Evoked Potential Monitoring Identifies Possible Neurological Injury During Positioning for Craniotomy

Anastasian, Zirka H., MD*; Ramnath, Brian, EEG/EP T, CNIM; Komotar, Ricardo J., MD; Bruce, Jeffrey N., MD; Sisti, Michael B., MD; Gallo, Edward J., EEG T, CNIM; Emerson, Ronald G., MD; Heyer, Eric J., MD, PhD*†

doi: 10.1213/ane.0b013e3181b086bd
Technology, Computing, and Simulation: Case Report
Free

Somatosensory-evoked potential (SSEP) monitoring is commonly used to detect changes in nerve conduction and prevent impending nerve injury. We present a case series of two patients who had SSEP monitoring for their surgical craniotomy procedure, and who, upon positioning supine with their head tilted 30°–45°, developed unilateral upper extremity SSEP changes. These SSEP changes were reversed when the patients were repositioned. These cases indicate the clinical usefulness of monitoring SSEPs while positioning the patient and adjusting position accordingly to prevent injury.

From the Departments of *Anesthesiology, †Neurology, and ‡Neurological Surgery, New York-Presbyterian Hospital, Columbia University, New York City, New York.

Accepted for publication May 18, 2009.

EJH was supported in part by a grant from the NIA (RO1 AG17604).

Address correspondence and reprint requests to Zirka H. Anastasian, MD, Department of Anesthesiology, Columbia University, 622 West 168th St., New York City, NY 10032. Address e-mail to zh2114@columbia.edu.

Neurologic injury secondary to positioning is a significant perioperative problem and a common cause of patient injury in the practice of anesthesiology. Somatosensory-evoked potential (SSEP) monitoring is reproducible, reliable, and commonly used during surgical procedures to detect changes in electrophysiological conduction in peripheral nerves and central nerve pathways and, thus, to prevent nervous system damage.1 A significant change in the SSEP responses is indicated by a decrease in amplitude and/or an increase in latency. Changes in SSEP responses may be due to spinal instrumentation, hypoperfusion, hypothermia, anesthetic drugs, and positioning.

SSEP monitoring has also been noted to be useful in evaluating upper extremity conduction changes related to positioning. Changes in nerve conduction are expected to occur before permanent nerve injury, thus, repositioning that reverses SSEP conduction changes should prevent perioperative peripheral nerve injury.2–4

The following case series is the first, to our knowledge, to describe two cases of asymmetric SSEP changes in the brachial plexus after positioning for craniotomy that resolved after repositioning. It underlines the clinical usefulness of monitoring SSEPs while positioning the patient and adjusting the position accordingly to prevent peripheral neurological injury, a common cause of patient injury and malpractice claims.

Back to Top | Article Outline

CASE DESCRIPTIONS

Case 1

A 73-yr-old woman with a history of progressive loss of coordination and left-sided hearing presented for a suboccipital craniectomy for acoustic neuroma. Her medical history was significant for hypertension and diabetes. The magnetic resonance imaging showed an acoustic neuroma with enlarged ventricles and chronic displacement of the cerebellum and brainstem.

After induction of general anesthesia, the patient was positioned supine. Her head was turned approximately 45° toward the right and fixed with a Mayfield neurosurgical head holder. A “sandbag” was placed under the mattress on the left side of the operating table to extend the neck and left shoulder.

Electrophysiologic monitoring was performed bilaterally for SSEPs in upper (median nerve) and lower (posterior tibial nerve) extremities, brainstem auditory-evoked potentials, and motor-evoked potentials. SSEPs from the median nerves were recorded using needle electrodes placed at Erb’s point over the brachial plexus for assessing peripheral conduction and on the scalp at CP3 and CP4 locations, for assessing central conduction.

