The Technology of Processed Electroencephalogram Monitoring Devices for Assessment of Depth of Anesthesia

Fahy, Brenda G. MD, MCCM*; Chau, Destiny F. MD

doi: 10.1213/ANE.0000000000002331
Technology, Computing, and Simulation: Narrative Review Article
Continuing Medical Education

Commercial brain function monitors for depth of anesthesia have been available for more than 2 decades; there are currently more than 10 devices on the market. Advances in this field are evidenced by updated versions of existing monitors, development of new monitors, and increasing research unveiling the mechanisms of anesthesia on the brain. Electroencephalography signal processing forms an integral part of the technology supporting the brain function monitors for derivation of a depth-of-anesthesia index. This article aims to provide a better understanding of the technology and functionality behind these monitors. This review will highlight the general design principles of these devices and the crucial stages in electroencephalography signal processing and classification, with a focus on the key mathematical techniques used in algorithm development for final derivation of the index representing anesthetic state. We will briefly discuss the advantages and limitations of this technology in the clinical setting as a tool in our repertoire used for optimizing individualized patient care. Also included is a table describing 10 available commercial depth-of-anesthesia monitors.

Published ahead of print August 3, 2017.

From the *Department of Anesthesiology, University of Florida College of Medicine, Gainesville, Florida; and Division of Pediatric Anesthesiology, Department of Pediatrics, the Children’s Hospital of the King’s Daughters, Eastern Virginia Medical School, Norfolk, Virginia.

Published ahead of print August 3, 2017.

Accepted for publication June 5, 2017.

Funding: B.G.F. receives funding from the Foundation for Anesthesia Education and Research.

The authors declare no conflicts of interest.

Reprints will not be available from the authors.

Address correspondence to Brenda G. Fahy, MD, MCCM, Department of Anesthesiology, University of Florida College of Medicine, 1600 SW Archer Rd, PO Box 100254, Gainesville, FL 32610. Address e-mail to bfahy@anest.ufl.edu.

Brain function monitors for assessing depth of anesthesia or hypnosis have been commercially available since the 1990s. Promoted to measure anesthetic effects on the brain, the purpose of these devices is to improve anesthetic titration to provide adequate anesthetic depth for each patient. Clinical application would hopefully translate by providing appropriate depth during administration of general anesthesia to avoid the consequences of insufficient anesthetic depth (eg, intraoperative awareness) or excessive anesthetic depth (eg, delayed emergence, postoperative delirium, and cognitive issues). Currently, no standards exist for intraoperative brain monitoring of anesthetic effects with full-montage electroencephalogram (EEG), the most widely accepted modality for cerebral function monitoring, including for the anesthetic effects on brain function.1 Interpretation of a full-montage EEG during anesthesia requires additional specialized training, which not all anesthesia providers have. All depth-of-anesthesia monitors utilize processed EEG (pEEG) signals to derive numerical indices representing the depth of anesthesia. Initially, the simple application of this noninvasive technology and the potential benefit to prevent intraoperative awareness led to rapid clinical adoption. Further research revealed important limitations of this technology; so, the need continues for an effective brain function monitor to guide anesthetic titration for optimal clinical care and improved patient outcomes. The purpose of this article is to review the technology behind the commercially available depth-of-anesthesia monitors, briefly summarize their advantages and limitations in the clinical setting, and assess the current barriers and opportunities for improvement toward a gold standard in state-of-anesthesia brain monitoring.

Back to Top | Article Outline

PRINCIPLES OF EEG MONITORING

The EEG represents the net summation of postsynaptic cortical neuronal potentials. To be registered by the sensors, the total current generated must penetrate several tissue layers and the monitoring electrodes. The signals received at the skin surface are approximately 100 times smaller than electrocardiographic signals and, as a result, are very sensitive to electrical interference and artifacts.2 The close proximity of the sensor to the rostral structures of the brain allows it to detect the EEG signals correlated with neural functions of the cerebral cortex, which relate to wakefulness, awareness, and memory. An EEG recording comprises various waveforms with different characteristics containing a large amount of information for interpretation. Advantages of the EEG as a monitor are that changes can be captured milliseconds after occurrence, and that it has a capacity to reflect both normal and aberrant aggregated electrical activity of different brain regions.

