Secondary Logo

Journal Logo

Kidney Support/Dialysis/Vascular Access

Substitution-Free Hemodiafiltration

Lee, Kyungsoo*; Pino, Christopher J.; Humes, H. David*†

Author Information
doi: 10.1097/MAT.0b013e318260c5bb
  • Free

Abstract

The incidence of kidney disease is rapidly increasing worldwide,1 and maintenance hemodialysis (HD) has become an established protocol for treating patients with end-stage renal disease (ESRD). Remarkable improvements have been made in the technologies used for dialysis treatment, and they are matched by the evolution of various modalities for renal supportive treatments. Better outcomes achieved by convective treatment have encouraged the use of synthetic high-flux membranes in clinical setups,2 and hemodiafiltration (HDF) and volume-controlled high-flux HD are now regarded as preferred forms of convective therapy.

High-flux HD adequately clears mid-size solutes without sterile fluid infusion, because forward filtration exceeding desired volume removal is compensated for by backfiltration (BF)3; thus, HD has been entitled to a simplest form of dialysis treatment. The convective dose delivered during high-flux HD has been shown to reduce mortality in patients at risk, as defined by a serum albumin level of <4 g/dl,4 but overall patient survival remains comparable to that of low-flux HD.5 This is presumably due to the limited amount of internal filtration involved. In contrast, HDF, which describes an intermittent renal supportive therapy of combined simultaneous diffusive and convective solute transport, is characterized by a large filtration volume that far exceeds the desired amount. Hence, this modality has been reported to improve middle-to-large size molecular removal, allow better erythropoietin control, and reduce oxidative stress and inflammation.6–9 It has been even reported to have a positive influence on the survival.10 However, the far higher filtration volumes during HDF must be compensated for by external substitution in real time.

Despite various modifications of HDF techniques based on infusion modes, the inevitable complexities associated with the requirement for the exogenous substitution infusion are still unavoidable. Therefore, efforts have been continued to propose novel HDF strategies that do not require external infusion. This is achieved by spontaneous fluid reinfusion at a rate that matches convection. These modifications can be classified into three developmental categories: 1) to increase the internal filtration rate by increasing pressure gradients along the hemodialyzer, 2) to use independent domains of ultrafiltration and diffusion, and 3) to alternate forward and backward filtration procedures.

In this review, therefore, the trials on substitution-free HDF are described. Technical aspects, including in vivo and in vitro efficacies and applicability to clinical use, are discussed in detail.

Internal Filtration–Enhanced HDF

Internal filtration is defined as the total water flux across membranes within the closed blood and dialysate compartments of a dialyzer.11 Volume-controlled high-flux HD uses the internal filtration phenomenon, and convective mass transfer during HD is governed by the balance of forward and backward filtration. Thus, augmenting internal filtration provides a straightforward means of achieving enhanced convection, which is termed iHDF. The amount of internal filtration is directly regulated by pressure gradients through the hemodialyzer. A pressure drop (ΔP) is inevitable as fluid flows through a cylindrical tube, and it is expressed by Poiseuille’s equation:

where µ is the fluid viscosity, L and d are the length and diameter of the flow path, respectively, and Q is the flow rate. However, because blood and dialysate flow in opposite directions, these pressure drops occur with opposing gradients, and in some regions hydraulic blood and dialysate pressures overlap (Figure 1). In a normal countercurrent dialysis setup, the sum of hydraulic and osmotic pressures and transmembrane pressure (TMP) is positive in the proximal region of a hemodialyzer, which leads to forward filtration. However, fluid movement occurs in the opposite direction in the distal region because TMP becomes negative.

Figure 1
Figure 1:
Blood and dialysate pressure gradient along hemodialyzer length. The sum of hydraulic and osmotic pressures is termed TMP, as expressed by the equation TMP = ΔPb − ΔPd −Δπ (ΔPb, ΔPd, and Δπ, hydraulic blood pressure, hydraulic dialysate pressure, and osmotic pressure gradient through the hemodialyzer, respectively). TMP, transmembrane pressure.

Resulting TMP gradients can be readily increased by increasing blood or dialysate pressure drops.12,13 Poiseuille’s equation shows that the pressure drop is inversely proportional to the fourth power of tube diameter. Hence, many investigations of iHDF have focused on using hemodialyzers with smaller membrane diameters so as to increase blood pressure drop. In early clinical studies, a 175 µm lumen diameter dialyzer enhanced beta-2 microglobulin (β2M) clearances by two and four times, respectively, over 200 and 250 µm diameter hemodialyzers.11 Clearances of inulin and vitamin b12 were also significantly greater with 175 µm dialyzer than a 200 µm dialyzer, without changing the clearances of low-molecular-weight solutes.14 These results were also confirmed by an analytical model, in which myoglobin clearance was increased by 34% when a membrane of diameter 150 µm, rather than 200 µm, with the same surface area was used at the same blood flow rates.15,16 These dialytic benefits afforded by reducing membrane lumen diameter allow internal filtration–enhanced hemodialyzers to be used clinically.17,18 However, the underlying risk of hemoconcentration due to high levels of forward filtration may not be negligible. Pressure-driven filtration causes large volume losses from blood and promptly increases blood viscosity, which deteriorates membrane sieving and hydraulic capabilities, and inevitably diminishes membrane efficiency.

