In Vitro Evaluation of Endothelial Cell Loss Using the Neusidl Corneal Inserter

Davis-Boozer, David MPH*; Terry, Mark A. MD; Greiner, Mark A. MD; Holiman, Jeff CEBT*; Saad, Hisham A. MD; Alqudah, Asem A. MD; Li, Jennifer Y. MD§

doi: 10.1097/ICO.0b013e318274a72b
Basic Investigation

Purpose: To determine the immediate endothelial cell loss (ECL) resulting from insertion of a precut donor button using the Neusidl Corneal Inserter (NCI) and compare it with the previously published ECL resulting from insertion of a folded donor button using non-coapting forceps.

Methods: Ten corneas were precut for Descemet stripping automated endothelial keratoplasty and trephinated to a diameter of 8.0 mm (n = 5) or 8.5 mm (n = 5). Each tissue was placed onto the platform of a new NCI spatula and inserted into a cadaveric whole globe through a 5.2 mm incision. The tissue was carefully removed and stained with trypan blue and alizarin red to detect damaged endothelium. ECL was estimated using Adobe Photoshop planimetry. Mean ECL was compared with previously reported studies of forceps insertion with a one-sample t test, using SPSS v. 19. Geographic patterns of ECL were also documented.

Results: Mean ECL was 15.6% (95% confidence interval, 13.8–17.4). We were unable to detect a difference in ECL compared with previous insertion methods studied (P < 0.001). The pattern of damage from the NCI was different than that previously seen with forceps insertion.

Conclusion: Immediate endothelial damage resulting from use of the NCI for insertion of Descemet stripping automated endothelial keratoplasty tissue is comparable with that seen with a standard forceps technique, but with a different damage pattern.

*Lions VisionGift, Portland, OR

Devers Eye Institute, Portland, OR

Department of Ophthalmology, Tanta Faculty of Medicine, Tanta, Egypt

§UC Davis Eye Center, Sacramento, CA.

Reprints: David Davis-Boozer, Devers Eye Institute, 1040 NW 22nd Ave, Suite 200, Portland, OR 97210 (e-mail: ddboozer@deverseye.org).

Fischer Surgical, Inc, provided funds to Lions VisionGift to cover the cost of tissue, materials, and employee time relating to this study. M. A. Terry receives royalties from Bausch & Lomb Surgical (St Louis, MO) for the DSAEK instruments that he designed.

The authors have no conflicts of interest to disclose.

Received July 11, 2012

Accepted September 14, 2012

© 2013 by Lippincott Williams & Wilkins.