Anesthesia and other processes that inhibit higher cortical function decrease overall neuronal firing, slowing the waveforms and increasing its synchrony. Although the mechanism of anesthesia in the brain is unclear, progressive anesthetic depth is associated with both behavioral phenomena and characteristic EEG changes. The observed behavioral and EEG correlation with anesthetic depth are generally consistent and reproducible, and they provide the basis for EEG-based depth-of-anesthesia monitoring.

Figure 1

Figure 1

Currently available depth-of-anesthesia monitoring devices measure limited EEG activity from forehead electrodes and then process and analyze these signals with highly complex mathematical methods; the most promising correlating parameters are then compared with empirical data (a database of EEG and behavioral correlates mostly obtained from healthy patients) and tested clinically using various iterations that incorporate learned information in developing a depth-of-anesthesia monitor (Figure 1A). The complete algorithms for many monitors are proprietary. Advances in complex mathematical tools, coupled with improved computing power and speed, now allow for real-time processing of raw EEG signals.

Back to Top | Article Outline

EEG SIGNAL PROCESSING

The analog EEG signals received from the forehead surface electrodes are amplified and then filtered to remove noise and interference. Common interference includes electrical (eg, power line and electrocardiogram) and muscle activity (eg, eye and scalp movements). Forehead electromyograms (EMGs) contribute to the algorithm and are often isolated for separate display as another depth-of-anesthesia indicator.

During processing, EEG data are divided into time periods, or epochs. For digital conversion, the continuous analog signals are sampled at regular intervals and converted into discrete data points. Converting from a continuous signal to a series of discrete signals causes a loss of fidelity. To extract the desired from the undesired EEG signals, the data pass through additional filtering steps before undergoing mathematical processing. After artifact removal, the epoch is either salvaged, via approximation of missing data, or discarded if it is highly contaminated (Figure 1B).1 , 3 , 4

Back to Top | Article Outline

EEG ANALYSIS APPROACHES

Mathematical and statistical modeling is applied to the digitized raw EEG data to derive a depth-of-anesthesia measurement (Figure 2). This analysis involves EEG feature extraction and categorization into a final dimensionless index.

Figure 2

Figure 2

The following methods highlight the most common approaches applied by algorithms for the extraction of relevant EEG features.

Back to Top | Article Outline

Time Domain Analysis Methods

Time domain analysis examines EEG signals as a function of time. Characteristics of EEG morphology, associated randomness, and lack of precise predictability represent examples of important factors in algorithm development.

The algorithms used by the depth-of-anesthesia monitors often incorporate 1 or more time domain-based analytical approaches; examples include zero-order frequency, aperiodic analysis, wavelet analysis, rhythmicity autocorrelates, autoregressive models, and symbolic analysis, of which a detailed review is outside the scope of this article.3 , 5 , 6

Specifically, burst suppression rate is a very important time domain parameter used to quantify the time proportion spent in burst suppression; when seen on EEG during anesthesia, it indicates reduced cerebral activity and can represent deep anesthetic level or pathology (eg, ischemia).

Back to Top | Article Outline

Frequency Domain Analysis Methods

The analysis of complex biosignals is greatly enhanced by frequency domain analysis, which analyzes signals as a function of frequency with the following important techniques.

Back to Top | Article Outline

Fourier Transform.