Dialysate pressure is also regulated by increasing the flow resistance on the dialysate stream, such as, increasing membrane packing density, lengthening the hemodialyzer, or physically restricting flow. In one study, β2M and alpha-1 microglobulin were shown to be better cleared by a 250-mm dialyzer than a 195-mm dialyzer.19 Analytical and experimental studies also revealed that myoglobin clearances were slightly improved when using a hemodialyzer with 71.3% packing density versus 52% or 60.1%.16 However, high packing densities cause substantial degrees of dialysate channeling and flow mismatch between blood and dialysate, which eventually leads to a loss of effective surface area and impairs the diffusion process.20 Flow visualization studies in a dialyzer with a high packing density (75%) reconfirmed this disproportionate flow pattern of dialysate and consequent reductions in urea clearance.21,22

The convective efficiency of iHDF may be quantified by evaluating internal filtration rates. The flux across a membrane can be expressed by the product of membrane hydraulic permeability and TMP. However, the flow dynamics inside the hemodialyzer are so complex that precise determinations of internal filtration rates are not available clinically, and thus, fluxes and permeabilities become parameters beyond the operator’s control. Alternatively, a semiempirical model has been developed to determine internal filtration rates. Using this model, internal filtration volumes and reinfusion rates were determined during iHDF and postdilution HDF modes and revealed that differences between total convections (4.1 and 5.4 L/h for iHDF and HDF, respectively) well reflected differences between β2M clearances (123 ± 11 and 149 ± 26 ml/min for iHDF and HDF, respectively).17 In a study conducted to verify this semiempirical model, the model was found to show excellent accuracies of around 97% and a prediction error of only 3%.23

In addition to the mathematical model, methods for performing indirect measurements of the internal filtration have been proposed. Changes in nonpermeable molecular concentrations occur in response to the water content of blood, and thus, the kinetics of water transport across membrane can be evaluated by measuring the cumulative concentration changes of nonpermeable molecules.24 Radiolabeled albumin has been used to determine the amounts of convection for hemodialyzers with reduced fiber diameters or physical flow restriction of the dialysate stream.14,25 A series of in vitro experiments proved that this scintigraphic method was accurate for measuring internal filtration rates, but despite its precision, its clinical application is not plausible due to the safety concerns with the use of radiolabeled molecules and the added complexity required for the procedures and equipment. Another approach to determine internal filtration is offered by Doppler ultrasonography. In the absence of net filtration, blood volume depletion in the proximal portion of a hemodialyzer leads to a reduction in blood flow velocity, and after the lowest point has been reached, the blood velocity gradually increases due to BF. Thus, changes in blood velocity along a dialyzer provide information on blood volume changes. In one study, the internal filtration rate of a 250 mm dialyzer was found to be 37.7 ml/min, but only 11.1 ml/min for a standard 195 mm dialyzer.19 Doppler ultrasonography is noninvasive and easily used at bedside,26 but the method is still incapable of measuring blood flow velocity precisely; i.e., this method is based on velocities measured in peripheral membranes, which differ from velocities within centrally located fibers, and as a result, deviations from true values are unavoidable. Other techniques have also been explored in an effort to quantify the filtration phenomena or to visualize flow distributions inside hemodialyzers, such as magnetic resonance imaging,27 computed tomography,28,29 and a computerized scanning technique.30,31 However, the quantification of internal filtration using these techniques is not available clinically, due to concerns of patient safety and technical requirements.

In summary, iHDF can provide a straightforward means of convective treatment by increasing internal filtration rates using specifically designed hemodialyzers, and hence, this technique is simpler than other modalities. The literature suggests superior dialysis outcomes for iHDF versus standard HD mode, but the precise quantification of internal filtration remains to be determined. This information allows the comparison of this unit with other HDF techniques with respect to their convective capacities.

Double High-Flux HDF

Double high-flux HDF was first introduced in the early 1980s as a means of combining HD and hemofiltration (HF).32 This technique was particularly aimed at significantly increasing small (diffusion) and middle-size molecular removal (convection) to shorten overall treatment time, and therefore, much larger surface areas were used by arranging two high-flux hemodialyzers in series, in conjunction with a extremely high blood (400–500 ml/min) and dialysate flow rate (800–1000 ml/min), and a bicarbonate dialysate.33,34 This arrangement of two hemodialyzers enabled flow resistances through the two hemodialyzers to be manipulated, which permitted TMP gradients to be adjusted. Flow resistance applied to the dialysate tubing between the two dialyzers promptly increases dialysate pressures at the venous dialyzer (Figure 2). Hydraulic dialysate pressures exceed blood pressures, which leads to BF in the venous dialyzer. However, dialysate pressures rapidly fall in the arterial dialyzer due to flow restriction, which causes ultrafiltration in the arterial dialyzer. Hence, ultrafiltration at the arterial dialyzer can be promptly compensated for by BF at the venous dialyzer. In addition, the high blood and dialysate flow rates used are associated with larger pressure gradients of this technique.