SSEP needle electrodes were placed on the patient before positioning. Positioning required 8 min, after which, SSEPs were recorded from both upper and lower extremities. Standard filter setting for these SSEPs is 30–1500 Hz, and standard stimulation parameters were used including rate of 4.7/s, 20 mA, and 0.3 ms duration. At the start of recording, after a 1-min interval of acquiring a steady signal, the SSEP amplitude from stimulation of the left median nerve was noted to be reduced at Erb’s point on all recording channels in comparison with contralateral recordings. When the patient’s left shoulder was raised to reduce extension of the neck and shoulder, the SSEPs from stimulation of the left median nerve were improved at recordings from Erb’s point and subcortical and cortical electrodes and became equal in amplitude to the contralateral side in 4 min (Fig. 1). Surgical exposure was not compromised significantly with this adjustment.

Figure 1.

Figure 1.

Back to Top | Article Outline

Case 2

A 62-yr-old man with a history of transient diffuse extremity weakness presented for left temporal and suboccipital craniotomy for subtotal resection of a skull base meningioma. His medical history was significant for hypertension, gout, and diverticulitis. A magnetic resonance imaging showed the presence of a left-sided mass consistent with a meningioma, at the level of the clivus and petrous bone, with compression of the pons and ventricular dilation.

After inducing general anesthesia, the patient was positioned supine. His head was turned toward the right by 30°, slightly extended and fixed with a Mayfield neurosurgical head holder. A “sandbag” was placed under the mattress on the left side of the operating table to extend the neck and left shoulder.

Electrophysiologic monitoring was performed bilaterally for SSEPs in upper (median nerve) and lower (posterior tibial nerve) extremities, and brainstem auditory-evoked potentials. Motor-evoked potentials were not obtained because most of the surgery was performed under microscope guidance. SSEPs from the median nerves were recorded using needle electrodes placed at Erb’s point over the brachial plexus for assessing peripheral conduction and on the scalp at CP3 and CP4 locations, for assessing central conduction.

SSEP needle electrodes were placed on the patient before positioning. Positioning required 10 min, after which, SSEPs were recorded from both upper and lower extremities. Standard filter settings for these SSEPs were the same as the previous case. At the start of recording, after a 1-min interval of acquiring a steady signal, the SSEP amplitude from stimulation of the left median nerve was reduced at Erb’s point on all recording channels in comparison with contralateral recordings. When the patient’s left shoulder was raised to reduce extension of the neck and shoulder, the SSEPs from stimulation of the left median nerve were improved at recordings from Erb’s point and subcortical and cortical electrodes and became equal in amplitude to the contralateral side in 6 min (Fig. 2). Surgical exposure was not compromised significantly in this position.

Figure 2.

Figure 2.

Both patients gave witnessed verbal informed consent to publish any collected data.

Back to Top | Article Outline

DISCUSSION

The reversibility of upper extremity SSEP changes with repositioning in our patients establishes a cause and effect relationship between positioning and SSEP change. Although we cannot be certain that the observed SSEP changes would have been associated with a permanent deficit, if the changes indeed heralded ischemia of the nerves, clinical neuropathy would be likely after a lengthy procedure. In addition, although not routine in our center for cases not involving cervical instability, these cases identify the utility of preposition baseline SSEPs, especially because the patients will be monitored during the case. This would clearly identify positioning changes by having pre- and postrecordings.

SSEP changes caused by positioning patients have been reported for prone2,5,6 and supine spine surgery,7–10 cardiac,2,4 and orthopedic surgeries.10–12 Our case report shows a change in SSEPs of the brachial plexus during positioning for craniotomy.

The incidence of SSEP changes during positioning of the prone spine surgery patient ranges from 3.6% to 15%. Schwartz et al.2 found a 30% decrease in ulnar nerve SSEP amplitude in 3.6% of pediatric patients having scoliosis correction. O’Brien et al.5 reported a 15% prevalence of brachial plexopathy during lumbar and spinal deformity surgery (≥60% decrease in amplitude or ≥10% increase in latency). Kamel et al.6 found that the lateral decubitus position (7.5%) and prone “superman” position (7.0%) had a higher incidence of position-related upper extremity SSEP changes compared with other positions including supine, arms tucked, and arms out (1.8%–3.2%).

