Validation of the Globalcare GCE603 automated blood pressure monitor for self-measurement according to the European Society of Hypertension International Protocol revision 2010 : Blood Pressure Monitoring

Journal Logo

Devices and Technology

Validation of the Globalcare GCE603 automated blood pressure monitor for self-measurement according to the European Society of Hypertension International Protocol revision 2010

Song, Chenga; Yu, Yangb; Lu, Bao-Chuana; Yan, Xi-Lingc

Author Information
Blood Pressure Monitoring 25(5):p 291-294, October 2020. | DOI: 10.1097/MBP.0000000000000461

Abstract

Objective 

To evaluate the accuracy of the automated oscillometric upper-arm blood pressure (BP) monitor Globalcare GCE603 for home BP monitoring according to the International Protocol of the European Society of Hypertension (ESH) revision 2010.

Methods 

BP was sequentially measured in 33 adult participants and compared with a standard mercury sphygmomanometer. A total of 99 comparison pairs were obtained and analyzed according to the ESH international protocol.

Results 

For the 33 eligible participants, the device achieved 80/99, 98/99 and 99/99 measurements of absolute differences between device and observers within 5, 10 and 15 mmHg for systolic BP (SBP) and 87/99, 98/99 and 99/99 for diastolic BP (DBP), respectively. The average device-observer difference was 1.01 ± 4.11 mmHg for SBP and −0.69 ± 3.56 mmHg for DBP, respectively. The number of participants with two or three of the device-observer difference within 5 mmHg was 28 for SBP and 30 for DBP, and there was no subject with none of the device-observer difference within 5 mmHg.

Conclusion 

According to the validation results on the basis of the ESH international protocol revision 2010, the Globalcare GCE603 automated upper-arm BP monitor can be recommended for home measurement in general population.

Copyright © 2020 Wolters Kluwer Health, Inc. All rights reserved.