Skip Navigation LinksHome > October 1, 2006 - Volume 43 - Issue 2 > Validation of Primary CD4 Gating as an Affordable Strategy f...
JAIDS Journal of Acquired Immune Deficiency Syndromes:
doi: 10.1097/01.qai.0000242447.82403.c2
Clinical Science

Validation of Primary CD4 Gating as an Affordable Strategy for Absolute CD4 Counting in Cambodia

Lynen, Lut MD*; Teav, Syna BSc†; Vereecken, Chris BSc*; Munter, Paul De MD†*; An, Sokkab MD†; Jacques, Gary MD†; Kestens, Luc PhD*

Free Access
Article Outline
Collapse Box

Author Information

Received for publication December 8, 2005; accepted June 27, 2006.

From the *Clinical and Microbiology Department, Institute of Tropical Medicine, Antwerp, Belgium; and †Sihanouk Hospital Center of Hope, Phnom Penh, Cambodia.

Funded by the Belgian Directorate-General for Development Cooperation, project 95900.

Reprints: Luc Kestens, PhD, Unit of Immunology, Department of Microbiology, Institute of Tropical Medicine, Nationalestraat 155, B-2000 Antwerp, Belgium (e-mail: lkestens@itg.be).

Collapse Box

Abstract

Objective: To validate primary CD4 gating in lysed whole blood for absolute CD4 counts in fresh and aged blood using an affordable compact volumetric commercial flow cytometer.

Design: Comparison of CD4 counts between the FACSCount and the 2-parameter CyFlow SL Green.

Methods: One hundred twenty fresh blood samples from patients likely to be infected with HIV were simultaneously run on a FACSCount at the Pasteur Institute of Cambodia and on a CyFlow SL Green at the Sihanouk Hospital Center of Hope (SHCH), Phnom Penh, Cambodia. Intra- and interrun precision was assessed using 2 blood samples. Stability of CD4 counting in blood stored up to 96 hours at room temperature was assessed using 27 blood samples.

Results: CD4 counts on the CyFlow SL Green and on the FACSCount correlated well apart from a relative bias (R2 = 0.993, bias of −9.5%, 95% confidence interval [CI]: −11.8% to −7.1%, limits of agreement: −32.5% to 13.6%). Intra- and interrun variability ranged from 3% to 5% and from 5% to 6%, respectively. CD4 counts on aged blood using the CyFlow SL Green showed an interassay variability of <10%.

Conclusions: Primary CD4 gating in lysed whole blood using the CyFlow SL Green is an affordable and precise method for CD4 counting. Because the fluorescence (FL) and light scatter signals have to be analyzed manually, however, intensive training of the technician and/or operator is imperative.

With the reduced cost and complexity of antiretroviral therapy (ART), the number of people living with HIV/AIDS who have access to treatment is rapidly increasing in resource-limited settings. Unfortunately, optimal and affordable laboratory monitoring tools are often lacking in these settings. Clinical monitoring is considered a good basis but is too insensitive to identify patients who need ART and opportunistic infection (OI) prophylaxis.1 A rapid decline in total lymphocyte count (TLC) or hemoglobin was associated with progression to clinical AIDS in the Multicenter AIDS Cohort Study,2 but, in general, the TLC correlates only weakly with the CD4 count and scores even worse for monitoring of therapy.3-8

The classic multiparameter measurement of CD4+ T cells by flow cytometry is complicated and often too expensive for resource-limited countries.9-11 Less complex dedicated clinical instruments like the FACSCount require dedicated reagents and sample processing within 48 hours of venipuncture. This, together with numerous logistic problems, makes it is difficult for remote health centers in resource-limited settings to send their samples to a reference laboratory. The panleucogating approach has been an important step forward to extend sample storage before processing up to 96 hours after venipuncture.12 Alternative manual methods for CD4 counting are considered less appropriate in high-prevalence urban centers, where large numbers of samples have to be analyzed per day.13,14

