Heparin Concentration–Based Anticoagulation for Cardiac Surgery Fails to Reliably Predict Heparin Bolus Dose Requirements

Garvin, Sean MD*; FitzGerald, Daniel C. CCP; Despotis, George MD‡,§,∥; Shekar, Prem MD; Body, Simon C. MBChB, MPH*

doi: 10.1213/ANE.0b013e3181b79d09
Cardiovascular Anesthesiology: Research Reports
Free
SDC

BACKGROUND: Hemostasis management has evolved to include sophisticated point-of-care systems that provide individualized dosing through heparin concentration–based anticoagulation. The Hepcon HMS Plus system (Medtronic, Minneapolis, MN) estimates heparin dose, activated clotting time (ACT), and heparin dose response (HDR). However, the accuracy of this test has not been systematically evaluated in large cohorts.

METHODS: We examined institutional databases for all patients who underwent cardiac surgery with cardiopulmonary bypass (CPB) at our institution from February 2005 to July 2008. During this period, the Hepcon HMS Plus was used exclusively for assessment of heparin dosing and coagulation monitoring. Detailed demographic, surgical, laboratory, and heparin dosing data were recorded. ACT, calculated and measured HDR, and heparin concentrations were recorded. Performance of the Hepcon HMS Plus was assessed by comparison of actual and target ACT values and calculated and measured HDR.

RESULTS: In 3880 patients undergoing cardiac surgery, heparin bolus dosing to a target ACT resulted in wide variation in the postheparin ACT (r2 = 0.03). The postheparin ACT did not reach the target ACT threshold in 7.4%(i.e., when target ACT was 300 s) and 16.9% (i.e., when target ACT was 350 s) of patients. Similarly, the target heparin level calculated from the HDR did not correlate with the postbolus heparin level, with 18.5% of samples differing by more than 2 levels of the assay. Calculated and measured HDR were not linearly related at any heparin level.

CONCLUSIONS: The Hepcon HMS Plus system poorly estimates heparin bolus requirements in the pre-CPB period. Further prospective studies are needed to elucidate what constitutes adequate anticoagulation for CPB and how clinicians can reliably and practically assess anticoagulation in the operating room.

Published ahead of print October 27, 2009 Supplemental Digital Content is available in the text.

From the *Department of Anesthesiology, Perioperative and Pain Medicine, and Division of Cardiac Surgery, Brigham and Women's Hospital, Harvard Medical School, Boston, Massachusetts; and Departments of Pathology, §Immunology, and Anesthesiology, Washington University School of Medicine, St. Louis, Missouri.

Supported by departmental funds.

Address correspondence and reprint requests to Simon C. Body, MBChB, MPH, Brigham and Women's Hospital, Harvard Medical School, 75 Francis St., Boston, MA 02215. Address e-mail to body@zeus.bwh.harvard.edu.

Accepted June 22, 2009

Published ahead of print October 27, 2009

Administration of heparin, and monitoring of whole blood activated clotting time (ACT), is still the mainstay of anticoagulation for cardiopulmonary bypass (CPB).1,2 Hemostasis management has evolved to include more sophisticated point-of-care systems that provide individualized dosing of heparin and protamine, through heparin concentration–based estimation of heparin dosage, and measurement of ACT.3,4 Individualized heparin concentration–based anticoagulation management has been reported to cause less activation of thrombin, decreased fibrinolysis, reduced blood loss, fewer transfusions, and a decreased risk of reexploration for hemorrhage,58 although there is still a lack of consensus in this area.9,10 Previous studies of heparin-level monitoring devices, including the Hepcon HMS Plus, had varied results, with some studies showing good agreement between whole blood heparin concentration and either ACT before CPB11 or plasma anti-Xa levels before or during CPB,12 whereas others showed a lack of significant relationship between ACT and heparin concentration13 or whole blood versus anti-Xa heparin concentration.14 All of these studies were limited by their small sample sizes. Importantly, there is previously observed interindividual variation in measured heparin levels for a target ACT required for CPB.

Therefore, we examined the relationship between individualized in vitro estimation of the heparin level for a specific target ACT, and the subsequently observed ACT and heparin level measured after bolus administration of the heparin dose response (HDR)-estimated heparin dose. We hypothesized that: 1) there would be strong correlation between calculated and measured HDR for individual patients; 2) residual variation in the predicted HDR would be due to patient factors that are accounted for in the calculation of the HDR; and 3) the HDR-estimated heparin dose would achieve the desired ACT for CPB in all patients.

