Optometry and Vision Science:
October 2006 - Volume 83 - Issue 10 - pp 740-744
doi: 10.1097/01.opx.0000232806.06677.97
Original Article

Rigid Gas-Permeable Contact Lens Base Curve Radius and Transmissibility Effects on Corneal Oxygen Uptake

FINK, BARBARA A. OD, PhD, FAAO; MITCHELL, G LYNN MAS, FAAO; HILL, RICHARD M. OD, PhD, FAAO

Collapse Box

Abstract

Purpose. The purpose of this study is to determine the effects of rigid gas-permeable contact lens thickness, base curve radius, and material permeability on corneal hypoxic stress.

Methods. Corneal oxygen uptake rates were measured with a Clark-type polarographic electrode on the right eye of 10 human subjects for the normal open eye (air) and after 5 minutes of static wear of rigid contact lenses of four cornea-to-contact lens base curve fitting relationships: 0.2 mm steeper-than-K (STK), 0.1 mm STK, on K, and 0.1 mm flatter-than-K (FTK). There were also four materials (polymethylmethacrylate [Dk = 0], lotifocon B [OP-2, Dk = 15.9], lotifocon A [OP-3, Dk = 30], lotifocon C [OP-6; Dk = 60]) and three center thicknesses (0.14, 0.28, and 0.53 mm for the OP-6 lenses and 0.14 mm for all other materials) with all other parameters being constant. Each subject participated in two identical sessions. A repeated-measures analysis of variance was performed to compare the mean response across lens materials/thicknesses and the four curvature values.

Results. Significant differences were found only for lens material/thickness (p < 0.0001). Although OP-2 and OP-6 (0.53 mm) were manufactured to have the same Dk/t, post hoc comparisons showed that the oxygen uptake rate with OP-6 (0.53 mm) was significantly lower. The same is also true for OP-3 and OP-6 (0.28 mm), with OP-6 (0.28 mm) having a significantly lower oxygen uptake rate.

Conclusions. In contrast to past studies with PMMA, cornea-to-contact lens base curve fitting relationship, with the lens materials and designs studied here, did not affect corneal hypoxic stress, and thick, high Dk lenses resulted in less change in corneal response than did thin, lower Dk lenses of the same Dk/t. This is attributed to the lens reservoir effect that has been previously described.

© 2006 American Academy of Optometry