In high-prevalence settings, there is a need for reliable CD4 counting methods in terms of accuracy and precision but at a reduced cost and allowing for high throughput analysis. Primary CD4 gating on volumetric flow cytometers, using only a single antibody (anti-CD4) has been put forward as an affordable and accurate alternative for resource-limited settings.15 We have evaluated this method under field conditions (Phnom Penh, Cambodia) using a 2-parameter (1 fluorescence [FL] and 1 side scatter [SSC] detector) volumetric flow cytometer, the CyFlow SL Green (FL1-SSC; Partec GmbH, Münster, Germany), with a lyse-no-wash CD4 staining protocol. The CyFlow SL Green provides direct absolute CD4 counts without the use of expensive reference microbeads. In addition, we evaluated whether this method could be used to analyze CD4+ T cells in unfixed whole blood stored between 24°C and 30°C for up to 96 hours after phlebotomy. Our method differs from the no-lyse-no-wash protocol described by Cassens et al.16 The lyse-no-wash procedure, in contrast to the unlysed blood procedure, does not cause perturbation of the light scatter properties of white blood cells by the presence of red blood cells. Therefore, it allows for better differentiation between CD4+ T cells and CD4-dim monocytes, particularly in aged blood stored at room temperature.

Back to Top | Article Outline

MATERIALS AND METHODS

Assessment of CD4 Counting on the CyFlow SL Green in Cambodia

To evaluate CD4 counting on the CyFlow SL Green, CD4 counting was performed on 2 different flow cytometers: (1) a dedicated clinical instrument for CD4 counting: the FACSCount (Becton Dickinson, Franklin Lakes, NJ) installed at the Pasteur Institute of Cambodia in Phnom Penh, which was the only available reference method for CD4 counting in Phnom Penh at the time of evaluation, and (2) the CyFlow SL Green (Partec), which is located at the Sihanouk Hospital Center of Hope (SHCH), a nongovernmental organization (NGO)-run hospital in Phnom Penh. The Cyflow SL Green is a compact mobile flow cytometer able to perform many types of cell analysis and absolute volumetric counting. The instrument is equipped with a 532-nm green solid-state laser, 1 FL detector, and 1 light SSC detector. Data acquisition, analysis, and real-time display were performed with a laptop computer using Windows FloMax software. A Cambodian laboratory technician from the SHCH was familiarized with the 2-parameter CyFlow SL Green and FloMax software through intensive training at the Institute of Tropical Medicine in Antwerp for 2 months. The training included proper interpretation of the light scatter and FL signals, correct setting of the gates and regions around CD4+ T cells, and calculation of the absolute number of CD4+ T cells. After training, the CyFlow SL Green was shipped to Cambodia and installed in the SHCH laboratory.

Back to Top | Article Outline
Blood Samples

One hundred twenty fresh whole-blood samples from patients for whom CD4 counting was requested and performed at the Pasteur Institute of Cambodia in Phnom Penh in May through June 2003 were used in this evaluation. All samples were sent by various doctors and organizations to the Pasteur Institute with the request to perform a CD4 count. The samples were not (re)tested for HIV in this study. In Cambodia, CD4 counting is done to assess the immune status of HIV patients. Therefore, the samples were categorized as likely to be HIV-positive. The blood samples were collected in EDTA-Vacutainer tubes (Becton Dickinson) and processed within 5 hours after collection. CD4 counting was performed on the same day, in a masked way, in parallel on the FACSCount at the Pasteur Institute and on the CyFlow SL Green at the SHCH. Transport between the Pasteur Institute and SHCH was organized once a day in the afternoon by a laboratory technician using an air-conditioned car. Samples were put in a rack in a safe container, and transport lasted 10 to 15 minutes. On arrival at the SHCH, the samples were immediately processed for the CyFlow SL Green. At the end of the evaluation, the absolute CD4 counts provided by the FACSCount were compared with those of the CyFlow SL Green.

Back to Top | Article Outline
Accuracy of CD4 Counting on the CyFlow SL Green

The FACSCount was used as predicate instrument for CD4 counting in this study. CD4 counting was performed according to the instructions provided by the manufacturer. In short, 50 μL of EDTA-blood was added to the dedicated FACSCount vials, which were then vortexed and left to incubate for 60 minutes at room temperature in the dark. Subsequently, a fixative was added, and the samples were incubated for 30 minutes at room temperature and then run on the FACSCount.

Direct volumetric CD4 measurements were performed on the CyFlow SL Green as follows: 50 μL of whole EDTA-blood was pipetted into polystyrene test tubes. Ten microliters of phycoerythrin (PE)-conjugated anti-CD4 monoclonal antibody (Becton Dickinson) was added to the blood in the test tubes and left to incubate for 15 minutes at room temperature. Before acquisition, 2 mL of red blood cell lysing solution (Becton Dickinson) was added for direct lyse-no-wash volumetric CD4 measurements on the CyFlow SL Green. The acquisition and analysis of lysed whole blood were done according to a strategy based on CD4 versus SSC gating, described as primary CD4 gating by Janossy et al.15 The concentration of CD4+ T cells per microliter was calculated by multiplying the counted CD4+ T cells by the dilution factor (41.2 = 50 μL of blood in a final volume of 2060 μL) and dividing the result by 200 (the CyFlow SL Green counts cells in a fixed volume of 200 μL).

