Journal Logo

LETTERS TO THE EDITOR

Usefulness of the second-derivative curve of activated partial thromboplastin time on the ACL-TOP coagulation analyzer for detecting factor deficiencies

Tokunaga, Naoki; Inoue, Chihiro; Sakata, Toshiyuki; Kagawa, Kumiko; Abe, Masahiro; Takamatsu, Norimichi; Nakao, Takayuki; Doi, Toshio

Author Information
Blood Coagulation & Fibrinolysis: June 2016 - Volume 27 - Issue 4 - p 474-476
doi: 10.1097/MBC.0000000000000436

Prolongation of the activated partial thromboplastin time (APTT) may be caused by a factor deficiency, inhibitor of coagulation factors, lupus anticoagulant, liver dysfunction, or anticoagulants such as unfractionated heparin and direct oral anticoagulants. Waveform analyses of APTT assay results from automated coagulation analyzers have been used for assessing very low levels (<1 IU/dl) of factor VIII (FVIII) and factor IX (FIX) and the in-vivo clotting function in patients with various types of hemophilia A [1–3]. It was reported that the ACL-TOP series of hemostasis testing systems (Instrumentation Laboratory, Bedford, Massachusetts, USA), which monitor absorbance, can detect abnormalities not only in FVIII and FIX but also in factor XI (FXI) and factor XII (FXII) by analyzing the waveform of the second-derivative curve [4]. However, it is not clear what types of deficiency and what levels of each deficiency can be detected by the waveform of the second-derivative curve on an ACL-TOP system.

ACL-TOP analysis produces three types of curves. One is a curve showing the changes in absorbance observed during the measurement of APTT. The second is the first derivative of the absorbance, corresponding to the coagulation velocity. The third type is the second derivative of the absorbance, corresponding to the coagulation acceleration. It was reported that among APTT reagents including silica or ellagic acid as an activator of FXII, the reagent APTT-SP (Instrumentation Laboratory), which includes silica, is suitable for the detection of deficiencies of the intrinsic coagulation factors or lupus anticoagulant when using an ACL-TOP system [4]. In the present study, we sought to determine the usefulness of the second-derivative curve of APTT assayed by the combination of APTT-SP and ACL-TOP to detect deficiencies of the intrinsic coagulation factor.

We made mixtures using normal pooled plasma and deficient plasma of each coagulation factor including factor V, FVIII, FIX, FXI, and FXII, and then we analyzed the second-derivative curve of the APTT assay using APTT-SP. We used calibration plasma (Instrumentation Laboratory) as the normal pooled plasma. At the same time, we measured the levels of each coagulation factor using a one-stage APTT-based clotting assay and SynthASil (Instrumentation Laboratory) as an APTT reagent. All tests were performed on an ACL-TOP.

The results revealed two types of abnormal second-derivative curve: a shoulder type in which the peak appears as a notch, and a biphasic type with a double peak. The measured levels of each mixture and the types of the second-derivative curve are shown in Table 1. We detected biphasic abnormal curves for plasma deficient in factor V, FVIII, FIX, and FXI. We also detected a shoulder-type abnormal curve in plasma with FVIII and FIX levels of 2–10%. The shoulder-type abnormal curve was also found for plasma with calculated levels of factor V and FXI of 2–4 and 2%, respectively. We detected only a shoulder-type abnormal curve in the FXII-deficient plasma. In the same manner, we confirmed that there was nothing abnormal about the second-derivative curves for mixtures of FII, FVII, and factor X or in plasma derived from 30 healthy individuals. These results suggest that the analysis of the second-derivative curves of APTT on an ACL-TOP might be useful for detecting factor deficiencies in factor V, FVIII, FIX, FXI, and FXII – especially FVIII and FIX.

