenated), and the removal of toxic metabolites (in case the perfusate is renewed or filtered). In addition to an improved preservation, machine perfusion may allow real-time monitoring of functional and biochemical performance of the graft prior to transplantation and may provide a tool to select ‘transplantable’ grafts, to improve their condition, to reduce the incidence of ischemia/reperfusion injury (IRI), and possibly to safely increase the preservation time.

Different methods have been investigated experimentally. Techniques may vary among solid organs. Various ranges of temperature have been used from hypothermic perfusion at 4°C to normothermic perfusion at 37°C. The former method is intended to preserve the organ in a low metabolic state, while the latter will keep the organ in a more physiologic and metabolically active state with the advantage that organ repair and reconditioning may become possible. Solutions used as perfusate also vary among various organs from low potassium crystalloid solutions to blood-based solutions (full blood or packed red blood cells) mixed with colloids (dextran) and albumin. The power to circulate the perfusate through the vasculature is created by a roller pump (occlusive) or a centrifugal pump (nonocclusive) that delivers a continuous flow or a pulsatile flow mimicking the physiologic variations in systolic and diastolic pressure. Organs can be preserved in resting mode (preservation) or functioning mode (assessment). With normothermic perfusion, organs will resume their function. Kidneys will excrete urine and clear creatinine, and the liver will produce bile and coagulation factors and will clear metabolites. Hearts can be perfused while fibrillating or beating building up pressures. Lungs can be perfused in an atelectatic state or while ventilated, allowing gas exchange.

During machine perfusion, different physiological parameters specific to the organ can be measured and various biochemical markers released in the perfusate or excreted (in urine or bile) can be analyzed to assess the viability and the functional state of the graft. The exact value of these markers to decide whether to accept or to decline an organ and to predict its functional performance after transplantation is not clear yet and needs to be further investigated. For the kidney, vascular resistance correlates with delayed graft function, but the predictive value is low [3]. For the liver, the vascular resistance does not seem to correlate with viability [4]. In normothermic machine perfusion, the group of Friend at Oxford University has shown that the capacity of the liver to correct metabolic acidosis during machine perfusion correlates well with its function after transplant [5]. For the lung, oxygenation capacity taken alone may be misleading in assessing the ex-vivo lung. Other parameters like pulmonary vascular resistance, compliance, and airway pressures are equally important [6].

Several smaller and transportable commercial devices for machine perfusion have recently become available for all organs facilitating the clinical application of this new technique in the daily transplant practice. Randomized clinical trials with these devices have been completed (kidney), are ongoing (lung, heart), or are planned (liver). Costeffectiveness of machine perfusion needs to be analyzed for every organ in order to justify reimbursement by health authorities and insurance companies. Machine perfusion of kidney – by reducing the need for dialysis after transplant and improving graft survival – has already been shown to be costeffective [7▪▪].

Further research in all solid organs is ongoing with regard to the optimal perfusion technique and solutions.

Back to Top | Article Outline

MACHINE PERFUSION: EXPERIENCE AND DEVICES FOR CLINICAL USE

A thorough review summarizing current experimental knowledge and clinical experience with pretransplant graft viability assessment during machine perfusion for all solid organs was recently published [8▪▪].

Machine perfusion of kidneys historically has gained the largest interest as a prototype machine was developed in the late 1960s by Belzer. Several clinical studies including a large randomized trial [9,10,11▪], but not all [12], have shown a reduced rate of delayed graft function and better graft survival after hypothermic machine perfusion versus SCS. Machine perfusion is now believed to be the preferred method for preserving kidneys at higher risk originating from expanded criteria donors [13], donors older than 65 years [14], and donors after circulatory death [15]. The predictive capacity of machine perfusion parameters (vascular resistance and perfusate biomarkers) is not sufficient to be used as a sole indicator to accept or discard a given kidney [16–18]. Successful outcome following normothermic machine perfusion using a red-cell-based solution has been recently published by the team of Leicester [19,20]. Several companies have now marketed commercial devices for clinical kidney perfusion (Fig. 1): LifePort Kidney Transporter (Organ Recovery Systems, Des Plaines, Illinois, USA); RM3 (Waters Medical Systems, Birmingham, Alabama, USA); and Kidney Assist (Organ Assist, Groningen, The Netherlands).

FIGURE 1

FIGURE 1

Machine perfusion of livers was first attempted in the late 1960s by Bretschneider and Starzl. More recently, hypothermic and normothermic machine perfusion of the liver have been investigated in an experimental setting [5,21–28]. So far, only one clinical study has been published comparing 20 adult liver recipients after hypothermic machine perfusion with a historically matched group of recipients after SCS with a reduction in early allograft dysfunction (5% versus 25%; respectively; P = 0.08) [29,30]. Feasibility of normothermic liver machine perfusion in man has yet to be proven. Several companies have now marketed commercial devices for clinical liver perfusion (Fig. 2): Metranormothermic perfusion (OrganOx Ltd, Oxford, UK); LifePort Liver Transporter (Organ Recovery Systems); Liver Assist (Organ Assist).

