Scanning Electron Microscopy Findings With Energy-Dispersive X-ray Investigations of Cosmetically Tinted Contact Lenses

Hotta, Fumika M.D.; Eguchi, Hiroshi M.D., Ph.D.; Imai, Shoji Ph.D.; Miyamoto, Tatsuro M.D., Ph.D.; Mitamura-Aizawa, Sayaka M.D., Ph.D.; Mitamura, Yoshinori M.D., Ph.D.

doi: 10.1097/ICL.0000000000000122
Article

Objective: To investigate the surfaces and principal elements of the colorants of cosmetically tinted contact lenses (Cos-CLs).

Methods: We analyzed the surfaces and principal elements of the colorants of five commercially available Cos-CLs using scanning electron microscopy with energy-dispersive x-ray analysis.

Results: In two Cos-CLs, the anterior and posterior surfaces were smooth, and colorants were found inside the lens. One lens showed colorants located to a depth of 8 to 14 μm from the anterior side of the lens. In the other lens, colorants were found in the most superficial layer on the posterior surface, although a coated layer was observed. The colorants in the other three lenses were deposited on either lens surface. Although a print pattern was uniform in embedded type lenses, uneven patterns were apparent in dot-matrix design lenses. Colorants used in all lenses contained chlorine, iron, and titanium. In the magnified scanning electron microscopy images of a certain lens, chlorine is exuded and spread.

Conclusions: Cosmetically tinted contact lenses have a wide variety of lens surfaces and colorants. Colorants may be deposited on the lens surface and consist of an element that has tissue toxicity.

Department of Ophthalmology (F.H., H.E., T.M., S.M.-A., Y.M.), Institute of Health Biosciences, The University of Tokushima Graduate School, Tokushima, Japan; and Division of Chemistry (S.I.), Institute of Socio-Arts and Sciences, The University of Tokushima, Tokushima, Japan.

Address correspondence to Hiorshi Eguchi, M.D., Ph.D., Department of Ophthalmology, Institute of Health Biosciences, The University of Tokushima Graduate School, 3-18-15, Kuramoto-cho, Tokushima-shi 770-8503, Japan; e-mail: hiroegu0113@gmail.com

F. Hotta, H. Eguchi, T. Miyamoto, S. Mitamura-Aizawa, and Y. Mitamura received a grant from Seed Co., Ltd. S. Imai has no conflicts of interest to disclose.

This is an open-access article distributed under the terms of the Creative Commons Attribution-NonCommercial-NoDerivatives 3.0 License, where it is permissible to download and share the work provided it is properly cited. The work cannot be changed in any way or used commercially.

Accepted October 27, 2014

Cosmetically tinted contact lenses (Cos-CLs), including colored corrective lenses, are widely available to the public and are especially popular among teenage girls and younger women in Pacific Asians. Although the sale of Cos-CLs is regulated by prescription in most Western countries, they can be easily purchased without a prescription in other countries. In such regions, young wearers use Cos-CLs with little or no supervision from professional eye care providers. Many of those wearers are uninformed about appropriate lens use and care, including proper hygiene practices for handling and storage. Therefore, we may see an increasing trend in corneal disorders related to the unregulated use of Cos-CLs.1–5

Lens materials and surfaces play an important role in the adherence of bacteria to the lens surface6–9 and are also believed to influence the manifestation of microbial keratitis related to contact lens (CL) wear.7–9 Given the potential role of Cos-CLs in microbial infections, a better understanding of the nature of the tinted surfaces in such lenses is desirable. Additionally, if colorants, which are printed, dot-matrix patterned tint on the anterior/posterior lens surfaces, potentially induce changes in corneal epithelial cells,10,11 the elements of those colorants should be fully published; however, this information is mostly unknown, or the characteristics including the manufacturing methods used for colorants are unclear in the information provided on the Internet.

In the article reported by Begley and Waggoner,12 nodular deposits on soft CLs were investigated using scanning electron microscopy with energy-dispersive x-ray analysis (SEM-EDX). The results provide a breakthrough for analysis of CL surface and indicated that most CL deposits contained calcium. The aim of this study was to use SEM-EDX to observe the surfaces of several commercially available Cos-CLs and to analyze the chemical elements in the colorants used in those lenses.