Figure 2
Figure 2:
Schematic representation of the pressure profiles during double hemodiafiltration.

Therefore, together with these features, this modality achieved unmatched depurative outcomes, as demonstrated by far higher uremic molecular clearances regardless of molecular sizes.32,35–37 Furthermore, increased clearances allowed treatment times to be substantially shortened.33,34 These results were also confirmed by comparing treatment modalities. Double high-flux HDF attained significantly greater β2M reduction and Kt/V values than standard high-flux HD and showed comparable β2M clearance to that of online HDF.38,39 Furthermore, the beneficial effect of this technique on patient survival was also suggested in a long-term assessment. In this study, double HDF was compared with high-efficiency or high-flux HD modes in terms of treatment time, Kt/V, and standardized mortality ratio over 6 years. Kaplan-Meier survival analysis revealed a significantly lower mortality ratio for double HDF versus US Renal Data System (0.41 and 1, respectively) despite the shortened treatment time.40

However, concerns have been raised regarding the use of two hemodialyzers in this technique, such as possible increases in treatment cost and system complexity. One possible way of overcoming these issues involves the reuse of dialyzers, although regulatory guidelines on renal replacement practices in some countries do not permit the reuse. Another concern arises from the large amount of cross-membrane flux. In particular, a large quantity of BF should be assured by the strict and regular verification of water quality.41

Double high-flux HDF emerged as an effort to increase treatment efficiencies and shorten treatment times by maximizing diffusive and convective mass transfer. Many observations have confirmed the high solutes clearances across a wide spectrum of molecular weights, which are the results of the unique features of this unit. In particular, the unique control of hydraulic pressures possibly gives this unit the ability to regulate convective dose. However, the widespread implementation of this technique may require the identification of patients capable of tolerating treatment and the overcoming of the previously mentioned underlying concerns.

Paired Filtration Dialysis With Endogenous Hemofiltrate Reinfusion

Another two-chamber technique for obtaining efficient HDF treatment is the so-called paired filtration dialysis (PFD). Like double HDF, PFD is also a strategy of simultaneous HD and HF treatment aimed at increasing both diffusive and convective clearances, but its design principle separates convection from diffusion.42–44 A hemofilter with a small surface area is combined in a circuit with a hemodialyzer as shown in Figure 3. Ultrafiltration purely occurs at the hemofilter, and then blood is dialyzed continually through the hemodialyzer. The convection, which is not connected with diffusion, can minimize interactions between diffusion and convection and enable the precise quantification of convective clearance.45

Figure 3
Figure 3:
Hemofiltrate reinfusion circuit.

However, PFD still requires the exogenous substitution infusion, which is administered at the midpoint between the hemofilter and hemodialyzer, because of the ultrafiltrate exceeding those required. One unique feature of PFD is that the ultrafiltrate is not mixed with dialysate. However, the replacement fluid must possess a physiologic balance of electrolytes after taking into account preexisting deficits or excesses. These features of ultrafiltrate and infusate encouraged the regeneration of ultrafiltrate to replace exogenous infusate, and ultrafiltrate for replacement purposes was successfully regenerated using an uncoated charcoal column.46,47 As ultrafiltrate passes through the adsorbent column, solutes with a wide spectrum of molecular weights are adsorbed except for some small molecules (e.g., urea and phosphate), but electrolytes and bicarbonate freely pass through the column. In addition, because uncaptured small molecules can be removed by diffusion at the hemodialyzer, the regenerated ultrafiltrate is successfully applied as replacement fluid.48 Adsorption capacities were further increased by combining hydrophobic styrene-based resin along with uncoated charcoal, because the resin has a high binding affinity for several mid-molecular-weight species, such as β2M49 and homocysteine,50 or free immunoglobulin light chains.51

The other benefits of this regenerated ultrafiltrate include a better acid-base balance due to the reinfusion of endogenous bicarbonate52 and also the considerable advantage of combining high convection without physiologic molecule loss. Ultrafiltrate has a composition similar to that of plasma and contains huge numbers of polypeptides and other beneficial substances, such as hormones and amino acids,53 and regeneration of ultrafiltrate allows these beneficial nutrients to be reinfused.54 In terms of plasma amino acid levels, no significant changes in their intradialytic levels occur during hemofiltrate reinfusion (HFR), whereas a ~25% reduction occur during acetate-free biofiltration.55