FIGURE 2

FIGURE 2

Machine perfusion of pancreas has remained largely unexplored. The majority of studies aimed to optimize the preservation of pancreata in order to increase the islet yield after cold preservation [31,32]. In a very small human study, machine perfusion of pancreata resulted in better islet yield and viability compared with SCS [33]. The same devices currently available for kidney perfusion can be used and adapted for the perfusion of human pancreata.

Ex-vivo lung perfusion (EVLP) was studied in historical studies as a method to assess the quality of the graft [34] and as a preservation technique during distant thoracic organ procurement [35]. The first successful transplant after EVLP was published by Steen et al.[36] in 2001 stimulating many research groups worldwide to further investigate the technique in animal models [37–40] and in discarded human lungs [41–43]. Much experimental work was carried out at the University of Toronto [44,45]. Successful transplantation of questionable lungs after EVLP has recently been reported by different groups worldwide [46–52,53▪]. Results of an international multicenter trial (Inspire) randomizing standard donor lungs to be preserved with SCS versus machine perfusion are awaited [54]. Several companies have now marketed commercial devices for clinical lung perfusion (Fig. 3): OCS Lung (Transmedics, Andover, Massachusetts, USA); Vivoline LS1 (Vivoline Medical, Lund, Sweden); Lung Assist (Organ Assist); and XPS (XVIVO Perfusion AB, Göteborg, Sweden).

FIGURE 3

FIGURE 3

Hypothermic machine perfusion of heart was reported in animal experiments [55–58]. Left ventricular function seems better preserved after hypothermic [59–61] and normothermic [62,63] machine perfusion compared with SCS. Clinical trials were started in 2007 in USA (PROCEED) [64] and Europe (PROTECT) [65], with only preliminary data reported so far. One company has marketed a commercial device for clinical heart perfusion (Fig. 4): OCS Heart (Transmedics).

FIGURE 4

FIGURE 4

Back to Top | Article Outline

EX-VIVO RESUSCITATION OF ORGANS

As discussed above, experimental and clinical data have already shown that machine perfusion creates a ‘window’ between procurement and transplantation during which functional performance and viability of the graft under optimal conditions can be evaluated. Data collected during this period of ex-vivo preservation may provide information that can help clinicians to predict the risk of dysfunction after transplantation and that can assist them in deciding to accept or discard a given organ for transplantation. Also, machine perfusion has the potential to better preserve the quality of the graft by sustaining continued metabolism (at higher temperatures and with oxygen), providing energy and nutrients, and removing toxic waste products (in case the perfusate is regularly renewed). In the future, machine perfusion may offer a tool for ex-vivo repair of the organs and improvement of their quality prior to transplantation and perhaps for ‘immunomodulation’ of these organs in order to protect them from innate immune responses (IRI) and adaptive immune responses (acute and chronic rejection) in the recipient.

Many organs, excluding those with fixed structural damage related to previous injuries or life-style habits such as smoking or alcohol abuse, are currently declined because of acute – albeit recoverable – injuries. Potential grafts may get injured by several hits during the whole transplantation process in the transition phase from donor to recipient: the initial insult leading to brain damage, resuscitation maneuvers with intubation and ventilation, autonomic storm and systemic inflammation following brain-stem death, warm ischemia in donors after circulatory death, organ manipulation and surgical trauma during procurement and organ extraction, the cold preservation period, the implantation process, and finally the reperfusion phase. Altogether, organ damage may result from direct mechanical trauma and contusion, hemorrhage, inflammation, and infection.

Once the organs are recovered from the deceased body, ex-vivo treatment during machine perfusion becomes possible. Normothermic perfusion providing oxygen and other metabolic substrates under physiological conditions appears to be the way forward to improve the viability of suboptimal grafts. Research is ongoing for all organs to look for ways to administer agents with the potential to reduce the injury and to restore organ functionality. The easiest strategy to deliver drugs directly to the organs is by including them into the perfusion solution or by injecting them into the afferent tubing running to the vasculature of the graft. Theoretically, different drugs according to the type of injury or even a combination of them (cocktail approach) could be administered at intervals during machine perfusion: antibacterial, antiviral, and antifungal agents to treat infection, anti-inflammatory molecules to block proinflammatory responses, vasodilating agents to improve perfusion of the microvasculature, fibrinolytic agents to dissolve microthrombi, high osmotic agents to remove interstitial edema, etc. An advantage of this isolated machine perfusion setting is that these drugs could be given at higher doses than in vivo as there is no risk to harm other organs. A restriction, however, may be that certain drugs cannot be metabolized in the ex-vivo circuit and therefore active components would have to be given. On the other hand, toxic metabolites may accumulate over time. Therefore, repeated renewal of the perfusate or insertion of filters in the ex-vivo circuit may become necessary. Finally, machine perfusion also offers the possibility of direct and targeted genetic intervention by using viral vectors or silencing RNA technology.

Compared with kidney, liver, heart, and pancreas, the lung can be considered as a privileged organ as it not only carries a vascular, but also a bronchial tree providing direct access to the entire parenchyma. In that way, active drugs can be delivered to the pulmonary graft by instillation or nebulization via the airway as already practiced nowadays in vivo. Also, the liver is different from other solid organs in that it carries a dual vascular system allowing drug delivery to the parenchyma by perfusion via either the hepatic artery (that preferentially vascularizes the biliary tree which is particularly susceptible to ischemia) or the portal vein (that preferentially vascularizes the liver parenchyma) or both. Flow competition between these two systems has been observed experimentally by researchers from our group [66]. The potential relevance of this finding for drug administration warrants further study.