Back to Top | Article Outline

METHODS

The Hepcon HMS Plus (Medtronic, Minneapolis, MN) was introduced in our institution in February 2004, in concert with a change in heparin coagulation management. Use of this system was based on a belief that adequate anticoagulation for CPB using a heparin-coated circuit could be achieved with lower ACTs as long as reasonable, albeit lower heparin levels than had been historically used were maintained. Furthermore, there was a perception that reduced heparin administration may result in reduced postoperative bleeding.

After IRB approval, we examined institutional databases for all patients who underwent CPB for cardiac surgery at our institution, from February 2005 to July 2008 (n = 4667). Patients with incomplete perioperative anticoagulation data were excluded from analysis (n = 562). The majority of patients were excluded for having a target ACT of >350 s or subsequent administration of aprotinin (n = 157), no record of postheparinization heparin level (n = 144), no record of baseline ACT (n = 91), or no record of height or weight (n = 67). Furthermore, we excluded an additional 225 patients who were not anticoagulated using the institutional protocol described below, yielding 3880 analyzed patients.

Back to Top | Article Outline

Anticoagulation Management

The Hepcon HMS Plus was used for anticoagulation management according to the manufacturer's recommendations.15 The estimated blood volume for each patient was calculated using the manufacturer's instructions,15 according to the method described by Allen et al.14 After induction of anesthesia, baseline kaolin ACT, predicted HDR, predicted heparin concentration, and heparin bolus calculations were performed according to the manufacturer's instructions, using HDR cartridges encompassing whole blood heparin concentration ranging from 0.4 to 3.4 U/mL. The Hepcon HMS Plus recommended CPB prime heparin dose based on a 750- or 1000-mL prime volume was added to the calculated heparin bolus and administered via a central venous catheter. Heparin was not added to the CPB pump prime. Three minutes after United States Pharmacopeia porcine heparin administration and 10 min after onset of CPB, heparin concentration and ACT were remeasured. All patients received an ε-aminocaproic acid load of 7.5–10 g over 1 h, after the initial blood draw for baseline ACT, but before heparin administration and blood sampling for measurement of postheparin ACT. ε-Aminocaproic acid was subsequently infused at a rate of 1.25–1.5 g/h.

Throughout the entire study period, an ACT of more than 350 s and a minimum heparin concentration of 2 U/mL were used in patients undergoing non–coronary artery bypass graft (CABG) procedures or CABG with the use of cardiotomy suction. Patients undergoing primary CABG surgery without the use of cardiotomy suction before March 2007 were anticoagulated using a protocol that proscribed a minimum ACT of 300 s before the institution of CPB. After February 2007, to standardize anticoagulation management, the minimum ACT before CPB was increased to 350 s with a minimum heparin concentration of 2.0 U/mL for CABG surgery. These protocols were driven by a desire to reduce heparin and protamine doses to reduce blood loss but were not supported by prior clinical evidence of efficacy.

We also observed variability in anticoagulation practice that was not proscribed by a protocol. Some perfusionists administered additional heparin when the measured heparin level was <1.4 or 2.0 U/mL, according to personal preference. These patients were excluded from analysis, as previously described.

Back to Top | Article Outline

Data Collection and Statistical Analysis

Detailed demographic data, surgical indications, preoperative laboratory values, operative data, heparin bolus dose, ACT values, and measured heparin concentrations were routinely collected into a centralized database. The calculated HDR was calculated as the target ACT − baseline ACT divided by the target heparin level and reported as s · U−1 · mL−1. The calculated HDR differed from the method used by the Hepcon HMS Plus,15 but prospective comparison of the Hepcon HMS Plus and the above calculation in 15 patients showed excellent correlation (r2 = 0.991). This calculation uses the target heparin concentration identified from the HDR slope generated by the Hepcon HMS Plus, not the protocol-driven heparin level. Thus, the calculation of the HDR slope is independent of the clinical protocol and unaffected by the heparin level desired or obtained after heparin dosing.

