Optometry & Vision Science:
Articles: Original Article

Aberrations of the Human Eye in Visible and Near Infrared Illumination

LLORENTE, LOURDES OD; DIAZ-SANTANA, LUIS PhD; LARA-SAUCEDO, DAVID BSc; MARCOS, and SUSANA PhD

Free Access
Article Outline
Collapse Box

Author Information

Instituto de Óptica “Daza de Valdés”, Consejo Superior de Investigaciones Científicas, Serrano, Madrid, Spain (LL, SM), Applied Vision Research Centre, Department of Optometry and Visual Science, City University, Northhampton Square, London, United Kingdom (LDS), Blackett Laboratory, Imperial College of Science Technology and Medicine, London, United Kingdom (DLS)

Received April 30, 2002;

revision received July 26, 2002.

Lourdes Llorente

Instituto de Óptica “Daza de Valdés”

Consejo Superior de Investigaciones Científicas

Serrano, 121, Madrid, 28006

Spain

e-mail: lourdes.lle@io.cfmac.csic.es

Collapse Box

Abstract

Purpose. In most current aberrometers, near infrared light is used to measure ocular aberrations, whereas in some applications, optical aberration data in the visible range are required. We compared optical aberration measurements using infrared (787 nm) and visible light (543 nm) in a heterogeneous group of subjects to assess whether aberrations are similar in both wavelengths and to estimate experimentally the ocular chromatic focus shift.

Methods. Ocular aberrations were measured in near infrared and visible light using two different laboratory-developed systems: laser ray tracing (LRT) and Shack-Hartmann. Measurements were conducted on 36 eyes (25 and 11 eyes, respectively), within a wide range of ages (20 to 71 years), refractive errors (−6.00 to +16.50), and optical quality (root mean square wavefront error, excluding defocus, from 0.40 to 9.89 μm). In both systems, wave aberrations were computed from the ray aberrations by modal fitting to a Zernike polynomial base (up to seventh order in laser ray tracing and sixth order in Shack-Hartmann). We compared the Zernike coefficients and the root mean square wavefront error corresponding to different terms between infrared and green illumination.

Results. A Student’s t-test performed on the Zernike coefficients indicates that defocus was significantly different in all of the subjects but one. Average focus shift found between 787 nm and 543 nm was 0.72 D. A very small percentage of the remaining coefficients was found to be significantly different: 4.7% of the 825 coefficients (25 eyes with 33 terms) for laser ray tracing and 18.2% of the 275 coefficients (11 eyes with 25 terms) for Shack-Hartmann. Astigmatism was statistically different in 8.3% of the eyes, root mean square wavefront error for third-order aberrations in 16.6%, and spherical aberration (Z40) in 11.1%.

Conclusions. Aerial images captured using infrared and green light showed noticeable differences. Apart from defocus, this did not affect centroid computations because within the variability of the techniques, estimates of aberrations with infrared were equivalent to those measured with green. In normal eyes, the Longitudinal Chromatic Aberration of the Indiana Chromatic Eye Model can predict the defocus term changes measured experimentally, although the intersubject variability could not be neglected. The largest deviations from the prediction were found on an aphakic eye and on the oldest subject.

In the last few years, there has been a renewed interest in the measurement and understanding of the aberrations of the human eye. Along with studies addressing important basic questions on physiological optics (i.e., change of aberrations with accommodation, 1,2 age, 3,4 retinal eccentricity, 5 refractive error 6,7), clinical applications of aberrometry are rapidly increasing. For example, it has been shown to be a useful tool in assessing keratoconus 8,9 or corneal transplantation. 10,11 In particular, aberrometry is of great use in refractive surgery, both as a tool to assess the outcomes of refractive surgery 12–15 and as a guide to optimize ablation algorithms to eventually compensate for the ocular aberrations. 16 In addition, static 17 or dynamic aberration correction, 18,19 with great potential for high-resolution ophthalmoscopy, 20 relies on the accurate measurement of aberrations.

All aberrometers are based on the common principle of measuring the slopes of the wave aberration, either as a light enters the eye (i.e., laser ray tracing, 21,22 spatially resolved refractometer, 23,24 Tscherning aberrometer 25 or the crossed-cylinder aberroscope 26) or as it emerges from the eye (Shack-Hartmann ocular wavefront sensor 27,28). Apart from the spatially resolved refractometer, which is a psychophysical technique (and therefore visible light must be used) the rest of these techniques measure the light reflected by the retina. Most of the currently available wavefront-sensing techniques use infrared (IR) illumination, which has several advantages over visible light. It is more comfortable for the patient because the human eye is less sensitive to IR 29; pupil dilation is not strictly required; the retina reflects a higher percentage of the incident light compared with shorter wavelengths 30; and backscatter by the anterior optics 31 is reduced. Dynamic measurement of aberrations is then possible using IR illumination 32 with natural accommodation because mydriasis (and its associated cycloplegic effects) is not necessary.

Although current aberration measurements are typically done with IR light, in most applications, data from visible light are required. For direct comparison between optical measurements (estimated from the wave aberration) and visual performance we need to make sure that the results obtained in IR light are equivalent to those obtained with visible light. This is particularly important if the measured wave aberration is planned to be used to guide ablation in refractive surgery procedures, where the aim is to improve the patient’s visual performance. Knowledge of the defocus shift between IR and visible wavelengths is essential if the results are to be used to predict refraction.

Previous measurements of aberrations at different visible wavelengths using a spatially resolved refractometer showed slight differences in some aberration terms as a function of wavelength. 33 The chromatic difference of focus agreed with previous psychophysical results from the literature; however, the longitudinal chromatic aberration (LCA) based on reflectometric double-pass measurements 34,35 has been reported to be lower than conventional psychophysical estimates. These results prompted us to revisit the question of whether reflections at different retinal layers may be the cause for the discrepancy. The following questions hence arise: (1) Are the aberrations measured with IR and green light equivalent? (2) Is the focus difference between IR and green predictable by the LCA (and therefore reasonably predictable across subjects), or can the relative differences in reflectance and scattering across wavelengths be affecting the aberration measurements?

There are two previous studies that compare visible and near infrared optical quality in the human eye. 36,37 Double-pass measurements of modulation transfer functions in IR and green light appear to be similar. In this previous study, subtraction of background halos (noticeably different between IR and green) was critical. 36 The other study used an objective crossed-cylinder aberroscope to measure aberrations and reported that aberrations are virtually identical in near IR and green light. 37 However, the data analysis was mainly qualitative and limited to three eyes.