A number of clinical studies on patients with ESRD have revealed that HFR remarkably improve dialytic efficiencies and solute removal, such as the removal of uremic marker molecules (β2M, leptin, and free immunoglobulin light chains),56–58 cardiovascular risk factors (homocysteine),50 inflammatory cytokines (C-reactive protein [CRP], interleukin [IL]-1, and IL-6), and biomarkers of oxidative stress (oxidized low-density lipoprotein and IL-1β).58,59 In a comparison between HFR and online HDF, both were found to be highly biocompatible and to considerably reduce inflammatory markers, such as CRP and IL-6.60 One technical variance of HFR is the repositioning of convection and diffusion. The change of sequence significantly enhanced reductions in urea and β2M, and also cytokine levels, e.g., IL6 and TNFα, more than conventional HFR, possibly due to the less saturated use of adsorbents.61,62 Contrary results have also been presented. For patients with ESRD treated with bicarbonate HD and then switched to HFR, nutritional and inflammatory parameters remained unchanged over a year. Neither serum β2M nor parathyroid hormone levels varied over the course of time, which led to the conclusion that although the change to HFR from bicarbonate HD is safe and tolerated, it is not associated with an improvement in nutritional or inflammatory parameters or a reduction in β2M levels.63 Prolonged, large-scale clinical studies remain to be conducted for HFR.

More recently, a significant decrease was observed in cardiac troponin (cTnT) level throughout HFR sessions when using acetate-free dialysate, but cTnT increased after HFR using dialysate containing acetate, which requires further explanation for the correlation between cTnT and acetate.64 However, both hemoglobin levels and erythropoietic-stimulating agent doses were not related to the presence of acetate during the 9 month HFR period.65

Adsorption as a third mechanism for purifying blood has been used in HFR units. Adsorption during HFR allows convective treatments to be performed by the endogenous reinfusion of ultrafiltrate. The ultrafiltrate reinfusion minimizes the loss of beneficial substances. Another feature of ultrafiltrate regeneration is the guaranteed purity of substitution fluid. Substitution is continuously obtained from ultrafiltrate, but the ultrafiltration, adsorption, and reinfusion system is totally closed during HFR, and therefore, excludes any possibility of contamination and ensures superior biocompatibility.

Push/Pull Techniques

Push/pull (PP) systems rely on alternate repetitions of forward and backward filtration during dialysis treatment. When the HDF technique using a serial arrangement of two hemodiafilters was described in the early 1980s, the PP concept was devised to eliminate the need for two filters. It is obvious that repetitive ultrafiltration can increase total filtration volume. However, such a system also requires a means of repeating BF.66

In early trials of PP HDF, a redundant dialysate bag was integrated downstream of the hemodialyzer, which was connected to the dialysate stream by a bidirectional peristaltic pump. The pump alternated the evacuation and replenishment of the bag, which forced hydrostatic pressures through the dialysate compartment to be oscillated. Likewise, another bag and an additional pump were integrated into the venous chamber and conducted the pulling and pushing of blood. However, disposable bags and separate PP pumps make this circuit considerably complicated. Hence, a double-chamber cylinder pump was devised to achieve the PP procedures and substantially simplify the required circuit (Figure 4). The cylinder pump includes two independent chambers and a reciprocal piston. Each chamber is connected to either dialysate or the blood stream.67 When the piston squeezes the chamber on the dialysate side, the dialysate compartment is pressurized and BF begins. At this time, the blood-side chamber expands and blood is filled in the venous chamber. Because the blood volume filled in the chamber is equal to the BF volume, blood flow returning to patients remains constant. The piston then moves in the opposite direction and squeezes the blood-side chamber; the dialysate compartment begins to expand and becomes depressurized, which leads to ultrafiltration. However, blood flow in the venous line is maintained, because the ultrafiltrate removed in the hemodialyzer is replenished in the venous chamber. In addition, the reciprocating movement of the piston is regulated by pressure differences between the two chambers so that the TMP is set at maximally optimal level, i.e., 400 and −400 mm Hg during forward and backward filtration phases, respectively.68

Figure 4
Figure 4:
Push/pull hemodiafiltration and a double-chamber cylinder pump.

This optimized use of TMP and frequent alternations of forward and backward filtration are obviously accompanied with markedly larger total filtration volumes and higher solutes clearances.68 The PP HDF unit tends to relieve symptoms such as arthralgia (joint pain), irritability, pruritus, and insomnia more rapidly than conventional HD mode.69–71 Furthermore, the optimal maintenance of membrane permeabilities by prompt BF has the added benefit of considerably inhibiting albumin loss in addition to increasing convection and diffusion.72 During PP HDF, backward flushing of dialysate takes place within the time frame required for the protein layer to fully develop, and thus, it can effectively wash out the inner lumen and inhibit excessive albumin leakage.

