Pulse oximeter accuracy and precision affected by sensor location in cyanotic children

Sedaghat-Yazdi, Farshad MD; Torres, Adalberto Jr MD, FCCM; Fortuna, Randall MD; Geiss, Dale M. MD

Collapse Box

Abstract

Objective: Children's digits are often too small for proper attachment of oximeter sensors, necessitating sensor placement on the sole of the foot or palm of the hand. No study has determined what effect these sensor locations have on the accuracy and precision of this technology. The objective of this study was to assess the effect of sensor location on pulse oximeter accuracy (i.e., bias) and precision in critically ill children.

Design: Prospective, observational study with consecutive sampling.

Setting: Tertiary care, pediatric intensive care unit.

Patients: Fifty critically ill children, newborn to 2 yrs of age, with an indwelling arterial catheter. Forty-seven of 50 (94%) patients were postcardiac surgery.

Interventions: None.

Measurements and Main Results: Co-oximeter-measured arterial oxygen saturation (Sao2) was compared with simultaneously obtained pulse oximetry saturations (Spo2). A total of 98 measurements were obtained, 48 measurements in the upper extremities (finger and palm) and 50 measurements in the lower extremities (toe and sole). The median Sao2 was 92% (66% to 100%). There was a significant difference in bias (i.e., average Spo2 - Sao2) and precision (±1 sd) when the sole and toe were compared (sole, 2.9 ± 3.9 vs. toe, 1.6 ± 2.2, p = .02) but no significant difference in bias and precision between the palm and the finger (palm, 1.4 ± 3.2 vs. finger, 1.2 ± 2.3, p = .99). There was a significant difference in bias ± precision when the Sao2 was <90% compared with when Sao2 was ≥90% in the sole (6.0 ± 5.7 vs. 1.8 ± 2.1, p = .002) and palm (4.5 ± 4.5 vs. 0.7 ± 2.4, p = .006) but no significant difference in the finger (1.8 ± 3.8 vs. 1.1 ± 1.8, p = .95) or toe (1.9 ± 2.9 vs. 1.6 ± 1.9, p = .65).

Conclusions: The Philips M1020A pulse oximeter and Nellcor MAX-N sensors were less accurate and precise when used on the sole of the foot or palm of the hand of a child with an Sao2 <90%.

©2008The Society of Critical Care Medicine and the World Federation of Pediatric Intensive and Critical Care Societies

You currently do not have access to this article.

You may need to:

Note: If your society membership provides for full-access to this article, you may need to login on your society’s web site first.

Article Tools

You currently do not have access to this article.

You may need to:

Note: If your society membership provides for full-access to this article, you may need to login on your society’s web site first.

Search for Similar Articles
You may search for similar articles that contain these same keywords or you may modify the keyword list to augment your search.