In this paper, we compare ocular aberrations between near IR (786 nm for laser ray tracing [LRT] and 788 nm for Shack-Hartmann [SH]) and visible illumination (543 nm) measured with two objective techniques, LRT and SH. These are experimental systems developed at Instituto de Optica (CSIC), Madrid, Spain and Imperial College, London, UK, respectively, but the conclusions drawn here can be extrapolated to recent unrelated commercially available instruments, based on similar principles. We performed measurements on 36 subjects with a wide range of ages, refractions, and ocular conditions (including old and surgical eyes), thus covering a wide range of aberrations and, potentially, ocular and retinal structural differences.

Back to Top | Article Outline

METHODS

Laser Ray Tracing
Setup and Procedures.

The laser ray tracing technique (Fig. 1 A) has been described in detail elsewhere. 5,13,21,22,38 A set of 37 parallel laser pencils sequentially scans a 6.51-mm pupil in a 1-mm step-hexagonal pattern. Aerial images formed by the light reflected off the retina are simultaneously recorded on a high-resolution CCD camera. The centroid of each aerial image is estimated. The deviations of the centroids from the reference (which is the position of the centroid corresponding to the chief ray) are proportional to the local derivative of the wave aberration. The wave aberration is obtained from the sets of derivatives by means of a modal fitting to the Zernike polynomial basis (through seventh order). In previous studies using this technique, measurements were obtained using a 543-nm HeNe laser beam (Melles Griot, 5 mW). For this study, light from an IR (786 nm) laser diode coupled to an optical fiber (Schäfter + Kirchhoff, 15 mW) was inserted into the system using a pellicle beam splitter and coaligned to the green beam. Both lasers were attenuated, by means of neutral-density filters, and light exposure was at least one order of magnitude below safety limits. 39

Figure 1
Figure 1
Image Tools
Back to Top | Article Outline
Setting and Control Experiment.

Measurements were conducted at Instituto de Óptica, CSIC, Madrid, Spain. The system was calibrated to verify that it did not introduce chromatic aberration. For this purpose, we placed a calibrated aberrated phase plate 17 in front of a diffraction-limited artificial eye and measured its aberrations using green and IR light. Identical results were obtained for all aberration terms within the accuracy of the technique, including the defocus term, and replicated the nominal aberrations of the phase-plate.

Back to Top | Article Outline
Subjects.

We measured 25 eyes (labeled #1 to #25) from 16 subjects: 19 eyes were normal, one eye was aphakic (#8), and five eyes had undergone laser-assisted in situ keratomileuesis refractive surgery (#5, #6, #10, #12, and #13). Ages ranged from 20 to 71 years (mean ± SD, 33 ± 11 years), spherical error ranged from −6.00 to +16.50 D (mean, −1.62 ± 4.42 D), and astigmatism ranged from 3.78 to 0.07 D (mean, 1.07 ± 0.98 D).

Before the measurement, an informed consent form approved by institutional ethical committees was signed by each patient in accordance with the tenets of the Declaration of Helsinki. The pupil was dilated with one drop of Tropicamide 1%.

Back to Top | Article Outline
Measurements.

Subjects were stabilized with a dental impression and a forehead rest, and the eye was monitored with a CCD camera to ensure alignment of the pupil center to the optical axis of the instrument during the measurement. Spherical refractive errors were compensated with trial lenses when necessary.

Each session consisted of 10 runs, each run (37 images corresponding to the 37 rays sampling the pupil) lasted approximately 4 s. Five consecutive series were collected using green light (543 nm), and then five series were collected using near IR light (786 nm).

Back to Top | Article Outline
Shack-Hartmann
Setup and Procedures.

A schematic diagram of the SH wavefront sensor used in this study is shown in Fig. 1 B. A detailed description of a similar system can be found elsewhere 38,40–42 without the minor modifications introduced for this study. Light from an IR (788 nm) super luminescent diode (Anritsu, 10 μW) was introduced by means of a pellicle beam splitter and coaligned to the green (543 nm) He-Ne laser beam (Melles Griot, 1 mW) used in previous measurements. The He-Ne laser was spatially filtered and expanded before collimation, bringing the maximum power reaching the eye to <5 μW over an 8-mm diameter pupil. Further power reduction was achieved by reducing the beam diameter to 1.5 to 2 mm and by the use of neutral-density filters before spatial filtering. The super luminescent diode power was largely reduced after fiber coupling (to about 10% of its maximum nominal power), and further power reduction was electronically controlled with its driver. In all cases, the maximum power reaching the eye was at least one order of magnitude below the safety limits. 39 The principle of the SH system has been described extensively in the literature. A narrow collimated laser beam forms a spot on the retina, and the light reflected and emerging from the eye is sampled by a rectangular lenslet array placed on a plane conjugate to the eye pupil. A CCD camera, placed on the focal plane of the lenslet array and conjugate to the retina, is used to record the SH spot pattern. Deviations from the ideal SH spot pattern are proportional to the local slopes of the wave aberration.

For this study, wave aberration was estimated from measured slopes using a least-mean square procedure. Wave aberration was fitted to a sixth-order Zernike polynomial expansion (27 terms).

The size of each lenslet was 0.8 mm × 0.8 mm over the pupil plane, and the focal length was 35 mm. The pupil size was 6 mm.

Back to Top | Article Outline
Setting and Control Experiment.

Measurements were conducted at Imperial College of Science Technology and Medicine, London, United Kingdom. The system was calibrated to ensure that it did not introduce chromatic aberration. Two reference SH images using green and IR light were compared. The green reference was used to calculate the aberrations of the IR reference. The order of magnitude of every Zernike coefficient was always smaller than or equal to the standard deviations of any series of 10 measurements of ocular aberrations using only one wavelength. This procedure proves that no significant amount of chromatic aberration is introduced by the optics of the system.

Back to Top | Article Outline
Subjects.

We measured 11 normal eyes (labeled #26 to #36) (six subjects). Ages ranged from 22 to 26 years (mean, 23 ± 1.47 years), spherical error ranged from −6.00 to +0.75 D (mean, 2.51 ± 3.24 D), and astigmatism ranged from 0.07 to 4.00 (mean, 1.30 ± 1.5 D).

The institutional research and ethical committee approved the use of the wavefront sensor and the experimental design. Written consent was obtained from all subjects participating in the study according to the tenets of the Declaration of Helsinki. Pupils were dilated using Tropicamide 1% and Phenylephrine 2.5% 30 min before the beginning of the measurements.