The aforementioned treatment possibilities during machine perfusion are currently hypothetical and only very few papers have been published on organ therapy during machine perfusion.

In the kidney, Nicholson of the Leicester group investigated the effect of adding erythropoietin to the perfusion solution in a porcine model of isolated normothermic hemoperfusion. The authors found that erythropoietin promotes inflammatory cell apoptosis, drives inflammatory and apoptotic cells into tubular lumens. This eventually leads to inflammation clearance, renoprotection, and tissue remodeling through caspase-3 and IL-1β [67].

In the liver, our group has demonstrated in a pig transplant model (of donation after circulatory death) that a multifactorial biological modulation (cocktail approach) including administration of streptokinase and epoprostenol in the donor and administration of glycine, alpha1-acid glycoprotein, MAP kinase inhibitor, alpha-tocopherol, glutathione, and apotransferrin to the recipient eliminates primary nonfunction, improves liver function, and increases survival. Biochemically, TNF-alpha and redox-active iron were suppressed and biliary bile salt toxicity was reduced [68]. Furthermore, inflammation-regulating genes were suppressed [69]. It will be interesting to test the administration of this type of treatment (and others) not only in donors and recipients, but also directly to the graft ex-vivo during machine perfusion. Here too, normothermic liver perfusion may reveal superior to hypothermic perfusion as many of these interventions may require an active metabolism to operate.

In the lung, our group has previously investigated the role of the antioxidant N-acetyl cysteine nebulized in nonheart-beating donor pig lungs subjected to 3 h of warm ischemia. Functional performance [70] and inflammatory response [71] assessed during EVLP was attenuated compared to the nontreated control group. Aerosolized N-acetyl cysteine, therefore, has great potential to attenuate inflammation during EVLP. The Zurich group investigated the role of surfactant administration to porcine lungs injured by acid aspiration. Ex-vivo administration of surfactant via lavage resulted in improved graft function when compared with a control group [72]. The same group found that adding the fibrinolytic drug urokinase to the reperfusion solution resulted in improved graft function with decreased pulmonary vascular resistance and better oxygenation [73]. In a series of human donor lungs determined to be unsuitable for transplantation by the Toronto group, five lungs were subjected to 12 h of normothermic EVLP and treated by transbronchial gene therapy with the anti-inflammatory interleukin 10. Improvement in pulmonary function, restoration of alveolar–blood barrier integrity, and attenuation of lung inflammation were noticed [74].

Back to Top | Article Outline

EX-VIVO CONDITIONING OF ORGANS WITH MESENCHYMAL STEM CELLS

MSCs are multipotent self-renewing cells isolated from whole bone marrow. A paradigm shift has occurred in our concept of how cell therapies utilizing MSCs mediate their beneficial effects. It is now appreciated that, although MSCs can be described as having differentiation potential, their effector function is based less on in-situ differentiation, transdifferentiation, or fusion and more on paracrine effects and cross-talk with other cells within diseased tissues. Mechanistic hypotheses of MSCs as cell-based therapy are postulated on their immunomodulatory properties (interaction with the innate immunity and suppression of T-cell responses) and their ability to secrete soluble factors [75]. These properties of MSCs make them particularly interesting for use as a cellular therapy in solid-organ transplantation [76,77] and in tissue-engineered organ replacement [78] for currently intractable conditions. Experience with MSCs therapy in transplantation is emerging for all solid organs in an attempt to attenuate the severity of IRI, to help recovery from reperfusion injury, and to promote graft acceptance by reducing acute and chronic rejection. Machine perfusion offers a unique platform to selectively administer these MSCs directly into the donor organ, overcoming the issues of homing, trafficking, and safety. Especially, allogeneic MSCs are attractive because of their wide availability at the time of organ harvest. Autologous stem cells might be of less interest to modulate acute donor organ injury during machine perfusion as the isolation steps take longer time intervals and can never be planned in advance when a potential donor becomes available.

In kidney, it has been shown that administration of MSCs enhances recovery from ischemia–reperfusion-induced acute renal failure in rats [79]. Operating mechanisms are not known with certainty, but seem to involve anti-inflammatory properties and facilitation of tissue repair, and this through endocrine and paracrine interactions and cellular crosstalk between MSCs and dendritic cells, macrophages and other cell types. On the basis of their renoprotective properties and their capacity to promote tissue repair after injury, MSCs are considered as a new therapeutic tool in patients with acute kidney injury and ongoing trials are testing their effect [80]. The role of autologous MSCs as an induction therapy to promote graft acceptance has also been studied in a randomized controlled trial after living-related kidney transplantation. This landmark study demonstrated a lower incidence of acute rejection, a decreased risk of opportunistic infection, and a better estimated renal function at 1 year compared with the IL-2 receptor antibody treatment group [81].