The incidence of positioning changes in SSEP recordings during anterior spine surgery is lower than that of prone spine surgery.13 Schwartz et al.7 found that 1.8% of patients undergoing anterior cervical spine surgery showed intraoperative SSEP changes secondary to positioning. SSEP changes are more frequent in patients with myelopathy who were having anterior spine surgery.8

There has been only one study to describe positioning changes during SSEP monitoring for craniotomy surgery. Deinsberger et al.14 performed SSEP monitoring for positioning of patients for posterior fossa surgery in the semisitting position. Monitoring of the median and tibial nerve was recorded for 55 consecutive patients. SSEPs of the median nerve showed no changes while placing patients in the sitting position; however, tibial SSEP recordings were altered in 14 cases (25%).

We reviewed all cases of craniotomy for skull base tumors and acoustic neuromas in a 2-yr period to derive an incidence of SSEP changes during this category of neurosurgical procedure. At our institution, all craniotomy cases for acoustic neuromas and skull base tumors receive SSEP neuromonitoring. We found 83 cases performed in the 2-yr period between January 2007 and December 2008. Of these cases, we reviewed the neuromonitoring records and identified two cases with SSEP changes upon positioning, which are described in this case series: an incidence of 2.4%. This is approximately the incidence for SSEP changes for anterior cervical spine surgery (1.8%), but less than that found in prone and lateral decubitus positioning.

The clinical correlation of SSEPs and clinical findings has been the subject of several investigations. Prielipp et al.15 studied the onset of clinical symptoms and SSEP changes in the ulnar nerve in 50 volunteers with their arm in 30°–90° of abduction, as well as in supination, neutral orientation, and pronation. Half of the patients who developed SSEP changes did not develop clinical symptoms, two of which manifested severe SSEP changes (≥60% decrease in amplitude). In contrast, Lorenzini and Poterack16 monitored SSEPs at the median and ulnar nerves in awake volunteers placed in a prone position as their arms were moved in a step-wise cephalad position. Three of seven patients developed upper extremity symptoms described as tingling, numbness, or aching in the hand, forearm, or upper arm without changes in their SSEPs, suggesting that SSEP monitoring is imperfect in detecting positioning injury. However, volunteer studies do not apply directly to anesthetized patients, who are often placed in positions that awake patients could not tolerate.

These data demonstrate the false positive and negative predictive values of clinical peripheral neuropathic changes based on SSEP monitoring. There may, indeed, be instances in which SSEP changes, or the lack thereof, may be misleading in predicting postoperative clinical neuropathy. However, many studies that have evaluated SSEP monitoring with postoperative peripheral neurological changes studied SSEP changes over a short interval. In the study by Lorenzini and Poterack,16 recordings for SSEP signals were made 10–15 min after positioning the arm. The study by Prielipp et al.15 was performed for a maximum of 60 min, and on average, patients experienced clinical symptoms after 37 min, with a range of 20–59 min. In contrast, our first case took 6 h and 50 min, and the second case took 11 h and 33 min. It is unclear, but seems likely, that changes observed by SSEP monitoring would eventually lead to clinical neuropathy during these lengthy procedures.

To determine if we had clinical peripheral neuropathy changes that were not predicted by SSEPs, we reviewed the case records of our department obtained from resident postoperative visits, consults, and patient phone calls after hospital discharge. From January 2007 to December 2008, 65,041 cases were reviewed. Of all types of surgical cases, 10 cases were identified in which postoperative peripheral neuropathy was reported: an incidence of 0.02%. None of these cases were for craniotomy. The major limitation of this type of review is that it is retrospective, thus underreporting of a complication can underestimate the incidence of peripheral neuropathy. However, with SSEP monitoring for cases, we may be decreasing the incidence of peripheral neuropathy due to positioning.

Although the incidence of peripheral nerve injury is low, it does remain a common cause of professional liability in the practice of anesthesiology. Lee et al.17 reviewed the American Society of Anesthesiologists closed-claims analysis and found that peripheral nerve injury was the complication recorded in 21% of neurosurgical closed claims. This finding emphasizes the importance of taking measures to prevent peripheral nerve damage, including vigilance of neuromonitoring changes.

These cases demonstrate that changes in SSEPs of the brachial plexus can occur during positioning for skull base craniotomy and acoustic neuromas. This case series, in addition to the literature on this topic, underlines the clinical usefulness of monitoring SSEPs during and after positioning the patient and adjusting position accordingly to prevent neurological injury.