Back to Top | Article Outline
Measurements.

Subjects were stabilized with the help of a dental impression, and the pupil of the eye was aligned to the optical axis of the instrument while it was continuously monitored with a CCD camera. The illumination source was used as the fixation point. Spherocylindrical refractive errors were compensated when necessary.

At least six series of 10 SH images were collected, three using green illumination (543 nm) and the rest using IR illumination (788 nm). Images with the same wavelength were collected consecutively.

Back to Top | Article Outline
Equivalence of LRT and SH

The equivalence of the SH and LRT wavefront sensors has been demonstrated in previous studies. 22,43 Control measurements on two subjects showed that the SH and LRT systems used in this study (in Madrid and London, respectively) provided similar aberrations in normal eyes. 38

Back to Top | Article Outline

RESULTS

Raw Data

Raw data obtained from both techniques consist of a set of aerial images (in different frames for the LRT or a single frame for the SH). Each image corresponds to a pupil position (entry pupil position for LRT and exit pupil position for SH).

Fig. 2 A and B shows a set of aerial images obtained with LRT for eye #5, for green and IR light, respectively. Each image has been placed at the corresponding entry pupil position. The intensity patterns differ significantly across wavelengths. Fig. 2 C shows the spot diagram (joint plot of the position of the centroids of the same set) corresponding to the average data of three consecutive runs with green light (crosses) and four consecutive runs with IR light (circles) for eye #5. The error bars indicate the standard deviation of the positions of the centroid between runs. Chromatic defocus is responsible for the consistent shift between wavelengths, which increases with entry pupil eccentricity.

Figure 2
Figure 2
Image Tools

Fig. 2 D and E shows SH images for green and IR light, respectively, for eye #29. The presence of a halo surrounding the centroid is more evident for the image with IR illumination than for that with green illumination. The spots at the upper right and the lower left corners of the image appear dimmer (particularly for green illumination) due to the use of crossed polarization between illumination and recording. 38 Fig. 2 F shows the SH centroids corresponding to D (crosses) and E (circles). As in LRT, the shift between the green and IR spots increases toward the periphery of the image.

Back to Top | Article Outline
Wave Aberration Maps

Fig. 3 shows wave aberration maps from LRT measurements for both wavelengths for third- and higher-order aberrations. Eyes #9 and #22 were normal eyes, whereas #13 had undergone laser-assisted in situ keratomileuesis surgery. Each map is the average of at least three experimental runs. Contour lines have been plotted every 0.2 μm. Fig. 4 shows wave aberration maps for three normal eyes (#29, #30, and #31) measured with SH for both wavelengths, excluding tilt and defocus. Contour lines have been plotted every 0.5 μm.

Figure 3
Figure 3
Image Tools
Figure 4
Figure 4
Image Tools

For both systems, the wave aberration patterns corresponding to green and IR wavelengths for the same subject are very similar.

Back to Top | Article Outline
Zernike Coefficients and Root Mean Square Wavefront Error (RMS)

Fig. 5 shows plots of sets of Zernike coefficients for green (crosses) and IR (circles) light for the same eyes as in Figs. 3 and 4. The coefficient ordering and normalization follows the Optical Society of America standardization committee recommendations. 44 First- and second-order terms have been cancelled to allow a higher-resolution view of higher-order terms.

Figure 5
Figure 5
Image Tools

Error bars represent the standard deviation of the measurement. Mean variability (standard deviation) averaged across Zernike coefficients and subjects was 0.10 ± 0.06 for green light and 0.07 ± 0.04 for IR light for the measurements performed with LRT and 0.019 ± 0.009 for green light and 0.015 ± 0.009 for IR light for the measurements performed with SH. The differences between the Zernike coefficients measured with green or IR light shown in Fig. 6 are within the inherent variability of the techniques.

Figure 6
Figure 6
Image Tools

We performed a univariate statistical analysis (Student’s t-test) on each Zernike coefficient for each eye to detect which subjects and particular terms showed significant differences (p < 0.01) when measured in green and IR light. For the 25 eyes measured with LRT only 39 coefficients (excluding defocus) of 825 (25 × 33 terms), i.e., 4.7%, were statistically different. The defocus term (Z02) was statistically different in 24 of the 25 eyes (96% of the subjects). All the other statistically different terms were randomly distributed.

For the 11 eyes measured with SH, 22% (61 of 275 [25 coefficients × 11 eyes]) of the coefficients were statistically different. The defocus term (Z02) was statistically different in all of the eyes. Among terms other than defocus, 50 of 275 (18%) were found to be significantly different.

Fig. 6 shows defocus for IR wavelength vs. green wavelength in diopters for all subjects. There is a good linear correlation (R2 = 0.976), and the slope of the linear fit is close to one (0.9615). The focus shift between IR and green given by the fitting equation is 0.722. The experimental focus shift was 0.78 ± 0.29 D.

Bar diagrams in Fig. 7 compare individual terms (astigmatism and spherical aberration) and the RMS including different terms obtained with green (black bars) and IR (gray bars) wavelengths for all subjects. Eyes #1 to #25 were measured with LRT, and eyes #26 to #36 were measured with SH. Asterisks indicate eyes that showed statistically significant differences (p < 0.01). Astigmatism (Fig. 7 A) was statistically different in three of the 36 subjects (8.3%). RMS for third-order aberrations (Fig. 7 B) was statistically different in six of the 36 subjects (16.6%). RMS for third- and higher-order aberrations (Fig. 7 C) was statistically different in five of the 36 subjects (13.5%). Spherical aberration (Z40) (Fig. 7 D) was statistically different in four of the 36 subjects (11.1%)

Figure 7
Figure 7
Image Tools

Only one eye (#35) came out significantly different for all the terms or orders reported above (RMS for third- and higher-order aberrations, third-order aberrations, spherical aberration, and astigmatism).

Back to Top | Article Outline

DISCUSSION

This study shows that although the intensity distribution of LRT aerial images or SH images is notably different between green and IR illumination, both wavelengths provide aberration estimates within the experimental error (except for defocus). Our sample includes eyes with large differences in optical quality (from normal eyes to surgical eyes) and ages (20 to 71 years), suggesting that this conclusion holds for most of the population.