Back to Top | Article Outline
Assessment of Precision of CD4 Counting

The precision of CD4 counting on the CyFlow SL Green instrument was assessed using 2 different blood samples. Intrarun (instrument) precision was determined by preparing a large volume (10 mL) of CD4-stained blood and repeating sample acquisition on the instrument at least 13 times to determine the instrument's precision. Interrun precision, which includes the variation induced by pipetting errors made by the technician (tube-to-tube variability), was assessed by repeating the entire CD4 staining procedure at least 10 times: pipetting, sample preparation, staining, lysing of red blood cells, and sample acquisition. By comparing the intrarun and interrun precision, a fairly good impression of instrument's and technician's performance can be obtained. Interperson variation (between different technicians) was not assessed. Precision is expressed as the coefficient of variance obtained by dividing SD of all the measurements by the mean (CV% = SD × 100/mean).

Back to Top | Article Outline
CD4 Counting in Aged Blood

We analyzed the accuracy of CD4 counting in EDTA-whole-blood samples from HIV-positive patients that were stored for up to 4 days (96 hours) at room temperature (24°C-30°C) without fixatives. Two sets of blood samples, 12 blood samples in the first set and 15 blood samples in the second set, were analyzed at an interval of 2 weeks. They were stained and analyzed by primary CD4 gating (CD4 vs. SSC) on the CyFlow SL Green after 0, 1, 2, 3, or 4 days of storage.

Back to Top | Article Outline
Statistical Analysis

Correlations between the absolute CD4 counts obtained by the different methods were analyzed by the Passing and Bablok method, which, in common with all nonparametric methods, is less sensitive to outliers.17 This method provides a test of the agreement of 2 analytic methods. Difference plots are given as proposed by Bland and Altman.18,19 The Bland-Altman method examines, in a discriminative fashion, whether the methods agree sufficiently well to be used interchangeably. The average of values seen by the 2 methods is displayed on the x axis and plotted against the difference between the 2 methods shown on the y axis.

The average difference between the 2 methods, referred to as bias, was marked on the graph by a horizontal line, and the mean difference and the limits of agreement with a 95% confidence interval (CI) were also calculated. The level of significance for linear regression was set at α < 0.05. MedCalc (Medcalc Software, Mariakerke, Belgium) version 8.1.0.0 statistical software was used to perform the Bland-Altman analyses. A modified Bland-Altman plot calculating the difference between 2 results and expressing it as a percentage of the mean of the 2 measurements was performed as well.20

Blood stability testing was analyzed using 27 samples, by comparing the results of absolute CD4 counting on days 1, 2, 3, and 4 with the results of day 0. To this end, we applied a repeated-measures linear regression model on log-transformed data with random intercept for each patient fitted by restricted maximum likelihood using Stata version 9.0 statistical software (StataCorp, College Station, TX).

To assess the clinical impact of using the CyFlow SL Green instead of the FACSCount in this setting, we calculated the sensitivity and the specificity of the CyFlow SL Green to identify patients who had a CD4 count <200 cells/μL with the FACSCount. A nonparametric ROC (receiver operating characteristic) curve was constructed to identify the optimal CD4 cutoff point of the CyFlow SL Green to detect a CD4 count <200 cells/μL on the FACSCount with MedCalc version 8.1.0.0 statistical software.21

Back to Top | Article Outline

RESULTS

Accuracy of Direct Volumetric CD4 Measurements on the CyFlow SL Green Using Lysed Whole Blood