The measured HDR was calculated as postheparin ACT − baseline ACT divided by the measured heparin level and reported as s · U−1 · mL−1. By using the measured postheparinization ACT and measured heparin level, the measured HDR is independent of the clinical protocol because the HDR slope has been demonstrated to be linear over the clinical range of heparin concentrations.11

The Hepcon HMS Plus heparin assay has limited fidelity, reporting only 6 categorical heparin levels, with intermediate values of heparin reported to the closest level. Accordingly, we could not directly compare the calculated and measured HDR. Therefore, we report the calculated HDR for each measured postbolus heparin level. Pearson product-moment correlation was performed between calculated and measured values of HDR, and heparin concentration and ACT data. ACT data at each heparin level were compared with Wilcoxon's ranked sum test. Statistical analysis was performed with JMP version 7.02 (SAS, Cary, NC). All data are presented as mean ± SD or median and interquartile range, as appropriate. A 2-sided P < 0.05 was considered as showing statistical significance.

Back to Top | Article Outline

RESULTS

A total of 3880 patients were analyzed. Baseline demographics and perioperative data are summarized in Table 1. Target ACTs of 300 and 350 s were used in 23.4% and 76.6% of patients, respectively. Heparin dose, ACT, and heparin level after the heparin bolus for each target ACT are detailed in Table 2. The mean heparin dose calculated to achieve a target ACT of 300 s was 152 U/kg, and for a target ACT of 350 s was 179 U/kg. After administration of the HDR-estimated heparin dose, the target ACT was not achieved in 7.4% of patients with a target ACT of 300 s, and it was not reached in 16.9% of patients with a target ACT of 350 s (Fig. 1). Additional heparin was administered either before or immediately after initiation of CPB to 91% of patients who failed to achieve their target ACT of 300 s and to 77% of patients who failed to achieve their target ACT of 350 s.

Table 1

Table 1

Table 2-a

Table 2-a

Table 2-b

Table 2-b

Table 2-c

Table 2-c

Table 2-d

Table 2-d

Figure 1

Figure 1

There was wide variation among patients for the ACT seen at each heparin level (Fig. 2). Measured HDR varied substantially among patients (mean ± SD, 104.7 ± 24.3 s · U−1 · mL−1; median [10%–90% interquartile range], 102 [77–137]). Correlation (r2) between the calculated and measured HDR was poor at all heparin levels (r2 < 0.02; Fig. 3). To exclude an outlier effect, the analysis was repeated in a subgroup of patients within 1 SD of the mean postheparin ACT, with similar results.

Figure 2

Figure 2

Figure 3

Figure 3

The predicted heparin level expected after heparin bolus administration was compared with the heparin level measured after heparin bolus. Levels that differed by more than 0.3 U/mL differed by more than the sensitivity (1 channel) of the assay and were observed in 51.0% of samples (Fig. 4). Levels that differed by more than 0.7 U/mL differed by more than 2 channels in the assay and were observed in 18.5% of samples.

Figure 4

Figure 4

Back to Top | Article Outline

DISCUSSION

We examined the performance of the Hepcon HMS Plus to guide anticoagulation in 3880 patients undergoing cardiac surgical procedures. Data provided by the Hepcon HMS Plus were used to calculate heparin dosing throughout CPB and to direct the administration of protamine after CPB. We observed wide variability in heparin dose requirements to achieve an adequate ACT for CPB, as others have previously described.1 We also observed poor correlation of the calculated HDR with the measured HDR. This led to ACT values that were less than the target ACT values when heparin dosing was guided by estimation of HDR, in 7.4%–16.9% of patients. The imprecision for the calculated HDR may explain the high frequency of failure to reach the target ACT.

Interpatient variability in HDR is well described, and use of a dose-response plot recommended by Bull et al.3 provides the basis of Hepcon HMS Plus calculations. Our initial hypothesis that the HDR slope calculated by the Hepcon HMS Plus would correlate with the measured HDR slope is not supported by the data. The interindividual variability in heparin response could be attributable to a number of in vivo factors, including error in the estimation of blood volume, effects of release of tissue factor pathway inhibitor (TFPI) by heparin in vivo but not ex vivo, extravascular sequestration of heparin, plasma protein binding, circulating antithrombin, and platelet activation.16

Estimation of blood volume is a potential source of error in heparin dose calculation. The methodology used by several previous studies5,6,12 is predominantly based on the method described by Allen et al.14 in 1956. The magnitude of error in estimation of blood volume in individual patients undergoing cardiac surgery has not been systematically evaluated. It is probable that the severity and nature of cardiac disease may be a substantial source of variation in blood volume estimates, thereby contributing to imprecision of point-of-care anticoagulation management instruments. Further investigation aimed at improving estimation of blood volume in patients undergoing cardiac surgery may result in better performance of point-of-care heparin concentration–based anticoagulation systems.