Back to Top | Article Outline
Differences in Image Intensity Profiles

Fig. 2 shows relevant intensity differences between the aerial images obtained with IR and those obtained with green illumination. IR images are typically more spread and are surrounded by a broad halo. It has been suggested that most of the light contributing to the core of double-pass aerial images is probably due to the light captured and guided back from the photoreceptors. 36,45 The halo is probably produced by effects other than aberrations, such as retinal stray light scattered at the choroid. 46,47 Retinal scattering increases for longer wavelengths due to their deeper penetration within the retina and the choroid. 48,49

Some previous comparisons of optical quality in IR and green light were based on estimates from double-pass aerial images. We performed a computer simulation to evaluate the contribution to the aerial image spread caused by degradation other than the ocular aberrations and the influence of wavelength on this additional contribution. We simulated LRT double-pass aerial images from the estimated wave aberration function. LRT aerial images are the autocorrelation of the entry (first pass) and exit (second pass) point-spread function. The entry pupil is a narrow incoming Gaussian beam (variance = 0.1034 mm and 0.1332 mm, respectively, for green and IR illumination) and the exit pupil is a 3-mm circular pupil. The entry and exit pupil sizes correspond to the experimental values in the LRT setup. Insets in Fig. 8 show real images and simulated images corresponding to an entry pupil centered at coordinates (+1.5, −2.6 mm). Fig. 8 A and B shows experimental and simulated results for green and IR light, respectively, for eye #22. The plots represent the normalized radial intensity profile of the corresponding real (solid) and simulated (dashed) aerial images. The distance from the peak position to the zero position represents the centroid deviation from the chief ray (which is practically the same for the simulated and real images). The width of the simulated images accounts for the spread caused exclusively by the measured aberrations, whereas the real images are further enlarged by scattering and nonmeasured higher-order aberrations.

Figure 8
Figure 8
Image Tools

The SH images in Fig. 2 C also suggest a larger contribution of scattered light in IR. A crossed-polarization configuration was used, which explains the “polarization-cross” pattern observed in green light illumination. 38 Green illumination maximizes the light reflected by the photoreceptor outer segments, 50 which are thought to partly retain polarization. 51 Light multiply scattered by deeper layers (probably a significant component of the IR images 49) does not retain polarization, and, therefore, the SH spots will show little polarization-related intensity differences across the image. 38

The effects mentioned above affect the shape and intensity distribution of the aerial image and are critical in double-pass measurements of the optical quality of the eye. In this technique, modulation transfer function estimates are directly obtained from double-pass aerial images. An appropriate halo subtraction is critical to obtain modulation transfer functions in IR consistent to those measured in green light. 36 However, reflectometric techniques for wave aberration measurements only rely on centroid deviation computations, which, as we have shown, are not significantly affected by wavelength.

Back to Top | Article Outline
Chromatic Difference of Focus

The defocus term was significantly different across wavelengths in all but one subject. The mean focus difference between green and IR across subjects was 0.78 ± 0.29 D, close to the shift estimated by the linear fitting shown in Fig. 7 (0.72 D). This value agrees well, within the inherent variability, with the chromatic focus shift predicted by the Indiana chromatic reduced eye model. 52

where λG = 543 nm and λIR = 787 nm (mean between IR wavelength used was 786 nm for LRT and 788 nm for SH).

Thibos et al. 52 obtained the parameters of the eye model by fitting experimental data for a range of wavelengths between 400 nm and 700 nm and using Cornu’s expression for the dependence of the index of refraction with wavelength. Equation 1 agrees well with experimental data in the literature for wavelengths up to 760 nm (close to the wavelength used in this study), with variations close to the intersubject variability in our sample. 52 Whether this expression for the LCA still holds for longer wavelengths used in some commercial SH systems (i.e., 830 nm) remains to be studied. Typically, Cornu’s equation fails beyond the visible, and other expressions 53 should be used.

Equation 1
Equation 1
Image Tools

It has been frequently argued that differences in the retinal layer where light is reflected may cause differences between manifest refraction and retinoscopy. 54,55 Charman et al. 56 and Williams et al. 45 for red light, and later López-Gil and Artal 36 for near IR light showed that the differences between subjective and reflectometric focus were negligible and concluded that reflection contributing to the central core of the point-spread function occurred within the photoreceptor layer. Our results, based on the Zernike defocus term of wave aberration reflectometric estimates, also support this conclusion. The focus shift that we found is slightly lower than the chromatic shift prediction (by 0.10 D), which is consistent with a reflection plane behind the photoreceptor layer. However, this shift is of the order of the measurement error (0.12 D for green light and 0.08 D for IR light on average) and lower than the intersubject variability (0.29 D). We did not find any particular trend for the focus shift in normal, young subjects as a function of refractive error (coefficient of correlation, r = 0.166, p = 0.44). In addition, we did not find any particular difference for the focus shift in eyes with abnormal corneas by laser-assisted in situ keratomileuesis surgery. However, we found that the focus shift for the aphakic eye was much higher than the average (1.7 D). Our population did not sample different age groups homogeneously. However, we found a slight increase of focus shift with age (r = 0.45, p = 0.022). The majority of subjects were young or middle-aged (20 to 43 years old), and we could not find an aged-related trend (r = 0.26, p = 0.2).

Back to Top | Article Outline

CONCLUSION

We have shown the equivalence of high-order aberrations measured in visible or near infrared illumination with LRT and SH, at least within the accuracy of the techniques. The shift in the defocus term was consistent with the shift predicted by chromatic aberration. These results are relevant because typical commercial wavefront-sensing devices use infrared illumination. This wavelength has several advantages over visible illumination: it is more comfortable for the subject, pupil dilation is not essential, and light exposure can be lower due to the higher reflectance of the eye fundus and the better sensitivity of most of the photodetectors at this wavelength. We have shown that despite the longer tails of the aerial images at this wavelength, it can be successfully used in all the tested conditions, including old and surgical eyes.

We also provide an experimental value for the focus shift between near infrared (786 to 788 nm) and green (543 nm) illumination in two reflectometric aberrometers (LRT and SH). One of the most promising applications of wavefront-sensing devices is their use as sophisticated autorefractometers. They are now being applied for use in refractive surgery to guide ablation with the aim of compensating both low- (second order) and high-order (third and higher) aberrations. An accurate transformation of the IR estimates of spherical error into visible wavelengths is crucial to determine the actual correction that should be applied. We have shown that Thibos’s chromatic reduced eye model equation is a valid expression to predict focus shift for our wavelength. However, for longer wavelengths, there is no evidence of the validity of this equation, and new expressions for the refractive index and chromatic difference of refraction may need to be developed. In addition, we found that discrepancies can occur in aphakic eyes and that there might be age-dependent corrections to Equation 1. Several reports in the literature found differences in the LCA’s of aphakic eyes 57 and pseudoaphakic eyes 58 with respect to normal eyes. Possible age-related changes of LCA have been a matter of controversy. 59–62 Although much of these refractive discrepancies are small, their magnitude can be comparable to the higher-order aberrations, and therefore accurate predictions of spherical errors for visible light from IR measurements are important.