However, PP HDF still requires the use of a separate double-compartment piston pump to regulate dialysate pressures. Thus, a simpler strategy for achieving the PP actions has been recently introduced, which was implemented by using pulsatile devices for blood and dialysate.73–76 Respective pushing phases of blood and dialysate pulsations alternated, and TMPs cycled between positive and negative, which drove consecutive periods of ultrafiltration and BF; thus, the need for a separate double-compartment piston pump was eliminated. The hemodialytic efficiencies of the revised PP unit have been demonstrated in vitro and also in vivo, and these studies have shown that the devised PP unit substantially improves the clearances of uremic marker molecules, particularly for mid-sized molecules.73,77 Pressure profiles also showed obvious oscillations of TMPs throughout treatment, and their magnitudes were significantly larger than those observed in HD mode. More recently, due to concerns with the pulsatile circulation of blood, particularly in terms of blood supply and vascular access, this unit has been further improved by eliminating blood pulsation, but retaining the recurrent forward and backward filtration procedures, which was achieved by integrating a dual pulsatile device in the dialysate stream.78

Push/pull techniques described have been reported to deliver maximal permissible levels of volume exchange, which allows speculation on the capability of this unit in terms of convective efficiency. However, no clinical observation has been conducted to examine the long-term clinical effect of these PP techniques. In addition, the forward and backward filtration rates exceed blood flow rates,79 and the repeated BF and consequent blood dilution may translate into a dialytic efficiency drop. Nevertheless, the optimizations of PP procedures, in terms of the PP volumes and frequencies, may result in more efficient use of membranes, generate better convective efficiency, and as a result, achieve a simple but efficient substitution-free HDF.

Conclusions

Substitution-free HDF treatments must be accompanied by spontaneous fluid restoration, such as BF or ultrafiltrate regeneration (Table 1). A simpler way might be to increase internal filtration during HD sessions, although this can only be done to a limited extent. Much higher efficiencies can be achieved by the two-chamber techniques, which were developed in an effort to increase solute removal and shorten treatment times. However, these modalities appear to unavoidably increase overall system complexity and cost. Push/pull techniques were derived by considering phase separation for forward and backward filtration. In addition, the pulse PP scheme has adopted pulsatile devices for blood and dialysate circulation as an effort of modulating flow patterns. However, no clinical observation has been made using these PP units.

Table 1
Table 1:
Summary of Substitution-Free HDF Modalities

Modern dialysis machines offer HDF and HD as default therapies and are also equipped with outstanding monitoring facilities not only for patients but also for treatments.80 In particular, advances in water treatment allow ultrapure replacement fluid to be prepared in real time. These technical advancements are certainly lowering the barriers to higher convective HDF therapies. Therefore, the additional convective and diffusive clearances of these substitution-free HDF techniques should be translated into features that simplify overall dialysis treatments, such as a miniaturized unit that can be characterized as being user-friendly, light, and portable. A dialysis unit equipped with these features may also provide treatment alternatives beyond the current thrice weekly clinic dialysis for patients with ESRD.

