The Accuracy of Pulse Spectroscopy for Detecting Hypoxemia and Coexisting Methemoglobin or Carboxyhemoglobin

Kulcke, Axel PhD; Feiner, John MD; Menn, Ingolf PhD; Holmer, Amadeus Dipl-Ing; Hayoz, Josef PhD; Bickler, Philip MD, PhD

doi: 10.1213/ANE.0000000000001219
Technology, Computing, and Simulation: Research Report
Free

BACKGROUND: Pulse spectroscopy is a new noninvasive technology involving hundreds of wavelengths of visible and infrared light, enabling the simultaneous quantitation of multiple types of normal and dysfunctional hemoglobin. We evaluated the accuracy of a first-generation pulse spectroscopy system (V-Spec™ Monitoring System, Senspec, Germany) in measuring oxygen saturation (SpO2) and detecting carboxyhemoglobin (COHb) or methemoglobin (MetHb), alone or simultaneously, with hypoxemia.

METHODS: Nineteen volunteers were fitted with V-Spec probes on the forehead and fingers. A radial arterial catheter was placed for blood sampling during (1) hypoxemia with arterial oxygen saturations (SaO2) of 100% to 58.5%; (2) normoxia with MetHb and COHb increased to approximately 10%; (3) 10% COHb or MetHb combined with hypoxemia with SaO2 of 100% to 80%. Standard measures of pulse-oximetry performance were calculated: bias (pulse spectroscopy measured value − arterial measured value) mean ± SD and root-mean-square error (Arms).

RESULTS: The SpO2 bias for SaO2 approximately 60% to 100% was 0.06% ± 1.30% and Arms of 1.30%. COHb bias was 0.45 ± 1.63, with an Arms of 1.69% overall, and did not degrade substantially during moderate hypoxemia. MetHb bias was 0.36 ± 0.80 overall and stayed small with hypoxemia. Arms was 0.88 and was <3% at all levels of SaO2 and MetHb. Hypoxemia was also accurately detected by pulse spectroscopy at elevated levels of COHb. At elevated MetHb levels, a substantial negative bias developed, −10.3 at MetHb >10%.

CONCLUSIONS: Pulse spectroscopy accurately detects hypoxemia, MetHb, and COHb. The technology also accurately detects these dysfunctional hemoglobins during hypoxemia. Future releases of this device may have an improved SpO2 algorithm that is more robust with methemoglobinemia.

Published ahead of print April 22, 2016

From the *Senspec, Rostock, Germany; Department of Anesthesia and Perioperative Care, University of California San Francisco, San Francisco, California; and Sentec, Therwil, Switzerland.

Accepted for publication December 28, 2015.

Published ahead of print April 22, 2016

Funding: University of California, San Francisco, Department of Anesthesia and Perioperative Care. Oximeter and probes were supplied by Senspec.

Conflict of Interest: See Disclosures at the end of the article.

Reprints will not be available from the authors.

Address correspondence to John Feiner, MD, Department of Anesthesia and Perioperative Care, University of California San Francisco, 500 Parnassus, MUE 406, Box 0648, San Francisco, CA 94143. Address e-mail to john.feiner@ucsf.edu.

Two-wavelength (red and infrared) pulse oximetry is a mature technology that determines the functional saturation of hemoglobin with oxygen. Over the past decade, the addition of more wavelengths of light has enabled oximetry systems to detect dysfunctional hemoglobins (carboxyhemoglobin [COHb] and methemoglobin [MetHb]) and total hemoglobin (e.g., Masimo Pulse Co-Oximetry™; Masimo Corp., Irvine, CA)1–3 as well as tissue oxygenation (e.g., near-infrared spectroscopy, various manufacturers.)4

Five to 12 wavelengths of light are used in the Masimo rainbow® Pulse CO-Oximeters, that is, the Radical-7® (Masimo Corp.).5 The measured parameter for COHb concentration is named SpCO® and for MetHb, SpMet® by Masimo. Early studies on healthy volunteers for COHb concentrations demonstrated acceptable accuracy of the Masimo Pulse CO-Oximeter for detecting COHb during normoxia.1,2 Studies with patients with elevated COHb levels (>10% COHb)5,6 suggested limitations in accuracy.7,8 More recent studies showed an improved accuracy in volunteer studies and a more robust tolerance for hypoxic oxygen levels,9,10 but detecting COHb when SaO2 is approximately <87% is still not possible.