Back to Top | Article Outline

ACKNOWLEDGMENTS

This study was supported by grants CAM 08.7/0010./2000 from Comunidad Autónoma de Madrid, Spain and BFM 2002–02638, Ministerio de Ciencia y Tecnologia, Spain to S. Marcos, TCI98-0925-C02-01 from the Ministerio de Educación y Cultura, Spain to R. Navarro, CONACyT 150238 predoctoral fellowship, Mexico to D. Lara-Saucedo and CAM 03/0101/1999 to L. Llorente.

We thank S. Barbero for his valuable help with experimental measurements. E. Moreno Barriuso contributed in the early stages of this study. We are indebted to C. Dainty and the Photonics Optics Group at Imperial College for allowing the use and modifications of the wavefront sensor and to Y. Tsang from City University for his invaluable help during the experimental sessions.

Back to Top | Article Outline

REFERENCES

1. He JC, Burns SA, Marcos S. Monochromatic aberrations in the accommodated human eye. Vision Res 2000; 40: 41–8.

2. Atchison DA, Collins MJ, Wildsoet CF, Christensen J, Waterworth MD. Measurement of monochromatic ocular aberrations of human eyes as a function of accommodation by the Howland aberroscope technique. Vision Res 1995; 35: 313–23.

3. McLellan JS, Marcos S, Burns SA. Age-related changes in monochromatic wave aberrations of the human eye. Invest Ophthalmol Vis Sci 2001; 42: 1390–5.

4. Artal P, Berrio E, Guirao A, Piers P. Contribution of the cornea and internal surfaces to the change of ocular aberrations with age. J Opt Soc Am (A) 2002; 19: 137–43.

5. Navarro R, Moreno E, Dorronsoro C. Monochromatic aberrations and point-spread functions of the human eye across the visual field. J Opt Soc Am (A) 1998; 15: 2522–9.

6. Collins MJ, Wildsoet CF, Atchison DA. Monochromatic aberrations and myopia. Vision Res 1995; 35: 1157–63.

7. Marcos S, Moreno-Barriuso E, Llorente L, Navarro R, Barbero S. Do myopic eyes suffer from larger amount of aberrations? In: Thorn F, Troilo D, Gwiazda J, eds. Myopia 2000: Proceedings of the VIII International Conference on Myopia. Boston: The New England College of Optometry, 2000: 118–21.

8. Barbero S, Marcos S, Merayo-Lloves J, Moreno-Barriuso E. Validation of the estimation of corneal aberrations from videokeratography in keratoconus. J Refract Surg 2002; 18: 263–70.

9. Schwiegerling J. Cone dimensions in keratoconus using Zernike polynomials. Optom Vis Sci 1997; 74: 963–9.

10. López-Gil N, Marín JM, Castejón-Mochón J, Benito A, Artal P. Ocular and corneal aberrations after corneal transplantation. Invest Ophthalmol Vis Sci 2001; 42: S529.

11. Munson K, Hong X, Thibos LN. Use of a Shack-Hartmann aberrometer to assess the optical outcome of corneal transplantation in a keratoconic eye. Optom Vis Sci 2001; 78: 866–71.

12. Seiler T, Kaemmerer M, Mierdel P, Krinke HE. Ocular optical aberrations after photorefractive keratectomy for myopia and myopic astigmatism. Arch Ophthalmol 2000; 118: 17–21.

13. Moreno-Barriuso E, Lloves JM, Marcos S, Navarro R, Llorente L, Barbero S. Ocular aberrations before and after myopic corneal refractive surgery: LASIK-induced changes measured with laser ray tracing. Invest Ophthalmol Vis Sci 2001; 42: 1396–403.

14. Marcos S, Barbero S, Llorente L, Merayo-Lloves J. Optical response to LASIK surgery for myopia from total and corneal aberration measurements. Invest Ophthalmol Vis Sci 2001; 42: 3349–56.

15. Marcos S. Aberrations and visual performance following standard laser vision correction. J Refract Surg 2001; 17: S596–601.

16. MacRae SM, Schwiegerling J, Snyder R. Customized corneal ablation and super vision. J Refract Surg 2000; 16: S230–5.

17. Navarro R, Moreno-Barriuso E, Bará S, Mancebo T. Phase-plates for wave aberration compensation in the human eye. Optics Lett 2000; 25: 236–8.

18. Hofer H, Chen L, Yoon GY, Singer B, Yamauchi Y, Williams DR. Improvement in retinal image quality with dynamic correction of the eye’s aberrations. Optics Express 2001;8:631–3. Available at: http://www.opticsexpress.org/oearchive/source/31887.htm. Accessed August 28, 2002.

19. Fernandez EJ, Iglesias I, Artal P. Closed-loop adaptive optics in the human eye. Optics Lett 2001; 26: 746–8.

20. Burns SA, Marcos S, Elsnser AE, Bará S. Contrast improvement for confocal retinal imaging using phase correcting plates. Optics Lett 2002; 27: 400–2.

21. Navarro R, Losada MA. Aberrations and relative efficiency of light pencils in the living human eye. Optom Vis Sci 1997; 74: 540–7.

22. Moreno-Barriuso E, Marcos S, Navarro R, Burns SA. Comparing laser ray tracing, the spatially resolved refractometer, and the Hartmann-Shack sensor to measure the ocular wave aberration. Optom Vis Sci 2001; 78: 152–6.

23. He JC, Marcos S, Webb RH, Burns SA. Measurement of the wave-front aberration of the eye by a fast psychophysical procedure. J Opt Soc Am (A) 1998; 15: 2449–56.

24. Webb RH, Penney CM, Thompson KP. Measurement of ocular wavefront distortion with a spatially resolved refractometer. Appl Optics 1992; 31: 3678–86.

25. Mierdel P, Krinke HE, Wiegand W, Kaemmerer M, Seiler T. Measuring device for determining monochromatic aberration of the human eye. Ophthalmologe 1997; 94: 441–5.

26. Walsh G, Cox MJ. A new computerised video-aberroscope for the determination of the aberration of the human eye. Ophthalmic Physiol Opt 1995; 15: 403–8.

