Skip Navigation LinksHome > January 1992 - Volume 16 - Issue 1 > MR Measurement of Spinal CSF Flow with the RACE Technique.
Journal of Computer Assisted Tomography:
CENTRAL NERVOUS SYSTEM: PDF Only

MR Measurement of Spinal CSF Flow with the RACE Technique.

Kahn, Thomas; Müller, Edgar; Lewin, Jonathan S.; Mödder, Ulrich

Collapse Box

Abstract

The purpose of our study was the application and validation of a phase-sensitive pulse sequence that allowed real time CSF flow measurement without need for electrocardiographic (ECG) synchronization. After excitation of a slice perpendicular to the axis of the spine, projective data were obtained with a gradient echo sequence [contrast enhanced Fourier acquired steadystate technique (CE-FAST)] without spin warp gradient [real time acquisition and evaluation of motion technique (RACE)], allowing one-dimensional spatial resolution across the region of interest with a total sampling time of 20-30 ms. The sequence was calibrated with a spinal CSF phantom with oscillatory fluid motion. The calculated mean pulsation amplitudes of 20 healthy volunteers in the cervical region were 16 mm (range 9-36 mm), in the thoracic region 11 mm (5-21 mm), and in the lumbar region 3 mm (1-6 mm). The technique was capable of demonstrating physiologic alterations of CSF flow during respiratory maneuvers and may provide a tool to evaluate the altered CSF dynamics resulting from spinal block, inflammatory processes, or hemorrhage.

(C) Lippincott-Raven Publishers.

  

Login

Search for Similar Articles
You may search for similar articles that contain these same keywords or you may modify the keyword list to augment your search.