Back to Top | Article Outline

METHODS

Five commercially available Cos-CLs were investigated in this study. These lens types selected are widely used across Pacific Asia. The five Cos-CLs we assessed were obtained arbitrarily without a prescription from the Internet, a general pharmacy, or a mass merchandise outlet. The product names of the lenses and lens characteristics specified in the instructions or packaging are shown in Table 1.

TABLE 1

TABLE 1

For evaluation by SEM-EDX, the Cos-CLs were first washed by soaking for 10 min in their blister package with ultrapure water (Milli-Q academic; Merck Co., Ltd, Tokyo, Japan) after withdrawal of original saline in the package, cut to the suitable size, and then desiccated on a cellulose-mixed ester membrane filter (Advantec Toyo Co., Ltd, Tokyo, Japan) to flatten the lens. Specifically, the lenses were cut into approximate quarters before desiccation, and individual pieces were made and then desiccated. Afterward, the pieces were manually broken using clean, sterile forceps, and the piece that was apparently flat was observed using SEM. Both lens surfaces (both anterior and posterior) of all lenses were examined using the Tabletop Microscope TM3030 (Hitachi High-Technologies Corporation, Instruments Co., Ltd, Tokyo, Japan). After localizing the colorant, elemental analysis and their distribution maps were measured by an energy-dispersive x-ray spectroscopy using Quantax 70 (Bruker AXS Microanalysis GmbH, Berlin, Germany). When colorants were located inside the lenses, the cross sections of the lenses were examined to measure the depth, and the elements were also analyzed using the same instruments as described above.

Back to Top | Article Outline

RESULTS

Surface Observation and EDX-Mapping Analysis

Surface observations for each lens are shown in Table 2. Regarding EDX-mapping, carbon and oxygen are found to be diffusely distributed throughout lenses. Chlorine, titanium, and iron are found to be distributed where the colorants are embedded or printed.

TABLE 2

TABLE 2

Back to Top | Article Outline

1-Day Acuvue Define

The lens had a smooth surface on both sides (Fig. 1A,B). In the cross-sectional image, the colorants were found embedded inside the lens to a depth of 8 to 14 μm from the anterior surface (Fig. 1C,D).

FIG. 1

FIG. 1

Back to Top | Article Outline

Eye Coffret 1-Day UV

The anterior side was smooth (Fig. 2A). The print design was uniform (Fig. 2B). The posterior surface appeared wrinkled under high magnification (Fig. 2C). A thin-coated layer covering the colorants was detected, and the coated layer was wrinkled because of desiccation. The colorants were found embedded in the most superficial layer of the posterior surface of the lens at a depth of 1 μm. Only chlorine was found in the print (Fig. 2D).

FIG. 2

FIG. 2

Back to Top | Article Outline

Ever Color 1-Day Natural

Colorants were deposited on the posterior surface of the lens, although the anterior surface was smooth (Fig. 3A,B). The colorants were not printed uniformly, and small amounts of colorant were scattered between a dot-patterned print (Fig. 3B). Two kinds of print patterns were observed, and chlorine covered most of the print area, whereas titanium oxide seemed to be overglazed in the midperiphery of the print (Fig. 3C,D). In the magnified image, chlorine is exuded and spread (Fig. 3E,F).

FIG. 3

FIG. 3

Back to Top | Article Outline

Tutti Vanity Rich

Colorants were deposited on the anterior surface, although the posterior surface was smooth (Fig. 4A,B). Colorants on the anterior side were deposited heavily in the periphery, and a dot-patterned print was overglazed in the midperipheral area (Fig. 4A). Oxygen and iron were painted heavily on the peripheral side of the print and painted as a dot-pattern in the midperiphery (Fig. 4C,D).

FIG. 4

FIG. 4

Back to Top | Article Outline

Joy Pop Rich

Colorants were deposited on the posterior surface, and the anterior surface was smooth (Fig. 5A,B). In some areas of the lens, uneven coatings were apparent (Fig. 5B). Chlorine was printed heavily on the periphery, and prints were uneven. In some areas, iron seemed to be overglazed (Fig. 5C,D).

FIG. 5

FIG. 5

Back to Top | Article Outline

Elemental Analysis of Colorants

Results from the elemental analysis of colorants are shown in Table 2. Colorants detected in the cross section of 1-DAY ACUVUE DEFINE consisted of iron and oxygen (Fig. 1D). Colorants in EYE COFFRET 1-DAY UV consisted of chlorine (Fig. 2D). Colorants in EVER COLOR 1-DAY NATURAL consisted of chlorine and titanium (Fig. 3C,D). Colorants in TUTTI VANITY RICH consisted of iron and oxygen (Fig. 4C,D). Colorants in JOY POP RICH consisted of chlorine, iron, and oxygen (Fig. 5C,D).