27. MacRae S, Fujieda M. Slit skiascopic-guided ablation using the Nidek laser. J Refract Surg 2000; 16: S576–80.

28. Liang J, Grimm B, Goelz S, Bille JF. Objective measurement of wave aberrations of the human eye with the use of a Hartmann-Shack wave-front sensor. J Opt Soc Am (A) 1994; 11: 1949–57.

29. Liang J, Williams DR. Aberrations and retinal image quality of the normal human eye. J Opt Soc Am (A) 1997; 14: 2873–83.

30. Wyszecki G, Stiles WS. Color Science: Concepts and Methods, Quantitative Data and Formulae. New York: Wiley, 1982.

31. Delori FC, Burns SA. Fundus reflectance and the measurement of crystalline lens density. J Opt Soc Am (A) 1996; 13: 215–26.

32. van den Berg TJ. Light scattering by donor lenses as a function of depth and wavelength. Invest Ophthalmol Vis Sci 1997; 38: 1321–32.

33. Hofer HJ, Porter J, Williams DR. Dynamic measurement of the wave aberration of the human eye. Invest Ophthalmol Vis Sci 1998; 39: S203.

34. Marcos S, Burns SA, Moreno-Barriusop E, Navarro R. A new approach to the study of ocular chromatic aberrations. Vision Res 1999; 39: 4309–23.

35. Rynders MC, Navarro R, Losada MA. Objective measurement of the off-axis longitudinal chromatic aberration in the human eye. Vision Res 1998; 38: 513–22.

36. Charman WN, Jennings JA. Objective measurements of the longitudinal chromatic aberration of the human eye. Vision Res 1976; 16: 999–1005.

37. Lopez-Gil N, Artal P. Comparison of double-pass estimates of the retinal-image quality obtained with green and near-infrared light. J Opt Soc Am (A) 1997; 14: 961–71.

38. Lopez-Gil N, Howland HC. Measurement of the eye’s near infrared wave-front aberration using the objective crossed-cylinder aberroscope technique. Vision Res 1999; 39: 2031–7.

39. Marcos S, Diaz-Santana L, Llorente L, Dainty C. Ocular aberrations with ray tracing and Shack-Hartmann wave-front sensors: does polarization play a role? J Opt Soc Am (A) 2002; 19: 1063–72.

40. American National Standard Institute. American National Standard for the Safe Use of Lasers, Standard Z-136.1-1993. Orlando, FL: The Institute, 1993.

41. Diaz Santana HL, Dainty JC. Single-pass measurements of the wave-front aberrations of the human eye by use of retinal lipofuscin autofluorescence. Optics Lett 1999; 24: 61–3.

42. Diaz-Santana HL, Dainty JC. Single vs. symmetric and asymmetric double-pass measurements of the wavefront aberration of the human eye. In: Love GD, ed. Proceedings of the Second International Workshop on Adaptive Optics for Industry and Medicine: University of Durham, England, July 12–16, 1999. Singapore: World Scientific, 2000: 45–50.

43. Diaz-Santana HL. Wavefront sensing in the human eye with a Shack-Hartmann sensor. 2000. PhD Thesis. London: Imperial College of Science Technology and Medicine.

44. Moreno-Barriuso E, Navarro R. Laser ray tracing versus Hartmann-Shack sensor for measuring optical aberrations in the human eye. J Opt Soc Am (A) 2000; 17: 974–85.

45. Thibos LN, Applegate RA, Schwiegerling JT, Webb R. Standards for reporting the optical aberrations of eyes. In: Lakshminarayanan V, ed. OSA trends in optics and photonics. Vision Science and Its Applications, Vol 35. OSA Technical Digest Series. Washington, DC: Optical Society of America, 2000: 232–44.

46. Williams DR, Brainard DH, McMahon MJ, Navarro R. Double-pass and interferometric measures of the optical quality of the eye. J Opt Soc Am (A) 1994; 11: 3123–35.

47. Westheimer G, Liang J. Influence of ocular light scatter on the eye’s optical performance. J Opt Soc Am (A) 1995; 12: 1417–24.

48. Westheimer G, Campbell FW. Light distribution in the image formed by the living human eye. J Opt Soc Am 1962; 52: 1040–4.

49. Elsner AE, Burns SA, Hughes GW, Webb RH. Reflectometry with a scanning laser ophthalmoscope. Appl Optics 1992; 31: 3697–710.

50. Gorrand JM, Alfieri R, Boire JY. Diffusion of the retinal layers of the living human eye. Vision Res 1984; 24: 1097–106.

51. Elsner AE, Burns SA, Weiter JJ, Delori FC. Infrared imaging of sub-retinal structures in the human ocular fundus. Vision Res 1996; 36: 191–205.

52. Thibos LN, Ye M, Zhang XX, Bradley A. The chromatic eye: a new reduced-eye model of ocular chromatic aberration in humans. Appl Optics 1992; 31: 3594–600.

53. Herzberger M. Colour correction in optical systems and a new dispersion formula. Optica Acta 1969; 5: 197–215.

54. Millodot M, O’Leary DJ. On the artifact of retinoscopy and chromatic aberration. Am J Optom Physiol Opt 1980; 57: 822–4.

55. Millodot M, O’Leary D. The discrepancy between retinoscopic and subjective measurements: effect of age. Am J Optom Physiol Opt 1978; 55: 309–16.

56. Charman WN. Some sources of discrepancy between static retinoscopy and subjective refraction. Br J Physiol Opt 1975; 30: 108–18.

57. Millodot M. The influence of age on the chromatic aberration of the eye. Albrecht Von Graefes Arch Klin Exp Ophthalmol 1976; 198: 235–43.

58. Negishi K, Ohnuma K, Hirayama N, Noda T. Effect of chromatic aberration on contrast sensitivity in pseudophakic eyes. Arch Ophthalmol 2001; 119: 1154–8.

59. Howarth PA, Zhang XX, Bradley A, Still DL, Thibos LN. Does the chromatic aberration of the eye vary with age? J Opt Soc Am (A) 1988; 5: 2087–92.

60. Mordi JA, Adrian WK. Influence of age on chromatic aberration of the human eye. Am J Optom Physiol Opt 1985; 62: 864–9.

61. Morrell A, Whitefoot HD, Charman WN. Ocular chromatic aberration and age. Ophthalmic Physiol Opt 1991; 11: 385–90.

62. Calver RI, Cox MJ, Elliott DB. Effect of aging on the monochromatic aberrations of the human eye. J Opt Soc Am (A) 1999; 16: 2069–78.