The fundamental mathematical basis for EEG signal processing is Fourier transformation, which deconstructs the original time domain waveforms into a series of individual sinusoidal waves of different frequencies (cycles per second, Hertz), amplitudes, and phase (shift relative from origin or synchrony). An analogy is that of white light passing through a prism and decomposing into a spectrum of separate colors (spectrum of frequencies) with different intensities (amplitudes). A power spectrum is generated when the amplitude of the signals in the frequency spectrum is squared for mathematical analysis. The fast Fourier transform significantly improves computing efficiency from the original Fourier transform, allowing digital spectral results in real time, including visual modalities such a compressed spectral arrays, density spectral arrays, and spectral edge frequency.3

Back to Top | Article Outline

Bispectral Analysis.

Bispectral analysis, a higher-order mathematical manipulation, is used to gain additional useful information from the EEG power spectrum. The bispectrum quantifies the relationship between component EEG frequencies. The Fourier transform of waves at 2 primary frequencies (f1 and f2), along with their frequency sum (f1 + f2), forming a triplet (f1, f2, f1 + f2), are multiplied and the bispectrum magnitude is calculated. Additionally, the degree of phase coupling in that triplet (f1, f2, f1 + f2), called the bicoherence or bispectral coherence, is also quantified through further calculations. Although the physiologic significance of bispectral and bicoherence values is uncertain, high values would imply that those component frequencies are correlated, maybe originating from the same neural pacemaker. Also, bispectral analysis is useful to suppress certain sources of noise, thus improving the signal to noise ratio.3

Back to Top | Article Outline

Entropy

Entropy measures the randomness or irregularity of the signals. Increasing anesthesia depth correlates with decreased randomness, with the EEG signals displaying more regularity, implying more stability and predictability of the system. Entropy analysis can be performed in time or frequency domains (spectral entropy), and can also detect nonlinear signal correlations. Mathematical manipulations involving Fourier transform of the EEG signals produce a range from maximum irregularity to complete regularity, while operating in parallel with the time variable.4 , 7

Back to Top | Article Outline

Evoked Potentials

Auditory-evoked potentials (AEPs) measure time-locked EEG responses to repetitive auditory clicks, testing the neural pathways carrying information from the periphery to the cerebral cortex. The typical AEP response to increasing anesthetic concentrations is increased signal latency and decreased amplitude. The brainstem is relatively less sensitive to anesthetics, whereas the middle-latency AEPs or early cortical responses change predictably, with increasing concentrations of both volatile and intravenous anesthetics. The corresponding middle-latency AEP signals are discriminated from background EEG for index calculation8 (Figure 3). Combining AEPs with complex EEG processing techniques may improve the predictability of anesthetic effects on the brain.8

Figure 3

Figure 3

Table

Table

The Table shows examples of available monitors. For each monitor, the extracted EEG parameters are chosen by their ability to correlate to different stages of anesthesia (eg, burst suppression rate correlates with deep anesthetic level, while EMG activity correlates to light anesthetic level). During depth-of-anesthesia monitoring, those parameters undergo classification analysis that determines the extent of contribution from each toward the index value at that time point. Examples of classification analysis techniques include weighted sum, plausibility analysis, fuzzy logic inference systems, and neural network classifier systems. Each approach has advantages, disadvantages, and accuracy rates for discrimination of anesthetic depth levels. A detailed review is outside the scope of this article. Finally, most algorithms include a smoothing function to decrease rapid fluctuations in the displayed index; this is done by averaging the new index with immediate past values, adding a delay in reflection of the clinical state.5 , 6

Back to Top | Article Outline

ADVANTAGES AND INDICATIONS FOR pEEG MONITORS

Research suggests that depth-of-anesthesia monitors may reduce intraoperative awareness and may predict anesthesia outcomes in specific high-risk populations. At present, no monitor has been proven superior, and the current American Society of Anesthesiologists practice advisory on brain function monitoring says it “is not routinely indicated for patients undergoing general anesthesia, either to reduce the frequency of intraoperative awareness or to monitor depth of anesthesia … that the decision to use a brain function monitor should be made on a case-by-case basis by the individual practitioner for selected patients …”17 These recommendations are also shared by other professional international organizations.18 , 19 Depth-of-anesthesia monitoring has been recommended for those undergoing total intravenous anesthesia and neuromuscular-blocking agents for intraoperative awareness reduction.19 It is also recommended for the elderly patient to minimize postoperative delirium20 , 21 by avoiding deep anesthesia levels that may lead to unnecessary burst suppression, and recently received a grade A recommendation by the European Society of Anaesthesiologists.21