Parallel CD4 measurements on both instruments, the FACSCount and CyFlow SL Green, were available for 115 of 120 tested blood samples. CD4 counting was automatically aborted by the FACSCount during sample acquisition of 5 blood samples, which were excluded from further analysis. The average CD4 count for 115 blood samples tested on the FACSCount was 289 ± SD 336 CD4+ T cells/μL, and results ranged from 1 to 2000 cells/μL (median = 183 cells/μL). Primary CD4 gating on the CyFlow SL Green using the same blood samples resulted in an average CD4 count of 268 ± SD 310 cells/μL, ranging from 1 to 1792 cells/μL (median = 170 cells/μL). An excellent correlation (R2 = 0.993) was observed between the absolute CD4 counts obtained by the FACSCount and those by the CyFlow SL Green via primary CD4 gating (Fig. 1A). The CyFlow SL Green CD4 results, however, showed a significant relative bias of −9.5% (95% CI: −11.8 to −7.1%) as compared to the FACSCount with limits of agreement between −32.5% and 13.6% after excluding CD4 counts of <10 cells/μL (FACSCount). This was done to avoid interference of large relative differences at low CD4 counts with little or no clinical impact. As shown in Figure 1B, the bias and limits of agreement were significantly distorted by the inclusion of low CD4 counts (<10 cells/μL). Considering the performance over the most relevant range between 50 and 750 CD4+ T cells/μL (n = 78), the relative bias was −8.1% (95% CI: −10.1 to −6.1%) when compared with the FACSCount, with limits of agreement between −25.4% and 9.4%. To investigate whether this bias would have a significant impact on clinical decision making during ART, we calculated the sensitivity and specificity of primary CD4 counting on the CyFlow SL Green to identify HIV patients having less than (or more than) 200 CD4+ T cells/μL by ROC analysis. At a cutoff of 200 CD4+ T cells/μL, the CyFlow SL Green has a sensitivity of 100% and a specificity of 96% in this population to identify patients with CD4 counts <200 cells/μL when compared with the FACSCount. The optimal cutoff would be 184 CD4+ T cells/μL (100% sensitivity and specificity).

Figure 1
Figure 1
Image Tools
Back to Top | Article Outline
Precision of Direct Volumetric CD4 Measurements on the Cyflow SL Green Using Primary CD4 Gating of Lysed Whole Blood

The intra- and interrun precision were tested using 2 blood samples. The intrarun precision of the instrument, expressed as the coefficient of variance, was 2.9% for the first sample (mean: 324 ± SD 9 CD4+ T cells/μL) and 5.1% for the second sample (mean: 182 ± SD 9 CD4+ T cells/μL). The interrun precision performed on the same 2 samples was 6% (319 ± SD 20 CD4+ T cells/μL) and 5.5% (mean: 172 ± SD 9 CD4+ T cells/μL), respectively.

Back to Top | Article Outline
Effect on Blood Stability of Direct Volumetric CD4 Measurements on the CyFlow SL Green Using Primary CD4 Gating of Lysed Whole Blood

The absolute CD4 count was determined using 27 blood samples and resulted in an average CD4 count of 241 ± SD 186 CD4+ T cells/μL and a standard error of the mean (SEM) of 36. The average CD4 counts (±SEM) on days 1, 2, 3, and 4 were 235 (±37), 233 (±35), 230 (±35), and 236 (±34), respectively, but remained within the range of the SEM (8.5%) of the average CD4 count on day 0. Figure 2A represents the individual CD4 counting in aging blood of 27 individuals with various CD4 counts at day 0. A repeated-measures linear regression model applied on log-transformed data from all 27 patients showed no significant change over the 5 consecutive days (P = 0.572). Overall, a relative change of 0.42% (95% CI: −1.0% to 1.9%) per day was observed. Repeating the same analysis per CD4 stratum, <300 CD4+ T cells and ≥300 CD4+ T cells, resulted in nonstatistically significant (P = 0.260) and borderline significant (P = 0.052) relative daily changes of 1.3% (95% CI: −0.98% to 3.7%) and −0.91% (95% CI: −1.9% to 0.01%), respectively. The cumulative relative daily changes were within the range of interrun precision using fresh blood (<6%). The interassay precision of absolute CD4 counting using aged blood (0-4 days), expressed as the coefficient of variance, was superior to the interrun precision using fresh blood, except for the case in which CD4 counts were lower than 100 cells/μL (see Fig. 2B). Figure 3 illustrates the deterioration of the quality of aged blood. Separation between CD4+ T cells in region 1 (RN1) and CD4-dim monocytes located at the left of RN1 was excellent on day 0 but became less clear on subsequent days. The separation between CD4+ T cells (RN1) and CD4-dim monocytes located at the lower right of RN1 in aged blood in our study was clearly facilitated by the use of 2 parameter (anti-CD4 and SSC) dot plots (left). This separation would have been much more problematic in single-parameter (anti-CD4) analyses, as illustrated by the histograms (right) in the same figure. Here, the operator has to select the optimal position of the left marker to delineate the position of CD4+ T cells (RN1). Because of the important overlap between CD4+ T cells and CD4-dim monocytes in aged blood, improper positioning of the left marker would result in an over- or underestimation of the absolute CD4 count.