Back to Top | Article Outline

DISCUSSION

This study showed a wide variation in the location, surface quality/roughness, and chemical composition of the tinted portion of the lenses examined. Lorenz et al.13 analyzed seven Cos-CLs using SEM and atomic force microscopy, and they also revealed that pigment on the surface of six lenses. For multiple reasons, the fact that some Cos-CLs have scabrous surfaces on the posterior sides is a matter of critical concern. Because most Cos-CLs are made of Poly (2-hydroxyethyl methacrylate), which has a lower oxygen permeability than recent silicone hydrogel, which is the material commonly used for corrective lenses. Although the oxygen permeability values (Dk value) of most Cos-CLs are provided, the oxygen transmissibility values (Dk/t value) are not disclosed. Moreover, the deposited colorants are a cause of friction on the corneal epithelium. Heavily applied colorants may alter the rigidity of the lens, and this alteration may also cause friction on the cornea. Corneal epithelial disorders with low oxygen concentration are an important risk factor for corneal infections related to CL wear.13–16 Chan et al.17 reported that surface pigments on Cos-CL resulted in significantly higher bacterial adherence after the examination of 15 commercially available Cos-CLs. They also concluded that bacterial adherence to surface pigments in some two lenses exhibited significantly higher adherence compared with their clear counterpart.

Colorants should be made of noncytotoxic or low-cytotoxic materials if they are deposited on the surfaces of lenses and touch the corneal epithelium directly. It is a great matter of concern that publicly available information about the principal elements of colorants is limited. This study found chlorine in the prints of three lenses and obtained an image suggesting that prints on EVER COLOR 1-DAY NATURAL exuded chlorine. We also found iron in the prints of four lenses. Initially, we presumed that chlorine originated from the saline in the blister package. However, we determined that chlorine was a component of the colorants because sodium was not detected from the area where we detected chlorine in EDX-mapping analysis. A more detailed investigation capable of identifying the chlorine element of organic compounds and the chlorine ion of inorganic salts is necessary, because we found no evidence indicating the presence of the counter ion of chlorine.

Moreover, the print technique is likely not secured in some lenses because colorants were deposited unevenly in the print areas of the same pattern. Colorants printed in a variety of textures may cause exposure of colorant elements to the corneal epithelium, and also may cause stronger friction than regularly patterned prints. In this study, it is possible that the uneven prints originated from irrigation in our experimental manipulation with ultrapure water. In that case, discoloration suggests that elements in colorants can easily be exposed in the process of routine daily lens care. In either case, the uneven print pattern that we found may be one of the risk factors for corneal infections and/or corneal insults related to Cos-CL wear. The amount and/or percentage of iron should be released to the public because it has been reported that overloaded iron lead to many diseases18 and to toxicity and cell death through free radical formation.19

Although we believe that this study provides a new insight into Cos-CLs, two limitations remain. First, we desiccated all lenses for SEM investigation, and the desiccation altered the shapes of those lenses. Therefore, the images we obtained with SEM are not representative of the actual lens shape on the human cornea. For example, EYE COFFRET 1-DAY UV seemed to have a slightly scabrous surface on the posterior side, but the scabrous surface most likely resulted from the desiccation process; therefore, because it may have a smoother surface on the cornea than indicated by our image, we should be aware of the possibility of overestimation. Second, only five commercially available Cos-CLs were investigated in this study. Because there are many Cos-CLs widely distributed in the marketplace, it was impractical to assess all lenses available to determine if most Cos-CLs have results similar to the five lenses we investigated. Although some manufacturers, including 1-DAY ACUVUE DEFINE and EYE COFFRET 1-DAY UV (Table 1), readily provide more detailed lens information, others do not provide the same level of detail. Consequently, further investigations are required.

In conclusion, ophthalmologists must be aware that Cos-CLs have a wide variety of lens surfaces and colorants. Some lenses have scabrous surfaces with colorants deposited on the surfaces, and the colorants consist of elements with tissue toxicity. It is not unwarranted for ophthalmologists to have a skeptical attitude about the quality of most Cos-CLs. Additionally, it is preferable for eye care professionals to recommend creditable lenses to wearers after detailed examinations of the lens surface and colorants. Further scientific investigations of Cos-CLs that are distributed internationally are necessary.

