Optometry & Vision Science

Home Current Issue Previous Issues Published Ahead-of-Print For Authors Journal Info
Skip Navigation LinksHome > October 2006 - Volume 83 - Issue 10 > A Comparison of Wavefront Aberrations in Eyes Wearing Differ...
Optometry and Vision Science:
October 2006 - Volume 83 - Issue 10 - pp 769-774
doi: 10.1097/01.opx.0000236786.96023.9c
Original Article

A Comparison of Wavefront Aberrations in Eyes Wearing Different Types of Soft Contact Lenses

JIANG, HONGJUN MD; WANG, DAN MD; YANG, LINA MD; XIE, PEIYING MD; HE, JI C. PhD

Collapse Box

Abstract

Purpose. The purpose of this study is to investigate the optical quality of eyes wearing different types of soft contact lenses (sCLs) using wavefront aberrometry.

Method. A psychophysical aberrometer (WFA1000B; BriteEye, SuZhou, China) was used to measure wavefront aberrations for 56 eyes of 28 subjects under four conditions: 1) without contact lenses (non-CL), 2) with lathe-cut contact lenses (lathe-cut CL; Shuliang, Weicon Optics, Shanghai, China), 3) with cast-molded contact lenses (cast-molded CL; Vistavue, J&J Vision Care, U.S.), and 4) with spun-cast contact lenses (spun-cast CL; Lanfenzhiji, Justcome Optics, Beijing, China). The three types of sCLs were made from different materials and lens design. The root mean square (RMS) values of the wavefront aberrations and individual Zernike aberrations were calculated and compared among the CL-wearing conditions.

Results. Astigmatism was not significantly changed by any of the three soft contact lenses when compared with the non-CL condition. However, a significant difference in the RMS values of higher-order aberrations was observed between the lens types (F = 8.29, p < 0.0001). Induced higher-order aberrations were found for both the cast-molded CL and the spun-cast CL conditions. There were significant differences in the aberrations in each of the higher orders between the CL conditions (F = 3.77, p = 0.011 for third order; F = 5.31, p = 0.002 for fourth order; F = 13.86, p < 0.0001 for fifth and higher orders). More higher-order Zernike aberrations such as coma and spherical aberration were induced by cast-molded CLs and/or the spun-cast CLs.

Conclusion. Wavefront aberrations in the eyes with soft contact lenses vary from one individual lens type to another. Although the variation in wavefront aberrations could be attributed to the differences in the methods of manufacture, influences from other factors such as the lens material and lens design could not be excluded and deserve further investigation.

© 2006 American Academy of Optometry

Search for Similar Articles
You may search for similar articles that contain these same keywords or you may modify the keyword list to augment your search.