References

1. Centers for Disease Control and Prevention. . Incidence of end-stage renal disease attributed to diabetes among persons with diagnosed diabetes—United States and Puerto Rico, 1996–2007. MMWR Morb Mortal Wkly Rep. 2010;59:1361–1366
2. Woods HF, Nandakumar M. Improved outcome for haemodialysis patients treated with high-flux membranes. Nephrol Dial Transplant. 2000;15(suppl 1):36–42
3. Ofsthun NJ, Leypoldt JK. Ultrafiltration and backfiltration during hemodialysis. Artif Organs. 1995;19:1143–1161
4. Locatelli F, Martin-Malo A, Hannedouche T, et al.Membrane Permeability Outcome (MPO) Study Group. Effect of membrane permeability on survival of hemodialysis patients. J Am Soc Nephrol. 2009;20:645–654
5. Eknoyan G, Beck GJ, Cheung AK, et al.Hemodialysis (HEMO) Study Group. Effect of dialysis dose and membrane flux in maintenance hemodialysis. N Engl J Med. 2002;347:2010–2019
6. Vaslaki L, Major L, Berta K, et al. On-line haemodiafiltration versus haemodialysis: Stable haematocrit with less erythropoietin and improvement of other relevant blood parameters. Blood Purif. 2006;24:163–173
7. Lornoy W, Becaus I, Billiouw JM, Sierens L, Van Malderen P, D’Haenens P. On-line haemodiafiltration. Remarkable removal of beta2-microglobulin. Long-term clinical observations. Nephrol Dial Transplant. 2000;15(suppl 1):49–54
8. Ward RA, Schmidt B, Hullin J, Hillebrand GF, Samtleben W. A comparison of on-line hemodiafiltration and high-flux hemodialysis: A prospective clinical study. J Am Soc Nephrol. 2000;11:2344–2350
9. Kuragano T, Kida A, Furuta M, et al. The impact of beta2-microglobulin clearance on the risk factors of cardiovascular disease in hemodialysis patients. ASAIO J. 2010;56:326–332
10. Canaud B, Bragg-Gresham JL, Marshall MR, et al. Mortality risk for patients receiving hemodiafiltration versus hemodialysis: European results from the DOPPS. Kidney Int. 2006;69:2087–2093
11. Dellanna F, Wuepper A, Baldamus CA. Internal filtration–advantage in haemodialysis? Nephrol Dial Transplant. 1996;11(suppl 2):83–86
12. Fiore GB, Ronco C. Internal hemodiafiltration (iHDF): A possible option to expand hemodiafiltration therapy. Int J Artif Organs. 2004;27:420–423
13. Fiore GB, Ronco C. Principles and practice of internal hemodiafiltration. Contrib Nephrol. 2007;158:177–184
14. Ronco C, Brendolan A, Lupi A, Metry G, Levin NW. Effects of a reduced inner diameter of hollow fibers in hemodialyzers. Kidney Int. 2000;58:809–817
15. Mineshima M, Ishimori I, Ishida K, et al. Effects of internal filtration on the solute removal efficiency of a dialyzer. ASAIO J. 2000;46:456–460
16. Mineshima M. New trends in HDF therapies: Validity of internal filtration-enhanced hemodialysis. Blood Purif. 2004;22(suppl 2):60–66
17. Lucchi L, Fiore GB, Guadagni G, et al. Clinical evaluation of internal hemodiafiltration (iHDF): A diffusive-convective technique performed with internal filtration enhanced high-flux dialyzers. Int J Artif Organs. 2004;27:414–419
18. Righetti M, Filiberti O, Ranghino A, et al. Internal hemodiafiltration versus low-flux bicarbonate dialysis: Results from a long-term prospective study. Int J Artif Organs. 2010;33:796–802
19. Sato Y, Mineshima M, Ishimori I, Kaneko I, Akiba T, Teraoka S. Effect of hollow fiber length on solute removal and quantification of internal filtration rate by Doppler ultrasound. Int J Artif Organs. 2003;26:129–134
20. Gastaldon F, Brendolan A, Crepaldi C, et al. Effects of novel manufacturing technology on blood and dialysate flow distribution in a new low flux “alpha Polysulfone” hemodialyzer. Int J Artif Organs. 2003;26:105–112
21. Yamamoto K, Hiwatari M, Kohori F, Sakai K, Fukuda M, Hiyoshi T. Membrane fouling and dialysate flow pattern in an internal filtration-enhancing dialyzer. J Artif Organs. 2005;8:198–205
22. Fujimura T, Uchi Y, Fukuda M, Miyazaki M, Uezumi S, Hiyoshi T. Development of a dialyzer with enhanced internal filtration to increase the clearance of low molecular weight proteins. J Artif Organs. 2004;7:149–154
23. Fiore GB, Guadagni G, Lupi A, Ricci Z, Ronco C. A new semiempirical mathematical model for prediction of internal filtration in hollow fiber hemodialyzers. Blood Purif. 2006;24:555–568
24. Ronco C, Brendolan A, Feriani M, et al. A new scintigraphic method to characterize ultrafiltration in hollow fiber dialyzers. Kidney Int. 1992;41:1383–1393
25. Ronco C, Orlandini G, Brendolan A, Lupi A, La Greca G. Enhancement of convective transport by internal filtration in a modified experimental hemodialyzer: technical note. Kidney Int. 1998;54:979–985
26. Mineshima M. Estimation of internal filtration flow rate in high-flux dialyzers by Doppler ultrasonography. Contrib Nephrol. 2011;168:153–161
27. Hardy PA, Poh CK, Liao Z, Clark WR, Gao D. The use of magnetic resonance imaging to measure the local ultrafiltration rate in hemodialyzers. J Memb Sci. 2002;204:195–205