In liver, MSCs have also been shown to accelerate recovery from IRI in rodents [82]. In liver transplantation, the role of MSCs as immunomodulators to reduce rejection is currently being investigated [83]. A phase I safety and feasibility trial is set up at the University Medical Center Regensburg in recipients receiving two doses of third-party multipotent adult progenitor cells (MAPCs) at day 1 and 3 after liver transplantation, in addition to a calcineurin-inhibitor-free immunosuppressive regimen with basiliximab, mycophenolic acid, and steroids [84]. Results are awaited before a phase II/III trial will be conducted to assess their clinical efficacy.

In intestine too, it has been shown that MSCs can help to control IRI in rodents [85]. This property of MSCs in addition to their protolerogenic properties makes them a particularly attractive strategy to improve outcome after intestinal transplantation, a procedure still defeated by a profound IRI and a vigorous rejection response.

In lung, much research was done by the group of Matthay at the University of California San Francisco in a model of acute lung injury comparable to donor lung injury. In cultured human alveolar type II cells damaged by a mixture of cytokines, these investigators demonstrated the ability of allogeneic human MSCs to restore the epithelial permeability that is needed to limit edema formation after lung transplantation [86]. In one study from the same research group, allogeneic MSCs were administered to human donor lungs declined for transplantation. In an isolated reperfusion model, MSCs were administered directly into the airways to study their treatment potential in acute lung injury [87]. Several basic anti-inflammatory and antibacterial properties have been attributed to MSCs that may be beneficial to restore lung injury in patients with acquired respiratory distress syndrome [88]. The spectrum of possible MSC-based therapies for acute lung injury includes both targeted intrapulmonary and intravascular administration during EVLP.

In heart, pretransplant donor infusion of MSCs has been shown to prolong the survival of a semi-allogeneic heart transplant in a mouse model through the generation of regulatory T cells [89].

Back to Top | Article Outline

FUTURE RESEARCH

Many issues related to MSCs therapy remain unresolved and warrant further research: precise operating molecular and cellular mechanisms, influence of the timing and the site of administration, effect of various type of cells (autologous versus donor-specific versus third party versus xenogenic), trafficking of these cells and their homing to particular organs and their transendothelial migration in the parenchyma, and the influence of local inflammation on this process.

With the development of advanced organ regenerative strategies, the use of molecular and cellular treatments to repair organs ex vivo is now becoming conceivable. Machine perfusion provides a unique tool to deliver these therapies directly to the organs while they remain physiologically perfused and metabolically active in an isolated circuit ex vivo.

From an experimental point of view, direct administration of MSCs to an organ during machine perfusion in an isolated circuit ex vivo provides a unique experimental platform to study the potential therapeutic effects of these cells and their operating mechanisms. Perfusing an organ directly with MSCs may allow to better study the trafficking of these cells and their homing to particular organs and the factors influencing this process, their effect on the immunogenicity of the perfused organ (MHC expression, effect on endothelial cells and passenger antigen-presenting cells), their effect on intragraft innate immunity and local inflammation (cytokine, adhesion molecule, chemokine expression, cellular infiltration etc.), and their mode of action. Administration of MSCs during machine perfusion carries exciting potentials: first, the possibility to promote repair of damage endured prior to preservation because of donor-related injury and second, the possibility to downregulate local intragraft inflammation and reduce graft immunogenicity in order to make grafts less vulnerable to IRI and to rejection in the recipient.

To our knowledge, no experimental data on ex-vivo treatment with MSCs during machine perfusion to improve graft quality have been published so far. Similarly, to our knowledge, there are no reported data on the possibility to administer MSCs cells during machine perfusion to render organs more resistant to IRI and to diminish acute or chronic rejection after transplantation. To explore these possibilities, further research testing the effect of the direct administration of MSCs during ex-vivo organ perfusion is urgently needed. If the above would prove to be possible, this may revolutionize the practice of solid-organ transplantation by increasing the number of transplantable grafts and by improving their function and facilitating their acceptance after transplant, thereby reducing the need for immunosuppression and its attending complications (toxicity, infection, and malignancies) [90,91]. Access to transplantation would increase and results of transplantation would improve.

Back to Top | Article Outline

CONCLUSION

Machine perfusion of solid organs has regained much attention during the last decade. In addition to better preservation, it provides a promising tool that may allow ex-vivo assessment, repair, and conditioning of donor organs prior to transplantation. The largest clinical experience accumulated so far has been obtained in kidney and lung transplantation. It is hoped that this new technique in the near future will contribute to maximize the number of available organs for transplantation, to decrease the risk of primary graft dys(non)function after transplantation, and to improve long-term graft and patient survival. Further technological developments and research on the optimal technique and preservation solutions for long-term, ex-vivo organ perfusion are needed before it will become an established technique for all organs in the daily transplant practice.

The use of MSCs to treat human diseases has gained much attention because of their proregenerative, anti-inflammatory, and immunomodulatory properties. This is an emerging field and further experimental research and clinical trials are warranted. Early experience in the field of transplantation suggests that MSCs – when they are administered in vivo – ameliorate IRI by downregulating innate immunity, favor tissue repair at sites of inflammation, and may promote graft acceptance by downregulating the adaptive immunity. Machine perfusion provides a unique experimental platform to better study the function and properties of MSCs and a unique therapeutic tool to directly ‘treat’ an organ while it is perfused and maintained in a metabolically active state in an isolated ex-vivo circuit. Further studies are warranted to determine whether this strategy may favor repair of higher risk grafts that have been damaged prior to preservation and may condition and protect grafts from subsequent inflammatory and immune insults after transplantation (e.g. IRI and rejection).