Other multispectral (up to 8 wavelengths) approaches to measuring hemoglobin species have also been developed.3,11–15 Multiwavelength spectroscopy systems were initially developed for postmortem measurements in forensic medicine applications.16–18

Theoretically, the spectroscopic approach can reduce cross talk between the detection of multiple hemoglobin species, especially when the absorbances are in the same region of the spectrum.19–21 In the region of 600 to 700 nm, deoxyhemoglobin and MetHb are spectrally similar,20 and a photometer with only few discrete wavelengths (such as current light-emitting diode [LED]–based systems) could fail to discriminate multiple hemoglobin species.22,23 With the high spectral resolution, the separation of relevant absorption band signals from distortions such as painted fingernails, melanin, water, or subcutaneous fat is also possible by applying a derivative transform to the data of the original spectrum (derivative spectroscopy). Such capabilities are clinically important when mixed intoxications are present or when toxic exposures have secondarily caused hypoxemia from respiratory dysfunction.

In this article, we document and describe the first results with the pulse spectroscopic method in volunteer studies for hypoxemia, COHb, and MetHb concentration measurements.

Back to Top | Article Outline

METHODS

Table 1

Table 1

The University of California at San Francisco (UCSF) Committee on Human Research approved the study, and all subjects gave informed written consent. The pool of subjects comprised healthy, nonsmoking men and women, from 21 to 38 years of age, willing to volunteer for the study for a nominal payment. The selected group of subjects was gender and ethnically balanced, following the Food and Drug Administration (FDA) requirements for standard studies of pulse oximeter accuracy. The final group included 19 healthy adult subjects, 10 men and 9 women, with a range of skin pigmentation (Table 1). Ten subjects participated in a standard hypoxemia study with normal (low) levels of COHb and MetHb, 6 subjects were studied with elevated COHb levels, and 3 subjects were studied with elevated MetHb levels. The study size was based on prior studies and the size of standard studies of pulse oximeter accuracy for FDA 510(k) approval.

Back to Top | Article Outline

V-Spec Monitoring System

Pulse-spectrometer probes (V-Spec™ Sensors, Germany) and instruments were supplied by Senspec. V-Spec Sensor probes (reusable with soft finger clip or attachment tape) were used to measure oxygen saturation (SpO2), COHb levels (SpHbCO), and MetHb levels (SpHbMet). The software versions of V-Spec Sensors and V-Spec monitors used in the hypoxemia study were 1.01 and 1.2.4, respectively. During the COHb Study and the MetHb Study, V-Spec Sensors and V-Spec monitors with software versions 2.05 and 03.03.01 were used, respectively.

The V-Spec Sensor is a miniaturized and very fast two-dimensional pulsatile spectroscopy unit. The sensor integrates an LED-based lighting unit, a spectrometer unit, and a microelectronic system to transform the optical signals into digital electronic signals and to transfer the information with high speed to a clinical monitor (V-Spec Monitor). The actual size of the integrated sensor is about 44 mm × 16 mm × 24 mm. The sensor is applied to the patient with a sensor-specific medical attachment tape. In another setup (V-Spec Soft Sensor), the sensor is integrated in a soft finger clip to position the sensor at the patient’s fingertip.

The V-Spec Sensor uses a transflectance measurement, also called the remission measuring setup. The light enters the tissue at 1 point, gets scattered and absorbed during the path through the tissue, and then leaves the tissue at a second point, where it enters into the spectrometer unit. The spectrometric sensor splits the light in 350 wavelength points in the range from 500 to 850 nm, instead of 2 points for conventional pulse oximeter and 5 to 10 points for multispectral photometers (Fig. 1). The calculation of hemoglobin species is based on about 15-million, single, 12-bit photo sensor values generated from the sensor in each second. The spectra are preprocessed and white calibrated and then transferred to 40 averaged absorption spectra per second. These spectra are used to extract physiological parameters (e.g., MetHb, SaO2, etc.) by different mathematical and chemometric procedures.

Figure 1

Figure 1

The integrated lighting unit is LED based. The LED emission spectrum is broadband and specially designed for use with patient skin and human tissue containing typical amounts of hemoglobin. The LED emission efficiency is optimized for the low electrical input power of the light unit.

