Reliability of home blood pressure monitoring: in the context of validation and accuracy

Jung, Mi-Hyanga; Kim, Gee-Heea; Kim, Ji-Hoona; Moon, Keon-Woonga; Yoo, Ki-Donga; Rho, Tae-Hob; Kim, Chul-Mina

doi: 10.1097/MBP.0000000000000121
Analytical Methods and Statistical Analysis
Buy

Objective Home blood pressure (BP) monitoring offers clinically relevant information enriched with more abundant data. However, there are few studies addressing the reliability of home BP devices and the quality of its data. This study aimed to evaluate the current status of home BP devices in terms of validation and accuracy.

Participants and methods Hypertensive patients with automated upper arm-type BP devices were consecutively enrolled. First, the validation status of each device was assessed through the website. Next, the accuracy of the individual device was evaluated by comparing the mean BP values between the automated device and a mercury sphygmomanometer. Accuracy of the device was defined as having less than a 5 mmHg difference in the mean BP values.

Results A total of 212 individuals were analyzed; 38.7% (82 of 212) of the participants used validated devices and 85.4% (181 of 212) were accurate. Inaccuracy was more common with nonvalidated devices than validated devices [19.2% (n=25) vs. 7.3% (n=6), P=0.017]. The range of inaccuracy of the validated devices was 6–26 mmHg for the systolic BP and 6–11 mmHg for the diastolic BP.

Conclusion The present study showed that nonvalidated devices are used widely in clinical practice and a substantial portion is inaccurate. Therefore, recommendation of validated devices should be the first step. Furthermore, all devices need to be examined for accuracy before use irrespective of their validation status.

aDepartment of Internal Medicine, Division of Cardiology, St Vincent’s Hospital, College of Medicine, The Catholic University of Korea, Suwon

bDepartment of Internal Medicine, Division of Cardiology, Seoul St Mary’s Hospital, College of Medicine, The Catholic University of Korea, Seoul, Republic of Korea

Correspondence to Chul-Min Kim, MD, Department of Internal Medicine, Division of Cardiology, St Vincent’s Hospital, College of Medicine, The Catholic University of Korea, 93-6, Ji-dong, Paldal-ku, Suwon, Gyunggi-do 442-723, Republic of Korea Tel: +82 31 249 7139; fax: +82 31 247 7139; e-mail: cmkim@cmcnu.or.kr

Received December 8, 2014

Received in revised form February 18, 2015

Accepted March 12, 2015

Copyright © 2015 Wolters Kluwer Health, Inc. All rights reserved.