Skip Navigation LinksHome > May 1995 - Volume 20 - Issue 9 > Evaluation of Intrapedicular Screw Position Using Intraopera...
Spine:
Surgery: PDF Only

Evaluation of Intrapedicular Screw Position Using Intraoperative Evoked Electromyography.

Maguire, J MD; Wallace, S MD; Madiga, R MD; Leppanen, R PhD; Draper, V PhD

Collapse Box

Abstract

Study Design. This study analyzed the parameters needed for electrical stimulation of vertebral pedicle drill bits and screws.

Objectives. The feasibility of using electrically evoked electromyography was studied.

Summary of Background Data. Considerable potential for damage of adjacent nerve roots with incorrectly placed bits and screws exists with pedicular fixation of the lumbar spine.

Methods. Ninety-five drill bits, one hundred fortyfour screws, and thirty-four exposed nerve roots were electrically stimulated in thirty-six patients. Simultaneous evoked electromyographic activity was recorded from four muscle groups in each lower extremity during partial neuromuscular blockade.

Results. A constant current threshold of 6 mA or less correlated with misplaced drill bits and screws that broached the cortex. Evoked electromyography was 93% sensitive whereas radiography was only 63percnt; sensitive to detecting bits and screws that had broached the cortex.

Conclusions. Evoked electromyography is a valuable sensitive adjunct to radiographic examination of pedicle drill bit and screw placement.

(C) Lippincott-Raven Publishers.

Follow Us!

  

Login

Search for Similar Articles
You may search for similar articles that contain these same keywords or you may modify the keyword list to augment your search.