Cited By:

This article has been cited 47 time(s).

Optics Express
Ocular aberrations as a function of wavelength in the near infrared measured with a femtosecond laser
Fernandez, EJ; Unterhuber, A; Prieto, PM; Hermann, B; Drexler, W; Artal, P
Optics Express, 13(2): 400-409.

Vision Research
Longitudinal changes of optical aberrations in normal and form-deprived myopic chick eyes
de la Cera, EG; Rodriguez, G; Marcos, S
Vision Research, 46(4): 579-589.
10.1016/j.visres.2005.06.012
CrossRef
Journal of Refractive Surgery
Total and corneal optical aberrations induced by laser in situ keratomileusis for hyperopia
Llorente, L; Barbero, S; Merayo, J; Marcos, S
Journal of Refractive Surgery, 20(3): 203-216.

Journal of Cataract and Refractive Surgery
Off-axis refraction and aberrations following conventional laser in situ keratomileusis
Ma, LX; Atchison, DA; Charman, WN
Journal of Cataract and Refractive Surgery, 31(3): 489-498.
10.1016/j.jcrs.2004..05.059
CrossRef
Journal of Microelectromechanical Systems
Addressable microlens array to improve dynamic range of Shack-Hartmann sensors
Choo, H; Muller, RS
Journal of Microelectromechanical Systems, 15(6): 1555-1567.
10.1109/JMEMS.2006.886011
CrossRef
Journal of the Optical Society of America A-Optics Image Science and Vision
Optical aberrations of intraocular lenses measured in vivo and in vitro
Barbero, S; Marcos, S; Jimenez-Alfaro, I
Journal of the Optical Society of America A-Optics Image Science and Vision, 20(): 1841-1851.

Vision Research
Balance of corneal horizontal coma by internal optics in eyes with intraocular artificial lenses: Evidence of a passive mechanism
Marcos, S; Rosales, P; Llorente, L; Barbero, S; Jimenez-Alfaro, I
Vision Research, 48(1): 70-79.
10.1016/j.visres.2007.10.016
CrossRef
Journal of Modern Optics
Effects of higher-order aberrations on contrast acuity as a function of light level
Dalimier, E; Dainty, C; Barbur, JL
Journal of Modern Optics, 55(): 791-803.
10.1080/09500340701469641
CrossRef
Journal of Cataract and Refractive Surgery
Back-calculation to model strategies for pretreatment adjustment of the ablation sphere in myopic wavefront laser in situ keratomileusis
Arnalich-Montiel, F; Wilson, CM; Morton, SJ; Allan, BD
Journal of Cataract and Refractive Surgery, 35(7): 1174-1180.
10.1016/j.jcrs.2009.03.016
CrossRef
Journal of Vision
Accommodative lag and fluctuations when optical aberrations are manipulated
Gambra, E; Sawides, L; Dorronsoro, C; Marcos, S
Journal of Vision, 9(6): -.
ARTN 4
CrossRef
Journal of Modern Optics
Wavefront curvature sensing for the human eye
Torti, C; Gruppetta, S; Diaz-Santana, L
Journal of Modern Optics, 55(): 691-702.
10.1080/09500340701470037
CrossRef
Optical Review
Evaluating optical quality of a bifocal soft contact lens in near vision using a Shack-Hartmann wavefront aberrometer
Hirohara, Y; Mihashi, T; Suzaki, A; Kuroda, T; Kelly, JE; Maeda, N; Fujikado, T
Optical Review, 13(5): 396-404.

Optics Express
On the possibility of intraocular adaptive optics
Vdovin, G; Loktev, M; Naumov, A
Optics Express, 11(7): 810-817.

Adaptive Optics for Industry and Medicine, Proceedings
Wireless control of an LC adaptive lens
Vdovin, G; Loktev, M; Zhang, X
Adaptive Optics for Industry and Medicine, Proceedings, 102(): 45-51.

Journal of the Optical Society of America A-Optics Image Science and Vision
Measuring ocular aberrations in the peripheral visual field using Hartmann-Shack aberrometry
Atchison, DA; Scott, DH; Charman, WN
Journal of the Optical Society of America A-Optics Image Science and Vision, 24(9): 2963-2973.

Journal of the Optical Society of America A-Optics Image Science and Vision
Calculation of retinal image quality for polychromatic light
Ravikumar, S; Thibos, LN; Bradley, A
Journal of the Optical Society of America A-Optics Image Science and Vision, 25(): 2395-2407.

Optics Express
Temporal dynamics and statistical characteristics of the microfluctuations of accommodation: Dependence on the mean accommodative effort
Leahy, C; Leroux, C; Dainty, C; Diaz-Santana, L
Optics Express, 18(3): 2668-2681.

Investigative Ophthalmology & Visual Science
Postblink changes in the ocular modulation transfer function measured by a double-pass method
Montes-Mico, R; Alio, JL; Charman, WN
Investigative Ophthalmology & Visual Science, 46(): 4468-4473.
10.1167/iovs.05-0609
CrossRef
Japanese Journal of Ophthalmology
Evaluation of optical function using a new point spread function analysis system in cataractous and pseudophakic eyes: Preliminary results
Negishi, K; Kobayashi, K; Ohnuma, K; Ohno, K; Noda, T
Japanese Journal of Ophthalmology, 50(1): 12-19.
10.1007/s10384-005-0275-z
CrossRef
Journal of Comparative Physiology A-Neuroethology Sensory Neural and Behavioral Physiology
Aberrations of chick eyes during normal growth and lens induction of myopia
Kisilak, ML; Campbell, MCW; Hunter, JJ; Irving, EL; Huang, L
Journal of Comparative Physiology A-Neuroethology Sensory Neural and Behavioral Physiology, 192(8): 845-855.
10.1007/s00359-006-0122-9
CrossRef
Optics Express
Customized computer models of eyes with intraocular lenses
Rosales, P; Marcos, S
Optics Express, 15(5): 2204-2218.

Vision Research
Finite schematic eye models and their accuracy to in-vivo data
Bakaraju, RC; Ehrmann, K; Papas, E; Ho, A
Vision Research, 48(): 1681-1694.
10.1016/j.visres.2008.04.009
CrossRef
Vision Research
Optical aberrations in the mouse eye
de la Cera, EG; Rodriguez, G; Llorente, L; Schaeffel, F; Marcos, S
Vision Research, 46(): 2546-2553.
10.1016/j.visres.2006.01.011
CrossRef
Optics Express
Effect of optical correction and remaining aberrations on peripheral resolution acuity in the human eye
Lundstrom, L; Manzanera, S; Prieto, PM; Ayala, DB; Gorceix, N; Gustafsson, J; Unsbo, P; Artal, P
Optics Express, 15(): 12654-12661.

