Reliability of 2 Implant Stability Measuring Methods in Assessment of Various Periimplant Bone Loss: An In Vitro Study With the Periotest and Osstell Mentor

Choi, Han-Hee DDS, MSD*; Chung, Chae-Heon DDS, PhD; Kim, Su-Gwan DDS, PhD; Son, Mee-Kyoung DDS, PhD§

Implant Dentistry:
doi: 10.1097/ID.0000000000000000
Basic and Clinical Research
Abstract

Purpose: The aim of this in vitro study was to evaluate reliability of the Periotest and Osstell Mentor in assessment of periimplant vertical and circular bone loss.

Materials and Methods: Sixteen dental implants were embedded into acrylic resin blocks. The portion of resin around the implant neck was successively removed to mimic periimplant vertical and circular bone loss. Measurement values of 2 devices were compared by assessing the change of Periotest value (PTV) and Implant Stability Quotient (ISQ) values, by calculating correlation analysis and by means of regression analysis referring to increasing bone loss.

Results: Both devices were able to discriminate differences of circular bone loss. However, PTVs were not noticeably changed in buccal, buccal-mesial, and buccal-mesial-distal bone loss. The differences in buccal and buccal-mesial bone loss were not discriminated by Osstell Mentor. As the range of bone loss was increased, there was a noticeable correlation of the PTV and ISQ values.

Conclusions: The results of this study suggest that both diagnostic devices for implant stability are useful in detecting the circular bone loss. However, the clinical reliability of both devices for detecting the partial vertical bone loss is low.

Author Information

*Graduate Student, Department of Prosthodontics, School of Dentistry, Chosun University, Gwangju, Republic of Korea.

Professor and Chairman, Implant Center, School of Dentistry, Chosun University, Gwangju, Republic of Korea.

Professor and Director, Department of Oral and Maxillofacial Surgery, School of Dentistry, Chosun University, Gwangju, Republic of Korea.

§Assistant Professor and Director, Department of Prosthodontics, School of Dentistry, Chosun University, Gwangju, Republic of Korea.

Reprint requests and correspondence to: Mee-Kyoung Son, DDS, PhD, Department of Prosthodontics, School of Dentistry, Chosun University, 375, SeoSuk Dong, DongGu, GwangJu 501-759, Republic of Korea, Phone: 82-62-220-3825, Fax: 82-62-234-2119, E-mail: son0513@chosun.ac.kr

© 2014 by Lippincott Williams & Wilkins, Inc.