Figure 2
Figure 2
Image Tools
Figure 3
Figure 3
Image Tools
Back to Top | Article Outline

DISCUSSION

The World Health Organization has published guidelines for scaling up ART using simple monitoring tools.22 These guidelines state that an absolute CD4 count is desirable to start and monitor patients on ART. Therefore, the development of affordable CD4 counting using compact and robust low-cost cytometers is mandatory. Clinical state-of-the-art flow cytometers used in the West cost approximately $100,000 US, and companies easily charge $15 to $25 US per test. Many of these instruments routinely measure CD4 percentage (CD4%) and can only provide absolute CD4 counts by using reagents that contain expensive reference beads (single-platform result). Alternatively, absolute CD4 counts can be obtained by multiplying the CD4% with the TLC (divided by 100) obtained from a hematology analyzer (double-platform result). The new generation of simplified flow cytometers can operate as single-platform volumetric instruments without the use of expensive microbeads (eg, CyFlow, Guava [Guava Technologies, Hayward, CA], Pointcare [PointCare Technologies, Marlborough, MA], Apogee A40 Analyzer [Amplimedical SpA, Assago, Italy]).23-25 The CyFlow SL Green and CyFlow Counter measure absolute CD4 counts by using a single monoclonal antibody and, as a consequence, are more affordable per CD4 test than instruments that require multiple monoclonal antibodies. At the time of the study, the FACS-Count cost more than $30,000 US and required expensive dedicated bead-based reagents that cost $15 US per test. The investment cost for a CyFlow instrument with 1 or 2 parameters is currently between $21,800 US (CyFlow Counter) and $26,975 US (CyFlow SL Green). These 1- and 2-parameter CyFlow versions can be operated using reagents from various manufacturers to reduce the cost to as low as $1 to $2 US per test.

Compact low-range flow cytometers are much more affordable and mobile, and they require less training of personnel than the expensive and sophisticated high-end clinical flow cytometers.10 We have shown that primary CD4 gating using lysed whole blood is an affordable method that can be implemented under field conditions on a 2-parameter CyFlow SL Green. Furthermore, analysis of unfixed aged blood stored at room temperature for several days (96 hours) was proven to be feasible on this instrument. A well-trained technician can use this method with low intra- and interrun variability comparable to other published CyFlow data.9,16,26,27 The blood stability study was performed on 27 randomly selected blood samples. The observed relative change of −0.91% per day in aged blood with CD4 >300 cells/μL was close to statistical significance (P = 0.052). The cumulative relative change over 4 days was −3.64%, however, which remains within the 6% coefficient of variance we obtained during our precision measurements using fresh blood samples. Therefore, we consider this relative change over 4 days of blood storage to be biologically insignificant. A weakness of this study is that the HIV serostatus of the patients was not (re)tested. The stability of the CD4 results over time in the group of patients with CD4 counts <300 cells/μL, representing those individuals most likely to be HIV-seropositive, was not worse than in the group with CD4 counts >300 cells/μL.

The absolute CD4 counts obtained with the CyFlow SL Green are, on average, 8.1% lower than those from the FACSCount for CD4 counts between 50 and 750 cells/μL. The tendency of the CyFlow SL Green to give lower CD4 counts than the predicate instruments is in agreement with other published data and is independent of the method (lyse-no-wash vs. no-lyse-no-wash).26-28

Fortunately, the bias we observed did not adversely affect clinical management or clinical decision making. All patients who had a CD4 count less than 200 cells/μL were correctly identified as such, and the 2 patients who were misclassified as having low CD4 counts on the CyFlow SL Green had CD4 counts of 214 and 219 cells/μL on the FACSCount and would benefit from treatment with ART according to most international ART guidelines.

Partec is currently promoting the CyFlow Counter as a dedicated instrument for absolute CD4 counting, which is a simplified system in which laser alignment by the operator is not required (Partec's AlignFree technology) as opposed to the CyFlow SL Green. Laser alignment requires high technical skills and regular maintenance. This prerequisite would be a drawback if the CyFlow SL Green is used by inexperienced or insufficiently trained laboratory staff. An advantage of the CyFlow SL Green is that the standard version is equipped with at least 2 detectors, which, as we have shown, allows for CD4 counting in aged blood for up to 4 days after blood collection. This is similar to the performance of the panleucogating protocol, which is based on CD45/SSC and subsequent CD4 gating.12 Because we do not know whether transport of blood samples would affect their stability over time, this should be tested. It would be possible for a mobile laboratory with a skilled laboratory technician and an instrument like the CyFlow SL Green to travel between different sites and perform analyses at the point of care with an interval of 5 days, as suggested by the manufacturer. The robustness of the CyFlow SL Green during transport should be formally tested before implementing this strategy, however. If proven successful, 1 well-trained laboratory technician and a single CyFlow SL Green instrument could cover several sites, which would further reduce the cost of ART monitoring.