Optics Express
A wavelength tunable wavefront sensor for the human eye
Manzanera, S; Canovas, C; Prieto, PM; Artal, P
Optics Express, 16(): 7748-7755.

Journal of Modern Optics
Adaptive optics and vision
Hampson, KM
Journal of Modern Optics, 55(): 3425-3467.
10.1080/09500340802541777
CrossRef
Vision Research
Changes in through-focus spatial visual performance with adaptive optics correction of monochromatic aberrations
Guo, HQ; Atchison, DA; Birt, BJ
Vision Research, 48(): 1804-1811.
10.1016/j.visres.2008.04.033
CrossRef
Journal of Biomedical Optics
Laser-induced retinal damage threshold measurements with wavefront correction
Lund, BJ; Lund, DJ; Esdall, PR
Journal of Biomedical Optics, 13(6): -.
ARTN 064011
CrossRef
Clinical and Experimental Optometry
New methods and techniques for sensing the wave aberrations of human eyes
Lombardo, M; Lombardo, G
Clinical and Experimental Optometry, 92(3): 176-186.
10.1111/j.1444-0938.2009.00356.x
CrossRef
Journal of the Optical Society of America A-Optics Image Science and Vision
Chromatic dispersions of the ocular media of human eyes
Atchison, DA; Smith, G
Journal of the Optical Society of America A-Optics Image Science and Vision, 22(1): 29-37.

Journal of Vision
Influence of adaptive-optics ocular aberration correction on visual acuity at different luminances and contrast polarities
Marcos, S; Sawides, L; Gambra, E; Dorronsoro, C
Journal of Vision, 8(): -.
ARTN 1
CrossRef
Optics Express
Estimation of centroid positions with a matched-filter algorithm: relevance for aberrometry of the eye
Leroux, C; Dainty, C
Optics Express, 18(2): 1197-1206.

Journal of Modern Optics
Cone directionality from laser ray tracing in normal and LASIK patients
Marcos, S; Burns, SA
Journal of Modern Optics, 56(): 2181-2188.
10.1080/09500340902927074
CrossRef
Optics Express
Requirements for segmented correctors for diffraction-limited performance in the human eye
Miller, DT; Thibos, LN; Hong, X
Optics Express, 13(1): 275-289.

Journal of the Optical Society of America A-Optics Image Science and Vision
Characteristic functions of Hartmann-Shack wavefront sensors and laser-ray-tracing aberrometers
Bara, S
Journal of the Optical Society of America A-Optics Image Science and Vision, 24(): 3700-3707.

Journal of Modern Optics
RGB Shack-Hartmann wavefront sensor
Jain, P; Schwiegerling, J
Journal of Modern Optics, 55(): 737-748.
10.1080/09500340701467728
CrossRef
Journal of Refractive Surgery
Optical quality and depth-of-field of eyes implanted with spherical and aspheric intraocular lenses
Marcos, S; Barbero, S; Jimenez-Alfaro, I
Journal of Refractive Surgery, 21(3): 223-235.

Optics Express
Population study of the variation in monochromatic aberrations of the normal human eye over the central visual field
Sheehan, MT; Goncharov, AV; O'Dwyer, VM; Toal, V; Dainty, C
Optics Express, 15(): 7367-7380.

Journal of Vision
Myopic versus hyperopic eyes: axial length, corneal shape and optical aberrations
Llorente, L; Barbero, S; Cano, D; Dorronsoro, C; Marcos, S
Journal of Vision, 4(4): 288-298.
10.1167/4.4.5
CrossRef
Optics Express
Liquid crystal Adaptive Optics Visual Simulator: Application to testing and design of ophthalmic optical elements
Manzanera, S; Prieto, PM; Ayala, DB; Lindacher, JM; Artal, P
Optics Express, 15(): 16177-16188.

Journal of Vision
Visual performance with real-life tasks under Adaptive-Optics ocular aberration correction
Sawides, L; Gambra, E; Pascual, D; Dorronsoro, C; Marcos, S
Journal of Vision, 10(5): -.
ARTN 19
CrossRef
Journal of the Optical Society of America A-Optics Image Science and Vision
Wide-angle chromatic aberration corrector for the human eye
Benny, Y; Manzanera, S; Prieto, PM; Ribak, EN; Artal, P
Journal of the Optical Society of America A-Optics Image Science and Vision, 24(6): 1538-1544.

Journal of Vision
Experimental validation of a Bayesian model of visual acuity
Dalimier, E; Pailos, E; Rivera, R; Navarro, R
Journal of Vision, 9(7): -.
ARTN 12
CrossRef
Investigative Ophthalmology & Visual Science
In Vivo Chromatic Aberration in Eyes Implanted With Intraocular Lenses
Perez-Merino, P; Dorronsoro, C; Llorente, L; Duran, S; Jimenez-Alfaro, I; Marcos, S
Investigative Ophthalmology & Visual Science, 54(4): 2654-2661.
10.1167/iovs.13-11912
CrossRef
Optometry & Vision Science
On-Eye Measurement of Optical Performance of Rigid Gas Permeable Contact Lenses Based on Ocular and Corneal Aberrometry
DORRONSORO, C; BARBERO, S; LLORENTE, L; MARCOS, S
Optometry & Vision Science, 80(2): 115-125.

PDF (964)
Optometry & Vision Science
Evaluation of Image Quality Metrics for the Prediction of Subjective Best Focus
Kilintari, M; Pallikaris, A; Tsiklis, N; Ginis, HS
Optometry & Vision Science, 87(3): 183-189.
10.1097/OPX.0b013e3181cdde32
PDF (509) | CrossRef
Optometry & Vision Science
Interocular Differences in Higher-Order Aberrations on Binocular Visual Performance
JimÉnez, JR; Castro, JJ; JimÉnez, R; Hita, E
Optometry & Vision Science, 85(3): 174-179.
10.1097/OPX.0b013e31816445a7
PDF (210) | CrossRef
Back to Top | Article Outline
Keywords:

ocular aberrations; Shack-Hartmann; laser ray tracing; near infrared

© 2003 American Academy of Optometry

Login

Search for Similar Articles
You may search for similar articles that contain these same keywords or you may modify the keyword list to augment your search.