Potentially contributing to the imprecision of the Hepcon HMS Plus are the effects of release of TFPI by heparin in vivo but not ex vivo. TFPI in vivo is bound to endothelial cell surfaces with small amounts circulating in plasma and bound to platelets.17 TFPI is an endogenous serine protease inhibitor that exerts its action primarily through neutralizing factor Xa by complexing with factor Xa and forming a TFPI-FXa complex, while also providing feedback inhibition of the factor VIIa–tissue factor complex.18 Plasma TFPI is released from endothelial surface stores by administration of unfractionated and low-molecular-weight heparins,19 increasing plasma levels during CPB.20 Brodin et al.21 recently demonstrated significant synergy between antithrombin and TFPI in postheparin plasma with an almost equal contribution of each to the prolongation of anticoagulation in a laboratory setting. In the absence of an endothelial source of TFPI, the contribution of TFPI to heparin responsiveness may be underestimated by the initial HDR calculation performed by the Hepcon HMS Plus. This source of error in the calculated HDR would seemingly lead to increased heparin dosages and higher postheparin ACT, which is incongruous with our finding of 7.4%–16.9% of patients failing to achieve their target ACT, but congruous with the mean ACT being higher than the target ACT in our overall population. Although the exact contribution of TFPI to point-of-care measurements of heparin responsiveness has not been established, strong correlation between the TFPI-responsive Heptest (American Diagnostica, Stamford, CT) measurements of heparin level and ACT in the pre-CPB period22 has been observed, and release of TFPI correlates with the Heptest measurement of heparin response in normal volunteers receiving IV heparin.23 These observations reinforce the importance of TFPI in heparin responsiveness and point to a likely impact on point-of-care assessments of HDR.

Earlier studies of these heparin-level monitoring devices, including the Hepcon HMS Plus, had varied results with some studies showing good agreement between whole blood heparin concentration and either ACT before CPB11 or plasma anti-Xa levels,12 whereas others showed a lack of significant relationship between ACT and heparin concentration24 or whole blood versus anti-Xa heparin concentration during CPB.13,14 One study12 revealed substantial interpatient variability in HDR; however, the linear relationship between heparin concentration and kaolin ACT within individual patients was generally exceptional. Our retrospective analysis demonstrated significant variability in the responsiveness of the kaolin ACT and heparin in vivo versus kaolin ACT and heparin ex vivo (i.e., after IV administration of heparin), and calculated versus measured HDR, respectively (Fig. 3). This was paralleled by the finding that the Hepcon HMS Plus–guided heparin administration failed to result in adequate anticoagulation for CPB in 16% of patients in this study when the target ACT was 350 s, even with the CPB prime heparin dose being administered as part of the initial IV bolus dose.

Early work performed by Bull et al.25 and Young et al.26 continues to form the basis for the ACT target used to safely institute CPB, with many centers using an ACT of more than 480 s as a target. In 1981, Jobes et al.27 determined that a heparin level of 2 U/mL was associated with an ACT >300 s in >95% of patients. These data were not the basis for our use of a heparin level of 2 U/mL in the latter portion of the study period, but are in line with our observation of wide variation in ACT values at specific heparin levels. The theoretical benefit of using heparin concentration– based anticoagulation is that a stable heparin level may be achieved, thereby minimizing hemostatic activation7,28 and keeping the patient fully anticoagulated throughout the CPB period, which would lead to reduced blood loss and transfusion.5,6 We observed wide variability in ACT for any given heparin level (Fig. 2). Almost all studies using heparin concentration have used this historical reference as the target for defining adequate anticoagulation, but establishment of a “safe heparin level” has not been systematically studied.