Two pulse-spectrometer probes (V-Spec sensors) were placed at the forehead and were fixed with the sensor-specific attachment tape. Two to 4 pulse-spectrometer probes (V-Spec Soft Sensor) were placed on different fingers. The probe locations were randomized for each subject and were connected to V-Spec monitors. The monitors were equipped with additional memory and with a special software version to store the original spectral data online during the study. The original spectral data, the extracted pulsatile part of the spectra (representing the pulsing, arterial blood), and the continuous part of the spectra (representing the tissue) were recorded continuously during the study.

Back to Top | Article Outline

Study Protocol

A 22-gauge radial arterial cannula was placed in either the left or the right wrist of each subject. Arterial blood was analyzed with a multiwavelength optical blood gas analyzer (ABL90 FLEX; Radiometer Medical A/S, Copenhagen, Denmark, or OSM3, Radiometer Medical A/S) to determine arterial oxygen saturation (SaO2), COHb concentration (%COHb), and MetHb concentration (%MetHb).

Studies on each subject began with 1 arterial blood sample drawn as a baseline value while patients were breathing room air. Hypoxemia was then induced to 5 to 6 different targeted plateaus from 100% to 60% by having subjects breathe mixtures of nitrogen, air, and carbon dioxide according to a protocol previously detailed.24 Oxygen saturation is calculated from end-tidal PO2 and PCO2 breath-by-breath, which guides the gas mixtures. Each saturation plateau level was maintained for at least 60 seconds. When pulse oximeter values had stabilized, 2 arterial blood samples were obtained approximately 30 seconds apart. After the final SaO2 plateau, subjects breathed 100% O2 and then returned to breathing room air.

Elevated COHb was induced by having subjects breath carbon monoxide (CO) gas to produce a target COHb level of approximately 10% to 12% based on the volunteer study by Barker et al.1 and accumulated experience in volunteers in UCSF laboratory.7 CO (15–30 mL) was added to a 1-L bag prefilled with approximately 200 mL of oxygen. Subjects then briefly rebreathed this mixture from a mouthpiece, allowing us to produce approximately 2% stepwise changes in COHb. Blood samples were obtained 5 minutes after each administration of CO. When COHb reached target levels (10%–12%), hypoxemia was induced in steps from 100% to 80%, and blood samples were taken using the prior protocol.

In a different group of subjects, elevated MetHb levels were induced by slow IV administration of approximately 300 mg sodium nitrite to produce a target MetHb level of 10%. During sodium nitrite infusion, blood samples were obtained every 5 minutes to confirm elevated MetHb levels. Hypoxemia was then induced to in steps from 100% to 80% after reaching the target level of MetHb, and blood samples were taken using the prior protocol.

Figure 2

Figure 2

Figure 3

Figure 3

For the COHb and the MetHb part of the study, the analysis of pulse spectroscopy was performed separately and blinded to the results of blood gas analysis. In Figures 2 and 3, typical sequences for COHb and MetHb studies are shown.

Back to Top | Article Outline

Data Analysis and Statistics

Accuracy of measuring hypoxemia (SpO2) is specified by an international guideline (DIN EN ISO 80601-2-61:2011).25 For the measurement of tissue hypoxia by near-infrared spectroscopy, no standard is established.

There is also no current FDA standard for accurately determining noninvasively measured COHb and MetHb concentrations. For SpO2, root-mean-square error (Arms) <3% is the acceptable accuracy standard established by the FDA. We considered that the SpO2 accuracy would be degraded if elevated %COHb or %MetHb reached Arms >3%. Arms <3% would certainly represent acceptably accurate performance for determining COHb and MetHb concentration noninvasively although it may not be reasonable to expect the same accuracy and precision as for determining SpO2.

Pulse spectrometer performance was analyzed by calculating mean bias (SpO2 − SaO2; SpHbCO − COHb; or SpHbMet − MetHb), precision (SD of the bias), and Arms over different ranges of %SaO2, %COHb, and %MetHb. The distribution of bias was plotted and tested for normality with the Shapiro-Wilk test. Results ranged from P < 0.0001 to P = 0.27. Where data were not normally distributed, this was predominantly due to high and low outliers, so no data transformation was performed. Bias, precision, and Arms were determined and analyzed separately for SpO2, SpHbCO, and SpHbMet. Bias within different ranges of SaO2 was compared with a repeated-measures analysis of variance, and Tukey-Kramer honest significant difference was used for multiple comparison testing. Levene F test was used to compare SDs within various ranges specified in the tables. Limits of agreement were calculated according to Bland and Altman,26 with adjustments for multiple measurements for each individual according to the “Method Where The True Value Varies”. The 95% confidence intervals (CIs) of the limits of agreement were calculated by bootstrapping.