Back to Top | Article Outline

CHALLENGES OF pEEG MONITORS

The utilization of pEEG for brain monitoring has known limitations. Many factors can modulate the raw EEG signals during an anesthetic-induced altered state of consciousness, affecting the reliability of the EEG signals as indicators of anesthesia state. Some patient-dependent variables include age and pathophysiologic states (eg, hypothermia, hypoglycemia, acid-base abnormalities, previous brain pathology, brain ischemia, and seizures). Another challenge involves ketamine, nitrous oxide, and xenon not producing the typical EEG patterns seen during general anesthesia and with other agents (eg, opioids), although not significantly changing the pEEG index may significantly impact the anesthetic state of the patient. Other limitations include undetected artifacts and time lag for the index to display after changes in anesthesia affecting index reliability.

The most significant limitation resides in the neurophysiological basis of the effects of anesthesia and the associated end points of interest: unconsciousness/unawareness/amnesia, analgesia, and immobility. The EEG is deemed limited for spatial resolution and assessment of regional connectedness of brain processes. Currently, there is no unified standard to EEG features during anesthesia that correspond to all patients and anesthetic agents, hence the multiple unique algorithms. Although the correlation is positive, the EEG alone at this time seems improbable to provide all information needed to reliably and accurately assess unawareness and other anesthetic end points. Until the neurobiological principles of those end points are better understood, the devices can only indirectly infer the state of anesthesia through the EEG and surrogate parameters such as EMG. In addition, the definitions of unconsciousness and awareness need standardization across researchers and the published literature. Recent research in the biology of consciousness and anesthetic effects are yielding promising results for better understanding and development of a monitor.22–24

In conclusion, all current depth-of-anesthesia monitors rely on raw EEG signals for derivation of their indices. The technology behind these monitors is based on high-order mathematical analysis, using complex algorithms for processing the raw EEG signals enabled by gains in computing speed and power. As the understanding of the mechanisms of anesthesia on consciousness and related end points improves, a monitor that directly measures those neurobiological pathways can provide customized delivery of anesthesia for each patient with the goal of improving outcomes.

Back to Top | Article Outline

DISCLOSURES

Name: Brenda G. Fahy, MD, MCCM.

Contribution: This author helped write the manuscript.

Name: Destiny F. Chau, MD.

Contribution: This author helped write the manuscript.

This manuscript was handled by: Maxime Cannesson, MD, PhD.

