Static and Dynamic Accommodation Measured Using the WAM-5500 Autorefractor

Win-Hall, Dorothy M.; Houser, Jamie; Glasser, Adrian

Optometry & Vision Science:
doi: 10.1097/OPX.0b013e3181f6f98f
Original Article
Abstract

Purpose. This study was undertaken to compare static and dynamic accommodation measurements using the Grand Seiko WR-5500 (WAM) in young, phakic subjects.

Methods. Fifteen subjects, aged 20 to 28 years (23.8 ± 0.58 years; mean ± SD years) participated. Accommodation was stimulated with printed text presented at various distances. In static mode, three measurements were taken for each stimulus amplitude. In dynamic mode, 5-Hz recordings were started, and subjects alternately looked through a transparent near chart and focused on a letter chart at 6 m for 5 seconds and then focused on the near letter chart for 5 seconds for a total of 30 seconds. After smoothing the raw data, the highest three individual values recorded in each 5-s interval of focusing at near were averaged for each stimulus amplitude. Analysis of variance and Bland-Altman analysis were used to compare the static and dynamic measurements. A calibration was performed with +3.00 to −10.00 D trial lenses behind an infrared filter, in 1.00 D steps in 5 of the 15 subjects.

Results. Stimulus-response graphs from static and dynamic modes were not significantly different in the lower stimulus range (<5.00 D, p = 0.93), but differed significantly for the higher stimulus amplitudes (p = 0.0027). One of the 15 subjects showed a significant difference between the static and dynamic modes. Corresponding pupil diameter could be recorded along with the accommodation responses for the subjects, and pupil diameter decreased with increasing stimulus demand. Calibration curves for static and dynamic measurements were not significantly different from the 1:1 line or from each other (p = 0.32).

Conclusions. Slight differences between the dynamically and statically recorded response amplitudes were identified. This is attributed to differences in the accommodative responses in this population and not to the instrument performance. Dynamic measurement of accommodation and pupil constriction potentially provides additional useful information on the accommodative response other than simply the response amplitude.

Author Information

*OD

PhD, FAAO

College of Optometry, University of Houston, Houston, Texas.

This study was supported, in part, by National Eye Institute Grant 1 RO1 EY017076 (to AG) and National Eye Institute Core Grant P30 EY007551 and National Eye Institute Summer Research Grant T35 EY007088 to the College of Optometry, University of Houston, Houston, Texas.

The authors have no commercial interests in the instrument used in this study or in the company that manufactures the instrument. The instrument was donated by AIT, US, and Grand-Seiko, Japan.

Received January 29, 2010; accepted July 21, 2010.

Adrian Glasser; University of Houston; College of Optometry; 505 J Davis Armistead Bldg; Houston, TX 77204-2020; email: aglasser@uh.edu

© 2010 American Academy of Optometry