Bias was plotted against arterial blood percent COHb, percent MetHb, or SaO2. Arterial values were treated as a gold standard. For bias plots, linear regression was performed accounting for repeated measures. Slopes are reported with 95% confidence limits.

Receiver operating characteristics (ROC) were analyzed by setting percent COHb ≥10% and ≥5% as positive carboxyhemoglobinemia. Data from the 6 devices were combined for the analysis. Area under the curve (AUC) with 95% confidence limits was calculated.

Data are reported as mean ± SD or mean (95% CI) as indicated. All statistical tests were 2 sided, and P value <0.05 was considered significant. Data were analyzed with JMP 11.0 (SAS Institute, Cary, NC) and Stata 14.0 (StataCorp, College Station, TX).

Back to Top | Article Outline

RESULTS

Demographics

Table 1 summarizes the demographic information separately for each study type. Ten subjects of different gender, ethnicity, and skin color participated in the hypoxemia study. The COHb study was performed with 6 subjects and 3 subjects participated in the MetHb study.

Back to Top | Article Outline

Accuracy of Detecting Hypoxemia

For 4 different probes placed randomly on the forehead and fingers, 902 data points for 250 blood draws were obtained from the 10 subjects participating in the hypoxemia portion of the study. The SaO2 values ranged from 58.5% to 100.0%.

Table 2 shows a summary of pulse spectroscopy oxygen saturation bias (SpO2 − SaO2) within different SaO2 ranges. Although bias did differ significantly with SaO2 level, the overall mean bias, precision, and root-mean-square error (Arms) were all low. Mean bias on the finger was 0.5% (P < 0.0001) higher than the mean bias on the forehead; however, data are clinically so similar that the combined data are shown in the table. Missing data occurred from 2 causes. One was a combination of sweating and movement during the study so that the sensor was not positioned. The second case was the skin-detection algorithm in combination with very dark skin. In 1 subject, the forehead sensors detected only weak signals.

Table 2

Table 2

Figure 4

Figure 4

In Figure 4A, bias is plotted versus SaO2. While bias increases significantly at lower SaO2 (P < 0.0001), the effect is small, only 1.5% over the range of SaO2 from 100% to 50%. Figure 4B shows a standard scatterplot of SpO2 versus SaO2.

Back to Top | Article Outline

Accuracy of Detecting Elevated COHb During Normoxia and Hypoxemia

During the COHb study, 152 blood draws and 888 corresponding pulse spectroscopy readings with 6 devices were obtained. Observations spanned SaO2 from 73.7% to 100% and COHb from 0.5% to 11.0%. The MetHb was not varied and stayed low during this part of the study.

Table 3 summarizes the COHb bias (%SpHbCO − %COHb) statistics. While the bias was statistically different in different ranges of both SaO2 and COHb, the differences were small. Overall bias, precision, and Arms were low in all ranges of SaO2 and COHb. The overall Arms was 1.69%, and it was <3% even with hypoxemia.

Table 3

Table 3

An example of step changes in COHb induced by intermittently breathing small quantities of carbon monoxide is shown in Figure 2A. Figure 2B shows the percent COHb bias versus percent COHb. In Figure 2C, the bias is shown versus SaO2. While statistically significant (P < 0001), the effects of both percent COHb and percent SaO2 on the bias were small: a difference in bias of 2.5% over the range of COHb of 0 to 12% and 1.7% over the range of SaO2 from 70% to 100%.

Figure 5

Figure 5

ROC curves showed very good sensitivity and specificity for detecting carboxyhemoglobinemia. Figure 5A shows the ROC curve for carboxyhemoglobinemia defined as COHb ≥5%. The AUC was 0.99 (95% CI, 0.99–1.0). Figure 5B shows the ROC curve for a positive COHb of ≥10%. The AUC was 0.91 (95% CI, 0.89–0.92).

Back to Top | Article Outline

Accuracy of Detecting Elevated MetHb During Normoxia and Hypoxemia

During the MetHb study, 98 blood draws and 539 corresponding pulse spectroscopy readings were obtained. Observations spanned SaO2 from 78% to 100% SaO2 and MetHb from 0.3% to 10.8%. The COHb remained low (<1%) during this part of the study.