Back to Top | Article Outline

Acknowledgements

The authors thank Diethard Monbaliu, MD, PhD, and Ina Jochmans, MD, PhD, for help with references.

The authors appreciate the help from Luanne Rodgers, Peter J. Friend, MD, Arjan van der Plaats, Maria Sipos and Waleed Hassanein, MD, with illustrations of the various machine perfusion devices.

Back to Top | Article Outline

Conflicts of interest

There are no conflicts of interest. Funding: The first author is a senior clinical investigator supported by the grant FWO-Vlaanderen G.3C04.99 (Fund for Scientific Research-Flanders).

Disclosures: Dirk Van Raemdonck is a consultant for Transmedics, Andover, MA, currently acting as principal investigator of the Expand Registry. He received research support from VitroLife, Göteborg, Sweden in the past. Jacques Pirenne is recipient of an unrestricted grant from Astellas and Roche, Belgium.

Back to Top | Article Outline

REFERENCES AND RECOMMENDED READING

Papers of particular interest, published within the annual period of review, have been highlighted as:

  • ▪ of special interest
  • ▪▪ of outstanding interest

Additional references related to this topic can also be found in the Current World Literature section in this issue (p. 115).

Back to Top | Article Outline

REFERENCES

1. Ortega F. Organ transplantation in the 21st century. Adv Exp Med Biol 2012; 741:13–26.
2. Pirenne J. Time to think out of the (ice) box. Curr Opin Organ Transplant 2010; 15:147–149.
3. Jochmans I, Moers C, Ploeg R, Pirenne J. To perfuse or not to perfuse kidneys donated after cardiac death. Am J Transplant 2011; 11:409–410.
4. Monbaliu D, Liu Q, Libbrecht L, et al. Preserving morphology and evaluating quality of liver grafts by hypothermic machine perfusion, a proof of concept study using discarded human livers. Liver Transplant 2012; DOI: 10.1002/lt.23550.
5. Vogel T, Brockmann JG, Friend PJ. Ex-vivo normothermic liver perfusion. Curr Opin Organ Transplant 2010; 15:167–172.
6. Yeung JC, Cypel M, Machuca TN, et al. Physiologic assessment of the ex vivo donor lung for transplantation. J Heart Lung Transplant 2012; 31:1120–1126.
7▪▪. Groen H, Moers C, Smits JM, et al. Cost-effectiveness of hypothermic machine perfusion versus static cold storage in renal transplantation. Am J Transplant 2012; 12:1824–1830.

This is the first study that compared cost–effectiveness of hypothermic machine perfusion versus cold storage.

8▪▪. Balfoussia D, Yerrakalva D, Hamaoui K, Papalois V. Advances in machine perfusion graft viability assessment in kidney, liver, pancreas, lung, and heart transplant. Exp Clin Transplant 2012; 10:87–100.

This study summarizes the current experimental knowledge and clinical experience with graft viability assessment of all solid organs during machine perfusion prior to transplantation.

9. Schold JD, Kaplan B, Howard RJ, et al. Are we frozen in time? Analysis of the utilization and efficacy of pulsatile perfusion in renal transplantation. Am J Transplant 2005; 5:1681–1688.
10. Moers C, Smits JM, Maathuis MH, et al. Machine perfusion or cold storage in deceased-donor kidney transplantation. N Engl J Med 2009; 360:7–19.
11▪. Moers C, Pirenne J, Paul A, Ploeg RJ. Machine perfusion or cold storage in deceased-donor kidney transplantation. N Engl J Med 2012; 366:770–771.

This study gives an update of the first publication [10] with results up to 3 years after transplantation.

