Critical Care Medicine:
September 1996 - Volume 24 - Issue 9 - pp 1501-1506
Clinical Investigation

Comparison of infrared ear thermometer derived and equilibrated rectal temperatures in estimating pulmonary artery temperatures

Rotello, Leo C. MD FCCP; Crawford, Lisa RN; Terndrup, Thomas E. MD FACEP

Collapse Box

Abstract

Objectives: To investigate the clinical accuracy of infrared ear thermometer derived and equilibrated rectal temperatures in estimating core body temperature. The clinical bias (i.e., mean difference between body sites), and variability (SD of the differences) of simultaneous temperatures were compared with pulmonary artery temperatures. Clinical repeatability (pooled SD of triplicate reading differences) was also examined for three ear infrared thermometers.

Design: Prospective clinical study.

Setting: A multidisciplinary, adult intensive care unit.

Patients: Twenty patients with an existing pulmonary artery catheter were studied in a multidisciplinary, adult intensive care unit.

Interventions: A single operator using optimum ear infrared technique and masked to ear and rectal temperatures recorded triplicate measurements with each of three infrared ear thermometers, each over a 4-min period with each infrared thermometer, while an assistant recorded temperatures. Infrared and rectal temperatures were compared with a simultaneous pulmonary artery temperature.

Measurements and Main Results: Infrared ear thermometers and rectal thermometers were calibrated daily, and pulmonary artery catheters were calibrated on removal from the patient. Patients were grouped into afebrile and febrile groups, based on initial pulmonary artery temperature. Bias and variability were compared between thermometers using analysis of variance.

Clinical bias, but not variability, was significantly different between three ear infrared thermometers (0.16 +/- 0.46 degrees C, 0.07 +/- 0.38 degrees C, and -0.22 +/- 0.47 degrees C). The repeatability was not different between ear infrared thermometers (range 0.13 degrees C to 0.14 degrees C). Rectal temperature had a significantly greater bias (average 0.3 degrees C), but less variability (average 0.2 degrees C). Bias was increased, and variability decreased for both rectal and infrared ear temperatures when pulmonary artery temperature was increased.

Conclusions: The three infrared ear thermometers studied provided a closer estimate of core body temperature than equilibrated rectal temperature. Clinical bias was greatest in febrile vs. afebrile intensive care unit patients.

(Crit Care Med 1996; 24:1501-1506)

© Williams & Wilkins 1996. All Rights Reserved.

You currently do not have access to this article.

You may need to:

Note: If your society membership provides for full-access to this article, you may need to login on your society’s web site first.

Article Tools

You currently do not have access to this article.

You may need to:

Note: If your society membership provides for full-access to this article, you may need to login on your society’s web site first.