Table 4 summarizes the MetHb bias (percent SpHbMet − percent MetHb) statistics. While mean bias did differ significantly between different ranges of percent SaO2 and percent MetHb, the differences were small. Arms was <3% in all ranges.

Table 4

Table 4

Figure 3A shows an example of changes in MetHb levels during and after the infusion of sodium nitrite. Figure 3B shows the MetHb bias versus percent MetHb from arterial blood samples. In Figure 3C, the MetHb bias is shown versus percent SaO2. Hypoxemia had a statistically significant effect on the MetHb bias (P < 0.0001), but this amounted to a difference in bias of only 1.5% over the range of SaO2 from 100% to 75%.

Back to Top | Article Outline

Accuracy of Detecting Hypoxemia During Carboxyhemoglobinemia and Methemoglobinemia

Table 5 shows a summary of SpO2 bias statistics during carboxyhemoglobinemia. Higher levels of COHb caused slightly negative bias. Arms was still <3% except at lower SaO2 level. Figure 6 shows the SpO2 bias as a function of COHb in panel A and SaO2 in panel B.

Table 5

Table 5

Figure 6

Figure 6

Table 6

Table 6

Figure 7

Figure 7

Table 6 shows summarized SpO2 bias during methemoglobinemia. Mean bias was dramatically negative at higher levels of MetHb. Figure 7 shows SpO2 bias as a function of MetHb in panel A and SaO2 in panel B.

Back to Top | Article Outline

DISCUSSION

The primary purpose of this study was to assess the in vivo accuracy of a new pulse spectroscopy sensor system. The main findings were that the V-Spec Monitoring System accurately detects hypoxemia in normal levels of COHb and MetHb, that COHb and MetHb are detected accurately at normoxia, and, finally, that even in the presence of substantial arterial hypoxemia, the V-Spec Monitoring System is able to accurately measure both MetHb and COHb. The simultaneous and accurate detection of both COHb and MetHb during significant degrees of hypoxia could be of clinical utility in situations where simultaneous hypoxemia and dyshemoglobinemia coexist.

The clinical importance for noninvasive measurement of COHb and MetHb has been reviewed in numerous case reports.27–31 The ability to diagnose suspected cases of CO exposure or elevated MetHb levels in a timely manner and to avoid unnecessary invasive testing requires good sensitivity and specificity. Detecting elevated COHb and MetHb levels to enable rapid initiation of appropriate treatment may improve outcomes and increases patient safety. The measurement of elevated COHb levels, in particular, can lead to different treatment procedures, such as normal or hyperbaric oxygen administration. ROC curves for detecting abnormal level of COHb indicate very good sensitivity and specificity.

Detecting elevated COHb levels, even at levels that may not be clinically deleterious, may identify sources of CO exposure at home or at work that could cause serious harm. This, in turn, could lead to testing of others exposed to an event that might otherwise go unrecognized. Determining whether COHb levels are improving in patients requires more frequent measurements. In this case, noninvasive and continuous point-of-care technologies are preferable. Simultaneous hypoxemia would be likely in cases of smoke inhalation and loss of consciousness.

Limitations of our study included the small number of subjects for the COHb and MetHb portions of the study. Gender and ethnicity balance is harder to achieve with smaller numbers. However, repeated-measures data are so robust that we are able to detect highly significant effects that are actually not clinically significant. For example, the difference in SpO2 bias with a decrease in SaO2 to 70% amounts to <1%. This is also the first human testing of a pulse spectrometer device, which has not yet undergone multiple iterations, correcting the algorithms for multiple different simultaneous hemoglobin species, such as MetHb and deoxyhemoglobin. Initial versions of multiple-wavelength oximeters had some difficulty with separation of multiple different hemoglobin species before correction with further human testing.32,33 This is because only real recorded in vivo data for the calibration of the device can be used. The extreme bias of SpO2 under the influence of MetHb is an example of the need to include real absorbance information in the algorithm, as the first release was developed without considering any MetHb spectra. The in vivo data recorded during this study can be used to increase the algorithm’s performance, and further releases of the device could have a revised SpO2 algorithm that is more robust to MetHb.