28. Kim JC, Kim JH, Kim HC, et al. Effect of fiber structure on dialysate flow profile and hollow-fiber hemodialyzer reliability: CT perfusion study. Int J Artif Organs. 2008;31:944–950
29. Frank A, Lipscomb GG, Dennis M. Visualization of concentration fields in hemodialyzers by computed tomography. J Memb Sci. 2000;175:239–251
30. Ronco C, Brendolan A, Crepaldi C, Rodighiero M, Scabardi M. Blood and dialysate flow distributions in hollow-fiber hemodialyzers analyzed by computerized helical scanning technique. J Am Soc Nephrol. 2002;13(suppl 1):S53–S61
31. Ronco C, Brendolan A, Crepaldi C, et al. Dialysate flow distribution in hollow fiber hemodialyzers with different dialysate pathway configurations. Int J Artif Organs. 2000;23:601–609
32. Cheung AC, Kato Y, Leypoldt JK, Henderson LW. Hemodiafiltration using a hybrid membrane system for self-generation of diluting fluid. Trans Am Soc Artif Intern Organs. 1982;28:61–65
33. Miller JH, von Albertini B, Gardner PW, Shinaberger JH. Technical aspects of high-flux hemodiafiltration for adequate short (under 2 hours) treatment. Trans Am Soc Artif Intern Organs. 1984;30:377–381
34. von Albertini B, Miller JH, Gardner PW, Shinaberger JH. High-flux hemodiafiltration: Under six hours/week treatment. Trans Am Soc Artif Intern Organs. 1984;30:227–231
35. Shinzato T, Sezaki R, Usuda M, Maeda K, Ohbayashi S, Toyota T. Infusion-free hemodiafiltration: Simultaneous hemofiltration and dialysis with no need for infusion fluid. Artif Organs. 1982;6:453–456
36. von Albertini B, Miller JH, Gardner PW, Shinaberger JH. Performance characteristics of the hemoflow F 60 in high-flux hemodiafiltration. Contrib Nephrol. 1985;46:169–173
37. Pisitkun T, Eiam-Ong S, Tiranathanagul K, et al. Convective-controlled double high flux hemodiafiltration: A novel blood purification modality. Int J Artif Organs. 2004;27:195–204
38. Tiranathanagul K, Yossundharakul C, Techawathanawanna N, et al. Comparison of middle-molecule clearance between convective control double high-flux hemodiafiltration and on-line hemodiafiltration. Int J Artif Organs. 2007;30:1090–1097
39. Susantitaphong P, Tiranathanagul K, Katavetin P, et al. Efficacy of convective-controlled double high-flux hemodiafiltration versus on-line hemodiafiltration: 1-year prospective study. Blood Purif. 2010;29:35–43
40. Bosch JP, Lew SQ, Barlee V, Mishkin GJ, von Albertini B. Clinical use of high-efficiency hemodialysis treatments: Long-term assessment. Hemodial Int. 2006;10:73–81
41. Bosch JP, Mishkin GJ. Hemofiltration and double high flux dialysis: Risks and benefits. Curr Opin Nephrol Hypertens. 1998;7:643–647
42. Ghezzi PM, Sanz-Moreno C, Gervasio R, Nigrelli S, Botella J. Technical requirements for rapid high-efficiency therapy in uremic patients. Paired filtration-dialysis (PFD) with a two-chamber technique. ASAIO Trans. 1987;33:546–550
43. Ghezzi PM, Dutto A, Gervasio R, Botella J. Hemodiafiltration with separate convection and diffusion: Paired filtration dialysis. Contrib Nephrol. 1989;69:141–161; discussion 162
44. Ronco C, Feriani M, Brendolan A, et al. Paired filtration dialysis: Studies on efficiency, flow dynamics and hydraulic properties of the system. Blood Purif. 1990;8:126–140
45. Ghezzi PM, Zucchelli P, Ringoir S, et al. Paired filtration-dialysis (PFD): A separate convective-diffusive system for extracorporeal blood purification in uraemic patients. Adv Exp Med Biol. 1987;223:267–275
46. Ghezzi PM, Botella J, Sartoris AM, Gervasio R, Diez C. Use of the ultrafiltrate obtained in two-chamber (PFD) hemodiafiltration as replacement fluid. Experimental ex vivo and in vitro study. Int J Artif Organs. 1991;14:327–334
47. Ghezzi PM, Gervasio R, Tessore V, Sartoris AM, Botella J. Hemodiafiltration without replacement fluid. An experimental study. ASAIO J. 1992;38:61–65
48. Sanz-Moreno C, Botella J. Hemodiafiltration in two chambers without replacement fluid: A clinical study. Artif Organs. 1995;19:407–410
49. Marinez de Francisco AL, Ghezzi PM, Brendolan A, et al. Hemodiafiltration with online regeneration of the ultrafiltrate. Kidney Int Suppl. 2000;76:S66–S71
50. Splendiani G, De Angelis S, Tullio T, et al. Selective adsorption of homocysteine using an HFR-ON LINE technique. Artif Organs. 2004;28:592–595
51. Testa A, Dejoie T, Lecarrer D, Wratten M, Sereni L, Renaux JL. Reduction of free immunoglobulin light chains using adsorption properties of hemodiafiltration with endogenous reinfusion. Blood Purif. 2010;30:34–36
52. de Francisco AL, Botella J, Escallada R, et al. Haemodiafiltration with sorbent-regenerated ultrafiltrate as replacement fluid: A multicenter study. Nephrol Dial Transplant. 1997;12:528–534
53. Weissinger EM, Kaiser T, Meert N, et al. Proteomics: A novel tool to unravel the patho-physiology of uraemia. Nephrol Dial Transplant. 2004;19:3068–3077
