Correlation of Prosthetic Femoral Head Size and/or Design with Longevity of Total Hip Arthroplasty

RITTER, MERRILL, A.; STRINGER, ELIZABETH, A.; LITTRELL, DONNA, A.; WILLIAMS, JAMES, G.

Clinical Orthopaedics and Related Research: June 1983 - Volume 176 - Issue - p 252–257
II: Clinical Application: PDF Only
Free

Sixty-seven Mueller and 84 Charnley total hip arthroplasties were followed up for a minimum of seven years; pain, radiographic radiolucency, and prosthetic component loosening were evaluated. The patients who had Mueller total hip prostheses had significantly more pain at five (p < 0.021) and seven years (p < 0.001). No statistical differences were found between the two prosthetic groups in regard to incomplete or complete radiolucency or loosening of the femoral component, or incomplete or complete nonprogressive radiolucency of the acetabular component. The Mueller acetabular cup was loose in 15%, as compared with 4% for the Charnley group (p < 0.032). A survival analysis at five and seven years revealed that the success rate for Charnley total hip arthroplasties was 94% and 86%, respectively, as compared with 87% and 70%, respectively, for Mueller total hip arthroplasties (p < 0.003). This difference possibly is due to the large Mueller femoral head; if so, resurfaced hips would also experience earlier acetabular loosening.

Department of Medical Research, Methodist Hospital of Indiana, Inc. and Indiana University School of Medicine, Indianapolis, Indiana.