In conclusion, pulse spectroscopy, involving the spectral analysis of hundreds of wavelengths of light, demonstrates a high level of accuracy for measuring hypoxemia, COHb, and MetHb up to 11%, which did not degrade strongly during hypoxemia.

Back to Top | Article Outline

DISCLOSURES

Name: Axel Kulcke, PhD.

Contribution: This author helped design the study, conduct the study, analyze the data, and write the manuscript.

Attestation: Axel Kulcke has seen the original study data, reviewed the analysis of the data, and approved the final manuscript.

Conflicts of Interest: Axel Kulcke worked for Senspec.

Name: John Feiner, MD.

Contribution: This author helped design the study, conduct the study, analyze the data, and write the manuscript.

Attestation: John Feiner has seen the original study data, reviewed the analysis of the data, approved the final manuscript, and is the author responsible for archiving the study files.

Conflicts of Interest: John Feiner received research funding from Masimo Inc., Nonin Medical Inc., Bluepoint Medical, Xhale, and CAS Medical and various pulse oximeter manufacturers funded our laboratory to perform human accuracy studies.

Name: Ingolf Menn, PhD.

Contribution: This author helped design the study, conduct the study, analyze the data, and write the manuscript.

Attestation: Ingolf Menn has seen the original study data, reviewed the analysis of the data, and approved the final manuscript.

Conflicts of Interest: Ingolf Menn worked for Senspec.

Name: Amadeus Holmer, Dipl-Ing.

Contribution: This author helped design the study, conduct the study, and analyze the data.

Attestation: Amadeus Holmer has seen the original study data, reviewed the analysis of the data, and approved the final manuscript.

Conflicts of Interest: Amadeus Holmer worked for Senspec.

Name: Josef Hayoz, PhD.

Contribution: This author reviewed the data analysis and revised the manuscript.

Attestation: Josef Hayoz has seen the original study data, reviewed the analysis of the data, and approved the final manuscript.

Conflicts of Interest: Josef Hayoz works for Sentec.

Name: Philip Bickler, MD, PhD.

Contribution: This author helped design the study, conduct the study, analyze the data, and write the manuscript.

Attestation: Philip Bickler has seen the original study data, reviewed the analysis of the data, and approved the final manuscript.

Conflicts of Interest: Philip Bickler received research funding from Masimo Inc., Nonin Medical, Bluepoint Medical, Xhale, and CAS Medical and various pulse oximeter manufacturers funded our laboratory to perform human accuracy studies.

This manuscript was handled by: Maxime Cannesson, MD, PhD.