Back to Top | Article Outline

ACKNOWLEDGMENTS

The authors thank Mr. Takeshi Kamimura for providing technical support with the investigation of lenses using the Tabletop Microscope TM3030 with Quantax70.

Back to Top | Article Outline

REFERENCES

1. Singh S, Satani D, Patel A, et al.. Colored cosmetic contact lenses: An unsafe trend in the younger generation. Cornea 2012;31:777–779.
2. Sauer A, Bourcier T; French Study Group for Contact Lenses Related Microbial Keratitis. Microbial keratitis as a foreseeable complication of cosmetic contact lenses: A prospective study. Acta Ophthalmol 2011;89:439–442.
3. Spiteri N, Choudhary A, Kaye S. Pigmentation of the cornea secondary to tinted soft contact lens wear. Case Rep Ophthalmol Med 2012;2012:852304.
4. Choi HW, Moon SW, Nam KH, et al.. Late-onset interface inflammation associated with wearing cosmetic contact lenses 18 months after laser in situ keratomileusis. Cornea 2008;27:252–254.
5. Ray M, Lim DK. A rare polymicrobial keratitis involving Chryseobacterium meningosepticum and Delftia acidovorans in a cosmetic contact lens wearer. Eye Contact Lens 2013;39:192–193.
6. Giraldez MJ, Resua CG, Oliveira ME, et al.. Contact lens hydrophobicity and roughness effects on bacterial adhesion. Optom Vis Sci 2010;87:E426–E431.
7. Dart JK, Radford CF, Minassian D, et al.. Risk factors for microbial keratitis with contemporary contact lenses: A case-control study. Ophthalmology 2008;115:1647–1654.
8. Keay L, Edwards K, Naduvilath T, et al.. Factors affecting the morbidity of contact lens-related microbial keratitis: A population study. Invest Ophthalmol Vis Sci 2006;47:4302–4308.
9. Santos L, Rodrigues D, Lira M, et al.. The influence of surface treatment on hydrophobicity, protein adsorption and microbial colonisation of silicone hydrogel contact lenses. Cont Lens Anterior Eye 2007;30:183–188.
10. Steinemann TL, Pinninti U, Szczotka LB, et al.. Ocular complications associated with the use of cosmetic contact lenses from unlicensed vendors. Eye Contact Lens 2003;29:196–200.
11. Steinemann TL, Fletcher M, Bonny AE, et al.. Over-the-counter decorative contact lenses: Cosmetic or Medical Devices? A Case Series. Eye Contact Lens 2005;31:194–200.
12. Begley CG, Waggoner PJ. An analysis of nodular deposits on soft contact lenses. J Am Optom Assoc 1991;62:208–214.
13. Lorenz KO, Kakkassery J, Boree D, et al.. Atomic force microscopy and scanning electron microscopy analysis of daily disposable limbal ring contact lenses. Clin Exp Optom 2014;97:411–417.
14. Latkovic S, Nilsson SE. The effect of high and low Dk/L soft lenses on the glycocalyx layer of the corneal epithelium and on the membrane associated receptors for lectins. CLAO J 1997;23:185–191.
15. Solomon OD, Loff H, Perla B, et al.. Testing hypotheses for risk factors for contact lens-associated infectious keratitis in an animal model. Contact Lens Assoc Ophthalmol J 1994;20:109–113.
16. Ren DH, Petroll WM, Jester JV, et al.. The relationship between contact lens oxygen permeability and binding of Pseudomonas aeruginosa to human corneal epithelial cells after overnight and extended wear. Contact Lens Assoc J 1999;25:80–100.
17. Chan KY, Cho P, Boost M. Microbial adherence to cosmetic contact lenses. Cont Lens Anterior Eye 2014;37:267–272.
18. Yang Z, Li JP, Zhang JX, et al.. Review on research and development of magnesium alloys. Acta Metall Sinica 2008;21:313–328.
19. Eaton JW, Qian MW. Molecular bases of cellular iron toxicity. Free Radic Biol Med 2002;32:833–840.
Keywords:

Cosmetically tinted contact lens; Scanning electron microscopy energy-dispersive x-ray analysis; Colorant; Elemental analysis