Single-platform flow cytometers, such as the CyFlow Counter, the 2-parameter (SSC × FL) CyFlow SL Green, or the FACSCount, can only provide absolute CD4 counts. In children less than 6 years of age, however, it is recommended to assess the CD4% instead of the absolute CD4 count. For pediatric ART projects, it might be worthwhile to expand the basic configuration of the CyFlow SL Green instrument with a second FL detector (CD45) or a forward scatter (FSC) detector. This would enable the technician to obtain the TLC and total CD4 count from a single analysis on the same instrument, from CD45/SSC or FSC/SCC gating and subsequent CD4 gating. Although both gating strategies provide CD4%, CD45/SSC gating is preferred. It is more expensive (extra anti-CD45 antibody required) than FSC/SSC gating, but it allows for the use of aged blood.12

The CyFlow SL Green is a precise and fairly accurate instrument to measure absolute CD4 counts under field conditions if operated by well-trained technicians. Because intensive and lengthy training cannot always be provided to technicians working in the field, the analysis of the FL and SSC signals should ideally be performed by automated software and not manually by the technician. Clinical CD4 instruments, such as the FACSCount and the Pointcare, have built-in software that automatically analyses FL and SSC signals and only report CD4 results to the technician as CD4+ T cells/μL. In case the automated software cannot correctly identify CD4+ T-cell clusters, the instrument aborts the analysis and reports an error code. As a consequence, the results of these instruments are less dependent on training, which may be an important asset for ART roll-out. During our study we did not assess the technician-related variability of the CD4 results on the CyFlow SL Green. This could be done by comparing the CD4 results provided by different technicians with different levels of training who analyzed the same samples on the same instrument. To date, the CyFlow SL Green is one of the most affordable single-platform volumetric flow cytometers with low running costs. It has the potential to be an affordable solution for resource-limited settings in which high numbers of CD4 tests are needed at the lowest cost possible. Although the manufacturer states that the instrument does not require expensive preventive maintenance by engineers, daily maintenance and careful cleaning of the delicate flow cell by the operator is imperative for continued proper functioning of the instrument.

Whatever system is put in place, worldwide networks are urgently needed to validate the accuracy and, in particular, the robustness of this new generation of instruments under true field conditions so that technical support for these electronic and delicate instruments can be ascertained.

Back to Top | Article Outline

ACKNOWLEDGMENTS

The authors are grateful to the SHCH for making this evaluation possible and to the Pasteur Institute of Cambodia for performing the analyses on the FACSCount. They thank Professor Robert Colebunders for useful comments on the manuscript and Joris Menten for statistical review.

Back to Top | Article Outline

REFERENCES

1. Costello C, Nelson KE, Jamieson DJ, et al. Predictors of low CD4 count in resource-limited settings: based on an antiretroviral-naive heterosexual Thai population. J Acquir Immune Defic Syndr. 2005;39:242-248.

2. Lau B, Gange SJ, Phair JP, et al. Use of total lymphocyte count and hemoglobin concentration for monitoring progression of HIV infection. J Acquir Immune Defic Syndr. 2005;39:620-625.

3. Beck EJ, Kupek EJ, Gompels MM, et al. Correlation between total and CD4 lymphocyte counts in HIV infection: not making the good an enemy of the not so perfect. Int J STD AIDS. 1996;7:422-428.

4. Van der Ryst E, Kotze M, Joubert G, et al. Correlation among total lymphocyte count, absolute CD4+ count, and CD4+ percentage in a group of HIV-1-infected South African patients. J Acquir Immune Defic Syndr Hum Retrovirol. 1998;19:238-244.

5. Post FA, Wood R, Maartens G. CD4 and total lymphocyte counts as predictors of HIV disease progression. Q J Med. 1996;89:505-508.

6. Akanmu AS, Akinsete I, Eshofonie AO, et al. Absolute lymphocyte count as surrogate for CD4+ cell count in monitoring response to antiretroviral therapy. Niger Postgrad Med J. 2001;8:105-111.