12. Watson CJ, Wells AC, Roberts RJ, et al. Cold machine perfusion versus static cold storage of kidneys donated after cardiac death: a UK multicenter randomized controlled trial. Am J Transplant 2010; 10:1991–1999.
13. Treckmann J, Moers C, Smits JM, et al. Machine perfusion versus cold storage for preservation of kidneys from expanded criteria donors after brain death. Transplant Int 2011; 24:548–554.
14. Gallinat A, Moers C, Treckmann J, et al. Machine perfusion versus cold storage for the preservation of kidneys from donors ≥65 years allocated in the Eurotransplant Senior Programme. Nephrol Dial Transplant 2012 [Epub ahead of print].
15. Jochmans I, Moers C, Smits JM, et al. Machine perfusion versus cold storage for the preservation of kidneys donated after cardiac death: a multicenter, randomized, controlled trial. Ann Surg 2010; 252:756–764.
16. Jochmans I, Moers C, Smits JM, et al. Renal resistance during machine perfusion is a factor for delayed graft function and poorer graft survival. Am J Transplant 2010; 10 (Suppl. 4):107.
17. Moers C, Varnav OC, van Heurn E, et al. The value of machine perfusion perfusate biomarkers for predicting kidney transplant outcome. Transplantation 2010; 90:966–973.
18. Jochmans I, Pirenne J. Graft quality assessment in kidney transplantation: not an exact science yet!. Curr Opin Organ Transplant 2011; 16:174–179.
19. Hosgood SA, Nicholson ML. First in man renal transplantation after ex vivo normothermic perfusion. Transplantation 2011; 92:735–738.
20. Hosgood SA, Nicholson ML. Normothermic kidney preservation. Curr Opin Organ Transplant 2011; 16:169–173.
21. Monbaliu D, Brassil J. Machine perfusion of the liver: past, present, and future. Curr Opin Organ Transplant 2010; 15:160–166.
22. Monbaliu D, Heedfeld V, Liu Q, et al. Hypothermic machine perfusion of the liver: is it more complex than for the kidney? Transplant Proc 2011; 43:3445–3450.
23. Imber CJ, St Peter SD, Lopez de Cenarruzabeitia I, et al. Advantages of normothermic perfusion over cold storage in liver preservation. Transplantation 2002; 73:701–709.
24. Hessheimer AJ, Fondevilla C, Garcia-Valdecasas JC. Extracorporeal machine liver perfusion: are we warming up? Curr Opin Organ Transplant 2012; 17:143–147.
25. De Rougemont O, Breitenstein S, Leskosek B, et al. One hour hypothermic oxygenated perfusion (HOPE) protects nonviable liver allografts donated after cardiac death. Ann Surg 2009; 250:674–683.
26. Brockmann J, Reddy S, Coussios C, et al. Normothermic perfusion: a new paradigm for organ preservation. Ann Surg 2009; 250:1–6.
27. Fondevilla C, Hessheimer AJ, Maathuis MA, et al. Superior preservation of DCD livers with continuous normothermic perfusion. Ann Surg 2011; 254:1000–1007.
28. Jamieson RW, Zilvetti M, Roy D, et al. Hepatic steatosis and normothermic perfusion – preliminary experiments in a porcine model. Transplantation 2011; 92:289–295.
29. Guarrera JV, Henry SD, Samstein B, et al. Hypothermic machine preservation in human liver transplantation: the first clinical series. Am J Transplant 2010; 10:372–381.
30. Henry SD, Nachber E, Tulipan J, et al. Hypothermic machine preservation reduces molecular markers of ischemia/reperfusion injury in human liver transplantation. Am J Transplant 2012; 12:2477–2486.
31. Taylor MJ, Baicu S, Leman B, et al. Twenty-four hour hypothermic machine perfusion preservation of porcine pancreas facilitates processing for islet isolation. Transplant Proc 2008; 40:480–482.
32. Karcz M, Cook HT, Sibbons P, et al. An ex-vivo model for hypothermic pulsatile perfusion of porcine pancreata: hemodynamic and morphologic characteristics. Exp Clin Transplant 2010; 8:55–60.