Back to Top | Article Outline

REFERENCES

1. Isley MR, Edmonds HL Jr, Stecker M; American Society of Neurophysiological Monitoring. Guidelines for intraoperative neuromonitoring using raw (analog or digital waveforms) and quantitative electroencephalography: a position statement by the American Society of Neurophysiological Monitoring. J Clin Monit Comput. 2009;23:369–390.
2. Teplan M. Fundamentals of EEG measurement. Measurement Sci Rev. 2002;2:1–11.
3. Rampil IJ. A primer for EEG signal processing in anesthesia. Anesthesiology. 1998;89:980–1002.
4. Marchant N, Sanders R, Sleigh J, et al. How electroencephalography serves the anesthesiologist. Clin EEG Neurosci. 2014;45:22–32.
5. Musizza B, Ribaric S. Monitoring the depth of anaesthesia. Sensors (Basel). 2010;10:10896–10935.
6. Al-Kadi MI, Reaz MB, Ali MA. Evolution of electroencephalogram signal analysis techniques during anesthesia. Sensors (Basel). 2013;13:6605–6635.
7. Viertiö-Oja H, Maja V, Särkelä M, et al. Description of the entropy algorithm as applied in the Datex-Ohmeda S/5 Entropy Module. Acta Anaesthesiol Scand. 2004;48:154–161.
8. Plourde G. Auditory evoked potentials. Best Pract Res Clin Anaesthesiol. 2006;20:129–139.
9. Weber F, Zimmermann M, Bein T. The impact of acoustic stimulation on the AEP monitor/2 derived composite auditory evoked potential index under awake and anesthetized conditions. Anesth Analg. 2005;101:435–439.
    10. Cortínez LI, Delfino AE, Fuentes R, Muñoz HR. Performance of the cerebral state index during increasing levels of propofol anesthesia: a comparison with the bispectral index. Anesth Analg. 2007;104:605–610.
      11. Revuelta M, Paniagua P, Campos JM, et al. Validation of the index of consciousness during sevoflurane and remifentanil anaesthesia: a comparison with the bispectral index and the cerebral state index. Br J Anaesth. 2008;101:653–658.
        12. Kreuer S, Wilhelm W. The Narcotrend monitor. Best Pract Res Clin Anaesthesiol. 2006;20:111–119.
          13. Zikov T, Bibian S, Dumont GA, Huzmezan M, Ries CR. Quantifying cortical activity during general anesthesia using wavelet analysis. IEEE Trans Biomed Eng. 2006;53:617–632.
            14. Drover D, Ortega HR. Patient state index. Best Pract Res Clin Anaesthesiol. 2006;20:121–128.
              15. Willmann K, Springman S, Rusy D, Daily E. A preliminary evaluation of a new derived EEG index monitor in anesthetized patients. J Clin Monit Comput. 2002;17:345–350.
                16. Jensen EW, Valencia JF, López A, et al. Monitoring hypnotic effect and nociception with two EEG-derived indices, qCON and qNOX, during general anaesthesia. Acta Anaesthesiol Scand. 2014;58:933–941.
                  17. American Society of Anesthesiologists Task Force on Intraoperative Awareness. Practice advisory for intraoperative awareness and brain function monitoring: a report by the American society of anesthesiologists task force on intraoperative awareness. Anesthesiology. 2006;104:847–864.
                  18. Australian and New Zealand College of Anaesthetists (ANZCA). Guidelines on monitoring during anaesthesia. Available at: http://www.anzca.edu.au/documents/ps18-2015-guidelines-on-monitoring-during-anaesthe.pdf. Updated December 2015. Accessed April 22, 2017.
                  19. Checketts MR, Alladi R, Ferguson K, et al; Association of Anaesthetists of Great Britain and Ireland. Recommendations for standards of monitoring during anaesthesia and recovery 2015: Association of Anaesthetists of Great Britain and Ireland. Anaesthesia. 2016;71:85–93.
                  20. Inouye SK, Robinson T, Blaum C, et al. Postoperative delirium in older adults: best practice statement from the American Geriatrics Society. J Am Coll Surg. 2015;220:136–148.e1.
                  21. Aldecoa C, Bettelli G, Bilotta F, et al. European Society of Anaesthesiology evidence-based and consensus-based guideline on postoperative delirium. Eur J Anaesthesiol. 2017;34:192–214.
                  22. Purdon PL, Pierce ET, Mukamel EA, et al. Electroencephalogram signatures of loss and recovery of consciousness from propofol. Proc Natl Acad Sci U S A. 2013;110:E1142–E1151.
                  23. Mashour GA, Orser BA, Avidan MS. Intraoperative awareness: from neurobiology to clinical practice. Anesthesiology. 2011;114:1218–1233.
                  24. Hudetz AG, Mashour GA. Disconnecting consciousness: is there a common anesthetic end point? Anesth Analg. 2016;123:1228–1240.
                  © 2018 International Anesthesia Research Society