7. Mane A, Patel A, Pujari S, et al. Total lymphocyte counts (TLC) is a poor surrogate for CD4 counts amongst asymptomatic HIV infected patients in resource limited settings [abstract 481]. Presented at: Second IAS Conference on HIV Pathogenesis and Treatment; 2003; Paris.

8. Kamya MR, Semitala FC, Quinn TC, et al. Total lymphocyte count of 1200 is not a sensitive predictor of CD4 lymphocyte count among patients with HIV disease in Kampala, Uganda. Afr Health Sci. 2004;4:94-101.

9. Greve B, Cassens U, Westerberg C, et al. A new no-lyse, no-wash flow-cytometric method for the determination of CD4 T cells in blood samples. Transfusion Medicine and Hemotherapy. 2003;30:8-13.

10. Mandy F, Bergeron MG, Fernandez-Sola A, et al. How to select, evaluate and maintain the correct CD4 T-cell counting method for your HIV/AIDS treatment clinic? Skill building workshop. Presented at: XIV International AIDS Conference; 2002; Barcelona.

11. Diagbouga S, Chazallon C, Kazatchkine MD, et al. Successful implementation of a low-cost method for enumerating CD4+ T lymphocytes in resource-limited settings: the ANRS 12-26 study. AIDS. 2003;17:2201-2208.

12. Glencross D, Scott LE, Jani IV, et al. CD45-assisted panleucogating for accurate, cost-effective dual-platform CD4+ T-cell enumeration. Cytometry. 2002;50:69-77.

13. Crowe S, Turnbull S, Oelrichs R, et al. Monitoring of human immunodeficiency virus infection in resource-constrained countries. Clin Infect Dis. 2003;37 (Suppl):S25-S35.

14. Didier JM, Kazatchkine MD, Demouchy C, et al. Comparative assessment of five alternative methods for CD4+ T-lymphocyte enumeration for implementation in developing countries. J Acquir Immune Defic Syndr. 2001;26:193-195.

15. Janossy G, Jani I, Gohde W. Affordable CD4(+) T-cell counts on ‘single-platform’ flow cytometers I. Primary CD4 gating. Br J Haematol. 2000;111:1198-1208.

16. Cassens U, Gohde W, Kuling G, et al. Simplified volumetric flow cytometry allows feasible and accurate determination of CD4 T lymphocytes in immunodeficient patients worldwide. Antivir Ther. 2004;9:395-405.

17. Passing H, Bablok W. A new biometrical procedure for testing the equality of measurements from two different analytical methods. Application of linear regression procedures for method comparison studies in clinical chemistry, Part I. J Clin Chem Clin Biochem. 1983;21:709-720.

18. Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;1:307-310.

19. Bland JM, Altman DG. Measuring agreement in method comparison studies. Stat Methods Med Res. 1999;8:135-160.

20. Pollock MA, Jefferson SG, Kane JW, et al. Method comparison-a different approach. Ann Clin Biochem. 1992;29:556-560.

21. Zweig MH, Campbell G. Receiver-operating characteristic (ROC) plots: a fundamental evaluation tool in clinical medicine. Clin Chem. 1993;39:561-577.

22. World Health Organization (WHO). Scaling up antiretroviral therapy in resource-limited settings: treatment guidelines for a public health approach, 2003 revision. Available at: http://www.who.int/hiv/pub/prev_care/en/arvrevision2003en.pdf.

23. Scott L, Kirkpatrick D, Hansen P, et al. PointCare CD4 testing: the new kid on the block [abstract U-132]. Presented at: 12th Conference on Retroviruses and Opportunistic Infections; 2005; Boston.

24. Kandathil AJ, Kannangai R, David S, et al. Comparison of microcapillary cytometry technology and flow cytometry for CD4+ and CD8+ T-cell estimation. Clin Diagn Lab Immunol. 2005;12:1006-1009.

25. Barbesti S, Soldini L, Carcelain G, et al. A simplified flow cytometry method of CD4 and CD8 cell counting based on thermoresistant reagents: implications for large scale monitoring of HIV-infected patients in resource-limited settings. Cytometry B Clin Cytom. 2005;68:43-51.

26. Fryland M, Chaillet P, Zachariah R, et al. The Partec CyFlow Counter® could provide an option for CD4+ T-cell monitoring in the context of scaling-up antiretroviral treatment at the district level in Malawi. Trans R Soc Trop Med Hyg [Epub ahead of print]. March 2006.