33. Leeser DB, Bingaman AW, Poliakova L, et al. Pulsatile pump perfusion of pancreata before human islet cell isolation. Transplant Proc 2004; 36:1050–1051.
34. Jirsch DW, Fisk RL, Couves CM. Ex vivo evaluation of stored lungs. Ann Thorac Surg 1970; 10:163–168.
35. Hardesty RL, Griffith BP. Autoperfusion of the heart and lungs for preservation during distant procurement. J Thorac Cardiovasc Surg 1987; 93:11–18.
36. Steen S, Sjoberg T, Pierre L, et al. Transplantation of lungs from a nonheart-beating donor. Lancet 2001; 357:825–829.
37. Steen S, Liao Q, Wierup PN, et al. Transplantation of lungs from nonheart-beating donors after functional assessment ex vivo. Ann Thorac Surg 2003; 76:244–252.
38. Rega FR, Vandezande EJ, Jannis NC, et al. The role of leucocyte depletion in ex vivo evaluation of pulmonary grafts from (non)heart-beating donors. Perfusion 2003; 18 (Suppl. 1):13–21.
39. Erasmus ME, Fernhout MH, Elstrodt JM, et al. Normothermic ex vivo lung perfusion of nonheart-beating donor lungs in pigs: from pretransplant function analysis towards a 6-h machine preservation. Transpl Int 2006; 19:589–593.
40. Snell GI, Oto T, Levvey B, et al. Evaluation of techniques for lung transplantation following donation after cardiac death. Ann Thorac Surg 2006; 81:2014–2019.
41. Neyrinck A, Rega F, Jannis N, et al. Ex vivo reperfusion of human lungs declined for transplantation: a novel approach to alleviate donor organ shortage? J Heart Lung Transplant 2004; 23(S2):S173 [abstract].
42. Wierup P, Haraldsson A, Nilsson F, et al. Ex vivo evaluation of nonacceptable donor lungs. Ann Thorac Surg 2006; 81:460–466.
43. Egan TM, Haithcock JA, Nicotra WA, et al. Ex vivo evaluation of human lungs for transplant suitability. Ann Thorac Surg 2006; 81:1205–1213.
44. Cypel M, Yeung JC, Hirayama S, et al. Technique for prolonged normothermic ex vivo lung perfusion. J Heart Lung Transplant 2008; 27:1319–1325.
45. Cypel M, Rubacha M, Yeung J, et al. Normothermic ex vivo perfusion prevents lung injury compared to extended cold preservation for transplantation. Am J Transplant 2009; 9:2262–2269.
46. Ingemansson R, Eyjolfsson A, Mared L, et al. Clinical transplantation of initially rejected donor lungs after reconditioning ex vivo. Ann Thorac Surg 2009; 87:255–260.
47. Cypel M, Yeung J, Liu M, et al. Normothermic ex vivo lung perfusion in clinical lung transplantation. N Engl J Med 2011; 364:1431–1440.
48. Zych B, Popov AF, Stavri G, et al. Early outcomes of bilateral sequential single lung transplantation after ex-vivo lung evaluation and reconditioning. J Heart Lung Transplant 2012; 31:274–281.
49. Aigner C, Slama A, Hötzenecker K, et al. Clinical ex vivo lung perfusion – pushing the limits. Am J Transplant 2012; 12:1839–1847.
50. Moradiellos FJ, Naranjo JM, Cordoba M, et al. Clinical lung transplantation after ex vivo evaluation of uncontrolled non heart-beating donors lungs: initial experience. J Heart Lung Transplant 2011; 30(4S):S38 [abstract].
51. Wallinder A, Ricksten SE, Hansson C, et al. Transplantation of initially rejected donor lungs after ex vivo lung perfusion. J Thorac Cardiovasc Surg 2012; 144:1222–1228.
52. Warnecke G, Moradiellos J, Tudorache I, et al. Normothermic perfusion of donor lungs for preservation and assessment with the organ care system lung before bilateral transplantation: a pilot study of 12 patients. Lancet 2012; 380:1851–1858.
53▪. Sanchez PG, Bittle GJ, Burdorf L, et al. State of art: clinical ex vivo lung perfusion: rationale, current status, and future directions. J Heart Lung Transplant 2012; 31:339–348.