54. La Greca G, Brendolan A, Ghezzi PM, et al. The concept of sorbent in hemodialysis. Int J Artif Organs. 1998;21:303–308
55. Borrelli S, Minutolo R, De Nicola L, et al. Intradialytic changes of plasma amino acid levels: Effect of hemodiafiltration with endogenous reinfusion versus acetate-free biofiltration. Blood Purif. 2010;30:166–171
56. Bolasco PG, Ghezzi PM, Ferrara R, et al. Effect of on-line hemodiafiltration with endogenous reinfusion (HFR) on the calcium-phosphorus metabolism: Medium-term effects. Int J Artif Organs. 2006;29:1042–1052
57. Kim S, Oh KH, Chin HJ, et al. Effective removal of leptin via hemodiafiltration with on-line endogenous reinfusion therapy. Clin Nephrol. 2009;72:442–448
58. Testa A, Gentilhomme H, Le Carrer D, Orsonneau JL. In vivo removal of high- and low-molecular-weight compounds in hemodiafiltration with on-line regeneration of ultrafiltrate. Nephron Clin Pract. 2006;104:c55–c60
59. Calò LA, Naso A, Davis PA, et al. Hemodiafiltration with online regeneration of ultrafiltrate: Effect on heme-oxygenase-1 and inducible subunit of nitric oxide synthase and implication for oxidative stress and inflammation. Artif Organs. 2011;35:183–187
60. Panichi V, Manca-Rizza G, Paoletti S, et al. Effects on inflammatory and nutritional markers of haemodiafiltration with online regeneration of ultrafiltrate (HFR) vs online haemodiafiltration: A cross-over randomized multicentre trial. Nephrol Dial Transplant. 2006;21:756–762
61. Meloni C, Ghezzi PM, Cipriani S, et al. One year of clinical experience in postdilution hemofiltration with online reinfusion of regenerated ultrafiltrate. Blood Purif. 2004;22:505–509
62. Meloni C, Ghezzi PM, Cipriani S, et al. Hemodiafiltration with post-dilution reinfusion of the regenerated ultrafiltrate: A new on-line technique. Clin Nephrol. 2005;63:106–112
63. Bossola M, Muscaritoli M, Tazza L, et al. Switch from bicarbonate hemodialysis to hemodiafiltration with online regeneration of the ultrafiltrate (HFR): Effects on nutritional status, microinflammation, and beta-microglobulin. Artif Organs. 2005;29:259–263
64. Bolasco P, Ghezzi PM, Serra A, et al.Sardinian Polycentric Study on Acetate-Free Haemodiafiltration. Effects of acetate-free haemodiafiltration (HDF) with endogenous reinfusion (HFR) on cardiac troponin levels. Nephrol Dial Transplant. 2011;26:258–263
65. Bolasco PG, Ghezzi PM, Serra A, et al. Hemodiafiltration with endogenous reinfusion with and without acetate-free dialysis solutions: Effect on ESA requirement. Blood Purif. 2011;31:235–242
66. Usuda M, Shinzato T, Sezaki R, et al. New simultaneous HF and HD with no infusion fluid. Trans Am Soc Artif Intern Organs. 1982;28:24–27
67. Tsuruta K, Andoh F, Kurahara I, Kaku T, Fukushima J, Shimada H. A simple method for clinical application of push/pull hemodiafiltration. Contrib Nephrol. 1994;108:71–78
68. Shinzato T, Fujisawa K, Nakai S, et al. Newly developed economical and efficient push/pull hemodiafiltration. Contrib Nephrol. 1994;108:79–86
69. Maeda K, Kobayakawa H, Fujita Y, et al. Effectiveness of push/pull hemodiafiltration using large-pore membrane for shoulder joint pain in long-term dialysis patients. Artif Organs. 1990;14:321–327
70. Maeda K, Shinzato T. Push/pull hemodiafiltration: Technical aspects and clinical effectiveness. Nephron. 1995;71:1–9
71. Shinzato T, Miwa M, Kobayakawa H, et al. Effectiveness of new push/pull hemodiafiltration for arthralgia in long-term hemodialysis patients. Contrib Nephrol. 1995;112:111–118
72. Shinzato T, Miwa M, Nakai S, et al. Alternate repetition of short fore- and backfiltrations reduces convective albumin loss. Kidney Int. 1996;50:432–435
73. Lee K, Lee SR, Mun CH, Min BG. Pulse push/pull hemodialysis: in vitro study on new dialysis modality with higher convective efficiency. Artif Organs. 2008;32:406–411
74. Gura V, Macy AS, Beizai M, Ezon C, Golper TA. Technical breakthroughs in the wearable artificial kidney (WAK). Clin J Am Soc Nephrol. 2009;4:1441–1448
75. Gura V, Davenport A, Beizai M, Ezon C, Ronco C. Beta2-microglobulin and phosphate clearances using a wearable artificial kidney: A pilot study. Am J Kidney Dis. 2009;54:104–111
76. Davenport A, Gura V, Ronco C, Beizai M, Ezon C, Rambod E. A wearable haemodialysis device for patients with end-stage renal failure: A pilot study. Lancet. 2007;370:2005–2010
77. Lee K, Min BG, Mun CH, Lee SR, Won YS. Pulse push/pull hemodialysis in a canine renal failure model. Blood Purif. 2008;26:491–497
78. Lee K, Lee DW, Min BG, Lee KK, Yun YM. Development of a new method for pulse push/pull hemodialysis. ASME J Medical Devices. 2011;5:031004 (031006 pages)
79. Shinzato T, Maeda K. Push/pull hemodiafiltration. Contrib Nephrol. 2007;158:169–176
80. Polaschegg HD. Hemodialysis machine technology: A global overview. Expert Rev Med Devices. 2010;7:793–810
Copyright © 2012 by the American Society for Artificial Internal Organs