27. Pattanapanyasat K, Lerdwana S, Noulsri E, et al. Evaluation of a new single-parameter volumetric flow cytometer (CyFlow(green)) for enumeration of absolute CD4+ T lymphocytes in human immunodeficiency virus type 1-infected Thai patients. Clin Diagn Lab Immunol. 2005;12:1416-1424.

28. Dieye TN, Vereecken C, Diallo AA, et al. Absolute CD4 T-cell counting in resource-poor settings: direct volumetric measurements versus bead-based clinical flow cytometry instruments. J Acquir Immune Defic Syndr. 2005;39:32-37.

Cited By:

This article has been cited 9 time(s).

Cytometry Part B-Clinical Cytometry
CD4 T cell measurements in the management of antiretroviral therapy - A review with an emphasis on pediatric HIV-infected patients
O'Gorman, MRG; Zijenah, LS
Cytometry Part B-Clinical Cytometry, 74B(): S19-S26.
10.1002/cyto.b.20398
CrossRef
Cytometry Part A
Low-cost enumeration of CD4(+) T cells using a density-based negative selection method (RosetteSep (TM)) for the monitoring of HIV-Infected individuals in non-OECD countries
Bold, A; Wurth, R; Keller, T; Trahorsch, U; Voigt, P; Schubert, S; Sack, U
Cytometry Part A, 73A(1): 28-35.
10.1002/cyto.a.20494
CrossRef
Cytometry Part B-Clinical Cytometry
Simplified cytometry for routine monitoring of infectious diseases
Janossy, G; Shapiro, H
Cytometry Part B-Clinical Cytometry, 74B(): S6-S10.
10.1002/cyto.b.20405
CrossRef
Cytometry Part B-Clinical Cytometry
Clinical applications and availability of CD4+T cell count testing in sub-Saharan Africa
Taiwo, BO; Murphy, RL
Cytometry Part B-Clinical Cytometry, 74B(): S11-S18.
10.1002/cyto.b.20383
CrossRef
Cytometry Part B-Clinical Cytometry
Affordable CD4 T-cell enumeration for resource-limited regions: A status report for 2008
Mandy, F; Janossy, G; Bergeron, M; Pilon, R; Faucher, S
Cytometry Part B-Clinical Cytometry, 74B(): S27-S39.
10.1002/cyto.b.20414
CrossRef
Laboratoriumsmedizin-Journal of Laboratory Medicine
Determination of CD4 lymphocyte percentages using a new simplified and affordable no lyse flow cytometric protocol: a pilot study
Greve, B; Cassens, U; Baumann, C; Sibrowski, W; Kuling, G; Szabados, F; Schlenke, P; Reichelt, D; Gebauer, W; Gohde, W
Laboratoriumsmedizin-Journal of Laboratory Medicine, 32(3): 182-189.
10.1515/JLM.2008.026
CrossRef
Bmc Infectious Diseases
Computed CD4 percentage as a low-cost method for determining pediatric antiretroviral treatment eligibility
Callens, SFJ; Kitetele, F; Lusiama, J; Shabani, N; Edidi, S; Colebunders, R; Behets, F; Van Rie, A
Bmc Infectious Diseases, 8(): -.
ARTN 31
CrossRef
Plos One
CD4 T-Cell Enumeration in a Field Setting: Evaluation of CyFlow Counter Using the CD4 Easy Count Kit-Dry and Pima CD4 Systems
Wade, D; Diaw, PA; Daneau, G; Camara, M; Dieye, TN; Mboup, S; Kestens, L
Plos One, 8(9): -.
ARTN e75484
CrossRef
JAIDS Journal of Acquired Immune Deficiency Syndromes
Comparison of 5 Flow Cytometric Immunophenotyping Systems for Absolute CD4+ T-Lymphocyte Counts in HIV-1-Infected Patients Living in Resource-Limited Settings
Pattanapanyasat, K; Phuang-Ngern, Y; Sukapirom, K; Lerdwana, S; Thepthai, C; Tassaneetrithep, B
JAIDS Journal of Acquired Immune Deficiency Syndromes, 49(4): 339-347.
10.1097/QAI.0b013e31818c1721
PDF (544) | CrossRef
Back to Top | Article Outline
Keywords:

affordable flow cytometry; CD4 count; HIV; resource-limited settings

© 2006 Lippincott Williams & Wilkins, Inc.

Login

Search for Similar Articles
You may search for similar articles that contain these same keywords or you may modify the keyword list to augment your search.