Successful balloon dilatation of an infolded polytetrafluoroethylene graft: An unusual cause of early Fontan failure : Annals of Pediatric Cardiology

Secondary Logo

Journal Logo

IMAGES

Successful balloon dilatation of an infolded polytetrafluoroethylene graft: An unusual cause of early Fontan failure

Gupta, Saurabh Kumar1,; Mishra, Sataroopa1; Rajashekar, Palleti2; Kumar, Sanjeev3; Kothari, Shyam Sunder1

Author Information
Annals of Pediatric Cardiology 16(1):p 77-78, Jan–Feb 2023. | DOI: 10.4103/apc.apc_85_22
  • Open
  • Associated Video

Abstract

CASE DETAILS

An 8-year-old boy with a double outlet right ventricle, nonroutable ventricular septal defect, and severe pulmonary stenosis underwent bidirectional Glenn surgery at 2 years of age. Echocardiography showed normal ventricular function with no atrioventricular valvar regurgitation. Following cardiac catheterization, with mean pulmonary artery pressure measured at 11 mmHg, he underwent extracardiac total cavopulmonary connection (TCPC) using a 19 mm nonring enforced expanded polytetrafluoroethylene (ePTFE) graft. Even after 3 weeks of surgery, pleural drainage persisted despite optimal medical management, including diuretics, dual pulmonary vasodilators, and dietary modification. There was no evidence of diaphragmatic palsy or lung pathology. Initially, based on preoperative computed tomography (CT) angiography, he underwent coil embolization of aortopulmonary collaterals. The pleural drainage, however, persisted. Echocardiography showed obstruction at the lower end of the conduit with a gradient of 3 mmHg. The exact reason for obstruction, however, could not be clarified, and hence, a CT angiography was performed, which revealed narrowing at the lower end of the conduit with distorted noncircular lumen at the site of anastomosis with the inferior caval vein [Figure 1 and Video 1]. Infolding of the ePTFE graft was deemed to be the cause of obstructed TCPC circuit. The patient was initially considered for surgical revision. However, considering the risk of redo cardiac surgery, focal narrowing, and graft infolding being the possible cause of Fontan failure, the patient was taken up for balloon dilatation of the lower end of the ePTFE graft.

F1
Figure 1:
CT angiogram in the modified axial (a) and coronal (b) plane shows infolded lower end (arrows) of the ePTFE graft with partial thrombus. ePTFE: Expanded polytetrafluoroethylene, CT: Computed tomography

{"href":"Single Video Player","role":"media-player-id","content-type":"play-in-place","position":"float","orientation":"portrait","label":"Video Clip 1","caption":"","object-id":[{"pub-id-type":"doi","id":""},{"pub-id-type":"other","content-type":"media-stream-id","id":"1_gy66jmbi"},{"pub-id-type":"other","content-type":"media-source","id":"Kaltura"}]}

Cardiac catheterization confirmed the obstruction with a gradient of 3 mmHg across the lower end of the graft. The mean pulmonary artery pressure measured 14 mmHg, while pressure in the inferior caval vein was 17 mmHg. There was no gradient between the conduit and pulmonary arteries. The angiograms also confirmed the narrowing at the lower end of the conduit [Figure 2a and Video 2]. Balloon dilatation was performed using 18 mm × 40 mm MAXI LD™ (Cordis, Florida, USA) at 6 atm for 20 s [Figure 2b and Video 3]. Postprocedure mean pulmonary artery and inferior caval vein pressure fell to 12 mmHg with no residual gradient. The angiogram also showed relief of stenosis and improved flow in the Fontan circuit [Figure 2c and Video 4]. The pleural drainage gradually decreased over the next 2 weeks, and patient was discharged home. At 1-year follow-up, the child is doing well with good effort tolerance and no evidence of obstruction on echocardiography.

F2
Figure 2:
Catheter angiogram in posteroanterior projection before (a) and after balloon dilatation (c) at the lower end of the ePTFE graft. The narrowing (red arrows in panel a) disappeared following balloon dilatation (c). (b) The corresponding waist during balloon dilatation. ePTFE: Expanded polytetrafluoroethylene

{"href":"Single Video Player","role":"media-player-id","content-type":"play-in-place","position":"float","orientation":"portrait","label":"Video Clip 2","caption":"","object-id":[{"pub-id-type":"doi","id":""},{"pub-id-type":"other","content-type":"media-stream-id","id":"1_vjt1lt0y"},{"pub-id-type":"other","content-type":"media-source","id":"Kaltura"}]}

{"href":"Single Video Player","role":"media-player-id","content-type":"play-in-place","position":"float","orientation":"portrait","label":"Video Clip 3","caption":"","object-id":[{"pub-id-type":"doi","id":""},{"pub-id-type":"other","content-type":"media-stream-id","id":"1_sfjra1d5"},{"pub-id-type":"other","content-type":"media-source","id":"Kaltura"}]}

{"href":"Single Video Player","role":"media-player-id","content-type":"play-in-place","position":"float","orientation":"portrait","label":"Video Clip 4","caption":"","object-id":[{"pub-id-type":"doi","id":""},{"pub-id-type":"other","content-type":"media-stream-id","id":"1_79tlltw7"},{"pub-id-type":"other","content-type":"media-source","id":"Kaltura"}]}

In retrospect, we believe that the infolding of the PTFE graft could have happened secondary to an uneven suture line or infolding of the graft. The infolding was possibly more pronounced because the graft was nonringed. This also underscores the need for a thorough assessment of the conduit’s lie and contour at the end of surgery. Any indentation or furling of the graft should be promptly rectified by simple outward traction or modifying the suture line.

The index case also demonstrates the importance of careful assessment in the diagnosis and successful management of an unusual mechanism of obstructed Fontan circuit. Should the mechanism remain unidentified, the patient would undergo a much riskier surgical revision or stent implantation across hepatic veins.[1] Complete resolution of obstruction and sustained relief of obstruction at 1-year follow-up substantiate the efficacy of balloon dilatation to relieve infolding of the PTFE graft although a suboptimal result or recurrence of obstruction would have warranted the placement of a stent.

Declaration of patient consent

The authors certify that they have obtained all appropriate patient consent forms. In the form, the patient(s) has/have given his/her/their consent for his/her/their images and other clinical information to be reported in the journal. The patients understand that their names and initials will not be published and due efforts will be made to conceal their identity, but anonymity cannot be guaranteed.

Financial support and sponsorship

Nil.

Conflicts of interest

There are no conflicts of interest.

REFERENCE

1. Hagler DJ, Miranda WR, Haggerty BJ, Anderson JH, Johnson JN, Cetta F, et al. Fate of the Fontan connection:Mechanisms of stenosis and management. Congenit Heart Dis 2019;14:571–81.
Keywords:

Balloon dilatation; Fontan failure; polytetrafluoroethylene graft; underexpansion

Copyright: © 2023 Annals of Pediatric Cardiology