This study is limited by its retrospective nature. We were not able to capture all of the perioperative factors that may influence anticoagulation and an individual's response to heparin administration. In addition, important clinical outcomes are not part of this data set, including perioperative complications that may relate to anticoagulation management, such as bleeding, transfusion, reexploration, myocardial infarction, new or worsening renal insufficiency, and cerebral vascular accident. Furthermore, our comparison of the calculated and measured HDR is limited by the imprecision of the automated protamine titration method, which measures whole blood heparin concentration using discrete cutoffs using only the 6 channels of the cartridge. Thus, we are left with 4 likely sources of disparity between calculated and measured HDR: 1) inaccurate estimate of the patient's blood volume; 2) lack of fidelity of measured heparin concentration because only 6 categories of heparin level describe the linear range of heparin concentration; 3) inherent inaccuracy of the device; and 4) differences between the anticoagulant activity of heparin ex vivo compared with its several actions in vivo, notably upon the release of TFPI. We were unable to subset these possible causes further, but they seem to be significant.

We conclude that the Hepcon HMS Plus fails to consistently provide the therapeutic heparin bolus dose uniformly in all patients based on the wide discrepancy in calculated versus measured HDR. This can lead to inadequate heparin doses to achieve a target ACT for CPB in as much as 16.9% of patients. However, the Hepcon HMS Plus was able to identify an adequate heparin dose for the majority of the patients. Because this study, did not compare the Hepcon HMS Plus with empiric dosing regimens, we are uncertain whether empiric regimens can either under- or overperform when compared to this system. Further prospective studies are needed to elucidate what constitutes adequate anticoagulation for CPB and how clinicians can reliably and practically assess anticoagulation in the operating room.