Back to Top | Article Outline

REFERENCES

1. Epstein NE, Danto J, Nardi D. Evaluation of intraoperative somatosensory-evoked potential monitoring during 100 cervical operations. Spine 1993;18:737–47
2. Schwartz DM, Drummond DS, Hahn M, Ecker ML, Dormans JP. Prevention of positional brachial plexopathy during surgical correction of scoliosis. J Spinal Disord 2000;13:178–82
3. Jellish WS, Martucci J, Blakeman B, Hudson E. Somatosensory evoked potential monitoring of the brachial plexus to predict nerve injury during internal mammary artery harvest: intraoperative comparisons of the rultract and pittman sternal retractors. J Cardiothorac Vasc Anesth 1994;8:398–403
4. Hickey C, Gugino LD, Aglio LS, Mark JB, Son SL, Maddi R. Intraoperative somatosensory evoked potential monitoring predicts peripheral nerve injury during cardiac surgery. Anesthesiology 1993;78:29–35
5. O’Brien MF, Lenke LG, Bridwell KH, Padberg A, Stokes M. Evoked potential monitoring of the upper extremities during thoracic and lumbar spinal deformity surgery: a prospective study. J Spinal Disord 1994;7:277–84
6. Kamel IR, Drum ET, Koch SA, Whitten JA, Gaughan JP, Barnette RE, Wendling WW. The use of somatosensory evoked potentials to determine the relationship between patient positioning and impending upper extremity nerve injury during spine surgery: a retrospective analysis. Anesth Analg 2006;102:1538–42
7. Schwartz DM, Sestokas AK, Hilibrand AS, Vaccaro AR, Bose B, Li M, Albert TJ. Neurophysiological identification of position-induced neurologic injury during anterior cervical spine surgery. J Clin Monit Comput 2006;20:437–44
8. Kombos T, Suess O, Da Silva C, Ciklatekerlio O, Nobis V, Brock M. Impact of somatosensory evoked potential monitoring on cervical surgery. J Clin Neurophysiol 2003;20:122–8
9. Jones SC, Fernau R, Woeltjen BL. Use of somatosensory evoked potentials to detect peripheral ischemia and potential injury resulting from positioning of the surgical patient: case reports and discussion. Spine J 2004;4:360–2
10. Swenson JD, Hutchinson DT, Bromberg M, Pace NL. Rapid onset of ulnar nerve dysfunction during transient occlusion of the brachial artery. Anesth Analg 1998;87:677–80
11. Mills WJ, Chapman JR, Robinson LR, Slimp JC. Somatosensory evoked potential monitoring during closed humeral nailing: a preliminary report. J Orthop Trauma 2000;14:167–70
12. Porter SS, Black DL, Reckling FW, Mason J. Intraoperative cortical somatosensory evoked potentials for detection of sciatic neuropathy during total hip arthroplasty. J Clin Anesth 1989;1:170–6
13. Labrom RD, Hoskins M, Reilly CW, Tredwell SJ, Wong PK. Clinical usefulness of somatosensory evoked potentials for detection of brachial plexopathy secondary to malpositioning in scoliosis surgery. Spine 2005;30:2089–93
14. Deinsberger W, Christophis P, Jodicke A, Heesen M, Boker DK. Somatosensory evoked potential monitoring during positioning of the patient for posterior fossa surgery in the semisitting position. Neurosurgery 1998;43:36–40; discussion 40–2
15. Prielipp RC, Morell RC, Walker FO, Santos CC, Bennett J, Butterworth J. Ulnar nerve pressure: influence of arm position and relationship to somatosensory evoked potentials. Anesthesiology 1999;91:345–54
16. Lorenzini NA, Poterack KA. Somatosensory evoked potentials are not a sensitive indicator of potential positioning injury in the prone patient. J Clin Monit 1996;12:171–6
17. Lee LA, Posner KL, Cheney FW, Domino KB. ASA closed claims project: an analysis of claims associated with neurosurgical anesthesia. Anesthesiology 2003;99:A362
© 2009 International Anesthesia Research Society