T1-23
Table 1:
Factor levels and type of second-derivative curve in the activated partial thromboplastin time assay

To investigate the effect of von Willebrand factor (vWF) on the second-derivative curve, we monitored the serial changes of waveforms in an APTT assay after desmopressin (1-deamino-8-D-arginine vasopressin; DDAVP) infusion treatment for a patient with acquired von Willebrand syndrome showing less than 10% of both vWF ristocetin cofactor activity and vWF antigen (Fig. 1) [5]. DDAVP is the treatment of choice not only for type I von Willebrand disease, but also for acquired von Willebrand syndrome [6] because it can induce the release of normal vWF from cellular compartments. After 15 min of DDAVP infusion for this patient, the levels of vWF and FVIII were greatly increased to 54 and 84%, respectively. In accord with the increase of vWF and FVIII levels, the abnormal biphasic peaks on the second-derivative curve for the patient's plasma disappeared. Also, in accord with the decreases in the vWF and FVIII levels, an abnormal curve (shoulder type) reappeared at 2 and 6 h after the DDAVP loading test.

F1-23
Fig. 1:
Serial changes of the ACL-TOP second-derivative curve and the levels of factor VIII and von Willebrand factor. (a) Before the desmopressin loading test. (b) After 15 min. (c) After 2 h. (d) After 6 h. An abnormal biphasic curve on the second-derivative curve (light blue) was detected in a patient with von Willebrand factor deficiency. A shoulder-type abnormal curve was detected at 2 and 6 h after the desmopressin loading test. APTT, activated partial thromboplastin time; FVIII, factor VIII; vWF, von Willebrand Factor.

To assess the direct effects of vWF on the waveform analysis, we added FVIII concentrates (ADVATE, a recombinant, human, full-length coagulation factor VIII) into the plasma derived from patients with acquired vWD or hemophilia A (Table 2). In case of acquired vWD, the abnormality in the curve did not disappear despite the increased level of FVIII. On the contrary, that for the plasma of hemophilia A disappeared with an increase in FVIII. These results indicate that the plasma level of vWF directly influences the second-derivative curve in the APTT assay. Moreover, the strength might be higher than that of FVIII.

T2-23
Table 2:
Effect of increased level of factor VIII on the second-derivative curve in the activated partial thromboplastin time assay in plasma derived from patients with acquired von Willebrand disease or hemophilia A

These results suggest that the waveform analysis of second-derivative curves of an APTT assay might be a useful tool not only to detect the deficiency of vWF, but also to assess the effect of DDAVP infusion treatment.

In conclusion, a waveform analysis of the second-derivative curve of an APTT assay with the combined use of APTT-SP and an ACL-TOP system might be useful not only for detecting deficiencies of intrinsic coagulation factor and vWF, but also for assessing the results of DDAVP infusion treatment for patients with acquired von Willebrand syndrome. However, further investigations using plasma derived from patients with factor deficiencies are necessary to assess the usefulness of waveform analyses by this system.

Acknowledgements

Conflicts of interest

There are no conflicts of interest.

References

1. Shima M, Thachil J, Nair SC, Srivastava A. Towards standardization of clot waveform analysis and recommendations for its clinical applications. J Thromb Haemost 2013; 11:1417–1420.
2. Shima M, Matsumoto T, Fukuda K, Kubota Y, Tanaka I, Nishiya K, et al. The utility of activated partial thromboplastin time (aPTT) clot waveform analysis in the investigation of hemophilia A patients with very low levels of factor VIII activity (FVIII:C). Thromb Haemost 2002; 87:436–441.
3. Siegemund T, Scholz U, Schobess R, Siegemund A. Clot waveform analysis in patients with haemophilia A. Hamostaseologie 2014; 34 (Suppl 1):S48–S52.
4. Solano C, Zerefa P, Bird R. A study of atypical APTT derivative curves on the ACL-TOP coagulation analyser. Int J Lab Hematol 2011; 33:67–78.
5. James PD, Lillicrap D. von Willebrand disease: clinical and laboratory lessons learned from the large von Willebrand disease studies. Am J Hematol 2012; 87:S4–11.
6. Federici AB. The use of desmopressin in von Willebrand disease: the experience of the first 30 years (1977–2007). Haemophilia 2008; 14:5–14.
Copyright © 2016 YEAR Wolters Kluwer Health, Inc. All rights reserved.