Back to Top | Article Outline

REFERENCES

1. Barker SJ, Curry J, Redford D, Morgan S. Measurement of carboxyhemoglobin and methemoglobin by pulse oximetry: a human volunteer study. Anesthesiology 2006;105:892–7.
2. Zaouter C, Zavorsky GS. The measurement of carboxyhemoglobin and methemoglobin using a non-invasive pulse CO-oximeter. Respir Physiol Neurobiol 2012;182:88–92.
3. Al-Ali A, Diab MK, Panch A, Abdul-Hafiz Y, Macneish I, William J. Patent Application. WO 2008118993 A1. Multiple wavelength optical sensor, 2007.
4. Macnab AJ, Stothers L. Development of a near-infrared spectroscopy instrument for applications in urology. Can J Urol 2008;15:4233–40.
5. Roth D, Herkner H, Schreiber W, Hubmann N, Gamper G, Laggner AN, Havel C. Accuracy of noninvasive multiwave pulse oximetry compared with carboxyhemoglobin from blood gas analysis in unselected emergency department patients. Ann Emerg Med 2011;58:74–9.
6. Touger M, Birnbaum A, Wang J, Chou K, Pearson D, Bijur P. Performance of the RAD-57 pulse CO-oximeter compared with standard laboratory carboxyhemoglobin measurement. Ann Emerg Med 2010;56:382–8.
7. Shamir MY, Avramovich A, Smaka T. The current status of continuous noninvasive measurement of total, carboxy, and methemoglobin concentration. Anesth Analg 2012;114:972–8.
8. Maisel WH, Lewis RJ. Noninvasive measurement of carboxyhemoglobin: how accurate is accurate enough? Ann Emerg Med 2010;56:389–91.
9. Feiner JR, Rollins MD, Sall JW, Eilers H, Au P, Bickler PE. Accuracy of carboxyhemoglobin detection by pulse CO-oximetry during hypoxemia. Anesth Analg 2013;117:847–58.
10. Caboot JB, Jawad AF, McDonough JM, Bowdre CY, Arens R, Marcus CL, Mason TB, Smith-Whitley K, Ohene-Frempong K, Allen JL. Non-invasive measurements of carboxyhemoglobin and methemoglobin in children with sickle cell disease. Pediatr Pulmonol 2012;47:808–15.
11. Urpalainen K. Development of a fractional multi-wavelength pulse oximetry algorithm. MS Thesis, 2011Espoo.
12. Huiku M, Tolonen A, Ostroverkhov VP, Urpalainen K. Patent Application. US20120253147 A1. Calibration method and arrangement and sensor for non-invasively measuring blood characteristics of a subject, 2011.
13. Abo Alam K. Fuzzy Logic Hemoglobin Sensors. PhD Thesis, 2011Karlsruhe.
14. Kraitl J. Die nichtinvasive Bestimmung der Hämoglobinkonzentration im Blut mittels Pulsphotometrie. PhD Thesis, 2008Rostock.
15. Aldrich TK. Patent. US5810723 A. Non-invasive carboxyhemoglobin analyzer, 1998.
16. Bohnert M, Weinmann W, Pollak S. Spectrophotometric evaluation of postmortem lividity. Forensic Sci Int 1999;99:149–58.
17. Bohnert M, Walther R, Roths T, Honerkamp J. A Monte Carlo-based model for steady-state diffuse reflectance spectrometry in human skin: estimation of carbon monoxide concentration in livor mortis. Int J Legal Med 2005;119:355–62.
18. Belenki L, Sterzik V, Schulz K, Bohnert M. Analyzing reflectance spectra of human skin in legal medicine. J Biomed Opt 2013;18:17004.
19. Horecker BL. The absorption spectra of hemoglobin and its derivatives in the visible and near-infrared regions. J Biol Chem 1943;148:178–183.
20. Zijlstra WG, Buursma A, Meeuwsen-van der Roest WP. Absorption spectra of human fetal and adult oxyhemoglobin, de-oxyhemoglobin, carboxyhemoglobin, and methemoglobin. Clin Chem 1991;37:1633–8.
21. Zijlstra WG, Buursma A, van Assendelft OW. Visible and Near Infrared Absorption Spectra of Human and Animal Haemoglobin: Determination and Application. 2000Utrecht, Boston: VSP.
22. Kulcke A. Patent Application. WO2011161102 A1. Device and method for detecting and monitoring ingredients or properties of a measurement medium, in particular of physiological blood values, 2010.
23. Hall JW, Cadell TE. Patent. US5361758 A. Method and device for measuring concentration levels of blood constituents non-invasively, 1988.
24. Feiner JR, Severinghaus JW, Bickler PE. Dark skin decreases the accuracy of pulse oximeters at low oxygen saturation: the effects of oximeter probe type and gender. Anesth Analg 2007;105:S18–23.
25. International Organization for Standardization. ISO 80601-2-61:2011 Medical electrical equipment—Particular requirements for basic safety and essential performance of pulse oximeter equipment, 2011.
26. Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986;1:307–10.
27. Partridge R, Chee KJ, Suner S, Sucov A, Jay GD. Unsuspected carbon monoxide toxicity detected by noninvasive monitoring: a case report. Respir Care 2006;51:1324.
28. Deppe M. Pulsoximetrisches Monitoring von Methämoglobin bei Regionalanästhesien mit Prilocain. PhD Thesis, 2009Lübeck.
29. Boylston M, Beer D. Methemoglobinemia: a case study. Crit Care Nurse 2002;22:50–5.
30. Rehman HU. Methemoglobinemia. West J Med 2001;175:193–6.
31. McRobb C, Walczak R, Lawson S, Lodge A, Lockhart E, Bandarenko N III, Ing R. Carboxyhemoglobinemia in a pediatric cardiopulmonary bypass patient derived from a contaminated unit of allogenic blood. Perfusion 2011;26:302–7.
32. Feiner JR, Bickler PE. Improved accuracy of methemoglobin detection by pulse CO-oximetry during hypoxia. Anesth Analg 2010;111:1160–7.
33. Feiner JR, Bickler PE, Mannheimer PD. Accuracy of methemoglobin detection by pulse CO-oximetry during hypoxia. Anesth Analg 2010;111:143–8.