Back to Top | Article Outline

REFERENCES

1. Akl BF, Vargas GM, Neal J, Robillard J, Kelly P. Clinical experience with the activated clotting time for the control of heparin and protamine therapy during cardiopulmonary bypass. J Thorac Cardiovasc Surg 1980;79:97–102
2. Hattersley PG. Activated coagulation time of whole blood. JAMA 1966;196:436–40
3. Bull BS, Huse WM, Brauer FS, Korpman RA. Heparin therapy during extracorporeal circulation. II. The use of a dose-response curve to individualize heparin and protamine dosage. J Thorac Cardiovasc Surg 1975;69:685–9
4. Raymond PD, Ray MJ, Callen SN, Marsh NA. Heparin monitoring during cardiac surgery. Part 1: Validation of whole-blood heparin concentration and activated clotting time. Perfusion 2003;18:269–76
5. Despotis GJ, Joist JH, Hogue CW Jr, Alsoufiev A, Kater K, Goodnough LT, Santoro SA, Spitznagel E, Rosenblum M, Lappas DG. The impact of heparin concentration and activated clotting time monitoring on blood conservation. A prospective, randomized evaluation in patients undergoing cardiac operation. J Thorac Cardiovasc Surg 1995;110:46–54
6. Jobes DR, Aitken GL, Shaffer GW. Increased accuracy and precision of heparin and protamine dosing reduces blood loss and transfusion in patients undergoing primary cardiac operations. J Thorac Cardiovasc Surg 1995;110:36–45
7. Despotis GJ, Joist JH, Hogue CW Jr, Alsoufiev A, Joiner-Maier D, Santoro SA, Spitznagel E, Weitz JI, Goodnough LT. More effective suppression of hemostatic system activation in patients undergoing cardiac surgery by heparin dosing based on heparin blood concentrations rather than ACT. Thromb Haemost 1996;76:902–8
8. Bowie J, Kemma G. Automated management of heparin anticoagulation in cardiovascular surgery. Proc Am Acad Cardiovasc Perf 1985;6:1–10
9. Shore-Lesserson L, Reich DL, DePerio M. Heparin and protamine titration do not improve haemostasis in cardiac surgical patients. Can J Anaesth 1998;45:10–8
10. Slight RD, Buell R, Nzewi OC, McClelland DB, Mankad PS. A comparison of activated coagulation time-based techniques for anticoagulation during cardiac surgery with cardiopulmonary bypass. J Cardiothorac Vasc Anesth 2008;22:47–52
11. Despotis GJ, Alsoufiev AL, Spitznagel E, Goodnough LT, Lappas DG. Response of kaolin ACT to heparin: evaluation with an automated assay and higher heparin doses. Ann Thorac Surg 1996;61:795–9
12. Despotis GJ, Summerfield AL, Joist JH, Goodnough LT, Santoro SA, Spitznagel E, Cox JL, Lappas DG. Comparison of activated coagulation time and whole blood heparin measurements with laboratory plasma anti-Xa heparin concentration in patients having cardiac operations. J Thorac Cardiovasc Surg 1994;108:1076–82
13. Hardy JF, Belisle S, Robitaille D, Perrault J, Roy M, Gagnon L. Measurement of heparin concentration in whole blood with the Hepcon/HMS device does not agree with laboratory determination of plasma heparin concentration using a chromogenic substrate for activated factor X. J Thorac Cardiovasc Surg 1996;112:154–61
14. Allen TH, Peng MT, Chen KP, Huang TF, Chang C, Fang HS. Prediction of blood volume and adiposity in man from body weight and cube of height. Metabolism 1956;5:328–45
15. Medtronic Perfusion Systems. Hepcon HMS Plus operator's manual. Minneapolis, MN, 2001
16. Ranucci M, Isgro G, Cazzaniga A, Ditta A, Boncilli A, Cotza M, Carboni G, Brozzi S. Different patterns of heparin resistance: therapeutic implications. Perfusion 2002;17:199–204
17. Bajaj MS, Kuppuswamy MN, Saito H, Spitzer SG, Bajaj SP. Cultured normal human hepatocytes do not synthesize lipoprotein-associated coagulation inhibitor: evidence that endothelium is the principal site of its synthesis. Proc Natl Acad Sci USA 1990;87:8869–73
18. Broze GJ Jr, Miletich JP. Characterization of the inhibition of tissue factor in serum. Blood 1987;69:150–5
19. Hoppensteadt DA, Walenga JM, Fasanella A, Jeske W, Fareed J. TFPI antigen levels in normal human volunteers after intravenous and subcutaneous administration of unfractionated heparin and a low molecular weight heparin. Thromb Res 1995;77:175–85
20. Adams MJ, Cardigan RA, Marchant WA, Grocott MP, Mythen MG, Mutch M, Purdy G, Mackie IJ, Machin SJ. Tissue factor pathway inhibitor antigen and activity in 96 patients receiving heparin for cardiopulmonary bypass. J Cardiothorac Vasc Anesth 2002;16:59–63
21. Brodin E, Appelbom H, Osterud B, Hilden I, Petersen LC, Hansen JB. Regulation of thrombin generation by TFPI in plasma without and with heparin. Transl Res 2009;153:124–31
22. Hellstern P, Bach J, Simon M, Saggau W. Heparin monitoring during cardiopulmonary bypass surgery using the one-step point-of-care whole blood anti-factor-Xa clotting assay heptest-POC-Hi. J Extra Corpor Technol 2007;39:81–6
23. Hoffmann U, Harenberg J, Bauer K, Huhle G, Tolle AR, Feuring M, Christ M. Bioequivalence of subcutaneous and intravenous body-weight-independent high-dose low-molecular-weight heparin Certoparin on anti-Xa, Heptest, and tissue factor pathway inhibitor activity in volunteers. Blood Coagul Fibrinolysis 2002;13:289–96
24. Culliford AT, Gitel SN, Starr N, Thomas ST, Baumann FG, Wessler S, Spencer FC. Lack of correlation between activated clotting time and plasma heparin during cardiopulmonary bypass. Ann Surg 1981;193:105–11
25. Bull BS, Korpman RA, Huse WM, Briggs BD. Heparin therapy during extracorporeal circulation. I. Problems inherent in existing heparin protocols. J Thorac Cardiovasc Surg 1975;69:674–84
26. Young JA, Kisker CT, Doty DB. Adequate anticoagulation during cardiopulmonary bypass determined by activated clotting time and the appearance of fibrin monomer. Ann Thorac Surg 1978;26:231–40
27. Jobes DR, Schwartz AJ, Ellison N, Andrews R, Ruffini RA, Ruffini JJ. Monitoring heparin anticoagulation and its neutralization. Ann Thorac Surg 1981;31:161–6
28. Koster A, Fischer T, Praus M, Haberzettl H, Kuebler WM, Hetzer R, Kuppe H. Hemostatic activation and inflammatory response during cardiopulmonary bypass: impact of heparin management. Anesthesiology 2002;97:837–41