This review summarizes the past and present evidence, and provides important background for clinicians and investigators using the ex-vivo lung perfusion/isolated perfused lung system.

54. Warnecke G, Haverich A, Massard G, et al. The INSPIRE international lung trial with Organ Care System technology (OCS™). J Heart Lung Transplant 2012; 31 (4S):S115.
55. Proctor E, Parker R. Preservation of isolated heart for 72 h. Br Med J 1968; 4:296–298.
56. Wicomb WN, Cooper DK, Novitzky D, Barnard CN. Cardiac transplantation following storage of the donor heart by a portable hypothermic perfusion system. Ann Thorac Surg 1984; 37:243–248.
57. Wicomb WN, Novitzky D, Cooper DK, Rose AG. Forty-eight hours hypothermic perfusion storage of pig and baboon hearts. J Surg Res 1986; 40:276–284.
58. Wicomb WN, Cooper DK, Barnard CN. Twenty-four-hour preservation of the pig heart by a portable hypothermic perfusion system. Transplantation 1982; 34:246–250.
59. Rosenbaum DH, Peltz M, DiMaio JM, et al. Perfusion preservation versus static preservation for cardiac transplantation: effects on myocardial function and metabolism. J Heart Lung Transplant 2008; 27:93–99.
60. Ozeki T, Kwon MH, Gu J, et al. Heart preservation using continuous ex vivo perfusion improves viability and functional recovery. Circ J 2007; 71:153–159.
61. Stowe DF, Camara AK, Heisner JS, et al. Low-flow perfusion of guinea pig isolated hearts with 26 degrees C air-saturated Lifor solution for 20 h preserves function and metabolism. J Heart Lung Transplant 2008; 27:1008–1015.
62. Hassanein WH, Zellos L, Tyrrell TA, et al. Continuous perfusion of donor hearts in the beating state extends preservation time and improves recovery of function. J Thorac Cardiovasc Surg 1998; 116:821–830.
63. Garbade J, Krautz C, Aupperle H, et al. Functional, metabolic, and morphological aspects of continuous, normothermic heart preservation: effects of different preparation and perfusion techniques. Tissue Eng Part C Methods 2009; 15:275–283.
64. McCurry K, Jeevanandam V, Mihaljevic T, et al. Prospective multicenter safety and effectiveness evaluation of the organ care system device for cardiac use (PROCEED). J Heart Lung Transplant 2008; 27:S166 [abstract].
65. Tenderich G, El-Banayosy A, Rosengard B, et al. Prospective multicenter European trial to evaluate the safety and performance of the organ care system for heart transplants (PROTECT). J Heart Lung Transplant 2007; 26:S64 [abstract].
66. Monbaliu DR, Debbaut C, Hillewaert WJ, et al. Flow competition between hepatic arterial and portal venous flow during hypothermic machine perfusion preservation of porcine livers. Int J Artif Organs 2012; 35:119–131.
67. Yang B, Hosgood SA, Bagul A, et al. Erythropoietin regulates apoptosis, inflammation and tissue remodelling via caspase-3 and IL-1β in isolated hemoperfused kidneys. Eur J Pharmacol 2011; 660:420–430.
68. Monbaliu D, Vekemans K, Balligand E, et al. Multifactorial biological modulation of warm ischemia reperfusion injury in liver transplantation from nonheart-beating donors eliminates primary nonfunction and reduces bile salt toxicity. Ann Surg 2009; 250:808–817.
69. Vekemans K, Monbaliu D, Balligand E, et al. Improving the function of liver grafts exposed to warm ischemia by the Leuven drug protocol: exploring the molecular basis by microarray. Liver Transpl 2012; 18:206–218.
70. Rega FR, Wuyts WA, Vanaudenaerde BM, et al. Nebulized N-acetyl cysteine protects the pulmonary graft inside the nonheart-beating donor. J Heart Lung Transplant 2005; 24:1369–1377.
71. Geudens N, Wuyts WA, Rega FR, et al. N-acetyl cysteine attenuates the inflammatory response in warm ischemic pig lungs. J Surg Res 2008; 146:177–183.
72. Inci I, Ampollini L, Arni S, et al. Ex vivo reconditioning of marginal donor lung injured by acid aspiration. J Heart Lung Transplant 2008; 27:1229–1236.
73. Inci I, Zhai W, Arni S, et al. Fibrinolytic treatment improves the quality of lungs retrieved from nonheart-beating donors. J Heart Lung Transplant 2007; 26:1054–1060.
74. Cypel M, Liu M, Rubacha M, et al. Functional repair of human donor lungs by IL-10 gene therapy. Sci Transl Med 2009; 1:1–9.
75. English K, Mahon BP. Allogeneic mesenchymal stem cells: agents of immune modulation. J Cell Biochem 2011; 112:1963–1968.
76. Roemeling-van Rhijn M, Weimar W, Hoogduijn MJ. Mesenchymal stem cells: application for solid-organ transplantation. Curr Opin Organ Transplant 2012; 17:55–62.
77. Hoogduijn MJ, Popp FC, Grohnert A, et al. Advancement of mesenchymal stem cell therapy in solid organ transplantation (MISOT). Transplantation 2010; 90:124–126.
78. Baiguera S, Jungebluth P, Mazzanti B, Macchiarini P. Mesenchymal stroma cells for tissue-engineered tissue and organ replacements. Transplant Int 2012; 25:369–382.
79. Lange C, Tögel F, Ittrich H, et al. Administered mesenchymal stem cells enhance recovery from ischemia/reperfusion-induced acute renal failure in rats. Kidney Int 2005; 68:1613–1617.
80. Tögel F, Westenfelder C. Mesenchymal stem cells: a new therapeutic tool for AKI. Nat Rev Nephrol 2010; 6:179–183.
81. Tan J, Wu W, Xu X, et al. Induction therapy with autologous mesenchymal stem cells in living-related kidney transplants: a randomized controlled trial. JAMA 2012; 307:1169–1177.
82. Pan GZ, Yang Y, Zhang J, et al. Bone marrow mesenchymal stem cells ameliorate hepatic ischemia/reperfusion injuries via inactivation of the MEK/ERK signaling pathway in rats. J Surg Res 2012; 178:935–948.
83. Popp FC, Renner P, Eggenhofer E, et al. Mesenchymal cells as immunomodulators after liver transplantation. Liver Transpl 2009; 15:1192–1198.
84. Popp FC, Fillenberg B, Eggenhofer E, et al. Safety and feasibility of third-party multipotent adult progenitor cells for immunomodulation therapy after liver transplantation – a phase I study (MISOT-I). J Transl Med 2011; 9:124.
85. Jiang H, Qu L, Li Y, et al. Bone marrow mesenchymal stem cells reduce intestinal ischemia/reperfusion injuries in rats. J Surg Res 2011; 168:127–134.
86. Fang X, Neyrinck AP, Matthay MA, Lee JW. Allogeneic human mesenchymal stem cells restore epithelial protein permeability in cultured human alveolar type II cells by secretion of angiopoietin-1. J Biol Chem 2010; 20:26211–26222.
87. Lee JW, Fang X, Gupta N, et al. Allogeneic human mesenchymal stem cells for treatment of E. coli endotoxin-induced acute lung injury in the ex vivo perfused human lung. Proc Natl Acad Sci USA 2009; 106:16357–16362.
88. Gotts JE, Matthay MA. Mesenchymal stem cells and acute lung injury. Crit Care Med 2011; 27:719–733.
89. Casiraghi F, Azzolini N, Cassis P, et al. Pretransplant infusion of mesenchymal stem cells prolongs the survival of a semiallogeneic heart transplant through the generation of regulatory T cells. J Immunol 2008; 181:3933–3946.
90. Franquesa M, Hoogduijn MJ, Baan CC. The impact of mesenchymal stem cell therapy in transplant rejection and tolerance. Curr Opin Organ Transplant 2012; 17:355–361.
91. Casiraghi F, Noris M, Remuzzi G. Immunomodulatory effect of mesenchymal stromal cells in solid organ transplantation. Curr Opin Organ Transplant 2010; 15:731–737.
Keywords:

ex-vivo treatment; machine perfusion; mesenchymal stem cells; organ transplantation

© 2013 Lippincott Williams & Wilkins, Inc.